Search Results
Found 10 results
510(k) Data Aggregation
(57 days)
The hip prosthesis AMIS-K Long is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery.
Total hip replacement is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia
- Avascular necrosis of the femoral head
- · Acute traumatic fracture of the femoral head or neck
· Failure of previous hip surgery: joint reconstruction, arthrodesis, partial hip arthroplasty, hip resurfacing replacement, or total hip arthroplasty.
Partial hip arthroplasty is indicated in the following cases:
· Acute traumatic fracture of the femoral head or neck
- · Non-union of femoral neck fracture
- · Avascular necrosis of the femoral head.
- · Primary pathology involving the femoral head but with a non-deformed acetabulum.
The AMIS-K Long cemented stem is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery.
The AMIS-K Long is a straight, double-tapered cemented stem whose primary stability is ensured by bone cement.
The AMIS-K Long implants in this submission are comprised of the following size: from #2 to #5 with 2 different stem body size for #2, #3 and #4, 1 different stem body size for #5. Stem length: 200 - 250 - 300 mm (for each size).
The AMIS-K Long implants are part of the Medacta Total Hip Prosthesis system.
The Medacta Total Hip Prosthesis system consists of femoral stemoral heads, and acetabular components.
The AMIS-K Long are cemented stems manufactured from high nitrogen stainless steel (ISO 5832-9) with a mirror polished surface.
The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE), or Highcross highly crosslinked ultra-high polyethylene (HXUHMWE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352, K120531 and K122911), Mpact (K103721 and K132879), Mpact 3D Metal (K171966), and Medacta Bipolar Head (K091967).
The AMIS-K Long implants can be combined with the CoCr Ball Heads (K072857 and K080885). Endo Head (K111145), or MectaCer BIOLOX® Forte (K073337) or MectaCer BIOLOX® Delta Femoral Heads (K112115).
The provided text is a 510(k) summary for a medical device (Amis K Long hip prosthesis), which primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria in the context of AI/ML or diagnostic performance.
Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving a device meets these criteria in the context of AI/ML performance. Specifically:
- No acceptance criteria for AI/ML performance are listed. The acceptance criteria mentioned refer to non-clinical performance tests for the physical hip prosthesis (e.g., pull-off strength, fatigue testing).
- No study proving AI/ML device performance is described. The document explicitly states "No clinical studies were conducted." and describes non-clinical mechanical tests.
- The requested elements (sample size, data provenance, expert qualifications, etc.) are not applicable in this context as the document is about a physical implant, not an AI/ML diagnostic or assistive device.
The study described in the document primarily focuses on non-clinical mechanical tests to validate the physical properties of the hip prosthesis, such as pull-off strength, range of motion, and fatigue. These tests adhere to established international standards (ASTM, EN ISO, ISO).
In summary, the provided text does not contain the information necessary to answer the questions about acceptance criteria and a study proving device performance as they relate to AI/ML or diagnostic accuracy.
Ask a specific question about this device
(91 days)
The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia
· Avascular necrosis of the femoral head
· Acute traumatic fracture of the femoral head or neck
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.
The SMS femoral stem is a cementless bone preserving short stem designed for proximal fixation in total or partial hip arthroplasty for primary or revision surgery.
The SMS implants are comprised of the following products:
- SMS Cementless Solid Standard Stem (available in 11 sizes from size 3 to 13); and ●
- . SMS Cementless Solid Lateralized Stem (available in 11 sizes - from size 3 to 13).
Both are available on the US market via the clearance - K181693.
Concerning the new sizes of the solid version: 1V2, 2V2, size 14, and size 15; both STD and LAT versions have been introduced as a product range extension to the stems currently marketed as the SMS product line (K181693).
The SMS implants are line extensions to Medacta's Total Hip Prosthesis - AMIStem-H, Quadra-S and Quadra-H Femoral Stems (K093944), AMIStem and Quadra - Line Extension (K121011), AMIStem-P, AMIStem-P Collared and AMIStem-H Proximal Coating Femoral Stems (K173794), Quadra-H and Quadra-R Femoral Stems (K082792), AMIStem-H Proximal Coating (K161635), MiniMAX (K170845), and SMS (K181693).
The SMS implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966) and Medacta Bipolar Head (K091967).
The SMS stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).
This is a 510(k) premarket notification for the "SMS Cementless Stem" (K201673), a hip prosthesis. The document details the device, its intended use, and a comparison to predicate devices, along with performance data.
Here's an analysis of the provided text in relation to your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document outlines various non-clinical performance tests conducted for the SMS Cementless Stem. However, it does not provide specific acceptance criteria values or the quantitative results of these tests in a table format. It only states that the testing was conducted "according to written protocols with acceptance criteria that were based on standards."
