K Number
K250649
Manufacturer
Date Cleared
2025-09-19

(199 days)

Product Code
Regulation Number
870.2380
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Bunkerhill ECG-EF is software intended to aid in screening for Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for, but not already diagnosed with low LVEF.

Bunkerhill ECG-EF is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm. A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%.

Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.

Bunkerhill ECG-EF is adjunctive and must be interpreted in conjunction with the clinician's judgment, the patient's medical history, symptoms, and additional diagnostic tests. For a final clinical diagnosis, further confirmatory testing, such as echocardiography, is required.

Device Description

ECG-EF is a software-only medical device that employs deep learning algorithms to analyze 12-lead ECG data for the detection of low left ventricular ejection fraction (LVEF < 40%). The algorithm processes 10-second ECG waveform snippets, providing predictions to assist healthcare professionals in the early identification of patients at risk for heart failure.

ECG-EF algorithm receives digital 12-lead ECG data and processes it through its machine learning model. The output of the analysis is transmitted to integrated third-party software systems, such as Electronic Medical Records (EMR) or ECG Management Systems (EMS). The results are displayed by the third-party software on a device such as a smartphone, tablet, or PC.

ECG-EF algorithm produces a result indicating "Low EF Screen Positive - High probability of low ejection fraction based on the ECG", " Low EF Screen Negative - Low probability of low ejection fraction based on the ECG" or "Error – device input criteria not met " for cases that do not meet data input requirements. These results are not intended to be definitive diagnostic outputs but rather serve as adjunctive information to support clinical decision-making. A disclaimer accompanies the output, stating: "Not for diagnostic use. The results are not final and must be reviewed alongside clinical judgment and other diagnostic methods."

The Low Ejection Fraction AI-ECG Algorithm device is intended to address the unmet need for a point-of-care screen for LVEF less than or equal to 40% and is expected to be used by cardiologists, front-line clinicians at primary care, urgent care, and emergency care settings, where cardiac imaging may not be available or may be difficult or unreliable for clinicians to operate. Clinicians will use the Low Ejection Fraction AI-ECG Algorithm to aid in screening for LVEF less than or equal to 40% and making a decision for further cardiac evaluation.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Bunkerhill ECG-EF device, based on the provided FDA 510(k) Clearance Letter:


1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance CriteriaReported Device Performance (Value and 95% Confidence Interval)Pass/Fail
SensitivitySe ≥ 80%82.66% (80.90–84.30)Pass
SpecificitySp ≥ 80%83.20% (82.60–83.80)Pass
PPVPPV ≥ 25%37.20% (35.70–38.76)Pass
NPVNPV ≥ 95%97.54% (97.28–97.83)Pass

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for Test Set: 15,994 patient records.
  • Data Provenance:
    • Country of Origin: United States.
    • Source: Two health systems.
    • Type: Retrospective study.
    • Diversity: Representative of the U.S. population (65.5% White, 18.8% Hispanic, 5.7% American Indian or Alaska Native, 3.9% Asian, 3.0% Black/African American, 2.8% Other; 53% Male, 47% Female).
    • Geographical Distribution: Curated from 5 geographically distributed sites throughout the United States.

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number of experts used or their specific qualifications for establishing the ground truth. It only mentions that the ground truth was established from echocardiograms.


4. Adjudication Method for the Test Set

The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the test set. The ground truth was derived directly from echocardiogram measurements.


5. MRMC Comparative Effectiveness Study

The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size of human readers improving with AI vs. without AI assistance. The study focuses solely on the standalone performance of the AI algorithm.


6. Standalone Performance Study (Algorithm Only)

Yes, a standalone study evaluating the algorithm's performance without human-in-the-loop was conducted. The performance metrics (Sensitivity, Specificity, PPV, NPV) and the confusion matrix presented are for the algorithm's direct output.


7. Type of Ground Truth Used

The ground truth used was Transthoracic Echocardiogram (TTE) with disease, specifically using the Simpson's Biplane measurement method to determine Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. The echocardiogram was taken less than 15 days apart from the ECG scan.


8. Sample Size for the Training Set

The document does not explicitly state the sample size used for the training set. It only mentions the retrospective study for validation involved 15,994 patient records.


9. How Ground Truth for the Training Set Was Established

The document states that the "Ground Truth for Model Training" was Transthoracic echocardiogram (TTE) with disease. It can be inferred that this same method (TTE, likely Simpson's Biplane) was used to establish ground truth for the training data, similar to the test set, but specific details on the process for the training set are not provided.

