(83 days)
Bovine Dermis Collagen Dermal Matrix is indicated for the management of wounds, including:
- . Full thickness and Partial thickness wounds
- . Chronic wounds (e.g. pressure ulcers, venous ulcers, diabetic ulcers, chronic ulcers)
- . Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- . Trauma wounds (abrasions, lacerations, and skin tears)
- Draining wounds
- Partial thickness burns
Bovine Dermis Collagen Dermal Matrix is an absorbent, porous, collagen matrix engineered from purified collagen derived from bovine dermal tissue. Bovine Dermal Matrix should be applied directly to the wound, covering the entire wound surface.
Bovine Dermis Collagen Dermal Matrix is supplied sterile, non-pyrogenic and for single use only.
The Collagen Matrix, Inc. Bovine Dermis Collagen Dermal Matrix (K243071) is a medical device. The provided text outlines the device's characteristics and compares it to predicate devices to establish substantial equivalence for regulatory purposes.
Here's an analysis of the acceptance criteria and study information, based solely on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" in a go/no-go fashion with numerical targets for clinical performance. Instead, it compares the subject device to predicate devices across various technical characteristics. The implicit acceptance criterion is that the subject device's performance is either equivalent to or better than the predicate devices, or that any differences do not raise new questions of safety or effectiveness.
| Characteristic | Acceptance Criterion (Implicitly "Similar to Predicate" or acceptable range) | Reported Device Performance (K243071) | Predicate K040211 Performance | Predicate K152600 Performance |
|---|---|---|---|---|
| Material/Source | Bovine dermis | Bovine dermis | Same | Same |
| Available Sizes | Acceptable range, risk analysis for larger sizes performed. | Up to 700 cm² | Up to 720 cm² | Up to 8 cm² |
| Thickness | 3 mm | 3 mm | Same | Same |
| Absorption Capacity | >10 mL/g (predicate K040211), or comparable/better | 31.0 mL/g | 40.2 mL/g | Same (Implied, linked to K040211) |
| Cross-linked | Yes | Yes | No | Yes |
| Crosslinker | Formaldehyde (monitored for residuals) | Formaldehyde | N/A | Same |
| Sterilization Method | Gamma irradiation | Gamma irradiation | Same | Same |
| SAL | 10-6 | 10-6 | Same | Same |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Same | Same |
| Packaging | Single barrier (Tyvek pouch) | Single barrier (Tyvek pouch) | Same | Same or blister tray |
| Shelf Life | 36 months | 36 months | Same | Same |
| Residual Formaldehyde | Monitored and mitigated; risk analysis performed for largest size | Monitored and mitigated | N/A | Monitored and mitigated |
| In Vitro Characterization | Demonstrated substantial equivalence to predicates | Performed | N/A | N/A |
| Biocompatibility | Demonstrated safety for long term (>30 days) contact | Performed | N/A | N/A |
| Viral Inactivation | Ensured viral safety | Performed | N/A | N/A |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: The document does not specify a "test set" in the context of clinical or performance data with a specific number of cases or patients. The studies described are in vitro characterization tests, biocompatibility tests, and a viral inactivation study. These are conducted on samples of the device itself or in laboratory settings, not on patient data.
- Data Provenance: Not applicable in the context of clinical patient data. The studies are laboratory-based and conducted on the manufactured device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. The studies are not clinical studies requiring expert ground truth for interpretation of patient data. They are lab tests for material properties, biological safety, and viral inactivation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. This pertains to clinical studies involving human readers and adjudicated outcomes, which are not described here.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No. The document describes laboratory-based testing of a dermal matrix, not a diagnostic AI device requiring human reader analysis.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This pertains to AI algorithm performance studies, which are not described here.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable in the human clinical sense. For the in vitro tests (Absorbency, pH, Thermal Transition Temperature, Density, Crosslinking Agent Residual Content, Pyrogenicity), the "ground truth" would be established by validated analytical methods and reference standards. For biocompatibility, it would be based on established international standards (e.g., ISO 10993 series) and their respective endpoints. For viral inactivation, it would be based on virological assays.
