(113 days)
FIBERGRAFT® BG Putty - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Putty must be used with autograft in the posterolateral spine.
FIBERGRAFT® BG Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.
The FIBERGRAFT®BG Putty is an osteoconductive, resorbable, biocompatible bone graft substitute to be gently packed into defect sites and to be used as non-structural scaffolds. The FIBERGRATT® BG putty is made from 45S5 bioactive glass, where the bioactive glass components are mixed with an absorbable binder to form a cohesive material.
The provided text does not contain information about acceptance criteria or a study that proves a device meets specific performance criteria in terms of accuracy, sensitivity, or specificity, as would be relevant for an AI/ML device.
Instead, the document is a 510(k) premarket notification for a medical device called "FIBERGRAFT® BG Putty Bone Graft Substitute". The 510(k) clearance process focuses on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving performance against specific quantitative accuracy metrics for software.
The performance data discussed relates to:
- Physical and chemical property evaluation studies: These confirmed in vitro functionality and bioactivity, but it explicitly states, "The in vitro bioactivity test results have not been correlated to clinical performance."
- Biocompatibility tests (ISO 10993 testing): This demonstrates the material's safety for biological contact.
- Packaging evaluations, shelf life testing, and real-time aging testing.
- Bacterial endotoxin testing.
- Functional performance animal studies:
- Ovine model: Evaluated in critical size defects in skeletally mature sheep (total of 58 tested samples). This study demonstrated evidence of new bone formation with "substantially equivalent performance compared to the predicate device."
- Rabbit posterolateral spine fusion model: Compared to predicate devices and positive controls in 71 skeletally mature rabbits. This study also showed "substantially equivalently to the predicate devices."
The animal studies used radiographic, histomorphometric, and biomechanical data to evaluate performance, which are methods for assessing bone formation and structural integrity. However, the document does not present specific acceptance criteria in a table format with corresponding device performance metrics (e.g., bone volume percentages, fusion rates above a certain threshold) that could be definitively extracted and formatted as requested. The conclusion is based on "substantial equivalence" to predicate devices, meaning it performs as safely and effectively without raising new safety or effectiveness concerns, rather than meeting a predefined quantitative performance target.
Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance, nor details about sample sizes for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance, as these concepts are not addressed in the provided 510(k) summary for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human profile in triplicate.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
Prosidyan, Inc. % Ms. Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K170306
Trade/Device Name: FIBERGRAFT® BG Putty Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: April 11, 2017 Received: April 11, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Vincent J. Devlin -S
- for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known) K170306
Device Name
FIBERGRAFT® BG Putty Bone Graft Substitute
Indications for Use (Describe)
FIBERGRAFT® BG Putty - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Putty must be used with autograft in the posterolateral spine.
FIBERGRAFT® BG Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
□ Over-The-Counter Use (21 CFR 801
PLEASE DO NOT WRITE BELOW THIS LINE
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
510(k) SUMMARY
Prosidyan, Inc.'s FIBERGRAFT® BG Putty – Bone Graft Substitute
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Prosidyan, Inc. 30 Technology Drive Warren, NJ 07059 Phone: (610)-945-5640 Facsimile: (908) 396-1151 Contact Person: Charanpreet S. Bagga
Date Prepared: May 15, 2017
Name of Device and Name
FIBERGRAFT® BG Putty Bone Graft Substitute
Common or Usual Name
Bone Void Filler
Classification Name/CFR Requlation/Product Code
Resorbable Calcium Salt Bone Void Filler, 21 CFR 888.3045, product code MQV
Predicate Devices
- Prosidyan, Inc.'s FIBERGRAFT® BG Morsels Bone Graft Substitute (K151154, K141956, K132805) ● (Primary Predicate)
- Prosidyan, Inc.'s FIBERGRAFT® BG Putty Bone Graft Substitute (K143533) .
- . NovaBone Products, LLC's NovaBone Putty- Bioactive Synthetic Bone Graft (K060728, K080009, K101860, K110368, K112773)
Intended Use / Indications for Use
FIBERGRAFT® BG Putty - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Putty must be used with autograft in the posterolateral spine.
FIBERGRAFT® BG Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.
Device Description
The FIBERGRAFT®BG Putty is an osteoconductive, resorbable, biocompatible bone graft substitute to be gently packed into defect sites and to be used as non-structural scaffolds. The FIBERGRATT® BG putty is made from 45S5 bioactive glass, where the bioactive glass components are mixed with an absorbable binder to form a cohesive material.
{4}------------------------------------------------
Technological Characteristics
The technological characteristics of the FIBERGRAFT® BG Putty have not changed since the prior clearance of the product. FIBERGRAFT® BG Putty provides an osteoconductive, resorbable, biocompatible bone graft substitute made from 45S5 bioactive glass mixed with a polymer carrier.
