K Number
K250733

Validate with FDA (Live)

Date Cleared
2025-07-03

(114 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3DOSE 1ml Syringe is intended for use by health care professionals for general purpose fluid aspiration/ injection.

Device Description

The 3DOSE 1ml Syringe is a sterile, single-use, disposable and non-reusable manual syringe which is intended for injection of fluids into the body. The syringe can be used for pulling up any mixture, and after operating the locking handle for accurate dosing according to the ml/unit scale on the plunger.

The 3DOSE 1ml Syringe is used like a classic syringe, but includes a feature that provides tactile and audible feedback to the user during dosing.

Models: 3DOSE 1ml Syringe (125 GREEN)
3DOSE 1ml Syringe (100 ORANGE)

AI/ML Overview

The provided FDA 510(k) clearance letter for the 3DOSE 1ml Syringe does not detail any study involving human readers, AI, or diagnostic performance metrics typically associated with AI/ML-based medical devices. This clearance is for a piston syringe, a physical medical device, not an AI/ML software as a medical device (SaMD).

Therefore, I cannot provide information for the following points as they are not applicable to this device and its clearance documentation:

  • A table of acceptance criteria and the reported device performance (in terms of diagnostic metrics like sensitivity, specificity, AUC)
  • Sample sizes used for the test set and data provenance (for diagnostic performance of an algorithm)
  • Number of experts used to establish the ground truth for the test set and their qualifications
  • Adjudication method for the test set
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done
  • If a standalone (algorithm only) performance study was done
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
  • The sample size for the training set
  • How the ground truth for the training set was established

However, I can extract information related to the device's technical specifications, performance standards, and the non-clinical testing performed to establish its safety and effectiveness relative to a predicate device.


Here's the summary of acceptance criteria and the study that proves the device meets them, based on the provided document, specifically for a physical medical device (syringe):

1. Acceptance Criteria and Reported Device Performance (Non-Diagnostic)

For a physical device like a syringe, acceptance criteria relate to its physical properties, functionality, and compliance with recognized standards. The "performance" reported is its compliance with these standards, indicating mechanical functionality, material safety, and sterility.

CategoryAcceptance Criteria (Standard Compliance)Reported Device Performance
Functional Performance- ISO 7886-1:2017: Sterile Hypodermic syringes for single use - Part 1: Syringes for manual use - ISO 80369-7:2016: Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications - ISO 80369-20:2015: Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods- Complies with ISO 7886-1:2017 - Complies with ISO 80369-7:2016 - Complies with ISO 80369-20:2015 (Demonstrated through "Functional Performance Testing")
Sterility- ISO 11135: Sterilization of health-care products - Ethylene oxide - ISO 11607-1: Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems - ISO 11607-2: Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes - ISO 10993-7: Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals- Device is sterile, single-use, disposable. - Sterilized using Ethylene Oxide. - Validated shelf-life of 5 years. (Demonstrated through "Sterility" testing per listed ISO standards)
Biocompatibility- ISO 10993-1: Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process (including Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemocompatibility) - ISO 10993-4: Tests for interactions with blood - ISO 10993-5: Tests for in vitro cytotoxicity - ISO 10993-10: Tests for irritation and skin sensitization - ISO 10993-11: Tests for systemic toxicity - USP <788> Particulate Matter In Injections- Biocompatibility evaluation conducted in accordance with ISO 10993-1 and FDA recognized standards. - Syringe of testing included Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemocompatibility. - All a-forementioned tests passed, demonstrating the device is biocompatible. (Demonstrated through "Biocompatibility Testing")
Chemical Properties- ISO 7886-1 (Section for chemical properties: Limits for acidity or alkalinity, Limits for extractable metals)- Chemical performances inspection based on ISO 7886-1. - Results conform to ISO 7886-1 for limits for acidity or alkalinity and extractable metals. (Demonstrated through "Chemical properties" testing)
Material/Design Integrity- Functionality of the "audible/tactile feature" (ratcheting dose mechanism) - No negative effect on overall device performance due to new features (locking arm, plastic part, audible/tactile feedback)- In-process quality control tests performed to confirm the audio notification mechanism works correctly and does not have a negative effect on overall performance. - Performance testing conducted for the ratcheting lock mechanism (locking arm, plastic part) to ensure it does not raise new questions of safety or effectiveness. (Demonstrated through implicit in-house testing/QC)

2. Sample Sizes used for the Test Set and Data Provenance

The document does not specify exact sample sizes for each non-clinical test (e.g., how many syringes were tested for extractable metals or sterility). It generally states "The collective results of the nonclinical testing demonstrate..." This implies a sufficient number of samples were tested to meet the statistical requirements of the respective ISO standards.

