(11 days)
Flat Panel Digital X-ray Detector 17HK701G-W is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in general purpose diagnostic procedures all and not to be used for mammography.
The 17HK701G-W is the solid state x-ray imager, which can generate radiographic images of any part of the body. These devices intercept x-ray photons and the scintillator (CSI:TI) emits visible spectrum photons that illuminate an array of photo-detectors that create an electrical signals. After the electrical signals are generated, it is converted to digital value, and the images are displayed on monitors. The digital value can be communicated to the operator console via wiring connection. The 17HK701G-W consists of the following components: Flat Panel Detector, Control Box, battery Charger, 2 packs of battery, power adapter for charger, Calibration Software, power cord and cables. The 17HK701G-W can be used for general X-ray system excluding fluoroscopic, angiographic, and mammographic applications. The subject device is supported by software. The software is of Moderate level of concern and is identical to the predicate software.
This document pertains to the 510(k) premarket notification for the LG Electronics Inc. 17HK701G-W Flat Panel Digital X-ray Detector. The information provided outlines the device's characteristics and its substantial equivalence to a predicate device, K182348 (14HK701G-W).
Acceptance Criteria and Device Performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed independent study with specific acceptance criteria and a "reported device performance" table in the manner typically seen for new AI/CADe systems with diagnostic claims.
Instead, the acceptance for this device hinges on its technological characteristics being substantially equivalent to the predicate device, specifically regarding safety and effectiveness. The comparison table below highlights key technological characteristics:
Table of Acceptance Criteria and Reported Device Performance
| Characteristic (Acceptance Criteria Based on Predicate Equivalence) | Proposed Device (17HK701G-W) Performance | Predicate Device (14HK701G-W) Performance | Note/Outcome (Equivalence) |
|---|---|---|---|
| Indications for Use | Flat Panel Digital X-ray Detector 17HK701G-W is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in general purpose diagnostic procedures all and not to be used for mammography. | Flat Panel Digital X-ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography. | Same (Meets criteria) |
| Scintillator | CsI | CsI | Same (Meets criteria) |
| Pixel Pitch | 140 um | 140 um | Same (Meets criteria) |
| High Contrast Limiting Resolution (LP/mm) | 3.6 lp/mm | 3.6 lp/mm | Same (Meets criteria) |
| Wireless Communication | 802.11 a/b/g/n/ac compliance Frequency: 2.4 GHz/5GHz Bandwidth: 20MHz/40MHz/80MHz MIMO: 2x2 | Wireless (Standard: 802.11 a/b/g/n/ac) and Wired | Different, but both offer wireless communication. The specific specifications of the predicate's wireless were not fully detailed, but the equivalence implies acceptable performance. |
| DQE | Typ. 72% @ 0.1 lp/mm | Typ. 72% @ 0.1 lp/mm | Same (Meets criteria) |
| MTF | Typ. 89% @ 0.5 lp/mm | Typ. 89% @ 0.5 lp/mm | Same (Meets criteria) |
| Resolution | 3.6 lp | 3.6 lp | Same (Meets criteria) |
| Anatomical Sites | General | General | Same (Meets criteria) |
| Exposure Mode | Manual, Auto (AED) | Manual, Auto (AED) | Same (Meets criteria) |
| Electrical Rating | 24V --- 2.1A | 24V --- 2.1A | Same (Meets criteria) |
| Imaging Area | 17 x 17 inches | 13.7 x 16.8 inches | Different, larger imaging area. Deemed not to affect safety or performance adversely. |
| Pixel Matrix | 3,060 x 3,060 pixels | 2,488 x 3,040 pixels | Different, larger pixel matrix corresponding to larger imaging area. Deemed not to affect safety or performance adversely. |
The Study That Proves the Device Meets the Acceptance Criteria:
The document describes a submission for substantial equivalence to an existing predicate device (K182348, 14HK701G-W). This means the "study" is a collection of tests and comparisons designed to demonstrate that the new device (17HK701G-W) is as safe and effective as the predicate device, despite minor differences.
