Search Results
Found 4 results
510(k) Data Aggregation
(116 days)
The Disposable and Reusable SPO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg at hospital facilities.
The proposed device, Disposable and Reusable SpO2 Sensors are accessories to the oximeters, which are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive with U.S. legally marketed oximeters or patient monitors. The SSD-001-W09AN sensors is disposable while the SS-010-AF10 and SS-018-AF10 sensors are reusable. They are only intended for adult.
The sensor shall be connected with its corresponding monitor. Oxygenation of blood is measured by detecting the infrared and red light absorption characteristics of deoxygenated hemoglobin and oxygenated hemoglobin, which consists of a probe attached to the patient's finger. The sensor is connected to a data acquisition system which is used to calculate and display oxygen saturation levels and heart rate conditions.
Each sensor has two LEDs, emitting both red and infrared light, and a photodiode. Red and infrared light lit alternately according to certain sequence, when the fingertips of capillary repeatedly with the heart pumps blood congestion, light emitting diode after blood vessels and projected onto a photodiode, photodiode can be induced to change with pulse light intensity, the electrical signals in the form of change. Then the received signal is forwarded to the corresponding oximeter that amplifies the signal and an algorithm that calculates the ratio. By measuring the wave crest of the pulse wave and the absorbance of the trough, SpO2 is calculated to obtain the correct oxygen saturation value. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).
Here's an analysis of the provided information regarding the acceptance criteria and study for the Disposable SpO2 Sensors and Reusable SpO2 Sensors:
1. Table of Acceptance Criteria and Reported Device Performance
The document explicitly states the acceptance criteria for SpO2 and Pulse Rate (PR) accuracy, and it reports that the device meets these requirements.
Acceptance Criteria | Reported Device Performance |
---|---|
SpO2 Accuracy | ±3% @ 70-100% |
PR Accuracy | ±2 bpm |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 12 human adult volunteers were used for each clinical study (there were two studies). This means a total of 24 human adult volunteers were used for the two studies combined.
- Data Provenance: The studies were described as "Clinical hypoxia test results obtained in human adult volunteers." The document doesn't specify the country of origin, but it implies a prospective clinical study using induced hypoxia.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts or their qualifications for establishing the ground truth. It only states that the ground truth for arterial oxygen saturation (SaO2) was "determined by co-oximetry."
4. Adjudication Method for the Test Set
The document does not provide details on any adjudication method used for the test set.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a sensor, and the clinical study focuses on its accuracy against a gold standard (co-oximetry) rather than a comparison of human reader performance with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the accuracy tests described are effectively a standalone performance evaluation of the device. The device itself (sensor and its internal algorithms for calculating SpO2 and PR) is directly compared against the ground truth (co-oximetry) measurements. There's no human 'reading' or interpretation of the sensor's output being evaluated in this context, other than potentially reading the numerical display that the device provides.
7. The Type of Ground Truth Used
The ground truth used was arterial oxygen saturation (SaO2) as determined by co-oximetry. This is a recognized gold standard for measuring oxygen saturation in blood.
8. The Sample Size for the Training Set
The document does not mention the sample size for any training set. Given that this is a sensor (hardware with embedded algorithms) rather than a software-as-a-medical-device (SaMD) based on AI/Machine Learning that typically requires extensive re-training, it's possible that a distinct "training set" as understood in deep learning contexts was not highly relevant or explicitly documented. The development process likely involved calibration and verification, which might use internal datasets not explicitly labeled as "training."
9. How the Ground Truth for the Training Set Was Established
Since a "training set" is not explicitly mentioned, the method for establishing its ground truth is also not provided.
Ask a specific question about this device
(84 days)
TS-F-D
The Finger Sensor is a reusable sensor intended for use for continuous non-invasive arterial oxygen saturation (SpO2) and pulse rate monitoring. Patient weight range > 20 kg (> 44 pounds)
TS-E-D
The Ear Sensor is a reusable sensor intended for use for continuous non-invasive arterial oxygen saturation (SpO2) and pulse rate monitoring. The headband is single-patient use. Patient weight range > 10 kg (> 22 pounds)
TS-W-D
The Wrap Sensor is a reusable sensor intended for use for continuous non-invasive arterial oxygen saturation (SpO2) and pulse rate monitoring. The tape and foam wrap are single-patient use. Patient weight range > 3 kg (> 6.6 pounds)
TS-SE-3
The Sensitive Skin Sensor is a reusable sensor intended for use for continuous non-invasive arterial oxygen saturation (SpO2) and pulse rate monitoring. The tape and foam wrap are singlepatient use. Patient weight range All patients
TS-AF-10 and TS-AF-25
The AllFit Sensor is a single-patient use adhesive sensor intended for use for continuous non-invasive arterial oxygen saturation (SpO2) and pulse rate monitoring. Patient weight range All patients
Trusignal SpO2 Interconnect cables
The Interconnect Cable is a reusable cable intended for use for all patients for continuous non-invasive arterial oxygen saturation (SpO2) and pulse rate monitoring when used with a compatible SpO2 sensor.
