K Number
K171672

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-19

(135 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Zenith is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

Device Description

The InNeuroCo Zenith product consists of a catheter, a hemostasis valve, two split sheath introducers, and a Scout introducer. The Zenith catheter is a single lumen, variable stiffness catheter that has an outer polymer shaft and is reinforced with a Stainless Steel cross coil. The Zenith catheter has a radiopaque Platinum/lridium marker band on the distal end. The Zenith product line is available in two internal diameters: 0.065 inches and 0.074 inches. The 0.065 catheter has a tapered design and is 0.079 inches outer diameter in the distal end and a 0.081 inches outer diameter in the proximal end. The 0.074 catheter has a nominal outer diameter of 0.086 inches. There are three working lengths available: 115 cm, 125 cm, and 132 cm in either diameter. The Zenith catheter has a PTFE-lined lumen throughout the catheter shaft. The catheter is flexible and has a hydrophilic coating. The Zenith catheter is inserted through a quide catheter or vascular sheath, provides access to the target site, and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve with an extension luer and two split sheath introducers. There is another introducer, named Scout, which provides support to the Zenith catheter while tracking to target site. The Zenith product line is supplied sterile, nonpyrogenic, and intended for single use only.

AI/ML Overview

The provided text describes the regulatory clearance (510(k)) of a medical device, the "Zenith" percutaneous catheter, and a summary of non-clinical data supporting its substantial equivalence to predicate devices. The document does not describe an AI/ML-enabled device or a study involving human readers or comparative effectiveness with AI. Therefore, several aspects of your request (e.g., sample sizes for test set/training set, number of experts, adjudication method, MRMC study, standalone performance, provenance of data) are not applicable to this document.

However, I can extract the acceptance criteria and study conclusions for the non-clinical performance and biocompatibility tests described.