Here's the breakdown of the information requested:
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- The document states: "Clinical data has been provided according to FDA guidance document 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices'. The data was not necessary to establish substantial equivalence based on the modifications to the device but provided further evidence in addition to the laboratory performance data to show that the device works as intended."
- This implies that while clinical data was submitted, it wasn't the primary basis for the substantial equivalence determination due to the nature of the device (a digital X-ray detector, not an AI/CADe system with a diagnostic claim) and the similarity to the predicate. Therefore, detailed sample sizes, data provenance (country, retrospective/prospective), or ground truth methods for this clinical data are not explicitly provided in this summary. The focus was on non-clinical performance and technological equivalence.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- As detailed above, the clinical data was not the primary driver for substantial equivalence, and therefore, details on expert ground truth establishment for a test set are not provided nor expected in this type of submission.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/not provided as a formal reader study with adjudication for ground truth was not the primary method for demonstrating equivalence for this device.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is a diagnostic imaging component (a flat panel detector), not an AI/CADe system designed to assist human readers or make a diagnostic claim.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This question is generally relevant for AI/CADe systems. For this device (a flat panel detector), its "standalone performance" is assessed by electrical safety, EMC, and imaging performance tests (DQE, MTF, resolution), which were conducted. The device (detector) itself does not perform an "algorithm" making a diagnostic assessment.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the non-clinical performance tests (DQE, MTF, resolution), the "ground truth" is established by physical measurement standards (e.g., IEC 62220-1). For the clinical data, if it was descriptive (e.g., image quality assessment), the ground truth would likely be expert interpretation, but specific details are not provided.
-
The sample size for the training set:
- This device is a hardware component with associated firmware/software for image acquisition and processing. It is not an AI/Machine Learning model that undergoes "training" on a dataset in the conventional sense. The software validation refers to standard software development lifecycle processes, not machine learning model training. Therefore, a "training set sample size" is not applicable in this context.
-
How the ground truth for the training set was established:
- As above, a "training set" in the context of machine learning is not applicable for this device. The software validation follows established engineering principles, ensuring the software performs its intended functions correctly and reliably, rather than learning from labeled data.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
LG Electronics Inc. Jinhwan Jun Chief Research Engineer 22, LG-Ro Jinwi-Myeon Pyeongtaek-Si 17709 Gyeonggi-do, KOREA December 7, 2018
Re: K183286
Trade/Device Name: 17HK701G-W Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: November 20, 2018 Received: November 26, 2018
Dear Jinhwan Jun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug Administration | |
|---|---|
| Indications for Use | |
| Form Approved: OMB No. 0910-0120 | Expiration Date: January 31, 2017 |
| See PRA Statement below. | |
| 510(k) Number (if known) | K183286 |
| Device Name | 17HK701G-W |
| Indications for Use (Describe) | Flat Panel Digital X-ray Detector 17HK701G-W is indicated for digital imaging solutiondesigned for general radiographic system for human anatomy. It is intended to replace filmor screen based radiographic systems in general purpose diagnostic procedures all and notto be used for mammography. |
| Type of Use (Select one or both, as applicable) | |
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human ServicesFood and Drug Administration | |
| Office of Chief Information Officer | |
| Paperwork Reduction Act (PRA) Staff | |
| PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." |
{3}------------------------------------------------
510(k) Summary
[As Required by 21 CFR 807.92]
1. Date Prepared [21 CFR 807.92(a)(a)]
Nov. 12, 2018
2. Submitter's Information [21 CFR 807.92(a)(1)]
| • | Name of Manufacturer: | LG Electronics Inc. |
|---|---|---|
| • | Address: | 77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381Republic of Korea |
| • | Contact Name: | Jinhwan Jun / Chief Research Engineer |
| • | Telephone No.: | +82-31-8066-5641 |
| • | Email Address: | Jinhwan.jun@lge.com |
| • | Registration No.: | 3013501671 |
3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade/Device Name | 17HK701G-W |
