Search Results
Found 2 results
510(k) Data Aggregation
(295 days)
The iT20 telemetry transmitter is intended to monitor physiological parameters including: ECG, oxygen saturation of arterial blood (SpO2) and pulse rate (PR) for adults and pediatric patients. The iT20 requires the EDAN MFM-CMS (Central Monitoring Station) to provide full functionality of the device.
The iT20 telemetry transmitter is intended to be used in clinical divisions of hospital environments, including CCU and general wards (as Cardiology Dept.).
iT20 collects physiological parameters by ECG cables and SPO2 sensors, then achieves data analyzing and processing. After that, data will be sent to MFM-CMS via Wi-Fi. The parameters supported are ECG, SPO2 and PR.
The ECG monitor samples small voltages of about 1 mV that appear on the skin as a result of cardiac activity. Three or five electrodes arranged in standard configurations called leads, are placed on the skin to sense these voltages. At least two electrodes are required for an ECG leads; The third electrode is used as a reference to reduce electrical interference. Each lead presents a heart, producing ECG waveform whose P waves, QRS complex, and T waves vary in amplitude and polarity. The signals from the different leads provide the cardiologist with a complete representation of the electrical activity of the heart, including the HR, which is interpreted as the R-to-R Interval. The timing and wave shape of ECG provides information on whether the patient's HR is characterized by arrhythmia or other altered functions requiring treatment. The ECG is also used to monitor the effects of infusing antiarrhythmia or cardiotonic agents.
SpO2 is based on the absorption of pulse blood oxygen to red and infrared light by means of finger sensor and SpO2 measuring unit. The light-electronic transducer in finger sensor converts the pulse red and infrared light modulated by pulse blood oxygen into electrical signal, the signal is processed by hardware and software of the unit. The PLETH curve and numeral value of SpO2 will be obtained.
The provided text describes the Edan Instruments, Inc. Telemetry Transmitter, model iT20, and its substantial equivalence to predicate devices, but it contains limited information regarding a formal acceptance criteria study of the device's performance measures directly. Instead, it focuses on comparing the proposed device's specifications to those of predicate devices and adherence to various safety and performance standards.
The document indicates that "Bench testing was conducted on the iT20 Patient Monitor device, consisting of all the modules and accessories in the system. The system complies with the IEC 60601-1-8: 2006, IEC 60601-2-27: 2011, IEC 60601-2-49: 2011, ISO 80601-2-61: 2011, ANSI/AAMI EC57: 2012 standards for performance effectiveness." This implies that the acceptance criteria are met by complying with these standards.
Here's an attempt to extract the requested information based on the provided text, while noting significant gaps where the information is not present:
Acceptance Criteria and Device Performance Study for Edan Instruments, Inc. Telemetry Transmitter, model iT20
The evaluation of the Edan Instruments, Inc. Telemetry Transmitter, model iT20, primarily relies on demonstrating substantial equivalence to predicate devices and compliance with relevant international standards for safety and performance. No specific stand-alone study "proving" the device meets acceptance criteria as a new AI/algorithmic device with defined accuracy metrics is detailed, rather its performance is compared to established standards and predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the operational parameters and performance accuracy specified in the comparison table with the predicate device (TMS-6016) and compliance with the listed standards.
