K Number
K231974
Manufacturer
Date Cleared
2024-03-20

(261 days)

Product Code
Regulation Number
862.1215
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PATHFAST® hs-cTnl-II is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnl) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction (AMI). PATHFAST® hs-cTnI-II is for use in clinical laboratory or point of care (POC) settings.

Device Description

The PATHFAST® hs-cTnI-II test is a chemiluminescent enzyme immunoassay performed on the PATHFAST® instrument. Patient samples, whole blood or plasma, are dispensed by the operator into the designated area on the reagent cartridge. The instrument combines the patient sample, the antibody coated magnetic particles, and the alkaline phosphatase conjugate and incubates the mixture for 5 minutes at 37℃. During this incubation, the analyte in the patient sample binds to the antibody on the coated particles, and the alkaline phosphatase conjugate binds to the analyteantibody coated-particle. After the incubation, the instrument performs Bound/Free (B/F) separation using Magtration® technology to remove any excess unbound reagents. The chemiluminescent substrate is then added. The substrate is catalyzed by the bound alkaline phosphatase, which results in emission of photons. The photo-multiplier tube in the PATHFAST® instrument detects the photons that are emitted during the reaction. The chemiluminescent count is converted to analyte concentration values by the instrument based on the master calibration curve for the reagent lot. The PATHFAST® hs-cTnI-II test is supplied in reagent kits. Each kit contains sufficient materials for 60 determinations. The calibrator materials are included with the reagent kit and are also available separately. Calibration kits and diluent kits are also provided separately.

AI/ML Overview

The provided text describes the regulatory clearance of the PATHFAST® hs-cTnI-II device, a diagnostic test for cardiac Troponin I (cTnI), and its comparison to a legally marketed predicate device (PATHFAST® cTnI-II, K100130). The focus of this document is on the analytical performance and how modifications to the reporting range do not affect clinical performance, relying on data from the predicate device.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

For this device, the "acceptance criteria" are tied to its analytical performance and how it meets the claims for its intended use, especially concerning linearity, detection limits, and maintaining the clinical significance established by the predicate.

Acceptance CriterionReported Device Performance (PATHFAST® hs-cTnI-II)
Linearity/Reportable RangeLinearity for the interval from 4.1 to 50,000 ng/L, with deviations from linearity within ±10%. (Maximum absolute deviation observed was 9.7%).
Limit of Blank (LoB)1.466 ng/L (for EDTA WB, Plasma, LiHep WB, LiHep Plasma)
Limit of Detection (LoD)2.991 ng/L (EDTA WB), 2.958 ng/L (EDTA Plasma), 2.942 ng/L (LiHep WB), 3.002 ng/L (LiHep Plasma)
Limit of Quantitation (LoQ)4.1 ng/L (for EDTA WB, Plasma, LiHep WB, LiHep Plasma) - defined as the minimum cTnI concentration with % CV < 20 %.
99th Percentile CutoffRemains at 29 ng/L (as established by the predicate device and re-evaluated with the new LoQ).
Measurable Concentrations in Healthy IndividualsContinues to meet the criteria that measurable concentrations are above the limit of detection for over 50% of healthy individuals.
Clinical Sensitivity and SpecificityRemains unchanged from the predicate device.

2. Sample size used for the test set and the data provenance

The document details analytical studies for linearity and detection limits, but does not describe a specific "test set" for clinical performance in the context of a new study for the PATHFAST® hs-cTnI-II. Instead, it refers to the re-evaluation of data from the predicate device (K100130).

  • Linearity/Assay Reportable Range:

    • Sample Size: Unknown specific number of unique whole blood and plasma samples (EDTA and lithium heparin) with low and high values, diluted to produce up to 12 dilution levels per series. Three dilution series of each sample matrix were tested across three different lots of reagents. Each level was tested in replicates of three.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This is an analytical study, not a clinical study on patient outcomes.
  • LoB/LoD Studies:

    • Sample Size for LoB: Four plasma samples, replicates of two per sample, repeated twice a day for three consecutive business days, across four lots of reagents. Total 96 determinations (4 lots * 4 samples * 2 replicates * 3 days * 2 times/day, but text says 24 replicates per lot, so 4 lots * 24 = 96).
    • Sample Size for LoD: 4 low samples, twice a day with four lots of reagent over three days. Total 96 replicates.
    • Data Provenance: Not explicitly stated.
  • LoQ Study:

    • Sample Size: Two sets of three samples and two reagent lots; each tested in replicates of four, three runs per day. Total 24 replicates per sample per condition. Testing was repeated for each matrix.
    • Data Provenance: Not explicitly stated.
  • Clinical Performance (Reference to K100130): The document indicates that the clinical performance evaluation relied on data from the predicate device (K100130). The original reference range study that established the 99th percentile cutoff used numerical results, which were re-evaluated.

