Search Results
Found 3 results
510(k) Data Aggregation
(275 days)
The Zenith Flex System, including the Zetheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump, is indicated in the revascularization of patients with acute secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
The Zenith Flex System consists of the Zenith Flex Catheter, the VC-701 Cliq Aspiration Pump and the Aspiration Tubing Set. The InNeuroCo Zenith Flex Catheter is a variable stiffness catheter shaft reinforced with Stainless-Steel and Nitinol to provide support. The Stainless-Steel is wound as a double coil in the proximal section and the Nitinol is a single coil in the distal section. It has a radiopaque Platinum/Iridium marker band on the distal end. The distal 25 cm of the Zenith Flex Catheter has a hydrophilic coating. The Zenith Flex Catheter is available with an internal diameter of 0.0715 inch. The outer diameter is of 0.0850 inch in the proximal section, and tapers to a nominal of 0.082 inch in the distal section. The Zenith Flex Catheter is available in three working lengths: 115 cm, 125 cm, and 132 cm. Accessories included with the device are a Tuohy-Borst Hemostasis Valve and two peel-away Introducers and a Scout introducer. The Scout may be used to introduce the Zenith Flex Catheter into distal vasculature, thereby helping the device reach the target anatomy. The Zenith Flex Catheter is supplied sterile, nonpyrogenic, and intended for single use only. The off-the-shelf aspiration pump is provided non-sterile.
This document is a 510(k) summary for the Zenith Flex System, a medical device for revascularization in acute ischemic stroke. It primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical (bench and animal) testing rather than clinical study. Thus, it does not describe a study involving an AI component or complex ground truth establishment for diagnostic output.
Here's an analysis based on the provided text, addressing your points where information is available:
1. A table of Acceptance Criteria and Reported Device Performance
The document provides extensive tables summarizing non-clinical testing. Here's a consolidated and simplified version of the information, focusing on the Zenith Flex Catheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump:
Test Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Biocompatibility | ||
Pyrogenicity | Test article extracts must not cause a febrile reaction > 0.5°C in any individual subject. | Met acceptance criteria. (Leveraged from K171672, as materials/manufacturing are equivalent). |
Cytotoxicity (MEM Elution) | Cultures treated with the test article must not have a reactivity grade > 2. | Met acceptance criteria. (Leveraged from K171672 for Catheter, and directly tested for Aspiration Tubing Set). |
Hemolysis | Hemolytic index above negative control article must be |
Ask a specific question about this device
(113 days)
The 091 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasulature.
The InNeuroCo, Inc. 091 Long Sheath is a variable stiffness catheter that has a catheter shaft reinforced with a stainless steel double coil. It has a radiopaque Platinum/lridium marker band on the distal end. The distal 10 cm of the 091 Long Sheath Catheter has a hydrophilic coating. The catheter has a nominal outer diameter of 0.109 inches and a nominal inner diameter of 0.091 inches. It is available in three working lengths: 70 cm, and 90 cm. The 091 Long Sheath has a PTFE-lined lumen. The 091 Long Sheath is inserted at a vascular access point to provide access to the target site and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst Hemostasis Valve and a Vessel Dilator. The 091 Long Sheath is supplied sterile, non-pyrogenic, and intended for single use only.
The provided text is a 510(k) summary for the InNeuroCo 091 Long Sheath peripheral, coronary, and neuro vasculature catheter. It outlines the device's characteristics and its comparison to a predicate device, along with the results of performance testing to demonstrate substantial equivalence.
Based on the provided document, here's a description of the acceptance criteria and the study proving the device meets these criteria:
This document describes the validation of a medical device, the InNeuroCo 091 Long Sheath, for regulatory approval. It is not an AI/ML device, and therefore several of the requested sections (e.g., number of experts, adjudication methods, MRMC study, standalone performance, training set details) are not applicable as they relate to the validation of AI/ML algorithms, not traditional medical devices like catheters.
Acceptance Criteria and Device Performance Study for the InNeuroCo 091 Long Sheath
The InNeuroCo 091 Long Sheath is a percutaneous catheter intended for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The study conducted was a series of non-clinical (bench and animal) tests to demonstrate substantial equivalence to a legally marketed predicate device (Stryker AXS Infinity LS, K152876).