  • Data Provenance: The tests would have been performed in a laboratory setting, likely in Turkey (Bimed Teknik Aletler Sanayi & Ticaret A.S. is based in Istanbul, Turkey) or by accredited labs contracted by them. These are prospective tests performed specifically for this 510(k) submission.

3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts

This concept is not applicable here as there is no "ground truth" in the clinical diagnostic sense. The "truth" for device functionality, sterility, and biocompatibility is established by compliance with internationally recognized consensus standards (ISO, USP) and physical/chemical measurements by qualified laboratory personnel.

4. Adjudication method for the test set

Not applicable. This is not a study assessing human or AI diagnostic performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done

No. This type of study is for evaluating observer performance, typically in medical imaging or diagnostics, and is not relevant to a physical medical device like a syringe.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This device is not an algorithm or software. It is a manually operated piston syringe.

7. The type of ground truth used

For this device, the "ground truth" is defined by the specifications and limits set by the referenced international standards (e.g., ISO 7886-1, ISO 10993 series). For example:

  • Dimensional accuracy of volume: defined by ISO 7886-1 tolerances.
  • Sterility: defined by ISO 11135 and ISO 11607 series (e.g., SAL of 10^-6).
  • Biocompatibility: Defined by the absence of adverse biological reactions as per ISO 10993 series tests.
  • Chemical properties: Defined by limits for acidity/alkalinity and extractable metals in ISO 7886-1.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI/ML model that requires training data.

9. How the ground truth for the training set was established

Not applicable.

FDA 510(k) Clearance Letter - 3DOSE 1ml Syringe

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

July 3, 2025

Bimed Teknik Aletler Sanayi & Ticaret A.S.
Aysel Kilic
Medical Quality Manager
Beylikduzu OSB Mah. Leylak Cd. No:16 Beylikduzu/Istanbul
Istanbul, 34520, Turkey

Re: K250733
Trade/Device Name: 3DOSE 1ml Syringe
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF
Dated: June 4, 2025
Received: June 4, 2025

Dear Aysel Kilic:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K250733 - Aysel Kilic Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250733 - Aysel Kilic Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250733

Device Name
3DOSE 1ml Syringe

Indications for Use (Describe)
The 3 DOSE 1 ml Syringe is intended for use by health care professionals for general purpose fluid aspiration/ injection.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250733
Date Prepared: June 27, 2025

BİMED TEKNİK ALETLER SAN. VE TİC. A.Ş.
Bakır ve Pirinç Sanayi Sitesi Leylak Caddesi No: 16, 34520 Beylikdüzü, İstanbul/Türkiye | T: +90 0212 875 73 76 (Pbx) | F: +90 0212 875 70 52
Tic.Sic: 198367 - 145900 | Marmara Kurumlar VD 8430026332 bimedteknik.com - bimedmedical.com - info@bimedteknik.com

K250733 510(k) SUMMARY

Contact Details

Applicant Name: Bimed Teknik Aletler Sanayi & Ticaret A.S.
Applicant Address: Beylikdüzü OSB Mah. Leylak Cd. No:16 Beylikdüzü/Istanbul Istanbul - 34520 Turkey
Applicant Contact Telephone: +90 0212 875 73
Applicant Contact: Mrs. Aysel Kilic
Applicant Contact Email: aysel.kilic@bimedteknik.com

Device Name

Device Trade Name: 3DOSE 1ml Syringe
Common Name: Piston syringe
Classification Name: Syringe, Piston
Regulation Number: 880.5860
Product Code: FMF

Predicate Device

Trade Name: TK Sterile Piston Syringe without Needle
510(k) Number: K191642
Classification Name: Syringe, Piston
Regulation Number: 880.5860
Product Code: FMF

Device Description Summary

The 3DOSE 1ml Syringe is a sterile, single-use, disposable and non-reusable manual syringe which is intended for injection of fluids into the body. The syringe can be used for pulling up any mixture, and after operating the locking handle for accurate dosing according to the ml/unit scale on the plunger.

The 3DOSE 1ml Syringe is used like a classic syringe, but includes a feature that provides tactile and audible feedback to the user during dosing.