| Performance Metric (Acceptance Criteria per Standards/Predicate Comparison) | Reported Device Performance (iT20) |
|---|---|
| ECG Function: | |
| HR Calculation Range (Adult) | 15 bpm to 300 bpm (Similar to predicate) |
| HR Calculation Range (Pediatric) | 15 bpm to 350 bpm (Improved range compared to predicate's 15-300 bpm) |
| HR Accuracy | ±1% or 1 bpm, whichever is greater (Same as predicate) |
| HR Resolution | 1 bpm (Same as predicate) |
| ECG Sensitivity | ≥300 µVPP (Improved sensitivity compared to predicate's ≥200 µVPP) |
| ST Numeric Range | -2.0 mV to +2.0 mV (Same as predicate) |
| ECG Bandwidth (Diagnosis) | 0.05Hz to 150Hz (Improved range compared to predicate's 0.1 Hz to 40 Hz) |
| ECG Bandwidth (Monitor) | 0.5Hz to 40Hz (Same as predicate) |
| ECG Bandwidth (Surgery) | 1Hz to 20Hz (Same as predicate) |
| CMRR (Diagnosis) | >95dB (the Notch filter is off) (Similar to predicate's 105dB) |
| CMRR (Monitor/Surgery) | >105dB (the Notch filter is on) (Similar to predicate's 105dB) |
| Pace Pulse Indicator (Amplitude, Width, Ascending Time) | Meets IEC 60601-2-27: 2011, Sect. 201.12.1.101.12 requirements for specific ranges (Improved compared to stricter predicate's minimal amplitude) |
| SpO2 Function: | |
| Measurement Range | 0-100% (Same as predicate) |
| Accuracy (Adult/Pediatric) | ±2 % (70% to 100%), Undefined (0% to 69%) (Similar to predicate, with clarification for 0-69%) |
| Resolution | 1 % (Same as predicate) |
| Pulse Rate Measuring Range | 25 bpm to 300 bpm (Similar to predicate's 18-300 bpm) |
| Pulse Rate Accuracy | ±2bpm when MR: 25 bpm to 300bpm (Improved compared to predicate's varying ±3, ±6, ±9 bpm) |
| Pulse Rate Resolution | 1 bpm (Same as predicate) |
| Environmental Specifications (Operating): | |
| Temperature Range | 0°C to +40°C (Same as predicate) |
| Humidity Range | 15% to 95% (non-condensing) (Same as predicate) |
| Altitude Range | 86kPa to 106kPa (Similar to predicate's 70.0 to 106.0kPa) |
| Compliance Standards: | |
| Electrical Safety | Complies with IEC 60601-1:2005/A1: 2012 |
| EMC | Complies with IEC 60601-1-2:2007 |
| Alarms | Complies with IEC 60601-1-8: 2006 |
| ECG Performance | Complies with IEC 60601-2-27: 2011, ANSI/AAMI EC57: 2012 |
| Multi-parameter Patient Monitoring Equipment | Complies with IEC 60601-2-49: 2011 |
| Pulse Oximeter Equipment | Complies with ISO 80601-2-61: 2011 |
| Biocompatibility | Complies with ISO 10993-1, ISO 10993-5, ISO 10993-5-10 |
| Software | Verified and Validated per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (Concern Level: "major") |
2. Sample size used for the test set and the data provenance
The document does not specify a distinct "test set" in terms of patient data or case numbers. The performance data is stated to be derived from "Bench testing" and "non-clinical data" used for substantial equivalence determination and compliance with standards. There is no mention of geographical origin or whether data was retrospective or prospective in the context of device performance testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided. The compliance with standards and comparison to predicate device specifications serves as the basis for performance claims, rather than a specific expert-adjudicated ground truth for a test set.
4. Adjudication method for the test set
This information is not provided, as no expert-adjudicated test set is detailed.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned. The device is a telemetry transmitter for physiological parameters, not an AI-assisted diagnostic tool for interpretation by human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The document implies that the device's functional performance (e.g., accuracy of HR and SpO2 measurement) was assessed in a standalone manner during bench testing and compliance evaluations against technical standards. The context is measuring physiological parameters rather than an algorithm producing diagnostic outputs.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the performance metrics listed (e.g., HR accuracy, SpO2 accuracy), the ground truth is implicitly defined by the measurement standards and methodologies used in bench testing to assess compliance with the specified accuracy ranges (e.g., using calibrated simulators for vital signs).
8. The sample size for the training set
This information is not applicable and not provided. The device described is a hardware telemetry transmitter with embedded software for physiological parameter monitoring, not a machine learning or AI model that typically requires a distinct training set.
9. How the ground truth for the training set was established
This information is not applicable and not provided, as the device is not described as involving machine learning or AI that would require a ground truth for a training set.
Ask a specific question about this device
(234 days)
iM20 Patient Monitor (hereinafter called monitor) is intended to be used for monitoring, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.
The monitored physiological parameters include: ECG. respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2).