FDA 510(k) Clearance Letter - Bunkerhill ECG-EF

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

September 19, 2025

BunkerHill Health
℅ John Smith
Partner
Hogan Lovells LLP
555 Thirteenth Street, NW
Washington, District of Columbia 20004

Re: K250649
Trade/Device Name: Bunkerhill ECG-EF
Regulation Number: 21 CFR 870.2380
Regulation Name: Cardiovascular Machine Learning-Based Notification Software
Regulatory Class: Class II
Product Code: QYE
Dated: March 3, 2025
Received: March 4, 2025

Dear John Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250649 - John Smith Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250649 - John Smith Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jackson Hair -S Digitally signed by Jackson Hair - S Date: 2025.09.19 13:41:37 -04'00'

for
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K250649

Device Name
Bunkerhill ECG-EF

Indications for Use (Describe)

Bunkerhill ECG-EF is software intended to aid in screening for Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for, but not already diagnosed with low LVEF.

Bunkerhill ECG-EF is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm. A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%.

Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.

Bunkerhill ECG-EF is adjunctive and must be interpreted in conjunction with the clinician's judgment, the patient's medical history, symptoms, and additional diagnostic tests. For a final clinical diagnosis, further confirmatory testing, such as echocardiography, is required.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(K) SUMMARY

Bunkerhill ECG-EF

Bunkerhill, Inc. 436 Bryant Street
San Francisco CA 94107
Phone: (516) 305-3596
Contact Person: Eren Alkan
Date Prepared: September 19, 2025

Proposed Device

Proprietary NameBunkerhill ECG-EF
Classification NameCardiovascular Machine Learning-Based Notification Software
Regulation Number21 CFR 870.2380
Product CodeQYE
Regulatory ClassII

Predicate Device

Proprietary NameLow Ejection Fraction AI-ECG Algorithm
Premarket NotificationK232699
Classification NameCardiovascular Machine Learning-Based Notification Software
Regulation Number21 CFR 870.2380
Product CodeQYE
Regulatory ClassII

Reference Device

Proprietary NameEko Low Ejection Fraction Tool (ELEFT)
Premarket NotificationK233409
Classification NameCardiovascular Machine Learning-Based Notification Software
Regulation Number21 CFR 870.2380
Product CodeQYE
Regulatory ClassII

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 1 of 9

Page 6

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 2 of 9

Device Description

ECG-EF is a software-only medical device that employs deep learning algorithms to analyze 12-lead ECG data for the detection of low left ventricular ejection fraction (LVEF < 40%). The algorithm processes 10-second ECG waveform snippets, providing predictions to assist healthcare professionals in the early identification of patients at risk for heart failure.

ECG-EF algorithm receives digital 12-lead ECG data and processes it through its machine learning model. The output of the analysis is transmitted to integrated third-party software systems, such as Electronic Medical Records (EMR) or ECG Management Systems (EMS). The results are displayed by the third-party software on a device such as a smartphone, tablet, or PC.

ECG-EF algorithm produces a result indicating "Low EF Screen Positive - High probability of low ejection fraction based on the ECG", " Low EF Screen Negative - Low probability of low ejection fraction based on the ECG" or "Error – device input criteria not met " for cases that do not meet data input requirements. These results are not intended to be definitive diagnostic outputs but rather serve as adjunctive information to support clinical decision-making. A disclaimer accompanies the output, stating: "Not for diagnostic use. The results are not final and must be reviewed alongside clinical judgment and other diagnostic methods."

The Low Ejection Fraction AI-ECG Algorithm device is intended to address the unmet need for a point-of-care screen for LVEF less than or equal to 40% and is expected to be used by cardiologists, front-line clinicians at primary care, urgent care, and emergency care settings, where cardiac imaging may not be available or may be difficult or unreliable for clinicians to operate. Clinicians will use the Low Ejection Fraction AI-ECG Algorithm to aid in screening for LVEF less than or equal to 40% and making a decision for further cardiac evaluation.

Intended Use / Indications for Use

Bunkerhill ECG-EF is software intended to aid in screening for Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for, but not already diagnosed with low LVEF.

Bunkerhill ECG-EF is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm. A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%.

Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.

Bunkerhill ECG-EF is adjunctive and must be interpreted in conjunction with the clinician's judgment, the patient's medical history, symptoms, and additional diagnostic tests. For a final clinical diagnosis, further confirmatory testing, such as echocardiography, is required.

