Search Filters

Search Results

Found 14 results

510(k) Data Aggregation

    K Number
    K220231
    Device Name
    GyroStim
    Date Cleared
    2022-04-27

    (90 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GyroStim is intended to assist in the treatment of balance disorders and vestibular dysfunction in adult patients, age 18 and older, up to 400lb. It is intended to be used by a trained medical professional as prescribed by physician in a clinical environment.

    Device Description

    GyroStim is an AC powered computer-controlled multi-axial positional and rotational chair. The device consists of a patient chair mounted within a rotational frame assembly that is capable of rotating in the pitch and yaw axes. The rotational frames are automated with computer-controlled drive systems capable of executing preprogrammed positional and rotational motion profiles. The system provides clinicians with a safe and efficient means for administering a wide range of specific positional and rotational motions for inducing stimulation of the vestibular system.

    GyroStim operational software runs on a pc-based desktop computer. The software can capture, monitor, and display the patient's subjective response to each run. The response data may be used by the clinician for clinical decision support (CDS) to assist with progress evaluation, for profile selection and rate of therapy advancement, for individualizing therapy sessions, and for monitoring and preventing overstimulation.

    Patient data is collected and stored and may be accessed for graphing, analysis, and reports.

    The software includes libraries of prewritten profiles. A library of pre-written positional profiles located on the RUN BPPV page contains a collection of profiles for executing commonly used canalith repositioning maneuvers. A library of pre-written motion profiles located on the RUN GyroStim page contains a collection of profiles for executing motion induced vestibular stimulation.

    Specifically, the library of motion profiles contains 30 levels of motion induced vestibular stimulation intensity, ranging from Level 1 to Level 30. The lowest level of motion induced vestibular stimulation intensity is Level 1, which rotates at a rate of 1 revolution per minute (RPM) or 0.016667 Hz. The highest level of motion induced vestibular stimulation intensity is Level 30. which rotates at 30 RPM or 0.5Hz. The intensity of the 30 levels differs and increments linearly at a rate of 1 RPM or 0.01667Hz per level; i.e.. Level 1 rotates at 1 RPM or 0.01667 Hz, Level 2 rotates at 2 RPM or 0.03334 Hz, Level 3 rotates at 3 RPM or 0.05 Hz, and so on up to Level 30 which rotates at 30 RPM or 0.5 Hz. Additionally, each of the 30 Levels offer a selection of run durations at 15 seconds (s), 30s, 60s, and 180s.

    When configured with the integrated patient transfer system, GyroStim provides accessibility to the large population of patients with moderate to severe mobility challenges, many of whom are limited or unable to engage in the existing standard of care for balance disorders and vestibular dysfunction.

    The GyroStim assembly, operational software, and libraries of profiles provide clinicians with a safe, effective, and efficient medical device for assisting in the treatment of balance disorders and vestibular dvsfunction.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the GyroStim device, based on the provided text:

    Acceptance Criteria and Device Performance

    Item TestedTest MethodPurposeAcceptance CriteriaReported Device Performance
    V4-GYROSTIM III-04A (Electrical Safety)ANSI/AAMI/ES 60601-1, Ed 3.1Electrical safetyComplies with ANSI/AAMI/ES 60601-1, Ed 3.1Pass
    V4-GYROSTIM III-04A (EMC)ANSI/AAMI/IEC 60601-1-2, Ed 4EMCComplies with Class A, Professional Healthcare Environment of ANSI/AAMI/IEC 60601-1-2, Ed 4Pass
    V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Static Loading Safety Factors)Engineering Analysis per Clause 9.8.2 and Table 21 of IEC 60601-1, Ed 3.1Determination of static loading safety factorsSafety factor determined to be greater than the 2.5x requirements of Table 21 of IEC 60601-1Pass, Min. safety factor, 27.5x
    V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Floor Loading Distribution)Engineering AnalysisDetermination of floor loading distribution for installation into non-residential environmentsComplies with uniform and concentrated live load requirements per IBC, 2018*Pass
    V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Reliability and Fatigue)Engineering Analysis (including fatigue and dynamic loading, and retrospective data analysis)Reliability analysis, including fatigue and dynamic loading, and retrospective data analysisDevice shall remain reliable and not cause patient or end user injury for 10-year shelf lifePass
    V4-GYROSTIM III-04, V4-GYROSTIM III-04A, GyroStim software (Software Verification)IEC 62304:2006/A1:2016, clause 5.7 and FDA's Guidance for Industry and FDA Staff, "Guidance for the Content for the Premarket Submissions for Software Contained in Medical Devices".Software verificationSoftware complies with pre-defined requirements of the software requirement specification for a Class B / "moderate" level of concern due to software failurePass
    V4-GYROSTIM III-04, V4-GYROSTIM III-04A, GyroStim software (Accuracy of Controls)Measurement of rotational accuracy in pitch and yaw directionsAccuracy of ControlsSystem meets the defined accuracy requirements for RPM, acceleration, and degrees of rotationPass
    V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Summative Usability Study)IEC 62366-1, 2015 +A2020 and FDA's Guidance for Industry and FDA Staff, "Applying Human Factors and Usability Engineering to Medical Devices"Summative Usability StudyNo instances of use errors or close calls with potential for harmPass, none observed

    Study Details

    The provided document describes a preliminary clinical study and various non-clinical bench (performance) tests.

