K Number
K142395

Validate with FDA (Live)

Date Cleared
2014-09-09

(13 days)

Product Code
Regulation Number
892.1560
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging". The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal Intraoperative, Intraoperative, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications. The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

Device Description

The Diagnostic Ultrasound Systems are multi-purpose mobile, software controlled diagnostic ultrasound systems with on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, 3D Imaging, or Harmonic Imaging and 4D imaging.

AI/ML Overview

Here's a summary of the acceptance criteria and study information based on the provided text, focusing on the request details:

Acceptance Criteria and Device Performance

The document describes the Siemens SC2000/X300 Diagnostic Ultrasound System and its various transducers. The acceptance criteria essentially focus on demonstrating substantial equivalence to previously cleared devices. The "reported device performance" is implicitly that the device performs equivalently to the predicate devices and meets safety and performance standards.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria / Standard DemonstratedReported Device Performance (as implied by the document)
Substantial EquivalenceEquivalence in Intended UseDemonstrated against K132654 (SC2000), K121699 (X300), K123001 (X700), K141846 (X700), K140318 (SSe8F), K071234 (AN8F)
Equivalence in Technological CharacteristicsDemonstrated against predicate devices
Device SafetyHardware Safety - EN60601-1 CertifiedConforms
Acoustic Thermal Safety - IEC 60601-2-37 CertifiedConforms
Medical device software - Software Life Cycle Process - IEC 62304Conforms
UL 60601-1, Safety Requirements for Medical EquipmentComplies
CSA C22.2 No. 601-1, Safety Requirements for Medical EquipmentComplies
AIUM/NEMA UD-3Complies
AIUM/NEMA UD-2Complies
93/42/EEC Medical Devices DirectiveComplies
EN/IEC 60601-1, EN/IEC 60601-1-1, EN/IEC 60601-1-2, EN/IEC 62304, EN/IEC 62366, EN/IEC 60601-2-18, EN/IEC 60601-2-25Complies
BiocompatibilityISO 10993-1 BiocompatibilityCleared patient contact materials, electrical and mechanical safety are unchanged.
Cleaning & DisinfectionCleaning and disinfection effectivenessEvaluated and found to conform
Quality System21 CFR 820 Quality System Regulation & ISO 13485:2003Conforms

2. Sample size used for the test set and the data provenance

The document explicitly states: "Since the modified devices use the same technology and principles as existing devices, clinical data is not required." This indicates that no clinical test set with human patients was used specifically for this 510(k) submission. The evaluation was based on nonclinical tests and comparison to predicate devices. Therefore, there is no sample size, country of origin, or information on retrospective/prospective data for a clinical test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

As there was no clinical test set for this 510(k) submission, there is no mention of experts used to establish ground truth for a test set. The review focused on engineering and performance standards, and comparison to previously cleared devices.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable, as no clinical test set was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The document describes a diagnostic ultrasound system and its transducers, not an AI-assisted diagnostic tool. No MRMC study or AI performance evaluation is mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a diagnostic ultrasound system, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the nonclinical tests conducted, the "ground truth" would be established by the specifications and performance metrics outlined in the various referenced safety and performance standards (e.g., UL 60601-1, IEC 60601-2-37, etc.). For the substantial equivalence claim, the ground truth is the performance and safety profiles of the predicate devices.

8. The sample size for the training set

Not applicable, as this is a traditional diagnostic ultrasound system, not an AI/ML-based device requiring a training set.

9. How the ground truth for the training set was established

Not applicable, as no training set was used.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 9, 2014

Siemens Medical Solutions, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K142395

Trade/Device Name: SC2000/X300 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, OBJ Dated: August 21, 2014 Received: August 27, 2014

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

This determination of substantial equivalence applies to the following transducers intended for use with the SC2000/X300 Diagnostic Ultrasound System, as described in your premarket notification:

SC2000 Transducer Model Number
AcuNav 8FSoundstar eco 8F9L4
6C1HDV5M4V1c
8V3CW24Z1c
AcuNav 10FACUSON AcuNav™ V 10F SoundStar 10F
V7M10V4
X300 Transducer Model Number
AcuNav 8FSoundstar eco 8FP4-2
CH5-2VF10-5I 9-5

{1}------------------------------------------------

EC9-4EV9-4VF13-5
P8-4BE 9-4CW2
CW5AcuNav 10FV5Ms
4V1cVF13-5SPC8-5
8L310V4C7F2
EV9F4L13F5VF8-3
P5-1C6-2SoundStar 10F
P9-4

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

{2}------------------------------------------------

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Smh.7)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known): K142395

Device Name: ACUSON X300™ Diagnostic Ultrasound System SONOVISTA X300 Diagnostic Ultrasound System ACUSON X300 " Diagnostic Ultrasound System, premium edition

The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

Type of Use (Select one or both, as applicable)

X Prescription Use (21 CFG 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Page 1 of 31

{4}------------------------------------------------

510(k) Number (if known):

Device Name: Intended Use: ACUSON X300 Diagnostic Ultrasound Systems

Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks11& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Neuro)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Laparoscopic
FetalImaging& OtherPediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-rectalPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-vaginalPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-urethral
Trans-esoph.(non-Card.)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Convent.)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra -vascularPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other(Specify)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
CardiacCardiac AdultPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Cardiac PediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-esophageal(Cardiac)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-CardiacPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
PeripheralVesselPeripheral vesselPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)PPPPPPPBMDCNote2,3,4,5,7,8,9,11

N = new indication; P = previously cleared K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaging

