Search Results
Found 6 results
510(k) Data Aggregation
(444 days)
The iNAP One Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults in whom positive airway pressure is not the preferred treatment choice.
The iNAP One Sleep Therapy System consists of six (6) main components. The components include a console (Model R07-A), a saliva container, a saliva absorbent (iNAP DryPad), a flexible polymer tubing (iNAP Tubing Set), soft polymer oral interfaces (iNAP Oral Interface), a Muffler (Optional) and a software application for mobile devices (iNAP Lab+). One additional accessory set (102C) is included, which is a combination of the iNAP Oral Interface and iNAP Tubing Set. The function of iNAP One Sleep Therapy System (Console Model: R07-A) is to develop a negative pressure gradient in user's oral cavity which can be adjusted between -20 to -90 mmHg (-27 to -122 cmH2O) via the App (iNAP Lab+) to achieve or maintain proper vacuum pressure for users.
Here's a breakdown of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the iNAP One Sleep Therapy System are primarily focused on its pressure setting range, accuracy, and clinical performance in treating Obstructive Sleep Apnea (OSA). The reported device performance is based on the clinical study described.
Acceptance Criteria | Reported Device Performance |
---|---|
Adjustable Pressure Range | -20 to -90 mmHg (-27 to -122 cmH2O) |
Pressure Accuracy | ±5 mmHg (±6.8 cmH2O) |
Clinical Efficacy for OSA | In a subgroup of 18 subjects treated within the intended pressure range (-20mmHg to -90mmHg): |
- 72.2% (13 subjects) achieved a successful response (treated AHI 50% from baseline and treated AHI 5), baseline AHI, and post-treatment AHI after 28 days of therapy. While PSG is the gold standard for diagnosing OSA and establishing AHI, the document does not specify the number of experts (e.g., sleep specialists, polysomnographic technologists) used to interpret these PSG studies or their specific qualifications.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the ground truth from the PSG results. It can be inferred that the PSG interpretations likely followed standard clinical protocols for sleep studies, which inherently involve expert interpretation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance
There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any evaluation of human readers improving with AI assistance. The device is a "Sleep Therapy System" that generates negative pressure for OSA treatment, not an AI-assisted diagnostic tool for human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The device itself is a physical therapy system, not an algorithm. Therefore, a "standalone algorithm only" performance evaluation, as might be done for an AI diagnostic tool, is not applicable in this context. The clinical study evaluates the performance of the device in use by patients.
7. The Type of Ground Truth Used
The ground truth for evaluating the device's efficacy was established using Polysomnogram (PSG), which is considered the gold standard for diagnosing and quantifying Obstructive Sleep Apnea (OSA). The key metric used was the Apnea-Hypopnea Index (AHI).
8. The Sample Size for the Training Set
The document does not mention the existence of a "training set" in the context of machine learning. The clinical study described is an evaluation of the device's performance after its development, not a study to train an algorithm.
9. How the Ground Truth for the Training Set was Established
As no training set (in the machine learning sense) is mentioned, this question is not applicable. The clinical study serves to demonstrate the safety and effectiveness of the device itself.
Ask a specific question about this device
(162 days)
The iNAP One Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults in whom positive airway pressure is not the preferred treatment choice.
The iNAP One Sleep Therapy System consists of six (6) main components. The components are a console, a saliva container, a saliva absorbent (iNAP DryPad), a flexible polymer tubing (iNAP Tubing Set), a soft polymer oral interface (iNAP Oral Interface) and a software application for mobile devices (iNAP Care). One additional accessory is Oral Interface with Tubing, which is a combination of the Oral Interface and Tubing Set. The function of iNAP One Sleep Therapy System is developing a negative pressure gradient in the user's oral cavity, which is set as -40 mmHg.