Here's the information derived from the document, focusing on the available details:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance (Conclusion)
Biocompatibility - Material Mediated PyrogenThe test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject.Zenith test samples met the acceptance criteria for Material Mediated Pyrogen to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Cytotoxicity MEM ElutionThe cultures treated with the test article must not have a reactivity grade greater than 2.Zenith test samples met the acceptance criteria for Cytotoxicity MEM Elution to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Hemolysis ASTM Method, extract human bloodThe hemolytic index above the negative control article must be less than 5%.Zenith test samples met the acceptance criteria for Hemolysis ASTM Method, extract human blood to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Hemolysis, ASTM method, direct contact (human blood)The hemolytic index above the negative control article must be less than 5%.Zenith test samples met the acceptance criteria for Hemolysis, ASTM method, direct contact (human blood) to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Unactivated Partial Thromboplastin TimeThere must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control.Zenith test samples met the acceptance criteria for Unactivated Partial Thromboplastin Time to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Complement ActivationThere must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls.Zenith test samples met the acceptance criteria for Complement Activation to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Dog ThromboresistanceThe test articles must receive a thrombus formation score less than or equal to that of the control.Zenith test samples met the acceptance criteria for Dog Thromboresistance to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Maximization SensitizationThe test article must elicit a positive response in less than 10% of the test animals.Zenith test samples met the acceptance criteria for Maximization Sensitization to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Intracutaneous Toxicity/ReactivityThe test article extracts must not induce a significantly greater biological reaction than the control.Zenith test samples met the acceptance criteria for Intracutaneous Toxicity/Reactivity to demonstrate that the Zenith is substantially equivalent to the predicate device.
Biocompatibility - Acute Systemic Toxicity TestThe test article extracts must not induce a significantly greater biological reaction than the control.Zenith test samples met the acceptance criteria for Acute Systemic Toxicity Test to demonstrate that the Zenith is substantially equivalent to the predicate device.
Animal Testing - Angiographic EvaluationNo units tested can cause vessel injury.Zenith test samples met the acceptance criteria for Animal Testing- Angiographic Evaluation to demonstrate that the Zenith is substantially equivalent to the predicate device.
Chemical CompatibilityChemicals have no negative effect on the catheter or accessories.Zenith test samples met the acceptance criteria for Chemical Compatibility to demonstrate that the Zenith is substantially equivalent to the predicate device.
Radiographic DetectabilityProduct shall be visible under fluoro imaging.Zenith test samples met the acceptance criteria for Radiographic Detectability to demonstrate that the Zenith is substantially equivalent to the predicate device.
Visual InspectionCatheter shall appear free from damage, including a rounded tip and smooth transition points.Zenith test samples met the acceptance criteria for Visual Inspection to demonstrate that the Zenith is substantially equivalent to the predicate device.
Working LengthTest samples should be within existing working length specification.Zenith test samples met the acceptance criteria for Working Length to demonstrate that the Zenith is substantially equivalent to the predicate device.
Hub compatibilityHub shall meet existing Luer specifications.Zenith test samples met the acceptance criteria for Hub compatibility to demonstrate that the Zenith is substantially equivalent to the predicate device.
Outside diameterTest samples should be within existing outside diameter specification.Zenith test samples met the acceptance criteria for Outside diameter to demonstrate that the Zenith is substantially equivalent to the predicate device.
Leak – AirTest samples should be within existing Air – Leak specifications.Zenith test samples met the acceptance criteria for Leak – Air to demonstrate that the Zenith is substantially equivalent to the predicate device.
Leak – LiquidTest samples should be within existing Leak – Liquid specifications.Zenith test samples met the acceptance criteria for Leak – Liquid to demonstrate that the Zenith is substantially equivalent to the predicate device.
ParticulatesTest samples should be within existing Particulate specifications.Zenith test samples met the acceptance criteria for Particulates to demonstrate that the Zenith is substantially equivalent to the predicate device.
Simulated Use - BenchTest samples must meet predetermined user needs.Zenith test samples met the acceptance criteria for Simulated Use - Bench to demonstrate that the Zenith is substantially equivalent to the predicate device.
Catheter BurstTest sample burst pressures must meet or exceed existing minimum burst pressure specification.Zenith test samples met the acceptance criteria for Catheter Burst to demonstrate that the Zenith is substantially equivalent to the predicate device.
TensileTest sample ultimate tensile strength must meet or exceed existing tensile strength specifications.Zenith test samples met the acceptance criteria for Tensile to demonstrate that the Zenith is substantially equivalent to the predicate device.
CorrosionTest samples shall exhibit no evidence of corrosion.Test results for the predicate device were leveraged for the Zenith as the materials and manufacturing processes are equivalent. (Implies Zenith is expected to meet this, based on predicate data.)
Packaging - Dye LeakTest sample shall not exhibit any visual leaks or channels.Zenith test samples met the acceptance criteria for Packaging - Dye Leak to demonstrate that the Zenith is substantially equivalent to the predicate device.
Packaging - PeelTest sample tensile strength must meet or exceed existing tensile strength specifications.Zenith test samples met the acceptance criteria for Packaging - Peel to demonstrate that the Zenith is substantially equivalent to the predicate device.
Kink ResistanceTest sample kink resistance must meet or exceed existing Kink Resistance specifications.Zenith test samples met the acceptance criteria for Kink Resistance to demonstrate that the Zenith is substantially equivalent to the predicate device.
TorqueTest sample torque results must meet or exceed existing torque specifications.Zenith test samples met the acceptance criteria for Torque to demonstrate that the Zenith is substantially equivalent to the predicate device.
PTFE Liner inspectionTest sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications.Zenith test samples met the acceptance criteria for PTFE Liner inspection to demonstrate that the Zenith is substantially equivalent to the predicate device.
Hydrophilic Coating IntegrityTest sample results must meet or exceed existing Hydrophilic Coating Integrity specifications.Zenith test samples met the acceptance criteria for Hydrophilic Coating Integrity to demonstrate that the Zenith is substantially equivalent to the predicate device.
Labeling LegibilityTest samples shall demonstrate text legibility.Zenith test samples met the acceptance criteria for labeling legibility to demonstrate that the Zenith is substantially equivalent to the predicate device.
BarcodeTest samples shall demonstrate readily readable barcodes.Zenith test samples met the acceptance criteria for barcode testing to demonstrate that the Zenith is substantially equivalent to the predicate device.
SterilizationSterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads.Zenith sterilization load met the acceptance criteria for sterilization to demonstrate that the Zenith is substantially equivalent to the predicate device.
Shelf LifeAged test samples must meet or exceed existing specifications.Zenith test samples met the acceptance criteria for shelf life to demonstrate that the Zenith is substantially equivalent to the predicate device.

