Search Results
Found 2 results
510(k) Data Aggregation
(267 days)
The AnyOne External Implant System is intended to be surgically placed in the maxillary or mandibular molar areas for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patient's chewing function. Smaller implants (less than 6.0 mm) are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger implants are dedicated for the molar region and are indicated for delayed loading.
AnyOne External Implant System is an integrated system of endosseous dental implants which designed to support prosthetic devices for partially or fully edentulous patients and consists of fixtures and abutments. The dental implants which used in conjunction with other prosthesis restore the lost chewing ability and improve the appearance. AnyOne External Fixture is a substructure of a dental implant system made of CP Ti Grade 4 with the surface treated by SLA method. It is placed in the anterior or posterior site of maxillary or mandibular jawbone considering bone quality and bone quantity. Dental prosthesis is a superstructure of a dental implant system and connecting elements between the dental implant and the crown. It is made of Ti-6A1-4V ELI, Gold alloy, CCM alloy and intended to be placed on the fixture allows single & multiple prosthetic restore a patient's chewing function.
Here's a breakdown of the acceptance criteria and study information for the AnyOne External Implant System, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission details the equivalence to predicate devices rather than explicit, stand-alone acceptance criteria with specific numerical thresholds for each component. However, the core acceptance criteria revolve around demonstrating substantial equivalence to existing, legally marketed devices (predicates and reference devices). The reported performance across all components is that they met the pre-set criteria for the conducted tests.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Biocompatibility (following ISO 10993-1) Identical material, manufacturing, and patient-contacting parts as predicate device | The subject device has the same material composition, manufacturing process, and patient contacting parts as predicate and reference devices, indicating it meets biocompatibility. |
Surface Treatment (following 'Section 11 of Class II Special Controls Guidance Document...') Identical surface treatment and manufacturing as predicate device | The subject device has the same surface treatment (S.L.A for fixture, Anodizing for abutment, TiN coating for abutment) and manufacturing process as predicate and reference devices. |
Pyrogen and Endotoxin Levels (following USP 39 ) Endotoxin levels below 0.5 EU/mL | Endotoxin testing will be conducted on every batch with a testing limit of below 0.5 EU/mL. (Note: This is a statement of intent for ongoing quality control, not a direct performance result from the validation study, but implies the acceptance criterion). |
Sterility Assurance Level (SAL) (following ISO 11137 and ISO 17665-1, 2) SAL of 10⁻⁶ | Sterilization validation tests achieved a sterility assurance level of 10⁻⁶. |
Shelf Life (following ASTM F1980) 5 years shelf life | Test results validated a 5-year shelf life using the accelerated aging method. |
Performance Tests (following ISO 14801 and 'Class II Special Controls Guidance Document...') Static compression-strength and fatigue tests met pre-set criteria. | The test results met the pre-set criteria for static compression-strength and fatigue tests for all relevant components, demonstrating substantial equivalence to predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the numerical sample size for the test set used in the performance bench tests (static compression-strength and fatigue tests). It mentions "test was performed on the subject device and predicate device" and "representative specimen".
- Sample Size (Test Set): Not explicitly stated as a numerical value for each test (e.g., how many implants were tested for fatigue). It generally refers to "the subject device and predicate device" and "representative specimen" in singular form, implying at least one of each worst-case scenario. However, for a regulatory submission of this nature, multiple samples would typically be tested to establish statistical confidence, even if not explicitly detailed in this summary.
- Data Provenance: The study is non-clinical bench testing. The country of origin of the data is not specified, but the applicant and correspondent are located in the Republic of Korea. The tests are prospective as they were conducted specifically for this submission to demonstrate substantial equivalence.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
This is not applicable as the studies are non-clinical bench tests (physical properties, biocompatibility, sterilization, etc.). Ground truth in this context is established by adherence to recognized international standards (ISO, ASTM, USP) and FDA guidance documents. There is no mention of human experts interpreting test results for "ground truth" in the way it might apply to a clinical study involving medical image analysis, for example.
4. Adjudication Method for the Test Set
This is not applicable for non-clinical bench testing. Passing or failing criteria are determined by adherence to established engineering and material science standards (e.g., force applied, number of cycles, endotoxin concentration limits).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This is a submission for a dental implant system involving physical components. MRMC studies are typically performed for AI/CAD systems that assist human readers in tasks like medical image interpretation.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
No. As above, this is for a physical medical device. There is no algorithm involved.
7. Type of Ground Truth Used
The "ground truth" for the non-clinical tests is based on:
- Standardized Test Methods: Adherence to international standards such as ISO 14801 for mechanical testing, ISO 10993-1 for biocompatibility, ISO 11137 and ISO 17665-1, 2 for sterilization, ASTM F1980 for shelf life, and USP 39 for endotoxin testing.
- Predicate Device Comparison: Substantial equivalence is established by demonstrating that the subject device's performance, materials, and design are comparable to legally marketed predicate and reference devices. The predicate devices themselves serve as a form of "ground truth" for acceptable performance and safety within their intended use.
8. Sample Size for the Training Set
Not applicable. There is no "training set" as this is a physical medical device, not a machine learning algorithm.
9. How the Ground Truth for the Training Set Was Established
Not applicable. See point 8.
Ask a specific question about this device
(269 days)
The Cowell Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is not intended for immediate loading.
The implants with diameters larger than 5.0mm are intended to be surgically placed in the maxillary or mandibular molar areas for the purpose of providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. These implants are intended to be used where smaller implants have failed.
The Cowell Implant System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper and / or lower jaw arches. This system is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
The provided text describes a 510(k) summary for the Cowell Implant System, a dental implant. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a study with performance metrics.
Therefore, the requested information about acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment cannot be extracted from the provided text.
The document states:
- "The Cowell Implant System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics."
- "The Cowell Implant systems are similar in fundamental scientific technology to the predicate devices (K090825, K053353) in that they all have been designed, manufactured and tested in compliance with FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutment."
- "concludes that the ball abutment is safe and effective and substantially equivalent to predicate devices as described herein."
This indicates that the device was approved based on its similarity to existing, approved devices and compliance with regulatory guidelines, rather than a standalone performance study with quantifiable acceptance criteria.
Ask a specific question about this device
Page 1 of 1