Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K243731
    Manufacturer
    Date Cleared
    2025-07-09

    (217 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K190641, K222792

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hexim Implant - The HEXIM Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations. The HEXIM Implant System is for two stage surgical procedures. It is intended for delayed loading.

    Samwon General Abutments - Couple Abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete.

    Device Description

    The HEXIM Implant with Samwon General Abutments is a Root-form Endosseous Dental Implant and Endosseous Dental Abutment system designed for use in dental implant surgery. These devices have special controls as described by the document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments – Class II Special Controls Guidance for Industry and FDA Staff.

    The HEXIM Implant is a threaded root-form endosseous dental implant with an internal hexagon connection. The HEXIM implants are surgically inserted into the upper and/or lower jawbone to provide a stable foundation for restorations. Geometrically, the implant is screw type. It has external self-tapping threads with chip pocket vertical flutes to facilitate implantation. The HEXIM implant also includes a cavity which extends through the bottom of the implant and connects to 5 channels running perpendicular to the implant body long axis to aid in intra bone fixation. The bottom cavity is not directly interconnected to the internal hex connection nor to the internal threading feature used to fix abutments to the implant.

    The Samwon General Abutments are connected to the implant through the implant's internally threaded hole and an internal hex connection. Samwon General Abutments come in several types: Cover Screw, Healing Abutment, Hex type Couple Abutment, and Abutment Screw. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. Healing abutments are used after osseointegration, once the cover screw is removed, to make a natural soft tissue shape before setting up prosthetics. The Hex-type Couple Abutment is designed to mate with the internal hex cavity of the endosseous implant and is fixed to the implant via an abutment screw. The Hex-type Couple Abutment is a transgingival component, which serves as the support for the artificial tooth or other prosthetic.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the HEXIM Implant and Samwon General Abutments is for a medical device that includes physical implants and abutments, not an AI or software-based medical device. Therefore, the information typically associated with acceptance criteria and a study proving an AI device meets those criteria (such as sample size for test sets, data provenance, expert adjudication, MRMC studies, standalone performance, ground truth establishment, and training set details) is not applicable to this document.

    This 510(k) summary focuses on demonstrating substantial equivalence to predicate devices through comparisons of:

    • Intended Use and Indications for Use
    • Technological Characteristics (material, design, dimensions, sterilization, surface treatment, connection type, principle of operation)
    • Performance Testing (bench testing, biocompatibility, sterilization validation)

    The "study" that proves the device meets acceptance criteria in this context is the non-clinical performance testing and the demonstration of substantial equivalence to legally marketed predicate devices, rather than a clinical trial or AI model validation study.

    Here's a breakdown of the relevant information provided in the document concerning how the device meets its "acceptance criteria" (understood as demonstrating substantial equivalence and safety/effectiveness):


    1. A table of acceptance criteria and the reported device performance

    The document does not present explicit "acceptance criteria" in a tabulated form with specific numerical targets typical for AI model performance (e.g., AUC > X, Sensitivity > Y). Instead, the "acceptance criteria" are implicitly met by demonstrating substantial equivalence to predicate devices and adherence to relevant industry standards and guidance documents. The "reported device performance" is demonstrated through various tests and comparisons:

    Category of "Acceptance Criteria" (Implicit)Reported Device Performance / How it's Met
    Intended Use & Indications for Use EquivalenceSubject device indications are "Same" or "Similar" to predicate and reference devices, with justified differences (e.g., subject device for delayed loading only, which is a subset of predicate's capabilities and does not impact safety/efficacy).
    Technological Characteristics Equivalence (HEXIM Implant)- Material: CP Ti Gr4 ASTM F67 (Similar to predicate, same as reference; justified).
    • Surface Treatment: SLA (Similar to predicate, same as reference; justified differences do not affect substantial equivalence).
    • Sterilization: Gamma irradiation (Same as reference, predicate not specified).
    • Connection Type: Internal Hex (Same as both predicate and reference).
    • Fixture Body Diameter/Length: Differ slightly but fall within predicate/reference ranges or justified as not affecting safety/efficacy.
    • Bottom Cavity: Presence/absence compared and justification provided. |
      | Technological Characteristics Equivalence (Samwon General Abutments) | - Material: Ti-6Al-4V ELI of ASTM F136 (Same as predicate/reference).
    • Surface Treatment: None (Same as predicate; justified difference from reference).
    • Sterilization: End User (Steam) (Same as predicate/reference where specified).
    • Connection Type: Threaded/Internal Hex (Same or similar, justified).
    • Dimensions (Diameter, Length, G/H, Post Height): Differ but fall within predicate/reference ranges or justified as not affecting safety/efficacy; mechanical testing confirms suitability. |
      | Biocompatibility | Tested in accordance with ISO 10993 series for various biological endpoints (cytotoxicity, irritation, sensitization, systemic toxicity, implantation effects, chemical characterization). Results implicitly met acceptance criteria as they support substantial equivalence. |
      | Sterilization Validation | For HEXIM Implants (sterile): ISO 11137 series.
      For Abutments (non-sterile, end-user steam): ANSI/AAMI ST79:2017 & ISO 17665 series. LAL bacterial endotoxin testing also performed. Results implicitly met acceptance criteria. |
      | Shelf-Life | Established via accelerated aging per ASTM F1980-2021 at 5 years for HEXIM Implants. |
      | Mechanical Performance (Bench Testing) | - Fatigue Testing: Performed per ISO 14801:2007 using worst-case geometries. Results met criteria.
    • Implant-to-Abutment Compatibility: Determined through mechanical testing. Results implicitly met criteria.
    • Surface Analysis (SLA): Chemical and visual analysis performed to ensure no blasting particles/chemicals remain. Results implicitly met criteria. |
      | MRI Safety | Non-clinical worst-case MRI review performed using scientific rationale and published literature, addressing magnetically induced displacement force and torque. Justified as safe. |

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as this is a medical device clearance for physical implants and abutments, not an AI/software device. The "test set" here refers to physical samples used in bench testing.