Please find below the requested information about the acceptance criteria and study proving the device meets these criteria:
1. Table of Acceptance Criteria and Reported Device Performance
Category | Acceptance Criteria | Reported Device Performance (iNAP One Sleep Therapy System) |
---|---|---|
Biocompatibility | Compliance with ISO 10993-1 series for patient contact materials. | All biocompatibility tests (Cytotoxicity, Skin Sensitization, Oral Mucosa Irritation, Pyrogenicity, Leechable & Extractables) passed for all relevant components (Oral Interface, Tubing Set, Oral Interface with Tubing Set). |
Bench Testing | Substantial equivalence to predicate device (Winx) in negative pressure application and maintenance. | Substantially equivalent to Winx device in negative pressure application and maintenance. |
Acoustics | Acoustic power |
Ask a specific question about this device
(347 days)
The Winx Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
The Winx Sleep Therapy System consists of four (4) main components and two (2) sizing methods. The components are: a small electronic bedside console, a soft polymer mouthpiece, a flexible polymer tube that connects the mouthpiece to the console, and a physicians' software application (Winx Data Management Software) that allows clinicians to download usage data from the console and generate patient usage reports. The sizing methods include: a sizing template and an iPhone sizing application (Winx Mouthpiece Sizing Application). The mouthpiece is an intraoral device that is worn during sleep. The system is designed to increase airway patency and decrease airway obstruction.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance for Winx Sleep Therapy System (Winx+ Mouthpiece)
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary focuses on demonstrating substantial equivalence of the Winx+ mouthpiece to the predicate Winx mouthpiece, rather than setting entirely new, independent acceptance criteria for the Winx+ system. However, specific performance targets and safety metrics were evaluated.
Clinical Performance Acceptance Criteria (derived implicitly from non-inferiority claims):
Performance Metric | Acceptance Criteria (for Winx+ to be non-inferior to Winx) | Reported Device Performance (Winx+ Mouthpiece) |
---|---|---|
Effectiveness (Primary Endpoint): | ||
Clinical Success Rate (First Night) | Non-inferiority to Winx mouthpiece, defined as AHI reduction of >50% compared to control PSG AND treatment AHI 50% and treatment AHI 50% (comparing treatment PSG to control PSG) and a treatment AHI of |
Ask a specific question about this device
(82 days)
The Winx Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
The Winx Sleep Therapy System consists of four (4) main components and two (2) sizing methods. The components are: a small electronic bedside console, a soft polymer mouthpiece, a flexible polymer tube that connects the mouthpiece to the console, and a physicians' software application (Winx Data Management Software) that allows clinicians to download usage data from the console and generate patient usage reports. The sizing methods include: a sizing template and an iPhone sizing application (Winx Mouthpiece Sizing Application). The mouthpiece is an intraoral device that is worn during sleep. The system is designed to increase airway patency and decrease airway obstruction.
The provided text is a 510(k) summary for the Winx Sleep Therapy System, which is an intraoral device for obstructive sleep apnea. It describes the device, its intended use, and states that it is substantially equivalent to a predicate device based on performance data from bench testing.
However, the document does not contain the following information:
- A table of acceptance criteria and reported device performance. The document only states that "Results of bench testing demonstrate that the Winx Sleep Therapy System is safe and effective for its intended use and substantially equivalent to the predicate." It does not provide specific performance metrics or acceptance thresholds.
- Details about any clinical study, including sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies. The approval explicitly states it's based on "bench testing" and not clinical trial data.
- Sample sizes for training sets. As there's no mention of a clinical study or an AI/algorithm-only performance evaluation, this information is not applicable and not present.
Therefore, many of your specific questions cannot be answered from the provided text. The document refers to "Performance Data" but then immediately states that "Results of bench testing demonstrate..." This indicates that the evaluation for substantial equivalence was primarily based on non-clinical (bench) testing, not on human clinical trials that would involve the collection of efficacy data against specific acceptance criteria.
Summary of available information vs. requested information:
- A table of acceptance criteria and the reported device performance: Not provided. The document states "Results of bench testing demonstrate that the Winx Sleep Therapy System is safe and effective for its intended use and substantially equivalent to the predicate" but does not quantify performance or acceptance criteria.
- Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable/Not provided. The evaluation was based on bench testing, not a clinical test set.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable/Not provided. No clinical test set.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable/Not provided. No clinical test set.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable/Not provided. This device is for sleep therapy, not an AI diagnostic imaging tool; therefore, MRMC studies are not relevant in this context and not mentioned.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable/Not provided. No mention of an AI algorithm being evaluated.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable/Not provided. The evaluation was based on a comparison to a predicate device via bench testing, not clinical outcomes or expert ground truth.
- The sample size for the training set: Not applicable/Not provided. No AI algorithm is mentioned, and therefore no training set.
- How the ground truth for the training set was established: Not applicable/Not provided. No AI algorithm and no training set.
Ask a specific question about this device
(105 days)
The Winx Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
The Winx Sleep Therapy System consists of four (4) main components: a small electronic bedside console, a soft polymer mouthpiece, a flexible polymer tube that connects the mouthpiece to the console, and a physicians' software application (Winx Data Management Software) that allows clinicians to download usage data from the console and generate patient usage reports. The mouthpiece is an intraoral device that is worn during sleep. The system is designed to increase airway patency and decrease airway obstruction.