Power SieScape panoramic imaging Note 5

Contrast agent imaging Note 7

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE Note 2 Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging
For example: abdominal, vascular Note 4

Note 6

SieClear multi-view spatial compounding Note 8

Note 10 Intracardiac imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

{5}------------------------------------------------

510(k) Number (if known):

AcuNav 8F Intracardiac Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Device Name: Intended Use:

Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks1& 3)BMPWDCWDColorDopplerPowerDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatric
Imaging& OtherSmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascularPPPPPPBMDCNote 2,8,9,13,14,16
Other(Specify)
CardiacCardiac AdultPPPPPPBMDCNote 2,8,9,13,14,16
Cardiac PediatricPPPPPPBMDCNote 2,8,9,13,14,16
Intra-vascular(Cardiac)PPPPPPBMDCNote 2,8,9,13,14,16
Trans-esophageal(Cardiac)
Intra-cardiacPPPPPPBMDCNote 2,8,9,13,14,16
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)
N = new indication; P = previously cleared K080760, 121699

Note 1 state etc F olo: hr st testes thvroid i onic

Note 1For example: breast, testes, thyroid, penis, prostate, etc.Note 11syngo Arterial Health Package (AHP)
Note 2Dynamic TCE TechnologyNote 12syngo Auto OB Measurements
Note 3SieClearNote 13syngo Auto Left Heart (Auto LH) Technology
Note 4Advanced SieClearNote 14syngo Velocity Vector Imaging Technology
Note 53-Scape 3D ImagingNote 15CartoSound Communication
Note 6For example: abdominal, vascularNote 16Intracardiac Echocardiography (ICE) Imaging
Note 7Stress Echo ImagingNote 17syngo fourSight TEE View
Note 8Axius Edge Assisted Ejection FractionNote 18syngo Mitral Valve Assessment (MVA)
Note 9Clarify Vascular Enhancement TechnologyNote 19syngo fourSight 4D imaging
Note 10SieScape Panoramic ImagingNote 20Contrast Agent Imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)__

{6}------------------------------------------------

510(k) Number (if known):

Device Name:Soundstar eco 8F Intracardiac Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other( Track1Only)Specific(Tracks1& 3)BMPWDCWDColorDopplerPowerDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatric
SmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascularPPPPPPBMDCNote 2,8,9,13,14,16
Other(Specify)
Cardiac AdultPPPPPPBMDCNote 2,8,9,13,14,16
CardiacCardiac PediatricPPPPPPBMDCNote 2,8,9,13,14,16
Intra-vascular(Cardiac)PPPPPPBMDCNote 2,8,9,13,14,16
Trans-esophageal(Cardiac)
Intra-cardiacPPPPPPBMDCNote 2,8,9,13,14,16
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 2 Dynamic TCE Technology

Note 3 SieClear

Advanced SieClear Note 4 Note 5 3-Scape 3D Imaging

For example: abdominal, vascular Note 6

Note 7 Stress Echo Imaging

Note 8 Axius Edge Assisted Ejection Fraction

Note 9 Clarify Vascular Enhancement Technology

Note 10 SieScape Panoramic Imaging

syngo fourSight TEE View Note 17 Note 18 syngo Mitral Valve Assessment (MVA)

Note 19 syngo fourSinght 4D imaging Note 20 Contrast Agent Imaging

syngo Arterial Health Package (AHP)

syngo Auto Left Heart (Auto LH) Technology

Intracardiac Echocardiography (ICE) Imaging

syngo Velocity Vector Imaging Technology

syngo Auto OB Measurements

CartoSound Communication

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Note 11

Note 12

Note 13

Note 14

Note 15

Note 16

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

{7}------------------------------------------------

510(k) Number (if known):

ACUSON X300 Diagnostic Ultrasound Systems

Device Name: Intended Use:

ACUSON X300 Diagnostic Ultrasound Systems

Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks11& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Neuro)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
LaparoscopicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
FetalImaging& OtherPediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-rectalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-vaginalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-urethralPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-esoph.(non-Card.)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Convent.)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra -vascularPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other(Specify)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Cardiac AdultPPPPPPPBMDCNote2,3,4,5,7,8,9,11
CardiacCardiac PediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-esophageal(Cardiac)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-CardiacPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
PeripheralVesselPeripheral vesselPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)PPPPPPPBMDCNote2,3,4,5,7,8,9,11

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3
Note 5 3D imaging

Power SieScape panoramic imaging Contrast agent imaging Note 7

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

Note 2 Ensemble tissue harmonic imaging Note 4
Note 6 B&W SieScape panoramic imaging
For example: abdominal, vascular

te 6 For example: abdominal, vasc

Note 8 SieClear multi-view spatial compounding

Note 10 Intracardiac imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 5 of 31

{8}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

P4-2 Phased Sector Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPPBMDCNote2,3,5,6,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)
Neonatal CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
CardiacCardiac AdultPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Cardiac PediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

  • Note 3 3D imaging
  • Power SieScape panoramic imaging Note 5
  • Contrast agent imaging
    Tissue Equalization Technology Note 7

Note 9

  • Note 11 Dynamic TCE
    (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

  • Note 8
    Note 10
    For example: abdominal, vascular SieClear multi-view spatial compounding
    Intracardiac imaging

10 Intracardiac imaging

Note 2 Ensemble tissue harmonic imaging Note 4 B&W SieScape panoramic imaging Note 6

{9}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

CH5-2 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatricPPPPPPBMDCNote2,3,4,5,7,8,9,11
Imaging& OtherSmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPBMDCNote2.3.4.5.7.8.9,11
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac AdultPPPPPPBMDCNote2,3,4,5,7,8,9,11
CardiacCardiac PediatricPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vesselPPPPPPBMDCNote2,3,4,5,7,8,9,11
VesselOther (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