The arrhythmia detection and ST Segment analysis are intended for adult only.
The monitor is additionally intended for use during patient transport inside of the hospital environment.
The monitor is not intended for airplane, helicopter transport, home use and MRI environments.
The iM20 Patient Monitor System (hereinafter called iM20) can perform long-time continuous monitoring of multiple physiological parameters, including ECG, respiration (RESP), non-invasive blood pressure (NIBP), oxygen saturation of the blood (SpO2), pulse rate (PR), temperature (TEMP), invasive pressure (IBP), Carbon Dioxide (CO2). The system capabilities include storing, displaying, analyzing and controlling such data afterwards. When necessary, alarms will be produced so that doctors and nurses can manage patient care appropriately. The system is intended to be used during patient transport inside and outside of the hospital environment.
iM20 can be utilized in two ways, as an independent monitor and as a module of V series Patient Monitor (including models elite V5, elite V6 and elite V8). When used as an independent monitor, it can simultaneously monitor, store, review several parameters data. And transfer patient data to V series patient monitor only under the transport mode. As a highly portable monitor, its compact design makes it particularly appropriate inside hospital and vehicle ambulance transport environments.
When as a multi-measurement module, when the iM20 is directly connected to a V series patient monitor, it can provide the measurements, trends, and patient information. When connected, the V series Patient Monitor controls the connected iM20, including all alarm functionality. So no alarms are available on iM20 in such application and iM20 takes power from the V series Patient Monitor.
This document focuses on the Edan Instruments Inc. Patient Monitor, model iM20 (K152552), and its equivalency to predicate devices. It lists performance data based on non-clinical and clinical studies to demonstrate substantial equivalence.
Here's the breakdown of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document provides a detailed comparison table between the proposed device (iM20) and a predicate device (Philips Medizin.Systeme Boeblingen GmbH IntelliVue MP2 cleared under K102562), focusing on various physiological parameters and technical specifications. A second comparison table is provided for the reference device Edan Instruments, Inc. iM70 (K131971), specifically for PR from NIBP and CO2 function.
Since the provided text does not present a separate "acceptance criteria" table and "reported device performance" table, but rather a comparison with performance metrics of a predicate device, the following table synthesizes the information from the "Predicate Device Comparison" section to represent acceptance criteria (from the predicate) and the reported device performance (for the iM20). Differences are explicitly noted as "Different." Where the proposed device's performance aligns with the predicate, it is marked "Same."
Acceptance Criteria (Based on Predicate Device) and Reported Device Performance (iM20)
| Feature / Parameter | Acceptance Criteria (Predicate: Philips MP2) | Reported Device Performance (iM20) | Match? |
|---|---|---|---|
| ECG Function - HR Calculation | |||
| HR Range (Adult/Pedi/Neo) | Adult/pedi: 15 to 300 bpm, Neo: 15 to 350 bpm | ADU: 15 bpm to 300 bpm, PED/NEO: 15 bpm to 350 bpm | Same |
