AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MicroPort total hip systems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, 1. ankylosis, protrusio acetabuli, and painful hip dysplasia;
  • inflammatory degenerative joint disease such as rheumatoid arthritis; 2.
  • correction of functional deformity; and, 3.
  • revision procedures where other treatments or devices have failed 4.

Modular shells with porous metal bead coating may be used in either cemented or uncemented arthroplasty. Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.

Device Description

The DYNASTY® Acetabular System with Ceramic contains femoral heads in size 28mm manufactured from alumina matrix composite, mating acetabular liners manufactured from crosslinked polyethylene, and beaded acetabular shells manufactured from titanium alloy.

AI/ML Overview

The provided 510(k) summary for the DYNASTY® Acetabular System with Ceramic does not contain information related to acceptance criteria for a device, nor does it describe a study that proves the device meets specific acceptance criteria in the manner typically associated with AI/software devices.

Instead, this document describes a traditional medical device (an orthopedic implant) and focuses on demonstrating substantial equivalence to previously cleared predicate devices. The "study" mentioned is purely for mechanical and tribological (wear) evaluation, not clinical or performance criteria as would be relevant for software or AI.

Therefore, I cannot extract the requested information in the format provided because it's not present in the document.

Here's why and what the document does provide:

  • Type of Device: The DYNASTY® Acetabular System with Ceramic is a physical orthopedic implant (acetabular shell, liner, and femoral head for total hip replacement). It is not a software device or an AI algorithm.
  • Regulatory Pathway: This is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as for a PMA.
  • "Acceptance Criteria" in this context: For a physical implant like this, "acceptance criteria" relate to engineering specifications, material properties, biocompatibility, mechanical strength, wear resistance, and manufacturing quality. These are typically assessed through bench testing and adherence to recognized standards.
  • "Study" in this context: The "study" mentioned is non-clinical mechanical and tribological testing.

Specific answers based on the provided text, explaining why other points are not applicable:

  1. A table of acceptance criteria and the reported device performance:

    • Not Applicable. The document does not provide a table of acceptance criteria or performance metrics in the way one would for a software or AI device (e.g., sensitivity, specificity, AUC). It discusses mechanical and tribological evaluations and comparisons to predicate devices rather than specific performance targets against clinical gold standards.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not Applicable. This refers to mechanical testing, not a clinical study involving a test set of data. The "subject size 28mm ceramic head was used as a worst case" for burst, post-fatigue burst, rotational stability, and pull-off force testing, implying a limited number of physical samples (likely one or a few) were subjected to these mechanical tests. No patient data or clinical retrospective/prospective studies were conducted.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. This device is not a diagnostic or AI tool requiring expert ground truth for a test set.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. This is not a clinical study requiring adjudication of expert opinions.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a physical medical device, not an AI system.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device, not an AI algorithm.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not Applicable. For mechanical testing, the "ground truth" would be the physical properties derived from standardized test methods (e.g., force measured at failure, wear volume). The document states:
      • "The subject DYNASTY® Acetabular System with Ceramic was evaluated mechanically and tribologically."
      • "The mechanical testing on the subject and predicate devices was performed on wrought Cobalt Chrome modular neck spigots cleared under K091423 and K100866."
      • "The subject size 28mm ceramic head was used as a worst case in burst, post-fatigue burst, rotational stability, and pull off force testing."
      • "Wear testing compared the subject size 28mm femoral heads to the K130376 DYNASTY® Acetabular System."
      • "Range of motion evaluation compared the subject system to the predicate K052026 in three planes."
      • This implies physical engineering measurements as the "ground truth" for evaluating the device's mechanical characteristics.
  8. The sample size for the training set:

    • Not Applicable. This is not an AI/ML device requiring a training set.
  9. How the ground truth for the training set was established:

    • Not Applicable. This is not an AI/ML device requiring a training set.

In summary, the provided document describes a "Traditional 510(k)" pathway for a physical orthopedic implant. The demonstration of safety and effectiveness relies on substantial equivalence to predicate devices and non-clinical bench testing, not clinical studies or performance criteria typically associated with software/AI medical devices.

{0}------------------------------------------------

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the use of the DYNASTY® Acetabular System with Ceramic.

Submitted By:MicroPort Orthopedics Inc.5677 Airline Rd, Arlington TN, 38002(866) 872-0211
Date:April 4, 2014
Contact Person:Matt PaulProject Manager, Regulatory Affairs
Proprietary Name:DYNASTY® Acetabular System with Ceramic
Common Name:Acetabular Shell, Acetabular Liner, Femoral Head
Classification Name and Reference:888.3353 prosthesis, hip, semi-constrained,metal/ceramic/polymer, cemented or non-porous,uncemented Class II
Subject Product Code and Panel Code:Orthopedics/87/LZO
Predicate Devices:DYNASTY® Acetabular System with Ceramic510(k): K130376Shell: K002149; Liner: K052026; Head: K130376

DEVICE INFORMATION

A. Intended Use

MicroPort total hip systems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, 1. ankylosis, protrusio acetabuli, and painful hip dysplasia;
  • inflammatory degenerative joint disease such as rheumatoid arthritis; 2.
  • correction of functional deformity; and, 3.
  • revision procedures where other treatments or devices have failed 4.

Modular shells with porous metal bead coating may be used in either cemented or uncemented arthroplasty. Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.

B. Device Description

The DYNASTY® Acetabular System with Ceramic contains femoral heads in size 28mm manufactured from alumina matrix composite, mating acetabular liners manufactured from crosslinked polyethylene, and beaded acetabular shells manufactured from titanium alloy.

