K Number
K222792
Manufacturer
Date Cleared
2023-06-08

(265 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SNUCONE Tissue Level Implant System is inticated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations and terminal or internediate Abutment support for fixed bridge work. Snucone implant system is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.

Device Description

SNUCONE Implant System Fixture, also known as an endosseous implant, is surgical component that interfaces with the bone of the jaw or skull to support a dental prosthesis such as a crown, bridge, denture. Snucone's abutment and prosthetic components and tools are compatible with the Snucone's fixture only. The system includes Fixtures and Cover Screw, and Abutment and Component (Solid Abutment, InOcta Closing Screw, InOcta Healing Abutment, InOcta Abutment, InOcta Angled Abutment, InOcta Temporary Abutment, Platform Switching Abutment, and Abutment Screw).

AI/ML Overview

The provided text is a 510(k) premarket notification for a medical device (SNUCONE Tissue Level Implant System). It contains details about the device, its intended use, comparison to predicate devices, and a list of non-clinical tests performed. However, it does not contain information about acceptance criteria or a study that proves the device meets those criteria, especially in the context of an FDA clearance for an AI/ML-driven device.

The document explicitly states: "Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device." It details mechanical performance, sterilization, shelf-life, packaging, MR environment conditions, bacterial endotoxin, and biocompatibility.

THERE IS NO MENTION OF AN AI/ML COMPONENT IN THIS DEVICE.

Therefore, I cannot fulfill your request for information regarding:

  1. A table of acceptance criteria and the reported device performance for an AI/ML component. The document describes acceptance criteria for mechanical, material, and sterilization properties, but not AI/ML performance metrics (e.g., sensitivity, specificity, AUC).
  2. Sample size for test set and data provenance specific to an AI/ML model. The tests mentioned are laboratory bench tests, not data-driven performance studies on patient cohorts for an AI model.
  3. Number of experts and their qualifications for ground truth establishment. This is not relevant to the non-clinical tests described.
  4. Adjudication method for the test set. Not relevant.
  5. MRMC comparative effectiveness study or human reader improvement. Not relevant.
  6. Standalone (algorithm only) performance. Not relevant.
  7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.). Not relevant to the tests performed.
  8. Sample size for the training set. Not relevant, as no AI model is described.
  9. How ground truth for the training set was established. Not relevant, as no AI model is described.

In summary, the provided document is a 510(k) submission for a physical dental implant system and does not involve any AI/ML components or related performance studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

SNUCONE Co., Ltd. % Sanglok Lee Manager Wise Company Inc. #303, 142 Gasan digital 1-ro Geumcheon-gu. Seoul 08507 REPUBLIC OF KOREA

Re: K222792

Trade/Device Name: SNUCONE Tissue Level Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: Class II Product Code: DZE, NHA Dated: May 8, 2023 Received: May 8, 2023

Dear Sanglok Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K222792

Device Name

SNUCONE Tissue Level Implant System

Indications for Use (Describe)

SNUCONE Tissue Level Implant System is inticated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations and terminal or internediate Abutment support for fixed bridge work. Snucone implant system is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

The assigned 510(k) Number: K222792

01. Date Prepared: May 29, 2023

02. Applicant

Company name: SNUCONE Co., LTD. Address: 5, Seongseo-ro 75-gil, Dalseo-qu, Daegu, Korea TEL: 82.53.592.7525 FAX: 82.53.592.7524 Email: snucone@naver.com

03. Submission Correspondent

Sanglok, Lee Wise COMPANY Inc. #303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org

04. Proposed Device Identification

Trade Name: SNUCONE Tissue Level Implant System Common Name: Endosseous Dental Implant Classification Name: Implant, Endosseous, Root-Form Primary Product Code: DZE Secondary Product Code: NHA Panel: Dental Regulation Number: 21 CFR 872.3640 Device Class: Class II

05. Indication for use

SNUCONE Tissue Level Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture and terminal or intermediate Abutment support for fixed bridge work. Snucone implant system is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.

