Search Results
Found 3 results
510(k) Data Aggregation
(89 days)
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
Porous coated components may be used cemented or uncemented (biological fixation). All other femoral, tibial baseplate, and all-polyethylene (UHMWPE and VEXLPE) patella components are indicated for cemented use only.
The Persona™ Personalized Knee System is a semiconstrained modular knee prosthesis designed to resurface the articulating surface of the femoral, tibial and patellar bones. The Persona Knee System utilizes a modular design between the tibial plates and articular surfaces. The addition of the Trabecular Metal™ femoral component will provide surgeons with the ability to implant with, or without cement (biological fixation).
The Zimmer® Persona™ Personalized Knee System is a medical device for knee replacement. The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing an extensive study report with detailed acceptance criteria and performance metrics typically found in efficacy trials.
Here's a breakdown based on the information provided:
1. Table of Acceptance Criteria and Reported Device Performance:
The document states that "Clinical data and conclusions were not needed for this device" because it is a Class II device demonstrating substantial equivalence to existing predicate devices. Therefore, the acceptance criteria are met by demonstrating that the new device exhibits similar or identical performance characteristics to these predicate devices through non-clinical (bench) testing. The acceptance criteria are implicit in the FDA guidance documents followed for testing.
Property or Characteristic | Acceptance Criteria (Implicit from FDA Guidance) | Reported Device Performance |
---|---|---|
Posterior Condyle Fatigue Testing in a Cantilever Loading Condition of the Persona™ Trabecular Metal™ femoral component | Adequate posterior condyle fatigue strength as defined by FDA Guidance: Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses. (The specific quantitative acceptance value is not stated in this summary, but would be detailed in the full test report). | Demonstrated adequate posterior condyle fatigue strength of the uncemented Persona™ Trabecular Metal™ femoral component in the cantilever loading condition. |
Posterior Condyle Fatigue of Persona™ Trabecular Metal™ femoral component in three-point Bend Loading Condition | Adequate posterior condyle fatigue performance as defined by FDA Guidance: Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses. (The specific quantitative acceptance value is not stated in this summary). | Demonstrated adequate posterior condyle fatigue performance of the Persona™ Trabecular Metal™ femoral component in a 3-point bend loading condition. |
Modified Metallic Surface Characterization for the Persona™ Trabecular Metal™ Porous femoral knee component with Trabecular Metal™ | Mechanical, physical, and chemical analyses in accordance with the Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement. (Specific quantitative acceptance metrics for these analyses are not stated in this summary). | Mechanical, physical, and chemical analyses of Trabecular Metal™ were assessed in accordance with the Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement. (Implied that the results met the requirements of the guidance). |
Static Tensile, Static Shear and Shear Fatigue of Trabecular Metal™ Diffusion Bonded to Zimaloy-One Hour Cycles, 0.28mm Ti Sheet Interlayer | Bond strength that met the 20 MPa static tensile test and static shear acceptance criteria. (The specific acceptance criteria for shear fatigue, if numerical, are not explicitly stated, but are implied to be met). | Demonstrated that one hour diffusion bonding cycles for Trabecular Metal™ produces a bond that met the 20 MPa static tensile test and static shear acceptance criteria. |
Evaluation of Interactions of the Zimmer Legacy Knee and Persona Primary Implant Systems with the Magnetic Fields in the Magnetic Resonance Imaging (MRI) Environment | Safety and compatibility within the MRI environment, as defined by FDA Guidance: Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. (Specific quantitative acceptance metrics for MR safety, such as temperature rise or artifact size, are not stated in this summary). | Demonstrated safety and compatibility within the MRI environment. |
2. Sample Size Used for the Test Set and the Data Provenance:
- Sample Size: The document does not specify the exact sample sizes (number of components tested) for each non-clinical bench test. These details would typically be found in the full test protocols and reports, but are not included in this summary.