1. Table of Acceptance Criteria and Reported Device Performance
Test | Test Method Summary | Acceptance Criteria | Reported Device Performance (Conclusions) |
---|---|---|---|
Biocompatibility-Material Mediated Pyrogen Testing | Completed per ISO 10993-11 | The test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject. | Test results for the InNeuroCo Super Distal Access (K161262) and the Zenith Catheter (K171672) were leveraged as materials and manufacturing processes are equivalent. Met acceptance criteria. |
Biocompatibility-Cytotoxicity MEM Elution | Completed per ISO 10993-5 | The cultures treated with the test article must not have a reactivity grade greater than 2. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Hemolysis ASTM Method, extract human blood | Completed per ISO 10993-4 | The hemolytic index above the negative control article must be less than 5%. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Hemolysis, ASTM method, direct contact (human blood) | Completed per ISO 10993-4 | The hemolytic index above the negative control article must be less than 5%. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Unactivated Partial Thromboplastin Time | Completed per ISO 10993-4 | There must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Complement Activation | Completed per ISO 10993-4 | There must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Dog Thrombogenicity | Completed per ISO 10993-4 | The test articles must receive a thrombus formation score less than or equal to that of the control. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Maximization Sensitization | Completed per ISO 10993-10 | The test article must elicit a positive response in less than 10% of the test animals. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Intracutaneous Toxicity/Reactivity | Completed per ISO 10993-10 | The test article extracts must not induce a significantly greater biological reaction than the control. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Biocompatibility-Acute Systemic Toxicity Test | Completed per ISO 10993-11 | The test article extracts must not induce a significantly greater biological reaction than the control. | Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria. |
Animal Testing-Angiographic Evaluation | Product evaluated within a porcine model for clinically significant injury. | No units tested can cause vessel injury. | Test article met the acceptance criteria for Animal Testing. |
Bench Testing - Chemical Compatibility | Catheter exposed to chemicals readily available in a clinical setting. | Chemicals have no negative effect on the catheter or accessories. | Leveraged results from InNeuroCo Super Distal Access (K161262). Met acceptance criteria. |
Bench Testing - Visual Inspection | Completed per ISO 10555-1 | Catheter shall appear free from damage, including a rounded tip and smooth transition points. | 091 Long Sheath test samples met the acceptance criteria. |
Bench Testing - Dimensions | Completed per ISO 10555-1 | Test samples should be within existing dimensional specifications. | 091 Long Sheath test samples met the acceptance criteria. Leveraged results for Hemostasis Valve from InNeuroCo Super Distal Access (K161262), which also met the criteria. |
Bench Testing - Hub compatibility | Completed per ISO 594-1 and ISO 594-2 | Hub shall meet existing Luer specifications. | Leveraged results for the 091 Long Sheath and Dilator from InNeuroCo Intermediate Catheter (K152202) and for the Hemostasis Valve from Super Distal Access (K161262). All tested samples met the acceptance criteria. |
Bench Testing - Leak - Air | Completed per ISO 10555-1 | Test samples should be within existing Air – Leak specifications. | 091 Long Sheath test samples met the acceptance criteria. |
Bench Testing - Leak - Liquid | Completed per ISO 10555-1 | Test samples should be within existing Leak - Liquid specifications. | 091 Long Sheath test samples met the acceptance criteria. |
Bench Testing - Particulates | Completed per USP 788 | Test samples should be within existing Particulate specifications. | 091 Long Sheath test samples met the acceptance criteria. |
Bench Testing - Simulated Use | LS underwent simulated use testing by a physician in a benchtop model. | Test samples must meet predetermined user needs. | 091 Long Sheath test samples met the acceptance criteria. |
Bench Testing - Catheter Burst | Completed per ISO 10555-1 | Test sample burst pressures must meet or exceed existing minimum burst pressure specification. | 091 Long Sheath test samples met the acceptance criteria. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Bench Testing - Tensile | Completed per ISO 10555-1 | Test sample ultimate tensile strength must meet or exceed existing tensile strength specifications. | 091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bench Testing - Corrosion | Completed per ISO 10555-1 | Test samples shall exhibit no evidence of corrosion. | Leveraged results from InNeuroCo Intermediate Catheter (K152202). All samples met the acceptance criteria. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Bench Testing - Packaging - Dye Leak | Completed per ASTM F1929-12 | Test sample shall not exhibit any visual leaks or channels. | 091 Long Sheath test samples met the acceptance criteria. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Bench Testing - Packaging - Peel | Completed per ASTM F88-09 | Test sample tensile strength must meet or exceed existing tensile strength specifications. | 091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bench Testing - Kink Resistance | Samples are subjected to different diameters until kink is observed. | Test sample kink resistance must meet or exceed existing Kink Resistance specifications. | 091 Long Sheath test samples met the acceptance criteria. