(129 days)
No
The device description focuses on mechanical components for aspiration and does not mention any software, algorithms, or data processing that would indicate the use of AI/ML. The performance studies are non-clinical bench and animal testing, not studies evaluating AI/ML performance metrics.
Yes
The device is intended for "revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease," which is a direct treatment for a disease condition.
No
The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. It is a treatment device designed to restore blood flow, not to diagnose a condition.
No
The device description explicitly lists multiple hardware components including a catheter, tubing, pump, and collection canister.
Based on the provided information, the Riptide™ Aspiration System is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use is to physically remove blood clots from blood vessels in patients with acute ischemic stroke. This is a therapeutic intervention performed in vivo (within the living body).
- Device Description: The device components are designed for mechanical aspiration within the vascular system. They are not used to analyze biological samples in vitro (outside the living body) to provide diagnostic information.
- Lack of Diagnostic Function: The system's purpose is to restore blood flow, not to diagnose a condition or provide information about a patient's health status based on the analysis of samples.
IVD devices are typically used to examine specimens such as blood, urine, or tissue to detect diseases, conditions, or infections. The Riptide™ Aspiration System does not perform this function.
N/A
Intended Use / Indications for Use
The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Product codes
NRY
Device Description
The Riptide™ Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. The Riptide™ Aspiration System is designed for use within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries. The Riptide™ Aspiration System is composed of the following components:
- Arc™ Catheter .
- Riptide™ Aspiration Tubing .
- Riptide™ Aspiration Pump ●
- Riptide™ Collection Canister with Intermediate Tubing .
The Arc™ Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. A lubricous, tapered liner is used to create a structure that has both proximal stiffness and distal flexibility. The Arc™ Catheter has a radiopaque marker band encapsulated at the distal tip for visualization under fluoroscopy. The Arc™ Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion. The Arc™ Catheter is the only component of the Riptide™ Aspiration System that is used intravascularly.
The Riptide™ Aspiration Tubing serves as a conduit to supply vacuum from the Riptide™ Aspiration Pump to the distal tip of the Arc™ Catheter. The Riptide™ Aspiration Tubing provides a connection between the sterile and non-sterile environments. The proximal end of the Riptide™ Aspiration Tubing is connected to the Riptide™ Collection Canister (outside of the sterile environment) while the distal end of the Riptide™ Aspiration Tubing is connected to the Arc™ Catheter (inside the sterile environment). The Riptide™ Collection Canister is connected to the Riptide™ Aspiration Pump (also outside of the sterile environment) via the Intermediate Tubing.
The Riptide™ Aspiration Pump is designed to generate vacuum for the Riptide™ Aspiration System. The vacuum pressure of the Riptide™ Aspiration Pump is set by turning the vacuum control valve until the vacuum gauge reads a minimum of 20inHg but not exceeding 25inHg. The Riptide™ Aspiration Pump is reusable, non-sterile, and intended to be utilized outside of the sterile environment.
The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single-use and the repository for aspirated material. The Riptide™ Collection Canister is placed into the receptacle of the Riptide™ Aspiration Pump while the Intermediate Tubing is connected to the vacuum inlet port.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries)
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance Data - Bench:
Non-clinical bench testing was performed for each component of the Riptide™ Aspiration System to evaluate the performance of the design modifications. Key results indicate that the Arc™ Catheter, Riptide™ Aspiration Tubing, Riptide™ Aspiration Pump, Riptide™ Canister with Intermediate Tubing, and the integrated Riptide™ Aspiration System all met their respective acceptance criteria for various tests including microbial, packaging, and performance characteristics (e.g., lumen patency, dimensional inspection, vacuum resistance, flow rate, electrical safety, volume capacity, simulated clot retrieval).
Performance Data - Animal:
Non-clinical animal testing was performed to evaluate the safety, efficacy, and usability of the Riptide™ Aspiration System in comparison to the Penumbra System® and Penumbra Pump MAX™ at acute and chronic time points in a porcine model, both in the presence and absence of simulated clot. This testing was performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies.