    • Sample Size: Not specified for the original predicate study within this document.
    • Data Provenance: Not specified within this document, but pertains to the predicate device's original submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This document does not describe a new clinical study involving experts to establish a "ground truth" for a test set. The clinical claims (99th percentile cutoff, sensitivity, specificity) rely on the original predicate device's studies, where ground truth would have been established through clinical diagnosis of Acute Myocardial Infarction.

4. Adjudication method for the test set

Not applicable. No new clinical test set with human interpretation and adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an in vitro diagnostic test for measuring a biomarker, not an imaging device or algorithm requiring human interpretation or AI assistance in that context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This refers to the analytical performance of the PATHFAST® hs-cTnI-II instrument and reagents. The "standalone" performance is effectively captured by the linearity, LoB, LoD, and LoQ studies, which measure the assay's ability to quantitatively measure cTnI in samples directly, without human interpretation of subjective outputs. The performance described in the table above reflects this standalone analytical performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the analytical studies described in this document (linearity, LoB, LoD, LoQ), the "ground truth" is established through highly controlled laboratory experiments using known concentrations or dilutions, not clinical outcomes or expert consensus on patient cases.

For the clinical claims (99th percentile, clinical sensitivity/specificity), the document refers back to the predicate device (K100130). In such diagnostic tests for AMI, the ground truth for establishing clinical sensitivity and specificity would typically involve:

  • Clinical Diagnosis of AMI: Based on established clinical criteria, often including troponin levels, ECG changes, and clinical symptoms, potentially adjudicated by a panel of cardiologists.
  • Outcomes Data: Confirmation of AMI or exclusion of AMI through follow-up data.

The specifics of how this ground truth was established for K100130 are not detailed in this document.

8. The sample size for the training set

Not applicable. This device is a quantitative immunoassay, not an AI/ML device that requires a training set in the typical sense for algorithm development.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION".

March 20, 2024

PHC Corporation Helen Landicho, RAC SVP Regulatory Affairs Polymedco, Inc. 510 Furnace Dock Road Cortlandt Manor, New York 10567

Re: K231974

Trade/Device Name: PATHFAST®hs-cTnI-II Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI Dated: March 1, 2024 Received: March 4, 2024

Dear Helen Landicho:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Paula V. Caposino -S

Paula Caposino, Ph.D. Acting Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K231974

Device Name PATHFAST®hs-cTnI-II

Indications for Use (Describe)

PATHFAST® hs-cTnl-II is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnl) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction (AMI). PATHFAST® hs-cTnI-II is for use in clinical laboratory or point of care (POC) settings.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K231974 PATHFAST® hs-cTnI-II High Sensitive Troponin

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Federal Food, Drug, and Cosmetic Act and 21 CFR 807.92.

510(k) Owner:

PHC Corporation 1460-6 Aza-Mitodai, Mito, Tako-machi Katori-gun, Chiba 289-2247, Japan Misato Igarashi Ph.D. 81-3-6400-2115 (Tel) 81-3-5577-0451(Fax

Contact Person: Polymedco, Inc. Helen Landicho, RAC 914.293.1605 (Tel)

Name of the device

Trade name: PATHFAST® hs-cTnI-II Common name: high sensitive Troponin I Classification: 21 CFR 862.1215 Creatine Phosphokinase/creatine kinase or isoenzymes test system Product code: MMI

Legally marketed device claiming equivalence:

PATHFAST® cTnI-II (K100130)

Device description:

The PATHFAST® hs-cTnI-II test is a chemiluminescent enzyme immunoassay performed on the PATHFAST® instrument.

Patient samples, whole blood or plasma, are dispensed by the operator into the designated area on the reagent cartridge. The instrument combines the patient sample, the antibody coated magnetic particles, and the alkaline phosphatase conjugate and incubates the mixture for 5 minutes at 37℃. During this incubation, the analyte in the patient sample binds to the antibody on the coated particles, and the alkaline phosphatase conjugate binds to the analyteantibody coated-particle.

