Search Results
Found 4 results
510(k) Data Aggregation
(306 days)
The My3D® Personalized Pelvic Reconstruction system is indicated for use in patients requiring reconstruction of the pelvis and/or hip joint due to disease, deformity, trauma, or revision procedures where other treatments or revisions have failed. The device is a combination of single use guided osteotomy instruments, a pelvic implant, screws, and acetabular/ femoral components. The pelvic implant is intended for cementless application in individuals where bone quality or bony defect size cannot support a standard sized acetabular implant is intended to be fixed to the remaining pelvic anatomy using compatible bone screws to create a prosthetic acetabulum. The reconstructed prosthetic acetabulum is intended to be used with a compatible cemented acetabular cup, dual mobility head (POLARCUP Insert), and femoral components to restore hip function.
The My3D® Personalized Pelvic Reconstruction system is a patient specific combination of single use resection instruments, a pelvic implant, screws, acetabular, and femoral components. The system was developed to address conditions which require reconstruction of the acetabulum and hip joint.
This patient matched device is designed from inputs including imaging, diagnosis, and surgical approach. Together with the surgeon, these inputs are then translated via a design process to create patient specific implants and, if appropriate, instruments to reconstruct the patient's pelvis. If utilized, the patient specific instruments are used to resect the bone and allow for implantation of the patient matched pelvis. The joint is then reconstructed with a cemented acetabular cup, dual mobility head (POLARCUP Insert), and femoral components.
The implants and resection instruments are single use devices. Reusable instrumentation is provided non-sterile in surgical trays which are to be re-processed per validated instructions.
I am sorry, but the provided text content does not contain enough information to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as per your request. The document is an FDA 510(k) clearance letter and summary for the My3D® Personalized Pelvic Reconstruction device, which primarily addresses substantial equivalence to a predicate device.
Specifically, the document mentions "Performance Data" and lists several types of testing performed (biocompatibility, manufacturing residuals, sterilization validation, mechanical testing of porous structure, fatigue testing, screw performance, cadaveric evaluation of design process). However, it does not provide:
- A table of acceptance criteria and reported device performance. It only lists the types of tests.
- Sample size used for the test set and data provenance.
- Number of experts used to establish ground truth and their qualifications.
- Adjudication method.
- Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done or its effect size.
- Whether a standalone performance study was done.
- The type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
The document explicitly states: "Clinical data were not deemed necessary for the subject device." This further indicates that no human-centric validation studies (like MRMC or studies requiring expert ground truth and adjudication) were conducted or included in this submission summary. The testing mentioned appears to be primarily engineering and material performance tests.
Ask a specific question about this device
(88 days)
The Prime E-CLASS™ XLPE Liner is an acetabular liner intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. This device is indicated for the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankyloses, protrusion acetabuli, and painful hip dysplasia;
- inflammatory degenerative joint disease such as rheumatoid arthritis;
- correction of functional deformity; and,
- revision procedures where other treatments or devices have failed.
Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.
The Prime E-CLASS™ XLPE Liner is an additional liner option for MicroPort Orthopedics' Prime Acetabular Cup System (K170444; K180798). The subject Liner is intended to be used with Acetabular Shells and optional Cancellous Bone Screws as part of a total acetabular system.
The Prime E-CLASS™ XLPE Liner is manufactured from E-CLASS™, a vitamin E blended XLPE, conforming to ASTM F2695-12. The subject Liner is available in Standard, Lipped, and Lateralized/Face-changing configurations. The subject Liner has a two-part locking detail featuring 12 anti-rotational tabs and a lock flange, which is intended to mate with the 12 antirotational pockets and lock groove of the compatible Prime Acetabular Shells (K170444; K180798).
The provided document is a 510(k) Summary for a medical device (Prime E-CLASS™ XLPE Liner), which focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific acceptance criteria with quantifiable performance metrics.
Therefore, the information required for a table of acceptance criteria and proven device performance as typically expected for a diagnostic AI/ML device is not available in this document. The document describes non-clinical testing performed to establish substantial equivalence for a physical implant, not a software algorithm.