Page 7

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 3 of 9

Table 1: Indications for Use Comparison

Proposed Device: Bunkerhill ECG-EF AlgorithmPredicate Device: Low Ejection Fraction AI-ECG Algorithm (K232699)Reference Device: Eko Low Ejection Fraction Tool (K233409)
Intended use / Indications for useBunkerhill ECG-EF is software intended to aid in screening for Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for, but not already diagnosed with low LVEF.Bunkerhill ECG-EF is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm. A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%.Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.Bunkerhill ECG-EF is adjunctive and must be interpreted in conjunction with the clinician's judgment, the patient's medical history, symptoms, and additional diagnostic tests. For a final clinical diagnosis, further confirmatory testing, such as echocardiography, is required.The Anumana Low Ejection Fraction AI- ECG Algorithm is software intended to aid in screening for Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for heart failure. This population includes, but is not limited to:• patients with cardiomyopathies• patients who are postmyocardial infarction• patients with aortic stenosis• patients with chronic atrial fibrillation• patients receiving pharmaceutical therapies that are cardiotoxic, and• postpartum women.Anumana Low Ejection Fraction AI-ECG Algorithm is not intended to be a stand- alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm.A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.The Anumana Low Ejection Fraction AI-ECG Algorithm should be applied jointly with clinician judgment.Eko Low Ejection Fraction Tool (ELEFT) is a software intended to aid clinicians in identifying individuals with Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. ELEFT takes as input ECG and heart sounds and is intended for use on patients at risk for heart failure. This population includes, but is not limited to, patients with: coronary artery disease; diabetes mellitus; cardiomyopathy; hypertension; and obesity.The interpretations of heart sounds and ECG offered by the software are meant only to assist healthcare providers in assessing Left Ventricular Ejection Fraction ≤ 40% , who may use the result in conjunction with their own evaluation and clinical judgment. It is not a diagnosis or for monitoring of patients diagnosed with heart failure. This software is for use on adults (18 years and older).

Page 8

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 4 of 9

The intended use of the Bunkerhill ECG-EF is the same as that of the Anumana Low Ejection Fraction AI-ECG Algorithm, and reference device, Eko Low Ejection Fraction Tool (ELEFT) as the devices are software tools designed to aid in screening for Left Ventricular Ejection Fraction (LVEF) ≤40% in adults at risk for heart failure.

  • Both devices assist certified medical professionals with analyze 12-lead ECG data for the rapid detection of low left ventricular ejection fraction (LVEF ≤ 40%).
  • Both subject and predicate devices are not intended to be used stand-alone, can be used optionally by the physician and do not provide a definitive diagnosis. Both devices require the physician to use this information to decide next steps and/or additional diagnostic work up.
  • Verification and validation testing, including performance assessment demonstrates that the Bunkerhill ECG-EF Device is as safe and effective as its predicate device.

Summary of Technological Characteristics

There are no major changes to technological characteristics in the subject device compared to the predicate device. Both the predicate and the subject device A) use machine learning algorithms to analyze 12-lead ECG data for the rapid detection of low left ventricular ejection fraction (LVEF ≤ 40%). Both the algorithms process 10-second ECG waveform snippets, providing predictions to assist healthcare professionals in the early identification of patients at risk for heart failure. However, they do not replace clinical evaluation and do not alter the standard of care.

The subject device and the predicate device are Software as a Medical Device (SaMD) provided as a software module packaged in a Docker container. The Algorithms do not provide a graphical user interface (GUI) of their own. They are integrated with other medical systems such as Electronic Medical Record (EMR) systems or ECG Management Systems (EMS). The third-party integrating software furnishes a 12-lead ECG digital waveform as input to the Low Ejection Fraction AI-ECG Algorithm and records the algorithm output for display via the integrated medical system or for printing in an offline report.

There are no major differences between subject device and reference device as well. The reference device can also analyze PCG signals when available. Any minor differences in the technology characteristics do not raise different questions of safety and effectiveness.