    1. Clinical Study:

    • Sample size used for the test set and the data provenance:
      • The document states that the study "utilized retrospective data from the two treatment groups: GyroStim group, and SOC Group." However, it does not specify the exact sample size (number of patients) used in either group.
      • The data provenance is retrospective. The country of origin of the data is not specified.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      • The assessments and treatments were performed by a "vestibular and balance medical specialist (DPT)."
      • The document does not mention a number of experts for establishing ground truth or their specific qualifications beyond "DPT." It implies one such specialist conducted the assessments and treatments.
    • Adjudication method: Not described or applicable, as it was a retrospective comparison of treatment outcomes, not a diagnostic study requiring expert consensus on individual cases.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      • No, this was not an MRMC study. The GyroStim is a physical device for therapy, not an AI for image interpretation or diagnosis by human readers.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      • The document describes the performance of the device itself in both non-clinical and clinical contexts. The "algorithm" (software) of GyroStim is part of the device's control system, and its verification was done as a standalone test (software verification). However, the overall clinical study evaluated the device in use by a trained medical professional, so it's not a purely "algorithm-only" performance in the clinical sense.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      • Outcomes data obtained from "pre and post intervention assessments" using "quantitative equipment and methods, including SOT, DHI, ABC, DVA, and mDGI." The "ground truth" for effectiveness was based on the improvement in these objective and subjective balance scores.
    • The sample size for the training set: Not applicable. The clinical study was a comparison of treatment outcomes using an existing device (GyroStim) against standard of care. There is no mention of a machine learning model being trained on a separate dataset for clinical decision-making within the scope of this FDA summary. The "RUN GyroStim profiles" are pre-written profiles within the software, not learned models from a training set in the typical AI sense.
    • How the ground truth for the training set was established: Not applicable, as there was no explicit training set described for an AI/ML model for clinical decision-making.

    2. Non-Clinical Bench (Performance) Testing:

    • Sample size used for the test set and the data provenance: This refers to physical units of the device (V4-GYROSTIM III-04A, V4-GYROSTIM III-04, GyroStim software). The sample size implicitly refers to the device units tested to ensure compliance with the specified standards. The data provenance is from internal engineering analysis and testing.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: These tests rely on established engineering standards and test methods (e.g., IEC, ANSI/AAMI). The "ground truth" is compliance with these standards, verified by qualified engineers or test facilities. The specific number or qualifications are not detailed beyond adherence to the standards.
    • Adjudication method: Not applicable for bench testing. Pass/fail criteria are defined by the standards.
    • MRMC study: Not applicable.
    • Standalone performance: Yes, these are standalone performance tests of the device and its software components against engineering and safety standards.
    • Type of ground truth: Defined engineering and safety standards (e.g., ANSI/AAMI/ES 60601-1, IEC 60601-1-2, IEC 62304, IBC 2018).
    • Sample size for training set: Not applicable.
    • How ground truth for training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200529
    Device Name
    Orion
    Manufacturer
    Date Cleared
    2020-08-12

    (163 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orion rotary chair is an optional accessory for VisualEyes 515 eye movement recording systems.

    The VisualEyes™ system provides information to assist in the nystagmographic evaluation, diagnosis and documentation of vestibular disorders. VNG testing evaluates nystagmus using goggles mounted with cameras. These images are measured, recorded, displayed and stored in the software. This information can then be used by a trained medical professional to assist in diagnosing vestibular disorders. The target population for videonystagmography is five years of age and above.

    Device Description

    The Orion is a rotary chair designed to assess the Vestibular Ocula Reflex (VOR).

    Orion rotary chair includes these three variants

    • Orion Reclining
    • Orion Auto Traverse
    • Orion Comprehensive .

    The Orion is considered an accessory to vestibular examination system software designated VisualEyes 525 and VisualEyes 515 manufactured by Interacoustics (FDA 510(k) K200534).

    The Orion is an update of the System 2000 by Micromedical ((FDA 510(k) ID K922037).

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Orion Rotary Chair). It does not contain details about acceptance criteria or a study proving the device meets those criteria in a quantitative sense.

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device (System 2000 by Micromedical). The "performance tests" mentioned are primarily comparative, asserting that the new device performs "as specified" and is "safe and effective" based on these comparisons, rather than against predefined, quantitative acceptance criteria. There are no clinical tests performed.

