Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K133103
    Manufacturer
    Date Cleared
    2014-01-27

    (119 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aversion Pedicle Screw System is intended for posterior, noncervical (T1-S1) pedicle fixation to provide immobilization and stabilization in skeletally mature patients as an adjunct to fusion for the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, spinal stenosis, scoliosis, kyphosis, lordosis, spinal tumor, pseudarthrosis and failed previous fusion

    Device Description

    The Aversion Pedicle Screw System consists of straight and curved rods, polyaxial pedicle screws and crosslink connectors. These are available in a variety of sizes to accommodate differing patient anatomy.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device called the "Aversion Pedicle Screw System." It is not an AI/ML device, therefore the requested information (acceptance criteria, study details, sample sizes, ground truth, expert qualifications, adjudication, MRMC, or standalone performance) related to AI/ML device testing is not applicable.

    The document describes a traditional medical device, a pedicle screw system, and focuses on demonstrating its substantial equivalence to a legally marketed predicate device based on:

    1. Technological characteristics: Basic design (rod-based fixation), material (titanium alloy), anatomic location (non-cervical spine), and sizes.
    2. Performance Data: Mechanical testing (static compression bending and torsion, dynamic compression bending, and pullout testing) performed according to ASTM F1717 and ASTM F543 standards. The results demonstrated substantial equivalence in performance to the predicate devices.

    Therefore, I cannot fulfill the request for information on acceptance criteria, study details, sample sizes, ground truth, expert qualifications, adjudication, MRMC, or standalone performance as these concepts are specifically relevant to AI/ML device evaluations, which this document does not describe.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110046
    Date Cleared
    2011-12-07

    (335 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Athena Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. The system is intended for posterior, pedicle fixation as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The Athena Pedicle Screw System is comprised of polyaxial pedicle screws in various diameters and lengths, spinal rods in various lengths, cross-connectors and set screws. The Athena Pedicle Screw System can be used for single or multiple level fixations. All implant components are manufactured from titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Static and dynamic axial compression bending testing per ASTM F1717Performed; results demonstrate substantial equivalence to predicate devices.
    Static torsion testing per ASTM F1717Performed; results demonstrate substantial equivalence to predicate devices.
    Use of titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136Device manufactured from this material, consistent with predicates.
    Same intended use as predicate devicesConfirmed.
    Similar primary implant design as predicate devicesConfirmed.
    Equivalent material of manufacture as predicate devicesConfirmed.
    Overall substantial equivalence to predicate devicesConcluded based on non-clinical testing and shared characteristics.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text describes non-clinical mechanical testing of the device components. It does not refer to a "test set" in the context of patient data or clinical trials. Therefore, information regarding sample size for a test set and data provenance (e.g., country of origin, retrospective/prospective) is not applicable to this type of study as described. The testing involves physical samples of the pedicle screw system components.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    Given this is a non-clinical mechanical performance study rather than a clinical study evaluating diagnostic or treatment outcomes, there were no experts used to establish ground truth in the way one would for clinical data interpretation. The "ground truth" for this study is defined by the established standards of ASTM F1717 for mechanical testing. Compliance with these standards is objectively measured.

    4. Adjudication Method

    Since no experts were establishing ground truth for a clinical test set, an adjudication method is not applicable. The testing involves objective physical measurements against established engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This submission focuses on the mechanical substantial equivalence of a medical implant, not the effectiveness of a diagnostic or treatment method involving human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone "algorithm only" performance study was not done. This is a physical medical device, not a software algorithm. The "standalone performance" here refers to the mechanical integrity of the device components in isolation (i.e., without being implanted in a patient, but using standardized test fixtures). The study conducted was precisely that – standalone mechanical testing of the device components.

    7. The Type of Ground Truth Used

    The "ground truth" used for this study is based on established engineering standards (ASTM F1717) for mechanical performance and material specifications (ASTM F136). The performance of the Athena Pedicle Screw System was compared directly against these standards and to the mechanical performance of predicate devices that previously met these standards.