8. The sample size for the training set:
Not applicable. This device is a physical medical device, not a machine learning model, so there is no "training set."
9. How the ground truth for the training set was established:
Not applicable, as there is no training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.
December 19, 2024
Collagen Matrix, Inc. Devin Dragon Senior Regulatory Specialist 15 Thornton Road Oakland, New Jersey 07436
Re: K243071
Trade/Device Name: Bovine Dermis Collagen Dermal Matrix Regulatory Class: Unclassified Product Code: KGN Dated: September 27, 2024 Received: September 27, 2024
Dear Devin Dragon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yu-chieh Chiu -S
Yu-Chieh Chiu, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K243071
Device Name
Bovine Dermis Collagen Dermal Matrix
Indications for Use (Describe)
Bovine Dermis Collagen Dermal Matrix is indicated for the management of wounds, including:
- . Full thickness and Partial thickness wounds
- . Chronic wounds (e.g. pressure ulcers, venous ulcers, diabetic ulcers, chronic ulcers)
- . Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- . Trauma wounds (abrasions, lacerations, and skin tears)
- Draining wounds
- Partial thickness burns
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
1. Contact Details Applicant Name: Collagen Matrix, Inc. Applicant Address: 15 Thornton Road Oakland, New Jersey 07436 Applicant Contact Telephone: (732) 856-2674 Applicant Contact: Devin Dragon Sr. Regulatory Specialist ddragon@regenity.com December 10, 2024 Date Prepared: 2. Device Name Device Trade Name: Bovine Dermis Collagen Dermal Matrix K243071 Device Common Name(s): Collagen Dermal Matrix Device Classification Name: Dressing, Wound, Collagen KGN Unclassified 3. Legally Marketed Devices to Which Substantial Equivalence is Claimed Primary Predicate Device: Collagen Topical Wound Dressing K040211 KGN Secondary Predicate Devices: Collagen Dental Wound Dressings K152600 KGN Device Description Summary 4.
Bovine Dermis Collagen Dermal Matrix is an absorbent, porous, collagen matrix engineered from purified collagen derived from bovine dermal tissue. Bovine Dermal Matrix should be applied directly to the wound, covering the entire wound surface.
Bovine Dermis Collagen Dermal Matrix is supplied sterile, non-pyrogenic and for single use only.
5. Intended Use
Bovine Dermis Collagen Dermal Matrix is indicated for the management of wounds, including:
- Full thickness and Partial thickness wounds ●
- . Chronic wounds (e.g. pressure ulcers, venous ulcers, diabetic ulcers, chronic ulcers)
- Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- Trauma wounds (abrasions, lacerations, and skin tears)
- Draining wounds
- Partial thickness burns
{5}------------------------------------------------
6. Indications for Use Comparison
The subject device and the predicate device (Collagen Topical Wound Dressing, K040211) are both intended for the management of wounds. The indications of both devices align with devices within the KGN product code (Dressing, Wound, Collagen).
7. Technological Comparison
Although the subject and predicate devices are both made primarily from purified Type I bovine dermis collagen, the subject device has an additional formaldehyde crosslinking step and a lower sterilization dose than the predicate device, K040211. The subject device's manufacturing process and sterilization dose is identical to the secondary predicate device K152600. The process of crosslinking the device with formaldehyde has been used in many of Collagen Matrix's cleared 510(k)s, including Collagen Dental Wound Dressing, K152600 and is a well-recognized practice in similar medical devices.
Through the manufacturing process and product release testing, the level of residual formaldehyde is monitored and mitigated for safety. The subject device and is crosslinked using formaldehyde under the same processing conditions that is used for the secondary predicate K152600. Due to the larger maximum size of available dressing sizes for the subject device, a risk analysis was performed to demonstrate that the potential content of formaldehyde residuals in the largest available dressing size does not raise new questions of safety.