06. Predicate and Reference Devices

Primary Predicate

  • K181137, IT-III active System manufactured by NEOBIOTECH CO., LTD

Reference Devices

  • K141159, K3Pro Konus Dental Implant System. Argon Med. Productions & Vertriebs Gesellschaft mbh & Co. KG
  • K121585, TS Implant System by Osstem Implant Co., Ltd.
  • K181138, IS-III active System manufactured by NEOBIOTECH CO., LTD
  • K200189, Luna Dental Implant System-Healing Abutment, Shinhung MST CO., LTD
  • K120043, CSM Internal-R Implant System, CSM Implant
  • K201700, BEGO Semados® RS/RSX Implant System by BEGO Implant Systems GmbH & Co. KG
  • K190637, Fit & Brilliant Dental Implant System, F&B Technology CO., LTD
  • K210161, Any One Onestage Implant System by MegaGen Implant Co., Ltd.

{4}------------------------------------------------

07. Device Description

SNUCONE Implant System Fixture, also known as an endosseous implant, is surgical component that interfaces with the bone of the jaw or skull to support a dental prosthesis such as a crown, bridge, denture. Snucone's abutment and prosthetic components and tools are compatible with the Snucone's fixture only.

  1. Fixtures and Cover Screw

There is 1 type of fixture, and the dimensions and specification are as following:

ProductEFCover Screw
PictureImage: EF ImplantImage: Cover Screws
DiameterØ3.7, Ø3.8, Ø4.1, Ø4.3,Ø4.8, Ø5.3, Ø5.5, Ø5.8Ø5.5 / Ø6.7
LengthØ3.7: 7,8,9,10,11,12,13,14mmØ3.8: 7,8,9,10,11,12,13,14mmØ4.1: 7,8,9,10,11,12,13,14mmØ4.3: 7,8,9,10,11,12,13,14mmØ4.8: 7,8,9,10,11,12,13,14mmØ5.3: 7,8,9,10,11,12,13,14mmØ5.5: 7,8,9,10,11,12,13,14mmØ5.8: 7,8,9,10,11,12,13,14mm8mm
SurfaceTreatmentAcid etchingAnodizing
ConnectionInternal OctaN/A
MaterialTitanium Gr4 (ASTM F67)Ti 6Al-4V ELI (ASTM F136)
SterilizationGamma irradiationGamma irradiation
Shelf Life5 years5 years

Tolerance of dimension shall be within ± 1% range.

2) Abutment and Component

2-1) Abutment

Dimensions and features of abutment are as following:

Tolerance of dimension shall be within ± 1% range.

