K Number
K181270
Manufacturer
Date Cleared
2018-09-07

(116 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable and Reusable SPO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg at hospital facilities.

Device Description

The proposed device, Disposable and Reusable SpO2 Sensors are accessories to the oximeters, which are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive with U.S. legally marketed oximeters or patient monitors. The SSD-001-W09AN sensors is disposable while the SS-010-AF10 and SS-018-AF10 sensors are reusable. They are only intended for adult.

The sensor shall be connected with its corresponding monitor. Oxygenation of blood is measured by detecting the infrared and red light absorption characteristics of deoxygenated hemoglobin and oxygenated hemoglobin, which consists of a probe attached to the patient's finger. The sensor is connected to a data acquisition system which is used to calculate and display oxygen saturation levels and heart rate conditions.

Each sensor has two LEDs, emitting both red and infrared light, and a photodiode. Red and infrared light lit alternately according to certain sequence, when the fingertips of capillary repeatedly with the heart pumps blood congestion, light emitting diode after blood vessels and projected onto a photodiode, photodiode can be induced to change with pulse light intensity, the electrical signals in the form of change. Then the received signal is forwarded to the corresponding oximeter that amplifies the signal and an algorithm that calculates the ratio. By measuring the wave crest of the pulse wave and the absorbance of the trough, SpO2 is calculated to obtain the correct oxygen saturation value. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).

AI/ML Overview

Here's an analysis of the provided information regarding the acceptance criteria and study for the Disposable SpO2 Sensors and Reusable SpO2 Sensors:

1. Table of Acceptance Criteria and Reported Device Performance

The document explicitly states the acceptance criteria for SpO2 and Pulse Rate (PR) accuracy, and it reports that the device meets these requirements.

Acceptance CriteriaReported Device Performance
SpO2 Accuracy±3% @ 70-100%
PR Accuracy±2 bpm

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 12 human adult volunteers were used for each clinical study (there were two studies). This means a total of 24 human adult volunteers were used for the two studies combined.
  • Data Provenance: The studies were described as "Clinical hypoxia test results obtained in human adult volunteers." The document doesn't specify the country of origin, but it implies a prospective clinical study using induced hypoxia.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications for establishing the ground truth. It only states that the ground truth for arterial oxygen saturation (SaO2) was "determined by co-oximetry."

4. Adjudication Method for the Test Set

The document does not provide details on any adjudication method used for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a sensor, and the clinical study focuses on its accuracy against a gold standard (co-oximetry) rather than a comparison of human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the accuracy tests described are effectively a standalone performance evaluation of the device. The device itself (sensor and its internal algorithms for calculating SpO2 and PR) is directly compared against the ground truth (co-oximetry) measurements. There's no human 'reading' or interpretation of the sensor's output being evaluated in this context, other than potentially reading the numerical display that the device provides.

7. The Type of Ground Truth Used

The ground truth used was arterial oxygen saturation (SaO2) as determined by co-oximetry. This is a recognized gold standard for measuring oxygen saturation in blood.

8. The Sample Size for the Training Set

The document does not mention the sample size for any training set. Given that this is a sensor (hardware with embedded algorithms) rather than a software-as-a-medical-device (SaMD) based on AI/Machine Learning that typically requires extensive re-training, it's possible that a distinct "training set" as understood in deep learning contexts was not highly relevant or explicitly documented. The development process likely involved calibration and verification, which might use internal datasets not explicitly labeled as "training."

9. How the Ground Truth for the Training Set Was Established

Since a "training set" is not explicitly mentioned, the method for establishing its ground truth is also not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Orantech Inc. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China

Re: K181270

Trade/Device Name: Disposable SpO2 Sensors, Reusable SpO2 Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: July 23, 2018 Received: August 10, 2018

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Todd D. Courtney -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181270

Device Name

Disposable SpO2 Sensors, Reusable SpO2 Sensors

Indications for Use (Describe)

The Disposable and Reusable SPO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg at hospital facilities.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K181270

    1. Date of Preparation: 8/21/2018
  • Sponsor Identification 2.

Orantech Inc.

Zone#A, 4F, 1st Bld, 7th Industrial Zone, Yulv Community, GongMing, Guangming New District, Shenzhen, 518132, Guangdong, China.

Establishment Registration Number: Not yet registered

Contact Person: Yunxi Xiong Position: General Manager Tel: +86-0755-23699939 Fax: +86-0755-23243594 Email: yunxi.xiong@orantech.cn

3. Designated Submission Correspondent

Ms. Diana Hong (Primary Contact Person) Mr. Chengyu Wang (Alternative Contact Person)

Mid-Link Consulting Co.. Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 001-3609253199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Disposable SpO2 Sensors, Reusable SpO2 Sensors Model: SSD-001-W09AN, SS-010-AF10 and SS-018-AF10 Classification Name: Oximeter (Accessory-sensor) Classification: II Product Code: DQA Regulation Number: 21 CFR 870.2700 Review Panel: Anesthesiology

Indications for Use:

The Disposable and Reusable SPO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg at hospital facilities.

