K Number
K130754

Validate with FDA (Live)

Date Cleared
2013-03-25

(6 days)

Product Code
Regulation Number
892.1550
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications.

Device Description

The Acuson SC2000™ has been designed to meet the following product safety standards: UL 60601-1, Safety Requirements for Medical Equipment; IEC 60601-2-37 Diagnostic Ultrasound Safety Standards; CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment; AIUM/NEMA UD-3. Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment; AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound; Safety and EMC Requirements for Medical Equipment; IEC 60601-1; IEC 60601-1-1; IEC 60601-1-2; IEC 1157 Declaration of Acoustic Power; ISO 10993-1 Biocompatibility.

AI/ML Overview

This 510(k) summary for the ACUSON SC2000 Diagnostic Ultrasound System is a Special 510(k), indicating a minor change to a previously cleared device. Therefore, a comprehensive de novo clinical study demonstrating performance against acceptance criteria for a novel device is not part of this submission. Instead, the performance data section explicitly states: "No performance data required to support this modification, as there is no change to any hardware or software, or method of use."

The submission focuses on demonstrating substantial equivalence to a previously cleared device (K123622) by only renaming three transducers. As such, the typical elements of a study proving a device meets acceptance criteria (like a test set, ground truth, experts, and statistical results) are not provided in this document.

However, I can extract the acceptance criteria implied by this type of submission, which is the determination of substantial equivalence to the predicate device for the specified indications for use.

Here's a breakdown based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied for a Special 510(k) for Renaming Transducers): The modified device, specifically the renamed transducers, must demonstrate substantial equivalence to the previously cleared predicate device concerning intended use and technological characteristics, with no changes to hardware, software, or method of use.

Acceptance CriterionReported Device Performance
Substantial Equivalence to PredicateThe submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics. This modification is intended only to rename three transducers.
No Change in Hardware/Software/Method of UseConfirmed: "No performance data required to support this modification, as there is no change to any hardware or software, or method of use."
Continued Adherence to Safety StandardsThe Acuson SC2000™ has been designed to meet the following product safety standards: UL 60601-1, IEC 60601-2-37, CSA C22.2 No. 601-1, AIUM/NEMA UD-3, AIUM/NEMA UD-2, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 1157, ISO 10993-1.
Intended Use Remaining ConsistentThe Indications for Use for the overall system and the individual transducers remain largely consistent with previously cleared indications. (P = Previously Cleared)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not applicable. No new performance testing with a specific test set was conducted for this Special 510(k) given the nature of the modification (renaming transducers).
  • Data Provenance: Not applicable. No new data was generated or analyzed for this submission. The basis for substantial equivalence relies on the established performance of the predicate device and the claim that the changes are confined to nomenclature.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable. No new ground truth establishment was required as no new performance data was generated.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No. This submission does not describe an MRMC comparative effectiveness study.
  • Effect Size of Human Readers Improvement: Not applicable.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Standalone Performance: Not applicable. This device is an ultrasound imaging system, not an AI algorithm requiring standalone performance evaluation in the context of this submission.
  • The document states there is "no change to any hardware or software," implying no new algorithms or software features are being introduced that would require such testing.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Not applicable for new performance evaluation. The "ground truth" for this Special 510(k) is implicitly the established safety and effectiveness of the predicate device (K123622), to which the modified device (with renamed transducers) is claimed to be substantially equivalent without any functional changes.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. This document does not describe the development or training of any new algorithms or software components.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth Establishment for Training Set: Not applicable.

{0}------------------------------------------------

ACUSON SC2000 Diagnostic Ultrasound System Special 510(k)

K130754
Page 1 of 3

SECTION 11

MAR 2 5 2013

510(k) Summary Prepared March 6, 2013

Siemens Medical Solutions USA, Inc., Sponsor: Ultrasound Division 685 East Middlefield Road Mountain View, California 94043

Contact Person: Patrick Lynch Telephone: (650) 694-5658

March 6, 2013 Submission Date:

Modified Device Name: ACUSON SC2000 Diagnostic Ultrasound System

Diagnostic Ultrasound System Common Name:

Classification:

Regulatory Class: ll Review Category: Tier II Classification Panel: Radiology

Ultrasonic Pulsed Doppler Imaging SystemFR # 892.1550Product Code 90-IYN
Ultrasonic Pulsed Echo Imaging SystemFR # 892.1560Product Code 90-IYC
Diagnostic Ultrasound TransducerFR # 892.1570Product Code 90-ITX
Diagnostic Intravascular CatheterFR # 870.1200Product Code OBJ

A. Legally Marketed Predicate Devices

The modified Diagnostic Ultrasound System is substantially equivalent to the company's previously cleared device (K123622).