    Therefore, many of the requested items cannot be extracted from the provided text.

    Here is a summary of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or a formal table comparing them against reported performance. Instead, it offers a "Comparison table" between the Orion device and its predicate (System 2000) for various descriptive and technical characteristics, aiming to show equivalence rather than adherence to specific numeric performance thresholds.

    DescriptionOrion Model (Reclining) Performance (Reported as 'Same' or 'Equivalent')System 2000 Model (Reclining) Performance (Predicate)Equivalence Justification
    Intended useThe Orion rotary chair is an optional accessory for VisualEyes 525/VisualEyes 515 eye movement recording systemsThe System 2000 rotary chair is an optional accessory for eye movement recording systemsSame
    Software to supportVisualEyes 515/525VisualEyes 515/525 and SpectrumSame (Spectrum is the replaced software). Evaluated in K163149.
    Chair and ControlleryesyesSame
    Video Goggle connectoryes USByes FirewireEquivalent - USB and Firewire are communication protocols
    Patient Weight Maximum350 lbs (158kilo)350 lbs (158kilo)Same
    VOR, VVOR, VFXyes .01-.64 Hzyes .01-.64 HzSame
    Step testsmax 200max 200Same
    VOR AnalysisyesyesSame
    Equipment cartyesyesSame
    Ocularmotor TestsyesyesSame
    VORTEQ/VHIT/DVAOptionOptionSame
    EOGOptionOptionSame
    Mechanical Foot BrakeElectric Lockyes (manual foot brake)Equivalent - both provide an immobile state for patient safety; electronic lock allows chair to be immobile before test and prevents rotation during test.
    Sinusoidal Frequency0.01 to 0.64 Hz0.01 to 0.64 HzSame
    Step Velocity (max)200 deg/sec200 deg/secSame
    Acceleration (max)100 deg/sec²100 deg/sec²Same
    Auto Traverse / Comprehensive Models Differences
    Chair and ControllerOrion: Controller is inside the Chair baseSystem 2000: Controller is a separate unitEquivalent - merely a design difference.
    Patient Weight Maximum400 lbs400 lbsSame
    Step Velocity (max)350 deg/sec300 deg/secEquivalent - typical Step Velocity test parameters do not exceed 180 d/s; SVV test uses 300 d/s. Higher max velocity is not critical.
    Lateral movement speed0.8 cm/sec1 cm/secEquivalent - speed to go offset is not a critical parameter; lateral offset controls stimulation of Otolith organs, not the speed to attain it.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. No formal test set or clinical study is described. The "performance tests" involved comparisons of the physical device features and operational parameters against the predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No ground truth was established by experts for a test set, as no clinical study or diagnostic performance assessment involving human interpretation is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a rotary chair for vestibular testing, not an AI-assisted diagnostic imaging device that would typically undergo an MRMC study. No AI component is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device (rotary chair) that requires operation by a trained medical professional; it is not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The basis for safety and effectiveness is substantial equivalence to a predicate device, demonstrated through comparative technical specifications and internal performance testing, not against a clinical ground truth.

    8. The sample size for the training set

    Not applicable. This document describes a physical medical device, not a machine learning model; therefore, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192652
    Device Name
    TRV
    Manufacturer
    Date Cleared
    2020-05-27

    (246 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRV is intended to assist in the diagnosis and treatment of balance disorders and vertigo, including benign paroxysmal positional vertigo.

    Device Description

    The TRV is a multi-axial chair that can rotate 360° around both the horizontal and vertical axes. With the patient secured by a four-point harness, a head mount, a leg strap, and shoulder pads, an examiner is able to rotate the patient around the plane of each of the 6 semicircular canals and hold the patient in any position for detailed examination of the semicircular canals. The TRV chair is manually handled by the healthcare professional. The axes of rotation are lockable in preset positions. The primary axis has a battery-powered electromagnetic lock controlled by a footswitch and the secondary axis is manually locked with a mechanical lever. A mechanical system with an adjustable counterweight ensures that the weight of the chair and the patient are balanced during the maneuvers.

    AI/ML Overview

    The provided text describes the TRV device and its substantial equivalence to a predicate device, the Epley Omniax™. However, it does not contain information related to a study that establishes acceptance criteria and then proves the device meets those criteria, especially in the context of an AI-powered device or for clinical accuracy.

    Instead, the document details a 510(k) submission for a physical medical device (a multi-axial chair) used in the diagnosis and treatment of balance disorders. The performance data presented focuses on electrical safety, electromagnetic compatibility, and mechanical testing, not on predictive performance, accuracy, or diagnostic effectiveness in the way an AI/software device would be evaluated.

    Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" in an AI/software context are not present in the provided document.

