Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K250652
    Manufacturer
    Date Cleared
    2025-07-28

    (146 days)

    Product Code
    Regulation Number
    870.2380
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    QYE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ECG-AI LEF 12-Lead algorithm is software intended to aid in earlier detection of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for heart failure. This population includes, but is not limited to:

    • patients with cardiomyopathies
    • patients who are post-myocardial infarction
    • patients with aortic stenosis
    • patients with chronic atrial fibrillation
    • patients receiving pharmaceutical therapies that are cardiotoxic, and
    • postpartum women.

    The ECG-AI LEF 12-Lead algorithm is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm.

    A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.

    The ECG-AI LEF 12-Lead Algorithm should be applied jointly with clinician judgment.

    Device Description

    The ECG-AI LEF 12-Lead algorithm interprets 12-lead ECG voltage times series data using an artificial intelligence-based algorithm. The device analyzes 10 seconds of a single 12-lead ECG acquisition, and within seconds provides likelihood of LVEF (ejection fraction less than or equal to 40%) to third party software. The results are displayed by the third party software on a device such as a smartphone, tablet, or PC. The ECG-AI LEF 12-Lead algorithm was trained to detect Low LVEF using positive and control cohorts, and the detection of Low LVEF in patients is generated using defined conditions and covariates.

    The ECG-AI LEF 12-Lead algorithm device is intended to address the unmet need for a point-of-care screen for LVEF less than or equal to 40% and is expected to be used by cardiologists, frontline clinicians at primary care, urgent care, and emergency care settings, where cardiac imaging may not be available or may be difficult or unreliable for clinicians to operate. Clinicians will use the ECG-AI LEF 12-Lead algorithm to aid in earlier detection of LVEF less than or equal to 40% and making a decision for further cardiac evaluation.

    The software module can be integrated into a client application to be accessed by clinicians and results viewed through an Electronic Medical Record (EMR) system or an ECG Management System (EMS) accessed via a PC, mobile device, or another medical device. In each case, the physician imports 12-lead ECG data in digital format. The tool analyzes the 10 seconds or longer duration of voltage data collected during a standard 12-lead ECG and outputs a binary result of the likelihood of low ejection fraction as an API result.

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for the Anumana, Inc. ECG-AI Low Ejection Fraction (LEF) 12-Lead Algorithm ([K250652](https://510k.innolitics.com/search/K250652)). While it describes the device, its intended use, and substantial equivalence to a predicate device, it does not contain the detailed performance study results, acceptance criteria tables, sample sizes, or ground truth establishment methods that would typically be found in the clinical study section of a full 510(k) submission.

    The document discusses a "Predetermined Change Control Plan (PCCP)" which mentions future performance enhancement validation studies, but it doesn't present the specific results of the validation study that led to this clearance ([K250652](https://510k.innolitics.com/search/K250652)). It only states that "The performance characteristics for the ECG-AI LEF 12-Lead algorithm were evaluated through software verification and labeling verification," which refers to non-clinical data.

    Therefore, many of the requested details cannot be extracted from the provided text. I will populate the table and answer the questions based only on the information available in the given document.


    Acceptance Criteria and Device Performance Study (Extracted from provided 510(k) Summary)

    The provided 510(k) summary (K250652) serves as an update to a previously cleared device (K232699). It focuses on expanding compatibility and minor changes, asserting substantial equivalence based on the predicate's performance rather than detailing a new, comprehensive clinical study for this specific submission. The document emphasizes "software verification and labeling verification" as the evaluation methods for performance characteristics for this particular submission, rather than a clinical performance study with specific metrics for acceptance criteria.

    The Predetermined Change Control Plan (PCCP) section alludes to future performance enhancements and their validation, stating: "To be implemented, a modified version must demonstrate improved performance by meeting pre-specified acceptance criteria. These criteria require the new version's sensitivity and specificity point estimates to be greater than or equal to the previous version, with an improvement shown by either an increased point estimate or a tighter confidence interval lower bound for at least one of these metrics." However, these are future criteria for updates, not the current acceptance criteria for the clearance of K250652 based on a new clinical study.

    Therefore, the specific quantitative acceptance criteria and reported device performance for the clinical study supporting the K250652 clearance are not explicitly stated in the provided text. The clearance is largely based on demonstrating substantial equivalence to the predicate (K232699) and software/labeling verification.

