K Number
K130200
Device Name
PICCOLO TOTAL CHOLESTEROL - CAPILLARY TEST SYSTEM
Manufacturer
Date Cleared
2013-02-15

(18 days)

Product Code
Regulation Number
862.1175
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Piccolo Total Cholesterol - Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of total cholesterol in capillary (fingerstick) heparinized whole blood in a clinical laboratory setting or point-of-care location. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein disorders.
Device Description
The Piccolo® Lipid Panel - Capillary Reagent Disc (which contains the Piccolo® Total Cholesterol - Capillary Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer.
More Information

No
The summary describes a chemical analysis system based on reagent discs and photometric measurement, with no mention of AI or ML in the device description, performance studies, or key metrics.

No
The device is used for the in vitro quantitative determination of total cholesterol, which is a diagnostic measurement, not a therapeutic intervention.

Yes

The "Intended Use / Indications for Use" section explicitly states that "Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein disorders." This indicates that the device provides information crucial for diagnosis.

No

The device description clearly outlines a physical reagent disc and its interaction with a chemistry analyzer, indicating hardware components are integral to the device's function.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The "Intended Use / Indications for Use" section explicitly states that the system is intended for the "in vitro quantitative determination of total cholesterol in capillary (fingerstick) heparinized whole blood". "In vitro" means "in glass" or "outside the body," which is a key characteristic of IVDs.
  • Device Description: The description details how the device processes a blood sample (separating plasma, mixing with reagents) to perform a chemical analysis. This process is performed on a sample taken from the body, not directly on the body itself.
  • Clinical Purpose: The intended use also states that the measurements are used in the "diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein disorders." This indicates a clinical purpose for the test results, which is another defining feature of IVDs.

N/A

Intended Use / Indications for Use

The Piccolo Total Cholesterol - Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of total cholesterol in capillary (fingerstick) heparinized whole blood in a clinical laboratory setting or point-of-care location.

Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein disorders.

Product codes (comma separated list FDA assigned to the subject device)

CHH

Device Description

The Piccolo® Lipid Panel - Capillary Reagent Disc (which contains the Piccolo® Total Cholesterol - Capillary Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Professional use / Clinical laboratory setting or point-of-care location.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Linearity:
The dynamic range of the Piccolo® Total Cholesterol - Capillary Test System is from 20 to 520 mg/dL. This was established using samples across the measuring range, see previously cleared submission K023642. Testing was done using a total of 20 Piccolo analyzers to ensure that the recoveries were not instrument-specific.
Slope: 0.989
Intercept: 19.7
Correlation Coefficient (r): 0.999

Precision:
Precision studies were designed to evaluate within-run and total precision of total Cholesterol included on the Piccolo® Lipid Panel Reagent Disc when run on the Abaxis analyzer.
Within-Run (n =160) for Serum 1: Mean = 223.7 mg/dL, SD = 3.0, %CV = 1.3
Total (n =160) for Serum 1: Mean = 223.7 mg/dL, SD = 5.7, %CV = 2.6
Within-Run (n =160) for Serum 2: Mean = 202.2 mg/dL, SD = 3.1, %CV = 1.5
Total (n =160) for Serum 2: Mean = 202.2 mg/dL, SD = 4.4, %CV = 2.2

Whole Blood Precision (tested seven times on four analyzers over three hours; 20 analyzers used):
Sample 1: Mean = 210.8 mg/dL, SD = 3.6, %CV = 1.7
Sample 2: Mean = 216.8 mg/dL, SD = 3.1, %CV = 1.4
Sample 3: Mean = 239.1 mg/dL, SD = 3.8, %CV = 1.6
Sample 4: Mean = 184.2 mg/dL, SD = 3.2, %CV = 1.7
Sample 5: Mean = 236.2 mg/dL, SD = 3.6, %CV = 1.5

Method Comparison:
Site 1: N = 216 for both Piccolo and Roche.
Piccolo Mean = 186.5 mg/dL, Std. Dev. = 53.6, Range = 21 - 283 mg/dL.
Roche Mean = 190.6 mg/dL, Std. Dev. = 55.3, Range (Plasma) = 23.5 - 289.5 mg/dL.
Linear Regression: Slope = 0.97, Intercept = 2.42, R2 = 0.991, SEE = 5.2.
Deming Regression: Slope = 0.97, Intercept = 1.57, R2 = 0.991, SEE = 5.2.

