Search Results
Found 148 results
510(k) Data Aggregation
(184 days)
KGN
Helios Dermal Scaffold is intended for the management of wounds that include:
- Partial- and full- thickness wounds
- Pressure, diabetic, and venous ulcers
- Partial thickness burns
- Surgical wounds – donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence
- Trauma wounds – abrasions, lacerations and skin tears
- Tunneled/undermined wounds
- Draining wounds
Helios Dermal Scaffold is an acellular dermal tissue matrix derived from fetal bovine dermis. The single-use device is supplied sterile and is provided in sheet form in a variety of sizes and configurations, including meshed, solid, and fenestrated designs ranging from 1.5 - 500 cm² to be trimmed by the surgeon to meet the individual patient's needs.
The provided FDA 510(k) clearance letter for the Helios Dermal Scaffold does not contain any information regarding acceptance criteria or the study details typically associated with AI/ML device performance validation.
This document is for a medical device (a dermal scaffold), which is a biological product, not a software or AI/ML-driven diagnostic device. The performance data section refers to physical and biological characteristics of the scaffold, such as:
- Device Characterization Testing: Dimensional verification, color and appearance, suture retention strength, acellularity (histology), collagen denaturation.
- Design Validations: Viral inactivation, sterilization.
- Biocompatibility
- Shelf-life
- Packaging testing
These tests are standard for biological and implantable materials, ensuring their physical integrity, safety, and compatibility with the human body. They do not involve AI/ML algorithms, image analysis, or diagnostic accuracy.
Therefore, it is not possible to extract the requested information (acceptance criteria, sample size, ground truth, expert opinions, MRMC studies, etc.) from this document, as those concepts are not applicable to the type of device described.
To answer your prompt with the requested information, the input document would need to be a 510(k) clearance letter for an AI/ML medical device, typically referencing a clinical study or performance study that evaluates diagnostic accuracy, sensitivity, specificity, or reader agreement.
Ask a specific question about this device
(30 days)
KGN
Kerecis® Marigen Wound Extra Autologous Hydration, Kerecis Silicone Autologous Hydration and Kerecis Parvus Autologous Hydration
Management of wounds including:
- Partial thickness wounds
- Full thickness wounds
- Pressure ulcers
- Venous ulcers
- Chronic vascular ulcers
- Diabetic ulcers
- Trauma wounds (abrasions, lacerations, partial thickness burns, skin tears)
- Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence)
- Draining wounds
All three subject devices of this bundled submission are part of a family of devices manufactured by Kerecis Limited. The subject devices can be seen in Table 1. They are lyophilized, terminally sterilized, fish skin medical devices comprised of biocompatible, resorbable fish skin (Wild North Atlantic Cod) for wound management. The devices are intended for single use only. The devices are applied to the wound bed to maintain a moist wound environment. The primary predicate device is Marigen Wound Extra (K190528) and the additional predicate devices are Kerecis Silicone (K213231), and Kerecis Parvus (K241080). Marigen Wound Extra is commercially available under the names Kerecis MariGen, Kerecis GraftGuide, and Kerecis SurgiClose. Kerecis Silicone is commercially available under the names Kerecis Shield and Kerecis SurgiClose Silicone. For clarity, this submission will refer to the devices under their commercially available names, except when specifically referring to the primary predicate device. This information is also shown in Table 1.
Although the subject devices differ from each other in terms of device indications and dimensional specifications, each one remains physically identical to its primary predicate device, both in design and packaging, as well as for indications for use. The only difference between each subject device and its respective primary predicate device is in the device labeling, with the subject devices having additional rehydration fluid options included in their instructions for use (IFUs).
The provided FDA 510(k) clearance letter and summary describe the acceptance criteria and a study to prove the device meets these criteria. However, it's important to note that this submission is for a modification to an existing device (Kerecis Marigen Wound Extra, Kerecis Silicone, Kerecis Parvus) and not for an entirely novel device. The modification specifically addresses the inclusion of additional rehydration fluids. Therefore, the "study" is focused on verifying the device's performance with these new rehydration fluids, rather than establishing initial clinical effectiveness.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the device's performance with the original rehydration fluid (saline) and the need for the device to perform comparably with the new rehydration fluids. The performance is assessed through specific bench tests.
Acceptance Criteria Category | Specific Criteria/Test | Reported Device Performance |
---|---|---|
Rehydration | Device rehydration with lactated Ringer's solution | Performance remains consistent and comparable to primary predicate devices. |
Suture Retention | Device suture retention with autologous body fluid | Performance remains consistent and comparable to primary predicate devices. |
General Device Performance | Maintenance of intended use, materials, manufacturing processes, and fundamental scientific technology despite new rehydration fluid options. | Device performance remains consistent and comparable to primary predicate devices, supporting substantial equivalence. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size for the rehydration and suture retention tests. It mentions "bench testing" was performed.
