K Number
K152552
Device Name
Patient Monitor
Date Cleared
2016-04-29

(234 days)

Regulation Number
870.1025
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

iM20 Patient Monitor (hereinafter called monitor) is intended to be used for monitoring, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG. respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2).

The arrhythmia detection and ST Segment analysis are intended for adult only.

The monitor is additionally intended for use during patient transport inside of the hospital environment.

The monitor is not intended for airplane, helicopter transport, home use and MRI environments.

Device Description

The iM20 Patient Monitor System (hereinafter called iM20) can perform long-time continuous monitoring of multiple physiological parameters, including ECG, respiration (RESP), non-invasive blood pressure (NIBP), oxygen saturation of the blood (SpO2), pulse rate (PR), temperature (TEMP), invasive pressure (IBP), Carbon Dioxide (CO2). The system capabilities include storing, displaying, analyzing and controlling such data afterwards. When necessary, alarms will be produced so that doctors and nurses can manage patient care appropriately. The system is intended to be used during patient transport inside and outside of the hospital environment.
iM20 can be utilized in two ways, as an independent monitor and as a module of V series Patient Monitor (including models elite V5, elite V6 and elite V8). When used as an independent monitor, it can simultaneously monitor, store, review several parameters data. And transfer patient data to V series patient monitor only under the transport mode. As a highly portable monitor, its compact design makes it particularly appropriate inside hospital and vehicle ambulance transport environments.
When as a multi-measurement module, when the iM20 is directly connected to a V series patient monitor, it can provide the measurements, trends, and patient information. When connected, the V series Patient Monitor controls the connected iM20, including all alarm functionality. So no alarms are available on iM20 in such application and iM20 takes power from the V series Patient Monitor.

AI/ML Overview

This document focuses on the Edan Instruments Inc. Patient Monitor, model iM20 (K152552), and its equivalency to predicate devices. It lists performance data based on non-clinical and clinical studies to demonstrate substantial equivalence.

Here's the breakdown of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides a detailed comparison table between the proposed device (iM20) and a predicate device (Philips Medizin.Systeme Boeblingen GmbH IntelliVue MP2 cleared under K102562), focusing on various physiological parameters and technical specifications. A second comparison table is provided for the reference device Edan Instruments, Inc. iM70 (K131971), specifically for PR from NIBP and CO2 function.

Since the provided text does not present a separate "acceptance criteria" table and "reported device performance" table, but rather a comparison with performance metrics of a predicate device, the following table synthesizes the information from the "Predicate Device Comparison" section to represent acceptance criteria (from the predicate) and the reported device performance (for the iM20). Differences are explicitly noted as "Different." Where the proposed device's performance aligns with the predicate, it is marked "Same."

Acceptance Criteria (Based on Predicate Device) and Reported Device Performance (iM20)