Site 2: N = 210 for both Piccolo and Roche.
Piccolo Mean = 185.9 mg/dL, Std. Dev. = 56.0, Range = 21 - 412 mg/dL.
Roche Mean = 190.8 mg/dL, Std. Dev. = 58.1, Range (Plasma) = 24 - 442.5 mg/dL.
Linear Regression: Slope = 0.96, Intercept = 2.83, R2 = 0.991, SEE = 5.4.
Deming Regression: Slope = 0.96, Intercept = 1.99, R2 = 0.991, SEE = 5.4.

Site 3: N = 213 for both Piccolo and Roche.
Piccolo Mean = 184.9 mg/dL, Std. Dev. = 51.2, Range = 22 - 320 mg/dL.
Roche Mean = 190.9 mg/dL, Std. Dev. = 52.9, Range (Plasma) = 23.5 - 340.5 mg/dL.
Linear Regression: Slope = 0.96, Intercept = 1.30, R2 = 0.987, SEE = 6.0.
Deming Regression: Slope = 0.97, Intercept = 0.09, R2 = 0.987, SEE = 6.0.

Combined Data: N = 639 for both Piccolo and Roche.
Piccolo Mean = 185.8 mg/dL, Std. Dev. = 53.5, Range = 21 - 412 mg/dL.
Roche Mean = 190.8 mg/dL, Std. Dev. = 55.4, Range (Plasma) = 23.5 - 442.5 mg/dL.
Linear Regression: Slope = 0.96, Intercept = 2.20, R2 = 0.989, SEE = 5.6.
Deming Regression: Slope = 0.97, Intercept = 1.23, R2 = 0.989, SEE = 5.6.

Accuracy:
Accuracy of the Piccolo method for Cholesterol was established by completing the certification process of the CRMLN.

Conclusions:
The clinical and non-clinical tests performed using the Piccolo® Total Cholesterol -Capillary Test System, when run on the Piccolo® xpress Chemistry Analyzer, demonstrate that the test system is as safe, effective and performs as well as the legally marketed devices identified above.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found. Key metrics presented as correlation coefficient, slope, intercept, standard deviation, %CV, and standard error of the estimate.

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K031824

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K023642

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 862.1175 Cholesterol (total) test system.

(a)
Identification. A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.

0

ABAXIS

3240 Whipple Road, Union City, CA 94587 Phone 510 . 675-6500 Fax 510 . 441-6150

FEB 1 5 2013

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

K130200 The assigned 510(k) number is: _

1. Applicant Information:

Date Prepared:February 8, 2013
Name:Abaxis, Inc.
Address:3240 Whipple Road
Union City, CA 94587
Contact Person:Dennis M. Bleile, PhD
Phone Number:(510) 675-6515
Fax Number:(510) 405-8871

2. Device Information:

ClassificationClass I- Reserved (point-of-care)
Trade Name:Piccolo® Total Cholesterol -Capillary Test System
Classification Name:Total Cholesterol Test system
862.1175

3. Identification of Legally Marketed Device to which the Submitter Claims Equivalence:

The following table identifies the legally marketed device to which Abaxis claims equivalence:

Predicate Device
Predicate DeviceManufacturer510(k) NumberDate of SE Determination
Cobas Cholesterol
Gen.2Roche Diagnostics
Indianapolis, INK0318247/09/2003

. . . . .

" ﻣﺤ

area

i

1

4. Description of the Device:

The Piccolo® Lipid Panel - Capillary Reagent Disc (which contains the Piccolo® Total Cholesterol - Capillary Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer.

5. Statement of Intended Use:

The Piccolo Total Cholesterol - Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of total cholesterol in capillary (fingerstick) heparinized whole blood in a clinical faboratory setting or point-of-care location.

Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein disorders.

Summary of the Technological Characteristics of the New Device in હ. Comparison to those of the Predicate Device:

Table 1 outlines the technological characteristics of the Piccolo® Total Cholesterol -Capillary Test System in comparison to the legally marketed predicate device.