    8. The Sample Size for the Training Set

    Since this is a non-clinical mechanical study rather than a machine learning or clinical study, the concept of a "training set" is not applicable. The device's design is developed through engineering principles, not through training on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    As the concept of a "training set" does not apply to this type of study, the question of how ground truth was established for a training set is not applicable. The design and manufacturing processes are validated against engineering specifications and industry standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102853
    Date Cleared
    2011-01-25

    (118 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The flamenco™ is a spinal fixation system intended to provide immobilization and stabilization of thoracic, lumbar, and sacral spinal segments as an adjunct to fusion.

    When used as posterior, pedicle screw fixation, the system is intended for the treatment of severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients. In addition, when used as a pedicle screw fixation system in the non-cervical posterior spine (T1 to S2), the system is intended for the treatment of the following acute and chronic instabilities or deformities: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, spinal stenosis, scoliosis, kyphosis, lordosis, spinal tumor, pseudarthrosis and failed previous fusion in skeletally mature patients.

    When used as a posterior, non-cervical, hook and/or sacral/iliac fixation system (i.e., nonpedicle screw), the flamenco™ is intended for the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (as conformed by patient history and radiographic studies), spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, lordosis, spinal stenosis, spinal tumor, pseudarthrosis and failed previous fusion.

    Device Description

    The flamenco™ consists of rods, monoaxial and polyaxial pedicle screws, hooks and transverse connectors with locking set screws. The components are available in various sizes to accommodate differing patient anatomy. Rods are available in one diameter and a variety of lengths. Monoaxial and polyaxial screws are available in a variety of diameter-length combinations. Hooks are offered in a variety shapes and sizes.

    AI/ML Overview

    The provided text describes the "flamenco™ spinal fixation system," a Class II medical device. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a comprehensive study report for standalone performance or comparative effectiveness with AI.

    Based on the provided information, the following can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical integrity under "worst case" conditions based on ASTM F1717 for static and dynamic tests."The mechanical test results demonstrated that flamenco™ performs as well as or better than the predicate devices."

    Note: The acceptance criteria are implied to be achieving performance at least on par with the predicate devices under the specified ASTM F1717 mechanical tests.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document reports mechanical performance data, not clinical performance data from a test set of patient cases. Therefore, information regarding "sample size used for the test set" (referring to patient data) and "data provenance" (country of origin, retrospective/prospective) is not applicable in this context. The testing was laboratory-based mechanical testing of "worst case" device constructs.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    As this is a mechanical testing study, there were no human experts involved in establishing "ground truth" in the clinical sense. The "ground truth" for the mechanical tests would be the physical properties and performance measured by standardized mechanical testing equipment.

    4. Adjudication Method for the Test Set:

    Not applicable, as this was a mechanical test, not a clinical study requiring adjudication of expert interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This device is a spinal fixation system, not an AI diagnostic or therapeutic software that would typically be evaluated with MRMC studies comparing human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, in spirit, a standalone performance was done for the device's mechanical properties. The mechanical tests were performed on the device "constructs" themselves, independent of human interaction during the test. This is not "standalone algorithm performance" in the context of AI, but rather standalone performance of the physical device.

    7. The Type of Ground Truth Used:

    The "ground truth" for the mechanical performance study was physical measurements and adherence to engineering standards (ASTM F1717). The performance was compared against the measured performance of predicate devices.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical medical implant, not an AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for a physical device.