Feature / ParameterAcceptance Criteria (Predicate: Philips MP2)Reported Device Performance (iM20)Match?
ECG Function - HR Calculation
HR Range (Adult/Pedi/Neo)Adult/pedi: 15 to 300 bpm, Neo: 15 to 350 bpmADU: 15 bpm to 300 bpm, PED/NEO: 15 bpm to 350 bpmSame
HR Accuracy±1% of range± 1% or 1 bpm, whichever is greaterDifferent
HR Resolution1 bpm1 bpmSame
Sensitivity≥ 200 µVpeak≥ 300 µVPPDifferent
PVC Rate Range0 to 300 bpmADU: 0 to 300 PVCs/ min, PED/NEO: 0 to 350 PVCs/ minDifferent
PVC Resolution1 bpm1 PVCs/minDifferent
ST Numeric Range-20 to +20 mm-2.0 mV to +2.0 mVDifferent
ST Accuracy± 0.5mm or 15%, whichever is greater-0.8 mV to +0.8 mV: ± 0.02 mV or 10%, whichever one is greater. Beyond this range: not specified.Different
ST Resolution0.1mm0.01 mVDifferent
Range of Sinus and SV Rhythms (Brady)Adult: 15 to 60 bpm, Pedi: 15 to 80 bpm, Neo: 15 to 90 bpmAdult: RR interval for 5 consecutive QRS complex ≥ 1.5 s. Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≥ 1 s.Different
Range of Sinus and SV Rhythms (Normal)Adult: 60 to 100 bpm, Pedi: 100 to 160 bpm, Neo: 90 to 180 bpmAdult: 0.5s 100 bpm, Pedi: >160 bpm, Neo: >180 bpmAdult: RR interval for 5 consecutive QRS complex ≤ 0.5 s. Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≤ 0.375 s.
BandwidthDiagnosis: 0.05 to 150 Hz, Monitor: Adult: 0.5 to 40 Hz, Neo/pedi: 0.5 to 55 Hz, Filter Mode: 0.5 to 20 HzDiagnosis: 0.05Hz to 150Hz, Monitor: 0.5Hz to 40Hz, Surgery: 1Hz to 20HzDifferent
CMRRDiagnostic mode: >86 dB, Filter mode: >106 dBDiagnosis: >95dB (Notch filter off), Monitor: >105dB (Notch filter on), Surgery: >105dB (Notch filter on)Different
Differential Input Impedance>2 MΩ RA-LL leads (Resp), >5 MΩ at all other leads>5MΩDifferent
Input Signal Range±5mV (peak-to-peak value)±10mVPPDifferent
Electrode Offset Potential Tolerance±500mV±500mVSame
Auxiliary Current (detect)Active electrode: 1200ms.Different
Accuracy of HR Meter and Response to Irregular RhythmVentricular bigeminy: 80 bpm, Slow alternating ventricular bigeminy: 60 bpm, Rapid alternating ventricular bigeminy: 120 bpm, Bidirectional systoles: 90 bpmVentricular bigeminy: 80 bpm±1 bpm, Slow alternating ventricular bigeminy: 60 bpm±1 bpm, Rapid alternating ventricular bigeminy: 120 bpm±1 bpm, Bidirectional systoles: 91 bpm±1 bpmDifferent
Response time of HR Meter to Change in HR (80 to 120 bpm)Range: [6.4 to 7.2 s], Average: 6.8 sWithin 11 sDifferent
Response time of HR Meter to Change in HR (80 to 40 bpm)Range: [5.6 to 6.4 s], Average: 6.0 sWithin 11 sDifferent
RESP Function
Respiration excitation waveformSinusoidal signal, 260 µA, 39 kHzSinusoid, 62.8 kHz (±10%), 300 mmHg (40 kPa) >2 sec297 ± 3mmHg (EDAN), 300 mmHg (40 kPa) >2 sec
AwRR Accuracy±1 rpm± 1 rpmSame
Sample Gas Flow rate50±10ml/min50±10 ml/minSame
O2 Compensation Range0 ~ 100%0~100%Same

2. Sample Sizes and Data Provenance for Test Set

  • Clinical Testing Sample Size: Not explicitly stated in terms of patient numbers or case count.
  • Data Provenance: The NIBP clinical testing was conducted as per ISO 81060-2: 2013, "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation of Automated Measurement Type." The document does not specify the country of origin for the clinical data or whether it was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth of Test Set

The document does not provide information on:

  • The number of experts used to establish ground truth for the test set.
  • The qualifications of those experts.

4. Adjudication Method for the Test Set

The document does not provide information on the adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No information about a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is provided. The study focuses on the device's technical and clinical performance against standards and a predicate device, not on human reader performance with or without AI assistance.

6. Standalone Performance Study (Algorithm Only)

The document primarily describes the performance of the entire "iM20 Patient Monitor device, consisting of all the modules and accessories in the system." While it details performance parameters of various physiological monitoring algorithms (ECG, NIBP, SpO2, CO2), it does not explicitly describe a standalone (algorithm only without human-in-the-loop performance) study in the detailed sense of a specific data set applied only to the algorithm for evaluation. However, the bench testing and NIBP clinical validation are evaluations of the device's inherent measurement capabilities.

7. Type of Ground Truth Used

  • For ECG, RESP, NIBP, SpO2, IBP, TEMP, CO2: The measurements of these physiological parameters are compared against the expected performance defined by international standards (e.g., ISO, IEC) or against the performance of a predicate device. For NIBP, the ground truth is established through clinical validation against "automated measurement type" as per ISO 81060-2:2013, which typically involves comparison to a reference standard measurement method (e.g., auscultatory method).
  • For Biocompatibility: Compliance with ISO 10993-1, (Cytotoxicity, Skin Sensitization, Skin Irritation).
  • For Electrical Safety and EMC: Compliance with IEC 60601-1:2005/A1:2012 and IEC 60601-1-2:2007.
  • For General Performance (Bench Testing): Compliance with various IEC/ISO standards specific to each parameter.

8. Sample Size for the Training Set

The document does not provide any information regarding a training set size, as it describes a medical device clearance submission based on equivalence to predicate devices and adherence to performance standards, not a machine learning model requiring a training set.

9. How Ground Truth for Training Set was Established

Not applicable, as no training set is mentioned in the context of this device's submission.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.