    Here's a breakdown of what can and cannot be answered based on the provided text:


    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (General Safety and Performance)Reported Device Performance
    Electrical Safety: Comply with IEC 60601-1Passed: Device complies with IEC 60601-1 for electrical safety.
    Electromagnetic Compatibility (EMC): Comply with IEC 60601-1-2Passed: Device complies with IEC 60601-1-2 standard for EMC.
    Mechanical Testing: Demonstrate long-term safety and withstand worst-case stress loads.Passed: Wear observed after testing is considered not to have any effect on the safety of patients or operators.
    Technological Equivalence (to predicate device): Similar indications for use, use environment, user, patient population, test options, maneuvers, movement range, and power supply (with supported differences).Achieved: Found to be as safe, as effective, and performs comparably to the predicate device, with technological differences supported by non-clinical data.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable / Not Provided: The document does not describe a "test set" in the context of data for an algorithm. The "testing" refers to physical device safety and mechanical performance. There is no data provenance for a diagnostic or predictive model.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable / Not Provided: No "ground truth" establishment for a diagnostic output is described, as this is a physical chair, not an AI diagnostic tool.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable / Not Provided: No adjudication method for a data test set is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No: A multi-reader multi-case (MRMC) study was not done. The device is a physical chair, not an AI assistant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No: This is not an algorithm. The device is a physical chair, and human healthcare professionals manually operate it. Its function is to assist professionals in performing maneuvers, not to make a standalone diagnosis or treatment decision without human input.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable / Not Provided: See point 3. The "ground truth" in this context pertains to compliance with safety regulations and functional equivalence to a predicate device, not diagnostic accuracy.

    8. The sample size for the training set

    • Not Applicable / Not Provided: No training set is described, as this is a physical device and not an AI model.

    9. How the ground truth for the training set was established

    • Not Applicable / Not Provided: No training set or ground truth establishment for it is described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183661
    Date Cleared
    2019-10-22

    (299 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearEdge® Balance System is intended for use to assess sway as an indicator of balance. Individual suitability for assessment must be judged on a case by case basis, by a qualified including those certified and/or licensed in their state to prescribe and/or use balance devices such as certified athletic trainers, physical therapists, chiropractors, nurses and physicians. The ClearEdge® Balance System can be used wherever compatible Android mobile operating devices can be used.

    Device Description

    The ClearEdge® Balance System is a mobile software system that analyzes balance through measurements of postural sway. The System is comprised of application software, and accompanying hardware consisting of a mobile computing device, proprietary sensor, balance pad, and friction pad. The test data is collected, uploaded to a HIPAA compliant server, analyzed by Motion Intelligence developed algorithms and presented in report form for clinician's review. The reports indicate how a subject's balance score may have changed between testing sessions, and whether the measured change is likely due to measurement error and normal human variation in performance of the selected balance stance or a real change in performance caused by an external factor.

    Using proprietary software on an Android tablet and the Edge Sensor, the ClearEdge Balance System tests balance by administering eight select balance stances.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the ClearEdge Balance System:

    This document focuses on demonstrating substantial equivalence to a predicate device (Sway™ Balance System) rather than absolute performance against pre-defined, numerical acceptance criteria for a novel device. Therefore, the "acceptance criteria" presented are implicitly derived from the comparison to the predicate and statistical demonstrations of equivalence.


    1. Table of Acceptance Criteria and Reported Device Performance

    The core "acceptance criteria" for demonstrating substantial equivalence are based on demonstrating that the ClearEdge Balance System's performance is equivalent enough to the predicate device (Sway™ Balance System) to not raise new questions of safety or effectiveness. This is shown through a combination of laboratory testing and human performance testing.

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (ClearEdge Balance System)
    Laboratory Equivalence (Scoring)Inverted Pendulum Testing:
    (Regression Analysis) Scores highly correlated with predicate.Linear Regression: r-values for all tests were greater than 0.97. Slopes of linear regression ranged between 0.92 and 1.05 (p
    Ask a Question

    Ask a specific question about this device

    K Number
    K190735
    Device Name
    EQ Balance
    Date Cleared
    2019-08-16

    (148 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EQ Balance is intended to assess body tilt as an indicator of balance. Individual suitability for assessment must be judged on a case by case basis, by a qualified including those certified and/or licensed in their state to prescribe and/or use balance devices such as certified athletic trainers and coaches, physical therapists, nurses and physicians. Interpretation of EQ Balance results should only be made by a healthcare professional, as part of a multifaceted assessment of balance. Conditions affecting postural sway include nausea, headache, orthopedic injury, ear infection, medications, head injury, dehydration and fatigue. EO Balance should be used in an indoor setting free from obstructions and free from excessive noise or visual distractions. EQ Balance should be only be used as directed by your healthcare practitioner. EQ Balance may be used on mobile devices running iOS or Android operating systems.

    Device Description

    EQ Balance is a mobile balance test to be used on mobile phone and tablet devices with iOS version 10.3 or later, or Android OS version 6 or later. EQ Balance measures balance using the accelerometer within the device.