The device is designed as an osteoconductive space-filling device to be gently packed into defect sites and used as a non-structural scaffold for the body's natural healing and bone regeneration process. The device indications are the same as for the predicate devices.
Bioactive glass is defined as a group of dlasses which has a compositional range that allows the formation of hydroxyapatite (HA) as a surface layer when exposed to an aqueous phosphatecontaining solution such as simulated body fluid. The HA layer that forms in an aqueous phosphatecontaining solution plays a significant role in forming a strong bond with natural bone. The bioactive glass in the FIBERGRAFT® BG Putty is the same as the bioactive glass used in the FIBERGRAFT BG Morsels and Novabone Putty products. In addition, both the BG Putty and the Novabone Putty are provided in putty form using a resorbable polymer carrier.
Performance Data
The performance of the FIBERGRAFT® BG Putty has been established through physical and chemical property evaluation studies, functional performance animal studies and biocompatibility tests. The physical and chemical property studies confirmed the in vitro functionality and bioactivity of the BG Putty. The in vitro bioactivity test results have not been correlated to clinical performance. The biocompatibility of the FIBERGRAFT® BG Putty is demonstrated by ISO 10993 testing and the long history of clinical use of the bioactive glass material for the same intended use. In addition, the BG Putty is composed of the same bioactive glass material with the same bioactive glass chemical composition and the same type and duration of patient contact as the predicates. Packaging evaluations, shelf life testing and real time aging testing were performed with passing results. Bacterial endotoxin testing was performed using the limulus amebocyte lysate (LAL) method and showed that the device meets the endotoxin limits of established quidelines.
The functional performance of the FIBERGRAFT® BG Putty was evaluated in critical size defects in an ovine model, consistent with FDA's recommendations for Class II synthetic bone graft substitutes. Animal testing demonstrated evidence of new bone formation in critical size defects, with substantially equivalent performance compared to the predicate device. A total of 58 tested samples were evaluated in the study from skeletally mature sheep, including radiographic, histomorphometric, and biomechanical data.
Testing of the BG Putty in the ovine model is representative of the indications for use and range of anatomical sites proposed for the subject device. The study was conducted for a duration of 24 weeks with interim evaluation points. The results of the study demonstrated that the BG Putty device performs as safely and as effectively as the predicate device, and any differences between the results of the device groups do not raise new types of safety or effectiveness concerns.
In addition, the FIBERGRAFT® BG Putty product was also evaluated in a rabbit study to further support device performance for its indications for use. The FIBERGRAFT® BG Putty product was compared to its predicate devices as well as positive controls. The animal study evaluated device performance in a rabbit posterolateral spine fusion model in 71 skeletally mature rabbits. The performance was evaluated using radiographic, histomorphometric, and biomechanical data. Testing of the FIBERGRAFT® BG Putty in the rabbit posterolateral spine fusion model is representative of the indications for use and range of anatomical sites proposed for the subject device. The results of the study through 26 week follow up demonstrated that the FIBERGRAFT® BG Putty device performs substantially equivalently to the predicate devices, and any minor technological differences between the device groups do not raise new types of safety or effectiveness concerns.
{5}------------------------------------------------
Therefore, performance testing demonstrated that the FIBERGRAFT® BG Putty device functions as intended and meets the requirements of class II bone void fillers as compared to the predicate devices.
Substantial Equivalence
FIBERGRAFT® BG Putty has the same technological characteristics as previously cleared by FDA; the sole change in this submission is the addition of an indication for use in posterolateral fusion. This indication is the same as other predicate materials with similar composition. The FIBERGRAFT® BG Putty is composed of the same bioactive glass material with the same chemical composition and the same type and duration of patient contact as the Prosidyan FIBERGRAFT® BG Putty Bone Graft Substitute and NovaBone predicates. The minor technological differences between BG Putty and its predicate devices do not raise any new issues of safety or effectiveness. As demonstrated in performance testing, including testing in an animal model, BG Putty performs in a manner that is substantially equivalent to the predicate devices when used for posterolateral fusion.
Conclusion
FIBERGRAFT® BG Putty is an osteoconductive, resorbable, biocompatible bone graft substitute composed of bioactive glass, mixed with a polymer carrier. The FIBERGRAFT® BG Putty is substantially equivalent to its predicate devices for its intended use as a synthetic bone void filler. Performance testing, including in vivo data, demonstrated that the device functions as intended without raising new safety or effectiveness questions.
§ 888.3045 Resorbable calcium salt bone void filler device.
(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.