(301 days)
The Attune Sleep Apnea System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
The Attune Sleep Apnea System consists of three (3) main components: a small electronic bedside console, a soft polymer mouthpiece, and a flexible polymer tube that connects the mouthpiece to the console. A mouthpiece holder is provided for mouthpiece storage and use during weekly system cleaning.
The mouthpiece is an intraoral device that is worn during sleep. The system is designed to increase airway patency and decrease airway obstruction.
Console: The console generates a gentle negative pressure, collects excess saliva, records patient use time, and monitors pressure. The console is provided with a power cord, which connects to a standard electrical outlet. An optional laboratory console is available for use in a sleep laboratory. The laboratory console includes wires that extend from the console. The wires connect to the sleep laboratory's polysomnography (PSG) system, allowing the sleep technicians to view the console's pressure on the same monitor as the other PSG channels.
Mouthpiece: The mouthpiece is provided in ten (10) discrete sizes. Sleep technicians use a bite wax to obtain an impression of each patient's teeth and a sizing template to determine the best mouthpiece size.
Tubing: The tubing connects to the console by screwing into the console base and to the mouthpiece with luer connectors.
Use: The patient connects the system and places the mouthpiece in his or her mouth. The mouthpiece is worn during sleep. The console generates a gentle, negative pressure, which is delivered through the mouthpiece into the oral cavity and holds the tongue and soft palate out of the airway.
Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Prospectively Defined) | Reported Device Performance |
|---|---|
| AHI reduction of > 50% | Clinical Success observed in 41.3% of the Primary Endpoint Cohort (based on AHI reduction and treated AHI) |
| Treated AHI < 20 | Included in the definition of Clinical Success |
Note: The document explicitly states "Clinical Success was prospectively defined as AHI reduction of > 50% and treated AHI < 20." The reported device performance for these criteria is intertwined within the "Clinical Success" rate.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 63 subjects in the Primary Endpoint Cohort.
- Data Provenance:
- Country of Origin: Not explicitly stated, but the study was described as a "multi-center" trial, suggesting it likely took place across several locations, possibly within the United States given the FDA submission.
- Retrospective or Prospective: Prospective. The study is described as "a four-week, multi-center, prospective, open label, randomized first-night order of control vs. treatment, single-arm trial."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The text does not detail how the AHI (Apnea-Hypopnea Index) measurements were established or adjudicated for the clinical trial, nor does it mention the involvement or qualifications of experts for this purpose.
4. Adjudication Method for the Test Set
This information is not provided in the document. The text does not describe any specific adjudication method (e.g., 2+1, 3+1) used for the clinical trial results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
This information is not applicable/not provided. The device described is the "Attune Sleep Apnea System," which is an intraoral device for treating sleep apnea, not an AI or imaging diagnostic tool that would typically involve human readers or MRMC studies.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This information is not applicable/not provided. The device is a physical intraoral system, not an algorithm. Therefore, a standalone algorithm performance study is not relevant.
7. The Type of Ground Truth Used
The ground truth used for assessing the device's effectiveness was based on physiological measurements of sleep apnea, specifically:
- AHI reduction of > 50%
- Treated AHI < 20
These measurements are standard metrics derived from polysomnography (PSG) studies, which are considered the gold standard for diagnosing and assessing the severity of sleep apnea. The document mentions "polysomnography (PSG) system" in relation to the laboratory console, implying PSG was the method for collecting relevant data.
8. The Sample Size for the Training Set
This information is not explicitly provided as this type of device (physical medical device) typically undergoes traditional clinical trials, not machine learning model training. The concept of a "training set" in the context of AI/ML is not directly relevant here. However, the document does mention "two feasibility studies" that preceded the pivotal trial.
9. How the Ground Truth for the Training Set Was Established
As with point 8, the concept of a "training set" and its ground truth establishment in an AI/ML context is not directly applicable. For the clinical trials, the "ground truth" (i.e., the actual AHI values) would have been established through standard clinical diagnostic procedures, likely polysomnography, as accepted in sleep medicine.
{0}------------------------------------------------
2. 510(K) SUMMARY
MAR 3 0 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(a).
Date Summary Prepared: March 29, 2012
510(k) Number: K111549
510(k) Owner Information:
ApniCure, Inc. 900 Chesapeake Drive Redwood City, CA 94063 Phone Number: (650) 361-9300 (650) 361-9399 Fax Number:
Contact Person:
Cindy Domecus, R.A.C. (US & EU) Principal, Domecus Consulting Services LLC (650) 343-4813
Device Information:
Trade Name: Attune Sleep Apnea System · Intraoral device for obstructive sleep apnea Common Name: Classification: Class II Intraoral devices for snoring and intraoral devices for snoring Classification Name: and obstructive sleep apnea (21 CFR 872.5570, Product Code OZR, Intraoral Pressure Gradient Device)
Physical Description:
The Attune Sleep Apnea System consists of three (3) main components: a small electronic bedside console, a soft polymer mouthpiece, and a flexible polymer tube that connects the mouthpiece to the console. A mouthpiece holder is provided for mouthpiece storage and use during weekly system cleaning.