    Based on the provided text, the specific details regarding the clinical performance study (including acceptance criteria, reported performance values, sample sizes, expert details, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for the test set) are NOT available.


    1. A table of acceptance criteria and the reported device performance

    As noted above, the provided text does not contain a table of explicit quantitative acceptance criteria or reported device performance metrics (e.g., sensitivity, specificity, AUC) from a clinical study for K250652. The document claims "The performance characteristics for the ECG-AI LEF 12-Lead algorithm were evaluated through software verification and labeling verification" for this submission, indicating that a new, detailed clinical performance study with such metrics was not the basis for this specific clearance. The PCCP section specifies criteria for future updates, but not for this clearance.

    MetricAcceptance CriteriaReported Device Performance
    Quantitative Performance Metrics (e.g., Sensitivity, Specificity, AUC)Not specified in the provided document for this clearance (K250652). The PCCP mentions that future updates must show sensitivity and specificity point estimates $\ge$ previous version, or improved confidence interval.Not specified in the provided document for this clearance (K250652). The clearance is based on substantial equivalence to a predicate and non-clinical verification.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in the provided document.
    • Data Provenance: Not specified in the provided document. The PCCP mentions "multi-center retrospective clinical study" for future validations, but this isn't linked to the original clearance's test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not specified in the provided document.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not specified in the provided document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not specified in the provided document. The current indication is "to aid in earlier detection" and "applied jointly with clinician judgment," which implies human-in-the-loop, but an MRMC study comparing performance with and without AI assistance is not detailed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • The document states: "The ECG-AI LEF 12-Lead algorithm is not intended to be a stand-alone diagnostic device for cardiac conditions," and "should be applied jointly with clinician judgment." This implies the device is not intended for standalone use in practice. However, whether a standalone performance study was conducted to assess its raw diagnostic capability (e.g., area under the curve) is not explicitly stated. The statement "outputs a binary result of the likelihood of low ejection fraction as an API result" suggests a standalone algorithm output, but the FDA's clearance is for an "aid," not a primary diagnostic tool.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The document mentions the device "was trained to detect Low LVEF using positive and control cohorts." For LVEF, the common ground truth is often echocardiography (measuring ejection fraction), but the specific method used for ground truth (e.g., echocardiography, MRI, or a combination/adjudication) is not specified.

    8. The sample size for the training set

    • Not specified in the provided document.

    9. How the ground truth for the training set was established

    • The document states the device "was trained to detect Low LVEF using positive and control cohorts," but it does not describe how the ground truth was established for these training cohorts (e.g., type of diagnostic test, clinical adjudication process).
    Ask a Question

    Ask a specific question about this device

    K Number
    K250119
    Manufacturer
    Date Cleared
    2025-07-15

    (180 days)

    Product Code
    Regulation Number
    870.2380
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    QYE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tempus ECG-Low EF is software intended to analyze resting, non-ambulatory 12-lead ECG recordings and detect signs associated with having a low left ventricular ejection fraction (LVEF less than or equal to 40%). It is for use on clinical diagnostic ECG recordings collected at a healthcare facility from patients 40 years of age or older at risk of heart failure. This population includes but is not limited to patients with atrial fibrillation, aortic stenosis, cardiomyopathy, myocardial infarction, diabetes, hypertension, mitral regurgitation, and ischemic heart disease.

    Tempus ECG-Low EF only analyzes ECG data and provides a binary output for interpretation. Tempus ECG-Low EF is not intended to be a stand-alone diagnostic tool for cardiac conditions, should not be used for patient monitoring, and should not be used on ECGs with paced rhythms. Results should be interpreted in conjunction with other diagnostic information, including the patient's original ECG recordings and other tests, as well as the patient's symptoms and clinical history.

    A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of low LVEF. Patients receiving a negative result should continue to be evaluated in accordance with current medical practice standards using all available clinical information.

    Device Description

    Tempus ECG-Low EF is a cardiovascular machine learning software intended for analysis of 12-lead resting ECG recordings using machine-learning techniques to detect signs of cardiovascular conditions for further referral or diagnostic follow-up. The software employs machine learning techniques to analyze ECG recordings and detect signs associated with a patient experiencing low left ventricular ejection fraction (LVEF), less than or equal to 40%. The device is designed to extract otherwise unavailable information from ECGs conducted under the standard of care, to help health care providers better identify patients who may be at risk for undiagnosed LVEF in order to evaluate them for further referral or diagnostic follow up.