|---|---|
| Common Name | Flat Panel Digital X-ray Detector |
| Device Classification Name | Stationary X-ray System |
| Regulation Number | 21 CFR 892.1680 |
| Classification Product Code | MQB |
| Device Class | II |
| 510(k) Review Panel | Radiology |
{4}------------------------------------------------
4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow;
Predicate Device
| ● | 510(k) Number: | K182348 |
|---|---|---|
| ● | Applicant: | LG Electronics Inc. |
| ● | Trade/Device Name: | 14HK701G-W |
| ● | Common Name: | Digital Diagnostic X-ray System |
| ● | Classification Name: | System. X-ray, Stationary |
| ● | Regulation Number: | 21 CFR 892.1680 |
| ● | Classification ProductCode | MQB |
| ● | Device Class: | II |
| ● | 510(k) Review Panel: | Radiology |
The predicate devices have not been subject to a design-related recall
{5}------------------------------------------------
5. Description of the Device [21 CFR 807.92(a)(4)]
The 17HK701G-W is the solid state x-ray imager, which can generate radiographic images of any part of the body. These devices intercept x-ray photons and the scintillator (CSI:TI) emits visible spectrum photons that illuminate an array of photo-detectors that create an electrical signals. After the electrical signals are generated, it is converted to digital value, and the images are displayed on monitors. The digital value can be communicated to the operator console via wiring connection.
The 17HK701G-W consists of the following components: Flat Panel Detector, Control Box, battery Charger, 2 packs of battery, power adapter for charger, Calibration Software, power cord and cables. The 17HK701G-W can be used for general X-ray system excluding fluoroscopic, angiographic, and mammographic applications. The subject device is supported by software. The software is of Moderate level of concern and is identical to the predicate software.
{6}------------------------------------------------
6. Indications for use [21 CFR 807.92(a)(5)]
Flat Panel Digital X-ray Detector 17HK701G-W is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in general purpose diagnostic procedures all and not to be used for mammography.
7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
There are no significant differences in the technological characteristics of these devices compared to the predicate devices which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the 17HK701G-W and the predicate devices:
| Proposed Device | Predicate Device | Note | |
|---|---|---|---|
| K Number | Not known | K182348 | |
| Manufacturer | LG Electronics Inc. | LG Electronics Inc. | Same |
| Trade Name | 17HK701G-W | 14HK701G-W | Same |
| Common Name | Flat Panel Digital X-ray Detector | Flat Panel Digital X-ray Detector | Same |
| Product Code | MQB | MQB | Same |
| RegulationNumber | 21 CFR 892.1680 | 21 CFR 892.1680 | Same |
| 510(k) ReviewPanel | Radiology | Radiology | Same |
| Indications forUse | Flat Panel Digital X-ray Detector17HK701G-W is indicated for digitalimaging solution designed for generalradiographic system for humananatomy. It is intended to replacefilm or screen based radiographicsystems in general purposediagnostic procedures all and not tobe used for mammography. | Flat Panel Digital X-ray Detector isindicated for digital imaging solutiondesigned for general radiographicsystem for human anatomy. It isintended to replace film or screenbased radiographic systems in allgeneral purpose diagnosticprocedures. Not to be used formammography. | Same |
| Detector | |||
| Scintillator | CsI | CsI | Same |
| Imaging Area | 17 x 17 inches | 13.7 x 16.8 inches | Different |
| Pixel Matrix | 3,060 x 3,060 pixels | 2,488 x 3,040 pixels | Different |
| Pixel Pitch | 140 um | 140 um | Same |
| High ContrastLimitingResolution(LP/mm) | 3.6 lp/mm | 3.6 lp/mm | Same |
| Communication | Wired/Wireless | Wired/Wireless | Different |
| DQE | Typ.72% @0.1lp/mm | Typ.72% @0.1lp/mm | Same |
| MTF | Typ.89% @0.5lp/mm | Typ.89% @0.5lp/mm | Same |
| Resolution | 3.6lp | 3.6lp | Same |
| Anatomical Sites | General | General | Same |
| Exposure Mode | Manual, Auto(AED) | Manual, Auto(AED) | Same |
| Wireless | Standard: | - |
[Table 1. Comparison of Proposed Device to Predicate Device]
{7}------------------------------------------------
| Proposed Device | Predicate Device | Note | |
|---|---|---|---|
| 802.11 a/b/g/n/ac complianceFrequency: 2.4 GHz/5GHzBandwidth: 20MHz/40MHz/80MHzMIMO: 2x2 | |||
| Rating | 24V --- 2.1A | 24V --- 2.1A | Same |
| Gap Analysis | There is some difference in the 'Imaging Area' and 'Pixel matrix'. But This functionsare not related to the 'safety' and 'performance' of the device. So Proposed device(17HK701G-W) and Predicate Device (14HK701G-W) are substantially same. |
There are no significant differences between the 17HK701G-W and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use. The proposed device, 17HK701G-W has been tested about electrical safety, EMC and performance, and the software has been validated. In addition, the clinical data has been provided to support the substantial equivalence to the predicate devices.