{1}------------------------------------------------

K140043

C. Substantial Equivalence Information

The subject DYNASTY® Acetabular System with Ceramic is indicated to be paired with the following existing acetabular components summarized in Table 1. These shell types were cleared in K 130376 for use with larger diameter BIOLOX® DELTA ceramic-polyethylene bearings.

Table I. Compatible Shells, Including 510(k) Information. All shells were cleared for use with Biolox Delta heads and polyethylene liners in K130376.

510(k)Device Name
K122382DYNASTY® 10 Hole Revision Shells
K122382DYNASTY® BIOFOAM® 3-Hole Shells
K122382DYNASTY® BIOFOAM® Solid Shells
K122382DYNASTY® BIOFOAM® 5 Hole Shells
K082924DYNASTY® BIOFOAM® Shells

The subject DYNASTY® Acetabular System with Ceramic is indicated to be paired with the following femoral components summarized in Table 2.

Table 2. Compatible Femoral Components, Including 510(k) Information. All femoral stems were cleared for
use with Biolox Delta heads in K130376, except K130984.
-510(k)Device Name
K003016PRO-FEMUR R
K012091PRO-FEMUR
K021346STEM HIP REPLACEMENT SYSTEM
K041114PROFEMUR TAPERED HIP STEM
K041586PROFEMUR S HIP STEM
K051995PROFEMUR RENAISSANCE HIP STEM
K052915PROFEMUR XTR HIP STEM
K053588PROFEMUR LX HIP STEM
K060358PROFEMUR TL HIP STEM
K080663PROFEMUR LX REVISION 5/8 COATED HIP STEM
K081090PROFEMUR LX 5/8 COATED HIP STEM
K091423PROFEMUR HIP SYSTEM MODULAR NECKS
K100866
K110399GLADIATOR PLASMA CLASSIC HIP STEM
K111698PROFEMUR(R) E CEMENTLESS HIP STEM
K111699PROFEMUR(R) Z TITANIUM PLASMA SPRAYED HIP STEM
K111910GLADIATOR HIP STEM
K112080PRESERVE HIP STEM
K112150PROFEMUR GLADIATOR HA HIP STEM
K121221PROFEMUR Z REVISION HIP STEM
K123434PROFEMUR Z CLASSIC STEM
K123688PROFEMUR TL CLASSIC STEM
K130984PROFEMUR RENAISSANCE CLASSIC STEM

The design features of the subject femoral head devices are substantially equivalent to those of the predicate DYNASTY® Acetabular system devices cleared under K130376. The design features of the subject acetabular liner and shell devices are substantially equivalent to those of the predicate LINEAGE® Acetabular system devices cleared under K052026 and K002149. The indications of the subject device are identical to the predicate. Specific warnings are added in the package insert. The fundamental scientific technology of the modified devices has not changed relative to the predicate

Tab 008 - Page 2 of 3

{2}------------------------------------------------

K140043 DYNASTY® Acetabular System with Ceramic Traditional 510(k)

Tab: 510(k) Summary of Safety and Effectiveness

devices. Validation of sterilization residuals by ethylene oxide was conducted on devices cleared in K893685 and K002149; this validation was submitted in K130376, is applicable to the subject devices and is included again in this notification. The safety and effectiveness of the DYNASTY® Acetabular System with Ceramic is adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.

D. Nonclinical Testing

The subject DYNASTY® Acetabular System with Ceramic was evaluated mechanically and The mechanical testing on the subject and predicate devices was performed on tribologically. wrought Cobalt Chrome modular neck spigots cleared under K091423 and K100866. The subject size 28mm ceramic head was used as a worst case in burst, post-fatigue burst, rotational stability, and pull off force testing. Wear testing compared the subject size 28mm femoral heads to the K130376 DYNASTY® Acetabular System. Range of motion evaluation compared the subject system to the predicate K052026 in three planes.

E. Clinical Testing

Clinical data was not provided for the subject devices.

F. Conclusion

The design features of the subject devices are substantially equivalent to the predicate devices. The instrument list and materials remain identical to those cleared under K122382 and K130376. The safety and effectiveness of the DYNASTY® Acetabular System with Ceramic is adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 3, 2014

MicroPort Orthopedics, Inc. Max Mortensen, Ph.D. VP Global Quality Systems, Regulatory and Clinical Affairs, Reimbursement and Market Access 5677 Airline Road Arlington, Tennessee 38002

Re: K140043

Trade/Device Name: Dynasty Acetabular System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO Dated: December 17, 2013 Received: January 8, 2014

Dear Dr. Mortensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{4}------------------------------------------------

Page 2 - Max Mortensen, Ph.D.

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Lori A. Wiggins

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

Indications for Use

510(k) Number (if known)

K140043

Device Name Dynasty Acetabular System

Indications for Use (Describe)

The DYNASTY® Accabular System is intended for use in total hip arthroplasty for reduction or relief of pain and or immoved hip function in skeletally mature patients.

  1. non-inflammatory degenerative joint disease such as osteoartinitis, avascular necrosis, ankylosis, protrusio acceabuli, and painful hip dysplasja:

  2. inflammatory degenerative joint disease such as rheumatoid arthritis;

  3. correction of functional deformity; and,

  4. revision procedures where other treatments or devices have failed

Shells with BIOFOAM® metal four costing are intented arthroplasty. Modular shells with porcus metal bead coating may be used in either cemented or uncemented arthroplasty.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

ices (30 )3 441

Please do not write below this line - continue on a separate page if needed.

CONSULTION CONTROLLER FOR FORIEDAUSE ONLY CONCELLER CONCELLERS CONCELLENT Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Elizabeth | @grank -S

Division of Orthopedic Devices

FORM FDA 3881 (9/13)

§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.

(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.