510(k) Data Aggregation

    K Number
    K233436
    Device Name
    IlluminOss Photodynamic Bone Stabilization System
    Date Cleared
    2024-01-19

    (98 days)

    Product Code
    Regulation Number
    888.3023
    Why did this record match?
    Product Code :

    QAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    The IlluminOss Photodynamic Bone Stabilization System is indicated for use in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, ulvis, fibula, metacarpals, metatarsals, and phalanges. The IlluminOss Photodynamic Bone Stabilization System can also be used in conjunction with FDA-cleared fracture fixation systems to provide supplemental fixation in thes. The IlluminOss System may be used in the femur and tibia to provide supplemental fixation to an anatomically appropriate FDA-cleared fracture fixation system.
    Device Description
    The IlluminOss Photodynamic Bone Stabilization System provides an important treatment option in the fixation and stabilization of fractures through a minimally invasive procedure. The system uses a catheter to deploy an inflatable, noncompliant, thin wall PET balloon into the medullary canal of the bone across the fracture site. The balloon is infused using a syringe with a photodynamic (light cured) monomer that causes the balloon to slowly expand and fill the intramedullary canal of the fractured bone. Activation of the light system allows for visible spectrum light to be delivered through a radially emitting light fiber that is temporarily positioned into a central lumen of the catheter that runs the length of the balloon. With this design, the liquid monomer within the balloon is exposed to light along the entire length of the balloon during the curing process.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202887
    Device Name
    IlluminOss Bone Stabilization System
    Date Cleared
    2020-10-27

    (29 days)

    Product Code
    Regulation Number
    888.3023
    Why did this record match?
    Product Code :

    QAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    The IlluminOss Photodynamic Bone Stabilization System is indicated for use in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, ulna, clavicle, pelvis, fibula, metacarpals, and phalanges. The IlluminOss Photodynamic Bone Stabilization System can also be used in conjunction with FDAcleared fracture fixation systems to provide supplemental fixation in these anatomic sites. The IlluminOss System may be used in the femur and tibia to provide supplemental fixation to an anatomically appropriate FDA-cleared fracture fixation system.
    Device Description
    The IlluminOss Photodynamic Bone Stabilization System provides an important treatment option in the fixation and stabilization of fractures through a minimally invasive procedure. The system uses a catheter to deploy an inflatable, noncompliant, thin wall PET balloon into the medullary canal of the bone across the fracture site. The balloon is infused using a syringe with a photodynamic (light cured) monomer that causes the balloon to slowly expand and fill the intramedullary canal of the fractured bone. Activation of the light system allows for visible spectrum light to be delivered through a radially emitting light fiber that is temporarily positioned into a central lumen of the catheter that runs the length of the balloon. With this design, the liquid monomer within the balloon is exposed to light along the entire length of the balloon during the curing process. The system is currently indicated for use in the humerus, radius, ulna, clavicle, metacarpal, metatarsal, phalanges, ulna, fibula, and anterior ring of the pelvis. The purpose of this Special 510(k) is to expand the indications of the IlluminOss Photodynamic Bone Stabilization System (PBSS) to include use in the femur and tibia in conjunction with FDA-cleared fracture fixation systems.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201961
    Device Name
    IlluminOss Photodynamic Bone Stabilization System
    Date Cleared
    2020-08-13

    (30 days)

    Product Code
    Regulation Number
    888.3023
    Why did this record match?
    Product Code :

    QAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    The IlluminOss Photodynamic Bone Stabilization System is in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, pelvis, fibula, metacarpals, metatarsals, and phalanges. The IlluminOss Photodynamic Bone Stabilization System can also be used in conjunction with FDA-cleared fracture fixation systems to provide supplemental fixation in these anatomic sites.
    Device Description
    The IlluminOss Photodynamic Bone Stabilization System provides an important treatment option in the fixation and stabilization of fractures through a minimally invasive procedure. The system uses a catheter to deploy an inflatable, noncompliant, thin wall PET balloon into the medullary canal of the bone across the fracture site. The balloon is infused using a syringe with a photodynamic (light cured) monomer that causes the balloon to slowly expand and fill the intramedullary canal of the fractured bone. Activation of the light system allows for visible spectrum light to be delivered through a radially emitting light fiber that is temporarily positioned into a central lumen of the catheter that runs the length of the balloon. With this design, the liquid monomer within the balloon is exposed to light along the entire length of the balloon during the curing process. The system is currently indicated for use in the humerus, radius, ulna, clavicle, metacarpal, metatarsal, phalanges, ulna, and anterior ring of the pelvic. The purpose of this Special 510(k) is to expand the indications for use to include the fibula.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200295
    Device Name
    IlluminOss Bone Stabilization System
    Date Cleared
    2020-06-25