    EQ Balance utilizes an algorithm that calculates a balance score based on the analysis of subject movements detected by the device's internal accelerometer. Balance is measured during a series of stances with the mobile device held to the chest by the user. Users follow the step-by-step voice guidance that runs after launching the app during the entirety of the test to ensure consistency in obtaining balance measures. EQ Balance does not function as a diagnostic test, and EQ Balance results should be interpreted by qualified individuals only.

    AI/ML Overview

    The provided text details the 510(k) summary for the EQ Balance device, which is intended to assess body tilt as an indicator of balance. The study's primary goal was to demonstrate substantial equivalence to the predicate device, Sway Balance.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the performance testing verified that "EQ Balance and the Sway™ Balance System (Sway) are in statistical agreement." The agreement was primarily assessed using Spearman's analysis and Deming Regression Model.

    Acceptance Criteria / Performance MetricAcceptance CriteriaReported Device PerformanceComments
    Statistical Agreement (3-pose average) with Predicate (Sway Balance)Not explicitly stated, implied to be demonstrable statistical agreementSpearman's analysis: "the 3-pose average measures calculated by EQ Balance were in statistical agreement with the 3-pose average obtained with Sway Balance for both normal and balance-impaired groups, and for all three stances (double leg, tandem right and left)."
    Deming Regression Model: "showed agreement between the 3-pose average balance measures (the primary device output), as well as 23 of the 27 additional analyses that were conducted."The study aimed to show agreement rather than a specific statistical threshold for acceptance.
    Intraclass Correlation (ICC) for consistency of EQ Balance measureNot explicitly stated. Implied to be strong consistency.ICC for the cohort was 0.75 (p
    Ask a Question

    Ask a specific question about this device

    K Number
    K173669
    Device Name
    K-D Balance
    Date Cleared
    2018-04-02

    (123 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    K-D Balance is an objective balance assessment tool and is intended to be used for individuals for whom a balance measurement is a desired outcome. K-D Balance is intended for use to assess balance performance. Individual suitability for assessment must be judged on a case by case basis, by a qualified individual including those certified or licensed in their state to prescribe and use balance devices including but not limited to athletic trainers and coaches, physical therapists, nurses, physicians, and other health care providers. Patients with could affect balance include, but are not limited to, vestibular dysfunction, nausea, headaches, orthopedic injury, inner ear infection or dysfunction, neurological conditions, head injury, medication side effects, dehydration and fatigue. K-D Balance can be used wherever an iOS mobile operating device is available. K-D Balance may only be used by health care professionals.

    Device Description

    K-D Balance is a mobile balance test to be used on the 6th generation iPhone or iPod touch with iOS versions 9.0 or later. K-D Balance measures balance using motion sensors already built-in to the iPhone or iPod touch devices. K-D Balance utilizes a proprietary algorithm that calculates a balance score based on the analysis of subject movements detected by the device's internal accelerometer. Balance is measured during a series of stances with the mobile device secured to the subject with a device holder. Health care professionals follow the step-by-step voice guidance that runs after launching the app during the entirety of the test to ensure consistency in obtaining balance measures. K-D Balance does not function as a diagnostic test, and K-D Balance results should be interpreted by healthcare professionals only.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the K-D Balance device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria with specific numerical thresholds for performance metrics. Instead, the primary acceptance criterion appears to be statistical agreement with the predicate device (Sway Balance™) for balance outcomes in both normal and balance deficit groups, across various stances.

    Acceptance Criteria (Inferred)Reported Device Performance (K-D Balance)
    Statistical agreement with predicate device (Sway Balance) for balance outcomes in normal subjects.K-D Balance measures were in statistical agreement with Sway Balance outcomes for the normal group for the double leg, tandem right, and tandem left leg stances, according to linear regression analysis and Deming Regression Model. This supports substantial equivalence.
    Statistical agreement with predicate device (Sway Balance) for balance outcomes in subjects with balance deficits.K-D Balance measures were in statistical agreement with Sway Balance outcomes for the balance deficit group (including subjects with brain injury) for the double leg, tandem right, and tandem left leg stances, according to linear regression analysis and Deming Regression Model. Poorer balance scores were observed for subjects with balance deficits. This supports substantial equivalence.
    Conformance to product specification (general).K-D Balance met specification.
    Sensitivity of the software to access data from the accelerometer.Bench testing analyzed the sensitivity of the software to access data from the ST Microelectronics MEMS Accelerometer. Sensitivity scores with K-D Balance were comparable to the predicate.
    Safety (no adverse effects or complications during or after testing).No adverse effects or complications were encountered during or after performance testing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 79 subjects (25 males, 54 females; mean age: 38.63 ± 12.88, range: 18-65).
      • 34 subjects had brain injury or other neurological conditions (balance deficits).
      • 45 subjects were non-injured (normal group).
    • Data Provenance: The study was conducted by the University of Texas, Southwestern Medical Center, Neurology and Sports Medicine Department. The data is prospective as testing was performed for the purpose of this device evaluation. The country of origin is the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This study does not appear to use human experts to establish "ground truth" in the traditional sense of diagnostic interpretation. Instead, the "ground truth" for balance performance is based on the objective measurements derived from the predicate device (Sway Balance™) and the clinical categorization of subjects (injured vs. non-injured).