Performance Data - Clinical:
Not Applicable. A determination of substantial equivalence is based upon non-clinical bench and animal testing.
Conclusion:
The non-clinical bench and animal testing supports a determination that the subject Riptide™ Aspiration System is substantially equivalent to the predicate Penumbra System® and Penumbra Pump MAX™.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Penumbra System® and Penumbra Pump MAX™ 510(k)#: K160449
Reference Device(s)
Arc™ Intracranial Support Catheter 510(k)#: K150107, Penumbra System®/Penumbra System MAX™ 510(k)#: K133317, Penumbra Pump MAX™ 510(k)#: K122756, Penumbra Reperfusion Catheter 054 510(k)#: K090752
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
December 21, 2017
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Ryan Kenney Senior Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618
Re: K172448
Trade/Device Name: Riptide™ Aspiration System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: November 20, 2017 Received: November 21, 2017
Dear Mr. Ryan Kenney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K172448
Device Name Riptide™ Aspiration System
Indications for Use (Describe)
The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Type of Use (Select one or both, as applicable) | |
---|---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary: K172448
510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular | |
---|---|---|
9775 Toledo Way | ||
Irvine, CA 92618 | ||
Establishment Registration No.: 2029214 | ||
Contact Person: | Ryan Kenney | |
Senior Regulatory Affairs Specialist | ||
Telephone: (949) 297-5489 | ||
Email: ryan.j.kenney@medtronic.com | ||
Date Summary Prepared: | December 18, 2017 | |
Trade Name of Device: | Riptide™ Aspiration System | |
Common Name of Device: | Catheter, Thrombus Retriever | |
Regulation Description: | Percutaneous Catheter | |
Review Panel: | Neurology | |
Product Code: | NRY | |
Regulation Number: | 21 CFR 870.1250 | |
Device Classification: | Class II | |
Predicate Device(s): | Penumbra System® and Penumbra Pump MAX™ | |
510(k)#: K160449 | ||
Reference Device(s): | Arc™ Intracranial Support Catheter | |
510(k)#: K150107 | ||
Penumbra System®/Penumbra System MAX™ | ||
510(k)#: K133317 | ||
Penumbra Pump MAX™ | ||
510(k)#: K122756 | ||
Penumbra Reperfusion Catheter 054 | ||
510(k)#: K090752 |
Device Description:
The Riptide™ Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. The Riptide™ Aspiration System is designed for use within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries. The Riptide™ Aspiration System is composed of the following components:
- Arc™ Catheter .
- Riptide™ Aspiration Tubing .
- Riptide™ Aspiration Pump ●
- Riptide™ Collection Canister with Intermediate Tubing .
4
The Arc™ Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. A lubricous, tapered liner is used to create a structure that has both proximal stiffness and distal flexibility. The Arc™ Catheter has a radiopaque marker band encapsulated at the distal tip for visualization under fluoroscopy. The Arc™ Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion. The Arc™ Catheter is the only component of the Riptide™ Aspiration System that is used intravascularly.
The Riptide™ Aspiration Tubing serves as a conduit to supply vacuum from the Riptide™ Aspiration Pump to the distal tip of the Arc™ Catheter. The Riptide™ Aspiration Tubing provides a connection between the sterile and non-sterile environments. The proximal end of the Riptide™ Aspiration Tubing is connected to the Riptide™ Collection Canister (outside of the sterile environment) while the distal end of the Riptide™ Aspiration Tubing is connected to the Arc™ Catheter (inside the sterile environment). The Riptide™ Collection Canister is connected to the Riptide™ Aspiration Pump (also outside of the sterile environment) via the Intermediate Tubing.
The Riptide™ Aspiration Pump is designed to generate vacuum for the Riptide™ Aspiration System. The vacuum pressure of the Riptide™ Aspiration Pump is set by turning the vacuum control valve until the vacuum gauge reads a minimum of 20inHg but not exceeding 25inHg. The Riptide™ Aspiration Pump is reusable, non-sterile, and intended to be utilized outside of the sterile environment.
