K Number
K220907

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-06-16

(444 days)

Product Code
Regulation Number
872.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iNAP One Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults in whom positive airway pressure is not the preferred treatment choice.

Device Description

The iNAP One Sleep Therapy System consists of six (6) main components. The components include a console (Model R07-A), a saliva container, a saliva absorbent (iNAP DryPad), a flexible polymer tubing (iNAP Tubing Set), soft polymer oral interfaces (iNAP Oral Interface), a Muffler (Optional) and a software application for mobile devices (iNAP Lab+). One additional accessory set (102C) is included, which is a combination of the iNAP Oral Interface and iNAP Tubing Set. The function of iNAP One Sleep Therapy System (Console Model: R07-A) is to develop a negative pressure gradient in user's oral cavity which can be adjusted between -20 to -90 mmHg (-27 to -122 cmH2O) via the App (iNAP Lab+) to achieve or maintain proper vacuum pressure for users.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the iNAP One Sleep Therapy System are primarily focused on its pressure setting range, accuracy, and clinical performance in treating Obstructive Sleep Apnea (OSA). The reported device performance is based on the clinical study described.

Acceptance CriteriaReported Device Performance
Adjustable Pressure Range-20 to -90 mmHg (-27 to -122 cmH2O)
Pressure Accuracy±5 mmHg (±6.8 cmH2O)
Clinical Efficacy for OSAIn a subgroup of 18 subjects treated within the intended pressure range (-20mmHg to -90mmHg):- 72.2% (13 subjects) achieved a successful response (treated AHI < 5).- 22.2% (4 subjects) achieved a partial response (AHI reduction > 50% from baseline and treated AHI < 15).- Mean Baseline AHI: 36.01 ± 15.87- Mean Treated AHI: 5.53 ± 9.53 (P-value < 0.001 based on paired-t test).
Safety ProfileNo adverse events reported in the clinical study.
Non-inferiority to Predicate Device (K193460)Performance determined to be non-inferior.

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for Test Set: 30 OSA patients were initially enrolled in the clinical study. A subgroup analysis was performed on 18 subjects who received iNAP treatment within the intended pressure range of -20mmHg to -90mmHg.
  • Data Provenance: Not explicitly stated, but the submission is from Somnics, Inc., located in Taiwan, and the regulatory consultant is in San Juan Capistrano, CA, USA. The study is described as a "clinical study," which implies it was a prospective study conducted for the purpose of evaluating the device. The specific country of origin for the patient data is not mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The text states that Polysomnogram (PSG) was used to evaluate inclusion criteria (AHI > 5), baseline AHI, and post-treatment AHI after 28 days of therapy. While PSG is the gold standard for diagnosing OSA and establishing AHI, the document does not specify the number of experts (e.g., sleep specialists, polysomnographic technologists) used to interpret these PSG studies or their specific qualifications.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the ground truth from the PSG results. It can be inferred that the PSG interpretations likely followed standard clinical protocols for sleep studies, which inherently involve expert interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any evaluation of human readers improving with AI assistance. The device is a "Sleep Therapy System" that generates negative pressure for OSA treatment, not an AI-assisted diagnostic tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

The device itself is a physical therapy system, not an algorithm. Therefore, a "standalone algorithm only" performance evaluation, as might be done for an AI diagnostic tool, is not applicable in this context. The clinical study evaluates the performance of the device in use by patients.

7. The Type of Ground Truth Used

The ground truth for evaluating the device's efficacy was established using Polysomnogram (PSG), which is considered the gold standard for diagnosing and quantifying Obstructive Sleep Apnea (OSA). The key metric used was the Apnea-Hypopnea Index (AHI).

8. The Sample Size for the Training Set

The document does not mention the existence of a "training set" in the context of machine learning. The clinical study described is an evaluation of the device's performance after its development, not a study to train an algorithm.