The technological characteristics of the subject detector 17HK701G-W are similar to the predicate 14HK701G-W, and the only major difference is the larger detector size
8. Integration Specifications / Requirements for 17HK701G-W
| No. | Item | Specification |
|---|---|---|
| 1 | Power frequency | 50Hz ~ 60Hz |
| 2 | KV | 40kVp ~ 150kVp |
| 3 | mA Range | 10 to 500mA |
| 4 | Exposure Time | 0.001 to 4sec |
| 5 | mAs Range | 0.1 to 500mAs |
| 6 | Accuracy | kV < ±(1%+1kV), mA < ±(3%+1mA) |
The proposed device is provided with the recommended generator specification as follows:
9. Non-Clinical Test summary
The 17HK701G-W comply with voluntary standards for electrical safety, electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:
- Electrical Safety, Electromagnetic Compatibility and Performance:
The 17HK701G-W comply with the electrical safety and electromagnetic compatibility requirements established by the standards.
| StandardsNo. | StandardsOrganization | Standard Title | Version | PublicationYear |
|---|---|---|---|---|
| ES60601-1 | AAMI | Medical Electrical Equipment - Part 1: GeneralRequirements For Basic Safety And EssentialPerformance (IEC 60601-1:2005, MOD) | ES60601-1:2005(R)2014 | 2014 |
{8}------------------------------------------------
| 2 andA1:2012 | ||||
|---|---|---|---|---|
| 60601-1-2 | IEC | Medical Electrical Equipment - Part 1-2: GeneralRequirements for Safety – Collateral Standard:Electromagnetic Compatibility - Requirementsand Tests | 60601-1-2Edition 4.02014-02 | 2016 |
| - | FDA | Radio Frequency Wireless Technology in MedicalDevices | August 14 | 2013 |
2) Software Validation
The 17HK701G-W contains MODERATE level of concern software as firmware. The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance:
- . The content of premarket submissions for software contained in medical devices, on May 11, 2005
3) Biocompatibility
- ISO 10993-1 and series, Biological evaluation of medical devices ●
4) Performance Test
Imaging performance test has been conducted according to:
- . IEC 62220-1, Medical Electrical Equipment – Characteristics of Digital X-ray Imaging Devices – Part 1-1: Determination of the Detective Quantum Efficiency – Detectors Used in Radiographic Imaging.
5) Cybersecurity
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, on October 18, 2018
- Postmarket Management of Cybersecurity in Medical Devices, on December 28, 2016
6) Label
- CFR Part 801
- Pediatric Information for X-ray Imaging Device Premarket Notifications, on November 28, 2017
10. Clinical Test Summary
Clinical data has been provided according to FDA quidance document "Guidance for the Submission of 510(k)s for Solid Sate X-ray Imaging Devices". The data was not necessary to establish substantial equivalence based on the modifications to the device but provided further evidence in addition to the laboratory performance data to show that the device works as intended.
{9}------------------------------------------------
11. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between 17HK701G-W and the predicate device, K182348 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.
12. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 17HK701G-W is substantially equivalent in safety and effectiveness to the predicate device as described herein.
N/A