Models: 3DOSE 1ml Syringe (125 GREEN)
3DOSE 1ml Syringe (100 ORANGE)

Page 6

K250733
Date Prepared: June 27, 2025

BİMED TEKNİK ALETLER SAN. VE TİC. A.Ş.
Bakır ve Pirinç Sanayi Sitesi Leylak Caddesi No: 16, 34520 Beylikdüzü, İstanbul/Türkiye | T: +90 0212 875 73 76 (Pbx) | F: +90 0212 875 70 52
Tic.Sic: 198367 - 145900 | Marmara Kurumlar VD 8430026332 bimedteknik.com - bimedmedical.com - info@bimedteknik.com

Material used:

  • 1ml luer lock syringe : polypropylene
  • Syringe plunger-100-a: polyacetal
  • Syringe plunger-125-b: polyacetal
  • Syringe locking arm-orange: polyacetal
  • Syringe locking arm-green: polyacetal
  • Plastic part-a: polyacetal
  • Plastic part-b:polyacetal
  • Spring: stainless steel
  • Colours (white ym-by-107, blue mv-74542 (p.649 u), orange 3682, green yl-83030)

Intended Use/Indications for Use

The 3DOSE 1ml Syringe is intended for use by health care professionals for general purpose fluid aspiration/ injection.

Technological Comparison

Characteristic3DOSE 1ml Syringe K250733TK Sterile Piston Syringes without Needle K191642Discussion
Common NamePiston syringePiston syringeSame – both devices are identified as piston syringes.
Classification NameSyringe, PistonSyringe, PistonSame – both devices fall under the classification "Syringe, Piston."
Product CodeFMFFMFSame – both devices share the product code FMF.
ClassificationClass IIClass IISame – both devices are classified as Class II.
Regulation Number880.5860880.5860Same – both devices fall under regulation number 880.5860.
Indications for useThe 3DOSE 1ml Syringe is intended for use by health care professionals for general purpose fluid aspiration/ injectionTK Sterile Piston Syringe without Needle is intended for use by health care professionals for general purpose fluid aspiration/ injection.Same-both devices are intended to be used by health care professionals for general fluid aspiration and injection
Operation ModeFor manual use onlyFor manual use onlySame – both devices are manually operated by the user.
Intended userThese syringes are intended for use by health care professionals.These syringes are intended for use by health care professionals.Same – both devices are intended to be used by trained health care professionals.
Syringe Volume1 ml1ml, 2ml, 3ml, 5ml, 10ml, 20ml, 30ml, 50ml, 60mlDifferent - the subject device is only offered in 1 ml capacity. The limitation of capacity does not introduce new concerns and is evaluated for performance

Page 7

K250733
Date Prepared: June 27, 2025

BİMED TEKNİK ALETLER SAN. VE TİC. A.Ş.
Bakır ve Pirinç Sanayi Sitesi Leylak Caddesi No: 16, 34520 Beylikdüzü, İstanbul/Türkiye | T: +90 0212 875 73 76 (Pbx) | F: +90 0212 875 70 52
Tic.Sic: 198367 - 145900 | Marmara Kurumlar VD 8430026332 bimedteknik.com - bimedmedical.com - info@bimedteknik.com

Characteristic3DOSE 1ml Syringe K250733TK Sterile Piston Syringes without Needle K191642Discussion
Nozzle typeLuer LockLuer Slip and Luer LockDifferent – the nozzle type for the subject device only includes Luer Lock. The predicate device's nozzle type covers that of subject device. Therefore, the differences do not raise new questions about safety and effectiveness.
Sterilization MethodEthylene OxideEthylene OxideSame – both devices are sterilized using ethylene oxide (EO) sterilization method.
Shelf – life5 years5 yearsSame – both devices have a validated shelf-life of 5 years.
Single UseYesYesSame – both devices are intended for single use only.
Performance specificationsComplies with ISO 7886-1:2017 Sterile Hypodermic syringes for single use-Part 1: Syringes for manual use Complies with ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications-- Part 7: Connectors for intravascular or hypodermic applications Complies with ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications-- Part 20: Common test methods- Complies with ISO 7886-1:2017 Sterile Hypodermic syringes for single use-Part 1: Syringes for manual use - Complies with ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications-- Part 7: Connectors for intravascular or hypodermic applications - Complies with ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications-- Part 20: Common test methodsSame – both devices comply with the applicable ISO standards for sterile hypodermic syringes and small-bore connector performance.
Configuration and materialBarrel - Polypropylene (PP) Plunger - Polyacetal Piston - Polyisoprene RubberBarrel - Polypropylene (PP) Plunger - Polypropylene (PP) Piston - Polyisoprene RubberDifferent – the materials of the plungers are different between the proposed device and predicate device. The biocompatibility test of the proposed device was conducted to demonstrate that the subject device is biocompatible. This difference does not raise any new safety and effectiveness questions.
BiocompatibilityThe biocompatibility evaluation for the 3DOSE 1ml Syringe was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices − Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA and the "Use of International Standard ISO 10993-1 "Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process". The syringe of testing included the following tests: Cytotoxicity Sensitization Irritation Acute Systemic Toxicity Pyrogenicity HemocompatibilityThe biocompatibility evaluation for the TK Sterile Piston Syringe without needle was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices − Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA and the "Use of International Standard ISO 10993-1 "Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process". The syringe of testing included the following tests: Cytotoxicity Sensitization Irritation Acute Systemic Toxicity Pyrogenicity HemocompatibilitySame – both devices were evaluated per ISO 10993 standards and passed the full panel of required biocompatibility tests.