9. How the Ground Truth for the Training Set was Established

As no training set (in the machine learning sense) is mentioned, this question is not applicable. The clinical study serves to demonstrate the safety and effectiveness of the device itself.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 16, 2023

Somnics Inc. % Feng-Yu Lee Principal Regulatory Consultant IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212 San Juan Capistrano, California 92675

Re: K220907

Trade/Device Name: The iNAP One Sleep Therapy System Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: OZR Dated: May 17, 2023 Received: May 18, 2023

Dear Feng-Yu Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220907

Device Name The iNAP One Sleep Therapy System

Indications for Use (Describe)

The iNAP One Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults in whom positive airway pressure is not the preferred treatment choice.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SOMNICS" in a bold, sans-serif font. The letters are all capitalized and in a dark blue color. The letters are evenly spaced and the word is centered in the image. The background is white.

510 (k) Summary

Date Prepared: June 16, 2023

510(k) number: K220907

1. Applicant Information

Somnics, Inc. 5F, No. 22, Sec. 2, Shengyi Rd. Zhubei City, Hsinchu County, 30261 Taiwan Contact Person: Chung Chu Chen Chief of Executive Officer Somnics, Inc. Email: chungchu1@somnics.com Tel.: +886-3-550-9623

c/o IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212, San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee TEL: 1-949-218-0929 FAX: 1-949-218-0928

2. Device Information

Trade Name:The iNAP One Sleep Therapy System
Common Name:Intraoral Pressure Gradient Device
Classification:Class II
Classification name:Intraoral devices for snoring and intraoral devices forsnoring and obstructive sleep apnea (21 CFR 872.5570)
Product Code:OZR

3. Predicate Device

Trade Name:The iNAP One Sleep Therapy
System Manufacturer:Somnics, Inc.
510(k)#:K193460

4. Indications for Use:

The iNAP One Sleep Therapy System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults in whom positive airway pressure is not the preferred treatment choice.

{4}------------------------------------------------

5. Device Description:

The iNAP One Sleep Therapy System consists of six (6) main components. The components include a console (Model R07-A), a saliva container, a saliva absorbent (iNAP DryPad), a flexible polymer tubing (iNAP Tubing Set), soft polymer oral interfaces (iNAP Oral Interface), a Muffler (Optional) and a software application for mobile devices (iNAP Lab+). One additional accessory set (102C) is included, which is a combination of the iNAP Oral Interface and iNAP Tubing Set. The function of iNAP One Sleep Therapy System (Console Model: R07-A) is to develop a negative pressure gradient in user's oral cavity which can be adjusted between -20 to -90 mmHg (-27 to -122 cmH2O) via the App (iNAP Lab+) to achieve or maintain proper vacuum pressure for users.

iNAP One Console

The console generates a gentle negative pressure, collects excess saliva and is driven by a built-in rechargeable Li-ion battery.

iNAP Saliva Container

The saliva container is attached directly to the console and retains up to 100 ml of saliva. An opening with the membrane is at the bottom of the container to connect the console.

iNAP DryPad (Saliva Absorbent)

The saliva absorbent is inserted into the saliva container to minimize foaming formed from saliva. The saliva absorbent retains over 30ml of saliva and is to be disposed off after each use regardless full or not.

iNAP Tubing Set

The Tubing Set (T01) is the means of connecting between the console and oral interface with custom connectors. Different styles of Oral Interface are used with this Tubing Set.

iNAP Oral Interface

The Oral Interface is provided in four (4) styles. Patients can choose one with optimum fitting and result.

Muffler (Optional)

The Muffler is to reduce the noise of the device by stabilizing the air flow from the iNAP outlet. It is an optional accessory for users to decrease the noise of iNAP device during use. To use the Muffler, the user attaches the Muffler to the bottom of the iNAP device.

iNAP Lab+ (Mobile App)

The mobile app for patients is intended to keep usage records for personal reference, including pressure setting, intensity setting and usage time (compliance records). It also allows the patients to set intensity and authorized users (physician) to set pressure.

iNAP Oral Interface with Tubing

The Oral Interface with Tubing (102C) is a combination of the Oral Interface and Tubing Set with an adjustable function. The material of the oral interface is silicone rubber.