Here's a breakdown of why the requested information cannot be fully provided based on the input text:
-
Acceptance Criteria & Device Performance: The document does not define specific clinical or diagnostic acceptance criteria (e.g., sensitivity, specificity, accuracy thresholds) that the device must meet. Instead, it details non-clinical tests (bacterial endotoxin, mechanical testing, material properties, wear analysis, fatigue, range of motion) to show the device performs similarly to predicate devices. The "reported device performance" are the results of these non-clinical tests, which aim to demonstrate equivalence rather than meeting independent performance criteria.
-
Study Type: This is a premarket notification (510(k)) for a physical medical implant, not a study of an AI/ML diagnostic or prognostic device. The "study" here refers to non-clinical bench testing and analysis to show equivalence.
Given the nature of the document, most of the requested fields are not applicable or the information is not present.
However, I can extract what is available regarding the "studies" (non-clinical testing) performed:
1. A table of acceptance criteria and the reported device performance
As mentioned, specific acceptance criteria in terms of clinical performance metrics are not provided. The "acceptance criteria" for this 510(k) submission are implied by successful completion of tests demonstrating substantial equivalence to predicate devices. The "reported device performance" refers to the outcomes of these non-clinical tests.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Bacterial endotoxin limit (USP endotoxin limit of 20 EU/device) | Endotoxins found to be less than the USP endotoxin limit of 20 EU/device. |
Mechanical performance of lock detail (Push-out, Lever-out, Torque-out, Axial long-term fatigue) per ASTM F1820, ASTM STP1301, FDA Guidance | Testing performed "through Pre- and Post- Fatigue Pushout Lever-out and Torque-out of the subject Liner from a compatible Shell... and testing found in literature." Axial long-term fatigue also evaluated. (Results affirm substantial equivalence, specific numeric results are not provided in this summary.) |
Material properties of E-CLASS™ per various ASTM/ISO standards | Considered applicable from K171181 (identical material). (Implies material meets these standards.) |
Smooth wear per ASTM F1714-96 (2003), ISO 14242-2, etc. | Considered applicable from K171181 (identical material and articulating surface). |
Abrasive wear per ASTM F1714-96 (2003), ISO 14242-3, etc. | Considered applicable from K171181 (identical material and articulating surface). |
Wear particle analysis per ASTM F1877-05 | Considered applicable from K171181. |
Long-term fatigue per FDA Draft Guidance Documents | Considered applicable from K171181. |
Deformation and frictional torque per ISO 7206-12 | Considered applicable from K171181. |
Range of motion per ISO 21535 | Considered applicable from K170444 (identical articulating surface and configurations). |
Biocompatibility (extractables, cytotoxicity, irritation, sensitization) | Performed for K171181 (identical E-CLASS™ material). (Implies results were acceptable.) |
Sterilization residuals per AAMI TIR 28:2009 Annex A | Evaluated. "Sterealization residuals are within the limits determined by the worst case, which was presented in K140043." |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated for each non-clinical test. For mechanical tests, standard testing protocols define sample sizes, but these are not enumerated here.
- Data Provenance: Not applicable. This refers to bench testing of physical components, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. Ground truth for clinical data is not relevant to this type of non-clinical device testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. Adjudication is typically for human assessment of images or clinical outcomes, not for mechanical bench testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI/ML diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an AI/ML diagnostic device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For non-clinical testing, the "ground truth" is established by adherence to recognized national and international standards (ASTM, ISO, FDA Guidance) and laboratory measurements using calibrated equipment. For example, the ground truth for endotoxin levels is defined by the USP limit. For mechanical properties, it's the measured values compared to established benchmarks or predicate device performance.
8. The sample size for the training set
- Not applicable. This is not an AI/ML diagnostic device, and thus no "training set" of data in that context. The "training" for the device's design comes from engineering principles, material science, and prior designs.
9. How the ground truth for the training set was established
- Not applicable. No "training set" in the context of AI/ML.