The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single-use and the repository for aspirated material. The Riptide™ Collection Canister is placed into the receptacle of the Riptide™ Aspiration Pump while the Intermediate Tubing is connected to the vacuum inlet port.
Indications for Use:
The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
| | Predicate Penumbra System® and
Penumbra Pump MAX™
(K160449) | Subject Riptide™ Aspiration System |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use
(IFU) Statement | Reperfusion Catheters and
Separators:
As part of the Penumbra System®,
the Reperfusion Catheters and
Separators are indicated for use in the
revascularization of patients with
acute ischemic stroke secondary to
intracranial large vessel occlusive
disease (within the internal carotid,
middle cerebral – M1 and M2 | The Riptide™ Aspiration System is
intended for use in the
revascularization of patients with acute
ischemic stroke secondary to
intracranial large vessel occlusive
disease (within the internal carotid,
middle cerebral – M1 and M2
segments, basilar, and vertebral
arteries) within 8 hours of symptom
onset. Patients who are ineligible for |
| | Predicate Penumbra System® and
Penumbra Pump MAX™
(K160449) | Subject Riptide™ Aspiration System |
| | segments, basilar, and vertebral
arteries) within 8 hours of symptom
onset. | intravenous tissue plasminogen
activator (IV t-PA) or who fail IV t-PA
therapy are candidates for treatment. |
| | Penumbra Aspiration Tubing:
As part of the Penumbra System®,
the Penumbra Sterile Aspiration
Tubing is indicated to connect the
Penumbra Reperfusion Catheters to
the Penumbra Pump MAX™.
Penumbra Pump MAX™:
The Penumbra Pump MAX™ is
indicated as a vacuum source for
Penumbra Aspiration Systems. | |
| | 5MAX™ ACE Reperfusion Catheter | Arc™ Catheter |
| Materials | Biocompatible, commonly utilized
for interventional devices. | Same |
| Coating | Hydrophilic | Same |
| Markerband | Radiopaque | Same |
| Guidewire | Yes | Same |
| Compatible | | |
| Dimensions | | |
| Usable Length | 132cm | 132-135cm |
| Proximal Inner
Diameter | 0.068" | 0.069" |
| Proximal Outer
Diameter | 0.083" | 0.0825" |
| Distal Inner
Diameter | 0.060" | 0.061" |
| Distal Outer
Diameter | 0.074" | 0.071" |
| Sterilization | | |
| Method | Ethylene Oxide (EO) | Same |
| Packaging | | |
| Pouch | Polyethylene | Nylon-Tyvek® |
| Packaging Hoop | Polyethylene | Same |
| Packaging Card | Polyethylene | Same |
| Carton | Solid Bleached Sulfate (SBS)
Paperboard | Same |
| Shelf Life | 36 Months | Same |
| | Penumbra MAX™ Aspiration Tubing | Riptide™ Aspiration Tubing |
| Materials | Biocompatible, commonly utilized | Same |
| | Predicate Penumbra System® and
Penumbra Pump MAX™
(K160449) | Subject Riptide™ Aspiration System |
| | for interventional devices. | |
| Dimensions | | |
| Usable Length | 112.0" | Same |
| Distal Length | 7.0" | Same |
| Inner Diameter | 0.088" | Same |
| Outer Diameter | 0.188" | Same |
| Sterilization | | |
| Method | Ethylene Oxide (EO) | Same |
| Packaging | | |
| Shelf Life | 36 Months | 3 Months |
| | Penumbra Pump MAX™ | Riptide™ Aspiration Pump |
| Performance Characteristics | | |
| Vacuum Range | 0-29 inHg (0-98.2 kPa) | Same |
| Flow Rate | 0-0.8 SCFM (0-23 LPM) | Same |
| Dimensions | | |
| Length | 15.5" | 16.1" |
| Depth | 11.2" | 13.2" |
| Height | 13.2" | 12.3" |
| Weight | 22.3lbs | 23.6lbs |
| Electrical Requirements | | |
| Voltage | 100-115 Vac | 110-115 Vac |
| Frequency | 50 Hz/60 Hz | 60 Hz |
| Duty Cycle | Non-continuous
97.8% (45 minutes on, 1 minute off) | Non-continuous