Page 8

K250733
Date Prepared: June 27, 2025

BİMED TEKNİK ALETLER SAN. VE TİC. A.Ş.
Bakır ve Pirinç Sanayi Sitesi Leylak Caddesi No: 16, 34520 Beylikdüzü, İstanbul/Türkiye | T: +90 0212 875 73 76 (Pbx) | F: +90 0212 875 70 52
Tic.Sic: 198367 - 145900 | Marmara Kurumlar VD 8430026332 bimedteknik.com - bimedmedical.com - info@bimedteknik.com

Characteristic3DOSE 1ml Syringe K250733TK Sterile Piston Syringes without Needle K191642Discussion
Chemical propertiesChemical performances inspection are based on ISO 7886-1, inspection items are as follows: Limits for acidity or alkalinity, Limits for extractable metals. Results conform to ISO7886-1.Chemical performances inspection are based on ISO 7886-1, inspection items are as follows: Limits for acidity or alkalinity, Limits for extractable metals. Results conform to ISO7886-1.Same – both devices were evaluated for chemical properties per ISO 7886-1 and met the acceptance criteria.
Graduation lines and audible/tactile featureThe piston syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger. Additionally, there are extra parts to use the audible/tactile feature (locking arm, plastic part). There are graduation lines and teeth on the plunger.The piston syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger.Different – There are graduation lines on the barrel and piston. The extra parts integrated into the device enable the operation of the ratcheting dose mechanism, which provides both audible and tactile feedback to the user during injection. In addition, the additional audio notification feature of our device provides feedback to the user during the injection process and does not change the basic technological functionality and purpose of use of the device. In order to ensure the safety and effectiveness of the audio notification feature, in-process quality control tests were performed. These tests confirmed that the audio notification mechanism works correctly and does not have a negative effect on the overall performance of the device. For these reasons, it was concluded that our device is 'Substantially Equivalent' (SE) with the equivalent product in terms of safety and effectiveness.
Ratcheting Lock MechanismIncluded – the device contains a ratcheting lock mechanism to initiate injection and provide tactile/audible feedback.Not includedDifferent – The subject device includes a locking arm and a plastic part not present in the predicate device. The locking arm is set to a "free" mode before drawing up the medicine, and once the syringe is filled, it is switched to "operation" mode, making the syringe ready for injection. The plastic part functions as a housing element, connecting the locking mechanism to the syringe body. This feature enhances ease of use but does not alter the fundamental operating principle of the device. It was evaluated through performance testing and determined not to raise new questions of safety or effectiveness.

Page 9

K250733
Date Prepared: June 27, 2025

BİMED TEKNİK ALETLER SAN. VE TİC. A.Ş.
Bakır ve Pirinç Sanayi Sitesi Leylak Caddesi No: 16, 34520 Beylikdüzü, İstanbul/Türkiye | T: +90 0212 875 73 76 (Pbx) | F: +90 0212 875 70 52
Tic.Sic: 198367 - 145900 | Marmara Kurumlar VD 8430026332 bimedteknik.com - bimedmedical.com - info@bimedteknik.com

Non-Clinical and/or Clinical Tests Summary & Conclusions

Functional Performance Testing

  • ISO 7886-1
  • ISO 80369-7

Sterility

  • ISO 11135
  • ISO 11607-1
  • ISO 11607-2
  • ISO 10993-7

Biocompatibility Testing

  • ISO 10993-1
  • ISO 10993-4
  • ISO 10993-5
  • ISO 10993-10
  • ISO 10993-11
  • USP<788>

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the 3DOSE 1ml Syringe meet the established specifications necessary for safety and performance during its intended use. In addition, the collective bench testing demonstrates that the subject device does not raise different questions of safety or effectiveness when compared to the predicate device.

Conclusions:

The 3DOSE 1 ml Syringe is substantially equivalent to the TK Sterile Piston Syringe without Needle. The materials, performance, and operational features of both the subject device and the predicate device are substantially equivalent.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).