Here's what can be inferred for a table, but actual criteria values and results are missing in the document:
Test (Standard) | Acceptance Criteria (Not provided explicitly, generally "meet standard requirements") | Reported Device Performance (Not provided explicitly, implied "met criteria") |
---|---|---|
Range of Motion (EN ISO 21535:2009) | Based on standard EN ISO 21535:2009 | Conducted |
Fatigue Testing - Stemmed Femoral Components (ISO 7206-4:2010 + AME 1:2016) | Based on standard ISO 7206-4 | Conducted |
Fatigue Testing - Head and Neck Region of Stemmed Femoral Components (ISO 7206-6:2013) | Based on standard ISO 7206-6 | Conducted |
Pull-off Force Testing - Taper Connections (ASTM F2009-00 (Reapproved 2011)) | Based on standard ASTM F2009-00 | Conducted |
Coating Tests - Hydroxyapatite (ISO 13779-1:2008) | Based on standard ISO 13779-1 | Conducted |
Coating Tests - Tension Testing of Calcium Phosphate and Metal Coatings (ASTM F1147-99) | Based on standard ASTM F1147-99 | Conducted |
Pyrogenicity - Bacterial Endotoxin Test (European Pharmacopoeia §2.6.14 / USP chapter ) | Based on European Pharmacopoeia §2.6.14 / USP chapter | Conducted |
Pyrogenicity - Pyrogen Test (USP chapter ) | Based on USP chapter | Conducted |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "No clinical studies were conducted." Therefore, there is no test set in the context of human clinical data or patient outcomes. The "performance data" refers to non-clinical mechanical and material tests. The sample sizes for these mechanical tests are not provided in this summary. Data provenance is not applicable here as no human data was used.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable as no clinical studies were conducted and no ground truth in the clinical sense was established. The "ground truth" for the non-clinical tests would be the specifications and requirements of the referenced ISO and ASTM standards.
4. Adjudication Method for the Test Set
Not applicable as no clinical studies were conducted which would require adjudication of expert opinions or outcomes.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size
No MRMC study was mentioned or performed, as "No clinical studies were conducted." This device is a mechanical implant, not an AI or diagnostic tool that would typically undergo MRMC studies.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a hip implant, not an algorithm or AI system.
7. The Type of Ground Truth Used
For the non-clinical performance and coating tests, the "ground truth" implicitly refers to the
requirements and specifications outlined in the referenced international standards (ISO and ASTM) and the European Pharmacopoeia/USP chapters. The device is expected to meet these pre-defined engineering and material standards. This is a form of engineering/material specification compliance as ground truth.
8. The Sample Size for the Training Set
Not applicable. This device is a mechanical implant, not an AI or machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of medical device's approval.
Ask a specific question about this device
(29 days)
The hip prosthesis MiniMAX is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:
• Severely painful and/or disabled joint as a result of arthritis, traumatoid polyarthritis, or congenital hip dysplasia.
· Avascular necrosis of the femoral head.
· Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
MiniMax anatomical stems are coated cementless stems intended to be used in total or partial hip arthroplasty in primary or revision surgery.
The anatomical design of the MiniMax stems results in a 9° anteversion of the neck which is mirror polished with 12/14 Eurocone taper and a 127° neck-shaft angle. The macrostructures are negative medially and positive laterally which increases the contact area. The lateral flare is rounded and non-invasive. The MiniMAX stems' distal tip has a 5° curvature to follow the contours of the femoral canal to aid in the insertion and avoiding distal interference.
MiniMAX stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).
MiniMAX stems are made of Titanium Aluminum Niobium Alloy (Ti-6Al-7Nb). The surface treatment consists of titanium plasma spray coating, Ra 300um, in the proximal 2/3 of the shaft to improve proximal fixation and HA (Hydroxyapatite) coating, Ra 80um, along the entire length of the shaft.
This submission seeks add Size 9 left and right anatomical stems to the currently marketed MiniMAX product line.
The provided text describes a 510(k) premarket notification for a medical device (MiniMAX hip prosthesis) and does not contain information about the acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML powered device. The document is a traditional medical device submission focused on substantial equivalence to a predicate device, primarily through mechanical testing and material characterization.
Therefore, most of the requested information regarding acceptance criteria, device performance tables, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and MRMC/standalone studies for an AI/ML device is not present in the provided text.
However, I can extract the information that is present regarding the device's performance data and how it was evaluated.
1. A table of acceptance criteria and the reported device performance
The document does not provide a table of acceptance criteria with numerical targets and reported performance for an AI/ML device. Instead, it lists the types of performance tests conducted for a mechanical hip prosthesis and states that the "subject devices do not represent a new worst case when compared to the previously cleared devices."