  • Note 3 3D imaging
  • Note 5 Power SieScape panoramic imaging

Contrast agent imaging
Tissue Equalization Technology Note 7

Note 9 Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 7 of 31

Note 6 For example: abdominal, vascular Note 8

Note 10

For example: abdominal,
SieClear multi-view spatial

SieClear multi-view spa
a Intracardiac imaging

SieClear multi-view spatial compounding
Intracardiac imaging

Steclear multi-view sp
0 Intracardiac imaging

Note 2 Ensemble tissue harmonic imaging Note 4 B&W SieScape panoramic imaging

{10}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

VF10-5 Linear Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)PPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)PPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular
Other(Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)
N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3

3D imaging
Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)

Page 8 of 31

Note 4 B&W SieScape panoramic imaging Note 6 For example: abdominal, vascular Note 8

Note 2

SieClear multi-view spatial compounding Note 10 Intracardiac imaging

Ensemble tissue harmonic imaging

{11}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

L9-5 Linear Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track 1Only)Specific(Tracks 1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatricPPPPPBMDCNote2,3,4,5,7,8,9,11
Imaging& OtherSmallOrgan(Note 1)PPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)PPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-Cardaic
Other (Specify)
PeripheralPeripheral vesselPPPPPBMDCNote2,3,4,5,7,8,9,11
VesselOther (Specify)

N = new indication; P = previously cleared by K080760, 121699

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

  • Note 3
    3D imaging
    Power SieScape panoramic imaging Note 5

Contrast agent imaging Note 7

Note 9 Tissue Equalization Technology

Dynamic TCE Note 11

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 9 of 31

Note 4 B&W SieScape panoramic imaging Note 6

For example: abdominal, vascular

Ensemble tissue harmonic imaging

Note 8 SieClear multi-view spatial compounding

Note 10 Intracardiac imaging

Note 2

{12}------------------------------------------------

510(k) Number (if known):

Device Name: Intended Use:

EC9-4 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote2,3,4,5,7,8,9,11
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricSmallOrgan(Note 1)PPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectalPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-vaginalPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-Cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)
N = new indication; P = previously cleared by K080760, 121699

N = new indication, F = previously cleared by K060760, 121055

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

  • Note 3 3D imaging
    Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging Note 9 Tissue Equalization Technology

  • Note 11 Dynamic TCE
    (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

  • For example: abdominal, vascular
    SieClear multi-view spatial compounding Intracardiac imaging Note 10
  • Note 2 Note 4 Note 6 Note 8

{13}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

EV9-4 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1 & 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPBMDCNote2,3,4,5,7,8,9,11
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatric
SmallOrgan(Note 1)PPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectalPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-vaginalPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-Cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

  • Note 3 3D imaging
  • Note 5 Power SieScape panoramic imaging
  • Note 7 Contrast agent imaging

Note 9 Tissue Equalization Technology

Dynamic TCE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Note 11

Page 11 OF 31

Intracardiac imaging

Note 2 Ensemble tissue harmonic imaging Note 4 B&W SieScape panoramic imaging

  • Note 6 For example: abdominal, vascular
    SieClear multi-view spatial compounding

Note 8 Note 10

{14}------------------------------------------------

510(k) Number (if known):

Device Name:VF13-5 Linear Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows.
Clinical ApplicationMode of Operation
Other(Track 1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)PPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)PPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)
N = new indication; P = previously cleared by K080760, 121699

For example: breast, testes, thyroid, penis, prostate, etc.
3D imaging Note 1

Note 3

  • Note 5 Power SieScape panoramic imaging
  • Note 2 Note 4 Note 6 Note 8
    Note 10

Ensemble tissue harmonic imaging
B&W SieScape panoramic imaging

Intracardiac imaging

For example: abdominal, vascular

SieClear multi-view spatial compounding

Note 7 Contrast agent imaging Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 12 OF 31

{15}------------------------------------------------

510(k) Number (if known):

Device Name:P8-4 Phased Sector Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other(Track 1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPPBMDCNote2,3,5,6,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)
Neonatal CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
CardiacCardiac AdultPPPPPPPBMDCNote2,3,4,5,7,8,9,11
CardiacCardiac PediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

3D imaging Note 3

Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging

Tissue Equalization Technology Note 9

Note 11 Dynamic TCE Note 2 Ensemble tissue harmonic imaging

Note 4
Note 6 B&W SieScape panoramic imaging For example: abdominal, vascular

Note 8 SieClear multi-view spatial compounding

Intracardiac imaging Note 10

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

{16}------------------------------------------------

510(k) Number (if known):

Device Name:BE 9-4 Convex Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows
Other(Track1Only)Specific(Tracks1& 3)Mode of Operation
ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPBMDCNote 2,3,4,5,7,8,9
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatric
SmallOrgan(Note 1)PPPPPBMDCNote 2,3,4,5,7,8,9
Neonatal CephalicPPPPPBMDCNote 2,3,4,5,7,8,9
Adult Cephalic
Trans-rectalPPPPPBMDCNote 2,3,4,5,7,8,9
Trans-vaginalPPPPPBMDCNote 2,3,4,5,7,8,9
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaging

Power SieScape panoramic imaging Note 5 Note 7

Contrast agent imaging
Tissue Equalization Technology Note 9

Note 11

Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 14 OF 31

Note 6 For example: abdominal, vascular Note 8 SieClear multi-view spatial compounding