Data provenance: Not explicitly stated, but bench testing typically involves laboratory-controlled conditions. It is not patient data from a specific country, nor is it referred to as retrospective or prospective in a clinical trial sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
This type of information (expert consensus for ground truth) is typically associated with clinical studies or diagnostic device clearances where a "truth" is established through expert review of patient data (e.g., radiologist opinions on images). Since this submission focuses on bench testing for a modification to rehydration fluids, this information is not relevant or provided. The "ground truth" here is the prior established performance of the predicate device under saline rehydration.
4. Adjudication Method for the Test Set
Adjudication methods (e.g., 2+1, 3+1) are relevant for studies involving human interpretation or clinical endpoints. As this submission describes bench testing for material and process compatibility, an adjudication method is not applicable and therefore not provided.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is typically for diagnostic devices where multiple readers evaluate cases to assess performance with and without AI assistance. This submission is for wound dressings and focuses on physical and biological compatibility with different rehydration fluids.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a wound dressing, not a software algorithm or an AI-based system. Therefore, there is no "algorithm only" performance to evaluate.
7. The Type of Ground Truth Used
For the specific tests conducted for this modification (rehydration and suture retention with new fluids), the ground truth is implicitly the established performance characteristics of the predicate device when rehydrated with saline. The goal of the new tests was to show that these characteristics are maintained or are comparable when using lactated Ringer's solution and autologous body fluids.
8. The Sample Size for the Training Set
This concept is not applicable as this is not a machine learning or AI device. The "training set" for a traditional medical device would refer to the data used to design and develop the device prior to its initial submission. The summary doesn't provide this detail for the original device development, only that performance testing was "leveraged from the Kerecis primary predicate devices."
9. How the Ground Truth for the Training Set was Established
Again, "training set" and its "ground truth" are terms typically used in AI/ML contexts. For a medical device like a wound dressing, the "ground truth" during initial development (analogous to a training phase) would involve extensive material testing, biocompatibility studies, mechanical property evaluations, and potentially pre-clinical and clinical studies to establish its safety and effectiveness for wound management. The summary indicates that for this modification, the ground truth is based on the previously established performance of the predicate devices.
Ask a specific question about this device
(140 days)
KGN
Collymer Self-Assembling Scaffold is intended for the management of wounds including: partial and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/ undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, skin tears) and draining wounds.
Collymer Self-Assembling Scaffold is a wound management device composed of a purified, self-assembling (scaffold-forming) collagen derived from porcine dermis. This two-part device consists of a collagen solution and a self-assembly reagent. Combining the two solutions initiates collagen self-assembly. The resulting collagen scaffold is suitable for cellularization and vascularization, providing a moist wound environment.
The provided FDA 510(k) Clearance Letter for the GeniPhys Collymer Self-Assembling Scaffold (K250329) does not include information about any clinical studies or the performance of a device based on acceptance criteria.
The clearance letter explicitly states:
"Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of Collymer Self-Assembling Scaffold to the predicate device."
Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as no such clinical study data is presented in this document.
The document focuses on demonstrating substantial equivalence through:
- Comparison of Indications for Use and Technological Characteristics to a predicate device (Integra Flowable Wound Matrix, K072113) and several reference devices.
- Non-Clinical Testing: Biocompatibility assessments, viral inactivation studies, sterility assessments, shipping studies, and stability evaluations. For these non-clinical tests, it is stated that "All results met the acceptance criteria" or "The acceptance criteria were met," but the specific numerical acceptance criteria themselves are not detailed, nor are the specific performance results beyond a qualitative statement of meeting the criteria.
To directly answer your specific points based on the provided document:
- A table of acceptance criteria and the reported device performance: This information is not explicitly provided in the document for any clinical performance. For non-clinical tests, only a qualitative statement that acceptance criteria were met is given.
- Sample sizes used for the test set and the data provenance: Not applicable as no clinical test set is described. For non-clinical tests, sample sizes are not specified.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no clinical test set is described.
- Adjudication method for the test set: Not applicable as no clinical test set is described.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, the document states clinical testing was not necessary.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical medical device (wound scaffold), not a software algorithm.
- The type of ground truth used: Not applicable as no clinical study or test set with a "ground truth" (in the context of AI/diagnostic device performance) is described.