Regarding the specific requested points that are relevant to this type of device clearance:

2. Sample size used for the test set and the data provenance:

  • The document does not explicitly state the sample sizes for each non-clinical test performed (e.g., how many catheters were burst-tested, how many animals in the angiographic evaluation). It refers to "test samples" or "units tested."
  • The data provenance is from non-clinical laboratory and animal studies (e.g., ISO standards, ASTM standards, porcine model). It is not human subject data, so concepts like "country of origin" or "retrospective/prospective" as they apply to clinical studies are not applicable.
    • For biocompatibility tests, it mentions "extract human blood" for some hemolysis tests.
    • For the Angiographic Evaluation, it explicitly states it was a "porcine model."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This is not applicable as the studies described are non-clinical (laboratory and animal testing). The "ground truth" for these tests comes from adherence to established scientific and regulatory standards (e.g., ISO 10993, ASTM, USP) and predefined physical/chemical limits.
  • For "Simulated Use - Bench," it mentions "by a physician," implying one or more physicians were involved in the benchtop evaluation, but their number and specific qualifications are not detailed beyond "physician."

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable to non-clinical laboratory or animal studies. Ground truth is determined by test results against defined acceptance criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a percutaneous catheter, not an AI/ML diagnostic or assistive device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an algorithm/AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For non-clinical tests, the "ground truth" is determined by meeting predefined technical specifications, industry standards (e.g., ISO, ASTM), and biological safety parameters. For example, a burst pressure test's ground truth is a numerical value compared to a specification. For biocompatibility, it's the absence of adverse biological reactions per standard protocols. For the animal study, it's the absence of "clinically significant injury."

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device that requires training data.

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI/ML device that requires a training set or its associated ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized depiction of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 19, 2017

InNeuroCo, Inc. Marianne Grunwaldt Director, Quality Assurance & Regulatory Affairs 4635 NW 103rd Avenue Sunrise, Florida 33351

Re: K171672

Trade/Device Name: Zenith (065 and 074) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: September 15, 2017 Received: September 18, 2017

Dear Ms. Grunwaldt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/5 description: The image shows the text "Carlos L. Pena -S" in a simple, sans-serif font. The text is left-aligned and appears to be part of a signature or nameplate. The letters are clear and legible, with a consistent font size and style. The background is plain and does not distract from the text.

Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171672

Device Name Zenith

Indications for Use (Describe)

The Zenith is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Submitter's Name and Address

InNeuroCo, Inc. 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-954-742-5988 Facsimile: 1-954-742-5989

Contact Information

Marianne Grunwaldt Director, Quality Assurance & Requlatory Affairs InNeuroCo, Inc 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-305-495-3883 Facsimile: 1-954-742-5989 E-Mail: Marianne@InNeuroCo.com

Date Prepared

September 14, 2017

Device Trade or Proprietary Name

Zenith Hemostasis Valve Split Sheath Introducer Scout Introducer

Device Common or Classification Name:

Catheter, Percutaneous, 21CFR870.1250, Class II Hemostasis Valve, 21CFR870.4290, Class II Catheter Introducer, 21 CFR 870.1340, Class I

{4}------------------------------------------------

Product Code:

DQY (Catheter) DTL (Hemostasis Valve) DYB (Catheter Introducer)

ldentification of the Leqally Marketed Devices to which Equivalence is Being Claimed

Name of Predicate DeviceName of Manufacturer510(k) Number
Navien Intracranial Support CatheterMedtronicK110055
INTERMEDIATE CATHETERInNeuroCoK152202

Device Description

The InNeuroCo Zenith product consists of a catheter, a hemostasis valve, two split sheath introducers, and a Scout introducer. The Zenith catheter is a single lumen, variable stiffness catheter that has an outer polymer shaft and is reinforced with a Stainless Steel cross coil. The Zenith catheter has a radiopaque Platinum/lridium marker band on the distal end. The Zenith product line is available in two internal diameters: 0.065 inches and 0.074 inches. The 0.065 catheter has a tapered design and is 0.079 inches outer diameter in the distal end and a 0.081 inches outer diameter in the proximal end. The 0.074 catheter has a nominal outer diameter of 0.086 inches. There are three working lengths available: 115 cm, 125 cm, and 132 cm in either diameter. The Zenith catheter has a PTFE-lined lumen throughout the catheter shaft. The catheter is flexible and has a hydrophilic coating. The Zenith catheter is inserted through a quide catheter or vascular sheath, provides access to the target site, and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve with an extension luer and two split sheath introducers. There is another introducer, named Scout, which provides support to the Zenith catheter while tracking to target site. The Zenith product line is supplied sterile, nonpyrogenic, and intended for single use only.