Page 9

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 5 of 9

Table 2: Technological Comparison

Proposed Device: Bunkerhill ECG-EF AlgorithmPredicate Device: Low Ejection Fraction AI-ECG Algorithm (K232699)Reference Device: Eko Low Ejection Fraction Tool (K233409)Summary
Product codeQYEQYEQYESame
Regulation number21 CFR 870.238021 CFR 870.238021 CFR 870.2380Same
Regulation NameCardiovascular machine learning-based notification softwareCardiovascular machine learning-based notification softwareCardiovascular machine learning- based notification softwareSame
Operational ModeSpot CheckSpot CheckSpot CheckSame
Patient populationAdultsAdultsAdults 18 and olderSame
Environment of UsePrimary care, urgent care, and emergency care settingsPrimary care, urgent care, and emergency care settingsPrimary care, urgent care, and emergency care settingsSame- Both devices are to be used in professional healthcare settings
AlgorithmMachine learning based algorithmMachine learning based algorithmMachine learning based algorithmSame
Algorithm Calculation and OutputDetection of LVEF (Left Ventricular Ejection Fraction less than or equal to 40%) from an ECG signalDetection of LVEF (Left Ventricular Ejection Fraction less than or equal to 40%) from an ECG signalThe Eko Low Ejection Fraction Tool (ELEFT) to identify individuals with Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% by analyzing ECG and heart sounds from patients at risk for heart failure.Same
Ground Truth for Model TrainingTransthoracic echocardiogram (TTE) with diseaseTransthoracic echocardiogram (TTE) with diseasegold standard echocardiogramSame- Both devices rely upon established clinical diagnostic methods as ground truth
Physiological Parameter Inputs12-Lead ECG waveform in digital format12-Lead ECG waveform in digital formatECG recording signals from single lead PCG recording signals (when available)Same- Both devices utilize ECG digital waveforms as input
Data DisplayedAlgorithm output is provided to third party software that displays the result to clinicians. Output provided for each ECG is "Low LVEF Detected" "LowAlgorithm output is provided to third party software that displays the result to clinicians. Output provided for each ECG is "Low LVEF Detected""Low LVEF Not Detected" Application Programming Interface (API) only, no user interfaceSame- Both devices provide data suggesting the likelihood of the same cardiovascular disease or condition for further referral or diagnostic follow-up.

Page 10

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 6 of 9

Proposed Device: Bunkerhill ECG-EF AlgorithmPredicate Device: Low Ejection Fraction AI-ECG Algorithm (K232699)Reference Device: Eko Low Ejection Fraction Tool (K233409)Summary
LVEF Not Detected" or "Error - device input criteria not met").or "Error").
HardwareCompatible 12-Lead diagnostic ECG machines with 500Hz digital outputCompatible 12-Lead diagnostic ECG machines with 500Hz digital outputECG recording signals from single lead PCG recording signals (when available)Same
SoftwareBunkerhill proprietary algorithm and applicationBunkerhill proprietary algorithm and applicationproprietary algorithm and applicationSame
Type of InterpretationAdjunctive informationAdjunctive informationSame Type of Interpretation
Intended UserAppropriately trained medical specialists such as cardiologists, front- line clinicians at primary care, urgent care, and emergency care settingsAppropriately trained medical specialists such as cardiologists, front- line clinicians at primary care, urgent care, and emergency care settingsSame Intended User
Rx or OTCRxRxSame Rx or OTC

Performance Data

Safety and performance of the Bunkerhill ECG-EF algorithm has been evaluated and verified in accordance with software specifications and applicable performance standards through Software Development and Validation & Verification Process to ensure performance according to specifications, User Requirements and Federal Regulations and Guidance documents, "Content of Premarket Submissions for Device Software Functions" and a thorough cybersecurity assessment was performed per FDA Guidance "Cybersecurity in medical devices: Quality System Considerations and Content of Premarket Submissions".

The performance of the ECG-EF algorithm was validated through a retrospective study involving 15,994 patient records sourced from two health systems across the United States. The study aimed to assess the diagnostic accuracy of the algorithm in identifying patients with an ejection fraction (EF) < 40% within a clinically and demographically diverse population. The study sample was representative of the U.S population and was 65.5 % White, 18.8% Hispanic, 5.7% American Indian or Alaska Native, 3.9% Asian, 3.0% Black/African American, and 2.8% Other. The sample consisted of 53% Male and 47% Female. The pivotal dataset was be curated from 5 sites geographically distributed throughout the United

Page 11

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 7 of 9

States. The ground truth was established from an echocardiogram using the Simpson's Biplane measurement method taken less than 15-days apart from the ECG scan.

