Search Results
Found 3 results
510(k) Data Aggregation
(13 days)
The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging". The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.
The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal Intraoperative, Intraoperative, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications. The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.
The Diagnostic Ultrasound Systems are multi-purpose mobile, software controlled diagnostic ultrasound systems with on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, 3D Imaging, or Harmonic Imaging and 4D imaging.
Here's a summary of the acceptance criteria and study information based on the provided text, focusing on the request details:
Acceptance Criteria and Device Performance
The document describes the Siemens SC2000/X300 Diagnostic Ultrasound System and its various transducers. The acceptance criteria essentially focus on demonstrating substantial equivalence to previously cleared devices. The "reported device performance" is implicitly that the device performs equivalently to the predicate devices and meets safety and performance standards.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria / Standard Demonstrated | Reported Device Performance (as implied by the document) |
---|---|---|
Substantial Equivalence | Equivalence in Intended Use | Demonstrated against K132654 (SC2000), K121699 (X300), K123001 (X700), K141846 (X700), K140318 (SSe8F), K071234 (AN8F) |
Equivalence in Technological Characteristics | Demonstrated against predicate devices | |
Device Safety | Hardware Safety - EN60601-1 Certified | Conforms |
Acoustic Thermal Safety - IEC 60601-2-37 Certified | Conforms | |
Medical device software - Software Life Cycle Process - IEC 62304 | Conforms | |
UL 60601-1, Safety Requirements for Medical Equipment | Complies | |
CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment | Complies | |
AIUM/NEMA UD-3 | Complies | |
AIUM/NEMA UD-2 | Complies | |
93/42/EEC Medical Devices Directive | Complies | |
EN/IEC 60601-1, EN/IEC 60601-1-1, EN/IEC 60601-1-2, EN/IEC 62304, EN/IEC 62366, EN/IEC 60601-2-18, EN/IEC 60601-2-25 | Complies | |
Biocompatibility | ISO 10993-1 Biocompatibility | Cleared patient contact materials, electrical and mechanical safety are unchanged. |
Cleaning & Disinfection | Cleaning and disinfection effectiveness | Evaluated and found to conform |
Quality System | 21 CFR 820 Quality System Regulation & ISO 13485:2003 | Conforms |
2. Sample size used for the test set and the data provenance
The document explicitly states: "Since the modified devices use the same technology and principles as existing devices, clinical data is not required." This indicates that no clinical test set with human patients was used specifically for this 510(k) submission. The evaluation was based on nonclinical tests and comparison to predicate devices. Therefore, there is no sample size, country of origin, or information on retrospective/prospective data for a clinical test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
As there was no clinical test set for this 510(k) submission, there is no mention of experts used to establish ground truth for a test set. The review focused on engineering and performance standards, and comparison to previously cleared devices.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable, as no clinical test set was used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The document describes a diagnostic ultrasound system and its transducers, not an AI-assisted diagnostic tool. No MRMC study or AI performance evaluation is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a diagnostic ultrasound system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the nonclinical tests conducted, the "ground truth" would be established by the specifications and performance metrics outlined in the various referenced safety and performance standards (e.g., UL 60601-1, IEC 60601-2-37, etc.). For the substantial equivalence claim, the ground truth is the performance and safety profiles of the predicate devices.
8. The sample size for the training set
Not applicable, as this is a traditional diagnostic ultrasound system, not an AI/ML-based device requiring a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set was used.
Ask a specific question about this device
(6 days)
The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications.
The Acuson SC2000™ has been designed to meet the following product safety standards: UL 60601-1, Safety Requirements for Medical Equipment; IEC 60601-2-37 Diagnostic Ultrasound Safety Standards; CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment; AIUM/NEMA UD-3. Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment; AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound; Safety and EMC Requirements for Medical Equipment; IEC 60601-1; IEC 60601-1-1; IEC 60601-1-2; IEC 1157 Declaration of Acoustic Power; ISO 10993-1 Biocompatibility.
This 510(k) summary for the ACUSON SC2000 Diagnostic Ultrasound System is a Special 510(k), indicating a minor change to a previously cleared device. Therefore, a comprehensive de novo clinical study demonstrating performance against acceptance criteria for a novel device is not part of this submission. Instead, the performance data section explicitly states: "No performance data required to support this modification, as there is no change to any hardware or software, or method of use."