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Bench Testing - Torque | Conditioned samples are torqued to failure. | Test sample torque results must meet or exceed existing torque specifications. | 091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bench Testing - PTFE Liner inspection | LS was challenged to demonstrate liner adherence. | Test sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications. | 091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bench Testing - Hydrophilic Coating Integrity | Conditioned samples were repeatedly exposed to friction to demonstrate that the hydrophilic coating is not affected. | Test sample results must meet or exceed existing Hydrophilic Coating Integrity specifications. | 091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bench Testing - Labeling Legibility | Label is legible after printing. | Test samples shall demonstrate text legibility. | 091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bench Testing - Barcode | Barcode is readable with a standard barcode reader. | Test samples shall demonstrate readily readable barcodes. | 091 Long Sheath test samples met the acceptance criteria. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Sterilization | ISO 11135 and AAMI TIR 28 | Sterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads. | 091 Long Sheath sterilization load met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bacterial Endotoxin | AAMIST72 | Test samples must meet the bacterial endotoxin acceptance criteria. | The 091 Long Sheath samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Shelf Life | ASTM F1980 | Aged test samples must meet or exceed existing specifications. | 091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The exact sample sizes for each specific test are not explicitly enumerated in the provided summary table. The statements typically indicate "Test samples" or "All samples," implying that a sufficient number were tested according to the referenced ISO or ASTM standards.
- Data Provenance: The data is a mix of:
- Direct testing: For many bench tests (e.g., Visual Inspection, Leaks, Particulates, Simulated Use, Catheter Burst, Tensile, Kink Resistance, Torque, PTFE Liner inspection, Hydrophilic Coating Integrity, Labeling Legibility, Barcode, Sterilization, Bacterial Endotoxin, Shelf Life). These tests were conducted on the 091 Long Sheath.
- Leveraged data: For all biocompatibility tests, Chemical Compatibility, Hub compatibility, and Corrosion, data was leveraged from previously approved InNeuroCo devices (Super Distal Access K161262, Zenith Catheter K171672, and Intermediate Catheter K152202). This leveraging is justified by stating that the materials and manufacturing processes are equivalent.
- Animal Testing: Product evaluated within a porcine model for angiographic evaluation.
- Retrospective/Prospective: All described testing appears to be prospective (new tests conducted for this device and leveraged results from other devices), not retrospective analysis of existing clinical data.
- Country of Origin: Not specified in the document, but given the FDA 510(k) submission, it would be expected to follow US or internationally recognized standards. The company address is in Florida, USA.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
- Not Applicable. This is a traditional medical device (catheter) validation, not an AI/ML algorithm. Ground truth for device performance is established through standardized physical, chemical, biological, and animal testing, not human expert interpretation of data.
- For "Simulated Use" testing, it mentions "by a physician," indicating one or more physicians were involved in that specific assessment, but their number and specific qualifications are not detailed beyond "physician."
4. Adjudication Method for the Test Set
- Not Applicable. As this is not an AI/ML study, there is no need for adjudication of conflicting expert opinions on a dataset. Test results are objective measurements against defined acceptance criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No. This type of study is specifically designed for evaluating the impact of AI on human reader performance, which is not relevant for the validation of a physical medical device like a catheter.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not Applicable. There is no AI algorithm being validated in this submission. The "device" is a physical catheter.
7. The Type of Ground Truth Used
- The "ground truth" for this device's performance is established by:
- Standardized Test Methods: Adherence to established international (ISO, ASTM, USP) and industry standards for material properties, mechanical performance, biocompatibility, and sterility.
- Physical Measurements: Objective measurements of dimensions, burst pressure, tensile strength, etc.
- Chemical/Biological Assays: Results from biocompatibility tests (e.g., cytotoxicity, hemolysis, pyrogenicity, sensitization) and bacterial endotoxin tests.
- Functional Demonstrations: Successful performance in simulated use environments and in an animal model (porcine model for angiographic evaluation, showing no vessel injury).
- Predetermined Specifications: Comparison against established product specifications and the performance of the predicate device.
8. The Sample Size for the Training Set
- Not Applicable. This is not an AI/ML device, so there is no "training set."
9. How the Ground Truth for the Training Set was Established
- Not Applicable. As there is no AI/ML training set, this question is not relevant.
