Search Results
Found 13 results
510(k) Data Aggregation
K Number
K233775Device Name
Vivatmo pro
Manufacturer
Bosch Healthcare Solutions GmbH
Date Cleared
2024-02-22
(90 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Vivatmo pro nitric oxide test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment of FeNO by Vivatmo pro is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti- inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO measurements are to be used as an adjunct to established clinical assessments.
Device Description
Vivatmo pro nitric oxide test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath.
Ask a Question
K Number
K213611Device Name
Fenom Pro
Manufacturer
CAIRE Diagnostics Inc.
Date Cleared
2023-06-07
(569 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Fenom Pro is a portable, non-invasive device to measure fractional exhaled nitric oxide (FENO) in human breath. FENO is increased in some airway inflammatory processes, such as asthma, and decreases in response to anti-inflammatory treatment. Fenom Pro measures fractional exhaled nitric oxide according to guidelines established by the American Thoracic Society.
Measurement of FENO by Fenom Pro is a non-invasive quantitative method to measure the decrease in FENO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FENO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro is suitable for adults and children 6 years and older.
Fenom Pro should be used in a point-of-care healthcare setting under professional supervision. Fenom Pro should not be used in critical care, emergency care or in anesthesiology.
Device Description
Fenom Pro is a non-invasive point-of-care breath analyzer for the quantitative measurement of fractional exhaled nitric oxide (FENO) in expired human breath.
Fenom Pro is comprised of four major components. The main unit contains a touch screen interface for the user and houses the nitric oxide sensor and pneumatics needed to sample the patient's breath. The patient exhales into the Fenom Pro though a single use disposable mouthpiece which is attached to the handpiece. The handpiece is connected to the main unit via a breath tube.
For testing, the patient inhales to total lung capacity and then slowly exhales into the mouthpiece for 6 or 10 seconds, depending on the mode of operation. Approximately 28 seconds after a successful breath maneuver, the FENO concentration is displayed in parts per billion (ppb) on the device touch screen. The device has a daily quality control procedure to ensure consistent device performance.
Ask a Question
K Number
K203695Device Name
NObreath®
Manufacturer
Bedfont Scientific Ltd
Date Cleared
2021-12-17
(364 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The NObreath® is a portable. non-invasive device for the measurement of Fractional Exhaled Nitric Oxide (FeNO) in human breath. The production of nitric oxide is often found to be increased in inflammatory conditions such as asthma. Measurement of FeNO by NObreath® is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels.
The fractional NO concentration in expired breath (FeNO), can be measured by NObreath® according to guidelines for NO measurement established by the American Thoracic Society.
NObreath® is intended for children, 7- 17 years, and older. NObreath® 12 second test mode is for age 7 and up
NObreath® 10 second test mode is for ages 7-10 only who cannot successfully complete a 12 second test.
FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti- inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NObreath® cannot be used with infants or by children under the age of 7 as measurement requires patient cooperation.
NObreath® should not be used in critical care, emergency care or in anesthesiology.
Device Description
NObreath® is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO). The NObreath® system is comprised of the main unit with AC adapter, a rechargeable battery, an electrochemical NO sensor, disposable patient mouthpiece with filter. The device can connect to the PC via a standard USB cable or wirelessly via Bluetooth.
For testing, the patient inhales deeply and slowly exhales for 10 or 12 seconds through the patient filter. In approximately 12 seconds the NO concentration is displayed in parts per billion (ppb). Results are processed using dedicated software. The device has built-in system control procedures and a calibration to be performed every 12 months.
Wireless Bluetooth Low Energy (BLE) is used as a means of communication between the monitor and FeNOchart™ software running on a PC. The FeNOchart™ software is a charting program that retrospectively collects data from the NObreath® monitor when it is not monitoring. It is not time critical, there are no alarms
Ask a Question
K Number
K182874Device Name
Fenom Pro Nitric Oxide Test
Manufacturer
Spirosure, Inc.
Date Cleared
2019-02-13
(124 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Fenom Pro™ Nitric Oxide Test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment of FeNO by Fenom Pro™ is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older.
Testing using the Fenom Pro™ should only be done in a point-of-care healthcare setting under professional supervision. Fenom Pro™ should not be used in critical care, emergency care or in anesthesiology.
Device Description
Fenom Pro™ is a point-of-care breath analyzer that uses solid-state electrochemical technology to measure the fraction of exhaled nitric oxide (FeNO), a marker for airway inflammation, in human exhaled breath. Measurement of FeNO by Fenom Pro is a quantitative and non-invasive method to indicate therapeutic effects of anti-inflammatory pharmacological therapy in patients with elevated FeNO levels. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older.