Device Description

The proposed device, Disposable and Reusable SpO2 Sensors are accessories to the oximeters, which are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive with U.S. legally marketed oximeters or patient monitors. The SSD-001-W09AN sensors is disposable while the SS-010-AF10 and SS-018-AF10 sensors are reusable. They are only intended for adult.

The sensor shall be connected with its corresponding monitor. Oxygenation of blood is measured by detecting the infrared and red light absorption characteristics of deoxygenated hemoglobin and oxygenated hemoglobin, which consists of a probe attached to the patient's finger. The sensor is connected to a data acquisition system which is used to calculate and display oxygen saturation levels and heart rate conditions.

Each sensor has two LEDs, emitting both red and infrared light, and a photodiode. Red and infrared light lit alternately according to certain sequence, when the fingertips of capillary repeatedly with the heart pumps blood congestion, light emitting diode after blood vessels and projected onto a photodiode, photodiode can be induced to change with pulse light intensity, the electrical signals in the form of change. Then the received signal is forwarded to the corresponding oximeter that amplifies the signal and an algorithm that calculates the ratio. By measuring the wave crest of the pulse wave and the absorbance of the trough, SpO2 is calculated to obtain the correct oxygen saturation value. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).

{5}------------------------------------------------

  • న. Identification of Predicate Device
    Predicate Device 510(k) Number: K153184 Product Name: Caremed Reusable & Disposable SpO2 Sensors

6. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ♪ IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance
  • ♪ IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests
  • A ISO 80601-2-61: 2011 Medical Electrical Equipment - Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.
  • ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity A
  • ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization

The proposed device have also been demonstrated to comply with the requirements of US electrical safety differences according to US National standard ANSI/AAMI ES60601-1: 2005 / A2:2010.

The accuracy test on proposed device demonstrated that the accuracy of pulse and SpO2 meets the specified requirements.

The accuracy test on proposed device under low perfusion conditions demonstrated that the accuracy of pulse and SpO2 under low perfusion conditions meets the specified requirements.

The proposed device belongs to skin contact, and the contact duration is less than 24h. Biocompatibility tests have been conducted on proposed device, including cytotoxicity, sensitization, and skin irritation. The test results show that the proposed device has no cytotoxicity, sensitization, or skin irritation.

    1. Clinical Test Conclusion
      Clinical studies were conducted to verify the accuracy of proposed device. All of three models were included in the studies. The clinical studies were conducted per following standards:

{6}------------------------------------------------

  • ISO 80601-2-61: 2011 Medical Electrical Equipment - Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.

  • Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and Drug Administration Staff

The clinical studies are conducted with proposed device and compatible oximeter, the device included in the clinical studies are listed as follow:

Proposed Sensor ModelCompatible OximeterK Number of Oximeter
SSD-001-W09ANNellcor NPB-40K963707
SS-010-AF10GE TuffSatK001688
SS-018-AF10GE TruSatK040831

Table 1 Device in Clinical Studies

There are two clinical studies for proposed device, one is conducted on SS-010-AF10 and SSD-001-W09AN, and the other is conducted on SS-018-AF10.

Clinical studies were performed under an approved protocol with subject informed consent. Clinical hypoxia test results were obtained in 12 human adult volunteers for each study, to validate the accuracy of proposed device versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.

    1. Substantially Equivalent (SE) Comparison
ITEMProposed DevicePredicate DeviceK153184
Product CodeDQADQA
Regulation Number21 CFR 870.270021 CFR 870.2700
ClassIIII
Indications for UseThe Disposable and Reusable SPO2Sensors are indicated for continuousnon-invasive monitoring of functionaloxygen saturation of arterialhemoglobin (SpO2) and pulse rate (PR)for adult patients weighing greater than40 kg at hospital facilities.Caremed Reusable & Disposable SPO2Sensors are indicated for continuous non-invasive monitoring of functional oxygensaturation of arterial hemoglobin (SpO2)and pulse rate(PR) for adult patientsweighing greater than 40 kg and pediatricpatients weighing 10 -50 kg athospital facilities
Principle of2-wavelength Relative Optical2-wavelength Relative Optical Absorption

Table 2 Substantially Equivalent Comparison

{7}------------------------------------------------

OperationAbsorption
Light EmittingRed:660, 661 and 663nmInfrared:890, 904 and 940nmRed:660-666nmInfrared:880-950nm
Signal DetectionMethodPhotodetectorPhotodetector
SpO2 Range70%-100%70-100%
SpO2 Accuracyf3%f3% @ 70-100%
PR Range35-240bpm30-250 bpm
PR Accuracyf2bpmf3 bpm
SterileNoNo
UsageReusable & DisposableReusable & Disposable
Electrical SafetyComplied with IEC 60601-1Complied with IEC 60601-1
EMCComplied with IEC 60601-1-2Complied with IEC 60601-1-2
PerformanceComplied with ISO 80601-2-61Complied with ISO 80601-2-61
Biocompatibility
CytotoxicityComplied with ISO 10993-5Complied with ISO 10993-5
Skin IrritationComplied with ISO 10993-10Complied with ISO 10993-10
SensitizationComplied with ISO 10993-10Complied with ISO 10993-10
  • Substantially Equivalent (SE) Conclusion 9.
    Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).