AbutmentSolid AbutmentInOcta Closing ScrewInOcta Healing AbutmentInOcta Abutment
PictureImage: Solid AbutmentImage: InOcta Closing ScrewImage: InOcta Healing AbutmentImage: InOcta Abutment
UseThis device is a one-pieceabutment that is securedto the Fixture without othercomponent.This device is connected tothe Fixture and is usedduring healing period.This device is used toconnect with Fixture tohelp gum tissue aroundthe implant site heal faster.This device is a two-pieceabutment that is securedto the Fixture with anabutment screw.
Principle ofThis product is aThis device is used forThis product is healingThis product is a
Operationsuperstructure which isconnects with the fixture. Itis a one-body typeAbutment of screw-retained that does notrequire an Abutmentscrew. It replaces thefunctions of the missingteeth as a dentalAbutment.protecting inner hole andconnecting part withexposed upper part ofstructure during the healingperiod after inserting dentalimplant fixture. Wheninserting the Abutment, Thisproduct is removed.Abutment to formationappropriate gingival shapeduring the soft tissuehealing period combinedwith implant. This productshould be removed whenthe superstructure is setup.superstructure which isconnects with the fixtureusing the enclosedAbutment screw. Itreplaces the functions ofthe missing teeth as adental abutment.
DiameterØ3.5, Ø4.3Ø3.5Ø5.5, Ø6.1, Ø6.3Ø4.8, Ø5.2, Ø6.2
Total Length9.5~13.0mm5.9mm9.0~11.5mm6.0~13.0mm
Gingival Height--4.0~6.5mm0.0~4.0mm
Post Height4.0~7.0mm--4.0~7.0mm
Angle(°)N/AN/AN/AN/A
AnodizingN/AYellowYellowN/A
ProstheticretentionCement-retained--Cement-retained
RestorationSingle-unitMulti-unit--Single-unitMulti-unit
MaterialTi 6Al-4V ELI(ASTM F136)Ti 6Al-4V ELI(ASTM F136)Ti 6Al-4V ELI(ASTM F136)Ti 6Al-4V ELI(ASTM F136)
SterilizationEnd user sterilizedEnd user sterilizedEnd user sterilizedEnd user sterilized
Shelf LifeN/AN/AN/AN/A
AbutmentInOcta AngledAbutmentInOcta Temporary AbutmentPlatform SwitchingAbutment-
PictureImage: InOcta Angled AbutmentImage: InOcta Temporary AbutmentImage: Platform Switching Abutment-
UseThis device is a two-pieceabutment that is securedto the Fixture with anabutment screw.This device is a two-pieceabutment that is temporarilyfixed to the Fixture with anabutment screw. This deviceis intended to be used for amaximum timeframe of 6months.This device is a one-pieceabutment that is securedto the Fixture without othercomponent.-
Principle ofOperationThis product is asuperstructure which isconnects with the fixtureusing the enclosedAbutment screw. Itreplaces the functions ofthe missing teeth as adental abutment.It is dental Abutmentsdesigned to serve as atemporary dental prosthesisduring the healing processuntil a permanent crown ismade.This product is asuperstructure which isconnects with the fixtureusing the enclosedAbutment screw. Itreplaces the functions ofthe missing teeth as adental abutment.
DiameterØ3.8Ø5.2Ø4.5, Ø5.5, Ø6.5-
Total Length9.5mm12.5~14.5mm9.9~16.9mm
Gingival Height-1.0~3.0mm0.0~5.5mm
Post Height7.0mm9.5mm5.5mm-
Angle(°)15° 25°N/AN/A-
AnodizingN/AN/AN/A-
ProstheticretentionCement-retainedCement-retainedCement-retained
RestorationSingle-unitMulti-unitSingle-unitMulti-unitSingle-unitMulti-unit
MaterialTi 6Al-4V ELI(ASTM F136)Ti 6Al-4V ELI(ASTM F136)Ti 6Al-4V ELI(ASTM F136)
SterilizationEnd user sterilizedEnd user sterilizedEnd user sterilized
Shelf LifeN/AN/AN/A-

{5}------------------------------------------------

{6}------------------------------------------------

2-2) Abutment Screw

Dimensions and features of component are as following:

Abutment ScrewInOcta Abutment Screw
PictureImage: Abutment Screw
UseThis product is a screw used to connect both two-pieceAbutment to the Fixture
Principle ofOperationThis product is a screw for connected with Abutmentand fixture.
SizeD : Ø2.5 / L : 4.5~8.0mm
MaterialTi 6Al-4V ELI (ASTM F136)
SterilizationEnd user sterilized

Tolerance of dimension for InOcta Abutment screw shall be within ± 1% range.

Fixture is packaged with Cover Screw, Abutment is packaged with single-packing or components, in some cases.

