K Number
K153537

Validate with FDA (Live)

Date Cleared
2016-06-17

(190 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.

Device Description

The Sol-M Syringe is a sterile, single use, standard 3 piece piston syringe with a fixed needle. The Sol-M 0.3, 0.5 and 1 ml Insulin Syringe will be labeled with a Units Scale calibrated for U-100 Insulin. The 0.5 and 1 ml TB Syringe will be labeled with a mL Scale. The piston type syringe is a plastic disposable syringe made of the following components: 1. Barrel – The barrel has a scale showing the capacity of the syringe as well as the company brand name with the Do Not Reuse symbol. 2. Plunger - The plunger is used to aspirate fluid into the barrel or push fluid out of the barrel. 3. Stopper - The Stopper maintains the fluid in the barrel between the base of the barrel and Plunger. 4. Cap The Cap is used to cover and protect the tip of the barrel from being damaged.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (syringes), not an AI/ML device. Therefore, the requested information about acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, MRMC studies, and standalone performance is generally not applicable in the context of AI/ML device evaluations. Traditional medical device clearances like this one focus on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical bench testing.

Based on the provided document, here's what can be extracted related to performance and testing, framed in the closest possible way to your request, but acknowledging the difference in context:

1. A table of acceptance criteria and the reported device performance

For the Sol-M Insulin Syringes:

Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
Delivery AccuracyISO 8537:2007 Sterile single use syringes, with and without needles, for insulinComplies with ISO 8537:2007
Stainless Steel Needle TubingISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/Amendment:2001Complies with ISO 9626:1991

For the Sol-M TB Syringes:

Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
Manual Use SyringesISO 7886-1:1993 Sterile Hypodermic syringes for single use Part 1: Syringes for manual useComplies with ISO 7886-1:1993
Hub/Needle Bond StrengthPer ISO 7886 (likely referring to the relevant section within 7886-1 or similar)Complies with ISO 7886
Stainless Steel Needle TubingISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/Amendment:2001Complies with ISO 9626:1991

For both devices (general characteristics):

Acceptance Criteria CategoryStandard/RequirementReported Device Performance (Compliance)
Lubricant amount/cm2Predicate: 1 ml – 0.078 Mg/cm2Subject device: 1ml-0.078mg/cm2 (Matches predicate)
BiocompatibilityPer ISO 10993-1Complies with ISO 10993-1
LabelingPer 21 CFR 801Complies with 21 CFR 801
MaterialsBarrel, Plunger, Cap: Polypropylene; Gasket: SantopreneMatches predicate
Gradations LegibilityBold markingsBold markings (Matches predicate)
Barrel TransparencyClearClear (Matches predicate)

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "Bench tests were conducted to verify that the proposed devices met all design specifications." However, it does not specify the sample sizes for these bench tests. The data provenance is not explicitly mentioned as country of origin, but these are non-clinical tests performed to regulatory standards. The testing is prospective in the sense that new samples of the device were tested against established standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as this is a traditional medical device (syringes) and not an AI/ML device relying on expert-annotated ground truth for performance evaluation in the way an AI would. Performance is measured against engineering and material standards (e.g., ISO standards).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable in the context of traditional device bench testing against ISO standards. Test results are typically objective measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable, as this is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable, as this is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by international and national standards for medical syringes (e.g., ISO 8537, ISO 7886-1, ISO 9626, ISO 10993-1, 21 CFR 801). The device's physical and functional characteristics are objectively measured against these established parameters.

8. The sample size for the training set

Not applicable, as this is not an AI/ML device and does not involve training data.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI/ML device and does not involve a training set or its ground truth.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a bird or abstract human figures in profile, with three distinct head shapes layered on top of each other. The image is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 17, 2016

Sol-Millennium Medical, Inc. Mr. Jim Barley Director of Regulatory Affairs 1735 North Brown Road, Suite 100 Lawrenceville, Georgia 30043

Re: K153537

Trade/Device Name: Sol-M Insulin Syringe and Sol-M TB Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: II Product Code: FMF Dated: May 2, 2016 Received: May 10, 2016

Dear Mr. Barley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Jim Barley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang -

for Erin I. Keith, M.S Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
510(k) Number (if known)K153537
Device NameSol-M TB Syringe
Indications for Use (Describe)The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
510(k) Number ( if known )K153537
Device NameSol-M Insulin Syringe
Indications for Use ( Describe )The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.
Type of Use ( Select one or both, as applicable ) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14) Page 1 of 1

{4}------------------------------------------------

510(k) Summary for 510(k) # K153537

(As required by 21 CFR 807.92(a))

510(k) Summary for the Sol-M Insulin and TB Syringes

Date Prepared: June 16, 2016 A. Submitter Information

Sol-Millennium Medical, Inc. 1735 North Brown Road Suite 120 Lawrenceville, GA 30043 Contact: Phone Number:

Trade Names:

  • B. Device Information Trade/Proprietary Name: Common name of device: Classification Name: Product Code: Regulatory Class: Classification Number: Reason for 510(k):
    C. Predicate Device:

Predicate 510(k) #:

Predicate product code:

Reference Devices:

Jim Barley 404-973-2200

Sol-M Insulin Syringe Sol-M TB Syringe

Sol-M Insulin and TB Syringe Piston syringe Piston syringe 80 FMF II 880.5860 New device

KDL Syringes and Needles

K112057

FMF

Sol-Guard Insulin and Tuberculin Safety Syringe 510(k) # K132681

1 ml Fixed Needle, 3 ml Luer Lock, 5 ml Luer Lock, 10 ml Luer Lock, 20 ml LuerLock Safety Syringe 510(k) # K092430

Invirostripe Luer Lock Syringes (1, 3, 10, 20, 30 And 60 ml 510(k) #K101359

{5}------------------------------------------------

D. Device Description

The Sol-M Syringe is a sterile, single use, standard 3 piece piston syringe with a fixed needle. The Sol-M 0.3, 0.5 and 1 ml Insulin Syringe will be labeled with a Units Scale calibrated for U-100 Insulin. The 0.5 and 1 ml TB Syringe will be labeled with a mL Scale.

The piston type syringe is a plastic disposable syringe made of the following components:

    1. Barrel – The barrel has a scale showing the capacity of the syringe as well as the company brand name with the Do Not Reuse symbol.
    1. Plunger - The plunger is used to aspirate fluid into the barrel or push fluid out of the barrel.
    1. Stopper - The Stopper maintains the fluid in the barrel between the base of the barrel and Plunger.
  • Cap The Cap is used to cover and protect the tip of the barrel from being 4. damaged.

Following is a listing of model numbers and product descriptions included in this submission:

Sol-M Insulin Syringes with Fixed Needle

Sol-M 0.3 ml Insulin Syringe with Fixed Needle
REF/Catalog #Description
163315180.3ml, 31G x 5/16"/10 Pack PE Bag
163305180.3ml, 30G x 5/16"/10 Pack PE Bag
163301280.3ml, 30G x 1/2"/10 Pack PE Bag
163291280.3ml, 29G x 1/2"/10 Pack PE Bag
Sol-M 0.3 ml Insulin Syringe with Fixed Needle
REF/Catalog #Description
163305160.3ml, 30G x 5/16"/1 ea. Blister Pack
16330120.3ml, 30G x 1/2"/ 1 ea. Blister Pack
16329120.3ml, 29G x 1/2"/ 1 ea. Blister Pack
Sol-M 0.5 ml Insulin Syringes with Fixed Needle
REF/Catalog #Description
165315180.5ml, 31G x 5/16"/10 Pack PE Bag
165305180.5ml, 30G x 5/16"/10 Pack PE Bag
165301280.5ml, 30G x 1/2"/10 Pack PE Bag
165291280.5ml, 29G x 1/2"/10 Pack PE Bag
161305160.5ml, 30G x 5/16"/1 ea. Blister Pack
16130120.5ml, 30G x 1/2"/ 1 ea. Blister Pack
16129120.5ml, 29G x 1/2"/ 1 ea. Blister Pack

{6}------------------------------------------------

Sol-M 1 ml Insulin Syringes with Fixed Needle
REF/Catalog #Description
161315181ml, 31G x 5/16"/10 Pack PE Bag
161305181ml, 30G x 5/16"/10 Pack PE Bag
161301281ml, 30G x 1/2"/10 Pack PE Bag
161291281ml, 29G x 1/2"/10 Pack PE Bag
165305161.0ml, 30G x 5/16"/1 ea. Blister Pack
16530121.0ml, 30G x 1/2"/ 1 ea. Blister Pack
16529121.0ml, 29G x 1/2"/ 1 ea. Blister Pack

Sol-M TB Syringes with Fixed Needle

Sol-M 0.5 ml TB Syringes with Fixed Needle
REF/Catalog #Description
1805270.5ml, 27G x 1/2"/1 each Blister Pack
180526IDB0.5ml, 26G x 3/8" IDB/1 each Blister Pack
180527T0.5ml, 27G x 1/2"/25 each Tray Pack
180527IDB0.5ml, 27G x 3/8"/25 Tray Pack
Sol-M 1 ml TB Syringes with Fixed Needle
REF/Catalog #Description
1810271ml, 27G x 1/2"/1 ea. Blister Pack
181026IDB1ml, 26G x 3/8"IDB /1 ea. Blister Pack
181026T1ml, 26G x 1/2"/25 each Tray Pack
181126IDB1ml, 26G x 1/2"IDB/25 each Tray Pack
181026IDB1ml, 26G x 3/8”IDB/25 each Tray Pack
181027T1ml, 27G x 1/2"/25 each Tray Pack
181127IDB1ml, 27G x 1/2"IDB/25 each Tray Pack
181027IDB1ml, 27G x 3/8”IDB/25 each Tray Pack
181023T1ml. 23G x 1/2"/25 ea. Blister Pack

The Sol-M Syringes are sterilized by Ethylene Oxide Gas and supplied sterile in blister pack, tray pack or 10 pack.