Test Type | Standard/Reference | Reported Outcome |
---|---|---|
Range of Motion (ROM) | EN ISO 21535:2009 (Joint Replacement Implants - Specific Requirements for Hip-Joint Replacement Implants) | "subject devices do not represent a new worst case when compared to the previously cleared devices (K170845)." |
Fatigue Testing (Stem) | ISO 7206-4 Third Edition 2010-06-15 (Implants for Surgery Partial and Total Hip Joint Prostheses - Part 4: Determination of Endurance Properties and Performance of Stemmed Femoral Components) | "subject devices do not represent a new worst case when compared to the previously cleared devices (K170845)." |
Fatigue Testing (Head and Neck Region) | ISO 7206-6 Second Edition 2013-11-15 (Implants for Surgery Partial and Total Hip Joint Prostheses - Part 6: Determination of Endurance Properties of Head and Neck Region of Stemmed Femoral Components) | "subject devices do not represent a new worst case when compared to the previously cleared devices (K170845)." |
Static Fatigue Testing (Modular Femoral Heads) | ISO 7206-10:2003 (Implants for Surgery -- Partial and Total Hip-Joint Prostheses -- Part 10: Determination of Resistance To Static Load of Modular Femoral Heads) | "subject devices do not represent a new worst case when compared to the previously cleared devices (K170845)." |
Pull Off Force Testing (Taper Connections) | ASTM F2009-00 (Reapproved 2011) (Standard Test Method O for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses) | "subject devices do not represent a new worst case when compared to the previously cleared devices (K170845)." |
Coating Characterization Testing | Not explicitly stated, implied to be standard material characterization. | "subject devices do not represent a new worst case when compared to the previously cleared devices (K170845)." |
Pyrogenicity | USP and USP | "subject devices do not represent a new worst case when compared to the previously cleared devices (K170845)." |
Additional Evaluation (Stem and Neck Fatigue) | ISO 21535 (via Finite Element Analysis) | The analysis concluded that the "additional size (longer stem length and larger neck offset) does not introduce a new worst case with respect to biomechanical or clinical performance." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "worst-case device" testing for the predicate device, K170845. This typically implies a very small sample size, potentially N=1 or N=3, for each test as per engineering standards for mechanical devices. It's not a 'test set' in the sense of an AI/ML validation. There's no information on data provenance (country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is a mechanical device submission, not an AI/ML study requiring expert ground truth for interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a mechanical device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For mechanical testing, the "ground truth" is defined by the passing criteria of the specified international standards (e.g., ISO, ASTM). For example, fatigue tests would have an endurance limit or cycles to failure requirement specified in the standard.
8. The sample size for the training set
Not applicable. This is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(154 days)
The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia
· Avascular necrosis of the femoral head
· Acute traumatic fracture of the femoral head or neck
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthroplasty, hip resurfacing replacement or total hip arthroplasty.
The SMS femoral stem is a cementless bone preserving short stem designed for proximal fixation in total or partial hip arthroplasty for primary or revision surgery. The SMS implants subject of this submission are comprised of the following products:
- SMS Cementless Solid Standard Stem (available in 11 sizes); and ●
- SMS Cementless Solid Lateralized Stem (available in 11 sizes).
The SMS implants are line extensions to Medacta's Total Hip Prosthesis - AMIStem-H, Quadra-S and Quadra-H Femoral Stems (K093944), AMIStem and Quadra - Line Extension (K121011), AMIStem-P, AMIStem-P Collared and AMIStem-H Proximal Coating Femoral Stems (K173794), Quadra-H and Quadra-R Femoral Stems (K082792), AMIStem-H Proximal Coating (K161635), and MiniMAX (K170845).
The SMS implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polvethylene (HXUHMWPE).
Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966) and Medacta Bipolar Head (K091967).
The SMS stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).
This document describes the premarket notification (510(k)) for the SMS Femoral Stem, a cementless bone-preserving short stem designed for proximal fixation in total or partial hip arthroplasty.
Here's an analysis of the provided text in relation to acceptance criteria and supporting studies, though it's important to note that this is a medical device submission, not an AI/ML device submission, so the questions regarding human readers, training sets, etc., are not directly applicable here. The document focuses on mechanical performance and biocompatibility for a physical implant.
1. A table of acceptance criteria and the reported device performance
The document lists performance tests and the standards they adhere to, implying that meeting these standards constitutes the acceptance criteria. However, it does not explicitly state numerical acceptance criteria and does not present specific performance data values in a table format. It merely states that "Testing was conducted according to written protocols with acceptance criteria that were based on standards." and that the studies "demonstrated substantial equivalence."
