K Number
K192652
Device Name
TRV
Manufacturer
Date Cleared
2020-05-27

(246 days)

Product Code
Regulation Number
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The TRV is intended to assist in the diagnosis and treatment of balance disorders and vertigo, including benign paroxysmal positional vertigo.
Device Description
The TRV is a multi-axial chair that can rotate 360° around both the horizontal and vertical axes. With the patient secured by a four-point harness, a head mount, a leg strap, and shoulder pads, an examiner is able to rotate the patient around the plane of each of the 6 semicircular canals and hold the patient in any position for detailed examination of the semicircular canals. The TRV chair is manually handled by the healthcare professional. The axes of rotation are lockable in preset positions. The primary axis has a battery-powered electromagnetic lock controlled by a footswitch and the secondary axis is manually locked with a mechanical lever. A mechanical system with an adjustable counterweight ensures that the weight of the chair and the patient are balanced during the maneuvers.
More Information

Not Found

No
The device description focuses on mechanical and manual operation, with no mention of AI/ML terms or functionalities.

Yes
The intended use statement explicitly states that the device is "intended to assist in the diagnosis and treatment of balance disorders and vertigo." The term "treatment" indicates a therapeutic function.

Yes
The 'Intended Use / Indications for Use' section explicitly states, "The TRV is intended to assist in the diagnosis and treatment of balance disorders and vertigo."

No

The device description clearly describes a physical chair with mechanical and electrical components, not a software-only device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to "assist in the diagnosis and treatment of balance disorders and vertigo." This describes a device used directly on a patient to perform physical maneuvers and examinations, not a device used to examine specimens in vitro (outside of the body).
  • Device Description: The description details a mechanical chair used to position a patient. There is no mention of reagents, test strips, or any components used to analyze biological samples.
  • Lack of IVD Characteristics: The information does not mention any of the typical characteristics of an IVD, such as:
    • Analysis of biological specimens (blood, urine, tissue, etc.)
    • Detection or measurement of analytes
    • Use of reagents or assays

The device is a physical therapy and diagnostic tool used for positional testing and treatment of balance disorders.

N/A

Intended Use / Indications for Use

The TRV is intended to assist in the diagnosis and treatment of balance disorders and vertigo, including benign paroxysmal positional vertigo.

Product codes

LXV

Device Description

The TRV is a multi-axial chair that can rotate 360° around both the horizontal and vertical axes. With the patient secured by a four-point harness, a head mount, a leg strap, and shoulder pads, an examiner is able to rotate the patient around the plane of each of the 6 semicircular canals and hold the patient in any position for detailed examination of the semicircular canals.

The rehabilitation of patients diagnosed with positional vertigo commonly involves traditional maneuvers such as the Epley, Semont, etc. These maneuvers can be performed efficiently and safely with the TRV chair. Kinetic energy can be added for specific maneuvers by driving the main arm of the TRV chair against a hydraulic stop in each sequence in the maneuver. This added kinetic energy accelerates the movement of the smaller otoconia.

The TRV chair is manually handled by the healthcare professional. The axes of rotation are lockable in preset positions. The primary axis has a battery-powered electromagnetic lock controlled by a footswitch and the secondary axis is manually locked with a mechanical lever. A mechanical system with an adjustable counterweight ensures that the weight of the chair and the patient are balanced during the maneuvers.

Healthcare professionals may use the TRV in combination with videonystagmography goggles to detect positional induced nystagmus while moving the patient through a series of positions.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Trained healthcare professional / In a medical practice

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Electrical safety and electromagnetic compatibility (EMC): The device complies with the IEC 60601-1 for electrical safety and the IEC 60601-1-2 standard for EMC.
Mechanical testing: To evaluate the worst-case stress loads and simulate device-use lifetime and demonstrate long-term safety of the TRV, the independent test institute Danish Technological Institute has tested the TRV based on the agreed test criteria from Interacoustics. After the testing of the TRV, the wear observed is considered not to have any effect on the safety on the patients or operator of the TRV.
Animal and clinical studies: No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate, nor its safety and effectiveness.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K071973

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

N/A

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 27, 2020

Interacoustics A/S Erik Nielsen Director, Regulatory & Compliance Audiometer Alle 1 Middelfart, 5500 DK

Re: K192652

Trade/Device Name: TRV Regulation Number: 21 CFR N/A Regulation Name: N/A Regulatory Class: Class II Product Code: LXV Dated: Not dated Received: April 29, 2020