2

:

Table 1: Specification Comparison for Piccolo Total Cholesterol – Capillary Test System and the Roche Total Cholesterol Test on the Cobas 6000 Analyzer

| | Piccolo Total Cholesterol -
Capillary Test on Abaxis
Chemistry Analyzer | Roche Total Cholesterol Test
on the Cobas 6000 Analyzer |
|------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Intended Use | Quantitative analysis of Total
Cholesterol | Quantitative analysis of Total
Cholesterol |
| Methodology | Enzymatic endpoint reaction | Enzymatic endpoint reaction |
| Sample Type | Lithium heparinized capillary whole
blood | Lithium and potassium heparinized
plasma and serum |
| Dynamic Range
Lower Limit | 20 mg/dL | 3.86 mg/dL |
| Reagents | Dry test-specific reagent beads
and liquid diluent; reconstitution
performed by analyzer | Liquid reagent |
| Temperature of
Reaction | 37°C | 37°C |
| Calibration | Bar code with factory
calibrated lot specific data | Calibrated periodically using
calibrators supplied by vendor |
| Assay Range | 20 - 520 mg/dL | 3.86 - 800 mg/dL |
| Testing Environment | Professional use | Professional use |
| Sample Size | Approx 100 µL | 2 µL |

. "

, "

3

7. Brief Discussion of the Clinical and Nonclinical Tests Relied on for a Determination of Substantial Equivalence.

The following information summarizes the results of clinical and non-clinical tests performed using the Piccolo® Total Cholesterol - Capillary Test System. This device is for the addition of capillary sampling on the previously cleared device K023642.

Linearity:

The dynamic range of the Piccolo® Total Cholesterol - Capillary Test System is from 20 to 520 mg/dL. This was established using samples across the measuring range, see previously cleared submission K023642. Testing was done using a total of 20 Piccolo analyzers to ensure that the recoveries were not instrument-specific.

| | Total Cholesterol -
Capillary |
|-----------------------------|----------------------------------|
| Slope | 0.989 |
| Intercept | 19.7 |
| Correlation Coefficient (r) | 0.999 |

Table 2: Summary of Linearity

Precision:

Precision studies were designed to evaluate within-run and total precision of total Cholesterol included on the Piccolo® Lipid Panel Reagent Disc when run on the Abaxis analyzer.

4

Table 3: Within-Run and Total Precision of Total Cholesterol Assayed on the Abaxis Analyzer

| | Within-Run
(n =160) | Total
(n =160) |
|----------------------------------|------------------------|-------------------|
| Total Cholesterol (mg/dL) | | |
| Serum 1 | | |
| Mean | 223.7 | 223.7 |
| SD | 3.0 | 5.7 |
| %CV | 1.3 | 2.6 |
| Serum 2 | | |
| Mean | 202.2 | 202.2 |
| SD | 3.1 | 4.4 |
| %CV | 1.5 | 2.2 |

Table 4: Whole Blood Precision for Total Cholesterol Assayed on the Abaxis Analyzer: Five fresh whole blood samples were tested seven times on four analyzers over a period of three hours; a total of 20 analyzers were used.

Total Cholesterol (mg/dL)

Sample 1Sample 2Sample 3Sample 4Sample 5
Mean210.8216.8239.1184.2236.2
SD3.63.13.83.23.6
%CV1.71.41.61.71.5
ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ2828282828

Page 5

.

5

Method Comparison:

Table 5: Method Comparison Data for Total Cholesterol Assayed by the Abaxis Piccolo Total Cholesterol - Capillary Test System and the Roche Cholesterol Test - Site 1

ParametersStatistics
Piccolo Total Cholesterol - Capillary: Singlicate Values, N216
Roche Cholesterol Assay: Average of Duplicates, N216
Piccolo Total Cholesterol - Capillary: Mean186.5
Roche Cholesterol Assay: Mean190.6
Piccolo Total Cholesterol - Capillary: Std. Dev.53.6
Roche Cholesterol Assay: Std. Dev55.3
Piccolo Total Cholesterol - Capillary: Range of Samples21 - 283
Roche Cholesterol Assay: Range of Samples (Plasma)23.5 - 289.5
Parameters (Roche on X Axis)Linear RegressionDeming Regression
N216216
Slope (95% CI)0.97 (0.95 to 0.98)0.97 (0.96 to 0.98)
Intercept (95% CI)2.42 (-0.10 to 4.95)1.57 (-0.63 to 3.78)
Correlation Coefficient (R2)0.9910.991
Std. Error of the Estimate (SEE)5.25.2

Abaxis «

6

Table 6: Method Comparison Data for Total Cholesterol Assayed by the Abaxis Piccolo Total Cholesterol - Capillary Test System and the Roche Cholesterol Test - Site 2