Indications for Use

The Zenith is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

{5}------------------------------------------------

Comparison to Predicate Device

Predicate DeviceMedtronic NavienInNeuroCo, Inc.ZenithReferenceDeviceInNeuroCo IC
510(k) NumberK110055K171672K152202
Classification21CFR870.1250,Class II21CFR870.1250,Class II21CFR870.1250,Class II
Product CodeDQYDQY
Review PanelCardiovascularCardiovascularCardiovascular
Indications ForUseThe NavienIntracranial SupportCatheter isindicated for theintroduction ofinterventionaldevices into theperipheral andneuro vasculature.The Zenith isindicated for theintroduction ofinterventional devicesinto the peripheral andneuro vasculature.The IntermediateCatheter isindicated for theintroduction ofinterventionaldevices into theperipheral andneuro vasculature.
ComponentsSuppliedNavien Catheter,Introducer SheathZenith Catheter, PeelAway Introducers,Hemostasis Valve,Scout IntroducerIntermediateCatheter, PeelAway Introducer,HemostasisValve
Catheter ShaftMaterialPolymeric CatheterPolymeric CatheterPolymericCatheter
Inner LinerPTFEPTFEPTFE
Catheter ShaftReinforcementNitinolStainless SteelStainlessSteel/Nitinol
ReinforcementpatternCoilCross CoilBraid
Predicate DeviceMedtronic NavienInNeuroCo, Inc.ZenithReferenceDeviceInNeuroCo IC
LubriciousCoatingHydrophilicCoatingHydrophilic CoatingHydrophilicCoating
RadiopaqueMarker BandPlatinum/ IridiumPlatinum/ IridiumPlatinum/ Iridium
PackagingPET/PE/TyvekPouch, Hoop,packaging card,SBS cartonTyvek/Nylon Pouch,polyethylene supporttube, packaging card,SBS cartonTyvek/NylonPouch,polyethylenesupport tube,packaging card,SBS carton
WorkingLengths105,115, 125, 130115, 125, 132 cm105, 115, 125 cm
ProximalInternalDiameter (ID)Navien0580.058inchesZenith0.0650.065inches0.062 inchesproximal
Navien0720.072inchesZenith0.0740.074inches
Distal IDNavien0580.058inchesZenith0.0650.065inches0.060 inchesdistal
Navien0720.072inchesZenith0.0740.074inches
Proximal OuterDiameterNavien0580.070inchesZenith0.0650.081inches0.079 inches
Navien0720.084inchesZenith0.0740.086inches
Distal OuterDiameterNavien0580.070inchesZenith0.0650.079inches0.072
Navien0720.084inchesZenith0.0740.086inches
Peel AwayIntroducerAid in catheter tipintroduction intohemostasis valveAid in catheter tipintroduction intohemostasis valveAid in catheter tipintroduction intohemostasis valve
Predicate DeviceMedtronic NavienInNeuroCo, Inc.ZenithReferenceDeviceInNeuroCo IC
HemostasisValveN/AMinimizes blood lossMinimizes bloodloss
Luer TaperedHubYesYesYes
CompatibleGuidewire0.0380.038 inches0.038 inches
ScoutIntroducerYesNoNo
SterilizationEO SAL 10-6EO SAL 10-6EO SAL 10-6
PyrogenicityNonpyrogenicNonpyrogenic
BiocompatibilityMeets ISO 10993-1:2009Meets ISO 10993-1:2009Meets ISO10993-1:2009

{6}------------------------------------------------

{7}------------------------------------------------

Summary of Non-Clinical Data

Biocompatibility tests conducted with the Zenith catheter and its accessories were selected in accordance with ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) for limited duration (< 24 hours), external communicating devices, contacting blood. Studies were conducted pursuant to 21 CFR 58, Good Laboratory Practices. Biocompatibility testing found the Zenith to be biocompatible and non-pyrogenic.

The conclusions drawn from the physical, mechanical, and performance testing of the subject Zenith demonstrates that the product is Substantially Equivalent to the legally marketed predicate device.