Within this diverse dataset, a total of 1725 LVEF ≤ 40% cases were identified based on the echocardiogram from 15,994 samples (prevalence of 10.8%). The Bunkerhill ECG-EF device achieved a sensitivity of 82.66% (80.90–84.30), a specificity of 83.20% (82.60–83.80), a PPV of 37.20% (35.70–38.76), and NPV of 97.54% (97.28–97.83). The confidence intervals are available in the table below. The confusion matrix summarizing the results are available in Figure 1: Bunkerhill ECG EF Confusion Matrix below.

Performance MetricAcceptance CriteriaValueBootstrapClopper-PearsonWilsonPass/Fail
SensitivitySe ≥ 80%82.66%(80.90–84.30)(80.80 - 84.43)(80.81 - 84..38)Pass
SpecificitySp ≥ 80%83.20%(82.60–83.80)(82.56 - 83.80)(82.56 - 83.79)Pass
PPVPPV ≥ 25%37.20%(35.70–38.76)(35.75 - 38.84)(35.76 - 38.38)Pass
NPVNPV ≥ 95%97.54%(97.28–97.83)(97.25 - 97.81)(97.25 - 97.80)Pass

Figure 1: Bunkerhill ECG-EF Confusion Matrix

Sensitivity
82.66% (80.90-84.30)

Specificity
83.20% (82.60-83.80)

Low LVEF (+) | Not Low LVEF (-)

Predicted: Low LVEF (+) | TP 1,426 | FP 2,399 | PPV 37.20% (35.70-38.76)

Predicted: Not Low LVEF (-) | FN 299 | TN 11,870 | NPV 97.54% (97.28-97.83)

Ground Truth

Figure 1: Bunkerhill ECG-EF Confusion Matrix

Page 12

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 8 of 9

The primary endpoint was Detection of Low Left Ventricular Ejection Fraction and the acceptance criteria were Se/Sp ≥ 80, PPV ≥ 25%, NPV ≥ 95%. All the primary acceptance criteria were successfully met. The secondary acceptance criteria for Detection of Low Left Ventricular Ejection Fraction using AUROC plot was also met.

Subgroup assessments of diagnostic performance were conducted to determine if there was heterogenicity in device performance across clinical sites, demographics, clinical characteristics, co-morbidities, ECG manufacturer and ECG devices. The findings are summarized in Table 3: Subgroup Analysis:

Table 3: Sub-group Analysis

Subgroup AnalysisResult of Test for Heterogeneity
Clinical SiteNot statistically significant
SexNot statistically significant
Race/EthnicityNot statistically significant
Age GroupDiagnostic performance varied across age group strata (p < 0.01). The diagnostic odds ratio was higher than the overall estimate in patients aged 18-49 and 70-79 while it's lower in patients age group of 60-69 and 80+. In younger patients, sensitivity was higher while specificity was lower, whereas in patients of advanced age, sensitivity was lower while specificity was higher.
Clinical CharacteristicsDiagnostic performance varied across certain elements of medical history derived from ICD9/ICD10 codes in patient medical records. The diagnostic odds ratio was higher in patients with Myocardial Infarction (p<0.01) and lower for Coronary Revascularization (p < 0.01) and Atrial Fibrillation/Flutter (p < 0.01). In each of these cases, there was a higher presence of low ejection fraction, the algorithm had higher sensitivity and lower specificity, and as a result there were robust PPV and NPV results.
Device / ManufacturerNot statistically significant
Conduction DisorderNot statistically significant

Note:
This device only should be used with 12-lead ECGs acquired using Ag-AgCl electrodes in the standard lead configuration on the following validated ECG acquisition systems: Philips PageWriter TC70, GE Dash 3000, GE MAC 5500 HD, GE MAC VU360.

Warning (Safety):
Use of non-validated ECG devices, alternative electrode types, or non-standard acquisition conditions may result in inaccurate results.

Conclusions

The Bunkerhill ECG-EF algorithm is as safe and effective as the predicate Low Ejection Fraction AI-ECG Algorithm (K232699). The subject device has the same intended uses and similar indications,

Page 13

K250649 - BunkerHill ECG-EF - Low Ejection Fraction AI-ECG Algorithm 9 of 9

technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended diagnostic use of the device and do not affect its safety and effectiveness when used as labeled. In summary, any minor differences between the Low Ejection Fraction AI-ECG Algorithm and the Bunkerhill ECG-EF Device do not raise any issues of safety or effectiveness.