- The sample size for the training set: Not applicable, as the document refers to a physical device and not a machine learning model that would have a training set.
- How the ground truth for the training set was established: Not applicable, as no training set is described.
In summary, this 510(k) clearance relied on substantial equivalence to a predicate device and extensive non-clinical testing, rather than a clinical study demonstrating performance against specific acceptance criteria.
Ask a specific question about this device
(267 days)
KGN
CollOvine™ Wound Dressing is indicated for the management of exuding wounds including:
- full-thickness & partial-thickness wounds
- pressure ulcers
- venous ulcers
- ulcers caused by mixed vascular etiologies
- diabetic ulcers
- superficial second-degree burns
- donor sites and other bleeding surface wounds
- abrasions
- traumatic wounds healing by secondary intention
- dehisced surgical incisions
CollOvine Wound Dressing is a sterile advanced wound care device composed of medical grade ovine collagen. CollOvine Wound Dressing is soft, absorbent, and readily conforms to the wound bed. CollOvine Wound Dressing is available in multiple sizes; 1"x1", 2"x2" and 4"x4". CollOvine wound dressing is intended to maintain the moist microenvironment. CollOvine Wound Dressing is a primary wound dressing that can be cut with sterile scissors to ensure good contact with the wound bed. CollOvine Wound Dressing should be used with suitable non-occlusive absorbent secondary dressing and secured with standard non-irritating fixations such as medical grade tape or semi-occlusive dressing as appropriate. CollOvine Wound Dressing is a primary dressing, to maintain the required moist wound bed environment. CollOvine Wound Dressing is soft and pliable. The ovine collagen for CollOvine Wound dressing is obtained from sheep in compliance with ISO22442, including controls of animal tissue sourcing, collection, storage and viral inactivation.
The provided FDA 510(k) clearance letter for the CollOvine™ Wound Dressing does not contain any information about acceptance criteria or a study that proves the device meets such criteria.
The document primarily focuses on establishing substantial equivalence to a predicate device (Fibracol Plus Collagen wound dressing with Alginate) based on non-clinical testing in a side-by-side comparison.
Therefore, I cannot fulfill your request for the specific details you've asked for, as they are not present in the provided text.
Here is a breakdown of what is contained in the document related to testing, and why it doesn't align with your request for acceptance criteria and a study proving device performance:
1. A table of acceptance criteria and the reported device performance
- Not present. The document lists the Indications for Use for both the CollOvine Wound Dressing and the predicate device, showing they are identical. It also presents a "Technological Comparison" table that details attributes like classification, product code, regulatory class, indications for use, manufacturer, material, origin, device characteristics, biocompatibility, reusability, and sterilization method. This table focuses on comparing attributes rather than defining performance acceptance criteria or reporting specific performance metrics against those criteria. The “Device Characteristics” for both are simply listed as “Collagen” and “Collagen with 10% calcium alginate” respectively, which is a material description, not a performance metric.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not present. The document describes non-clinical biocompatibility testing (Acute systemic toxicity, Primary skin irritation, Dermal sensitization, Pyrogenicity, Subacute Systemic toxicity, Subcutaneous implant, In vitro Cytotoxicity, Chemical Characterization, Toxicological Risk assessment). For these tests, sample sizes and data provenance are not provided. These are generally standard biological safety tests, not clinical performance studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not present. This information is typically relevant for studies involving human assessment or interpretation (e.g., radiology AI devices). This is not applicable to the non-clinical biocompatibility and physiochemical testing described.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not present. Adjudication methods are typically used in clinical studies where expert consensus is needed. This is not mentioned for the non-clinical tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not present. An MRMC study is a type of clinical study, often used for diagnostic devices involving human interpretation (like AI in radiology). The document describes a wound dressing, and the testing conducted is non-clinical.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not present. This concept is specific to AI or software-as-a-medical-device (SaMD). The CollOvine Wound Dressing is a physical product, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Not present. For the non-clinical biocompatibility tests, "ground truth" would be established by the results of standardized biological assays comparing the device's effects to known safety profiles or controls. For pH and absorbency tests, the "ground truth" is the measured physical property itself, compared to the predicate. No expert consensus or pathology is mentioned.
8. The sample size for the training set
- Not present. Training sets are relevant for AI/machine learning models. This is a physical wound dressing device.
9. How the ground truth for the training set was established
- Not present. As above, not relevant for this type of device.
Summary of what the document does provide regarding testing:
- Study Objective (Implied): To demonstrate substantial equivalence to the predicate device (Fibracol Plus Collagen wound dressing with Alginate, K982597) through non-clinical testing.