Zenith Performance Testing

TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
Biocompatibility-Material MediatedPyrogenTestingcompleted perISO 10993-11The test articleextracts must notcause a febrilereaction greaterZenith test samplesmet the acceptancecriteria for MaterialMediated PyrogenYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
than 0.5°C in anyindividual subject.to demonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.
Biocompatibility-Cytotoxicity MEMElutionTestingcompleted perISO 10993-5The cultures treatedwith the test articlemust not have areactivity gradegreater than 2.Zenith test samplesmet the acceptancecriteria forCytotoxicity MEMElution todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Biocompatibility-Hemolysis ASTMMethod, extracthuman bloodTestingcompleted perISO 10993-4The hemolytic indexabove the negativecontrol article mustbe less than 5%.Zenith test samplesmet the acceptancecriteria forHemolysis ASTMMethod, extracthuman blood todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Biocompatibility-Hemolysis, ASTMmethod, directcontact (humanblood)Testingcompleted perISO 10993-4The hemolytic indexabove the negativecontrol article mustbe less than 5%.Zenith test samplesmet the acceptancecriteria forHemolysis, ASTMmethod, directcontact (humanblood) todemonstrate thatthe Zenith issubstantiallyYes
TestTest Method SummaryAcceptance CriteriaConclusionsPredicate Device Testing (Yes/Unknown)
Biocompatibility- Unactivated Partial Thromboplastin TimeTesting completed per ISO 10993-4There must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control.equivalent to the predicate device. Zenith test samples met the acceptance criteria for Unactivated Partial Thromboplastin Time to demonstrate that the Zenith is substantially equivalent to the predicate device.Yes
Biocompatibility- Complement ActivationTesting completed per ISO 10993-4There must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls.Zenith test samples met the acceptance criteria for Complement Activation to demonstrate that the Zenith is substantially equivalent to the predicate device.Yes
Biocompatibility- Dog ThromboresistanceTesting completed per ISO 10993-4The test articles must receive a thrombus formation score less than or equal to that of the control.Zenith test samples met the acceptance criteria for Dog Thromboresistance to demonstrate that the Zenith is substantially equivalent to the predicate device.Yes
Biocompatibility- Maximization SensitizationTesting completed per ISO 10993-10The test article must elicit a positive response in less than 10% of the test animals.Zenith test samples met the acceptance criteria for Maximization Sensitization toYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
Biocompatibility-IntracutaneousToxicity/ReactivityTestingcompleted perISO 10993-10The test articleextracts must notinduce asignificantly greaterbiological reactionthan the control.Zenith test samplesmet the acceptancecriteria forIntracutaneousToxicity/Reactivityto demonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Biocompatibility-Acute SystemicToxicity TestTestingcompleted perISO 10993-11The test articleextracts must notinduce asignificantly greaterbiological reactionthan the control.Zenith test samplesmet the acceptancecriteria for AcuteSystemic ToxicityTest to demonstratethat the Zenith issubstantiallyequivalent to thepredicate device.Yes
Animal Testing-AngiographicEvaluationProductevaluated withina porcine modelfor clinicallysignificant injury.No units tested cancause vessel injury.Zenith test samplesmet the acceptancecriteria for AnimalTesting-AngiographicEvaluation todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
ChemicalCompatibilityCatheterexposed toChemicals have nonegative effect onZenith test samplesmet the acceptanceUnknown
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
chemicalsreadily availablein a clinicalsetting.the catheter oraccessories.criteria for ChemicalCompatibility todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.
RadiographicDetectabilityTestingcompleted perISO 10555-1Product shall bevisible under fluoroimaging.Zenith test samplesmet the acceptancecriteria forRadiographicDetectability todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes – part of in vivotesting
Visual InspectionTestingcompleted perISO 10555-1Catheter shallappear free fromdamage, including arounded tip andsmooth transitionpoints.Zenith test samplesmet the acceptancecriteria for VisualInspection todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Working LengthTestingcompleted perISO 10555-1Test samplesshould be withinexisting workinglength specification.Zenith test samplesmet the acceptancecriteria for WorkingLength todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