    The document mentions that the study was conducted by the University of Texas, Southwestern Medical Center, Neurology and Sports Medicine Department, implying medical professionals at that institution were involved in classifying subjects (e.g., identifying those with brain injury or neurological conditions). However, specific numbers or qualifications of individual experts for this categorization are not provided.

    4. Adjudication Method for the Test Set

    Not applicable. This was a comparative performance study against a predicate device and clinical categories, not an expert-driven adjudication of AI output. The "ground truth" for comparison was the predicate device's output and the subject's clinical status (injured/non-injured).

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not explicitly described. This study focused on the performance of the K-D Balance device itself in comparison to a predicate device, not on how the K-D Balance assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The performance testing directly evaluated the K-D Balance application's ability to measure balance and its statistical agreement with the predicate device (Sway Balance™). The balance assessment is performed by the application's algorithm, and the output is a score. Human involvement is in administering the test and securing the device, but the balance measurement itself is algorithmic.

    7. The Type of Ground Truth Used

    The ground truth for this comparison study was:

    • Objective measurements from a legally marketed predicate device: The balance scores generated by the Sway Balance™ application.
    • Clinical categorization: Subjects were categorized as "normal" or "balance deficit" (which included those with brain injury or other neurological conditions). Poorer balance scores were expected and observed in the deficit group, confirming the clinical relevance of the measurements.

    8. The Sample Size for the Training Set

    The document does not provide information regarding a specific training set size or how the K-D Balance algorithm was developed (i.e., whether it used a training set in machine learning terms). The description focuses on its proprietary algorithm and its use of motion sensors.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described, the method for establishing ground truth for a training set is not available in the provided text. The evaluation is focused on the device's performance against a predicate and clinical categories rather than a machine learning model's training process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121590
    Date Cleared
    2012-09-20

    (112 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sway™ Balance System is intended for use to assess sway as an indicator of balance. Individual suitability for assessment must be judged on a case by case basis, by a qualified individual including those certified and/or licensed in their state to prescribe and/or use balance devices such as certified athletic trainers and coaches, physical therapists, nurses and physicians. Conditions affecting postural sway include nausea, headache, orthopedic injury, ear infection, medications, head injury, dehydration and fatigue. The Sway™ Balance System can be used wherever an iOS mobile operating device is available.

    Device Description

    The Sway Balance™ System is a mobile measurement system that analyzes balance through thoracic sway, using the built in accelerometer of a mobile device. The Sway Balance™ System is a stand-alone mobile operating system software application that does not include any peripheral hardware add-ons.

    AI/ML Overview

    The provided text is a 510(k) summary for the Sway Balance™ System, a mobile application designed to assess balance. Based on this document, here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Equivalence/Accuracy: The device should accurately measure sway as an indicator of balance, comparable to established methods.Bench Testing: Analyzed the sensitivity of the software program to access data from the built-in accelerometer. "Sensitivity scores using the Sway Balance™ Software were comparable" to those of a force platform.
    Clinical Testing (Correlation): Compared the Sway Balance™ System to force platform assessment tools. "Results showed no significant difference between the two data sets (p =
    Ask a Question

    Ask a specific question about this device

    K Number
    K102364
    Device Name
    NYDIAG 200
    Manufacturer
    Date Cleared
    2010-12-23

    (125 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nydiag 200 rotary chair is an optional accessory for the Interacoustics eye movement recording systems. The rotary chair is used to elicit Vestibulo-Ocular Reflexes in patients aged 2 years or older who weigh less than 135 kg / 300 lbs who are being tested for bilateral or unilateral lesions in the peripheral vestibular organs.

    Device Description

    The Nydiag 200 is a rotation chair, including a chair unit and a motor pedestal unit. The Nydiag 200 is controlled by software. The rotating chair is an optional accessory and is used for stimulation of the patients balance organs. The chair is supplied from the mains and controlled from a computer through an USB connection. The chair unit includes the chair itself attached to a frame assembly, seat belts and other restraints, patient discomfort signal device and goggle connections (Firewire and USB). The pedestal includes a power unit and motor and gearing units.

    AI/ML Overview

    The Nydiag 200 Rotary Chair is a medical device, and the provided document is a 510(k) premarket notification summary. This type of submission focuses on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving efficacy or performance against specific, quantifiable acceptance criteria in the same way a new drug or novel high-risk device might.