    As input, the software takes data from a patient's 12-lead resting ECG (including age and sex). It is only compatible with ECG recordings collected using 'wet' Ag/AgCl electrodes with conductive gel/paste, and using FDA authorized 12-lead resting ECG machines manufactured by GE Medical Systems or Philips Medical Systems with a 500 Hz sampling rate. It checks the format and quality of the input data, analyzes the data via a trained and 'locked' machine-learning model to generate an uncalibrated risk score, converts the model results to a binary output (or reports that the input data are unclassifiable), and evaluates the uncalibrated risk score against pre-set operating points (thresholds) to produce a final result. Uncalibrated risk scores at or above the threshold are returned as 'Low LVEF Detected,' and uncalibrated risk scores below the threshold are returned as 'Low LVEF Not Detected.' This information is used to support clinical decision making regarding the need for further referral or diagnostic follow-up. Typical diagnostic follow-up could include transthoracic echocardiogram (TTE) to detect previously undiagnosed LVEF, as described in device labeling. Results should not be used to direct any therapy against LVEF itself. Tempus ECG-Low EF is not intended to replace other diagnostic tests.

    Tempus ECG-Low EF does not have a dedicated user interface (UI). Input data comprising ECG tracings, tracing metadata (e.g., sample count, sample rate, patient age/sex), is provided to Tempus ECG-Low EF through standard communication protocols (e.g., file exchange) with other medical systems (e.g., electronic health record systems, hospital information systems, or other data display, transfer, storage, or format-conversion software). Results from Tempus ECG-Low EF are returned to users in an equivalent manner.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study that proves the Tempus ECG-Low EF device meets them, based on the provided FDA 510(k) clearance letter:

    Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance CriteriaReported Device Performance
    Sensitivity (for LVEF ≤ 40%)≥ 80% (lower bound of 95% CI)86% (point estimate); 84% (lower bound of 95% CI)
    Specificity (for LVEF > 40%)≥ 80% (lower bound of 95% CI)83% (point estimate); 82% (lower bound of 95% CI)

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Greater than 15,000 ECGs (specifically, 14,924 patient records are detailed in Table 1, with each patient having one ECG).
    • Data Provenance: Retrospective observational cohort study. The data was derived from 4 geographically distinct US clinical sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts used or their qualifications for establishing the ground truth. It mentions that a clinical diagnosis of Low EF (LVEF ≤ 40%) was determined by a Transthoracic Echocardiogram (TTE), which is considered the gold standard for LVEF measurement. The interpretation of these TTE results to establish the ground truth would typically be done by cardiologists or trained echocardiography specialists, but the specific number and qualifications are not provided in this document.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (such as 2+1 or 3+1) for the ground truth of the test set. The ground truth was established by correlating ECGs with TTEs to determine the presence or absence of a clinical diagnosis of low EF. It is implied that the TTE results themselves, as the gold standard, served as the definitive ground truth without a further adjudication process by multiple human readers for the TTE results in the context of this AI device validation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    The document does not indicate that an MRMC comparative effectiveness study was performed, nor does it provide an effect size for human reader improvement with AI assistance. The study focuses on the standalone performance of the AI device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone study was done. The described clinical performance validation evaluated the device's ability to "detect signs associated with a clinical diagnosis of low LVEF" and provided sensitivity and specificity metrics for the algorithm's output. The device "only analyzes ECG data and provides a binary output for interpretation," indicating a standalone performance assessment.

    7. The Type of Ground Truth Used

    The ground truth used was established by Transthoracic Echocardiogram (TTE), specifically used to determine the presence or absence of a clinical diagnosis of Low EF (LVEF ≤ 40%). This is a form of outcomes data / reference standard as TTE is the established clinical diagnostic method for LVEF.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: More than 930,000 ECGs (specifically, 930,689 ECGs are detailed in Table 1).

    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. However, given that the model was trained to "detect signs associated with having a low left ventricular ejection fraction (LVEF less than or equal to 40%)" and the validation set used TTE for ground truth, it is highly probable that the training set also used LVEF measurements (likely from echocardiograms) as the ground truth. The description states the model was trained "on data from more than 930,000 ECGs," but does not detail the specific methodology for establishing the LVEF ground truth for each of these training examples.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233409
    Manufacturer
    Date Cleared
    2024-03-28

    (174 days)

    Product Code
    Regulation Number
    870.2380
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    QYE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Eko Low Ejection Fraction Tool (ELEFT) is a software intended to aid clinicians in identifying individuals with Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. ELEFT takes as input ECG and heart sounds and is intended for use on patients at risk for heart failure. This population includes, but is not limited to, patients with: coronary artery disease; diabetes mellitus; cardiomyopathy; hypertension; and obesity.