Here's a generalized representation based on the provided information:
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
Conformance to EN ISO 21535:2009 (Range of Motion) | Testing conducted, supporting substantial equivalence. |
Conformance to ISO 7206-4:2010 (Fatigue Testing - Stem) | Testing conducted, supporting substantial equivalence. |
Conformance to ISO 7206-6:2013 (Fatigue Testing - Head/Neck) | Testing conducted, supporting substantial equivalence. |
Conformance to ISO 7206-10:2003 (Static Fatigue - Modular Heads) | Testing conducted, supporting substantial equivalence. |
Conformance to ASTM F2009-00 (Axial Disassembly Force) | Testing conducted, supporting substantial equivalence. |
Conformance to ISO 13779-1:2008 (Coating - Hydroxyapatite) | Testing conducted, supporting substantial equivalence. |
Conformance to ASTM F1147-99 (Tension Testing - Coatings) | Testing conducted, supporting substantial equivalence. |
Conformance to European Pharmacopoeia §2.6.14/USP (BET) | Testing conducted (LAL test), supporting substantial equivalence. |
Conformance to USP (Pyrogen Test) | Testing conducted, supporting substantial equivalence. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document lists various non-clinical mechanical and coating tests, which typically involve physical samples of the device components. It does not specify the sample sizes (number of stems tested for fatigue, etc.) for each test. The provenance of the data is implicit: it's generated from laboratory testing of the device components. There is no clinical data or patient-derived data mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is not applicable to this type of medical device submission. Ground truth and expert adjudication are relevant for AI/ML diagnostic devices, not for physical implants that undergo mechanical and material testing. The "ground truth" for these tests are the established scientific principles and measurement techniques outlined in the specified ISO and ASTM standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable for the same reasons as #3. Mechanical testing results are typically objectively measured and evaluated against a standard, not subject to expert adjudication in the same way as diagnostic interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable as this is not an AI/ML device. No MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable as this is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical tests, the "ground truth" is defined by the specifications and acceptable ranges outlined in the referenced international (ISO) and American (ASTM) standards. For instance, a fatigue test determines if the device can withstand a certain number of cycles at a specific load without failure, as per the standard's requirements. Pyrogenicity is determined by established biological testing methods (LAL test, USP pyrogen test).
8. The sample size for the training set
This question is not applicable as this is not an AI/ML device. There is no concept of a "training set" for a physical implant undergoing mechanical and material characterization.
9. How the ground truth for the training set was established
This question is not applicable as this is not an AI/ML device.
Ask a specific question about this device
(164 days)
The hip prosthesis QUADRA-P is designed for cementless use in total or partial hip arthroplasty; for primary or revision surgery. Hip replacement is indicated in the following cases:
• Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia
· Avascular necrosis of the femoral head
· Acute traumatic fracture of the femoral head or neck
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthroplasty, hip resurfacing replacement or total hip arthroplasty.
The Quadra-P implants are line extensions to Medacta's Total Hip Prosthesis – AMIStem-H, Quadra-S, and Quadra-H Femoral Stems (K093944); AMIStem and Quadra - Line Extension (K121011); AMIStem-P, AMIStem-P Collared, and AMIStem-H Proximal Coating Femoral Stems (K173794); Quadra-H and Quadra-R Femoral Stems (K082792); and AMIStem-H Proximal Coating (K161635).
The Quadra-P implants subject of this submission are comprised of the following products:
- . Quadra-P STD. Stem sizes 0 - 10;
- Quadra-P LAT, Stem sizes 0 - 10.
The Quadra-P implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads, and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966), and Medacta Bipolar Head (K091967).
The Quadra-P implants can be combined with the CoCr Ball Heads (K072857 and K080885), Endo Head (K111145), MectaCer Biolox Option Heads (K131518), or with the MectaCer BIOLOX® Forte (K073337) or MectaCer BIOLOX® Delta Femoral Heads (K112115).
This document is a 510(k) premarket notification for the "Quadra-P" hip prosthesis, specifically describing the physical and mechanical testing performed to demonstrate substantial equivalence to previously cleared predicate devices. It does not contain information about the performance or acceptance criteria of an AI/Software device, nor does it describe a study involving humans or the establishment of ground truth for such a device.
Therefore, I cannot extract the information required to answer your questions about acceptance criteria and studies proving the device meets those criteria, as the provided document pertains to a physical medical device (hip prosthesis) and its mechanical engineering performance tests, not an AI or software-based medical device.
The document discusses:
- Device Name: Quadra-P (hip prosthesis)
- Regulation Number: 21 CFR 888.3353 (Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis)
- Performance Data (Non-Clinical Studies):
- Range of Motion (ROM) testing (EN ISO 21535:2009)
- Fatigue testing (ISO 7206-4 and ISO 7206-6)
- Pull off force testing (ASTM F2009-00)
- Pyrogenicity testing (European Pharmacopoeia §2.6.14/USP and USP )
- Clinical Studies: "No clinical studies were conducted."
Since the request is specifically about a device proving it meets acceptance criteria through a study involving a test set, expert ground truth, adjudication methods, and potentially human readers (which implies an AI/Software device or diagnostic device), the provided text is entirely irrelevant to those questions.
Ask a specific question about this device
(206 days)
MicroPort total hip systems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
-
non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli. and painful hip dysplasia;
-
inflammatory degenerative joint disease such as rheumatoid arthritis;
-
correction of functional deformity; and,
-
revision procedures where other treatments or devices have failed.