- Tests Performed:
- Biocompatibility testing in compliance with ISO 10993 Part 1 Biological Evaluation of Medical Devices, including:
- Acute systemic toxicity (in mice)
- Primary skin irritation (in rabbits)
- Dermal sensitization (in Guinea pigs)
- Pyrogenicity (in rabbits)
- Subacute Systemic toxicity (in rats; 30 days intraperitoneal route)
- Subcutaneous implant (in rats, 1-week and 4-weeks)
- In vitro Cytotoxicity (Direct cell contact)
- Chemical Characterization
- Toxicological Risk assessment
- pH and absorbency testing (results stated as "found to be similar to that of the primary predicate Fibracol Plus wound dressing").
- Biocompatibility testing in compliance with ISO 10993 Part 1 Biological Evaluation of Medical Devices, including:
- Conclusion: Based on the non-clinical testing, CollOvine wound dressing is substantially equivalent to the predicate device.
In essence, the document serves as a regulatory clearance (510(k)) based on demonstrating that the new device is as safe and effective as a legally marketed predicate device, primarily through non-clinical, benchtop, and biocompatibility studies, rather than a clinical performance study with predefined acceptance criteria for patient outcomes.
Ask a specific question about this device
(189 days)
KGN
DermiSphere™ hDRT hydrogel Dermal Regeneration Template is a wound dressing indicated for the management of wounds including: partial and full-thickness wounds, pressure ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds (donor sites/grafts, post Moh's surgery, post laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, second degree burns, skin tears) and draining wounds. This device is intended for one-time use.
DermiSphere™ hDRT hydrogel Dermal Regeneration Template is an advanced wound dressing comprised of crosslinked bovine collagen microspheres embedded in bovine collagen hydrogel matrix. This composite biodegradable wound dressing provides a scaffold for native tissue cellular invasion and capillary growth.
The provided text does not contain information about acceptance criteria and the study that proves the device meets them. The text primarily focuses on the FDA's clearance of the DermiSphere™ hDRT (K241904) as substantially equivalent to a predicate device (Avagen Matrix Wound Dressing, K022127).
It mentions:
- Biocompatibility testing
- Bench testing including dimensions, drapability, and pyrogenicity.
However, it does not provide:
- A table of acceptance criteria and reported device performance.
- Sample sizes for test sets or data provenance.
- Number of experts or their qualifications for establishing ground truth.
- Adjudication methods.
- Information about multi-reader multi-case (MRMC) comparative effectiveness studies or effect sizes.
- Information about standalone (algorithm only) performance studies.
- The type of ground truth used (e.g., expert consensus, pathology, outcomes data).
- Sample size for the training set.
- How ground truth for the training set was established.
The conclusion states that nonclinical tests demonstrated substantial equivalence, with safety and effectiveness demonstrated through bench testing, biocompatibility, and comparison to the predicate device. However, the specifics of these tests and their results against predefined acceptance criteria, as requested, are not detailed in this document.
Ask a specific question about this device
(83 days)
KGN
Bovine Dermis Collagen Dermal Matrix is indicated for the management of wounds, including:
- . Full thickness and Partial thickness wounds
- . Chronic wounds (e.g. pressure ulcers, venous ulcers, diabetic ulcers, chronic ulcers)
- . Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- . Trauma wounds (abrasions, lacerations, and skin tears)
- Draining wounds
- Partial thickness burns
Bovine Dermis Collagen Dermal Matrix is an absorbent, porous, collagen matrix engineered from purified collagen derived from bovine dermal tissue. Bovine Dermal Matrix should be applied directly to the wound, covering the entire wound surface.
Bovine Dermis Collagen Dermal Matrix is supplied sterile, non-pyrogenic and for single use only.
The Collagen Matrix, Inc. Bovine Dermis Collagen Dermal Matrix (K243071) is a medical device. The provided text outlines the device's characteristics and compares it to predicate devices to establish substantial equivalence for regulatory purposes.
Here's an analysis of the acceptance criteria and study information, based solely on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" in a go/no-go fashion with numerical targets for clinical performance. Instead, it compares the subject device to predicate devices across various technical characteristics. The implicit acceptance criterion is that the subject device's performance is either equivalent to or better than the predicate devices, or that any differences do not raise new questions of safety or effectiveness.