Ask a specific question about this device
(104 days)
The Prime Acetabular Cup System is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
-
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankyloses, protrusio acetabuli, and painful hip dysplasia:
-
- inflammatory degenerative joint disease such as rheumatoid arthritis;
-
- correction of functional deformity; and,
-
- revision procedures where other treatments or devices have failed.
Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.
The Prime Acetabular Cup System includes acetabular shells, acetabular liners, and optional cancellous bone screws. The design features are summarized below:
- · Acetabular Shells
- Manufactured from titanium alloy о
- BIOFOAM® coated o
- Available in Solid or Quad configurations o
- Outer Diameter sizes 42mm to 68mm in 2mm increments O
- Acetabular Liners ●
- Manufactured from A-CLASS® (highly crosslinked ultra-high molecular o weight polyethylene)
- Lipped Available in in Standard, or Face-changing Lateralized O configurations with 12 anti-rotational tabs
- Inner Diameter sizes 22mm to 44mm O
- Bone Screws ●
- Manufactured from titanium alloy o
- 6.5mm diameter O
- 0 Available in lengths 15mm to 80mm in 5mm increments
The subject Acetabular Liner has undergone design changes in regard to the locking detail; all other design aspects, intended use, fundamental scientific technology, and material remain identical to the predicate K170444 Acetabular Liner.
This looks like a 510(k) summary for a medical device (Prime Acetabular Cup System), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical study results with acceptance criteria.
As such, it does not contain the specific information requested in your prompt regarding acceptance criteria, a study proving device meets acceptance criteria, sample sizes for test/training sets, data provenance, expert adjudication, MRMC studies, standalone performance, or ground truth establishment.
The document primarily provides:
- Device Description and Intended Use: Details the components of the Prime Acetabular Cup System (Acetabular Shells, Liners, Bone Screws) and their intended applications in total hip arthroplasty.
- Substantial Equivalence Justification: States that the device is substantially equivalent to predicate devices (PROCOTYL® PRIME Acetabular Cup System, K170444) based on identical indications for use, similar materials (for the liner, drawing on K052026), and unchanged design features for shells and screws since K170444.
- Nonclinical Testing: Lists bench tests performed on the subject device (liner push-out, lever-out, torque-out per ASTM F1820, ASTM STP1301) and references tests performed on the predicate device and deemed applicable (deformation, frictional torque, wear, fatigue, range of motion, screw properties).
- Clinical Testing: Explicitly states that clinical data was not provided for the subject devices.
- Biocompatibility: Justifies not performing specific biocompatibility tests on the subject implants as their materials and patient contact are identical to predicate devices.
- Compatibility Tables: Details compatibility with other MicroPort Orthopedics products.
- Sterilization Residuals: Discusses previous evaluations and current applicability.
Therefore, I cannot populate the requested table and answer your questions directly from the provided text. The document does not describe a study with acceptance criteria in the context of clinical performance or diagnostic accuracy, which is implied by your prompt's questions. It focuses on engineering and material equivalence testing for an orthopedic implant.
Ask a specific question about this device
(132 days)
The PROCOTYL® PRIME Acetabular Cup System is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli. and painful hip dysplasia;
- inflammatory degenerative joint disease such as rheumatoid arthritis;
- correction of functional deformity; and,
- revision procedures where other treatments or devices have failed.
Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.
The PROCOTYL® PRIME Acetabular Cup System includes acetabular shells, acetabular liners, and optional cancellous bone screws. The design features are summarized below:
- Acetabular Shells
- Manufactured from titanium alloy
- BIOFOAM® coated
- Available in Solid or Quad configurations
- Outer Diameter sizes 42mm to 68mm in 2mm increments
- Acetabular Liners
- Manufactured from A-CLASS® highly crosslinked ultra high molecular weight polyethylene)
- Available in Standard. Lipped Face-changing or Lateralized configurations
- Inner Diameter sizes 22mm to 44mm
- Bone Screws
- Manufactured from titanium alloy
- 6.5mm diameter
- Available in lengths 15mm to 80mm in 5mm increments
This document is a 510(k) summary for the PROCOTYL® PRIME Acetabular Cup System, a medical device for total hip arthroplasty. It describes the device, its intended use, and argues for its substantial equivalence to previously cleared predicate devices.