    Therefore, many of the typical performance study details (like sample size, ground truth establishment, expert qualifications, and MRMC studies) are not present in this type of submission for this particular device. The "acceptance criteria" here are primarily about meeting safety standards and demonstrating equivalence to the predicate device in terms of technological characteristics and intended use.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Demonstration for 510(k))Reported Device Performance
    Safety and EfficiencyTested according to appropriate standards; found safe and efficient for its purpose.
    Technological Characteristics EquivalenceFunctionally similar to the predicate device (System 2000 ROTATIONAL VESTIBULAR CHAIR). Both are rotary chairs controlled by software, used for vestibular stimulation.
    Indications for Use EquivalenceSame indications for use as the predicate device: eliciting Vestibulo-Ocular Reflexes in patients (2+ years old,
    Ask a Question

    Ask a specific question about this device

    K Number
    K082634
    Manufacturer
    Date Cleared
    2009-04-07

    (209 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VertiGONE goggle is indicated for the treatment of Benign Paroxysmal Positional Vertigo (BPPV). The VertiGONE goggle guides the practitioner to move the patient's head accurately through the Epley, or canalith repositioning maneuver.

    Device Description

    The VertiGONE BPPV Goggle is a head mounted medical device worn by the patient. The VertiGONE Goggle is designed as a non-intrusive device with nonsubstantial risk to the subjects. The subject device consists of a molded plastic goggle, an elastic strap to secure the goggle to the head, and a visual reference platform including two visual reference mechanisms. The VertiGONE Goggle provides a visual reference platform for the practitioner to guide the patient to accurately conduct the Epley, or canalith repositioning, maneuver. When using the device, the patient and/or the practitioner are each provided with visual references. Thus, the practitioner can help the patient to complete the Epley maneuver more accurately.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study conducted for the VertiGONE BPPV Goggle based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
    EffectivenessImprovement in rotary nystagmus scores."The results of the clinical trial suggest moderate improvement in BPPV outcomes in patients when the Epley maneuver was performed correctly using VertiGONE verse outcomes when a less accurate Epley maneuver was performed."
    SafetyNo new questions of safety raised compared to predicate devices."Although there are differences in the technological characteristics... those differences do not raise new questions of safety."
    Intended UseDevice accurately guides the practitioner to perform the Epley maneuver."The controlled clinical design was a prospective, single-blind crossover study... to confirm that VertiGONE goggles improve rotary nystagmus scores, and thus BPPV, by aiding the physician in accurately performing the Classical Epley, or canalith repositioning, maneuver."
    Substantial EquivalenceDevice is substantially equivalent to legally marketed predicate devices."VertiGONE, Inc. believes that, based on the results of clinical performance data, the VertiGONE BPPV Goggle is substantially equivalent to other legally marketed devices for the treatment of benign paroxysmal positional vertigo (BPPV)."

    Note: The provided document does not explicitly state quantifiable acceptance criteria (e.g., "rotary nystagmus scores must improve by X%"). The acceptance criteria are implicitly derived from the discussion of the study's purpose and its findings.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample size used for the clinical study (test set). It is referred to as "patients" in the context of the Epley maneuver.

    • Sample Size: Not explicitly stated.
    • Data Provenance: The study was conducted in a "controlled practitioner office environment." The country of origin is not specified but given the submitter and contact information are in the U.S., it is highly probable the data is from the U.S. The study was prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not provide details on the number of experts used to establish ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience").

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The study design was a "prospective, single-blind crossover study" which compared performing the Epley maneuver with the VertiGONE Goggle versus a "less accurate Epley maneuver performed." This indicates a comparison of a procedure aided by the device versus a non-aided procedure, not a comparison of multiple readers' interpretations of cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    This device is not an algorithm-only device. It is a physical goggle that aids a human practitioner. Therefore, a standalone (algorithm only) performance study was not applicable and not performed. The performance is inherently tied to human use.

    7. Type of Ground Truth Used

    The ground truth implicitly used for the study was the improvement in rotary nystagmus scores and BPPV outcomes. This is a clinical outcome observed by the practitioner. It's an objective physical manifestation of the condition.

    8. Sample Size for the Training Set

    The document does not mention any "training set." This device is a physical medical device, not an AI/ML algorithm that requires a training set in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is discussed for this physical device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081602
    Device Name
    DIZZYFIX
    Date Cleared
    2008-09-19

    (105 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LXV

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DizzyFIX is indicated for the treatment of vertigo where a physician has diagnosed posterior canal BPPV and recommended one of the exercise variations which are used to treat this type of vertigo. The DizzyFIX should be used as a component to assist in correct performance of therapeutic exercises for the treatment of vertigo.