Hub compatibilityTestingcompleted perISO 594-1 andISO 594-2Hub shall meetexisting Luerspecifications.Zenith test samplesmet the acceptancecriteria for Hubcompatibility todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Outside diameterTestingcompleted perISO 10555-1Test samplesshould be withinexisting outsidediameterspecification.Zenith test samplesmet the acceptancecriteria for Outsidediameter todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Leak – AirTestingcompleted perISO 10555-1Test samplesshould be withinexisting Air – Leakspecifications.Zenith test samplesmet the acceptancecriteria for Leak –Air to demonstratethat the Zenith issubstantiallyequivalent to thepredicate device.Yes
Leak – LiquidTestingcompleted perISO 10555-1Test samplesshould be withinexisting Leak –Liquidspecifications.Zenith test samplesmet the acceptancecriteria for Leak –Liquid todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
equivalent to thepredicate device.
ParticulatesTestingcompleted perUSP 788Test samplesshould be withinexisting Particulatespecifications.Zenith test samplesmet the acceptancecriteria forParticulates todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Simulated Use -BenchZenithunderwentsimulated usetesting by aphysician in abenchtop modelTest samples mustmeet predetermineduser needsZenith test samplesmet the acceptancecriteria forSimulated Use -Bench todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
Catheter BurstTestingcompleted perISO 10555-1Test sample burstpressures mustmeet or exceedexisting minimumburst pressurespecification.Zenith test samplesmet the acceptancecriteria for CatheterBurst todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
TensileTestingcompleted perISO 10555-1Test sampleultimate tensilestrength must meetor exceed existingZenith test samplesmet the acceptancecriteria for Tensileto demonstrate thatYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
tensile strengthspecifications.the Zenith issubstantiallyequivalent to thepredicate device.
CorrosionTestingcompleted perISO 10555-1Test samples shallexhibit no evidenceof corrosion.Test results for thepredicate devicewere leveraged forthe Zenith as thematerials andmanufacturingprocesses areequivalent.Yes
Packaging - DyeLeakTestingcompleted perASTM F1929-12Test sample shallnot exhibit anyvisual leaks orchannelsZenith test samplesmet the acceptancecriteria forPackaging - DyeLeak todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Unknown
Packaging - PeelTestingcompleted perASTM F88-09Test sample tensilestrength must meetor exceed existingtensile strengthspecifications.Zenith test samplesmet the acceptancecriteria forPackaging - Peel todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Unknown
Kink ResistanceSamples aresubjected todifferentTest sample kinkresistance mustmeet or exceedexisting KinkZenith test samplesmet the acceptancecriteria for KinkResistance toYes
TestTest MethodSummaryAcceptance CriteriaConclusionsPredicate DeviceTesting(Yes/Unknown)
diameters untilkink is observed.Resistancespecifications.demonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.
TorqueConditionedsamples aretorqued to failureTest sample torqueresults must meetor exceed existingtorquespecifications.Zenith test samplesmet the acceptancecriteria for Torque todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Yes
PTFE LinerinspectionZenith waschallenged todemonstrateliner adherence.Test sample lineradhesion mustmeet or exceedexisting PTFE Linerinspectionspecifications.Zenith test samplesmet the acceptancecriteria for PTFELiner inspection todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Unknown
HydrophilicCoating IntegrityConditionedsamples wererepeatedlyexposed tofriction todemonstrate thatthe hydrophiliccoating is notaffected.Test sample resultsmust meet orexceed existingHydrophilic CoatingIntegrityspecifications.Zenith test samplesmet the acceptancecriteria forHydrophilic CoatingIntegrity todemonstrate thatthe Zenith issubstantiallyequivalent to thepredicate device.Unknown
Labeling LegibilityLabel is legibleafter printing.Test samples shalldemonstrate textlegibility.Zenith test samplesmet the acceptancecriteria for labelinglegibility todemonstrate thatUnknown
TestTest Method SummaryAcceptance CriteriaConclusionsPredicate Device Testing (Yes/Unknown)
BarcodeBarcode is readable with a standard barcode reader.Test samples shall demonstrate readily readable barcodesZenith test samples met the acceptance criteria for barcode testing to demonstrate that the Zenith is substantially equivalent to the predicate device.Unknown
SterilizationISO 11135 and AAMI TIR 28Sterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads.Zenith sterilization load met the acceptance criteria for sterilization to demonstrate that the Zenith is substantially equivalent to the predicate device.Yes
Shelf LifeASTM F1980Aged test samples must meet or exceed existing specificationsZenith test samples met the acceptance criteria for shelf life to demonstrate that the Zenith is substantially equivalent to the predicate device.Yes

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).