The BIOLOX® delta Option and Extra-long Heads include femoral heads and neck sleeves. The design features are summarized below:
- Femoral Heads
- BIOLOX® delta Option:
- Manufactured from alumina matrix composite
- Head sizes: 28mm, 32mm, 36mm, 40mm, 44mm
- Conical Bore: 16/18 Taper
- BIOLOX® delta Extra-long:
- Manufactured from alumina matrix composite
- Head Sizes: 32mm, 36mm, 40mm
- Conical Bore: 12/14 Taper
- BIOLOX® delta Option:
- Neck Sleeves
- BIOLOX® delta Option:
- Manufactured from titanium alloy
- Neck Sleeve sizes: Short, Medium, Long, and Extra-long
- Conical Bore: 16/18 External Taper, 12/14 Internal Taper
- BIOLOX® delta Option:
The provided document describes a 510(k) premarket notification for the BIOLOX® delta Option and Extra-long Heads, which are components of total hip arthroplasty systems. Per the document, the device did not undergo clinical testing, and therefore, no acceptance criteria based on clinical performance metrics are provided. The clearance is based on substantial equivalence to predicate devices and non-clinical testing.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
Since this is a 510(k) submission based on substantial equivalence and non-clinical testing, specific acceptance criteria in terms of clinical performance metrics (e.g., sensitivity, specificity, accuracy, or reduction in error rate) are not stated in the document. The non-clinical tests assess the physical and mechanical properties of the device. The document states that the safety and effectiveness are "adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification." The reported "performance" for these non-clinical tests would typically be within the expected range for such devices and comparable to predicate devices, but the specific numerical results or acceptance thresholds are not detailed in this summary.
Acceptance Criteria Category | Reported Device Performance (Summary) |
---|---|
Mechanical Performance | |
Static (Pre-fatigue) Burst Strength | Evaluated (results not specified) |
Post-fatigue Burst Strength | Evaluated (results not specified) |
Taper System Stability | |
Axial Pull-off Test (Distraction Forces) | Evaluated (results not specified) |
Rotational Stability (Torque) | Evaluated (results not specified) |
Corrosion | |
Fretting Corrosion (Sleeve & Tapers) | Tested (results not specified) |
Biocompatibility | Referencing Master File and predicate/reference devices (no new testing) |
Component Compatibility | Compatible with listed MicroPort Orthopedics products (See Tables 1 & 2) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document primarily discusses non-clinical testing. For these mechanical tests, the sample sizes are not specified, nor is the country of origin of the data mentioned. The data provenance is non-clinical (laboratory testing). No human test set data is presented.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. The document describes non-clinical testing of a hip implant device, not a diagnostic or AI-based device requiring expert ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a study requiring adjudication of expert interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document pertains to the mechanical and biological safety of an orthopedic implant, not an AI-assisted diagnostic tool. No MRMC study was conducted or mentioned.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical tests, the "ground truth" is based on established engineering principles and material science standards for evaluating the performance and safety of medical devices. For instance, burst strength tests have defined failure points based on material properties and structural integrity. For biocompatibility, the ground truth is established through validated testing methods outlined in international standards (e.g., ISO 10993) or by referencing previously cleared devices with known biological responses.
8. The sample size for the training set
Not applicable. This is a physical medical device, not an AI model requiring a training set.
9. How the ground truth for the training set was established
Not applicable. As above, this is a physical medical device, not an AI model.
Ask a specific question about this device
(140 days)
The hip prosthesis MiniMAX is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery.
Hip Replacement is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthritis, traumatoid polyarthritis, or congenital hip dysplasia.
· Avascular necrosis of the femoral head.
· Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
The purpose of this submission is to gain clearance for the new MiniMAX, which are anatomical HA coated cementless stems intended to be used in total or partial hip arthroplasty in primary or revision surgery.
The anatomical design of the MiniMAX stems results in a 9° anteversion of the neck which is mirror polished with 12/14 Eurocone taper and a 127° neck-shaft angle. The macrostructures are negative medially and positive laterally which increases the contact area. The lateral flare is rounded and non-invasive. The MiniMAX stems' distal tip have a 5° curvature to follow the contours of the femoral canal to aid in the insertion and avoiding distal interference.
MiniMAX stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).
MiniMAX stems are made of Titanium Aluminum Niobium Alloy (Ti-6Al-7Nb). The surface treatment consists of titanium plasma spray coating, Ra 300μm, in the proximal 2/3 of the shaft to improve proximal fixation and HA (Hydroxyapatite) coating, Ra 80um, along the entire length of the shaft.
MiniMAX stems are similar to predicate devices Howmedica Osteonics' ABG III Monolithic Hip Stem (K123604) and Medacta's MasterLoc Stem (K151531).