{7}------------------------------------------------

08. Substantial Equivalence Comparison

  1. Fixture
Subject DevicePrimary PredicateReference Device
CompanySNUCONE Co., LTD.Neobiotech Co., LtdArgon Med Productions&Vertriebs GesellschaftmbH&Co.KG
Device NameEF Fixture for SNUCONE Tissue Level Implant SystemIT-III active SystemK3Pro Konus Dental Implant
510(k) NumberNot Assigned Yet.K181137K141159
DeviceClassificationNameImplant, Endosseous,Root-FormImplant, Endosseous,Root-FormImplant, Endosseous,Root-Form
Product CodeDZEDZEDZE
Regulation Number872.3640872.3640872.3640
Indications for UseSNUCONE Tissue Level Implant System is indicated foruse in partially or fully edentulous mandibles andmaxillae, in support of single or multiple-unit restorationsincluding; cemented retained, or overdenturerestorations and terminal or intermediate Abutmentsupport for fixed bridge work. Snucone implant systemis dedicated for two stage surgical procedures and forimmediate loading when there is good primary stabilityand an appropriate occlusal load. Also, implants withdiameters larger than 5mm are indicated for molarregions.The IT-III active system is indicated for use in partially orfully edentulous mandibles and maxillae, in support ofsingle or multiple-unit restorations including; cementedretained, or overdenture restorations and terminal orintermediate Abutment support for fixed bridge work. IT-III active system is dedicated for two stage surgicalprocedures and for immediate loading when there isgood primary stability and an appropriate occlusalloading.The Konus K3Pro and K3Pro Rapid Implant is designedfor use in edentulous sites in the mandible or maxilla forsupport of a complete denture prosthesis, a terminal orintermediate abutment for fixed bridgework or for partialdentures, or as a single tooth replacement.
MaterialTitanium Gr4 (ASTM F67)Pure Titanium of ASTM F67Pure Titanium Grade 4
DesignImage: Two titanium implantsImage: Two titanium implantsImage: Two titanium implants
ConnectionInternal OctaInternal OctaInternal Hex
Diameters(ø)$Ø$ 3.7, $Ø$ 3.8, $Ø$ 4.1, $Ø$ 4.3, $Ø$ 4.8, $Ø$ 5.3, $Ø$ 5.5, $Ø$ 5.83.5/4.0/4.5/5.0/5.5/6.0/7.03.0-6.0mm
Lengths(mm)$Ø$ 3.7: 7,8,9,10,11,12,13,14mm$Ø$ 3.8: 7,8,9,10,11,12,13,14mm$Ø$ 4.1: 7,8,9,10,11,12,13,14mm$Ø$ 4.3: 7,8,9,10,11,12,13,14mm$Ø$ 4.8: 7,8,9,10,11,12,13,14mm7.0/8.5/10.0/11.5/13.0/15.07.5-17mm
Ø5.3: 7,8,9,10,11,12,13,14mm
Ø5.5: 7,8,9,10,11,12,13,14mm
Ø5.8: 7,8,9,10,11,12,13,14mm
Surface TreatmentAcid etchingSLAAcid etching
SterilizationGamma irradiationGamma irradiationGamma irradiation
Principle ofOperationThis product is a root-type fixture which is inserted in thealveolar bone. It replaces the functions of the missingteeth as a dental implant fixture.This product is a root-type fixture which is inserted in thealveolar bone. It replaces the functions of the missingteeth as a dental implant fixture.Unknown
SimilaritiesSNUCONE Tissue Level Fixture has the same device characteristics with the primary predicate such as Indication for use, material, design, connection, diameter, length,sterilization method, and principle of operation.The surface treatment is applied to the threaded part only of the subject device and primary predicate. The upper portion of EF Fixture is pure titanium without surfacetreatment
DifferencesThe difference between the subject device and the Predicate device are surface treatment. The surface treatment method of the subject fixture is Acid etching while theprimary predicate is SLA. To support this inconsistency, K141159 is added as a reference device to support the difference in surface treatment method.