B. Device Description:

The Acuson SC2000™ has been designed to meet the following product safety standards:

  • UL 60601-1, Safety Requirements for Medical Equipment ■
  • 제 IEC 60601-2-37 Diagnostic Ultrasound Safety Standards
  • 1 CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment
  • AIUM/NEMA UD-3. Standard for Real Time Display of Thermal and ■ Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
  • 해 AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound
  • Safety and EMC Requirements for Medical Equipment
  • IEC 60601-1
  • IEC 60601-1-1
  • IEC 60601-1-2
  • IEC 1157 Declaration of Acoustic Power
  • ISO 10993-1 Biocompatibility

C. Intended Use

{1}------------------------------------------------

ACUSON SC2000 Diagnostic Ultrasound System Special 510(k)

K130754
Page 2 of 3

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Intraoperative, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients.

The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs; and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic capture of ultrasound images.

The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

Intraoperative Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Dooder (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Dopoler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology.

The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

D. Substantial Equivalence

The submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics. This modification is intended only to rename three transducers, as described in the table below.

{2}------------------------------------------------

K130754
Page 3 of 3

.

. :

.

:

:

ACUSON SC2000 Diagnostic Ultrasound System Special 510(k)

·

PredicateDevice(K123622)SubmissionDevice
Current NameModified Name
8V3c8V3
V5M TEEV5Ms
AUX CW2CW2

E. Performance Data

No performance data required to support this modification, as there is no change to any hardware or software, or method of use.

.

:

:

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three arms or branches extending upwards.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2013

Siemens Medical Solutions, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street. NW BUFFALO MN 55313

Re: K130754

Trade/Device Name: SC2000™ Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, OBJ Dated: March 18, 2013 Received: March 19, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Acuson SC2000™ Diagnostic Ultrasound System, as described in your premarket notification:

Transducer Model Number

9L4AcuNav 8F and 10F Ultrasound Catheter
V5MsACUSON AcuNavTM V 10F UltrasoundCatheter
4V1c
8V3SoundStar 10F Ultrasound Catheter
CW2V7M TEE
4Z1c10V4

{4}------------------------------------------------

Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Sm.h.7)

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K130754

Device Name: SC2000™ Diagnostic Ultrasound System

Indications For Use:

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients.

The system transmits ultrasound energy from either a transthoracic or transesophageal approach in adult and pediatric patients; and from a transthoracic approach in neonatal and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs; and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

{6}------------------------------------------------

Intraoperative Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during neurological intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology.

The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Page 2 of 14

{7}------------------------------------------------

510(k) Number (if known):

Device Name: SC2000 Diagnostic Ultrasound System

Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationABMPWDCWDColor DopplerPower (Amplitude) DopplerColor Velocity ImagingCombined (Specify)Other: Harmonic ImagingOther: 3DOther: Real Time 3D
Ophthalmic
FetalPPPPPPP*PP
AbdominalPPPPPPP*P
Intraoperative AbdominalPPPPPPP*P
Intraoperative NeurologicalPPPPPPP*P
PediatricPPPPPPP*PPP
Small Organ (specify)
Neonatal Cephalic
Adult CephalicPPPPPPP*P
CardiacPPPPPPP*PPP
Trans-esophagealPPPPPP*P
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPP*P
Peripheral VesselPPPPPPPP*P
Laparoscopic
Musculo-skeletal ConventionalPPPPPPP*P
Musculo-skeletal SuperficialPPPPPPP*P
Other (Neonatal Cardiac)PPPPPPP*P
Other (Intra-Cardiac)PPPPPP*P

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179, K123622

Additional Comments:


*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)

Page 3 of

{8}------------------------------------------------

510(k) Number (if known):

Device Name:

9L4

Indications for Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurologicalPPPPPPP*P
Pediatric
Small Organ(specify)
NeonatalCephalic
Adult Cephalic .
Cardiac
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral VesselPPPPPPP*P
·Laparoscopic
Musculo-skeletalConventionalPPPPPPP*P
Musculo-skeletalSuperficialPPPPPPP*P
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179, K123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)

Page 4 of 14

{9}------------------------------------------------

510(k) Number (if known):

Device Name:

VSM TEE

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPP*P
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacPPPPPP*P
Trans-esophagealPPPPPP*P
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179, K123622

Additional Comments:


*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWQ+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)


{10}------------------------------------------------

510(k) Number (if known):

Device Name: .