- Data Provenance: The data provenance is "Non-Clinical Performance and Conclusions," meaning the tests were conducted in a laboratory setting (bench testing) rather than on human subjects. This is not retrospective or prospective clinical data. The location of the testing laboratories is not specified.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This is not applicable. For bench testing of medical devices like knee implants, ground truth is established by engineering specifications, material properties, and relevant industry standards (e.g., ISO, ASTM), rather than expert medical consensus on patient cases. The "experts" involved are likely engineering and materials science professionals who design and conduct the tests according to the FDA guidance.
4. Adjudication Method for the Test Set:
This is not applicable. Adjudication methods like 2+1 or 3+1 refer to how discrepancies in expert opinions on clinical data are resolved. Since this submission relies on non-clinical bench testing, such an adjudication method is not relevant. The "adjudication" is essentially the adherence to pre-defined test protocols and acceptance criteria based on FDA guidance and engineering standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a knee implant, not an AI-powered diagnostic tool requiring MRMC studies to assess the impact of AI on human reader performance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. As above, this is a physical implant, not an algorithm.
7. The Type of Ground Truth Used:
The ground truth used for these non-clinical tests is established by:
- Engineering Specifications: Designed tolerances, material strengths, and fatigue limits.
- Industry Standards: ASTM, ISO, and other relevant standards for orthopedic implant testing.
- FDA Guidance Documents: The specific guidance documents cited (e.g., Class II Special Controls Guidance Document for Knee Joints, Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces, Guidance for MR Safety) define the methods and performance expectations for validating the device's properties.
8. The Sample Size for the Training Set:
Not applicable. This device is not an AI/ML algorithm that requires a training set. The "training" for such a device would be the extensive research and development phases involving finite element analysis, material science studies, and repeated experimental bench testing, but these are not considered "training sets" in the context of AI.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no "training set" in the AI sense for this device.
Ask a specific question about this device
(133 days)
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
This device is intended for cemented use only.
The Zimmer® Persona™ Personalized Knee System is a semi-constrained modular knee prosthesis designed to resurface the articulating surface of the femoral, tibial and patellar bones. The Persona Knee System utilizes a modular design between the tibial plates and articular surfaces.
This document describes the non-clinical performance testing for the Zimmer® Persona™ Personalized Knee System, a semi-constrained modular knee prosthesis. Since this is a medical device (knee implant) rather than an AI/software device, many of the requested fields (multi-reader multi-case study, ground truth establishment, training set details) are not applicable.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for each test are implicitly demonstrating "sufficient" or "adequate" strength/wear/constraint, or being "comparable" to predicate devices. The "Test Results" column describes how the device performance met these implicit criteria.
Property or Characteristic | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Fatigue Test of the Persona Tibia Keel and Stem Extension | Sufficient fatigue strength to survive expected worst-case loading conditions. | Demonstrated that the Persona tibia keel and stem extension taper junction provide sufficient fatigue strength to survive expected worst-case loading conditions. |
Cantilever Fatigue Test of the Persona Cemented Tibia | Adequate fatigue strength in the cantilever loading condition. | Demonstrated adequate fatigue strength in the cantilever loading condition. |
Wear Testing of Persona CR Conventional Articular Surfaces Under Load and Motion Curves From the ISO 14243 Standard | Sufficient wear characteristics to survive expected in-vivo loading conditions. | Demonstrated that the wear characteristics of the Persona CR articular surfaces, when articulated against the Persona CR femoral component, are sufficient to survive expected in-vivo loading conditions. |
Wear Testing of Persona PS Conventional UHMWPE Articular Surfaces Under Load and Motion Curves from the ISO 14243 Standard | Sufficient wear characteristics to survive expected in-vivo loading conditions. | Demonstrated that the wear characteristics of the Persona PS articular surfaces, when articulated against the Persona PS femoral component, are sufficient to survive expected in-vivo loading conditions. |