Characteristic | Acceptance Criterion (Implicitly "Similar to Predicate" or acceptable range) | Reported Device Performance (K243071) | Predicate K040211 Performance | Predicate K152600 Performance |
---|---|---|---|---|
Material/Source | Bovine dermis | Bovine dermis | Same | Same |
Available Sizes | Acceptable range, risk analysis for larger sizes performed. | Up to 700 cm² | Up to 720 cm² | Up to 8 cm² |
Thickness | 3 mm | 3 mm | Same | Same |
Absorption Capacity | >10 mL/g (predicate K040211), or comparable/better | 31.0 mL/g | 40.2 mL/g | Same (Implied, linked to K040211) |
Cross-linked | Yes | Yes | No | Yes |
Crosslinker | Formaldehyde (monitored for residuals) | Formaldehyde | N/A | Same |
Sterilization Method | Gamma irradiation | Gamma irradiation | Same | Same |
SAL | 10-6 | 10-6 | Same | Same |
Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Same | Same |
Packaging | Single barrier (Tyvek pouch) | Single barrier (Tyvek pouch) | Same | Same or blister tray |
Shelf Life | 36 months | 36 months | Same | Same |
Residual Formaldehyde | Monitored and mitigated; risk analysis performed for largest size | Monitored and mitigated | N/A | Monitored and mitigated |
In Vitro Characterization | Demonstrated substantial equivalence to predicates | Performed | N/A | N/A |
Biocompatibility | Demonstrated safety for long term (>30 days) contact | Performed | N/A | N/A |
Viral Inactivation | Ensured viral safety | Performed | N/A | N/A |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: The document does not specify a "test set" in the context of clinical or performance data with a specific number of cases or patients. The studies described are in vitro characterization tests, biocompatibility tests, and a viral inactivation study. These are conducted on samples of the device itself or in laboratory settings, not on patient data.
- Data Provenance: Not applicable in the context of clinical patient data. The studies are laboratory-based and conducted on the manufactured device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. The studies are not clinical studies requiring expert ground truth for interpretation of patient data. They are lab tests for material properties, biological safety, and viral inactivation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. This pertains to clinical studies involving human readers and adjudicated outcomes, which are not described here.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No. The document describes laboratory-based testing of a dermal matrix, not a diagnostic AI device requiring human reader analysis.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This pertains to AI algorithm performance studies, which are not described here.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable in the human clinical sense. For the in vitro tests (Absorbency, pH, Thermal Transition Temperature, Density, Crosslinking Agent Residual Content, Pyrogenicity), the "ground truth" would be established by validated analytical methods and reference standards. For biocompatibility, it would be based on established international standards (e.g., ISO 10993 series) and their respective endpoints. For viral inactivation, it would be based on virological assays.
8. The sample size for the training set:
Not applicable. This device is a physical medical device, not a machine learning model, so there is no "training set."
9. How the ground truth for the training set was established:
Not applicable, as there is no training set.
Ask a specific question about this device
(147 days)
KGN
BIOBRANE Dressings are indicated for:
- Covering clean partial thickness burn wounds
- Split thickness donor sites
BIOBRANE Glove is indicated for:
- Covering clean partial thickness burn wounds of the hand.
BIOBRANE Dressing and BIOBRANE Glove are wound dressings made from an ultrathin, semipermeable, perforated silicone membrane that is mechanically bonded to a flexible knitted tri-filament nylon fabric. Denatured porcine dermal collagen is bonded to the silicone-nylon membrane to provide a flexible and conformable dressing with adherence properties and a hydrophilic surface.
This document is a 510(k) K242146 premarket notification for the BIOBRANE Dressing and BIOBRANE Glove. It asserts substantial equivalence to the predicate device BIOBRANE II (K901369).
Here's an analysis of the provided information relating to acceptance criteria and device performance:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not provide a table of acceptance criteria with specific performance metrics such as sensitivity, specificity, accuracy, or any quantitative measures for the device's diagnostic or predictive capabilities. This is likely because the device is a wound dressing, not a diagnostic AI/ML device.
Instead, the submission focuses on demonstrating substantial equivalence to a predicate device through shared characteristics and verification tests. The "acceptance criteria" can be inferred as successful completion of these tests and demonstration that the device's characteristics are similar to the predicate.
Characteristic / Test | Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|---|
Intended Use | Same as predicate | Same |
Indications for Use | Within scope of predicate's indications | More precise, but substantially equivalent |
Principles of Operation | Same as predicate | Same |
Materials and Structure | Same as predicate | Same (Silicone membrane, nylon, collagen) |
Silicone Membrane Function | Same as predicate | Same |
Nylon Function | Same as predicate | Same |
Collagen Function | Same as predicate | Same |
Principal Operator | Same as predicate | Same (Trained HCP only) |
Environment of Use | Same as predicate | Same (Healthcare facility and home use) |
Single-Use or Reusable | Same as predicate | Same (Single-Use) |
Sterilization | Same as predicate | Same (Moist heat) |
Sterility | Same as predicate | Same (Sterile, SAL 10-6) |
Shelf Life | Same as predicate | Same (3 years) |
Biocompatibility | Complies with ISO 10993 | Complies with ISO 10993 |
Device Dimensions | Equivalent to predicate sizes | Equivalent to predicate sizes |
Anatomical sites | Same as predicate | Same |
Bench Performance Verification Tests | Meets design specifications | Completed and confirm specifications met |
Product Characterization | Confirms product characteristics | Completed and confirm characteristics |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "No clinical data is relied upon in this submission to determine substantial equivalence."