{5}------------------------------------------------

Product NameThe iNAP One Sleep TherapySystem(Subject Device/K220907)The iNAP One Sleep TherapySystem(Predicate Device/K193460)
Intended UseThe iNAP One Sleep TherapySystem is indicated for home usein the treatment of obstructivesleep apnea (OSA) in adults inwhom positive airway pressure isnot the preferred treatment choice.Same
Console ModelR07-AR07-B
CompatibleAppiNAP Lab+ App(for R07-A and R07-B)iNAP Care App
Functions onAppCompliance recordPressure SettingIntensity SettingCompliance record
PressureAdjustable pressure range -20 to 90 mmHg (-27 to -122 cmH2O)Set as -40 mmHg. (-54 cmH2O)
CompatibleOral InterfaceI07/ I07S/ I07M/ I02C /I07HI07/ I07S/ I07M/ I02C
MufflerOptionalN/A
Mechanism ofActionPressure gradient developed in oralcavity supplied and maintained viathe oral interface to the patientmouth.Same
Single/MultipleUseSingle Person / Multiple UseSame

6. Comparison of Technological Characteristics with the Predicate Device

Comparison of Technological Characteristics A.

ProductNameThe iNAP One SleepTherapy System (SubjectDevice/K220907)The iNAP One SleepTherapy System(PredicateDevice/K193460)Comment
Energy SourceOperation mode:Rechargeable Lithium ionbattery Charge mode: 5VDC, 2ASameIdentical
Human FactorsUse during the sleepperiod. A user operates theconsole, saliva container,and oral interface duringuseSameIdentical
Design –ComponentsConsoleSaliva ContainerSaliva AbsorbentsSameIdentical
Tubing SetSame
Oral interfaceSame
Oral interfaceAdditional accessory: Oral Interface with TubingSame
App (iNAP Lab+)with adjustable settingsApp (iNAP Care)without adjustable settingsSimilar, with adjustable settings
Design - EnergyUsed andDeliveredThe adjustable pressure is -20 to -90 mmHg (-27 to -122 cmH2O), and theaccuracy is$±5 mmHg (±6.8 cmH2O)$The setting of negativepressure is -40 mmHg, andthe accuracy is ±10%Similar
Design -Console size5.98" x 3.14" x 1.41"(152mm x 80mm x 36mm)SameIdentical
Design - Weight Weight: 0.47 lbs. (0.21kg)with batteriesSameIdentical
Design - DatastorageFlash memory in MCU(64K bytes, storage data:operation time/ duration,pumping duration/leaking)SameIdentical
Design - Salivacontainervolume100 mlSameIdentical
Design - OralInterfaceStructure(Materials)Oral Interface Structure:1. Oral Interface connector2. Flexible tubewith vacuum port3. Lip shield4. Tongue shieldSameIdentical
Design -VacuumDeliveringLocation in OralCavityOne vacuum port betweenupper palate and tongueSameIdentical
Design -OralInterface -PatientContactingMaterials107 series Oral Interface:Polymers (thermoplasticelastomer, polycarbonate)I02C Oral Interface withTubing: Silicone &polypropylene107 series Oral Interface:Polymers (thermoplasticelastomer, polycarbonate)I02C Oral Interface withTubing: Silicone &polypropyleneSame materialsand sameproductionprocesses.107Hhassimilarstructure.
SterilityNon-sterileSameIdentical
Product NameThe iNAP One SleepTherapy System (SubjectDevice/K220907)The iNAP One SleepTherapy System(PredicateDevice/K193460)Comment
Principle ofOperationThe iNAP One SleepTherapy System provides anegative pressure gradientwithin the oral cavity andthe pressure gradient pullsthe tongue toward the upperpalate and also pulls the softpalate forward as well toreduce or eliminate theobstruction of the upperairway, allowing a patient tobreathe from nosenaturally during sleep.SameIdentical
BiocompatibilityBiocompatibility evaluationbased on ISO 10993-1SameIdentical
OperatingconditionsOperating Temperature &Humidity: 5 to 40°C and 15to 93% humidity(noncondensing) based onIEC 60601-1SameIdentical
Storage conditionsStorage Temperature &Humidity: -20 to 50℃ and15 to 93% humidity(noncondensing) based onIEC 60601-1SameIdentical
AcousticsAcoustic power < 30 dB perISO 7779SameIdentical
ElectromagneticCompatibility(EMC)Compliance: IEC 60601-1-2AIM 7351731SameIdentical
Electrical SafetyElectrical safety (FollowIEC 60601-1):Class II Equipment, TypeBF, IP22, ContinuousOperationSameIdentical
Safety for homeenvironmentCompliance:IEC 60601-1-11SameIdentical
Performance -Negative pressureSetting andaccuracy-20 to -90 mmHg ± 5mmHg(-27 to -122 ± 6.8 cmH2O)40 mmHg (±10%)Similar, withuser adjustablesettings. Safetyand performancevalidated usingclinical testing.