Fenom Pro uses solid state, potentiometric, sensor technology sensitive to nitric oxides (NO) compounds. The solid state sensor is fluidly preceded by a reactive filter material that renders (oxidizes) potentially confounding species such as carbon monoxide (CO), ammonia (NH4), and methanol (CH4O) inactive, or inert, to the NO sensor. Fenom Pro provides visual and audible feedback during its use. The visual and audible feedback is especially important during the FeNO measurement such that the user can modulate their breath speed within the flow parameters required by the American Thoracic Society (ATS) and the European Respiratory Society (ERS) standards.
Fenom Pro is comprised of four major components. The main unit contains a touch screen interface for the use as well as houses the nitric oxide sensor and pneumatics needed to sample the patient's breath. The patient interfaces with Fenom though the mouthpiece which is attached to the handpiece. The handpiece is connected to the main unit via a breath tube. The handpiece contains a breath conditioning cartridge which prepares the breath sample from the patient for proper analysis in the main unit. Both the mouthpiece and the breath conditioning cartridge are consumables.
Ask a Question
K Number
K170983Device Name
NIOX VERO
Manufacturer
Circassia AB
Date Cleared
2017-11-22
(233 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
NIOX VERO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some airway inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX VERO according to guidelines for NO measurement established by the American Thoracic Society.
Measurement of FeNO by NIOX VERO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX VERO is suitable for children, 7- 17 years, and adults 18 years and older.
NIOX VERO 10 second test mode is for age 7 and up
NIOX VERO 6 second test mode is for ages 7-10 only who cannot successfully complete a 10 second test.
FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NIOX VERO is intended for prescription use and should only be used as directed in the NIOX VERO User Manual by trained healthcare professionals. NIOX VERO cannot be used with infants or by children under the age of 7 as measurement requires patient cooperation.
NIOX VERO should not be used in critical care, emergency care or in anesthesiology.
Device Description
NIOX VERO is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO). NO levels increase during allergic airway inflammation. Measurement of changes in FeNO concentration is used in evaluating a patient's response to antiinflammatory therapy, as an adjunct to established clinical and laboratory assessments.
The NIOX VERO system is comprised of the NIOX VERO unit with AC adapter, a rechargeable battery, an electrochemical NO sensor, disposable patient filters, and an exchangeable handle containing an internal NO scrubber filter. The NIOX Panel is an optional PC application for operation of the NIOX VERO from a PC and access to electronic medical record systems. The device can connect to the PC via a standard USB cable or wirelessly via Bluetooth.
The patient empties their lungs, inhales deeply through the patient filter to total lung capacity and then slowly exhales for 10 seconds. A 6 second mode is available for children aged 7 – 10 who cannot perform a 10 second exhalation. In approximately one minute, the NO concentration is displayed in parts per billion (ppb).
The NIOX VERO unit includes a sampling and gas conditioning system and a man-machine interface (MMI). The instrument controls the inhaled sample appropriately via the electronics and software program. Filtering of inhaled air elimination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s so that it is standardized for all patients. The sample enters an electromechanical sensor and interacts with an electrolyte. A chemical reaction takes place where electrons are generated proportional to the number of NO molecules.
Results are processed using dedicated software. In order to verify the device's performance and reliability of measurements, there are built-in system control procedures and a Quality Control procedure to be performed on a daily basis.
Ask a Question
K Number
K150233Device Name
NIOX VERO Airway Inflammation Monitor
Manufacturer
AEROCRINE AB
Date Cleared
2015-02-26
(24 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
NIOX VERO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX VERO according to guidelines for NO measurement established by the American Thoracic Society.
Measurement of FeNO by NIOX VERO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX VERO is suitable for children, approximately 7 - 17 years, and adults 18 years and older.
FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NIOX VERO is intended for prescription use and should only be used as directed in the NIOX VERO User Manual by trained healthcare professionals. NIOX VERO cannot be used with infants or by children approximately under the age of 7, as measurement requires patient cooperation. NIOX VERO should not be used in critical care, emergency care or in anesthesiology.
Device Description
NIOX VERO is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO). NO levels increase during allergic airway inflammation. Measurement of changes in FeNO concentration is used in evaluating a patient's response to antiinflammatory therapy, as an adjunct to established clinical and laboratory assessments.
The NIOX VERO system is comprised of the NIOX VERO unit with AC adapter, a rechargeable battery, an electrochemical NO sensor, disposable patient filters, and an exchangeable handle containing an internal NO scrubber filter. The NIOX Panel is an optional PC application for operation of the NIOX VERO from a PC and access to electronic medical record systems. The device can connect to the PC via a standard USB cable or wirelessly via Bluetooth.