| HR Accuracy | ±1% of range | ± 1% or 1 bpm, whichever is greater | Different |
| HR Resolution | 1 bpm | 1 bpm | Same |
| Sensitivity | ≥ 200 µVpeak | ≥ 300 µVPP | Different |
| PVC Rate Range | 0 to 300 bpm | ADU: 0 to 300 PVCs/ min, PED/NEO: 0 to 350 PVCs/ min | Different |
| PVC Resolution | 1 bpm | 1 PVCs/min | Different |
| ST Numeric Range | -20 to +20 mm | -2.0 mV to +2.0 mV | Different |
| ST Accuracy | ± 0.5mm or 15%, whichever is greater | -0.8 mV to +0.8 mV: ± 0.02 mV or 10%, whichever one is greater. Beyond this range: not specified. | Different |
| ST Resolution | 0.1mm | 0.01 mV | Different |
| Range of Sinus and SV Rhythms (Brady) | Adult: 15 to 60 bpm, Pedi: 15 to 80 bpm, Neo: 15 to 90 bpm | Adult: RR interval for 5 consecutive QRS complex ≥ 1.5 s. Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≥ 1 s. | Different |
| Range of Sinus and SV Rhythms (Normal) | Adult: 60 to 100 bpm, Pedi: 100 to 160 bpm, Neo: 90 to 180 bpm | Adult: 0.5s < RR interval for 5 consecutive QRS complex < 1.5 s. Pedi/Neo: 0.375s < RR interval for 5 consecutive QRS complex < 1 s. | Different |
| Range of Sinus and SV Rhythms (Tachy) | Adult: >100 bpm, Pedi: >160 bpm, Neo: >180 bpm | Adult: RR interval for 5 consecutive QRS complex ≤ 0.5 s. Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≤ 0.375 s. | Different |
| Bandwidth | Diagnosis: 0.05 to 150 Hz, Monitor: Adult: 0.5 to 40 Hz, Neo/pedi: 0.5 to 55 Hz, Filter Mode: 0.5 to 20 Hz | Diagnosis: 0.05Hz to 150Hz, Monitor: 0.5Hz to 40Hz, Surgery: 1Hz to 20Hz | Different |
| CMRR | Diagnostic mode: >86 dB, Filter mode: >106 dB | Diagnosis: >95dB (Notch filter off), Monitor: >105dB (Notch filter on), Surgery: >105dB (Notch filter on) | Different |
| Differential Input Impedance | >2 MΩ RA-LL leads (Resp), >5 MΩ at all other leads | >5MΩ | Different |
| Input Signal Range | ±5mV (peak-to-peak value) | ±10mVPP | Different |
| Electrode Offset Potential Tolerance | ±500mV | ±500mV | Same |
| Auxiliary Current (detect) | Active electrode: <100nA, Reference electrode: <900nA | Active electrode: <100nA, Reference electrode: <900nA | Same |
| Pace Pulse Indicator | ±2 mV to ±700 mV and widths from 0.1 ms to 2.0 ms | Amplitude: ±2 mV to ±700 mV, Width: 0.1 ms to 2.0 ms, Ascending time: 10 us to 100 us | Different |
| Pace Pulse Rejection | ±2 mV to ±700 mV and widths from 0.1 ms to 2.0 ms | Amplitude: ±2 mV to ±700 mV, Width: 0.1 ms to 2.0 ms, Ascending time: 10 us to 100 us | Different |
| Max Start-up alarm time for Ventricular Tachycardia (1mV 206bpm) | Gain 1.0 Range 6.1 to 6.9 seconds, Average 6.5 seconds | Gain 1.0: 10 s | Different |
| Max Start-up alarm time for Ventricular Tachycardia (2mV 195bpm) | Gain 1.0, Range 5.7 to 6.5 seconds, Average 6.1 seconds | Gain 1.0: 10 s | Different |
| Tall T-wave Rejection | Exceeds ANSI/AAMI EC 13 Sect. 3.1.2.1(c) 1.2 mV T-Wave amplitude | Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17 minimum recommended 1.2 mV T-Wave amplitude | Same |
| HR Averaging Method | Three different methods used, typically 12 most recent RR intervals | Method 1: excluding min/max values from 12 most recent RR intervals and averaging the residual 10. Method 2: 4 most recent RR intervals if 3 consecutive > 1200ms. | Different |