Dear Erik Nielsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K192652

Device Name TRV

Indications for Use (Describe)

The TRV is intended to assist in the diagnosis and treatment of balance disorders and vertigo, including benign paroxysmal positional vertigo.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

TRV - Multi-axial chair

Erik Nielsen, Director, Regulatory & Compliance

Interacoustics A/S

Audiometer Allé 1 5500 Middelfart

+45 6371 3555

April 17, 2020

Multi-axial chair

Unclassified

06/20/2008

Apparatus, vestibular analysis

erni@demant.com

Denmark

TRV

LXV

Submitter Information:

Company Name Address

Phone e-mail Contact Person Date Summary Prepared

Device Identification:

Trade Name Common Name Classification Name Product Code Classification Panel Device Class

Predicate Device:

Predicate Device Manufacturer 510(k) No. Date Cleared

Epley Omniax™ Vesticon inc K071973

Ear Nose & Throat

Device description

The TRV is a multi-axial chair that can rotate 360° around both the horizontal and vertical axes. With the patient secured by a four-point harness, a head mount, a leg strap, and shoulder pads, an examiner is able to rotate the patient around the plane of each of the 6 semicircular canals and hold the patient in any position for detailed examination of the semicircular canals.

The rehabilitation of patients diagnosed with positional vertigo commonly involves traditional maneuvers such as the Epley, Semont, etc. These maneuvers can be performed efficiently and safely with the TRV chair. Kinetic energy can be added for specific maneuvers by driving the main arm of the TRV chair against a hydraulic stop in each sequence in the maneuver. This added kinetic energy accelerates the movement of the smaller otoconia.

4

The TRV chair is manually handled by the healthcare professional. The axes of rotation are lockable in preset positions. The primary axis has a battery-powered electromagnetic lock controlled by a footswitch and the secondary axis is manually locked with a mechanical lever. A mechanical system with an adjustable counterweight ensures that the weight of the chair and the patient are balanced during the maneuvers.

Healthcare professionals may use the TRV in combination with videonystagmography goggles to detect positional induced nystagmus while moving the patient through a series of positions

Indications for Use

The TRV is intended to assist in the diagnosis and treatment of balance disorders and vertigo, including benign paroxysmal positional vertigo.

Technological characteristics

The Technological Characteristics are the same as the predicate device. See Substantial Equivalence table below for characteristics.

Substantial Equivalence

Precise, multiaxial patient positioning for the analysis of nystagmus and the application of maneuvers used to treat vestibular disorders is the technological principal for both the subject and predicate devices.

At a high level, the subject and predicate devices are based on the same technological elements as presented in the table below.

The information regarding the videonystagmography system of the Epley Omniax™ has not been included as this function is not an integral part of the TRV.