4V1cc

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
FetalPPPPPPP*P
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPP*P
Small Organ(specify)
NeonatalCephalic
Adult CephalicPPPPPPP*P
CardiacPPPPPPP*P
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (NeonatalCardiac)PPPPPPP*P

N=new indication. Previously Cleared in 510(k) K072365, K102017. K113179, K123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Cglor Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)

Page 6 of

{11}------------------------------------------------

510(k) Number (if known):

Device Name:

8V3c

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
FetalPPPPPPP *P
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPP *P
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPP *P
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (NeonatalCardiac)PPPPPPP *P

N=new indication. Previously Cleared in 510(k) K102017, K113179, K123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, _______________________________________________________________________________________ B+CWD+Power Doppler, B+Clarify VE

{12}------------------------------------------------

510(k) Number (if known):

Device Name:

AUX CW2

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricP
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacP
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral VesselP
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. Previously Cleared in 510(k) K072365, K102017, K113179, K123622

Additional Comments:

(PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109)

Page 8 of ___ 14______________________________________________________________________________________________________________________________________________________________

{13}------------------------------------------------

510(k) Number (if known):

Device Name:

4Z1c

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
FetalPPPPPP*PP
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPP*PP
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPP*PP
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179, K123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, J B+CWD+Power Doppler、B+Clarify VE

{14}------------------------------------------------

510(k) Number (if known):

Device Name:

AcuNav 8F and 10F Ultrasound Catheter

Intended Use:

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPPP*
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPPP*
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPPP*
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPPP*

N=new indication. P = Previously Cleared in 510(k) K071234, K093812, K113179, K123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler

{15}------------------------------------------------

510(k) Number (if known):

Device Name:

ACUSON AcuNav™ V 10F Ultrasound Catheter

Intended Use:

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPP*P
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPP*P
Trans-esophageal
Transrectal
Transvaginal
Transurethra!
Intra-LuminalPPPPPPP*P
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPP*P

N=new indication. P = Previously Cleared in 510(k) K081808, K 113179, K 123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

{16}------------------------------------------------

510(k) Number (if known):

Device Name:

SoundStar 10F Ultrasound Catheter

Intended Use:

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:RealTime3D
Ophthalmic
Fetal
Abdominal
IntraoperativeAbdominal
IntraoperativeNeurological
Pediatric
Small Organ(specify) **
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPP*
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-LuminalPPPPPPP*
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (Intra-Cardiac)PPPPPPP*

N=new indication. P = Previously Cleared in 510(k) K070242, K 113179, K 123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler

{17}------------------------------------------------

510(k) Number (if known):

Device Name:

V7M TEE

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImagingOther:3D
Ophthalmic
Fetal
AbdominalPPPPPPPP*PP
IntraoperativeAbdominal
IntraoperativeNeurological
PediatricPPPPPPP*P
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPP*P
Trans-esophagealPPPPPPP*P
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (specify)

N=new indication. P = Previously Cleared in 510(k) K 111674, K 123622

Additional Comments:

  • Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color, _________________________________________________________________________________ Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)

Page 13 of 14

{18}------------------------------------------------

510(k) Number (if known):

Device Name:

10V4

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

ClinicalApplicationABMPWDCWDColorDopplerPower(Amplitude)DopplerColorVelocityImagingCombined(Specify)Other:HarmonicImaging
Ophthalmic
FetalPPPPPPP *P
AbdominalPPPPPPP *P
IntraoperativeAbdominalPPPPPPP *P
IntraoperativeNeurologicalPPPPPPP *P
PediatricPPPPPPP *P
Small Organ(specify)
NeonatalCephalic
Adult Cephalic
CardiacPPPPPPP *P
Trans-esophageal
Transrectal
Transvaginal
Transurethral
Intra-Luminal
Peripheral Vessel
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletalSuperficial
Other (NeonatalCardiac)PPPPPPP *P

N=new indication. Previously Cleared in 510(k) K111674, K123622

Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

Smh.p).

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k) K130754

of 14 Page 14

N/A