After the incubation, the instrument performs Bound/Free (B/F) separation using Magtration® technology to remove any excess unbound reagents. The chemiluminescent

{5}------------------------------------------------

510(k) Submission PATHFAST® hs-cTnI-II

substrate is then added. The substrate is catalyzed by the bound alkaline phosphatase, which results in emission of photons.

The photo-multiplier tube in the PATHFAST® instrument detects the photons that are emitted during the reaction. The chemiluminescent count is converted to analyte concentration values by the instrument based on the master calibration curve for the reagent lot.

The PATHFAST® hs-cTnI-II test is supplied in reagent kits. Each kit contains sufficient materials for 60 determinations. The calibrator materials are included with the reagent kit and are also available separately. Calibration kits and diluent kits are also provided separately.

ComponentQuantity
Reagent Cartridge6 cartridges x 10 trays
Calibrator 12 vials
Calibrator 22 vials
Calibrator diluent4 vials of 1.0 mL each

Contents of the PATHFAST® hs-cTnI-II reagent kit

Reagent Cartridge: The reagent cartridge contains 16 wells 1, 6, 8, 9, 10, 12, 14, 15, 16 are empty. The other wells are filled with the following reagents:

Reagent DescriptionVolumeCartridgeWell
Alkaline phosphatase (calf intestine) conjugated anti cTnImonoclonal antibody (mouse) in MES buffer (pH 6.0) with0.007% zinc chloride, and 0.06% sodium azide aspreservative50 µl2
Washing Buffer: Tris buffer (pH 7.5) with 0.05% sodiumazide as preservative400 µl3, 4, 5
Magnetic particles coated with anti cTnI monoclonalantibody (mouse) in MOPS buffer50 µl
Sample Dilution Buffer: Tris buffer (pH 8.2) with 0.05%sodium azide as preservative25 ul11
Chemiluminescent substrate: CDP-Star100 µl13

Contents of the PATHFAST® hs-cTnI-II reagent cartridge

Calibrator 1: Lyophilized preparation containing MES pH 6.0, lactose, and enzyme free human serum. DTT

Calibrator 2: Lyophilized preparation containing cTnI complex, MES pH 6.0, lactose, and enzyme free human serum, DTT

Calibrator diluent: Aqueous solution with 0.05% sodium azide used for reconstituting Calibrators 1 and 2

Calibrator 1 and Calibrator 2 contain human serum obtained from donors who were confirmed negative for anti-HIV-1/2, HbsAg and Anti-HCV.

{6}------------------------------------------------

Intended use:

PATHFAST® hs-cTnl-II is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I ( c Tnl) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction (AMI). PATHFAST® hs-cTnl-II is for use in clinical laboratory or point of care (POC) settings.

The indications for use are identical to the predicate assay. The PATHFAST® hs-cTnl-II modified the reporting units from ng/mL to the conventional ng/L and extended the reportable range from 4.1 ng/L to 50,000 ng/L. The modification of the units and reportable range of the device does not affect the diagnostic or clinical utility of the device or the safety and effectiveness of the device when used as labeled.

Similarities
ItemCandidate DevicePATHFAST hs-cTnI-IIPredicate DevicePATHFAST cTnI-IIK100130
Intended UseAssist in the diagnosis ofacute myocardialinfarction. For use inclinical laboratory orpoint of care (POC)settings.Same
Storage2-8° C2-8° C
Calibration Levels66
MethodologyChemiluminescent enzymeimmunoassaySame
Indications for useAssist in the diagnosis ofacute myocardialinfarction. For use inclinical laboratory or pointof care (POC) settings. Notfor risk stratificationSame
Sample TypesEDTA and Lithiumheparinwhole blood and plasmaSame
ItemDifferences
Candidate DevicePATHFAST hs-cTnI-IIPredicate DevicePATHFAST cTnI-IIK100130
Reportable range4.1 to 50,000 ng/L0.019 to 50 ng/mL

Comparison with the predicate device

Modification of the reporting units from ng/mL to ng/L meets conventional laboratory reporting for troponin.

{7}------------------------------------------------

Non clinical testing:

Precision, reproducibility, point of care precision reproducibility studies, high dose hook, traceability, stability, expected values, analytical specificity, assay cut-off, comparison studies, matrix studies, clinical sensitivity, clinical specificity, clinical cut-off and reference ranges are reported in K100130 PATHFAST® cTnI-II IFU.