{6}------------------------------------------------

{7}------------------------------------------------

B. Principle of Operation and Non-Clinical Performance Data

CONFIDENTIAL

{8}------------------------------------------------

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

Electrical Safety and Battery Safety: Testing was conducted per:

  • IEC 60601-1:2012: Medical electrical equipment Part 1: General requirements for ● basic safety and essential performance.
  • IEC 60601-1-11:2015: Medical electrical equipment Part 1-11: General . requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
  • . IEC 62133 Edition 2.0 2012-12: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications.

Electromagnetic compatibility (EMC): Testing was conducted per:

  • IEC 60601-1-2:2014: Medical electrical equipment Part 1-2: General ● requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • AIM Standard 7351731 Rev. 2.00 2017-02-23 Medical Electrical Equipment and . System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers - An AIM Standard.

Software and Cybersecurity:

  • Software verification and validation testing were conducted, and documentation . was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since a failure or latent flaw in the software could lead to delay in delivery of appropriate medical care that would likely lead to minor injury.
  • Cybersecurity concerns for monitoring system were addressed in accordance with . FDA's "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices: Guidance for Industry and Food and Drug Administration Staff" document issued October 2, 2014.

Biocompatibility:

  • . The patient contacting components for the subject device come under the category of surface contact (mucosal membrane) with permanent contact duration (> 30 days). No new biocompatibility testing was deemed necessary due to the lack of significant changes in patient contacting components compared to the predicate (Testing was conducted based on the FDA guidance on Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process").
    Functional Testing:

  • Functional and mechanical testing of oral interfaces after shelf life and repeated cleaning. This includes pressure stability testing.

  • Device noise evaluation with and without optional muffler. ●

{9}------------------------------------------------

8. Discussion of Clinical Tests Performed

A clinical study with 30 OSA patients was conducted to evaluate the outcomes of iNAP treatment following the pressure configuration process. Polysomnogram (PSG) was used to evaluate the inclusion criteria (AHI > 5), baseline AHI, and post-treatment AHI after 28 days of therapy. In this study, a subgroup analysis revealed that 18 subjects received iNAP treatment within the intended pressure range of -20mmHg to - 90mmHg. The rest received higher negative pressure in order to assess device safety and therapy tolerance at high negative pressures. Of these 18 subjects. 72.2% (13 subjects) experienced successful response (defined as treated AHI < 5) and 22.2% (4 subjects) experienced partial response (defined as AHI reduction > 50% from baseline and treated AHI < 15) with iNAP at pressures below -90mmHg. An improvement in mean treated AHI was observed for patients who receive therapy pressures below -90mmHg: Baseline AHI = 36.01 ±15.87; Treated AHI: 5.53 ± 9.53 (P value < 0.001 based on paired-t test). No adverse events were reported during this study. The performance of the subject device was determined to be non-inferior to the FDA-cleared predicate device (K193460).

9. Conclusions

Based on the indications for use, product performance, and clinical information provided in this notification, the subject device has been shown to be substantially equivalent to the currently marketed predicate device.

§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.

(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”