Search Results
Found 2 results
510(k) Data Aggregation
(261 days)
K100130 PATHFAST® cTnI-II IFU, K 100130
PATHFAST® hs-cTnl-II is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnl) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction (AMI). PATHFAST® hs-cTnI-II is for use in clinical laboratory or point of care (POC) settings.
The PATHFAST® hs-cTnI-II test is a chemiluminescent enzyme immunoassay performed on the PATHFAST® instrument. Patient samples, whole blood or plasma, are dispensed by the operator into the designated area on the reagent cartridge. The instrument combines the patient sample, the antibody coated magnetic particles, and the alkaline phosphatase conjugate and incubates the mixture for 5 minutes at 37℃. During this incubation, the analyte in the patient sample binds to the antibody on the coated particles, and the alkaline phosphatase conjugate binds to the analyteantibody coated-particle. After the incubation, the instrument performs Bound/Free (B/F) separation using Magtration® technology to remove any excess unbound reagents. The chemiluminescent substrate is then added. The substrate is catalyzed by the bound alkaline phosphatase, which results in emission of photons. The photo-multiplier tube in the PATHFAST® instrument detects the photons that are emitted during the reaction. The chemiluminescent count is converted to analyte concentration values by the instrument based on the master calibration curve for the reagent lot. The PATHFAST® hs-cTnI-II test is supplied in reagent kits. Each kit contains sufficient materials for 60 determinations. The calibrator materials are included with the reagent kit and are also available separately. Calibration kits and diluent kits are also provided separately.
The provided text describes the regulatory clearance of the PATHFAST® hs-cTnI-II device, a diagnostic test for cardiac Troponin I (cTnI), and its comparison to a legally marketed predicate device (PATHFAST® cTnI-II, K100130). The focus of this document is on the analytical performance and how modifications to the reporting range do not affect clinical performance, relying on data from the predicate device.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
For this device, the "acceptance criteria" are tied to its analytical performance and how it meets the claims for its intended use, especially concerning linearity, detection limits, and maintaining the clinical significance established by the predicate.
| Acceptance Criterion | Reported Device Performance (PATHFAST® hs-cTnI-II) |
|---|---|
| Linearity/Reportable Range | Linearity for the interval from 4.1 to 50,000 ng/L, with deviations from linearity within ±10%. (Maximum absolute deviation observed was 9.7%). |
| Limit of Blank (LoB) | 1.466 ng/L (for EDTA WB, Plasma, LiHep WB, LiHep Plasma) |
| Limit of Detection (LoD) | 2.991 ng/L (EDTA WB), 2.958 ng/L (EDTA Plasma), 2.942 ng/L (LiHep WB), 3.002 ng/L (LiHep Plasma) |
| Limit of Quantitation (LoQ) | 4.1 ng/L (for EDTA WB, Plasma, LiHep WB, LiHep Plasma) - defined as the minimum cTnI concentration with % CV < 20 %. |
| 99th Percentile Cutoff | Remains at 29 ng/L (as established by the predicate device and re-evaluated with the new LoQ). |
| Measurable Concentrations in Healthy Individuals | Continues to meet the criteria that measurable concentrations are above the limit of detection for over 50% of healthy individuals. |
| Clinical Sensitivity and Specificity | Remains unchanged from the predicate device. |
2. Sample size used for the test set and the data provenance
The document details analytical studies for linearity and detection limits, but does not describe a specific "test set" for clinical performance in the context of a new study for the PATHFAST® hs-cTnI-II. Instead, it refers to the re-evaluation of data from the predicate device (K100130).
-
Linearity/Assay Reportable Range:
- Sample Size: Unknown specific number of unique whole blood and plasma samples (EDTA and lithium heparin) with low and high values, diluted to produce up to 12 dilution levels per series. Three dilution series of each sample matrix were tested across three different lots of reagents. Each level was tested in replicates of three.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This is an analytical study, not a clinical study on patient outcomes.
-
LoB/LoD Studies:
- Sample Size for LoB: Four plasma samples, replicates of two per sample, repeated twice a day for three consecutive business days, across four lots of reagents. Total 96 determinations (4 lots * 4 samples * 2 replicates * 3 days * 2 times/day, but text says 24 replicates per lot, so 4 lots * 24 = 96).
- Sample Size for LoD: 4 low samples, twice a day with four lots of reagent over three days. Total 96 replicates.
- Data Provenance: Not explicitly stated.
-
LoQ Study:
- Sample Size: Two sets of three samples and two reagent lots; each tested in replicates of four, three runs per day. Total 24 replicates per sample per condition. Testing was repeated for each matrix.
- Data Provenance: Not explicitly stated.
-
Clinical Performance (Reference to K100130): The document indicates that the clinical performance evaluation relied on data from the predicate device (K100130). The original reference range study that established the 99th percentile cutoff used numerical results, which were re-evaluated.
- Sample Size: Not specified for the original predicate study within this document.