    Summary of Study:

    The study described is a mechanical performance study conducted according to ASTM F1717. This standard specifies methods for static and dynamic testing of spinal implant constructs. The "flamenco™" system's "worst case" construct was subjected to static compression bending, torsion, and dynamic compression bending. The objective of the study was to demonstrate that the mechanical performance of the flamenco™ system is equivalent to or superior to that of legally marketed predicate devices. The study concluded that the "flamenco™ performs as well as or better than the predicate devices," thereby meeting the implied acceptance criteria for mechanical equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081108
    Date Cleared
    2008-07-14

    (87 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K974749, K031585, K053573, K061577

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MonoPoly Pedicle Screw System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar and/or sacral spine (T1-S1), specifically as follows:

    When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the is indicated for one or more of the following: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) (2) degenerative spondylolisthesis with objective evidence of neurologic impairment, (4) dislocation, (5) spinal tumor, and/or (6) failed previous fusion (pseudarthrosis).

    When used as a non-cervical and non-pedicle screw fixation system, the MonoPoly Pedicle Screw System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) trauma (fracture, dislocation), (5) pseudarthrosis, (6) tumor resection and/or (7) failed previous fusion.

    Device Description

    The MonoPoly Pedicle Screw System is comprised of a variety of monoaxial and polyaxial pedicle screws sizes, couplers, a set screw, cross links, a washer, rods and hooks. All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-Eli).

    AI/ML Overview

    This document describes a medical device, the MonoPoly Pedicle Screw System, and its regulatory filing, but it does not contain acceptance criteria for device performance or a study demonstrating that the device meets such criteria.

    Instead, the provided text is a 510(k) summary and associated FDA correspondence, which focuses on demonstrating substantial equivalence to previously cleared predicate devices. This means the manufacturer is asserting their device is as safe and effective as existing legally marketed devices, rather than establishing de novo performance criteria.

    Therefore, I cannot populate the requested table or answer the specific questions about acceptance criteria and performance studies. The "Performance Testing" section explicitly states: "The pre-clinical testing performed indicated that the MonoPoly Pedicle Screw System is substantially equivalent to predicate devices." This implies the testing was comparative in nature against existing devices, not against a predefined set of acceptance criteria for a new device.

    Here's a breakdown of what is and is not present in the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance:

    • Not present. The document does not define specific performance acceptance criteria (e.g., in terms of strength, fatigue life, or other biomechanical properties). It also does not report specific quantitative performance data for the device against such criteria. The "Performance Testing" section only states that the testing indicated substantial equivalence to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not present. The document does not describe a clinical "test set" for performance evaluation in the way you're asking. The testing mentioned is "pre-clinical," likely referring to mechanical or biomechanical bench testing. No sample sizes for any test sets are provided, nor is any information about data provenance (e.g., country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not applicable. Ground truth is generally relevant for diagnostic or AI-driven devices where a human expert's judgment is used to label data. This document describes a mechanical implant, not a diagnostic device.

    4. Adjudication Method:

    • Not applicable. See point 3.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Not applicable. This type of study is used for diagnostic imaging devices to assess human reader performance with and without AI assistance. The MonoPoly Pedicle Screw System is a surgical implant.

    6. Standalone (Algorithm Only) Performance Study:

    • Not applicable. This is a mechanical implant, not an algorithm or AI system.

    7. Type of Ground Truth Used:

    • Not applicable. See point 3.

    8. Sample Size for the Training Set:

    • Not applicable. A "training set" is relevant for machine learning algorithms. This document describes a physical medical device.

    9. How Ground Truth for the Training Set Was Established:

    • Not applicable. See point 8.

    Summary from the provided text:

    The MonoPoly Pedicle Screw System is a Class III spinal implant comprising various screws, rods, and connectors made from titanium alloy. It is intended to aid in spinal segment immobilization and stabilization as an adjunct to fusion for various spinal conditions. The pre-clinical testing indicated substantial equivalence to four predicate devices: Cross Medical Synergy VLS System (K974749), U&I Optima Xia System (K031585), GSS Medical GS System (K053573), and Signus Medical OvalTwist System (K061577). The document does not provide details of specific performance criteria or a study demonstrating the device meets such criteria, as its 510(k) clearance path relies on equivalence to existing devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1