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

Intracardiac imaging Note 10

Note 2

Note 4

{17}------------------------------------------------

510(k) Number (if known):

Device Name:CW2 Continuous Wave Doppler Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalP
AbdominalP
Intra-operative(Note 6)P
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricP
SmallOrgan(Note 1)P
Neonatal CephalicP
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)P
Musculo-skel.(Superfic)
Intra-vascularOther
(Specify)
Cardiac AdultP
CardiacCardiac PediatricP
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselP
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3

3D imaging
Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

Note 4B&W SieScape panoramic imaging
Note 6For example: abdominal, vascular

Ensemble tissue harmonic imaging

Note 6 Note 8 SieClear multi-view spatial compounding

Note 2

Note 10 Intracardiac imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 15 OF 31

{18}------------------------------------------------

510(k) Number (if known):

CW5 Continuous Wave Doppler Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Device Name: Intended Use:

Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalP
AbdominalP
FetalImaging& OtherIntra-operative(Note 6)P
Intra-operative(Neuro)
Laparoscopic
PediatricP
SmallOrgan(Note 1)P
Neonatal CephalicP
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)P
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
CardiacCardiac AdultP
Cardiac PediatricP
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-Cardiac
Other (Specify)
PeripheralVesselPeripheral vesselP
Other (Specify)
N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaging

Note 5 Power SieScape panoramic imaging

Note 7 Contrast agent imaging

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 16 OF 31

Ensemble tissue harmonic imaging

Intracardiac imaging

Note 4 B&W SieScape panoramic imaging Note 6 For example: abdominal, vascular Note 8 SieClear multi-view spatial compounding

Note 2

Note 10

{19}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

AcuNav 10F Intracardiac Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems

Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPPPBMDCNote2,3,7,8,9,10,11
SmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascularPPPPPPPBMDCNote2,3,7,8,9,10,11
Other(Specify)
Cardiac AdultPPPPPPPBMDCNote2,3,7,8,9,10,11
CardiacCardiac PediatricPPPPPPPBMDCNote2,3,7,8,9,10,11
Intra-vascular(Cardiac)PPPPPPPBMDCNote2,3,7,8,9,10,11
Trans-esophageal(Cardiac)
Intra-cardiacPPPPPPPBMDCNote2,3,7,8,9,10,11
Other (Specify)
PeripheralPeripheral vessel
VesselOther (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaqing Note 5

Power SieScape panoramic imaging Note 7 Contrast agent imaging

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

((Please do not write below this line-continue on another Page if needed)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Note 2 Ensemble tissue harmonic imaging Note 4 B&W SieScape panoramic imaging

Note 6 For example: abdominal, vascular

Note 8 SieClear multi-view spatial compounding

Note 10 Intracardiac imaging

{20}------------------------------------------------

510(k) Number (if known):

Device Name:V5Ms TEE Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks11& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatric
Imaging& OtherSmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
CardiacCardiac AdultPPPPPPPBMDCNote 2,3,7,8,9,11
Cardiac PediatricPPPPPPPBMDCNote 2,3,7,8,9,11
Trans-esophageal(Cardiac)PPPPPPPBMDCNote 2,3,7,8,9,11
Intra-cardiac
Other (Specify)
PeripheralPeripheral vessel
VesselOther (Specify)

N = new indication; P = previously cleared by K090276

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

  • Note 3
    3D imaging
    Power SieScape panoramic imaging Note 5 Note 7 Contrast agent imaging

  • Note 9 Tissue Equalization Technology
    Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 18 OF 31

B&W SieScape panoramic imaging

SieClear multi-view spatial compounding

Note 2 Ensemble tissue harmonic imaging For example: abdominal, vascular

te 6 For example: abdominal, va

  • 0 Cis-Clear multi-view spatial

Note 10 Intracardiac imaging

Note 4 Note 6 Note 8

{21}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

4V1c Phased Sector Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote 2,3,4,5,7,8,9
AbdominalPPPPPPBMDCNote 2,3,5,6,7,8,9
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatricPPPPPPBMDCNote 2,3,4,5,7,8,9
Imaging& OtherSmallOrgan(Note 1)
Neonatal CephalicPPPPPPBMDCNote 2,3,4,5,7,8,9
Adult CephalicPPPPPPBMDCNote 2,3,4,5,7,8,9
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac AdultPPPPPPBMDCNote 2,3,4,5,7,8,9
CardiacCardiac PediatricPPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vesselPPPPPPBMDCNote 2,3,4,5,7,8,9
VesselOther (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaging

Note 5 Power SieScape panoramic imaging

Note 7 Contrast agent imaging Note 8 SieClear multi-view spatial compounding

Note 9 Tissue Equalization Technology

Intracardiac imaging Note 10

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 19 OF 31

6 For example: abdominal, v

Note 2 Ensemble tissue harmonic imaging

e 4
B&W SieScape panoramic in
e 6
For example: abdominal, vas

te 4 B&W SieScape pane
ote 6 For example: abdom

Note 4 Note 6

B&W SieScape panoramic imaging For example: abdominal, vascular

{22}------------------------------------------------

510(k) Number (if known):

Device Name:VF13-5SP Linear Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(Note 6)PPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Neuro)PPPPPPBMDCNote2,3,4,5,7,8,9.11
Laparoscopic
FetalPediatricPPPPPPBMDCNote2,3,4,5,7,8,9.11
Imaging& OtherSmallOrgan(Note 1)PPPPPPBMDCNote2,3,4,5,7,8,9.11
Neonatal CephalicPPPPPPBMDCNote2,3,4,5,7,8,9.11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPBMDCNote2,3,4,5,7,8,9.11
Musculo-skel.(Superfic)PPPPPPBMDCNote2,3,4,5,7,8,9.11
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vesselPPPPPPBMDCNote2,3,4,5,7,8,9.11
VesselOther (Specify)
N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaging

  • Note 5 Power SieScape panoramic imaging
  • Note 2 Note 4 Note 6 Note 8
    Note 10

Ensemble tissue harmonic imaging
B&W SieScape panoramic imaging

Intracardiac imaging

For example: abdominal, vascular

SieClear multi-view spatial compounding

Note 7 Contrast agent imaging Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 20 OF 31

{23}------------------------------------------------

510(k) Number (if known):

Device Name:C8-5 Tightly Curved Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks11& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote 2,3,4,5,7,8,9
AbdominalPPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatricPPPPPPBMDCNote 2,3,4,5,7,8,9
Imaging& OtherSmallOrgan(Note 1)PPPPPPBMDCNote 2,3,4,5,7,8,9
Neonatal CephalicPPPPPPBMDCNote 2,3,4,5,7,8,9
Adult CephalicPPPPPPBMDCNote 2,3,4,5,7,8,9
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPBMDCNote 2,3,4,5,7,8,9
Musculo-skel.(Superfic)PPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-vascular
Other(Specify)
Cardiac AdultPPPPPPBMDCNote 2,3,4,5,7,8,9
CardiacCardiac PediatricPPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPBMDCNote 2,3,4,5,7,8,9
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

  • Note 3 3D imaging
  • Power SieScape panoramic imaging
    Contrast agent imaging
    Tissue Equalization Technology Note 5
  • Note 7
  • Note 9
  • Note 11 Dynamic TCE

Note 2 Ensemble tissue harmonic imaging

Note 4 B&W SieScape panoramic imaging

For example: abdominal, vascular
SieClear multi-view spatial compounding Note 6

Intracardiac imaging

Note 8 Note 10

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 21 OF 31

{24}------------------------------------------------

510(k) Number (if known):

Device Name:8L3 Linear "Regel" Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows.
Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPBMDCNote 2,3,4,5,7,8,9
AbdominalPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPBMDCNote 2,3,4,5,7,8,9
SmallOrgan(Note 1)PPPPPBMDCNote 2,3,4,5,7,8,9
Neonatal CephalicPPPPPBMDCNote 2,3,4,5,7,8,9
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPBMDCNote 2,3,4,5,7,8,9
Musculo-skel.(Superfic)PPPPPBMDCNote 2,3,4,5,7,8,9
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPBMDCNote 2,3,4,5,7,8,9
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaging

Note 5 Power SieScape panoramic imaging

Contrast agent imaging
Tissue Equalization Technology Note 7

Note 9

Note 2 Ensemble tissue harmonic imaging

Note 4 B&W SieScape panoramic imaging

Note 6 For example: abdominal, vascular

Note 8 SieClear multi-view spatial compounding

Intracardiac imaging Note 10

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 22 OF 31

{25}------------------------------------------------

510(k) Number (if known):

Device Name:10V4 Phased Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows
Clinical ApplicationMode of Operation
Other( Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPPPBMDCNote 2,3,4,5,7,8,9
AbdominalPPPPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatricPPPPPPPPBMDCNote 2,3,4,5,7,8,9
Imaging& OtherSmallOrgan(Note 1)
Neonatal CephalicPPPPPPPPBMDCNote 2,3,4,5,7,8,9
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPPPBMDCNote 2,3,4,5,7,8,9
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac AdultPPPPPPPPBMDCNote 2,3,4,5,7,8,9
CardiacCardiac PediatricPPPPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vesselPPPPPPPPBMDCNote 2,3,4,5,7,8,9
VesselOther (Specify)

N = new indication; P = previously cleared by K080760, 121699

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

  • 3D imaging Note 3
    Power SieScape panoramic imaging
    Contrast agent imaging
    Tissue Equalization Technology Note 5

Note 7

Note 9

Note 11 Dynamic TCE Note 2 Ensemble tissue harmonic imaging

Note 4 B&W SieScape panoramic imaging

For example: abdominal, vascular
SieClear multi-view spatial compounding Note 6

Note 8 Note 10 Intracardiac imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 23 OF 31

{26}------------------------------------------------

510(k) Number (if known):

Device Name:C7F2 Curved Array Mechanical 3D/4D Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks1I& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatricPPPPPPBMDCNote2,3,4,5,7,8,9,11
Imaging& OtherSmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac AdultPPPPPPBMDCNote2,3,4,5,7,8,9,11
CardiacCardiac PediatricPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vesselPPPPPPBMDCNote2,3,4,5,7,8,9,11
VesselOther (Specify)

N = new indication; P = previously cleared by K080760, 121699

For example: breast, testes, thyroid, penis, prostate, etc. Note 1
Note 3

3D imaging
Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging Note 9 Tissue Equalization Technology

Note 9 Tissue Equalization Technique
Note 11 Dynamic TCE

Note 11 Dynamic TCE Note 8 SieClear multi-view spatial compounding

Note 2

Note 4

Note 6

Note 10 Intracardiac imaging

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging
For example: abdominal, vascular

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

{27}------------------------------------------------

510(k) Number (if known):

EV9F4 Curved Array Mechanical 3D/4D Transducer for use with: Device Name: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPBMDCNote2,3,4,5,7,8,9,11
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalPediatric
Imaging& OtherSmallOrgan(Note 1)PPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectalPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-vaginalPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vessel
VesselOther (Specify)