Here's an analysis of the provided text in relation to your request about acceptance criteria and studies:
Crucially, this document is a 510(k) Pre-Market Notification, which is a regulatory submission to demonstrate that a new device is substantially equivalent to a legally marketed predicate device. It is not a detailed clinical study report or a publication presenting performance metrics against specific acceptance criteria for an AI/ML device.
Therefore, most of the information you are requesting about acceptance criteria, specific performance metrics, sample sizes for test sets, expert ground truth, MRMC studies, or standalone algorithm performance for an AI/ML device is not present in this document. This document concerns a physical orthopedic implant.
However, I can extract the information that is available and clarify what is missing based on your request.
Summary of Device Acceptance Criteria and Performance (Based on available information):
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Mechanical Performance (Nonclinical Testing): | |
Lock detail strength (Push-out, Lever-out, Torque-out per ASTM 1820) | "The lock detail of the subject PROCOTYL® PRIME Acetabular Cup System was evaluated by testing Push-out, Lever-out and Torque-out of the Liner per ASTM 1820." (Implied: met relevant ASTM standards, but no specific values or thresholds are provided for "meeting" the criteria). |
Deformation and frictional torque under pinch-loading conditions | "The subject System underwent deformation and frictional torque testing to evaluate the properties under pinch-loading conditions." (Implied: results were acceptable, but no specific values or thresholds are provided.) |
Fatigue resistance (Long term axial cyclic loading) | "The subject System completed long term axial cyclic loading to evaluate fatigue resistance properties." (Implied: acceptable fatigue resistance, but no specific methodology or pass/fail criteria are defined.) |
Wear rate and wear particle analysis | "The subject System underwent wear testing to determine the wear rate and a wear particle analysis was performed on the wear debris." (Implied: acceptable wear characteristics, but no specific wear rate thresholds or particle analysis results are given.) |
Range of motion | "Additionally, a range of motion study was completed for the PROCOTYL® PRIME Acetabular Cup System." (Implied: acceptable range of motion, but no specific values or comparison to predicate is given.) |
Cancellous bone screw properties (per ASTM F543) | "Finally, per ASTM F543, the properties of the cancellous bone screws were examined." (Implied: met relevant ASTM standards, but no specific values or thresholds are provided.) |
Biocompatibility: | |
Pyrogen Limit Specification (per Limulus Amebocyte Lysate (LAL) test) | "The Limulus Amebocyte Lysate (LAL) test was used to determine the subject devices meets the specified pyrogen limit specification of ≤ 20 EU/device." (Explicitly states: "meets the specified pyrogen limit specification of ≤ 20 EU/device.") |
Substantial Equivalence (Overall Regulatory Goal): | "The design features and materials are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the subject devices has not changed relative to the predicate devices. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provide within this Premarket Notification." (This is the overarching "acceptance criterion" for a 510(k), demonstrated by comparing to predicate devices and performing nonclinical testing to support this claim). |
Missing Information (as per your request, because this is not an AI/ML device nor a typical clinical study report):
- Sample size used for the test set and the data provenance: Not applicable. The "test set" here refers to physical components tested in a lab, not a diagnostic dataset. The document does not specify sample sizes (e.g., number of cups, liners, or screws) for any of the nonclinical tests. Data provenance, in the sense of country of origin or retrospective/prospective for patient data, is also not applicable here.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an orthopedic implant's mechanical properties is typically established through recognized engineering standards (e.g., ASTM) and validated testing methodologies, not expert medical consensus on diagnostic images.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This applies to human review of diagnostic outputs, not mechanical testing.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML diagnostic device with human-in-the-loop.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): For mechanical testing, the "ground truth" is defined by the specifications and performance standards (e.g., ASTM standards) that the device must meet. For biocompatibility, it's defined by the pyrogen limit specification (≤ 20 EU/device). It's not based on expert consensus, pathology, or outcomes data in the conventional sense used for diagnostic devices.
- The sample size for the training set: Not applicable. This is not an AI/ML device that requires a training set.
- How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
Page 1 of 1