{8}------------------------------------------------

1-2) Cover Screw

Subject DevicePrimary PredicateReference Device
CompanySNUCONE Co., LTD.Neobiotech Co., LtdOsstem Implant Co., Ltd.
Device NameCover screw for SNUCONE Tissue Level ImplantSystemIT-III active SystemTS Implant System
510(k) NumberNot Assigned Yet.K181137K121585
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136
DesignImage: Yellow and blue cover screwImage: Gray cover screwImage: Green cover screw
Diameters(ø)Yellow: Ø5.5Sky blue: Ø6.75.5/6.94.0/4.5/5.0/6.0/7.0
Height (mm)3.0mmCuff Length: 6.0Height: 0.4, 1.4, 2.0
AnodizingAnodizingNon-AnodizingAnodizing
SterilizationGamma irradiationGamma Sterilization-
Principle ofOperationThis product connects with the Dental Implant Fixture toprotect the inside of the fixture during bone fusion periodafter installation of Dental Implant Fixture, and removesthe cover screw before inserting the Abutment.Cover screw as a set of medical devices is used forprotecting inner hole and connecting part with exposedupper part of structure during the healing period afterinserting dental implant fixture. When inserting theAbutment, Cover screw is removed.Cover screw as a set of medical devices is used forprotecting inner hole and connecting part with exposedupper part of structure during the healing period afterinserting dental implant fixture. When inserting theAbutment, Cover screw is removed.
SimilaritiesThe subject and primary predicate devices have the same material, design, principle of operation, and similar diameter.
DifferencesThere is a slightly different surface treatment between the subject device and primary predicate. To support this inconsistency, K121585 is added as a reference device tosupport the difference in surface treatment method.

{9}------------------------------------------------

2) Abutment, Abutment Screw and Component

2-1) Solid Abutment

Subject DevicePrimary Predicate
CompanySNUCONE Co., LTD.Neobiotech Co., Ltd
Device NameSolid Abutment for SNUCONE Tissue Level Implant SystemIT-III active System
510(k) NumberNot Assigned Yet.K181137
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136
DesignImage: Solid Abutment for SNUCONE Tissue Level Implant SystemImage: IT-III active System
Diameters(Ø)Ø3.5, Ø4.33.5/4.3
Post Heights (mm)4.0~7.0mm4.0/5.5/7.0
AnodizingNon-AnodizingAnodizing / Non-Anodizing
SterilizationEnd user sterilizedEnd user sterilized
Principle of OperationThis product is a superstructure which is connects with the fixture. It is a one-body type Abutment of screw-retained that does not require an Abutment screw. It replaces the functions of the missing teeth as a dental Abutment.This product is a superstructure which is connects with the fixture. It is a one-body type Abutment of screw-retained that does not require an Abutment screw. It replaces the functions of the missing teeth as a dental Abutment.
SimilaritiesThe subject and primary predicate devices have the same material, design, diameter, principle of operation and similar post height.

2-2) InOcta Closing Screw

Subject DeviceReference Device
CompanySNUCONE Co., LTD.Neobiotech Co., Ltd
Device NameInOcta Closing Screw for SNUCONE Tissue Level Implant SystemIS-III active System
510(k) NumberNot Assigned Yet.K181138
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136
DesignImage: Subject Device DesignImage: Reference Device Design
Diameters(Ø)$Ø$ 3.53.45/3.6
Lengths(mm)5.9mm5.85/6.85/7.45/ 6.4/7.4/8.0
AnodizingAnodizingAnodizing/ Non-Anodizing,

{10}------------------------------------------------

SterilizationEnd user sterilizedGamma irradiation
Principle ofOperationThis device is used for protecting inner hole and connecting part with exposed upper part of structure during the healing period after inserting dental implant fixture. When inserting the Abutment, This product is removed.It is used for protecting inner hole and connecting part with exposed upper part of structure during the healing period after inserting dental implant fixture. When inserting the Abutment, Cover screw is removed.
SimilaritiesThe subject device and reference devices have the same material, design, surface treatment, principle of operation and similar diameter and length.