    (141 days)

    Product Code
    Regulation Number
    888.3023
    Why did this record match?
    Product Code :

    QAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    The IlluminOss Photodynamic Bone Stabilization System is indicated for use in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, and ulna. The IlluminOss Photodynamic Bone Stabilization System can also be used in conjunction with an FDA-cleared fracture fixation system to provide supplemental fixation in the humerus, radius, and ulna. It is also indicated for use in skeletally mature patients in the treatment of the pelvis, clavicle, metacarpals, metatarsals, and phalanges.
    Device Description
    The IlluminOss Photodynamic Bone Stabilization System provides an important treatment option in the fixation and stabilization of fractures through a minimally invasive procedure. The system uses a catheter to deploy an inflatable, noncompliant, thin wall PET balloon into the medullary canal of the bone across the fracture site. The balloon is infused using a syringe with a photodynamic (light cured) monomer that causes the balloon to slowly expand and fill the intramedullary canal of the fractured bone. Activation of the light system allows for visible spectrum light to be delivered through a radially emitting light fiber that is temporarily positioned into a central lumen of the catheter that runs the length of the balloon. With this design, the liquid monomer within the balloon is exposed to light along the entire length of the balloon during the curing process. The purpose of this 510(k) is to add smaller, and intermediate sizes of the implant and expand the indications for use to include the pelvis, clavicle, metacarpals, metatarsals, and phalanges.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183145
    Device Name
    IlluminOss Photodynamic Bone Stabilization System
    Date Cleared
    2018-12-13

    (30 days)

    Product Code
    Regulation Number
    888.3023
    Why did this record match?
    Product Code :

    QAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    IlluminOss Photodynamic Bone Stabilization System (PBSS) is indicated for use in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, and ulna.
    Device Description
    The IlluminOss Photodynamic Bone Stabilization System provides an important treatment option in the fixation and stabilization of fractures through a minimally invasive procedure. The system uses a catheter to deploy an inflatable, noncompliant, thin wall PET balloon into the medullary canal of the bone across the fracture site. The balloon is infused using a standard 20cc syringe with a photodynamic (light cured) monomer that causes the balloon to slowly expand and fill the intramedullary canal of the fractured bone. Activation of the light system allows for visible spectrum light to be delivered through a radially emitting light pipe that is temporarily positioned into a central lumen of the catheter that runs the length of the balloon. With this design, the liquid monomer within the balloon is exposed to light along the entire length of the balloon during the curing process. The purpose of this Special 510(k) is to add smaller, intermediate, and larger sizes of the implant, an intermediate size of the Sheath/Dilator Set, and a smaller size of the Delivery Set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181228
    Device Name
    IlluminOss Bone Stabilization System
    Date Cleared
    2018-08-31

    (115 days)

    Product Code
    Regulation Number
    888.3023
    Why did this record match?
    Product Code :

    QAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    Device Description
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN160062
    Device Name
    IlluminOss Bone Stabilization System
    Date Cleared
    2017-12-19

    (356 days)

    Product Code
    Regulation Number
    888.3023
    Why did this record match?
    Product Code :

    QAD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    The IlluminOss Photodynamic Bone Stabilization System (PBSS) is indicated for skeletally mature patients in the treatment of impending and actual pathological fractures of the humerus, radius, and ulna, from metastatic bone disease.
    Device Description
    The IlluminOss Photodynamic Bone Stabilization System (PBSS) is intended to be used in the fixation and stabilization of actual and impending pathological fractures of the humerus, radius, and ulna through a minimally invasive procedure. The system uses a catheter to deploy an inflatable, noncompliant, thin wall PET balloon into the medullary canal of the bone across the fracture site. The balloon is infused using a standard 20cc syringe with a photodynamic (light cured) monomer that causes the balloon to slowly expand and fill the intramedullary canal of the fractured bone. Activation of the light system allows for visible spectrum light to be delivered through a radially emitting light pipe that is temporarily positioned within a central lumen of the catheter that runs the length of the balloon. The liquid monomer within the balloon is exposed to light along the entire length of the balloon during the curing process. Curing (and hardening) occurs only when the photo initiator within the monomer is exposed to a specific frequency of light causing rapid polymerization of the monomer resulting in a solid intramedullary (IM) rod. The time to cure the IM rod depends on the size of the balloon used to stabilize the fracture. A Timer Kev, included within each balloon catheter kit, determines the time the light source is activated during the curing process to ensure the appropriate cure time is used for each balloon size.