ParametersStatistics
Piccolo Total Cholesterol - Capillary: Singlicate Values, N210
Roche Cholesterol Assay: Average of Duplicates, N210
Piccolo Total Cholesterol - Capillary: Mean185.9
Roche Cholesterol Assay: Mean190.8
Piccolo Total Cholesterol - Capillary: Std. Dev.56.0
Roche Cholesterol Assay: Std. Dev58.1
Piccolo Total Cholesterol - Capillary: Range of Samples21 - 412
Roche Cholesterol Assay: Range of Samples (Plasma)24 - 442.5
Parameters (Roche on X Axis)Linear RegressionDeming Regression
N210210
Slope (95% CI)0.96 (0.95 to 0.97)0.96 (0.95 to 0.98)
Intercept (95% CI)2.83 (0.28 to 5.37)1.99 (-1.08 to 5.07)
Correlation Coefficient (R2)0.9910.991
Std. Error of the Estimate (SEE)5.45.4

7

Table 7: Method Comparison Data for Total Cholesterol Assayed by the Abaxis Piccolo Total Cholesterol – Capillary Test System and the Roche Cholesterol Test - Site 3

ParametersStatistics
Piccolo Total Cholesterol - Capillary: Singlicate Values, N213
Roche Cholesterol Assay: Average of Duplicates, N213
Piccolo Total Cholesterol - Capillary: Mean184.9
Roche Cholesterol Assay: Mean190.9
Piccolo Total Cholesterol - Capillary: Std. Dev.51.2
Roche Cholesterol Assay: Std. Dev52.9
Piccolo Total Cholesterol - Capillary: Range of Samples22 - 320
Roche Cholesterol Assay: Range of Samples (Plasma)23.5 - 340.5
Parameters (Roche on X Axis)Linear RegressionDeming Regression
N213213
Slope (95% CI)0.96 (0.95 to 0.98)0.97 (0.95 to 0.98)
Intercept (95% CI)1.30 (-1.73 to 4.33)0.09 (-2.48 to 2.65)
Correlation Coefficient (R2)0.9870.987
Std. Error of the Estimate (SEE)6.06.0

8

Table 8: Method Comparison Data for Total Cholesterol Assayed by the Abaxis Piccolo Total Cholesterol - Capillary Test System and the Roche Cholesterol Test - Combined Data

ParametersStatistics
Piccolo Total Cholesterol - Capillary: Singlicate Values, N639
Roche Cholesterol Assay: Average of Duplicates, N639
Piccolo Total Cholesterol - Capillary: Mean185.8
Roche Cholesterol Assay: Mean190.8
Piccolo Total Cholesterol - Capillary: Std. Dev.53.5
Roche Cholesterol Assay: Std. Dev55.4
Piccolo Total Cholesterol - Capillary: Range of Samples21 - 412
Roche Cholesterol Assay: Range of Samples (Plasma)23.5 - 442.5
Parameters (Roche on X Axis)Linear RegressionDeming Regression
N639639
Slope (95% CI)0.96 (0.95 to 0.97)0.97 (0.96 to 0.98)
Intercept (95% CI)2.20 (0.64 to 3.76)1.23 (-0.29 to 2.75)
Correlation Coefficient (R2)0.9890.989
Std. Error of the Estimate (SEE)5.65.6

8. Accuracy

Accuracy of the Piccolo method for Cholesterol was established by completing the certification process of the CRMLN.

9. Conclusions

The clinical and non-clinical tests performed using the Piccolo® Total Cholesterol -Capillary Test System, when run on the Piccolo® xpress Chemistry Analyzer, demonstrate that the test system is as safe, effective and performs as well as the legally marketed devices identified above.

.

9

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines extending from its body, resembling flowing ribbons or streams. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue

Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Public Health Service

February 15, 2013

Abaxis, Inc. c/o Dennis M. Bleile, Ph.D. 3240 Whipple Road Union City, CA 94587

Re: K130200

Trade/Device Name: Piccolo® Total Cholesterol - Capillary Test System Regulation Number: 21 CFR 862.1175 Regulation Name: Cholesterol (total) test system Regulatory Class: I, meets limitations of exemptions per 21 CFR 862.9 (c)(9) Product Code: CHH Dated: January 25, 2013 Received: January 28, 2013

Dear Dr. Bleile:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must . comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for

10

Page 2-Dr. Bleile

the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Carol C. Benson for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

11

Indications for Use Form

510(k) Number (if known):

Device Name: Piccolo® Total Cholesterol - Capillary Test System

Indications for Use:

The Piccolo Total Cholesterol - Capillary Test System used with the Piccolo xpress Chemistry Analyzer is intended for the in vitro quantitative determination of total cholesterol in capillary (fingerstick) heparinized whole blood in a clinical laboratory setting or point-of-care location.

Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein disorders.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Ruth A. @hesler

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

K130200 510(k) =

Page 1 of _ 1 _ _ _ _