| Accuracy of HR Meter and Response to Irregular Rhythm | Ventricular bigeminy: 80 bpm, Slow alternating ventricular bigeminy: 60 bpm, Rapid alternating ventricular bigeminy: 120 bpm, Bidirectional systoles: 90 bpm | Ventricular bigeminy: 80 bpm±1 bpm, Slow alternating ventricular bigeminy: 60 bpm±1 bpm, Rapid alternating ventricular bigeminy: 120 bpm±1 bpm, Bidirectional systoles: 91 bpm±1 bpm | Different |
| Response time of HR Meter to Change in HR (80 to 120 bpm) | Range: [6.4 to 7.2 s], Average: 6.8 s | Within 11 s | Different |
| Response time of HR Meter to Change in HR (80 to 40 bpm) | Range: [5.6 to 6.4 s], Average: 6.0 s | Within 11 s | Different |
| RESP Function | |||
| Respiration excitation waveform | Sinusoidal signal, 260 µA, 39 kHz | Sinusoid, 62.8 kHz (±10%), <500 µA | Different |
| RR Measuring Range | Adult/pedi: 0 to 120 rpm, Neo: 0 to 170 rpm | Adult: 0 rpm to120rpm, Pediatric/neonate: 0 rpm to 150rpm | Different |
| Accuracy | at 0 to 120 rpm ±1 rpm, at 120 to 170 rpm ±2 rpm | Adult: 6 to 120 rpm: ±2 rpm, 0 to 5 rpm: not specified. Neo/Ped: 6 to 150 rpm: ±2 rpm, 0 to 5 rpm: not specified. | Different |
| Resolution | 1 rpm | 1 rpm | Same |
| Waveform bandwidth | 0.3 to 2.5 Hz (-6 dB) | 0.2Hz to 2.5Hz (-3dB) | Different |
| Apnea Alarm Time Setup | 10 to 40 seconds 5 second steps | 10s, 15s, 20s, 25s, 30s, 35s, 40s | Different |
| NIBP Function | |||
| Principle of Operation | oscillation | oscillation (EDAN, SunTech NIBP) | Same |
| Measurement Range (Adult Systolic) | 30 to 270 mmHg | EDAN NIBP: 40 mmHg ~ 270 mmHg, SunTech NIBP: 60 mmHg ~ 250 mmHg | Different |
| Measurement Range (Adult Diastolic) | 10 to 245 mmHg | EDAN NIBP: 10 mmHg ~ 215 mmHg, SunTech NIBP: 30 mmHg ~ 190 mmHg | Different |
| Measurement Range (Adult Mean) | 20 to 255 mmHg | EDAN NIBP: 20 mmHg ~ 235 mmHg, SunTech NIBP: 40 mmHg ~ 210 mmHg | Different |
| Measurement Range (Pediatric Systolic) | 30 to 180 mmHg | EDAN NIBP: 40 mmHg ~ 230 mmHg, SunTech NIBP: 40 mmHg ~ 230 mmHg | Different |
| Measurement Range (Pediatric Diastolic) | 10 to 150 mmHg | EDAN NIBP: 10 mmHg ~ 180 mmHg, SunTech NIBP: 20 mmHg ~ 160 mmHg | Different |
| Measurement Range (Pediatric Mean) | 20 to 160 mmHg | EDAN NIBP: 20 mmHg ~ 195 mmHg, SunTech NIBP: 30 mmHg ~ 175 mmHg | Different |
| Measurement Range (Neonate Systolic) | 30 to 130 mmHg | EDAN NIBP: 40 mmHg ~135 mmHg, SunTech NIBP: 40 mmHg ~130 mmHg | Different |
| Measurement Range (Neonate Diastolic) | 10 to 100 mmHg | EDAN NIBP: 10 mmHg ~ 100 mmHg, SunTech NIBP: 20 mmHg ~ 90 mmHg | Same/Different |
| Measurement Range (Neonate Mean) | 20 to 120 mmHg | EDAN NIBP: 20 mmHg ~ 110 mmHg, SunTech NIBP: 30 mmHg ~ 100 mmHg | Different |
| Accuracy | Max. Std. Deviation: 8 mmHg, Max. Mean Error: ± 5 mmHg | Max standard deviation: 8 mmHg (EDAN, SunTech); Max mean error: ± 5 mmHg | Same |
| Pressure Resolution | 1mmHg | 1mmHg (EDAN, SunTech) | Same |
| Maximum measuring period | 180 seconds (adult/pedi), 90 seconds (neonates) | Adult/Pediatric 120s, Neonate 90s | Different |
| Overpressure protection (Adult) | >300 mmHg (40 kPa) >2 sec | 297 ± 3mmHg (EDAN), <300mmHg (SunTech) | Different |
| Overpressure protection (Pediatric) | >300 mmHg (40 kPa) >2 sec | 240 ± 3mmHg (EDAN), <300mmHg (SunTech) | Different |
| Overpressure protection (Neonate) | >150 mmHg (20 kPa) >2 sec | 147 ± 3mmHg (EDAN), <150mmHg (SunTech) | Different |
| Measuring interval in AUTO Mode | 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60 or 120 minutes | 1/2/3/4/5/10/15/30/60/90/120/240/480 min (EDAN), 1/2/3/4/5/10/15/30/60/90/120/240/480 min (SunTech) | Different |