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Non-Clinical Mechanical Tests | Based on standards and FDA guidance documents (specific values not provided in this summary). | |
Range of Motion (ROM) | Standards: EN ISO 21535:2009 | Testing was conducted to written protocols with acceptance criteria. The summary states the data supports substantial equivalence, implying the device met these criteria. |
Fatigue Testing (Stem Endurance) | Standards: ISO 7206-4 Third Edition 2010-06-15 (with Amendment 1, 2016) | Testing was conducted to written protocols with acceptance criteria. The summary states the data supports substantial equivalence, implying the device met these criteria. |
Fatigue Testing (Head and Neck Region Endurance) | Standards: ISO 7206-6 Second Edition 2013-11-15 | Testing was conducted to written protocols with acceptance criteria. The summary states the data supports substantial equivalence, implying the device met these criteria. |
Static Fatigue Testing (Modular Femoral Heads) | Standards: ISO 7206-10:2003 | Testing was conducted to written protocols with acceptance criteria. The summary states the data supports substantial equivalence, implying the device met these criteria. |
Pull Off Force Testing (Taper Connections) | Standards: ASTM F2009-00 (Reapproved 2011) | Testing was conducted to written protocols with acceptance criteria. The summary states the data supports substantial equivalence, implying the device met these criteria. |
Coating Tests | (Not explicitly stated in this summary, but would be related to coating characteristics, adhesion, etc.) | Coating characterization testing was performed. The summary states the data supports substantial equivalence, implying the device met these criteria. |
Pyrogenicity | (Based on European Pharmacopoeia §2.6.14 and USP for Bacterial Endotoxin Test, and USP for Pyrogen Test). The device is not labeled as non-pyrogenic or pyrogen-free. | The Bacterial Endotoxin Test (LAL test) and Pyrogen Test were conducted. The summary implies the results were acceptable for the intended labeling. |
2. Sample size used for the test set and the data provenance
- Sample size for test set: Not explicitly stated in the provided document. The document mentions "testing was conducted to written protocols," but does not detail the number of units tested for each mechanical or coating test.
- Data provenance:
- Country of origin: Not specified beyond the manufacturer being Medacta International SA (Switzerland). The testing standards are international (ISO, ASTM, European Pharmacopoeia, USP).
- Retrospective or Prospective: Not applicable in this context as these are laboratory-based non-clinical performance and material tests, not clinical data from patients.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. The "ground truth" for non-clinical performance tests is established by adhering to widely recognized engineering and medical device standards (ISO, ASTM, EN, USP, European Pharmacopoeia) and internal protocols. These tests measure objective properties (e.g., fatigue strength, range of motion, coating characteristics) rather than requiring expert interpretation of clinical data as a "ground truth."
4. Adjudication method for the test set
- Not applicable. This concept typically applies to clinical studies where different readers or experts might have varying interpretations. For non-clinical, objective testing, results are measured against predefined acceptance criteria from established standards. There is no "adjudication" in the sense of reconciling differing expert opinions on the test outcomes.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This is a medical device (hip prosthesis), not an AI-based diagnostic or assistive software. The submission focuses on the safety and effectiveness of the implant itself through material properties and mechanical performance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- No, a standalone (algorithm only) performance study was not done. This is a hardware medical device, not a software algorithm.
7. The type of ground truth used
- The "ground truth" for the performance data in this submission is based on established international and national standards (e.g., ISO 7206 series for hip prostheses, ASTM F2009 for taper connections, EN ISO 21535 for hip-joint implants, European Pharmacopoeia and USP for pyrogenicity). These standards define the acceptable performance parameters and testing methodologies for such devices.
8. The sample size for the training set
- Not applicable. This submission describes a physical medical device. There is no "training set" in the context of an algorithm or machine learning model. The relevant "data" for development would involve extensive engineering design, material science data, and prior knowledge from similar predicate devices.
9. How the ground truth for the training set was established
- Not applicable. As there is no training set in the AI/ML sense, there is no ground truth established for a training set. The device design and manufacturing process would be informed by decades of biomechanical research, clinical outcomes with similar devices, and engineering principles. The "ground truth" for its design and manufacturing would be validated through the performance testing described.
Ask a specific question about this device
(122 days)
The Modified Bencox Hip System is intended for use in total or partial hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
The Modified Bencox Mirabo Hip System consists of the following components,
- Acetabular Insert Bencox Mirabo PE Liner (Std. & Elevated), .
- Femoral Stem Bencox ID Stem (Offset), .
- Femoral Head Bencox Delta Head (XL), ●
- Instrumentation Bencox Total Hip System Instrumentation. .
The components are modification of Acetabular Insert cleared in K120924 (& K103431), Femoral Stem cleared in K112019 & specification inclusion of Femoral Heads cleared in K121665. The modified components are cementless, metal-onpolyethylene hip system for hip arthroplasty similar to devices cleared in respective mentioned 510(k)'s.
Modified Bencox Mirabo PE Liner is manufactured from similar material cleared under K120924 & K103431, conforming to ASTM F648 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants, Type 2 (GUR 1050), and irradiated with average dose of 10.0 Mrad of gamma radiation similar to predicate device Bencox Mirabo PE Liner and other predicates such as Stryker Crossfire.
Modified Bencox ID Stem (Offset) is similar to the predicate device Bencox ID Stem cleared in K112019, except for the slightly thicker neck geometry and horizontal offset.
Bencox Delta Head XL specification inclusion consists of diameters 32, 36 & 40 mm design is similar to the 510k cleared devices, Bencox Delta Head (K121665). The specification is also similar with other predicate devices and has the same supplier, CeramTec, AG.
This document describes the Modified Bencox Hip System, a hip implant. The information provided outlines the performance data rather than the acceptance criteria for an AI/ML device. Therefore, it is important to note that the following analysis is based on the provided text, which details the performance testing conducted for a medical device (hip implant components), not an AI/ML algorithm.