The patient empties their lungs, inhales deeply through the patient filter to total lung capacity and then slowly exhale for 10 seconds. In approximately one minute, the NO concentration is displayed in parts per billion (ppb).
The NIOX VERO unit includes a sampling and gas conditioning system and a man-machine interface (MMI). The instrument controls the inhaled sample appropriately via the electronics and software program. Filtering of inhaled air elimination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s so that it is standardized for all patients. The sample enters an electromechanical sensor and interacts with an electrolyte. A chemical reaction takes place where electrons are generated proportional to the number of NO molecules.
Results are processed using dedicated software. In order to verify the device's performance and reliability of measurements, there are built-in system control procedures and a Quality Control procedure to be performed on a daily basis.
Ask a Question
K Number
K133898Device Name
NIOX VERO AIRWAY INFLAMMATION MONITOR
Manufacturer
AEROCRINE AB
Date Cleared
2014-11-06
(321 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
NIOX VERO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX VERO according to guidelines for NO measurement established by the American Thoracic Society.
Measurement of FeNO by NIOX VERO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX VERO is suitable for children, approximately 7 - 17 years, and adults 18 years and older.
FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NIOX VERO is intended for prescription use and should only be used as directed in the NIOX VERO User Manual by trained healthcare professionals. NIOX VERO cannot be used with infants or by children approximately under the age of 7, as measurement requires patient cooperation. NIOX VERO should not be used in critical care, emergency care or in anesthesiology.
Device Description
NIOX VERO is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO). NO levels increase during allergic airway inflammation. Measurement of changes in FeNO concentration is used in evaluating a patient's response to anti-inflammatory therapy, as an adjunct to established clinical and laboratory assessments.
The NIOX VERO system is comprised of the NIOX VERO unit with AC adapter, a rechargeable battery, an electrochemical NO sensor, disposable patient filters, and an exchangeable handle containing an internal NO scrubber filter. The NIOX Panel is an optional PC application for operation of the NIOX VERO from a PC and access to electronic medical record systems.
The patient empties their lungs, inhales deeply through the patient filter to total lung capacity and then slowly exhale for 10 seconds. In approximately one minute, the NO concentration is displayed in parts per billion (ppb).
The NIOX VERO unit includes a sampling and gas conditioning system and a man-machine interface (MMI). The instrument controls the inhaled sample appropriately via the electronics and software program. Filtering of inhaled air elimination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s so that it is standardized for all patients. The sample enters an electromechanical sensor and interacts with an electrolyte. A chemical reaction takes place where electrons are generated proportional to the number of NO molecules.
Results are processed using dedicated software. In order to verify the device's performance and reliability of measurements, there are built-in system control procedures and a Quality Control procedure to be performed on a daily basis.
Ask a Question
K Number
K123683Device Name
NIOX PANEL FOR USE WITH THE NIOX MINO AIRWAY INFLAMMATION MONITOR
Manufacturer
AEROCRINE AB
Date Cleared
2012-12-27
(27 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
NIOX MINO measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX MINO according to guidelines for NO measurement established by the American Thoracic Society.
Measurement of FeNO by NIOX MINO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX MINO is suitable for children approximately 7 - 17 years, and adults 18 years and older.
FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. NIOX MINO should only be used as directed in the NIOX MINO User Manual and the NIOX MINO Quality Control Test User Manual, by trained physicians, nurses, respiratory therapists and laboratory technicians. NIOX MINO cannot be used with infants or by children approximately under the age of 7, as measurement requires patient cooperation. NIOX MINO should not be used in critical care, emergency care or in anaesthesiology.
Device Description
NIOX MINO is a small, hand-held, portable system for the non-invasive, online, quantitative measurement of fractional nitric oxide (NO) concentration in expired human breath (FeNO) measured in parts per billion (ppb). The device is intended for routine clinical use and laboratory assessments of the patient's condition.
Measurement of changes in FeNO concentration is used in evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to established clinical and laboratory assessments of asthma. All results are to be interpreted in conjunction with other clinical and laboratory assessments of the patient's condition.
The NIOX MINO unit includes a sampling and gas conditioning system. The valves and pumps of the instrument are automatically controlled to handle the inhaled sample appropriately via the instrument electronics and software program. Filtering of inhaled air eliminates contamination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s.
Results are processed using dedicated software and are expressed as the NO concentration in ppb. In order to verify the device's performance and reliability of measurements, there are builtin system control procedures and a special designed External Quality Test Program to be performed on a daily basis.
NIOX Panel is an optional software program accessory which provides an additional display for operating the NIOX MINO Airway Inflammation Monitor by allowing the user to operate the instrument from their personal computer (PC). Interaction with the NIOX Panel is performed with common human interface devices such as keyboards, mice, etc. A USB interface is used for communication with connected NIOX MINO instruments. Instrument supervision and measurement analysis is still performed by the NIOX MINO instrument, however. The NIOX Panel merely serves as an additional interface to the instrument thus, complementing the instrument's LCD screen.