The submission focuses on demonstrating substantial equivalence to a previously cleared device (K123622) by only renaming three transducers. As such, the typical elements of a study proving a device meets acceptance criteria (like a test set, ground truth, experts, and statistical results) are not provided in this document.
However, I can extract the acceptance criteria implied by this type of submission, which is the determination of substantial equivalence to the predicate device for the specified indications for use.
Here's a breakdown based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied for a Special 510(k) for Renaming Transducers): The modified device, specifically the renamed transducers, must demonstrate substantial equivalence to the previously cleared predicate device concerning intended use and technological characteristics, with no changes to hardware, software, or method of use.
Acceptance Criterion | Reported Device Performance |
---|---|
Substantial Equivalence to Predicate | The submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics. This modification is intended only to rename three transducers. |
No Change in Hardware/Software/Method of Use | Confirmed: "No performance data required to support this modification, as there is no change to any hardware or software, or method of use." |
Continued Adherence to Safety Standards | The Acuson SC2000™ has been designed to meet the following product safety standards: UL 60601-1, IEC 60601-2-37, CSA C22.2 No. 601-1, AIUM/NEMA UD-3, AIUM/NEMA UD-2, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 1157, ISO 10993-1. |
Intended Use Remaining Consistent | The Indications for Use for the overall system and the individual transducers remain largely consistent with previously cleared indications. (P = Previously Cleared) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable. No new performance testing with a specific test set was conducted for this Special 510(k) given the nature of the modification (renaming transducers).
- Data Provenance: Not applicable. No new data was generated or analyzed for this submission. The basis for substantial equivalence relies on the established performance of the predicate device and the claim that the changes are confined to nomenclature.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable. No new ground truth establishment was required as no new performance data was generated.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No. This submission does not describe an MRMC comparative effectiveness study.
- Effect Size of Human Readers Improvement: Not applicable.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: Not applicable. This device is an ultrasound imaging system, not an AI algorithm requiring standalone performance evaluation in the context of this submission.
- The document states there is "no change to any hardware or software," implying no new algorithms or software features are being introduced that would require such testing.
7. The Type of Ground Truth Used
- Type of Ground Truth: Not applicable for new performance evaluation. The "ground truth" for this Special 510(k) is implicitly the established safety and effectiveness of the predicate device (K123622), to which the modified device (with renamed transducers) is claimed to be substantially equivalent without any functional changes.
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This document does not describe the development or training of any new algorithms or software components.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth Establishment for Training Set: Not applicable.
Ask a specific question about this device
(20 days)
V5Ms trans-esophageal echocardiograph (TEE) ultrasound transducer is intended primarily for cardiology applications.
ACUSON Antares Ultrasound System: The Acuson Antares ultrasound imaging system is intended for the following applications: Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Intracardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Intravascular, Musculoskeletal, Superficial Musculoskeletal. Great Vessel, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
ACUSON CV70 Cardiovascular System: The CV70 ultrasound imaging system is intended for the following applications: Abdominal. Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Intracardiac, Transesophageal, Pelvic, Neonatal/Adult Gephalic, Vascular, Intravascular, Musculoskeletal, Superficial Musculoskeletal, Great Vessel, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
ACUSON Cypress Ultrasound System: The Cypress platform is intended for use in the following applications: General Imaging and Cardiology for Fetal. Abdominal. Intraoperative (cardiac), Pediatrics, Neonatal Cephalic, Cardiac (adult, pediatric), Trans-esophageal, Peripheral Vessel, Intraluminal and Intra-cardiac applications, and intended uses as defined in the FDA quidance document. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes
ACUSON S1000/S2000 Ultrasound System: The S2000™ ultrasound imaging systems are intended for the following applications: Fetal. Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal(conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging."
ACUSON SC2000 Ultrasound System: The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications.
ACUSON Sequoia Ultrasound System: The Siemens ACUSON Sequoia ultrasound imaging system is intended for the following applications: General Imaging and Cardiology for Fetal, Abdominal, Intraoperative (abdominal and neurological), Pediatrics, Small Organs (breast, testes, thyroid and penis), Neonatal/Adult Cephalic, Cardiac (adult, pediatric, and neonatal), Trans-esophageal, Transrectal, Transvaginal, Peripheral Vessel, and Musculo-sceletal (superficial and conventional) applications, and intended uses as defined in the FDA quidance document. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
ACUSON X300 Ultrasound System: The Siemens ACUSON X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides measurement of anatomical structures and analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients reqarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement: Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".