97% (58.2 minutes on, 1.8 minutes off) |
| IEC 60601-1
Compliance | Yes | Same |
| IEC 60601-1-2
Compliance | Yes | Same |
| | Penumbra MAX™ Canister | Riptide™ Collection Canister with
Intermediate Tubing |
| Dimensions | | |
| Volume | 1000ml | 1200ml |
Device Comparison:
5
6
Biocompatibility:
Biocompatibility was performed for the Arc™ Catheter and the Riptide™ Aspiration Tubing. The Arc™ Catheter is categorized as a limited exposure (in vitro Hemocompatibility
Assay | The Arc™ Catheter did not
result in a decrease in any
blood component as compared
to the reference material. These
results indicate that the cause of
thrombi is not related to the
materials exposed to human
blood during use | No adverse effect on leukocyte
or platelet counts. |
7
8
Riptide™ Aspiration Tubing | ||
---|---|---|
Test Description | Results | Conclusions |
ISO MEM Elution Using L-929 | ||
Mouse Fibroblast Cells | The test article scored '0' at 24, | |
48 and 72 ± 4 hours and is | ||
considered non-cytotoxic under | ||
the conditions of this test. | Non-cytotoxic. | |
ISO Guinea Pig Maximization | ||
Sensitization Test | Under the conditions of this | |
protocol, the test article did not | ||
elicit a sensitization response. | Non-sensitizer. | |
ISO Intracutaneous Irritation | ||
Test | The differences in the mean test | |
and control scores of the extract | ||
dermal observations were Microbial | ||
Ethylene Oxide (EO) Residual | EO Residual for the Arc™ Catheter is performed per ISO 10993-7. | The Arc™ Catheter met the acceptance criteria for EO Residual. |
Ethylene Chlorohydrin (ECH) | ECH for the Arc™ Catheter is performed per ISO 10993-7. | The Arc™ Catheter met the acceptance criteria for ECH. |
Bioburden | Bacterial Endotoxin for the Arc™ Catheter is performed per ANSI/AAMI ST72 and USP 161. | The Arc™ Catheter met the acceptance criteria for Bacterial Endotoxin. |
Packaging | ||
Arc™ Catheter | ||
Test | Test Method Summary | Results |
Visual Inspection | Visual Inspection for the Arc™ Catheter is performed per ASTM F1886. | The Arc™ Catheter met the acceptance criteria for Visual Inspection. |
Bubble Leak | Bubble Leak for the Arc™ Catheter is performed per ASTM F2096. | The Arc™ Catheter met the acceptance criteria for Bubble Leak. |
Seal Strength | Seal Strength for the Arc™ Catheter is performed per ASTM F88. | The Arc™ Catheter met the acceptance criteria for Seal Strength. |
Performance | ||
Lumen Patency | The proximal hub to the distal tip of the Arc™ Catheter must pass through a stainless-steel mandrel of required size. | The Arc™ Catheter met the acceptance criteria for Lumen Patency. |
Dimensional Inspection | The usable length, proximal and distal inner and outer diameter of the Arc™ Catheter is recorded. | The Arc™ Catheter met the acceptance criteria for Dimensional Inspection. |
Tip Buckling | The repeated distal tip buckling force under compressive load is evaluated for stiffness. | The Arc™ Catheter met the acceptance criteria for Tip Buckling. |
Injection Flow Rate | Injection Flow Rate for the Arc™ Catheter is performed per ISO 10555-1, Annex E with injection through the proximal hub. | The Arc™ Catheter met the acceptance criteria for Injection Flow Rate. |
Suction Flow Rate | Suction Flow Rate for the Arc™ Catheter is performed per ISO 10555-1, Annex E with injection through the distal tip. | The Arc™ Catheter met the acceptance criteria for Suction Flow Rate. |
Vacuum Resistance | Vacuum Resistance for the Arc™ Catheter is performed under static conditions using a 60cc syringe. | The Arc™ Catheter met the acceptance criteria for Vacuum Resistance. |
