Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K222744
    Date Cleared
    2022-10-11

    (29 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K101359

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sol-M Luer Lock Syringe is intended for aspiration of medications and fluids into the body.

    Device Description

    The Sol-M Luer Lock Syringe is used to aspirate and inject fluid/medication into the body. The syringe is a sterile, single-use, 3-part syringe with luer lock conical connection. The Sol-MTM Luer Lock Syringe consists of the following components: Barrel, Plunger, Gasket. In addition to being used manually, the Sol-MTM Luer Lock Syringe can also work with a powerdriven syringe pump, with the luer lock tip connected to the IV administration line to deliver specific volume of fluid/medication. The syringe is individually blister packaged, and EtO sterilized with SAL of 10-6. The Sol-M™ Luer Lock Syringe is provided in 1, 3, 5, 10, 20, 30, and 60ml volume.

    AI/ML Overview

    The provided text describes the Sol-M Luer Lock Syringe (K222744) and its acceptance criteria, particularly focusing on its suitability for use with power-driven syringe pumps.

    Here's an analysis of the provided information, framed by your requested criteria:

    The study primarily focuses on demonstrating the suitability of the Sol-M Luer Lock Syringe for use with power-driven syringe pumps, referring to ISO 7886-2 as the key standard.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states that the device was tested according to ISO 7886-2:2020. The acceptance criteria are implicitly those defined by this standard. The reported performance is that the device met these criteria.

    Acceptance Criteria Category (based on ISO 7886-2)Reported Device Performance
    Critical Dimensions (for fit in pump)Met acceptance criteria
    Short-term Flow Rate Error (with pump)Met acceptance criteria
    Pump Forces (required to move plunger)Measured and made available
    Syringe Compliance (fluid displaced)Met acceptance criteria
    Manual Use (ISO 7886-1)Previously demonstrated
    Material/Design/IndicationsUnchanged from predicate
    Shelf Life (3 years)Verified in testing
    BiocompatibilityLeveraged from predicate
    SterilizationUnchanged from predicate

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the specific number of syringes used for each test (e.g., for critical dimensions, flow rate, compliance tests). It only refers to "testing per ISO 7886-2."
    • Data Provenance: Not explicitly stated, but based on the context of a 510(k) submission, this would be prospective testing conducted by the manufacturer, Sol-Millennium Medical Inc. The country of origin of the data is generally where the manufacturer performs their testing, which for Sol-Millennium Medical Inc. is based in Suwanee, Georgia.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is a bench testing (non-clinical) submission involving a physical medical device (syringe). "Ground truth" in this context refers to standardized measurements and performance specifications outlined in ISO 7886-2, not expert consensus on diagnostic images or clinical outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies, especially those involving human interpretation of data where consensus is needed to establish a "ground truth" or resolve disagreements. For bench testing against a standard like ISO 7886-2, objective measurements are taken, compared against predefined limits, and either pass or fail.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret data (e.g., radiologists reading images) and the AI aims to improve their performance. This submission is for a physical medical device (syringe) and focuses on engineering performance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This is for a physical medical device, not an algorithm or AI. The testing described is "standalone" in the sense that the syringe's performance is measured directly, not in conjunction with human interpretation of its output.

    7. Type of Ground Truth Used

    The "ground truth" for the test set is established by international standards and specifications, specifically ISO 7886-2:2020 (for use with power-driven syringe pumps) and ISO 7886-1:2017 (for manual use). These standards define acceptable ranges for dimensional requirements, flow rate error, pump forces, and syringe compliance.

    8. Sample Size for the Training Set

    Not applicable. There is no AI/algorithm component requiring a training set mentioned in this submission. This is a traditional medical device (syringe) bench test.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153537
    Date Cleared
    2016-06-17

    (190 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K132681,K092430,K101359

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
    The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.

    Device Description

    The Sol-M Syringe is a sterile, single use, standard 3 piece piston syringe with a fixed needle. The Sol-M 0.3, 0.5 and 1 ml Insulin Syringe will be labeled with a Units Scale calibrated for U-100 Insulin. The 0.5 and 1 ml TB Syringe will be labeled with a mL Scale. The piston type syringe is a plastic disposable syringe made of the following components: 1. Barrel – The barrel has a scale showing the capacity of the syringe as well as the company brand name with the Do Not Reuse symbol. 2. Plunger - The plunger is used to aspirate fluid into the barrel or push fluid out of the barrel. 3. Stopper - The Stopper maintains the fluid in the barrel between the base of the barrel and Plunger. 4. Cap The Cap is used to cover and protect the tip of the barrel from being damaged.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (syringes), not an AI/ML device. Therefore, the requested information about acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, MRMC studies, and standalone performance is generally not applicable in the context of AI/ML device evaluations. Traditional medical device clearances like this one focus on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical bench testing.

    Based on the provided document, here's what can be extracted related to performance and testing, framed in the closest possible way to your request, but acknowledging the difference in context:

    1. A table of acceptance criteria and the reported device performance

    For the Sol-M Insulin Syringes:

    Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
    Delivery AccuracyISO 8537:2007 Sterile single use syringes, with and without needles, for insulinComplies with ISO 8537:2007
    Stainless Steel Needle TubingISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/Amendment:2001Complies with ISO 9626:1991

    For the Sol-M TB Syringes:

    Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
    Manual Use SyringesISO 7886-1:1993 Sterile Hypodermic syringes for single use Part 1: Syringes for manual useComplies with ISO 7886-1:1993
    Hub/Needle Bond StrengthPer ISO 7886 (likely referring to the relevant section within 7886-1 or similar)Complies with ISO 7886
    Stainless Steel Needle TubingISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/Amendment:2001Complies with ISO 9626:1991

    For both devices (general characteristics):

    Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
    Lubricant amount/cm2Predicate: 1 ml – 0.078 Mg/cm2Subject device: 1ml-0.078mg/cm2 (Matches predicate)
    BiocompatibilityPer ISO 10993-1Complies with ISO 10993-1
    LabelingPer 21 CFR 801Complies with 21 CFR 801
    MaterialsBarrel, Plunger, Cap: Polypropylene; Gasket: SantopreneMatches predicate
    Gradations LegibilityBold markingsBold markings (Matches predicate)
    Barrel TransparencyClearClear (Matches predicate)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Bench tests were conducted to verify that the proposed devices met all design specifications." However, it does not specify the sample sizes for these bench tests. The data provenance is not explicitly mentioned as country of origin, but these are non-clinical tests performed to regulatory standards. The testing is prospective in the sense that new samples of the device were tested against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as this is a traditional medical device (syringes) and not an AI/ML device relying on expert-annotated ground truth for performance evaluation in the way an AI would. Performance is measured against engineering and material standards (e.g., ISO standards).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable in the context of traditional device bench testing against ISO standards. Test results are typically objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable, as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by international and national standards for medical syringes (e.g., ISO 8537, ISO 7886-1, ISO 9626, ISO 10993-1, 21 CFR 801). The device's physical and functional characteristics are objectively measured against these established parameters.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device and does not involve training data.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device and does not involve a training set or its ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1