Linearity/assav reportable range:

Linearity studies were conducted on EDTA and lithium heparin whole blood and plasma samples. For the PATHFAST® hs-cTnI-II, the measurement procedure shows linearity for the interval from 4.1 to 50.000 ng/L, with deviations from linearity within +10%.

EDTA whole blood and plasma samples, and lithium heparin whole blood and plasma samples with low and high values were diluted to produce up to 12 dilution levels per each unique series with values ranging from 1 to 64,500 ng/L. Three-dilution series of each sample matrix were tested across three different lots of reagents. Each level was tested in replicates of three. The acceptable coefficient of variation for the replicates was set at <20%. The relative concentration, measured values, predicted values, deviations from linearity and percent deviations were analyzed for each sample in a series. The %CV for the replicates was set at <20% and the allowable deviation from linearity was acceptable within +10%. The maximum absolute deviation observed in the linearity study across all sample matrices was 9.7%

{8}------------------------------------------------

Detection limit:

LoB/LoD studies were conducted using EDTA and lithium heparin whole blood and plasma samples. The LoB was determined by testing four lots of reagents, four plasma samples, replicates of two per sample, repeated twice a day for three consecutive business days. Each lot had a total of 24 replicates for a combined number of 96 determinations. The LoD for EDTA and lithium heparin whole blood and plasma was determined by testing 4 low samples, twice a day with four lots of reagent over three days for a total of 96 replicates.

The LoQ study was conducted using EDTA and lithium heparin whole blood and plasma samples at the estimated concentration where the coefficient of variation for the replicates was < 20%. Two sets of three samples and two reagent lots; each tested in replicates of four, three runs per day, for a total of 24 replicates per sample. Testing was repeated for each matrix. The coefficient of variations and means were calculated and pooled for each sample per reagent lot. The LoQ for EDTA and lithium heparin whole blood and plasma was determined by averaging the values over two lots. Lowest hs-cTnI concentration of the samples that showed less than 20 % CV.

Summary:

EDTALiHep
WBPlasmaWBPlasma
LoB1.4661.4661.4661.466
LoD2.9912.9582.9423.002
LoQ4.14.14.14.1

LoD = LoB + coSDt, where co =1.645/1-(1/4*(L - J))

C LoQ = minimum cTnI concentration with % CV < 20 %

The measuring range of the cTnI-II assay is from 4.1 ng/L to 50,000 ng/L.

The analytical information in this premarket notification supports the measuring range of the PATHFAST® hs-cTnI of 4.1 ng/L to 50,000 ng/L

Clinical performance:

The effects of the modification of the LoQ were evaluated on the previously reported clinical claims from the predicate device. During the predicate reference range study used to establish the 99th percentile cutoff all values below 0.019 ng/mL were reported as numerical results. The data set used to non-parametrically calculate the 99th percentile consisted entirely of numerical results. The predicate LoQ was the lowest concentration at which there was an imprecision of 10% CV; the updated LoQ is the lowest concentration at which there was an imprecision of 20% CV, which is 4.1 ng/L.

There is no impact on the 99th percentile cutoff of the reference range validated in K 100130 with the updated LoQ of 4.1 ng/L, because the original numerical results for all testing were previously used in the calculation. The PATHFAST® hs-cTnI-II assay continues to meet the criteria that measurable concentrations are above the limit of detection for over 50% of healthy individuals. The 99th percentile cutoff remains at 29 ng/L.

{9}------------------------------------------------

510(k) Submission PATHFAST® hs-cTnI-II

The clinical sensitivity and specificity of the PATHFAST® hs-cTnI-II assay is not affected by lowering the LoQ to 4.1 ng/L, as all the numerical data was included in the predicate calculations. The clinical sensitivity and specificity remain unchanged when lowering the LoQ to 4.1 ng/L.

The evidence supports that the non-parametrically calculated cutoff at the 99th percentile remains at 29 ng/L and the assay continues to meet the criteria that measurable concentrations are above the limit of detection for over 50% of healthy individuals. The clinical sensitivity and specificity of the assay remain unchanged from the predicate device.

Summary conclusion:

The summary includes conclusions drawn from the nonclinical tests that demonstrate that the device is safe, effective and performs as wells as or better than the predicate device. The information provided in this submission is complete and demonstrates that the subject device is substantially equivalent to the predicate device.

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.