Air Aspiration Leak | Air Aspiration Leak for the Arc™ Catheter is performed per ISO 10555-1, Annex D. | The Arc™ Catheter met the acceptance criteria for Air Aspiration Leak. |
Hub/Shaft Peak Tensile Force | Hub/Shaft Peak Tensile Force for the Arc™ Catheter is performed per ISO 10555-1, Annex B. | The Arc™ Catheter met the acceptance criteria for Hub/Shaft Peak Tensile Force. |
Coating Integrity | Coating Integrity for the Arc™ Catheter was assessed using a staining method pre and post. | Characterization only. |
ArcTM Catheter | ||
Test | Test Method Summary | Results |
Coating Lubricity/Durability | simulation to estimated coating | |
coverage. | The ArcTM Catheter met the | |
acceptance criteria for Coating | ||
Lubricity/Durability. | ||
Particulate | Particulate for the ArcTM | |
Catheter was performed in a | ||
tortuous model per USP . | ArcTM Catheter met the | |
acceptance criteria for | ||
Particulate. | ||
Kink Resistance | Kink Resistance for the ArcTM | |
Catheter is performed with the | ||
device wrapped around a rod or | ||
known radius at which point | ||
the ArcTM Catheter is inspected | ||
for kinking. | The ArcTM Catheter met the | |
acceptance criteria for Kink | ||
Resistance. | ||
Liquid Leak | Liquid Leak for the ArcTM | |
Catheter is performed per ISO | ||
10555-1, Annex C. | The ArcTM Catheter met the | |
acceptance criteria for Liquid | ||
Leak. | ||
Static/Dynamic Burst | Static/Dynamic Burst for the | |
ArcTM Catheter is performed | ||
per ISO 10555-1, Annex F. | The ArcTM Catheter met the | |
acceptance criteria for | ||
Static/Dynamic Burst. | ||
Corrosion Resistance | Corrosion Resistance for the | |
ArcTM Catheter is performed | ||
per ISO 10555-1, Annex A. | The ArcTM Catheter met the | |
acceptance criteria for | ||
Corrosion Resistance. | ||
Torque to Failure | The ArcTM Catheter is | |
navigated through a tortuous | ||
model where the distal tip is | ||
secured, and the proximal end | ||
of the catheter is rotated 360°. | ||
The total number of 360° | ||
rotations prior to failure is | ||
recorded | Characterization only. |
10
11
The following non-clinical bench testing was performed for the Riptide™ Aspiration Tubing:
Riptide™ Aspiration Tubing | ||
---|---|---|
Test | Test Method Summary | Results |
Microbial | ||
Ethylene Oxide (EO) Residual | EO Residual for the Riptide™ Aspiration Tubing is performed per ISO 10993-7. | The Riptide™ Aspiration Tubing met the acceptance criteria for EO Residual. |
Ethylene Chlorohydrin (ECH) | ECH for the Riptide™ Aspiration Tubing is performed per ISO 10993-7. | The Riptide™ Aspiration Tubing met the acceptance criteria for ECH. |
Riptide™ Aspiration Tubing | ||
Test | Test Method Summary | Results |
Bioburden Recovery | Bioburden Recovery for the | |
Riptide™ Aspiration Tubing is | ||
performed per | ||
ANSI/AAMI/ISO 11737-1. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Bioburden | ||
Recovery. | ||
Bioburden | Bioburden for the Riptide™ | |
Aspiration Tubing is performed | ||
per ANSI/AAMI/ISO 11737-1. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Bioburden. | ||
Bacterial Endotoxin | Bacterial Endotoxin for the | |
Riptide™ Aspiration Tubing is | ||
performed per ANSI/AAMI | ||
ST72. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Bacterial Endotoxin. | ||
Packaging | ||
Aseptic Presentation | Aseptic Presentation for the | |
Riptide™ Aspiration Tubing is | ||
performed for delamination. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Aseptic | ||
Presentation. | ||