{7}------------------------------------------------

  • E. Statement of Indications for Use

Sol-M Insulin Syringes

The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.

Sol-M TB Syringe

The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

  • F. Comparison of Required Technological Characteristics:
    Information was submitted to demonstrate that there are no significant differences in technological characteristics between the Sol-M Insulin and KDL Insulin Syringes and the Sol-M TB and KDL Hypodermic Syringes. The following comparison chart shows that the subject device and the predicate device are substantially equivalent:

Sol-M Insulin and TB Syringe versus the KDL Syringes and Needle

ELEMENT OFCOMPARISONSol-M Insulin andTB SyringeKDL Syringes and Needle
Syringe typePistonPiston
InsulinInsulin
Indications forUseThe Sol-M Insulin Syringewith needle, with thecalibration unit of insulin forU-100, is a device intendedfor medical purposes for themanual aspiration of insulin,and for the injection of insulininto parts of the body belowthe surface skin.The sterile Insulin Syringefor single use with needle is adevice intended for medicalpurposes for themanual aspiration of insulin,and for the injection into partsof the body below the surfaceskin
General UseGeneral Use
Indications forUseThe Sol-M TB Syringe isintended to be used formedical purposes to injectfluids into or withdraw fluidsfrom the body.The Hypodermic Syringesare intended to be used formedical purposes to injectfluids into or withdrawfluids from the body.
Principle ofOperationThree piece piston syringe.Plunger is used to fill syringeas well as discharge the fluid.Three piece piston syringe.Plunger is used to fill syringeas well as discharge thefluid.
Specific DrugUseSol-M Insulin syringes -InsulinSol-M TB syringes - GeneralUseKDL Insulin syringes -InsulinKDL Hypodermic syringes -General Use
Tip TypeFixed NeedleFixed Needle
VolumeInsulin Syringes - 0.3, 0.5and 1mlTB Syringes - 0.5 and 1 mlInsulin Syringes - 0.3, 0.5and 1mlHypodermic Syringes – 0.5and 1 ml
GradationsLegibilityBold markingsBold markings
LubricantCompositionDow Corning MedicalSilicone 360Dow Corning MedicalSilicone 360
Lubricantamount/cm21ml-0.078mg/cm21 ml – 0.078 Mg/cm2
BarrelTransparencyClearClear
DeliveryAccuracyInsulin Syringes- Per ISO8537:2007Insulin Syringes- Per ISO8537:2007
Hypodermic Syringes - Per ISO 7886-1:1993Hypodermic Syringes - Per ISO 7886-1:1993
Hub/NeedleBond StrengthPer ISO 7886Per ISO 7886
BiocompatibilityPer ISO 10993-1Per ISO 10993-1
MaterialsBarrel, Plunger and Cap –PolypropyleneGasket - SantopreneBarrel, Plunger and Cap –PolypropyleneGasket - Santoprene
LabelingPer 21 CFR 801Per 21 CFR 801

SIDE BY SIDE COMPARISON TABLE

{8}------------------------------------------------

{9}------------------------------------------------

  • G. Summary and Conclusion of Nonclinical and Clinical Tests:
    Bench tests were conducted to verify that the proposed devices met all design specifications and were Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed devices comply with the following standards:

Sol-M TB Syringes

  • ISO 7886-1:1993 Sterile Hypodermic syringes for single use Part 1: Syringes for ● manual use;
  • ISO 9626:1991 Stainless steel needle tubing for the manufacture of medical ● devices/Amendment:2001

Sol-M Insulin Syringes

  • ISO 8537:2007 Sterile single use syringes, with and without needles, for insulin; ●
  • ISO 9626:1991 Stainless steel needle tubing for the manufacture of medical ● devices/Amendment:2001
  • H. Discussion of Clinical Tests:

None submitted

{10}------------------------------------------------

  • Substantially Equivalent Conclusion: I.
    The proposed device, Sol-M TB Syringes (various sizes), has been determined to be substantially equivalent to the predicate device, Sterile Hypodermic Syringe for single use, with needle, (various sizes), 510(k) # K112057.

The proposed device, Sol-M Insulin Syringes (various sizes), is determined to be substantially equivalent (SE) to the predicate device, Sterile Insulin Syringe for single use with needle, 510(k) # K112057.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).