The subject and predicate device have the same intended use, principle of operation, and identical starting materials. With the exception of the crosslinking step and lower sterilization dose, the subject and predicate devices have similar technical characteristics, manufacturing materials, performance specifications, and packaging materials.
A summary table comparing the subject device to the predicate devices is included on the next page.
{6}------------------------------------------------
| Subject Device &Predicate Device(s): | K243071 | K040211 | K152600 |
|---|---|---|---|
| General DeviceCharacteristics | Bovine Dermis CollagenDermal Matrix | Collagen Topical WoundDressing | Collagen Dental WoundDressing |
| Intended Use | For the management ofwounds, including:• Full thickness and Partialthickness wounds• Chronic wounds (e.g.pressure ulcers, venousulcers, diabetic ulcers,chronic ulcers)• Surgical wounds (donorsites/grafts, post-Moh'ssurgery, post-laser surgery,podiatric, wounddehiscence)• Trauma wounds(abrasions, lacerations, andskin tears)• Draining wounds• Partial thickness burns | For the management ofmoderately to heavilyexudating wounds, and tocontrol minor bleeding.Collagen Topical WoundDressing may be used for themanagement of exudatingwounds such as:• pressure ulcers• venous stasis ulcers• diabetic ulcers• acute wounds, for exampletrauma and surgical wounds• partial-thickness burns | For the management oforal wounds and sores,including:• Denture sores• Oral ulcers (non-infectedor viral)• Periodontal surgicalwounds• Suture sites• Burns• Extraction sites• Surgical wounds• Traumatic wounds |
| Tissue source | Bovine dermis | Same | Same |
| Available Sizes | Up to 700 cm²• 2 cm x 4 cm• 5 cm x 5 cm• 10 cm x 10 cm• 10 cm x 20 cm• 20 cm x 35 cm | Up to 114 in²(720 cm²)Width: 2 in - 8 in(5 cm - 20 cm)Length: 2 in - 14 in(5 cm - 36 cm) | Up to 8 cm²• 2 cm x 4 cm |
| Thickness | 3 mm | Same | Same |
| AbsorptionCapacity | Specification: >18 mL/gTest: 31.0 mL/g | Specification: >10 mL/gTest: 40.2 mL/g | Same |
| Cross-linked | Yes | No | Yes |
| Crosslinker | Formaldehyde | N/A | Same |
| Sterilization Method | Gamma irradiation | Same | Same |
| SAL | 10-6 | Same | Same |
| Pyrogenicity | Non-pyrogenic | Same | Same |
| Packaging | Single barrier (Tyvek pouch) | Same | Same or blister tray |
| Shelf Life | 36 months | Same | Same |
{7}------------------------------------------------
8. Non-Clinical and/or Clinical Tests Summary & Conclusions
Subject and predicate devices performance data were compared to support the safety of the subject device and demonstrate that the Bovine Dermis Collagen Dermal Matrix perform as intended. The substantial equivalence of Bovine Dermis Collagen Dermal Matrix to its predicates is demonstrated based on in vitro characterization studies and biocompatibility studies.
In vitro product characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate devices. A series of bench tests were conducted to evaluate material properties, biological properties, chemical and physical properties. These tests included: Absorbency, pH, Thermal Transition Temperature, Density, Crosslinking Agent Residual Content, and Pyrogenicity.
A series of biocompatibility testing was performed to assess the safety of the Bovine Dermis Collagen Dermal Matrix as a surface device categorized for long term (>30 days) contact with breach and/or compromised skin surfaces. These tests included: Physical and Chemical Information, Cytotoxicity, Sensitization, Irritation, Material Mediated Pyrogenicity, Acute / Subacute / Subchronic / Chronic Toxicity, Implantation, Genotoxicity, and Carcinogenicity.
A viral inactivation study was performed to ensure the viral safety of the product.
9. Conclusions
The conclusions drawn from the above documentation demonstrate that the proposed device is as safe, effective and performs as well as the legally marketed predicate devices.
N/A