    The interpretations of heart sounds and ECG offered by the software are meant only to assist healthcare providers in assessing Left Ventricular Ejection Fraction ≤ 40% , who may use the result in conjunction with their own evaluation and clinical judgment. It is not a diagnosis or for monitoring of patients diagnosed with heart failure. This software is for use on adults (18 years and older).

    Device Description

    Eko Low Ejection Fraction Tool (ELEFT) is an algorithm that is intended to aid clinicians to identify individuals with Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. ELEFT takes as input ECG and heart sounds from patients at risk for heart failure. The software uses signal processing as well as machine learning algorithms, to analyze the electrocardiogram (ECG) and heart sound/phonocardiogram (PCG) recording signals generated by FDA-cleared Eko Stethoscopes and saved as .WAV file recordings in the Eko Cloud. ELEFT is a machine learning based notification software which employs machine learning techniques to suggest the likelihood of LVEF

    AI/ML Overview

    The Eko Low Ejection Fraction Tool (ELEFT) is a software intended to aid clinicians in identifying individuals with Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. The device takes ECG and heart sound inputs and processes them using signal processing and machine learning algorithms.

    Here's an analysis of its acceptance criteria and the study proving its performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state "acceptance criteria" in a numerical target format (e.g., "Sensitivity must be >= X%"). However, the clinical performance results presented demonstrate the device's capability to detect Low EF. The acceptance effectively hinges on the presented sensitivity and specificity values.

    MetricAcceptance Criteria (Implicit from Study Results)Reported Device Performance (95% CI)
    SensitivityDemonstrated performance74.7% (69.4-79.6)
    SpecificityDemonstrated performance77.5% (75.9-79.0)
    PPVDemonstrated performance25.7% (22.8-28.7)
    NPVDemonstrated performance96.7% (95.9-97.4)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 3,456 unique subjects. After excluding 307 recordings due to poor ECG quality, the performance analysis was based on the remaining suitable recordings.
    • Data Provenance: Retrospective data collected from:
      • US, 5 sites: 2,960 patients.
      • India, 1 site: 496 patients.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not explicitly stated as a number, but the ground truth for ejection fraction was "overread by a board-certified cardiologist." This implies at least one, and potentially multiple, board-certified cardiologists were involved in reviewing the echocardiogram results.
    • Qualifications of Experts: Board-certified cardiologists.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method like 2+1 or 3+1 for resolving discrepancies in ground truth establishment. It states that the "subject's true ejection fraction was measured by the echocardiogram machine's integrated cardiac quantification software at the echocardiogram and then overread by a board-certified cardiologist." This suggests a single expert review after automated measurement, with no mention of multiple reviewers or a formal reconciliation process if initial measurements or interpretations differed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The study focuses solely on the standalone performance of the ELEFT algorithm without a human-in-the-loop component or evaluating the improvement of human readers with AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone (algorithm only) performance study was conducted. The results for sensitivity, specificity, PPV, and NPV presented in Table 2 and the subsequent text (page 9) are for the ELEFT algorithm's performance in differentiating between Low EF (≤40%) and Normal EF (>40%).

    7. Type of Ground Truth Used

    The type of ground truth used was expert consensus / pathology based on instrumental measurements and expert review:

    • Echocardiogram (Instrumental Measurement): The true ejection fraction was measured by the echocardiogram machine's integrated cardiac quantification software.
    • Expert Overread: This measurement was "overread by a board-certified cardiologist."
    • Categorization: Ejection status (Low EF or Normal EF) was then assigned based on these measured and reviewed values.

    8. Sample Size for the Training Set

    The sample size for the training set was 1,852 patients. This data was contributed from:

    • US, 7 sites: 1,515 patients.
    • India, 1 site: 337 patients.

    9. How Ground Truth for the Training Set Was Established

    The document does not explicitly detail the exact process for establishing ground truth for the training set. However, given the consistency in the data description and the validation methodology, it is highly probable that the ground truth for the training set was established using the same methodology as the test set: gold standard echocardiogram measurements, subsequently overread by board-certified cardiologists, and then categorized into Low EF (≤40%) or Normal EF (>40%).