- Data Provenance: Not specified within this document, but pertains to the predicate device's original submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This document does not describe a new clinical study involving experts to establish a "ground truth" for a test set. The clinical claims (99th percentile cutoff, sensitivity, specificity) rely on the original predicate device's studies, where ground truth would have been established through clinical diagnosis of Acute Myocardial Infarction.
4. Adjudication method for the test set
Not applicable. No new clinical test set with human interpretation and adjudication is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an in vitro diagnostic test for measuring a biomarker, not an imaging device or algorithm requiring human interpretation or AI assistance in that context.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to the analytical performance of the PATHFAST® hs-cTnI-II instrument and reagents. The "standalone" performance is effectively captured by the linearity, LoB, LoD, and LoQ studies, which measure the assay's ability to quantitatively measure cTnI in samples directly, without human interpretation of subjective outputs. The performance described in the table above reflects this standalone analytical performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the analytical studies described in this document (linearity, LoB, LoD, LoQ), the "ground truth" is established through highly controlled laboratory experiments using known concentrations or dilutions, not clinical outcomes or expert consensus on patient cases.
For the clinical claims (99th percentile, clinical sensitivity/specificity), the document refers back to the predicate device (K100130). In such diagnostic tests for AMI, the ground truth for establishing clinical sensitivity and specificity would typically involve:
- Clinical Diagnosis of AMI: Based on established clinical criteria, often including troponin levels, ECG changes, and clinical symptoms, potentially adjudicated by a panel of cardiologists.
- Outcomes Data: Confirmation of AMI or exclusion of AMI through follow-up data.
The specifics of how this ground truth was established for K100130 are not detailed in this document.
8. The sample size for the training set
Not applicable. This device is a quantitative immunoassay, not an AI/ML device that requires a training set in the typical sense for algorithm development.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for an AI/ML algorithm.
Ask a specific question about this device
(385 days)
The PATHFAST® cTnl Calibrators are for calibration of the PATHFAST® system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma.
The PATHFAST cTnl Calibrators are for calibration of the PATHFAST cTnl-II test performed on the PATHFAST instrument. PATHFAST cTnI-II test is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnl) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction.
Currently, PATHAST cTnl Calibrators 1 and 2 are provided as lyophilized products of two vials each. Calibrator 1 consists of buffer only and Calibrator 2 consists of cTnI complex in buffer. Four vials of calibrator diluent also are provided for reconstitution of the calibrators. The diluent consists of an aqueous solution with 0.05% sodium azide. This Special 510(k) is being submitted for a change to a liquid Calibrator 1 formulation and a new dropper bottle container. There are no changes to PATHFAST cTnl Calibrator 2. As a result of elimination of the reconstitution step for PATHFAST cTnI Calibrator 1, the number of bottles of Calibrator Diluent is being reduced from four bottles to two bottles.
The provided document describes the PATHFAST® cTnI Calibrator 1, which underwent a Special 510(k) submission for a change in formulation from lyophilized to liquid and a new container type. The purpose of the submission is to demonstrate substantial equivalence to the predicate device (PATHFAST® cTnI Calibrator 1, lyophilized format, K100130).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly detailed in a quantitative table. Instead, the document states that the modified device "met all specifications for the studies identified in the company's design control system Risk Analysis." The studies performed were:
| Study Area | Performance Metric | Reported Device Performance |
|---|---|---|
| Formulation | Assay sensitivity, accuracy, precision, method comparison, sample type and matrix comparison | Met all specifications |
| Format | Real-time stability | Met all specifications |
| Container | Elution and evaporation | Met all specifications |
2. Sample Sizes Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes for the test sets used in the validation/verification studies. It only mentions the types of studies performed (assay sensitivity, accuracy, precision, method comparison, sample type and matrix comparison, real-time stability, elution, and evaporation).
The data provenance (country of origin, retrospective/prospective) is not mentioned.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. This device is a calibrator, and its performance is evaluated against established analytical specifications and comparison to a predicate device, rather than through expert-established ground truth on clinical data.
4. Adjudication Method for the Test Set
Not applicable.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is a medical device calibrator, not an AI-assisted diagnostic tool for human readers.
6. Standalone Performance
Yes, the studies conducted (formulation, format, container evaluations) represent standalone performance of the modified calibrator against its own specifications and in comparison to the predicate device's performance characteristics. The studies directly assess the analytical performance of the calibrator itself.
7. Type of Ground Truth Used
The "ground truth" for the calibrator's performance is established by:
- Analytical specifications: The device is expected to meet predefined performance targets for sensitivity, accuracy, precision, etc.
- Comparison to predicate device: The modified calibrator's performance is compared to that of the legally marketed predicate device (lyophilized format) to demonstrate substantial equivalence.
- Reference materials/methods: Implicitly, the accuracy and precision studies would rely on reference materials or methods to determine true values.
8. Sample Size for the Training Set
Not applicable. This device is a calibrator and does not involve AI/machine learning models that require training sets in the conventional sense. Its purpose is to calibrate an existing diagnostic system, not to perform a diagnostic function itself based on learned patterns.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
Ask a specific question about this device
Page 1 of 1