The mouthpiece is an intraoral device that is worn during sleep. The system is designed to increase airway patency and decrease airway obstruction.
Console
The console generates a gentle negative pressure, collects excess saliva, records patient use time, and monitors pressure. The console is provided with a power cord, which connects to a standard electrical outlet. An optional laboratory console is available for use in a sleep laboratory. The laboratory console includes wires that extend from the console. The wires connect to the sleep laboratory's polysomnography (PSG) system, allowing the sleep technicians to view the console's pressure on the same monitor as the other PSG channels.
{1}------------------------------------------------
Mouthpiece
The mouthpiece is provided in ten (10) discrete sizes. Sleep technicians use a bite wax to obtain an impression of each patient's teeth and a sizing template to determine the best mouthpiece size.
Tubing
The tubing connects to the console by screwing into the console base and to the mouthpiece with luer connectors.
Use
The patient connects the system and places the mouthpiece in his or her mouth. The mouthpiece is worn during sleep. The console generates a gentle, negative pressure, which is delivered through the mouthpiece into the oral cavity and holds the tongue and soft palate out of the airway.
Indications for Use:
The Attune Sleep Apnea System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults.
Predicate Devices:
The Attune Sleep Apnea System is substantially equivalent in intended use, indications for use, and technological characteristics to the following devices:
| Name | Manufacturer | 510(k) # |
|---|---|---|
| SnoreSilencer Pro | Respironics | K033822 |
| Oasys Oral Airway System | Mark Abramson | K030440 |
| Provent Professional SleepApnea Therapy | Ventus Medical | K071560 |
| Repose Bone ScrewSystem | Influence, Inc. | K981677 |
| Prelude III™ TongueSuspension System | Siesta Medical, Inc. | K110127 |
| Sleepstyle 200 SeriesHC234 CPAP, Opus Mask | Fisher & Paykel | K040941,K063036 |
| S8 Aspen CPAP | Resmed Ltd. | K091947 |
Non-clinical Performance Data:
Non-clinical performance testing included functional testing, biocompatibility testing, software validation and cleaning validation. Results of non-clinical testing demonstrate that
{2}------------------------------------------------
the Attune Sleep Apnea System is safe and effective for its intended use and substantially equivalent to the predicates.
Clinical Performance Data:
Clinical performance testing included data from two feasibility studies and one pivotal trial. The pivotal trial was a four-week, multi-center, prospective, open label, randomized firstnight order of control vs. treatment, single-arm trial of the Attune Sleep Apnea System for the treatment of obstructive sleep apnea. The objective of the study was to demonstrate safety and effectiveness of the Attune Sleep Apnea System. Clinical Success was determined per patient and was prospectively defined as AHI reduction of > 50% and treated AHI < 20. There were 63 subjects in the Primary Endpoint Cohort. Clinical Success was observed in 41.3% (4/15 mild, 10/18 moderate, 12/30 severe) of the Primary Endpoint Cohort comparing First Treatment Night with Control. Results of clinical testing demonstrate that the Attune Sleep Apnea System is safe and effective for its intended use and substantially equivalent to the predicates.
Conclusion:
Based on the indications for use and non-clinical and clinical data provided in this 510(k) notification, the subject device has been shown to be safe, effective, and substantially equivalent to the currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The overall design is simple and conveys a sense of government authority and public service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Chris Daniel Executive Vice President Apnicure, Incorporated 900 Chesapeake Drive Redwood City, California 94063
MAR 3 0 2012
Re: K111549 Trade/Device Name: Attune Sleep Apnea System Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea. Regulatory Class: II Product Code: OZR Dated: March 15, 2012 Received: March 26, 2012
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Daniel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INDICATIONS FOR USE STATEMENT 1.
510(k) Number (if known):
Device Name: Attune Sleep Apnea System
Indications for Use:
The Attune Sleep Apnea System is indicated for home use in the treatment of obstructive sleep apnea (OSA) in adults. huller A
(Division Sign-Off) Division of Anesthesiology, General Hospital :nfection Control, Dental Devices
1111549 510(k) Number: _
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ApniCure, Inc.
CONFIDENTIAL
§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.
(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”