Therefore:
- Sample size for the test set: Not applicable, as no clinical test set was used for substantial equivalence determination.
- Data provenance: Not applicable. The evaluation was based on non-clinical (bench) testing and comparison of technological characteristics.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable, as no clinical test set with a ground truth established by experts was used. The substantial equivalence was based on non-clinical data and direct comparison to a predicate device.
4. Adjudication Method for the Test Set
Not applicable, as no clinical test set was used requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or relied upon. This is a wound dressing, not an AI/ML diagnostic device where such studies are typical.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This device is a wound dressing, not an algorithm or software-only device.
7. The Type of Ground Truth Used
Not applicable, as no clinical data or ground truth in the context of diagnostic performance was used. The "ground truth" for this submission is implicitly the established safety and efficacy profile of the legally marketed predicate device (BIOBRANE II, K901369) and the physical/chemical properties assessed through bench testing.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. This is not an AI/ML device that requires a training set or ground truth for training.
Ask a specific question about this device
(382 days)
KGN
ABCcolla® Collagen ADM Scaffold is intended to be used for management of wounds. including venous ulcers, pressure ulcers, chronic vascular ulcers, diabetic ulcers, tunneled undermined wounds (donor site/ grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, and skin tears), first and second-degree burns, draining wounds.
ABCcolla® Collagen ADM Scaffold is a decellularized porcine collagen biomaterial from porcine dermis. When applied on a wound, this product helps absorb wound exudates and maintain a moist wound environment.
This document is a 510(k) premarket notification for the ABCcolla® Collagen ADM Scaffold, a medical device for wound management. The core of the submission is to demonstrate the substantial equivalence of the new device to a legally marketed predicate device, ABCcolla® Collagen Matrix (K162348), and a reference device, Cook® ECM Powder (K152033).
Based on the provided text, the "acceptance criteria" and the "study that proves the device meets the acceptance criteria" are not related to an AI/ML-driven device's performance in terms of diagnostic accuracy or a clinical study in humans with a traditional statistical endpoint and acceptance criteria. Instead, the "acceptance criteria" for this specific device (a collagen scaffold for wound management) are primarily focused on benchmarking against a predicate device to demonstrate "substantial equivalence" as required by the FDA 510(k) pathway. The "study" here refers to the pre-clinical testing and characterization that demonstrates the new device has similar technological characteristics and performance to the predicate device, and any differences do not raise new questions of safety or effectiveness.
Here's an interpretation of the requested information based on the provided document:
1. A table of acceptance criteria and the reported device performance
For a 510(k) submission, "acceptance criteria" are not typically framed as specific performance metrics and thresholds like sensitivity/specificity for an AI device. Instead, the acceptance is based on demonstrating that the new device is "substantially equivalent" to a legally marketed predicate device. This is achieved by comparing various characteristics.