Ask a specific question about this device
(135 days)
The Zenith is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
The InNeuroCo Zenith product consists of a catheter, a hemostasis valve, two split sheath introducers, and a Scout introducer. The Zenith catheter is a single lumen, variable stiffness catheter that has an outer polymer shaft and is reinforced with a Stainless Steel cross coil. The Zenith catheter has a radiopaque Platinum/lridium marker band on the distal end. The Zenith product line is available in two internal diameters: 0.065 inches and 0.074 inches. The 0.065 catheter has a tapered design and is 0.079 inches outer diameter in the distal end and a 0.081 inches outer diameter in the proximal end. The 0.074 catheter has a nominal outer diameter of 0.086 inches. There are three working lengths available: 115 cm, 125 cm, and 132 cm in either diameter. The Zenith catheter has a PTFE-lined lumen throughout the catheter shaft. The catheter is flexible and has a hydrophilic coating. The Zenith catheter is inserted through a quide catheter or vascular sheath, provides access to the target site, and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve with an extension luer and two split sheath introducers. There is another introducer, named Scout, which provides support to the Zenith catheter while tracking to target site. The Zenith product line is supplied sterile, nonpyrogenic, and intended for single use only.
The provided text describes the regulatory clearance (510(k)) of a medical device, the "Zenith" percutaneous catheter, and a summary of non-clinical data supporting its substantial equivalence to predicate devices. The document does not describe an AI/ML-enabled device or a study involving human readers or comparative effectiveness with AI. Therefore, several aspects of your request (e.g., sample sizes for test set/training set, number of experts, adjudication method, MRMC study, standalone performance, provenance of data) are not applicable to this document.
However, I can extract the acceptance criteria and study conclusions for the non-clinical performance and biocompatibility tests described.
Here's the information derived from the document, focusing on the available details:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance (Conclusion) |
---|---|---|
Biocompatibility - Material Mediated Pyrogen | The test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject. | Zenith test samples met the acceptance criteria for Material Mediated Pyrogen to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Cytotoxicity MEM Elution | The cultures treated with the test article must not have a reactivity grade greater than 2. | Zenith test samples met the acceptance criteria for Cytotoxicity MEM Elution to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Hemolysis ASTM Method, extract human blood | The hemolytic index above the negative control article must be less than 5%. | Zenith test samples met the acceptance criteria for Hemolysis ASTM Method, extract human blood to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Hemolysis, ASTM method, direct contact (human blood) | The hemolytic index above the negative control article must be less than 5%. | Zenith test samples met the acceptance criteria for Hemolysis, ASTM method, direct contact (human blood) to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Unactivated Partial Thromboplastin Time | There must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control. | Zenith test samples met the acceptance criteria for Unactivated Partial Thromboplastin Time to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Complement Activation | There must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls. | Zenith test samples met the acceptance criteria for Complement Activation to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Dog Thromboresistance | The test articles must receive a thrombus formation score less than or equal to that of the control. | Zenith test samples met the acceptance criteria for Dog Thromboresistance to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Maximization Sensitization | The test article must elicit a positive response in less than 10% of the test animals. | Zenith test samples met the acceptance criteria for Maximization Sensitization to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Intracutaneous Toxicity/Reactivity | The test article extracts must not induce a significantly greater biological reaction than the control. | Zenith test samples met the acceptance criteria for Intracutaneous Toxicity/Reactivity to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Biocompatibility - Acute Systemic Toxicity Test | The test article extracts must not induce a significantly greater biological reaction than the control. | Zenith test samples met the acceptance criteria for Acute Systemic Toxicity Test to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Animal Testing - Angiographic Evaluation | No units tested can cause vessel injury. | Zenith test samples met the acceptance criteria for Animal Testing- Angiographic Evaluation to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Chemical Compatibility | Chemicals have no negative effect on the catheter or accessories. | Zenith test samples met the acceptance criteria for Chemical Compatibility to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Radiographic Detectability | Product shall be visible under fluoro imaging. | Zenith test samples met the acceptance criteria for Radiographic Detectability to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Visual Inspection | Catheter shall appear free from damage, including a rounded tip and smooth transition points. | Zenith test samples met the acceptance criteria for Visual Inspection to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Working Length | Test samples should be within existing working length specification. | Zenith test samples met the acceptance criteria for Working Length to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Hub compatibility | Hub shall meet existing Luer specifications. | Zenith test samples met the acceptance criteria for Hub compatibility to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Outside diameter | Test samples should be within existing outside diameter specification. | Zenith test samples met the acceptance criteria