2-3) InOcta Healing Abutment

Subject DevicePrimary PredicateReference Device
CompanySNUCONE Co., LTD.Neobiotech Co., LtdSHINHUNG MST Co., Ltd.
Device NameInOcta Healing Abutment for SNUCONE Tissue LevelImplant SystemIT-III active SystemLuna Dental Implant System - Healing Abutment
510(k) NumberNot Assigned Yet.K181137K200189
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136Ti-Gr4
DesignImage: [yellow abutment]Image: [silver abutment]Image: [purple and silver abutment]
Diameters(ø)Ø5.5, Ø6.1, Ø6.35.8/6.3/7.1/7.24.3, 4.8, 5.8, 6.3
GingivalHeights(mm)4.0~6.5mm1.5/2.5/3.5/4.5/5.5/6.5Lengths: 9.2, 10, 10.2, 11, 11.2, 12, 12.2, 13, 13.2, 14
AnodizingAnodizingN/AAnodizing
Principle ofOperationThis product is healing Abutment to formationappropriate gingival shape during the soft tissue healingperiod combined with implant. This product should beremoved when the superstructure is set up.This product is healing Abutment to formationappropriate gingival shape during the soft tissue healingperiod combined with implant. This product should beremoved when the superstructure is set up.Unknown
SterilizationEnd user sterilizedGamma SterilizationUnknown
SimilaritiesThe subject and primary predicate devices have the same material, design, principle of operation and similar diameter, gingival height.
DifferencesThe diameter of the subject device includes Ø5.5 that is slightly smaller than the primary predicate. To support this inconsistency, K200189 is added as a reference deviceto support the difference in diameter.

2-4) InOcta Abutment

Subject DevicePrimary Predicate
CompanySNUCONE Co., LTD.Neobiotech Co., Ltd
Device NameInOcta Abutment for SNUCONE Tissue Level Implant SystemIT-III active System
510(k) NumberNot Assigned Yet.K181137
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136

{11}------------------------------------------------

DesignImage: OctaImage: Non-OctaImage: OctaImage: Non-OctaImage: SCRP
Diameters(ø)Ø4.8, Ø5.2, Ø6.24.3/4.85/5.5/6.55
GingivalHeights(mm)0.0~4.0mm0/1/2/3/4
AnodizingNon-AnodizingTiN-Coating
Principle ofOperationThis product is a superstructure which is connects with the fixture using the enclosedAbutment screw. It replaces the functions of the missing teeth as a dental abutment.This product is a superstructure which is connects with the fixtureusing the enclosed Abutment screw. It replaces the functions ofthe missing teeth as a dental Abutment.
SterilizationEnd user sterilizedEnd user sterilized
SimilaritiesThe subject and primary predicate devices have the same material, design, principle of operation, gingival height, and similar diameter

2-5) InOcta Angled Abutment

Subject DevicePrimary Predicate
CompanySNUCONE Co., LTD.Neobiotech Co., Ltd
Device NameInOcta Angled Abutment for SNUCONE Tissue Level Implant SystemIT-III active System
510(k) NumberNot Assigned Yet.K181137
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136
DesignImage: OctaImage: OctaImage: Non-Octa
OctaOctaNon-Octa
Diameters(Ø)Ø3.83.7/4.3
Lengths(mm)9.5mm9.5/9.68
Angle(°)15° 25°15° 25°
AnodizingNon-AnodizingNon-Anodizing
Principle of OperationThis product is a superstructure which is connects with the fixture using the enclosed Abutment screw. It replaces the functions of the missing teeth as a dental abutment.This product is a superstructure which is connects with the fixture using the enclosed Abutment screw. It replaces the functions of the missing teeth as a dental Abutment.
SterilizationEnd user sterilizedEnd user sterilized

{12}------------------------------------------------

Similarities The subject and primary predicate devices have material, design, angle, surface treatment, principle of operation, and similar diameter, length.

2-6) InOcta Temporary Abutment

Subject DeviceReference Device
CompanySNUCONE Co., LTD.BEGO Implant Systems GmbH & Co. KG
Device NameInOcta Temporary Abutment for SNUCONE TissueLevel Implant SystemPS TTIA and PS TTIA NH
510(k) NumberNot Assigned Yet.K201700
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI (ASTM F136)
DesignImage: Two implantsImage: Two implants
Diameters(ø)05.23.2, 3.7, 4.1, 5.1mm
Lengths(mm)12.5~14.5mmOverall height: 12mm
AnodizingNon-AnodizingN/A
Principle ofOperationIt is dental Abutments designed to serve as a temporarydental prosthesis during the healing process until apermanent crown is made.Unknown
Maximumtimeframe for use≤ 6 month< 6 month
SterilizationEnd user sterilizedEnd user sterilized
Similarities