Here's a table based on the "Substantial Equivalence Comparison Table" in the document, interpreting "acceptance criteria" as demonstrating "sameness" or "differences that do not raise new questions of safety/effectiveness" compared to the predicate/reference devices:
Characteristic | Predicate Device (ABCcolla® Collagen Matrix) Performance | Reference Device (Cook® ECM Powder) Performance | New Device (ABCcolla® Collagen ADM Scaffold) Performance | Acceptance Criteria (Implicit for 510(k)) | Interpretation/Result (From Document) |
---|---|---|---|---|---|
1. 510(k) Number | K162348 | K152033 | N/A (New Submission) | Not Applicable (for comparison) | -- |
2. Product Code | KGN | KGN | KGN | Same | Same |
3. Classification | Unclassified | Unclassified | Unclassified | Same | Same |
4. Intended Use | For management of wounds including: partial and full thickness wounds, venous ulcers, pressure ulcers, chronic vascular ulcers, diabetic ulcers, tunneled/undermined wounds, surgical wounds (donor site/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, and skin tears), first and second-degree burns, draining wounds. | Similar wording for wound management categories | For management of wounds including: venous ulcers, pressure ulcers, chronic vascular ulcers, diabetic ulcers, tunneled/undermined wounds, surgical wounds (donor site/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, and skin tears), first and second-degree burns, draining wounds. | Substantially similar | ABCcolla® Collagen ADM Scaffold was same as predicate device and reference device. |
5. User | Professional surgical surgeon | Professional surgical surgeon | Professional surgical surgeon | Same | Same |
6. Material | Porcine small intestinal submucosa derived collagen material | Porcine small intestinal submucosa | Porcine dermis derived collagen material | Same source (porcine tissue) | The source of material was the same, from porcine tissue. |
7. Material Characterization | Type I collagen | Type I collagen | Type I collagen | Same | Same |
8. Structure | Sheet form | Powder | Powder | Differences do not raise new questions of safety/effectiveness | ABCcolla® Collagen ADM Scaffold was same as reference device. But, different compared to the predicate, the difference does not raise different questions of safety and effectiveness. |
9. Dimensions | 12 cm to 510 cm, 11 inch to 22 inch | Particles |
Ask a specific question about this device
(29 days)
KGN
Corplex P/Theracor P/Allacor P is indicated for use in the management of the following wounds:
- · Partial and full-thickness wounds
- Pressure ulcers
- Venous ulcers
- · Diabetic ulcers
- · Chronic vascular ulcers
- · Tunneled/undermined wounds
- · Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- · Trauma wounds (abrasions, lacerations, partial-thickness burns, and skin tears)
- · Draining wounds
Corplex P/Theracor P/Allacor P is derived from human umbilical cord extracellular matrix (ECM) and is indicated for the management of a range of acute and chronic wounds. As a resorbable particulate device, Corplex P/Theracor P/Allacor P is lyophilized and packaged in a sterile vial, allowing the device to be rehydrated and applied directly to the wound.
The provided text describes a 510(k) premarket notification for a medical device called Corplex P/Theracor P/Allacor P. It states that the device is identical to its predicate device (K231325) in most aspects, with a minor modification being the addition of an 8 cc configuration. The document explicitly states that this modification does not raise new questions of safety or effectiveness and therefore, no new clinical testing, biocompatibility testing, or extensive performance studies were required to demonstrate substantial equivalence.
Instead, the submission relies on the established performance of the predicate device and verification testing related to the minor design change.
Therefore, many of the requested details about acceptance criteria, detailed study design, sample sizes, expert involvement, and ground truth establishment, which are typically associated with performance studies, are not explicitly present in this summary document because such extensive studies were deemed unnecessary for this 510(k) submission.
Here's an attempt to answer your questions based on the provided text, indicating where information is not available due to the nature of this submission:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative format for all characteristics. Instead, it relies on the device being "identical" or having "same" or "does not raise new questions of safety and effectiveness" compared to the predicate device. For the one specific test mentioned (Bioburden Testing), the outcome is a simple "PASS".
Characteristic | Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance (K242828) |
---|---|---|
Size/Volume | Must not raise new questions of safety and effectiveness compared to 1 cc, 2 cc, and 4 cc configurations. | 1 cc, 2 cc, 4 cc, 8 cc (The additional 8cc configuration was deemed to "not raise new questions of safety and effectiveness") |
Nominal Particle Sizes | Same as predicate: Particles ranging from 0.1 mm to 2.0 mm (must pass through 2.00 mm aperture sieve and not pass through 0.106 mm aperture sieve, ASTM E11). | Particles ranging from 0.1 mm to 2.0 mm (Same as predicate: Particles must be able to pass through a 2.00 mm aperture calibrated sieve and not pass through a 0.106 mm aperture calibrated sieve (ASTM E11) when manually sieved.) |
Intended Use | Same as predicate: To cover, protect, and provide a moist wound environment. | Same |
Indications for Use | Same as predicate for specified wound types. | Same |
Configuration | Same as predicate: Particles. | Same |
Material Source | Same as predicate: Human umbilical cord (recovered per 21 CFR 1271). | Same |
Components | Same as predicate: Collagen and associated ECM components (collagen I, collagen III, collagen V). | Same |
Collagen (% total weight) | Same as predicate: 46% (Mean Value). | 46% (Mean Value) |
Total Glycosaminoglycans (mg/g) | Same as predicate: 20.3 mg/g (Mean Value). | 20.3 mg/g (Mean Value) |
Endotoxin (EU/device) | Same as predicate: |
Ask a specific question about this device
(124 days)
KGN
Management of wounds including:
- Partial and full-thickness wounds
- Pressure ulcers
- Venous ulcers
- Chronic vascular ulcers
- Diabetic ulcers
- Trauma wounds (abrasions, lacerations, partial thickness burns, skin tears)
- Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence)
- Draining wounds
Kerecis "Parvus™ is a lyophilized, terminally sterilized, acellular, particulate fish skin medical device comprised of biocompatible, non-crosslinked, resorbable, acellular fish skin (North Atlantic Cod) for wound management. The device is intended for single use only.