    Patients with a diagnosis of Posterior Canal Benign Paroxysmal Positional Vertigo (BPPV) who exhibit:
    a. Momentary spinning dizziness,
    b. Which lasts seconds to minutes,
    c. Which is associated with position change

    Device Description

    The DizzyFIX is a small, non powered, non-sterile head worn device which can assist in the performance of the treatment exercises for vertigo. The DizzyFIX is a device which assists in the performance of a correct "Particle Repositioning Maneuver". Whether being used by a patient at home, a doctor or other health professional, this device helps to treat BPPV by encouraging the correct angles and duration for conducting both diagnostic and treatment exercise maneuvers. With the DizzyFIX, persons who suffer from BPPV can repeat the repositioning exercises at home as often as needed to achieve resolution of symptoms.

    The DizzyFIX consists of a specially curved acrylic tube containing a non-toxic viscous fluid and a bead. The entire device is about the size of a person's fist and weighs only a few hundred grams. Factors such as fluid viscosity and tube shape were carefully designed so that motion of the bead within the tube guides the patient accurately through the series of vestibular exercises.

    The device clips onto the mid-brim of any hat such that it is visible to the wearer. Beginning with the bead at one end of the tube, the patient tilts his or her head at appropriate angles and time intervals to slowly move the bead to the other end of the tube. Thus, there is real-time feedback to the patient. If the bead arrives at the other end of the tube, the exercise maneuvers have been performed correctly. The exercise maneuver can be repeated as often as needed.

    The DizzyFIX device consists of a transparent tube filled with mineral oil. The device contains a bead that freely travels within the mineral oil. The tube is permanently sealed at both ends. The device has a clip which facilitates its attachment to the included hat such that it is visible to the user. The hat has a strap such that it is affixed snugly to the users head. The device, tube and bead have characteristics such that only by performing a correct vestibular exercise might the bead move from one end of the tube to the other.

    AI/ML Overview

    Here's an analysis of the DizzyFIX 510(k) submission, extracting the requested information:

    The document (K081602) describes the DizzyFIX, a device intended to assist in the performance of the particle repositioning maneuver for Benign Paroxysmal Positional Vertigo (BPPV).


    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on the clinical effectiveness of the device in assisting BPPV treatment, rather than discrete, quantifiable technical acceptance criteria with pre-defined thresholds. The key "acceptance criteria" is implied by the clinical trial outcome: successful treatment of BPPV.

    Acceptance Criteria (Implied)Reported Device Performance
    Successful treatment of BPPV88% of patients successfully treated their BPPV by performing these exercise maneuvers with the assistance of the DizzyFIX device.
    SafetyNo significant side effects.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states:

    • Sample Size (Test Set): Not explicitly stated numbers for individual studies. It mentions "Three clinical trials were performed." The reported performance of "88% of patients successfully treated" implies a test set was used across these trials.
    • Data Provenance:
      • Country of Origin: Canada (specifically, University of Western Ontario, London, Ontario, Canada).
      • Retrospective or Prospective: Not explicitly stated, but the description of "clinical trials were performed" typically implies a prospective study design.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the supplied text. The document refers to "physicians for the treatment of BPPV" and mentions "specialist physicians," but it does not specify how many experts established ground truth or their exact qualifications for the clinical trials.


    4. Adjudication Method for the Test Set

    This information is not provided in the supplied text.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The DizzyFIX is a non-powered, head-worn mechanical guidance device, not an AI-powered diagnostic or assistive tool. There is no mention of human readers or AI assistance in the context of an MRMC study.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. The DizzyFIX is a physical device that assists a human user in performing a maneuver. The device itself "does not treat BPPV," but rather "assists in the exercise while the user performs the standard exercise maneuver." Therefore, its performance is inherently human-in-the-loop.


    7. The Type of Ground Truth Used

    The ground truth used to determine "successful treatment" in the clinical trials was based on the resolution of BPPV symptoms after performing the Particle Repositioning Maneuver with the DizzyFIX's assistance. This typically involves assessing the patient's subjective experience of dizziness and objective clinical tests used to diagnose and confirm BPPV (though specific tests aren't detailed here).


    8. The Sample Size for the Training Set

    This information is not applicable (or not provided in the context of "training set"). The DizzyFIX is a mechanical device, not a machine learning model, so it doesn't have a "training set" in the conventional sense of AI/ML. Its design (fluid viscosity, tube shape) was likely developed through iterative engineering and biomechanical modeling, not data-driven machine learning training.


    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no mention of a "training set" for an AI/ML model for this device. The design principles were established using:

    • Known anatomical and physical properties of the human balance organ (vestibular apparatus).
    • Previously published mathematical and biomechanical models of the vestibular apparatus.
    • Clinical aspects like latency of vertigo onset, patient anxiety, and safety.
      The device's design was engineered so that "only by performing a maneuver essentially similar to Epley's maneuver... would it register success" (i.e., the bead moving correctly).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2