for Outside diameter to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Leak – Air | Test samples should be within existing Air – Leak specifications. | Zenith test samples met the acceptance criteria for Leak – Air to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Leak – Liquid | Test samples should be within existing Leak – Liquid specifications. | Zenith test samples met the acceptance criteria for Leak – Liquid to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Particulates | Test samples should be within existing Particulate specifications. | Zenith test samples met the acceptance criteria for Particulates to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Simulated Use - Bench | Test samples must meet predetermined user needs. | Zenith test samples met the acceptance criteria for Simulated Use - Bench to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Catheter Burst | Test sample burst pressures must meet or exceed existing minimum burst pressure specification. | Zenith test samples met the acceptance criteria for Catheter Burst to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Tensile | Test sample ultimate tensile strength must meet or exceed existing tensile strength specifications. | Zenith test samples met the acceptance criteria for Tensile to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Corrosion | Test samples shall exhibit no evidence of corrosion. | Test results for the predicate device were leveraged for the Zenith as the materials and manufacturing processes are equivalent. (Implies Zenith is expected to meet this, based on predicate data.) |
Packaging - Dye Leak | Test sample shall not exhibit any visual leaks or channels. | Zenith test samples met the acceptance criteria for Packaging - Dye Leak to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Packaging - Peel | Test sample tensile strength must meet or exceed existing tensile strength specifications. | Zenith test samples met the acceptance criteria for Packaging - Peel to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Kink Resistance | Test sample kink resistance must meet or exceed existing Kink Resistance specifications. | Zenith test samples met the acceptance criteria for Kink Resistance to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Torque | Test sample torque results must meet or exceed existing torque specifications. | Zenith test samples met the acceptance criteria for Torque to demonstrate that the Zenith is substantially equivalent to the predicate device. |
PTFE Liner inspection | Test sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications. | Zenith test samples met the acceptance criteria for PTFE Liner inspection to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Hydrophilic Coating Integrity | Test sample results must meet or exceed existing Hydrophilic Coating Integrity specifications. | Zenith test samples met the acceptance criteria for Hydrophilic Coating Integrity to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Labeling Legibility | Test samples shall demonstrate text legibility. | Zenith test samples met the acceptance criteria for labeling legibility to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Barcode | Test samples shall demonstrate readily readable barcodes. | Zenith test samples met the acceptance criteria for barcode testing to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Sterilization | Sterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads. | Zenith sterilization load met the acceptance criteria for sterilization to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Shelf Life | Aged test samples must meet or exceed existing specifications. | Zenith test samples met the acceptance criteria for shelf life to demonstrate that the Zenith is substantially equivalent to the predicate device. |
Regarding the specific requested points that are relevant to this type of device clearance:
2. Sample size used for the test set and the data provenance:
- The document does not explicitly state the sample sizes for each non-clinical test performed (e.g., how many catheters were burst-tested, how many animals in the angiographic evaluation). It refers to "test samples" or "units tested."
- The data provenance is from non-clinical laboratory and animal studies (e.g., ISO standards, ASTM standards, porcine model). It is not human subject data, so concepts like "country of origin" or "retrospective/prospective" as they apply to clinical studies are not applicable.
- For biocompatibility tests, it mentions "extract human blood" for some hemolysis tests.
- For the Angiographic Evaluation, it explicitly states it was a "porcine model."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This is not applicable as the studies described are non-clinical (laboratory and animal testing). The "ground truth" for these tests comes from adherence to established scientific and regulatory standards (e.g., ISO 10993, ASTM, USP) and predefined physical/chemical limits.
- For "Simulated Use - Bench," it mentions "by a physician," implying one or more physicians were involved in the benchtop evaluation, but their number and specific qualifications are not detailed beyond "physician."
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable to non-clinical laboratory or animal studies. Ground truth is determined by test results against defined acceptance criteria.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a percutaneous catheter, not an AI/ML diagnostic or assistive device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm/AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For non-clinical tests, the "ground truth" is determined by meeting predefined technical specifications, industry standards (e.g., ISO, ASTM), and biological safety parameters. For example, a burst pressure test's ground truth is a numerical value compared to a specification. For biocompatibility, it's the absence of adverse biological reactions per standard protocols. For the animal study, it's the absence of "clinically significant injury."
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device that requires training data.
9. How the ground truth for the training set was established:
- Not applicable. This is not an AI/ML device that requires a training set or its associated ground truth establishment.
Ask a specific question about this device
Page 1 of 1