Ask a Question
K Number
K101034Device Name
NIOX MINO MODEL 09-1000
Manufacturer
AEROCRINE AB
Date Cleared
2010-09-02
(141 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
NIOX MINO measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some inflammatory processes such as asthma. The fractional NO concentration in expired breath (FENO), can be measured by NIOX MINO according to guidelines for NO measurement established by the American Thoracic Society.
Measurement of FENO by NIOX MINO is a quantitative, non-invasive, simple and safe method to measure the decrease in FENO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FENO levels. NIOX MINO is suitable for children, approximately 7 - 17 years, and adults 18 years and older.
FENO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. NIOX MINO should only be used as directed in the User Manual and by physicians, nurses, respiratory therapists and laboratory technicians. NIOX MINO cannot be used with infants or by children approximately under the age of 7, as measurement requires patient cooperation. NIOX MINO should not be used in critical care, emergency care or in anaesthesiology.
Device Description
NIOX MINO is a small, hand-held, portable system for the non-invasive, online, quantitative measurement of the fractional nitric oxide (NO) concentration in expired human breath (FENO) measured in parts per billion levels (ppb). The device is intended for routine clinical use and is suitable for point of care settings.
Measurement of changes in the FENO concentration is used in evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to established clinical and laboratory assessments of asthma. All results are to be interpreted in conjunction with other clinical and laboratory assessments of the patient's condition.
The NIOX MINO unit includes a sampling and gas conditioning system and a man-machine interface (MMI). The user is guided by the built-in touch-screen display through the breathing maneuver by use of the interactive MMI. The valves and pumps of the instrument are automatically controlled to handle the inhaled sample appropriately via the instrument electronics and software program. Filtering of inhaled air eliminates contamination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s.
Results are processed using dedicated software and are expressed as the Nitric Oxide concentration in parts per billion (ppb).
To be able to verify the performance of the device and reliability of measurements, there are built-in system control procedures and a special designed External Quality Control Test Program, to be performed on a daily basis.
Ask a Question
K Number
K083617Device Name
APIERON INSIGHT ENO SYSTEM
Manufacturer
APIERON, INC.
Date Cleared
2009-01-27
(50 days)
Product Code
MXA
Regulation Number
862.3080Why did this record match?
Product Code :
MXA
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The intended use of the Apieron INSIGHT™ eNO System is to quantitatively measure exhaled nitric oxide (eNO) in expired human breath as a marker of inflammation in persons with asthma. Measurement of eNO in expired human breath by the Apieron INSIGHT eNO System is a non-invasive, simple and safe method to measure a decrease in eNO in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of the therapeutic effects in patients with elevated eNO levels. The Apieron INSIGHT eNO System is suitable for use in children ages 8 to 17 years of age, and in adults 18 years of age and older. eNO measurements, as an adjunct to established clinical assessments, provide the physician an objective marker to evaluate the patient's response to anti-inflammatory therapy. The Apieron INSIGHT eNO System can be used by trained operators in a physician's office laboratory setting. The Apieron INSIGHT eNO System should not be used in critical care, emergency care or in anesthesiology.
Device Description
The proposed Apieron INSIGHT™ eNO System consists of a Monitor, a multi-use disposable Sensor Cartridge and the same disposable test supplies as the cleared device. The only difference between the proposed devices and the cleared device is that the single-use disposable eNO Sensor Cartridge of the INSIGHT System has been tested and proven to support up to 10 patient uses and 5 QC tests. A modification has been made to the software which will allow for counting the uses of the Sensor Cartridge and prevention of uses beyond the specified limits. Two sensors are being proposed for clearance one for 10 patient uses and 5 QC tests and one for 5 patient uses and 5 QC tests. The Monitor functions the same and contains the same measurement and breath sampling hardware to provide for a user-friendly interface guiding the operator through the test sequence and the patient through the breath sampling maneuver. The 10-use and the 5-use disposable Sensor Cartridges contain the same Biosensor as in the cleared device which changes its optical transmission properties when it reacts with the nitric oxide in the breath sample. Apart for some software changes to deal with the number of sensor uses, and the addition of more air purges between tests to clear the sensor of nitric oxide from the previous test, the proposed Apieron INSIGHT eNO System is unchanged. It is transportable, operated by an AC outlet and designed for use with the Sensor Cartridge. The proposed Apieron INSIGHT eNO System has the same indication for use as the predicate device, the Apieron INSIGHT eNO System (K073265).
Ask a Question
Page 1 of 2