Wear Testing of Persona UC Conventional UHMWPE Articular Surfaces Under Displacement Control with Load and Motion Curves from the ISO 14243-1 Standard | Sufficient wear characteristics to survive expected in-vivo loading conditions. | Demonstrated that the wear characteristics of the Persona UC articular surfaces, when articulated against the Persona CR femoral component, are sufficient to survive expected in-vivo loading conditions. |
Spine Fatigue Evaluation of the Persona PS Conventional UHMWPE Articular Surfaces | Sufficient strength to survive expected in-vivo stress/strain loading conditions. | Demonstrated that the spine of the Persona PS articular surfaces has sufficient strength to survive expected in-vivo stress/strain loading conditions. |
Tibiofemoral Constraint Evaluation of the Persona Conventional UHMWPE Articular Surface | Constraint values comparable to similar NexGen articular surfaces to provide adequate constraint through needed tibiofemoral flexion angles. | Demonstrated that constraint values for the Persona articular surfaces are comparable to data from similar NexGen articular surfaces. Therefore, the Persona articular surfaces provide adequate constraint through the needed tibiofemoral flexion angles. |
Lateral Constraint Evaluation of the Persona Conventional All-Poly Patellar Component on the Persona CR and PS Femoral Components | Lateral subluxation force comparable to control testing on NexGen predicate devices. | Demonstrated that the lateral subluxation force of the Persona conventional all-poly patellar component on the Persona CR and PS femoral implants at tibiofemoral flexion angles 0° to 90° was comparable to control testing on NexGen predicate devices. |
Tibiofemoral Contact Area and Contact Pressure Evaluation of the Persona CR/UC/PS Conventional Articular Surfaces | Contact area and contact pressure comparable to previous testing on similar NexGen articular surfaces. | Demonstrated that the contact area and contact pressure of the Persona articular surfaces are comparable to data from previous testing on similar NexGen articular surfaces. |
Contact Area and Contact Stress Evaluation of the Persona Conventional All-Poly Patellar Component on the Persona Primary CR and PS Femoral Components | Contact areas similar between CR and PS femoral components for all flexion angles. | Demonstrated that, for all flexion angles, the contact areas were similar between the CR and PS femoral components. |
Anterior Liftoff Testing of the Persona Articular Surfaces | Sufficient locking mechanism strength to survive potential worst-case anterior liftoff loading conditions during deep flexion. | Demonstrated sufficient locking mechanism strength to survive potential worst-case anterior liftoff loading conditions during deep flexion. |
Posterior Liftoff Fatigue Strength of the Persona Articular Surfaces | Sufficient locking mechanism strength to survive potential worst-case shear loading conditions. | Demonstrated sufficient locking mechanism strength to survive potential worst-case shear loading conditions. |
Assembly Testing of the Persona Tibia Locking Mechanism | Successful assembly of the modular articular surfaces at normal and maximum interference conditions. | Demonstrated successful assembly of the modular articular surfaces at normal and maximum interference conditions. |
Static Shear Strength of the Persona Tibia Locking Mechanism | Adequate resistance of the modular articular surfaces to disassembly. | Demonstrated adequate resistance of the modular articular surfaces to disassembly. |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the sample sizes for each mechanical test. These "tests" are non-clinical (laboratory-based mechanical and wear simulations) rather than using patient data. Therefore, the concept of "data provenance" (country of origin, retrospective/prospective) is not applicable here as these are not studies involving human subjects or patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. This is a medical device, not an AI/software device that requires expert ground truth for its performance evaluation. The "ground truth" for these tests is based on established engineering principles, industry standards (e.g., ISO 14243), and comparison to predicate devices, not expert human interpretation.
4. Adjudication Method for the Test Set
Not applicable. As described above, this is non-clinical mechanical testing, not a clinical study requiring human adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
Not applicable. This is a physical medical device (knee implant), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and the concept of "human readers improving with AI assistance" are not relevant.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
Not applicable. As this is not an AI/software device, the concept of a "standalone algorithm" is not relevant. The device itself (the knee implant) is the "standalone" item being tested in these non-clinical performance studies.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the non-clinical performance tests is based on:
- Industry Standards: For instance, wear testing was conducted "Under Load and Motion Curves From the ISO 14243 Standard."
- Established Engineering Principles: Tests like fatigue, static shear strength, and contact mechanics rely on well-known engineering material and biomechanical principles.