    Ask a Question

    Ask a specific question about this device

    K Number
    K232699
    Manufacturer
    Date Cleared
    2023-09-28

    (23 days)

    Product Code
    Regulation Number
    870.2380
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    QYE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anumana Low Ejection Fraction AI-ECG Algorithm is software intended to aid in screening for Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for heart failure. This population includes, but is not limited to:
    · patients with cardiomyopathies

    • patients who are post-myocardial infarction
    • · patients with aortic stenosis
    • · patients with chronic atrial fibrillation
    • · patients receiving pharmaceutical therapies that are cardiotoxic, and
      • postpartum women.

    Anumana Low Ejection Fraction Al-ECG Algorthm is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm.

    A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.

    The Anumana Low Ejection Fraction AI-ECG Algorithm should be applied jointly with clinician judgment.

    Device Description

    The Low Ejection Fraction AI-ECG Algorithm interprets 12-lead ECG voltage times series data using an artificial intelligence-based algorithm. The device analyzes 10 seconds of a single 12lead ECG acquisition, and within seconds provides a prediction of likelihood of LVEF (ejection fraction less than or equal to 40%) to third party software. The results are displayed by the third-party software on a device such as a smartphone, tablet, or PC. The Low Ejection Fraction AI-ECG Algorithm was trained to predict Low LVEF using positive and control cohorts, and the prediction of Low LVEF in patients is generated using defined conditions and covariates. The Low Ejection Fraction AI-ECG Algorithm device is intended to address the unmet need for a point-of-care screen for LVEF less than or equal to 40% and is expected to be used by cardiologists, front-line clinicians at primary care, urgent care, and emergency care settings, where cardiac imaging may not be available or may be difficult or unreliable for clinicians to operate. Clinicians will use the Low Eiection Fraction AI-ECG Algorithm to aid in screening for LVEF less than or equal to 40% and making a decision for further cardiac evaluation.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided FDA 510(k) clearance letter for the Low Ejection Fraction AI-ECG Algorithm:


    Low Ejection Fraction AI-ECG Algorithm: Acceptance Criteria and Performance Study

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance (95% CI)
    Sensitivity80% or higher84.5% (82.2% to 86.6%)
    Specificity80% or higher83.6% (82.9% to 84.2%)
    Positive Predictive Value (PPV)Not specified (derived metric)30.5% (28.8% to 32.1%)
    Negative Predictive Value (NPV)Not specified (derived metric)98.4% (98.2% to 98.7%)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The clinical validation study included 16,000 patient records initially, though 2,040 records were excluded due to quality checks, resulting in a final analysis sample of 13,960 patient-ECG pairs.
    • Data Provenance: The data was retrospective, collected from 4 health systems across the United States.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts or their qualifications used to establish the ground truth for the clinical validation test set. The ground truth (LVEF 40%) was derived from transthoracic echocardiogram (TTE) measurements. While TTE interpretation requires expertise, the document doesn't detail the method of expert review or consensus for these TTE results themselves for the test set.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (e.g., 2+1, 3+1) for the ground truth for the test set. The ground truth was established by TTE measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study evaluated the standalone performance of the AI algorithm against a ground truth without human readers in the loop.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, a standalone performance study was done. The reported sensitivity and specificity values are for the algorithm's performance alone in detecting low LVEF.

    7. The Type of Ground Truth Used

    The type of ground truth used for both training and validation was objective clinical measurements from Transthoracic Echocardiogram (TTE), specifically the Left Ventricular Ejection Fraction (LVEF) measurement. An LVEF of $\le$ 40% was defined as the disease cohort, and > 40% as the control cohort.

    8. The Sample Size for the Training Set

    The training set for the algorithm development consisted of 93,722 patients with an ECG and TTE performed within a 2-week interval. These were split into:

    • Training dataset: 50% of the 93,722 patients.
    • Tuning dataset: 20% of the 93,722 patients.
    • Set-aside testing dataset: 30% of the 93,722 patients (used for internal validation during development, distinct from the independent clinical validation study).

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the training set was established using LVEF measurements obtained from transthoracic echocardiograms (TTE). Specifically, for each patient, the LVEF measurement from the earliest TTE within a 2-week interval of an ECG was paired with the closest ECG recording. LVEF $\le$ 40% defined the disease cohort, and LVEF > 40% defined the control cohort. This data was identified from a research-use authorized clinical database from Mayo Clinic.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1