| Continuous | 5 minutes | 5min, interval is 5s | Same |
| SpO2 Function (EDAN) | |||
| Measurement Range | 0 to 100 % | 0-100% | Same |
| Accuracy | Philips Reusable Sensors: M1191A, etc. = 2 % (70 % to 100 %); M1193A, etc. = 3 % (70 % to 100 %) | Adult/Pediatric: ± 2 % (70% to 100% SpO2), Undefined (0 to 69% SpO2). Neonate: ± 3 % (70% to 100% SpO2), Undefined (0 to 69% SpO2). | Different |
| Resolution | 1 % | 1 % | Same |
| Pulse Rate Measuring Range | 30 to 300 bpm | 25 bpm to 300 bpm | Different |
| Pulse Rate Accuracy | ±2 % or 1 bpm, whichever is greater | ±2bpm | Different |
| Pulse Rate Resolution | 1 bpm | 1 bpm | Same |
| SpO2 Function (Nellcor) | |||
| Measurement Range | 1% to 100% | 1% to 100% | Same |
| Accuracy | Philips Reusable Sensors: M1193A, etc. = 3 % (70 % to 100 %) (Adult); = 4 % (70 % to 100 %) (Neonate) | DS-100A, OXI-A/N (Adult): ± 3% (70% to 100% SpO2). OXI-A/N (Neonate): ± 4% (70% to 100% SpO2). | Same |
| Resolution | 1 % | 1 % | Same |
| Pulse Rate Measuring Range | 30 to 300 bpm | 20bpm to 300bpm | Different |
| Pulse Rate Accuracy | ±2 % or 1 bpm, whichever is greater | ± 3bpm (20bpm to 250bpm) | Different |
| Pulse Rate Resolution | 1 bpm | 1 bpm | Same |
| Temperature Function | |||
| Measurement Range | -1 to +49°C (30 to 113°F) | 0 to 50°C | Different |
| Accuracy | ±0.1°C (±0.2°F) | ±0.1 °C | Same |
| Resolution | 0.1°C (±0.1°F) | 0.1 °C | Same |
| Average Time Constant | Less than 10 seconds | <30 s | Different |
| IBP Function | |||
| Measurement Range | -40 to 360 mmHg | -50 to +300 mmHg | Different |
| Accuracy | 4 % of reading or ±4 mmHg, whichever is greater | ± 2 % or ±1 mmHg, whichever is greater | Different |
| Resolution | 1 mmHg | 1 mmHg | Same |
| Transducer Load/Output Impedance | Load Impedance: 200 to 2000 Ω, Output Impedance: 3000 Ω | 5 (μV/V/mmHg) 300 to 3000 Ω | Different |
| Zero Range | ±200 mmHg | ±200 mmHg | Same |
| General Specifications | |||
| Electrical Safety Anti-electroshock type | Class II | Class II | Same |
| Working system | Continuous operation equipment | Continuous operation equipment | Same |
| Operating Temp Range | 0 to 40°C (32 to 104°F) | 0°C to +40°C | Same |
| Storage Temp Range | -20 to 60°C (-4 to 140°F) | -30°C to +70°C | Different |
| Operating Humidity Range | 15 % to 95 % RH | 15% to 95% (non-condensing) | Same |
| Storage Humidity Range | 5 % to 95 % RH | 15% to 95% (non-condensing) | Different |
| Operating Altitude Range | -500 m to 3000 m (10000 ft) | 615hPa to 1060hPa (DC-DC power supply)(-400 m to 4000 m), 680hPa to 1060hPa (AC to DC power supply)(-400m to 3000m) | Different |
| Storage Altitude Range | -500 m to 4600 m (15000 ft) | 615hPa to 1060hPa (-400 m to 4000 m) | Different |
| Weight | 1.25 kg | iM20: <1.5kg, EFM: < 0.58 kg | Different |
| Dimensions | (W x H x D) 188 x 99 x 86 mm | iM20: 185 mm (L) × 85.3 mm (W)× 116 mm (H), EFM: 207 mm (L) × 93.4 mm (W)× 116 mm (H) | Different |
| Display Description | 72 x 54 mm (2.8 x 2.1 in) | 5 inches LCD | Different |
| Display Resolution | 320 x 240 | 800x480 | Different |
| Power consumption | <12 W average, <30 W while battery is loading | <25w (Max) | Different |
| Power input | 36 to 60 V DC floating | 11.1V to 19.8VDC | Different |
| Current | 0.7 to 0.4 A | 1.27~2.3A | Different |
| AC-DC Power supply Line Voltage | 100 to 240 V ~ | 100~240Vac | Same |
| AC-DC Power supply Output Voltage | 36 to 60 V DC floating | 15V±5%dc, 24VA | Different |