TRVEpley Omniax™SE Comparison
Indications for UseThe TRV is intended to
assist in the diagnosis
and treatment of balance
disorders and vertigo,
including benign
paroxysmal positional
vertigo.The Epley Omniax™ is
intended to assist in
the diagnosis and
treatment of balance
disorders and vertigo,
including benign
paroxysmal positional
vertigo.Same
Use environmentIn a medical practiceIn a medical practiceSame
UserTrained healthcare
professionalTrained healthcare
professionalSame
Patient populationPatients presenting with
imbalance, unsteadiness
or dizziness symptoms
suspected of originating
from dislocated otoconia
in one or more of the
semicircular canals in the
inner ear.Patients presenting
with imbalance,
unsteadiness or
dizziness symptoms
suspected of
originating from
dislocated otoconia in
one or more of the
semicircular canals in
the inner ear.Same
Test options
Video recordingYesYesSame
Nystagmus
analysisYesYesSame
Oculomotor testsYesYesSame
Session loggingYesYesSame
Positional testsYesYesSubstantial equivalent
Discussed below
Maneuvers
CanalithCanalith repositioningPre-set protocolSubstantial equivalent
repositioning(Epley) for Posterior"CanalithBoth devices can perform
(Epley)semi-circular canal
(PSC) BPPVrepositioning"same maneuver
SemontCanalith repositioning for
PSC BPPVCustomized protocol
using the Freestyle tabSubstantial equivalent
Both devices can perform
same maneuver
LorinCanalith repositioning for
Anterior semi-circular
canal (ASC) BPPVCustomized protocol
using the Freestyle tabSubstantial equivalent
Both devices can perform
same maneuver
BBQ Roll Left HSCCanalith repositioning for
Horizontal semi-circular
canal (HSC) BPPV
canalithiasis, Left sidePre-set protocol "3/4
Barrel roll" -
canalithiasis LeftSubstantial equivalent
Both devices can perform
same maneuver
BBQ Roll Right
HSCCanalith repositioning for
Horizontal semi-circular
canal (HSC) BPPV
canalithiasis, Right sidePre-set protocol "3/4
Barrel roll" -
canalithiasis RightSubstantial equivalent
Both devices can perform
same maneuver
BBQ Roll Left HSCCanalith repositioning for
Horizontal semi-circular
canal (HSC) BPPV
cupulolithiasis, Left sidePre-set protocol "3/4
Barrel roll" -
cupulolithiasis LeftSubstantial equivalent
Both devices can perform
same maneuver
BBQ Roll Right
HSCCanalith repositioning for
Horizontal semi-circular
canal (HSC) BPPV
cupulolithiasis, Right sidePre-set protocol "3/4
Barrel roll" -
cupulolithiasis RightSubstantial equivalent
Both devices can perform
same maneuver
360° Back Flip
"incremental"360° Back Flip
"incremental"Pre-set protocol 360°
Back Flip "incremental"Substantial equivalent
Both devices can perform
same maneuver
360° Back Flip
"continuous"360° Back Flip carried
out in slow, continuous
sweepPre-set protocol 360°
Back Flip "continuous"Substantial equivalent
Both devices can perform
same maneuver
360° Forward Flip360° Forward FlipPre-set protocol 360°
Forward FlipSubstantial equivalent
Both devices can perform
same maneuver
Technical
DimensionsWeight: 640 kg
Height: 1.9 m
Depth: 1.6 m
Width: 1.2 mWeight: 372 kg
Height: 2.1 m
Depth: 2.0 m
Width: 2.3 mThe TRV is slightly
smaller in size and
heavier than the predicate
device. This difference
does not raise new
questions of safety and
effectiveness.
Movement range2 axes, 360° each:
  • Primary axis is a roll
    movement with the
    patient in the load
    position.
  • Secondary axis is a
    yaw movement with the
    patient in the load
    position. | 2 axes, 360° each:
  • Alpha axis is a roll
    movement with the
    patient in the load
    position.
  • Beta axis is a yaw
    movement with the
    patient in the load
    position. | The range of motion is
    the same for both
    devices. |
    | Power supply | The TRV chair is
    equipped with a 24V
    rechargeable battery that
    powers the magnetic
    lock for the primary
    frame (Linak BAJ1 (24 V
    DC, 2,9 Ah)).
    A wall-mounted battery
    charger is also provided
    (Linak, CHJ2). | Mains power supply:
    Voltage input
    120V/240V
    (switchable) at 50-
    60Hz
    Current rating: 6A
    @120V, 3A @ 240V
    Fusing: 8A @120V, 4A
    @ 240V
    Isolation from main
    power designed to
    meet UL/cUL 2601 | Both devices have
    passed electrical security
    testing according to IEC
    60601-1 and EMC testing
    according to IEC 60601-
    1-2.
    The differences in the
    energy source do not
    raise different questions
    of safety and
    effectiveness and are
    supported by testing |

5

6

Positional tests discussion

The multi-axial positioning has been evaluated comparing the positions available from the TRV and the Omniax. Interacoustics has found that the positionings enables the same maneuvers and both devices offer multi-axial patient positioning with rotation around 2 axes.

Discussion of differences

Interacoustics did not find any essential or major differences between the devices.

7

Performance data

The following performance data were provided in support of the substantial equivalence determination.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the TRV. The device complies with the IEC 60601-1 for electrical safety and the IEC 60601-1-2 standard for EMC.

Mechanical testing

To evaluate the worst-case stress loads and simulate device-use lifetime and demonstrate long-term safety of the TRV, the independent test institute Danish Technological Institute has tested the TRV based on the agreed test criteria from Interacoustics. After the testing of the TRV, the wear observed is considered not to have any effect on the safety on the patients or operator of the TRV.

Animal and clinical studies

No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate, nor its safety and effectiveness.

Conclusion

Based on its intended use, design principles, and technological characteristics, the TRV was found to be as safe, as effective, and performs comparably to the predicate device.

The technological differences identified do not raise new questions of safety and effectiveness as the non- clinical data support the safety of the device and demonstrate that the TRV performs as intended in the specified use conditions.