- Predicate Device Performance: Many tests established comparability to existing, legally marketed predicate devices (e.g., NexGen articular surfaces, NexGen predicate devices). The performance of these predicate devices, which have a history of safe and effective use, serves as a benchmark or "ground truth" for acceptable performance.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device (knee implant), not a machine learning or AI algorithm. There is no concept of a "training set" for the device itself.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(74 days)
This device is indicated for patients with severe knee pain and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle, post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities, the salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
This device is intended for cemented use only.
Specific uses with CR-Flex GSF or LPS-Flex GSF femorals: Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range. The CR-Flex GSF femoral, when used with 90-prefix NexGen CR articular surfaces, 00-prefix 10, 12 or 14mm or 90prefix 17 or 20mm Prolong™ Highly Crosslinked Polyethylene CR articular surfaces, is designed for use with a functional posterior cruciate ligament and when load bearing range of motion (ROM) is expected to be less than or equal to 155 degrees. The LPS-Flex GSF femoral, when used with LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees.
The NexGen Knee GSF Femoral Components include both LPS-Flex GSF and CR-Flex GSF versions and are part of the Zimmer Flex-series of semiconstrained, nonlinked, condylar knee prostheses that are designed to have a maximum active flexion of 155 degrees. The GSF designation indicates that the design of the femoral component has been modified slightly to address specific anatomic features of the distal femur that can be seen in both male and female patients, but are more typical of a female patient.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the NexGen® Knee Gender Solutions Female (GSF) Femoral Components:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Substantial Equivalence | Demonstrates substantial equivalence to predicate devices (NexGen LPS-Flex knee, K991581; NexGen CR-Flex femoral component, K023211). | Mechanical testing and Finite Element Analysis of the NexGen Knee GSF Femoral Components indicate that they are substantially equivalent to the predicate devices. |
Functional Equivalence | Except for modifications to address specific anatomic features typical of a female patient, these components are identical to their respective predicate device. | The GSF designation indicates that the design of the femoral component has been modified slightly to address specific anatomic features of the distal femur that can be seen in both male and female patients, but are more typical of a female patient. |
Manufacturing/Sterilization | Packaged and sterilized using the same materials and processes as predicate devices. | The device is packaged and sterilized using the same materials and processes. |
Clinical Performance | (No specific clinical acceptance criteria stated for this device, as clinical data was deemed not needed.) | Clinical data and conclusions were not needed for this device. |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable. The submission explicitly states "Clinical data and conclusions were not needed for this device." The primary evaluation relied on non-clinical testing (mechanical testing and Finite Element Analysis) and comparison to predicate devices, rather than a clinical test set with human subjects.
- Data Provenance: Not applicable for a human subject test set. The data provenance for the non-clinical mechanical testing and Finite Element Analysis would be internal laboratory data generated by Zimmer, Inc.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. No human subject test set was used, and therefore no ground truth established by experts in this context.
4. Adjudication method for the test set
- Not applicable. No human subject test set was used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a total knee prosthesis, not an AI-assisted diagnostic tool or imaging device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is not an algorithm or software. The "standalone" performance here refers to the mechanical and analytical performance of the physical device components, which was assessed through non-clinical testing.
7. The type of ground truth used
- The implicit "ground truth" for the performance evaluation of this medical device is the established performance and safety characteristics of its predicate devices (NexGen LPS-Flex Knee and NexGen CR-Flex femoral component). The device is deemed safe and effective if it can demonstrate substantial equivalence to these legally marketed predicates through mechanical testing and component analysis.
8. The sample size for the training set
- Not applicable. This device is a physical implant, not an AI algorithm that requires a training set. The "training" in product development would refer to engineering design, CAD modeling, and material selection based on established bio-mechanical principles and previous device designs.
9. How the ground truth for the training set was established
- Not applicable. As noted above, this isn't an AI device with a training set. The design was informed by the established anatomical features and performance of prior generations of knee prostheses, particularly those with modifications for "female" anatomical characteristics. The "ground truth" for its design would be anatomical data, biomechanical principles, and the performance history of existing devices.
Ask a specific question about this device
Page 1 of 1