| AC-DC Power supply Frequency | 50/60 Hz ~ | 50/60 Hz ~ | Same |
| DC-DC Power supply Line Voltage | N/A | 10~16Vdc | Different |
| DC-DC Power supply Output Voltage | N/A | 15V±5%dc | Different |
| Standard Compliance (General) | IEC 60601-1:1988 + A1:1991 + A2:1995 | IEC 60601-1:2005 | Different |
| Out-Of-Hospital Transport Shock Tests | According to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-27 (peak acceleration up to 100g). | According to IEC TR 60721-4-7, Class 7M3. IEC/EN60068-2-27 (peak acceleration up to 100g). | Same |
| Out-Of-Hospital Transport Random Vibration | According to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-64 (RMS acceleration 5g). | According to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-64 (RMS acceleration 5g). | Same |
| Out-Of-Hospital Transport Sinusoidal Vibration | According to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-6 (acceleration up to amplitude 2g). | According to EN1789:2007+A: 2010, 6.4.1 vibration test, IEC/EN 60068-2-6 (acceleration up to amplitude 2g). | Different |
| Out-Of-Hospital Transport Bump | -- (Not specified for predicate) | According to EN1789:2007+A: 2010, 6.4.1 Bump test, IEC/EN 60068-2-6 (acceleration up to amplitude 15g). | N/A |
| Out-Of-Hospital Transport Drop Test | According to EN1789 (covers also IEC TR 60721-4-7 and Class 7M3). IEC/EN 60068-2-32 (height 0.75 m). | according to EN1789:2007+A:2010, IEC 60068-2-32 (height 1.2 m). | Different |
| Degrees of Protection | provided by enclosures according to IEC/EN 60529: IP32 | provided by enclosures according to IEC/EN 60529: IP44 | Different |
| Medical Vehicles and Equipments | EN 1789 +A1:2003 Medical vehicles and their equipment - Road ambulances (chapter 6 - Medical Devices). | EN1789:2007+A: 2010 Medical vehicles and their equipment - Road ambulances (chapter 6 – Medical Devices). | Same |
| Radiated Susceptibility | 20 V/m according to EN ISO 9919 (SpO2) and EN ISO 21647 (CO2). | 20 V/m according to EN ISO 9919 (SpO2) and EN ISO 21647 (CO2). | Same |
| Alarm Categories | Physiological alarms, Technical alarms, Prompt information | Physiological alarms, Technical alarms, Prompt information | Same |
| Alarm Levels | High level alarms, Medium level alarms, Low level alarms | High level alarms, Medium level alarms, Low level alarms | Same |
| Alarm Modes | Visual alarms, Audible alarms, Alarm messages, Parameter flashes | Visual alarms, Audible alarms, Alarm messages, Parameter flashes | Same |
Comparison to Reference Device iM70 (Specifically for PR from NIBP and CO2 Function)
| Feature / Parameter | Acceptance Criteria (Reference: iM70) | Reported Device Performance (iM20) | Match? |
|---|---|---|---|
| PR from NIBP | |||
| Measurement range | 40 to 240 bpm | 40 to 240 bpm (EDAN), 30 bpm ~220bpm (SunTech) | Same/Different |
| Accuracy | ±3bpm or 3.5%, whichever is greater | ±3bpm or 3.5%, whichever is greater (EDAN); ±3bpm or ±2%, whichever is greater (SunTech) | Same/Different |
| CO2 Function | |||
| EtCO2 Range | 0 to 150 mmHg | 0 mmHg ~ 150 mmHg | Same |
| FiCO2 Range | 3 mmHg to 50 mmHg | 3 mmHg~50 mmHg | Same |
| AwRR Range | 2 to 150 rpm | 2 rpm ~ 150 rpm (Sidestream) | Same |
| EtCO2 Resolution | 1.0 mmHg | 1mmHg | Same |
| FiCO2 Resolution | 1mmHg | 1mmHg | Same |
| AwRR Resolution | ±1 rpm | 1 rpm | Same |
| EtCO2 Accuracy (0-40 mmHg) | ±2.0 mmHg | ± 2 mmHg | Same |
| EtCO2 Accuracy (41-70 mmHg) | ±5 % of reading | ± 5 % | Same |