Given the context of a hip implant system, the "acceptance criteria" are typically related to the physical and mechanical properties of the device, ensuring it meets established international standards for safety and performance in the human body. The "study" refers to the mechanical and material testing performed on the device to demonstrate its substantial equivalence to previously cleared predicate devices.
Here's an interpretation based on the provided document:
Acceptance Criteria and Device Performance (Applied to a Medical Device)
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Bencox Mirabo PE Insert: | |
Conformity to ISO 14242 (Wear) | Performed comparable to predicate devices. |
Conformity to ASTM F1820 (Liner Torsion & Lever Out) | Performed comparable to predicate devices. |
Conformity to ASTM F2582 (Push Out and Impingement) | Performed comparable to predicate devices. |
Material Characterization (ASTM 2565) | Obtained values of various parameters were comparable to predicate devices. |
Bencox ID Stem (Offset): | |
Conformity to ISO 7204-4 (Fatigue) | Satisfied the standards and performed similar to predicate devices. |
Conformity to ISO 7204-6 (Range of Motion) | Satisfied the standards and performed similar to predicate devices. |
Bencox Delta Heads XL: | |
Reference to predicate device submission, K121665 and CeramTec Device Master File (Implies meeting the established standards for these components) | Performance refers to predicate device submission K121665 and CeramTec Device Master File. |
Overall: No new issues of safety or efficacy from technological differences | Differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy. |
Regarding the specific questions intended for AI/ML device evaluation, the provided document does not contain information for most of these points, as it describes a physical medical device. However, I will address them to the best of my ability based on the information not present or by inferring the closest equivalent for a physical device.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated in terms of "test set" samples as one would for an AI/ML model. For physical devices, "sample size" typically refers to the number of units tested. The document mentions "worst case combination" for the Bencox ID Stem, implying specific device configurations were selected for testing. The number of individual components or assemblies tested to meet standard requirements is not detailed.
- Data Provenance: Not applicable in the context of an AI/ML model's data. The testing was physical and performed to international standards. The manufacturer is Corentec Co., Ltd. from South Korea.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. "Ground truth" in this context would be the objective criteria defined by the ISO and ASTM standards and the measurements taken during mechanical and material testing. No human experts are used to "establish ground truth" in the way they would for labeling in an AI/ML diagnostic task.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. This concept pertains to resolving discrepancies in expert labels for AI/ML training/testing data. For medical device testing, adjudication is typically not performed in this manner; results are objectively measured against established standard thresholds.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is a framework for evaluating diagnostic AI/ML systems. No human readers or AI assistance are involved in the performance testing of a hip implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This refers to the performance of an AI algorithm alone. The "performance" described here is for a physical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For a physical medical device, the "ground truth" is defined by the objective, measurable criteria within the referenced ISO and ASTM international standards. These standards specify test methods (e.g., how to measure wear, fatigue strength, material properties) and acceptable performance limits. The device's performance is then compared directly to these limits.
8. The sample size for the training set
- Not applicable. There is no "training set" in the context of developing and testing this type of physical medical device. The device itself is manufactured, and specific samples are then tested for performance.
9. How the ground truth for the training set was established
- Not applicable. As there is no "training set" for an AI/ML model, there is no ground truth established for one. The "ground truth" for the device's performance is the adherence to the published standards as mentioned in point 7.
Ask a specific question about this device
(28 days)
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.
The patient should be skeletally mature.
The patient's condition should be due to one or more of:
· Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or
psoriactic arthritis, Congenital hip dysplasia, Ankylosing spondylitis.
· Avascular necrosis of the femoral head.
· Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
The Versafitcup CC Trio family of acetabular components is designed to be used with the Medacta Total Hip Prosthesis System. The Medacta Total Hip Prosthesis system includes the Quadra S, H, R, and C Stems and CoCrMo and ceramic ball heads (K072857, K073337, K080885, K082792, K083558, and K112115). The AMIStem femoral stems also work with the Medacta Total Hip Prosthesis System (K093944, K103189). The Medacta Total Hip Prosthesis System is a total hip replacement system consisting of the femoral stem made of metal, a modular femoral head made of metal or ceramic, and acetabular components. The Versafitcup CC Trio extension that are the subject of this 510(k) consist of new sizes of flat and hooded liners, new sizes of the two-hole acetabular shell, and no-hole acetabular shells. The liners are made from either ultra-high molecular weight polyethylene (UHMWPE) or HighCross® highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE) conforming to ISO 5834. The acetabular shells are made from titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3. The outside of the metal component has macrostructures in the equatorial region. The outer surface of the metallic cup has a dual laver of coatings: Ti (ASTM F 1580) plasma spray and Hydroxyapatite (ASTM F 1185).
All the Versafitcup CC Trio extension components are supplied sterile in single-use individual packages.
The provided text describes a 510(k) premarket notification for a medical device, the Versafitcup CC Trio extension, which is an acetabular component for total hip arthroplasty. The submission focuses on demonstrating substantial equivalence to predicate devices rather than a standalone clinical study with acceptance criteria in the typical sense of a diagnostic or AI device.