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3 3D imaging

Note 5 Power SieScape panoramic imaging

Note 7 Contrast agent imaging

Tissue Equalization Technology Note 9

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 25 OF 31

Note 8 Note 10 Intracardiac imaging

Note 2 Note 4

Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging

Note 6 For example: abdominal, vascular

3 SieClear Multi
10 Intracardiac im

SieClear multi-view spatial compounding

{28}------------------------------------------------

510(k) Number (if known):

L13F5 3D/4D Mechanical Wobbler Linear Transducer for use with: Device Name: ACUSON X300 Diagnostic Ultrasound Systems Intended Use: Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote 2,3,4,5,7,8,9
AbdominalPPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPPBMDCNote 2,3,4,5,7,8,9
SmallOrgan(Note 1)PPPPPPBMDCNote 2,3,4,5,7,8,9
Neonatal CephalicPPPPPPBMDCNote 2,3,4,5,7,8,9
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPBMDCNote 2,3,4,5,7,8,9
Musculo-skel.(Superfic)PPPPPPBMDCNote 2,3,4,5,7,8,9
Intra-vascular
Other(Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPBMDCNote 2,3,4,5,7,8,9
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

  • Note 3 3D imaging
    Power SieScape panoramic imaging
    Contrast agent imaging Note 5

Note 7

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE Note 2 Ensemble tissue harmonic imaging

Note 4 B&W SieScape panoramic imaging

SieClear multi-view spatial compounding

Intracardiac imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 26 OF 31

Note 6 For example: abdominal, vascular

Note 8

Note 10

{29}------------------------------------------------

510(k) Number (if known):

Device Name:VF8-3 Linear Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows
Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1 & 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)PPPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)PPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular
Other(Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)

For example: breast, testes, thyroid, penis, prostate, etc.

Note 1 Note 3 3D imaging

  • Note 5
    Power SieScape panoramic imaging Note 7 Contrast agent imaging

Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Note 8 Note 10 Intracardiac imaging

B&W SieScape panoramic imaging

Ensemble tissue harmonic imaging

Note 6

For example: abdominal, vascular SieClear multi-view spatial compounding

SleClear multi-view sp
Intracardiac imaging

Note 2 Note 4

{30}------------------------------------------------

510(k) Number (if known):

Device Name:P5-1 Phased Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks1I& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)
Neonatal CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
Cardiac AdultPPPPPPPBMDCNote2,3,4,5,7,8,9,11
CardiacCardiac PediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

  • Note 3
    Note 5
    3D imaging
    Power SieScape panoramic imaging
    C Note 7

Contrast agent imaging Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

e 6 For example: abc

8
10
SieClear multi-view spa
Intracardiac imaging

Note 2 Ensemble tissue harmonic imaging

B&W SieScape panoramic imaging
For example: abdominal, vascular

SieClear multi-view spatial compounding

Note 10 Intracardiac imaging

ON ANOTHER PAGE

Note 4
Note 6 Note 8

{31}------------------------------------------------

510(k) Number (if known):

Device Name:C6-2 Convex Array Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other(Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatricPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)PPPPPPBMDCNote2,3,4,5,7,8,9,11
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
CardiacCardiac AdultPPPPPPBMDCNote2,3,4,5,7,8,9,11
Cardiac PediatricPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)

N = new indication; P = previously cleared by K080760, 121699

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

Note 3

3D imaging
Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging Tissue Equalization Technology

Note 9 Note 11 Dynamic TCE

Note 11 Dynamic TCL

Note 2 Ensemble tissue harmonic imaging Note 4 B&W SieScape panoramic imaging

Note 6 For example: abdominal, vascular

SieClear multi-view spatial compounding

Note 10 Intracardiac imaging

Note 8

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

{32}------------------------------------------------

510(k) Number (if known):

Device Name:SoundStar 10F Transducer for use with:ACUSON X300 Diagnostic Ultrasound Systems
Intended Use:Diagnostic imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
Other( Track 1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(Note 6)
Intra-operative(Neuro)
Laparoscopic
FetalImaging& OtherPediatric
SmallOrgan(Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascularPPPPPPPBMDCNote2,3,7,8,9,10,11
Other(Specify)
Cardiac AdultPPPPPPPBMDCNote2,3,7,8,9,10,11
CardiacCardiac PediatricPPPPPPPBMDCNote2,3,7,8,9,10,11
Intra-vascular(Cardiac)PPPPPPPBMDCNote2,3,7,8,9,10,11
Trans-esophageal(Cardiac)
Intra-cardiacPPPPPPPBMDCNote2,3,7,8,9,10,11
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)
N = new indication; P = previously cleared by K080760, 121699
  1. non indication - 1 previously ordered by 10001-00/121000

Note 1 For example: breast, testes, thyroid, penis, prostate, etc.

  • Note 3 3D imaging
  • Note 5 Power SieScape panoramic imaging

Note 7 Contrast agent imaging

Tissue Equalization Technology Note 9

Note 11 Dynamic TCE Note 2 Ensemble tissue harmonic imaging Note 4

B&W SieScape panoramic imaging Note 6 For example: abdominal, vascular

Note 8 SieClear multi-view spatial compounding

Note 10 Intracardiac imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 30 OF 31

{33}------------------------------------------------

510(k) Number (if known):

P9-4 Phased Array Transducer for use with:

Device Name: Intended Use:

ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Other( Track1Only)Specific(Tracks1& 3)ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPPPPBMDCNote2,3,4,5,7,8,9,11
AbdominalPPPPPPPBMDCNote2,3,5,6,7,8,9,11
Intra-operative(Note 6)
Intra-operative(Neuro)PPPPPPBMDCNote2,3,4,5,7,8,9,11
Laparoscopic
FetalImaging& OtherPediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
SmallOrgan(Note 1)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
Neonatal CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Adult CephalicPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic)
Intra-vascular
Other(Specify)
CardiacCardiac AdultPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Cardiac PediatricPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Intra-vascular(Cardiac)
Trans-esophageal(Cardiac)
Intra-cardiac
Other(Cardiac Neonatal)PPPPPPPBMDCNote2,3,4,5,7,8,9,11
PeripheralVesselPeripheral vesselPPPPPPPBMDCNote2,3,4,5,7,8,9,11
Other (Specify)

N = new indication; P = previously cleared by FDA 121699

For example: breast, testes, thyroid, penis, prostate, etc. Note 1

Note 3 3D imaging

Power SieScape panoramic imaging Note 5

Note 7 Contrast agent imaging Note 9 Tissue Equalization Technology

Note 11 Dynamic TCE

Note 11 Dynamic TCE

Note 2 Ensemble tissue harmonic imaging

Note 4 B&W SieScape panoramic imaging Note 6 For example: abdominal, vascular

SieClear multi-view spatial compounding Note 8

Note 10 Intracardiac imaging

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

{34}------------------------------------------------

Indications for Use

510(k) Number (if known):

ACUSON SC2000 Diagnostic Ultrasound System Device Name:

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal Intraoperative, Intraoperative, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients.

The system transmits ultrasound energy from either a transthoracic or transesophageal approach in adult and pediatric patients; and from a transthoracic approach in neonatal and fetal cardiants creating 2D (B). 3D. M-Mode (M). Color Doppler (CD). Color Power Doppler (CPD). Pulsed Wave (PW) Dopler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms, legs and abdomen: and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Dopler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

{35}------------------------------------------------

Intraoperative Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B). Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology.

The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients

Type of Use (Select one or both, as applicable)

X Prescription Use (21 CFG 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Page 2 of 18

{36}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

SC2000 Diagnostic Ultrasound System

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:3DOther:RealTime3D
Ophthalmic
FetalPPPPPPPP*PP
AbdominalPPPPPPPP*P
IntraoperativeAbdominalPPPPPPPP*P
IntraoperativeNeurological
PediatricPPPPPPPP*PPP
Small Organ(specify)
NeonatalCephalic
Adult CephalicPPPPPPPP*P
CardiacPPPPPPPP*PPP
Trans-esophagealPPPPPPP*P
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPPP*P
Peripheral VesselPPPPPPPPP*P
Laparoscopic
Musculo-skeletalConventionalPPPPPPPP*P
Musculo-skeletalSuperficialPPPPPPPP*P
Other (NeonatalCardiac)PPPPPPPP*P
Other (Intra-Cardiac)PPPPPPP*P

N=new indication. P = Previously Cleared K072365, K102017, K113179, K132654

Additional Comments:

*Combinations include:

B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

{37}------------------------------------------------

510(k) Number (if known):

Device Name:

AcuNav 8F Intracardiac Transducer for use with:

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPPP*
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPPP*
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPPP*
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPPP*

N=new indication. P = Previously Cleared K072365, K102017, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)_

Page 4 of 18

{38}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

Soundstar eco 8F Intracardiac Transducer for use with: ACUSON SC2000 Diagnostic Ultrasound Systems Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPPP*
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPPP*
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPPP*
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPPP*

N=new indication. P = Previously Cleared

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)__

Page 5 of 18

{39}------------------------------------------------

510(k) Number (if known):

Device Name:

9L4

Indications for Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
Pediatric
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
Cardiac
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral VesselPPPPPPPP*P
Laparoscopic
Musculo-skeletalConventionalPPPPPPPP*P
Musculo-skeletalSuperficialPPPPPPPP*P
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)

Page 6 of 18

{40}------------------------------------------------

6C1HD

510(k) Number (if known):

Device Name:

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
FetalPPPPPPP*P
AbdominalPPPPPPP*P
IntraoperativeAbdominalPPPPPPP*P
Intraoperative
PediatricPPPPPPP*P
Small Organ(specify)PPPPPPP*P
NeonatalCephalic
Adult Cephalic
Cardiac
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral VesselPPPPPPP*P
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other(AbdominalVascular)PPPPPPP*P

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)

Page 8 of 18

{41}------------------------------------------------

510(k) Number (if known):

Device Name:

V5M

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPP*P
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPP*P
Trans-esophagealPPPPPPP*P
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)

Page 9 of 18

{42}------------------------------------------------

510(k) Number (if known):

Device Name:

4V1c

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
FetalPPPPPPPP *P
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPPP *P
Small Organ(specify)
NeonatalCephalic
Adult CephalicPPPPPPPP *P
CardiacPPPPPPPP *P
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (NeonatalCardiac)PPPPPPPP *P

N=new indication. Previously Cleared in 510(k) K072365, K102017, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 10 of 18

{43}------------------------------------------------

510(k) Number (if known):

Device Name:

8V3

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
FetalPPPPPPP *P
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPP *P
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPP *P
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (NeonatalCardiac)PPPPPPP *P

N=new indication. Previously Cleared in 510(k) K102017, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 11 of 18

{44}------------------------------------------------

510(k) Number (if known):

Device Name:

CW2

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricP
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacP
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral VesselP
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. Previously Cleared in 510(k) K072365, K102017, K113179, K132654

Additional Comments:

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 12 of 18

{45}------------------------------------------------

510(k) Number (if known):

Device Name:

4Z1c

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
FetalPPPPPP*PP
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPP*PP
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPP*PP
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 13 of 18