| EtCO2 Accuracy (71-100 mmHg) | ±8 % of reading | ± 8 % | Same |
| EtCO2 Accuracy (101-150 mmHg)| ±10 % of reading | ± 10 % | Same | |
| RESP > 80rpm | ±12 % of actual | ± 12 % | Same |
| AwRR Accuracy | ±1 rpm | ± 1 rpm | Same |
| Sample Gas Flow rate | 50±10ml/min | 50±10 ml/min | Same |
| O2 Compensation Range | 0 ~ 100% | 0~100% | Same |
2. Sample Sizes and Data Provenance for Test Set
- Clinical Testing Sample Size: Not explicitly stated in terms of patient numbers or case count.
- Data Provenance: The NIBP clinical testing was conducted as per ISO 81060-2: 2013, "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation of Automated Measurement Type." The document does not specify the country of origin for the clinical data or whether it was retrospective or prospective.
3. Number of Experts and Qualifications for Ground Truth of Test Set
The document does not provide information on:
- The number of experts used to establish ground truth for the test set.
- The qualifications of those experts.
4. Adjudication Method for the Test Set
The document does not provide information on the adjudication method used for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No information about a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is provided. The study focuses on the device's technical and clinical performance against standards and a predicate device, not on human reader performance with or without AI assistance.
6. Standalone Performance Study (Algorithm Only)
The document primarily describes the performance of the entire "iM20 Patient Monitor device, consisting of all the modules and accessories in the system." While it details performance parameters of various physiological monitoring algorithms (ECG, NIBP, SpO2, CO2), it does not explicitly describe a standalone (algorithm only without human-in-the-loop performance) study in the detailed sense of a specific data set applied only to the algorithm for evaluation. However, the bench testing and NIBP clinical validation are evaluations of the device's inherent measurement capabilities.
7. Type of Ground Truth Used
- For ECG, RESP, NIBP, SpO2, IBP, TEMP, CO2: The measurements of these physiological parameters are compared against the expected performance defined by international standards (e.g., ISO, IEC) or against the performance of a predicate device. For NIBP, the ground truth is established through clinical validation against "automated measurement type" as per ISO 81060-2:2013, which typically involves comparison to a reference standard measurement method (e.g., auscultatory method).
- For Biocompatibility: Compliance with ISO 10993-1, (Cytotoxicity, Skin Sensitization, Skin Irritation).
- For Electrical Safety and EMC: Compliance with IEC 60601-1:2005/A1:2012 and IEC 60601-1-2:2007.
- For General Performance (Bench Testing): Compliance with various IEC/ISO standards specific to each parameter.
8. Sample Size for the Training Set
The document does not provide any information regarding a training set size, as it describes a medical device clearance submission based on equivalence to predicate devices and adherence to performance standards, not a machine learning model requiring a training set.
9. How Ground Truth for Training Set was Established
Not applicable, as no training set is mentioned in the context of this device's submission.
Ask a specific question about this device
Page 1 of 1