2-7) Platform Switching Abutment

Subject DeviceReference Device
CompanySNUCONE Co., LTD.F&B Technology Co., Ltd.
Device NamePlatform Switching Abutment for SNUCONE Tissue Level Implant SystemSolid Abutment for Fit & Brilliant Dental Implant System
510(k) NumberNot Assigned Yet.K190637
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI of ASTM F136

{13}------------------------------------------------

DesignImage: Dental ImplantImage: Dental Implant
Diameters(ø)Ø4.5, Ø5.5, Ø6.5Ø4.0 Ø4.5 Ø5.0 Ø5.5 Ø6.0 Ø6.5
Post Heights(mm)5.5mm4, 5.5, 7mm
SterilizationEnd user sterilizedSteam Sterilization by user
SimilaritiesThe subject and primary predicate devices have the same material, design, diameter, and lengths.

2-8) Abutment Screw

Subject DeviceReference Device
CompanySNUCONE Co., LTD.MegaGen Implant Co., Ltd.
Device NameAbutment ScrewMulti Post Screw For AnyOne Onestage ImplantSystemAbutment Screw For AnyOne Onestage Implant System
510(k) NumberNot Assigned Yet.K210161
MaterialTi-6Al-4V ELI of ASTM F136Ti-6Al-4V ELI (ASTM F136-13)Ti-6Al-4V ELI (ASTM F136-13)
DesignImage: Abutment ScrewImage: Multi Post ScrewImage: Abutment Screw
Diameters(ø)Ø2.52.52.5, 2.6
Lengths(mm)4.5~8.0mm8.14.85, 5.5
Principle ofOperationThis product is a screw for connected with Abutmentand fixture.The Multi Post Screw is used for connecting the MultiPost, EZ Post Abutment, Gold Abutment and CCMAbutment to the FixtureThe Abutment Screw is used for connecting the AngledAbutment to the Fixture, and Healing Cap, TemporaryCylinder, EZ Post Cylinder, Gold Cylinder, CCM Cylinderto the Octa Abutment.
SterilizationEnd user sterilizedNon-sterileNon-sterile
SimilaritiesThe subject and reference device have the same material, design, diameter, principle of operation and similar diameter.

{14}------------------------------------------------

09. Non-Clinical Test

Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:

The subject device was tested to evaluate its substantial equivalence according to the following standards.

· Mechanical Performance

  • Fatigue testing for worst-case implant body EF1.8-3714S with 25° InOcta Angled Abutment was conducted according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" and "ISO 14801:2016 Dentisty - Fatigue test for endosseous dental implants" under the worst case scenario.
  • SEM/EDS analysis for worst-case implant body was conducted to confirm removal of the suface treatment media according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implant Abutment.".

■ Sterilization. Shelf-life and Packaging for Sterile Product

  • End User Sterilization Validation according to ISO 17665-1:2006. ISO 17665-2:2009. ISO 11737-1:2006. and ISO 11737-2:2009
  • Gamma sterilization validation according to ISO 11737-1, and ISO 11737-2
  • Accelerating aging test according to ASTM F1980-16, ASTM F1140-13, ASTM F1140-13, ASTM F1929-15, and ASTM F2096-11

■ MR Environment Condition

Non-clinical worst-case MRI review was performed to evaluate the metallic SNUCONE Tissue Level Implant System devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", including magnetically induced displacement force and torque.

▪ Bacterial Endotoxin

  • ANSI/AAMI ST72:2019 Bacterial endotoxins – Test methods, routine monitoring, and alternatives to batch testing

· Biocompatibility

Subject device has been evaluated for biocompatibility according to ISO 10993

The results of the above tests have met the standards and demonstrated the substantial equivalence with the predicate device

10. Conclusion

SNUCONE Tissue Level Implant System constitutes a substantially equivalent medical device. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, SNUCONE Tissue Level Implant System and its predicates are substantially equivalent.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.