{46}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

AcuNav 10F Ultrasound Catheter

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPPP*
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPPP*
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPPP*
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPPP*

N=new indication. P = Previously Cleared in 510(k) K071234, K093812, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)

Page 14 of 18

{47}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

ACUSON AcuNav™ V 10F Ultrasound Catheter

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPPP*P
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPPP*P
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPPP*P
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPPP*P

N=new indication. P = Previously Cleared in 510(k) K081808, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 15 of 18

{48}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

SoundStar 10F Ultrasound Catheter

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
Pediatric
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPPP*
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPPP*
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPPP*

N=new indication. P = Previously Cleared in 510(k) K070242, K113179, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

510(k)

Page 16 of 18

{49}------------------------------------------------

510(k) Number (if known):

Device Name:

V7M

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:3D
Ophthalmic
Fetal
AbdominalPPPPPPP*PP
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPP*P
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacPPPPPP*P
Trans-esophagealPPPPPP*P
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K111674, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 17 of 18

{50}------------------------------------------------

510(k) Number (if known):

Device Name:

10V4

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
FetalPPPPPPPP *P
AbdominalPPPPPPPP *P
IntraoperativeAbdominalPPPPPPPP *P
IntraoperativeNeurologicalPPPPPPPP *P
PediatricPPPPPPPP *P
Small Organ(specify)
NeonatalCephalicAdult Cephalic
CardiacPPPPPPPP *P
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (NeonatalCardiac)PPPPPPPP *P

N=new indication. Previously Cleared in 510(k) K111674, K132654

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) 510(k)_

Page 18 of 18

{51}------------------------------------------------

510(k) Summary Prepared August 19, 2014

Sponsor:Siemens Medical Solutions, Inc.,Ultrasound Division685 East Middlefield RoadMountain View, California 94043
Contact Person:Shelly PearceTelephone: (650) 694-5988
Submission Date:August 7, 2014
Device Name:SC2000, X300 Diagnostic Ultrasound Systems
Common Name:Diagnostic Ultrasound System with Accessories

Classification:

Regulatory Class: ll Review Category: Tier II Classification Panel: Radiology

Ultrasonic Pulsed Doppler Imaging SystemCFR # 892.1550Product Code 90-IYN
Ultrasonic Pulsed Echo Imaging SystemCFR # 892.1560Product Code 90-IYO
Diagnostic Intravascular CatheterCFR # 870.1200Product Code OBJ

A. Legally Marketed Predicate Devices

The device modifications described in this 510(k) are substantially equivalent to the company's own devices, previously cleared on K132654 (SC2000), K121699 (X3001 (X700), K141846 (X700). The catheter is substantially equivalent to K140318 (SSe8F), K071234 (AN8F).

B. Device Description:

The Diagnostic Ultrasound Systems are multi-purpose mobile, software controlled diagnostic ultrasound systems with on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, 3D Imaging, or Harmonic Imaging and 4D imaging.

{52}------------------------------------------------

C. Intended Use

SC2000 -

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neonatal and Fetal Cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adiunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juqqular veins in the neck; superficial and deep veins and arteries in the arms and legs and abdomen; and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

Intraoperative Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous

{53}------------------------------------------------

Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

X300:

The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

The Acuson Acunav and Soundstar Ultrsound Catheter are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

D. Substantial Equivalence

The device modifications described in this 510(k) are substantially equivalent to the company's own devices, previously cleared on K132654 (SC2000), K121699 (X300), K123001 (X700), K141846 (X700). The catheter is substantially equivalent to K140318 (SSe8F), K071234 (AN8F). The submission devices are substantially equivalent to the predicate devices with regard to both intended use and technological characteristics.

DescriptionThissubmission(AllDevices)AcusonX700K141846AcusonSC2000 TMK132654AcusonX300 - K121699 /X700 - K123001
System:
Hardware Safety - EN60601-1 CertifiedXXXX
Acoustic Thermal Safety - EN60601-2-37 CertifiedXXXX

{54}------------------------------------------------

Medical device software - Software Life CycleProcess - IEC 62304XXXX
Transducers:
As cleared in the device's previous 510(k)XXXX
AcuNav 8F CatheterXXXX
Soundstar eco 8F CatheterXXXX
Software Features
As cleared in the device's previous 510(k)XXXX

E. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence

The devices have been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform with applicable medical device safety standards. The systems comply with the following voluntary standards:

  • UL 60601-1, Safety Requirements for Medical Equipment
  • . IEC 60601-2-37 Diagnostic Ultrasound Safety Standards
  • . CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment

AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic ■ Output Indices on Diagnostic Ultrasound Equipment

  • AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound .
  • 93/42/EEC Medical Devices Directive ■
  • 트 Safety and EMC Requirements for Medical Equipment
  • 트 EN/IEC 60601-1
  • 트 EN/IEC 60601-1-1
  • . EN/IEC 60601-1-2
  • EN/IEC 62304
  • EN/IEC 62366 트
  • 트 EN/IEC 60601-2-18
  • EN/IEC 60601-2-25 트
  • ISO 10993-1 Biocompatibility ■

Cleared patient contact materials, electrical and mechanical safety are unchanged.

F. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.

Since the modified devices use the same technology and principles as existing devices, clinical data is not required.

G. Summary

Intended uses and other key features are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms with 21 CFR 820 Quality System Requlation and ISO 13485:2003 quality system standards. The product is designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore it is the opinion of Siemens Medical that the modified devices are substantially equivalent with respect to safety and effectiveness to devices currently cleared for market. The modified devices are verified and validated according to the company's design control process.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.