    • Sample Size: Not explicitly stated for each bench test (e.g., number of implants for fatigue testing). The tests were performed on "representative final manufactured samples."
    • Data Provenance: Not applicable. Tests are conducted in a lab environment (e.g., ISO-certified labs for mechanical and biocompatibility testing).
    • Retrospective or Prospective: Not applicable. Bench testing is a controlled, experimental process, not typically classified as retrospective or prospective clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. Ground truth establishment by experts pertains to diagnostic AI models. For physical implants, "ground truth" is established through physical and chemical measurements against engineering specifications and international standards, and results are interpreted by engineers and scientists.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are relevant for human interpretation of data, typically in diagnostic studies for AI. For bench testing, results are objective measurements against defined standards.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. MRMC studies are for evaluating diagnostic performance, particularly with AI assistance. This submission is for physical dental implants and abutments.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. Standalone performance refers to AI algorithm performance. This is a physical device.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this device, the "ground truth" for demonstrating safety and effectiveness includes:

    • Engineering Specifications: Design drawings, material specifications (e.g., ASTM F67, ASTM F136).
    • International Standards: ISO 14801 (fatigue), ISO 10993 series (biocompatibility), ISO 11137 / ANSI/AAMI ST79 / ISO 17665 series (sterilization).
    • FDA Guidance Documents: "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments" and "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment."
    • Pre-market Review Precedents (Predicate Devices): The performance and safety profile of the legally marketed predicate devices (K111889, K190641, K221866, K222792) serve as the established "ground truth" for substantial equivalence.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" for physical medical devices in the context of AI/machine learning.


    9. How the ground truth for the training set was established

    This information is not applicable. There is no "training set" for physical medical devices that requires ground truth establishment in the AI sense. The design and manufacturing processes follow established engineering principles and quality systems (21 CFR Part 820), which ensure the product meets its intended specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222142
    Manufacturer
    Date Cleared
    2023-06-24

    (339 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K190641

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BR SLA Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. BR SLA Type Implant System is for single and two-stage surgical procedures. It is intended for delayed loading.

    Device Description

    The BR SLA Type Implant System is a dental implant system made of CP Ti Gr 4, intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period. The implants may be used to replace one or more missing teeth. The system is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The surface of the implants has been treated with S.L.A.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "BR SLA Type Implant System." It aims to demonstrate substantial equivalence to legally marketed predicate devices. The document does not describe a study evaluating the performance of an AI/ML device in meeting acceptance criteria. Instead, it focuses on the equivalence of a dental implant system.

    Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert qualifications, or ground truth for an AI device. The document is about a conventional medical device (dental implant).

    Here's what the document does provide regarding the non-clinical testing performed for the BR SLA Type Implant System, which serves as evidence for its substantial equivalence:

    The non-clinical testing included:

    • Sterilization validation testing: For sterile devices (fixtures) in accordance with ISO 11137, ISO 11737-1 & ISO 11737-2 for gamma sterilization.
    • Steam sterilization validation: For non-sterile devices (abutments) in accordance with ISO 17665-1 and 17665-2.
    • Surface characteristics test report: Chemical and SEM image analysis performed to verify no residual after SLA treatment on the fixtures.
    • Cytotoxicity test: Performed according to ISO 10993-5:2009.
    • Sensitization test: Performed according to ISO 10993-10:2010.
    • LAL Endotoxin lot release testing: According to USP .
    • Shelf life testing: Performed according to ISO 11607.
    • Fatigue testing: Leveraged from the predicate device.

    The document states that "The result of the above tests has met the criteria of the standard and proved the substantial equivalence with the predicate device." This indicates that the various tests performed (as listed above) served as the acceptance criteria for demonstrating that the device is substantially equivalent to existing predicate devices. The study proving the device meets these criteria is the culmination of these non-clinical tests.

    No information is provided about:

    1. A table of acceptance criteria and reported device performance (in the context of AI/ML).
    2. Sample sizes used for a test set or data provenance (as this is not an AI/ML study).
    3. Number of experts or their qualifications for ground truth (as this is not an AI/ML clinical study).
    4. Adjudication method (not applicable for this type of device submission).
    5. Multi-reader multi-case (MRMC) comparative effectiveness study (not applicable).
    6. Standalone (algorithm only) performance (not applicable).
    7. Type of ground truth used (not applicable to the non-clinical tests described).
    8. Sample size for the training set (not applicable).
    9. How ground truth for the training set was established (not applicable).
    Ask a Question

    Ask a specific question about this device

    K Number
    K222144
    Manufacturer
    Date Cleared
    2023-06-14

    (329 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K190641

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IR SLA Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. IR SLA Type Implant System is for one stage surgical procedures. It is intended for delayed loading. IR SLA Type Implant System is intended only for straight placement with no correction of angulation.

    Device Description

    The IR SLA Type Implant System is a dental implant system made of CP Ti Gr 4, intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period. The implant may be used to replace one or more missing teeth. The system is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The system has been treated with S.L.A.

    AI/ML Overview

    The provided text does NOT describe an AI/ML powered medical device, therefore no information about acceptance criteria, study details, sample sizes, expert involvement, ground truth, or training sets is available as requested.

    The document is a 510(k) premarket notification for a dental implant system (IR SLA Type Implant System), which is a physical medical device and not an AI/ML product. The document details the device's indications for use, description, comparison to predicate devices, and non-clinical testing performed to demonstrate substantial equivalence. These tests primarily cover material properties, sterilization, and biocompatibility, not AI model performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1