The V5Ms Transesophageal Transducer consists of a gastroscope control housing where nosepiece articulation and transducer rotation are controlled. A flexible transesophageal guide tube with a nosepiece containing the acoustic array extends from one end of the control housing and the system cable/connector extends from the other end. The acoustic array has 64 elements and rotates 180 degrees to provide imaging planes from transverse view to inverse transverse view. Rotation is powered by a motor in the control housing and is controlled by the operator using a switch button on the control housing for clockwise and counterclockwise rotation. Nosepiece articulation is achieved by manipulating a vertebrae section adjacent to the nosepiece through a series of control wires attached to knobs on the control housing. The nosepiece can be articulated in both the anterior/posterior and left/right directions.
Here's a breakdown of the requested information, based on the provided document:
Acceptance Criteria and Study Details for V5Ms Transesophageal Transducer
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) submission document does not explicitly state quantitative acceptance criteria or specific device performance metrics in the way one might expect for a new or significantly modified device. Instead, the submission relies heavily on substantial equivalence to a previously cleared predicate device (Siemens V5Ms Transesophageal Transducer, K052021). The key "performance data" presented is that "Performance data is not required for this modification as there is no change to software or hardware." This implies that the device is expected to perform identically to its predicate because the changes are limited to patient contact materials.
Therefore, the acceptance criteria are implicitly met by demonstrating that the modifications do not negatively impact the established performance of the predicate device.
Implicit Acceptance Criteria and Reported Performance (Based on Substantial Equivalence):
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
Safety and Effectiveness | The modified device is safe and effective for its intended uses, as demonstrated by its substantial equivalence to the predicate device (Siemens V5Ms Transesophageal Transducer, K052021). No changes to software or hardware, thus performance is maintained. |
Biocompatibility | Meets ISO10993-1 standards for cytotoxicity, sensitization, and irritation. |
Intended Use (across various ultrasound systems) | The device is intended for the same cardiology applications and clinical applications (e.g., Transesophageal, Cardiac, various Doppler modes, 3D imaging, Harmonic Imaging) as the predicate when used with compatible ACUSON systems. The intended uses are identical to the predicate device. |
Technological Characteristics | The device possesses the same fundamental technological characteristics (e.g., phased array transducer, 64 elements, 180-degree rotation, nosepiece articulation) as the predicate, with modifications only to patient contact materials. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable (N/A). The document explicitly states, "Performance data is not required for this modification as there is no change to software or hardware." Therefore, there was no clinical performance test set used. The evaluation centered on non-clinical tests for biocompatibility and asserting substantial equivalence.
- Data Provenance: N/A for clinical performance data. For biocompatibility testing, the data would typically originate from a laboratory (likely in the USA or a facility that adheres to international standards like ISO). The origin is not specified beyond "Nonclinical tests contained in this submission include biocompatibility testing per ISO10993-1 (cytotoxicity, sensitization, irritation)."
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- N/A. As no clinical performance test set was conducted, there was no need for experts to establish ground truth. The primary "experts" involved in the submission would be Siemens' own regulatory and engineering teams, and subsequently, FDA reviewers.
4. Adjudication Method for the Test Set
- N/A. Since no clinical performance test set was conducted, no adjudication method was used.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. No MRMC study was conducted. The submission is a 510(k) for substantial equivalence based on minor modifications (patient contact materials) to an existing device, not a new clinical effectiveness study.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- N/A. The device is a diagnostic ultrasound transducer, not an AI algorithm. Therefore, the concept of "standalone performance" in the context of an algorithm does not apply.
7. The Type of Ground Truth Used
- N/A for clinical effectiveness. For biocompatibility, the "ground truth" is defined by the ISO10993-1 standard for cytotoxicity, sensitization, and irritation. This involves standardized biological tests.
8. The Sample Size for the Training Set
- N/A. This device is a hardware component (transducer) and does not involve machine learning or AI, and therefore does not have a "training set."
9. How the Ground Truth for the Training Set Was Established
- N/A. As there is no training set for this device, this question is not applicable.
Ask a specific question about this device