The subject device is packaged in the following weights:
100mg (4 cm2) 200mg (8 cm2) 500mg (19 cm²) 1000mg (38 cm²) 2,500mg (95 cm2 ) 3,000mg (114 cm2)
This document is a 510(k) Premarket Notification summary for Kerecis® Parvus™, a wound dressing. It is not an AI/ML device, and therefore does not contain acceptance criteria for device performance related to AI/ML or a study proving those criteria are met. The document states that the Kerecis® Parvus™ is substantially equivalent to a predicate device, Kerecis MariGen Wound Extra (K190528), in terms of indications for use, intended use, raw material origin and composition, device performance, packaging material, and sterilization methods. The primary difference is an additional cutting and sieving step in the manufacturing process to convert the fish skin from intact sheets to fragmented pieces ≤2.0mm.
Here's a breakdown of the requested information based on the provided document, noting that many items are not applicable (N/A) due to the nature of the device:
-
A table of acceptance criteria and the reported device performance
This document does not provide specific performance acceptance criteria or reported performance data in the traditional sense of a clinical or analytical study with defined metrics (e.g., accuracy, sensitivity, specificity). Instead, it asserts substantial equivalence to a predicate device based on material properties and manufacturing processes. The "Performance Characteristics" section for the subject device largely mirrors the predicate device, with differences primarily in the physical form (fragmented vs. sheets) and nominal sizes.
Characteristic Acceptance Criteria (Implied by Substantial Equivalence Goal) Reported Device Performance (as described) Source Origin Wild Caught Atlantic Cod Fish Wild Caught Atlantic Cod Fish (Same as predicate) Tissue source Fish Skin Fish Skin (Same as predicate) Nominal Sizes Fragmented, size controlled to ≤2.00mm Irregular shaped 3D fragmented fish skin, size controlled to ≤2.00mm and packaged in various weights (100mg to 3,000mg). (Differs from predicate's sheet sizes, but subject device is cut from predicate device sizes). Presentation Lyophilized, sterilized, fragmented fish skin Lyophilized, sterilized, Fragmented, fish skin in a Tyvek peel pouch. (Differs from predicate's sheet presentation due to cutting and sieving). Packaging Tyvek Single and Double Peel Pouch Tyvek Single and Double Peel Pouch (Same as predicate) Sterilization Ethylene Oxide Ethylene Oxide (Same as predicate) Sterility Assurance Level (SAL) SAL 10^-6^ SAL 10^-6^ (Same as predicate) Endotoxin limit ≤20 EU/device ≤20 EU/device (Same as predicate) Shelf Life At least 1 year (matching current reported) 1 year (Real-time shelf life testing in progress for the subject device; predicate has 3 years). Biocompatibility Demonstrated Additional testing for biocompatibility completed (for the subject device). Fragmented size characterization Demonstrated Additional testing for fragmented size characterization completed (for the subject device). Ethylene oxide residuals Demonstrated Additional testing for ethylene oxide residuals completed (for the subject device). Metal contamination Demonstrated Additional testing for metal contamination completed (for the subject device). Protein analysis Demonstrated Additional testing for protein analysis completed (for the subject device). Bioburden Demonstrated Additional testing for bioburden completed (for the subject device). Residual limits Demonstrated Additional testing for residual limits completed (for the subject device). -
Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
N/A. This is a 510(k) submission for a non-AI/ML medical device establishing substantial equivalence through material and manufacturing process comparisons, not a clinical study involving a test set for performance evaluation. The "additional testing" mentioned is likely laboratory bench testing on the device materials.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
N/A. Not an AI/ML device requiring expert-established ground truth.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
N/A. Not an AI/ML device requiring adjudication of a test set.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A. This device is not an AI/ML product and does not involve human readers or AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done
N/A. This device is not an AI/ML product.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
N/A. Not an AI/ML device requiring ground truth for performance evaluation. The substantial equivalence argument relies on material characteristics and manufacturing processes.
-
The sample size for the training set
N/A. This is not an AI/ML device with a training set.
-
How the ground truth for the training set was established
N/A. This is not an AI/ML device with a training set.
Ask a specific question about this device
Page 1 of 15