The provided text does not contain detailed acceptance criteria, device performance data in a quantitative format, or descriptions of studies that prove the device meets specific acceptance criteria. The document is primarily a 510(k) summary for regulatory clearance, focusing on substantial equivalence to a predicate device.
However, based on the information provided, here's what can be extracted and what is explicitly not available:
1. Table of Acceptance Criteria and Reported Device Performance:
No specific, quantitative acceptance criteria or corresponding device performance values are reported in the provided text. The document broadly states: "Performance testing demonstrate that the Winx Sleep Therapy System is safe and effective for its intended use and substantially equivalent to the predicate." This is a general statement, not a table of specific metrics.
2. Sample Size Used for the Test Set and Data Provenance:
This information is not provided in the text. There is no mention of a specific "test set" or its size, nor is there any information about the country of origin of data or whether it was retrospective or prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not provided in the text. There is no mention of experts, ground truth establishment methods, or their qualifications.
4. Adjudication Method for the Test Set:
This information is not provided in the text.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
This information is not provided in the text. There is no mention of a comparative effectiveness study involving human readers or the effect size of AI assistance. The device in question is a physical device (intraoral pressure device), not an AI algorithm for interpretation.
6. Standalone (Algorithm Only) Performance Study:
This information is not provided in the text. The device is a physical system, not an algorithm, so a "standalone" algorithmic performance study is not applicable in the way it would be for AI software.
7. Type of Ground Truth Used:
This information is not provided in the text. Given the nature of a physical device for OSA treatment, "ground truth" would typically refer to clinical outcomes (e.g., AHI reduction) measured by polysomnography, but this is not mentioned.
8. Sample Size for the Training Set:
This information is not provided in the text. There is no mention of a "training set" as this device is not an AI/ML algorithm that requires training data in the conventional sense.
9. How the Ground Truth for the Training Set Was Established:
This information is not provided in the text. As there's no training set mentioned, the establishment of ground truth for it is also not discussed.
Summary of what is present:
- Indications for Use: The Winx Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
- Substantial Equivalence: The device claims substantial equivalence to the 'Attune Sleep Apnea System' (K111549).
- Performance Claim: "Performance testing demonstrate that the Winx Sleep Therapy System is safe and effective for its intended use and substantially equivalent to the predicate." This is a general statement required for regulatory submission.
To reiterate, the provided text is a regulatory filing summary for a physical medical device, not a scientific publication detailing comprehensive performance studies with acceptance criteria, sample sizes, and ground truth methodologies often associated with AI/ML device evaluations.
Ask a specific question about this device
(301 days)
The Attune Sleep Apnea System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
The Attune Sleep Apnea System consists of three (3) main components: a small electronic bedside console, a soft polymer mouthpiece, and a flexible polymer tube that connects the mouthpiece to the console. A mouthpiece holder is provided for mouthpiece storage and use during weekly system cleaning.
The mouthpiece is an intraoral device that is worn during sleep. The system is designed to increase airway patency and decrease airway obstruction.
Console: The console generates a gentle negative pressure, collects excess saliva, records patient use time, and monitors pressure. The console is provided with a power cord, which connects to a standard electrical outlet. An optional laboratory console is available for use in a sleep laboratory. The laboratory console includes wires that extend from the console. The wires connect to the sleep laboratory's polysomnography (PSG) system, allowing the sleep technicians to view the console's pressure on the same monitor as the other PSG channels.
Mouthpiece: The mouthpiece is provided in ten (10) discrete sizes. Sleep technicians use a bite wax to obtain an impression of each patient's teeth and a sizing template to determine the best mouthpiece size.
Tubing: The tubing connects to the console by screwing into the console base and to the mouthpiece with luer connectors.
Use: The patient connects the system and places the mouthpiece in his or her mouth. The mouthpiece is worn during sleep. The console generates a gentle, negative pressure, which is delivered through the mouthpiece into the oral cavity and holds the tongue and soft palate out of the airway.
Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Prospectively Defined) | Reported Device Performance |
---|---|
AHI reduction of > 50% | Clinical Success observed in 41.3% of the Primary Endpoint Cohort (based on AHI reduction and treated AHI) |
Treated AHI 50% and treated AHI 50%** |
- **Treated AHI
Ask a specific question about this device
Page 1 of 1