Search Results
Found 1 results
510(k) Data Aggregation
(28 days)
The Riptide™ Aspiration Pump is intended as a vacuum source for Medtronic aspiration devices or systems for use in hospitals or clinics. Refer to the Instructions for Use included with the accessory aspiration device for indications related to the procedure for which this device will be used.
The Riptide™ Aspiration Pump is an externally powered electromechanical device capable of generating a vacuum designed for drawing fluids and thrombus into the Riptide™ Collection Canister during interventional procedures. It is intended for general suction use in hospitals or clinics and is not intended for transport or field applications. The Riptide™ Aspiration Pump includes a receptacle that holds the Riptide™ Collection Canister with Intermediate Tubing in place. The intermediate tubing is connected to the vacuum inlet port.
The Riptide™ Aspiration Pump package includes the following:
- Riptide™ Aspiration Pump (LMT-RAP)
- Region specific power cord for connection to earthed receptacle.
- Riptide™ Aspiration Pump User's Manual
The following item is required for use with Riptide™ Aspiration Pump (packaged separately):
- Riptide™ Collection Canister with Intermediate Tubing (LMT-RCT)
The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single use. The Riptide™ Collection Canister is the repository for aspirated material. The Riptide™ Canister is placed into the receptacle of the Riptide™ Aspiration Pump as shown in Figure 3. The Intermediate Tubing is then connected to the vacuum inlet port.
Here's an analysis of the Riptide™ Aspiration Pump's acceptance criteria and the study proving it, based on the provided document:
This document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive de novo clinical trial data. Therefore, the device's acceptance criteria are framed in terms of meeting established standards and matching the performance of a legally marketed predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Riptide™ Aspiration Pump are primarily based on compliance with international standards for medical electrical equipment and medical suction equipment, as well as demonstrating equivalent performance to the predicate device.
| Acceptance Criteria (Standard / Characteristic) | Reported Device Performance |
|---|---|
| Electrical Safety (IEC 60601-1, Edition 3.2 2020-08) | Pass |
| Electromagnetic Compatibility (IEC 60601-1-2, Edition 4.1 2020-09) | Pass |
| User Manual Design Validation (FDA Guidance & ISO 14971, ANSI/AAMI HE75) | Pass (evaluated per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices" and recognized consensus standards ISO 14971 and ANSI/AAMI HE75) |
| Medical Suction Equipment (ISO 10079-1, Fourth Edition 2022-03) | Pass |
| Medical Suction Equipment (ISO 10079-4, First Edition 2021-08) | Pass |
| Vacuum Range | 0-29 inHg (Same as predicate device Penumbra Pump MAX™) |
| Flow Rate (60 Hz US) | 0-0.8 SCFM (0-23 LPM) (Same as predicate device Penumbra Pump MAX™) |
| Voltage | 110-115 Vac (Comparable to predicate device 100-115 Vac) |
| Frequency | 60 Hz (US) (Same as predicate device Penumbra Pump MAX™) |
| Duty Cycle | Non-continuous: 97% (58.2 minutes on, 1.8 minutes off) (Compared to predicate device 97.8% (45 minutes on, 1 minute off) - changed to match reference device Riptide™ Aspiration System) |
| Applied Part Classification | Type CF (Same as predicate device Penumbra Pump MAX™; change from Reference Device Type BF to match predicate) |
| Noise | <70 dBa (Compared to predicate device <60 dBa and reference device 60.5 dBa) |
| Materials of Construction (Collection Canister) | Commonly used medical grade plastics (Same as predicate device Penumbra Pump MAX™) |
| Volume (Collection Canister) | 1200 mL (Compared to predicate device 1000 mL - matched to reference device Riptide™ Aspiration System) |
| Shelf Life (Collection Canister) | 3 years (Compared to predicate device "Unknown" - matched to reference device Riptide™ Aspiration System) |
| Operating Life | 500 hours (Same as predicate device Penumbra Pump MAX™) |
| Biocompatibility | No impact to biocompatibility (due to no change in materials associated with the proposed changes) |
2. Sample Size Used for the Test Set and Data Provenance
The document indicates that the determination of substantial equivalence is based upon non-clinical bench testing. It does not specify a "test set" in the context of clinical data, as no clinical studies were performed.
- Sample Size for Test Set: Not applicable as no clinical test set was used. The performance evaluation was based on bench tests of the device itself.
- Data Provenance: The data provenance is from bench testing conducted by Micro Therapeutics Inc. d/b/a ev3 Neurovascular (the applicant). This is internal testing, not data from a specific country or clinical setting, and is prospective in nature (tests performed on the device to establish its performance).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. For this type of 510(k) submission based on non-clinical bench testing, "ground truth" is not established by medical experts for a test set of patient cases. Instead, compliance with established engineering and medical device standards (e.g., IEC, ISO, FDA guidance) serves as the "ground truth" for technical performance.
4. Adjudication Method for the Test Set
Not applicable. There was no clinical test set requiring expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the intended use, fundamental scientific technology, or materials of construction." Therefore, there is no effect size of human readers improving with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
While the device is an electromechanical pump (not an AI algorithm), the performance data presented is standalone performance of the device itself, evaluated through bench testing without human-in-the-loop performance as part of its primary performance evaluation for this submission. The "User Manual Design Validation" does involve human factors, but this is about the usability of the manual, not the device's functional performance itself.
7. The Type of Ground Truth Used
The "ground truth" for this submission is:
- Compliance with recognized international standards: IEC 60601-1, IEC 60601-1-2, ISO 14971, ANSI/AAMI HE75, ISO 10079-1, and ISO 10079-4.
- Equivalence to a legally marketed predicate device: Penumbra Pump MAX™ (K122756) in terms of fundamental technological characteristics and performance attributes (vacuum range, flow rate, materials, operating life).
- Equivalence to a reference device: Riptide™ Aspiration System (K172448) for certain updated characteristics like duty cycle, volume, and shelf life of the collection canister, where these characteristics are deemed equivalent to or improve upon the predicate without raising new safety/effectiveness concerns.
There is no pathology, outcomes data, or expert consensus on patient cases used as ground truth for this mechanical device submission.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device, an aspiration pump, not a machine learning or AI algorithm. Therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1