Pulse oximeter sensors and interconnect cables connecting to patient monitors
Here's a breakdown of the acceptance criteria and the study details for the TruSignal® SpO2 Sensors and Interconnect Cables, based on the provided 510(k) summary:
1. Acceptance Criteria and Device Performance
Acceptance Criteria (Target) | Reported Device Performance |
---|---|
A_RMS (Accuracy Root Mean Square) |
Ask a specific question about this device
(241 days)
OxiScan II Oximetry Data Management Software (OxiScan II) is a central server based software system that uses a personal computer based application written in Microsoft Dot Net protocol to capture oximetry data from an oximeter, and then transmit this data, in a secure encrypted file, via internet, to the central server. At the central server the data is used to render a standard report that is then transmitted via fax and/or email to the prescribing physician. The intended use of the report is to provide a physician with information to help determine the best pulmonary treatment.
The OxiScan II Oximetry Data Management Software is intended to collect, report and archive oximetry trend data to provide information to a medical professional, as a supplemental tool to assist in the timely identification of pulmonary needs.
The OxiScan II Oximetry Data Management Software is intended to (1) transfer oximetry data from a pulse oximeter to a central server data base in order to maintain unique records per patient and test of this pulse oximetry data, and (2) to generate and archive standard reports drawn from this data. A list of approved oximeters appears in the Capture software.
The OxiScan II Oximetry Data Management Software is not a diagnosis tool. It is a decision management support tool that allows medical personnel to securely and accurately upload and view data related to pulse oximetry and to provide output reports as feedback which may be used by a Heathcare professional to form a patient history.
OxiScan II Oximetry Data Management Software is an accessory for use with compatible pulse oximeters. OxiScan II Oximetry Data Management Software collects and stores patient information containing raw oximetery data captured by a pulse oximeter. This information is transferred by a personal computer to an internet Web server where a report which summarizes and graphically presents the data is prepared. This report is delivered to a Healthcare Professional who then uses it along with other information to determine a course of pulmonary treatment. The report may be retrieved, reviewed, and retransmitted via the OxiScan II Oximetry Data Management Software web site.
The OxiScan II Oximetry Data Management Software is a software system intended to collect, report, and archive oximetry trend data from compatible pulse oximeters to provide information to a medical professional. It acts as a supplemental tool for timely identification of pulmonary needs and to form a patient history, but it is not a diagnostic tool.
The acceptance criteria for the OxiScan II Oximetry Data Management Software are primarily focused on its functional characteristics, safety, and effectiveness in meeting its stated intended use, and its substantial equivalence to a predicate device. The study demonstrating that the device meets these criteria is described as "Functional and Safety Testing."
Here's the breakdown of the information requested, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|
Functional Characteristics: | Representative samples of OxiScan II successfully performance tested to verify compliance to appropriate functional characteristics. |
- Ability to capture oximetry data from compatible oximeters. | The device uses a PC-based application to capture oximetry data from an oximeter. A list of approved oximeters appears in the Capture software. |
- Secure and encrypted data transmission to central server. | Data is transmitted "in a secure encrypted file, via internet, to the central server." |
- Generation of standard reports. | At the central server, data is used to "render a standard report." |
- Transmission of reports to prescribing physician. | Reports are transmitted "via fax and/or email to the prescribing physician." |
- Data archiving for unique patient records and tests. | Intended to "maintain unique records per patient and test of this pulse oximetry data." |
- Accessibility and review of reports by healthcare professionals. | Reports can be "retrieved, reviewed, and retransmitted via the OxiScan II Oximetry Data Management Software web site." |
Safety: | Representative samples of OxiScan II successfully performance tested to verify compliance to appropriate safety characteristics. |
Effectiveness: | Representative samples of OxiScan II successfully performance tested to verify compliance to appropriate effectiveness characteristics. |
Substantial Equivalence: | "OxiScan II Oximetry Data Management Software has been concluded as substantially equivalent to the predicate device [Nonin Medicals "nVISION" software (K033307)]" based on similar intended uses, operation methods, and performance specifications. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a numerical sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only states that "Representative samples of OxiScan II Oximetry Data Management Software were successfully performance tested."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not provide information on the number of experts used or their qualifications for establishing ground truth. The testing mentioned appears to be software performance testing rather than clinical validation with expert review.