Therefore, the information requested for AI/diagnostic devices, such as sample size for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment, is not applicable or available in this document.
Instead, the performance testing described is focused on mechanical characteristics, comparing the new device components to "worst case liners and acetabular shells of the predicate devices."
Here's an attempt to extract the relevant information based on the provided text, while acknowledging the limitations:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Reported Device Performance |
---|---|
Range of Motion | Compared to predicate, found to be acceptable. |
Instability of connection between liner and acetabular shell | Compared to predicate, found to be acceptable. |
Wear | Compared to predicate, found to be acceptable. |
Safety & Effectiveness | "does not introduce any new issues in regards to safety and effectiveness." |
Note: The document does not provide specific numerical acceptance criteria or performance metrics (e.g., maximum allowable wear per year, specific range of motion values). The acceptance is implicitly based on being comparable to or not worse than the predicate devices' "worst case" performance for these mechanical tests.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not specified. The testing is described as mechanical testing of the device components, not involving human subjects or clinical data in the form of a test set.
- Data Provenance: Not applicable in the context of clinical data provenance. The testing is laboratory-based mechanical testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. The "ground truth" here is based on engineering specifications and mechanical test results, not expert interpretation of clinical data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This refers to adjudication of expert opinions for clinical ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This type of study is for diagnostic or AI-assisted interpretation, which is outside the scope of this device's submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is an implanted medical device, not a diagnostic algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- The "ground truth" or basis for evaluation is compliance with mechanical testing standards and comparison to the performance of predicate devices for parameters like range of motion, connection stability, and wear. These are engineering and material science metrics.
8. The sample size for the training set
- Not applicable. This refers to training data for an AI algorithm.
9. How the ground truth for the training set was established
- Not applicable. This refers to establishing ground truth for AI training data.
Ask a specific question about this device
(56 days)
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.
The patient should be skeletally mature.
The patient's condition should be due to one or more of:
· Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis. rheumatoid arthritis or
psoriactic arthritis. Congenital hip dysplasia, Ankylosing spondylitis.
· Avascular necrosis of the femoral head.
· Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
The Versafitcup® CC Trio family of acetabular components is designed to be used with the Medacta Total Hip Prosthesis System. The Medacta Total Hip Prosthesis system includes the Quadra S, H, R, and C Stems and CoCrMo and ceramic ball heads (K072857, K073337, K080885, K082792, K083558, and K112115). The AMStem femoral stems also work with the Medacta Total Hip Prosthesis System (K093944, K103189). The Medacta Total Hip Prosthesis System is a total hip replacement system consisting of the femoral stem made of metal, a modular femoral head made of metal or ceramic, and acetabular components. The Versafitcup® CC Trio acetabular components that are the subject of this 510(k) consist of two new flat fixed liners that are made of HighCross® highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE).
All the Versafitcup® CC Trio components are supplied sterile in single-use individual packages.
The provided document is a 510(k) summary for a medical device (Versafitcup CC Trio - Additional Liners) and, as such, it focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study report with specific acceptance criteria and performance metrics typically found in clinical trials for new drug or high-risk device approvals.
Therefore, many of the requested elements (e.g., sample size for test sets and training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone AI performance) are not applicable or not provided in this type of regulatory submission. The device is a physical medical device (hip implant liners), not a software/AI device, which explains why many of these questions about AI performance and expert review are not relevant to this document.
Here's a breakdown of the available information:
Acceptance Criteria and Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical Tests applicable to hip implant liners: | The Versafitcup CC Trio - Additional Liners were compared to the worst-case liners of the predicate devices in regards to mechanical tests. |
- Range of Motion | The liners met the performance requirements for range of motion, as they were compared to the worst-case predicate devices and determined to introduce no new issues. |
- Instability of connection between liner and acetabular shell | The liners met the performance requirements for connection stability, as they were compared to the worst-case predicate devices and determined to introduce no new issues. |
- Wear | The liners met the performance requirements for wear, as they were compared to the worst-case predicate devices and determined to introduce no new issues. |
Safety and Effectiveness | The manufacturer concluded that "the Versafitcup CC Trio - Additional Liners do not introduce any new issues in regards to safety and effectiveness" compared to the predicate devices. This implies they met the equivalent safety and effectiveness standards demonstrated by the predicate devices. |
Study Details (Based on available information)
- Sample size used for the test set and the data provenance: Not applicable. This is a physical device. Performance was assessed through mechanical testing, not a test set of data. The document refers to "worst-case liners of the predicate devices" for comparison.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's mechanical properties is established through standardized engineering tests, not expert consensus in the clinical sense.
- Adjudication method for the test set: Not applicable. Mechanical tests for physical devices do not involve clinical adjudication.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical hip implant liner, not an AI diagnostic or assistive device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical hip implant liner, not a software device.
- The type of ground truth used: Mechanical test results, compared against established standards or predicate device performance.
- The sample size for the training set: Not applicable. This is a physical device, not an AI model.
- How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
Page 1 of 1