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
The document does not describe any adjudication method for a test set. The validation described focuses on functional, safety, and effectiveness testing of the software itself and its equivalence to a predicate device.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. The device is data management software, not an AI-assisted diagnostic tool. Its purpose is to collect, report, and archive oximetry data for review by healthcare professionals, not to provide interpretations or assist human readers in making diagnoses in the way an AI algorithm might.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The device, OxiScan II, is described as "Oximetry Data Management Software." Its standalone performance would relate to its ability to accurately capture, transmit, process, and report oximetry data as per its specifications. The document states that "Representative samples of OxiScan II Oximetry Data Management Software were successfully performance tested to verify compliance to appropriate functional characteristics." This implies standalone testing of the software's data management capabilities. However, it explicitly states it is "not a diagnosis tool" and relies on a "Healthcare Professional" for interpretation, meaning its intended use always involves human-in-the-loop for clinical decision-making.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The document does not detail the type of "ground truth" in the sense of clinical diagnoses or outcomes. Given the nature of the device as data management software, the "ground truth" for its performance testing would likely involve:
- Expected data values/formats: Verifying that captured oximetry data matches the source.
- Report accuracy: Confirming that generated reports correctly reflect the stored data.
- System functionality: Ensuring features like secure transmission, archiving, and retrieval work as designed.
- Compliance with specifications: Testing against pre-defined functional and technical requirements.
8. The Sample Size for the Training Set
The document does not mention a training set, as the device is not described as utilizing machine learning or artificial intelligence that would typically require a training set. The "Functional and Safety Testing" refers to performance verification, not an AI model's training.
9. How the Ground Truth for the Training Set was Established
As no training set is mentioned or applicable to this type of software, this information is not provided.
Ask a specific question about this device
(16 days)
The CARESCAPE V100 Vital Signs Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature.
The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac stepdown. The CARESCAPE V100 Vital Signs Monitor can also be used in satellite areas, physicians' office, or alternate care settings.
The CARESCAPE V100 Vital Signs Monitor is a small, portable monitor for use in a sub-acute hospital and non-hospital settings. The CARESCAPE V100 is for use on adult, pediatric, or neonatal patients - one at a time. The battery operated monitor provides noninvasive determination of systolic blood pressure, diastolic blood pressure, mean arterial pressure, pulse rate, oxygen saturation, and temperature.
The CARESCAPE V100 is a multi-parameter device that provides Sp02 monitoring for use in spot-checking or continuous monitoring. The CARESCAPE V100 also measures NIBP using the oscillometric method. There are no features of CARESCAPE V100 that can be controlled by the patient.
The CARESCAPE V100 monitor is available in four standard configurations with or without an integrated printer:
- CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or . SuperSTAT), Pulse
- CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or . SuperSTAT), Pulse, and Temperature
- . CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or SuperSTAT), Pulse, and Sp02 (Ohmeda TruSignal, Nellcor, or Masimo)
- . CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or SuperSTAT), Pulse, Temperature, and Sp02 (Ohmeda TruSignal, Nellcor, or Masimo)
The provided FDA 510(k) summary for the CARESCAPE V100 Vital Signs Monitor (K073203) describes the device and its intended use, but it does not contain information regarding specific acceptance criteria, study details (like sample sizes, ground truth establishment, or expert qualifications), device performance against those criteria, or multi-reader multi-case studies.
Instead, the document states: "The subject of this 510(k) is a design modification for the ProCare V1 Vital Signs Monitor. The CARESCAPE V100 Vital Signs Monitor complies with the voluntary standards as detailed in Section 4.2 of this submission. The following quality assurance measures were applied to the development of the CARESCAPE V100 Vital Signs Monitor: Risk Analysis, Requirements Reviews, Design Reviews, Subsystem Verification, Integration testing (System verification), Final acceptance testing (Validation), Performance testing, Safety testing, Environmental testing."
The conclusion provided is: "The results of these measurements demonstrated that the CARESCAPE V100 Vital Signs Monitor is as safe, as effective, and performs as well as the predicate device(s)," which is a general statement of substantial equivalence rather than a detailed report of performance against specific acceptance criteria.
Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, or details about the study, sample sizes, ground truth, expert qualifications, adjudication methods, or MRMC studies, as this information is not present in the provided text.
The document refers to "voluntary standards as detailed in Section 4.2 of this submission," which would presumably contain the technical performance specifications and acceptance criteria, but Section 4.2 itself is not included in the provided text.
Ask a specific question about this device
Page 1 of 1