Seal Width | Seal Width for the Riptide™ | |
Aspiration Tubing is performed | ||
per ASTM F2203. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Seal Width. | ||
Dye Leak | Dye Leak for the Riptide™ | |
Aspiration Tubing is performed | ||
per ASTM F1929. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Dye Leak. | ||
Visual Inspection | Visual Inspection for the | |
Riptide™ Aspiration Tubing is | ||
performed per ASTM F1886. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Visual Inspection. | ||
Bubble Leak | Bubble Leak for the Riptide™ | |
Aspiration Tubing is performed | ||
per ASTM F2096. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Bubble Leak. | ||
Legibility | Legibility for the Riptide™ | |
Aspiration Tubing is performed | ||
for smearing or degradation. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Legibility. | ||
Foreign Material | Foreign Material on the | |
Riptide™ Aspiration Tubing is | ||
performed via visual | ||
inspection. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Foreign Material. | ||
Seal Strength | Seal Strength for the Riptide™ | |
Aspiration Tubing is performed | ||
per ASTM F88. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Seal Strength. | ||
Performance | ||
Dimensional Inspection | The overall and distal length, | |
proximal and distal inner and | ||
outer diameter of the Riptide™ | ||
Aspiration Tubing is recorded. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Dimensional | ||
Inspection | ||
Riptide™ Aspiration Tubing | ||
Test | Test Method Summary | Results |
Identification of Heat Shrink | ||
Bands | The heat shrink bands of the | |
Riptide™ Aspiration Tubing | ||
are visually inspected. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Identification of the | ||
Heat Shrink Bands. | ||
Identification of Flow Switch | The flow switch of the | |
Riptide™ Aspiration Tubing is | ||
visually inspected. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Identification of the | ||
Flow Switch. | ||
Joint Tensile Strength | The Riptide™ Aspiration | |
Tubing is affixed to upper and | ||
lower grips of the tensile test | ||
fixture and stretched. The | ||
maximum value before failure | ||
is recorded. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Joint Tensile | ||
Strength. | ||
Leak | The Riptide™ Aspiration | |
Tubing is pressurized to | ||
specified vacuum pressure and | ||
any decay in pressure is | ||
recorded | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Leak (Vacuum and | ||
Pressure Decay). | ||
Degree of Collapse | Degree of Collapse for the | |
Riptide™ Aspiration Tubing is | ||
performed per ISO 10079-3. | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Degree of Collapse. | ||
Compatibility | The distal end of the Riptide™ | |
Aspiration Tubing is connected | ||
to the proximal hub of the | ||
Arc™ Catheter. The proximal | ||
end of the Riptide™ Aspiration | ||
Tubing is connected to the | ||
horizontal port of the Riptide™ | ||
Collection Canister | The Riptide™ Aspiration | |
Tubing met the acceptance | ||
criteria for Compatibility to the | ||
Arc™ Catheter and Riptide™ | ||
Collection Canister. |
12
13
The following non-clinical bench testing was performed for the Riptide™ Aspiration Pump:
Riptide™ Aspiration Pump | ||
---|---|---|
Test | Test Method Summary | Results |
Degrees of Tilt | The test axes of the Riptide™ | |
Aspiration Pump are pivoted | ||
45° and does not tip over. | The Riptide™ Aspiration Pump | |
met the acceptance criteria for | ||
Degrees of Tilt. | ||
Durability | The Riptide™ Aspiration Pump | |
is functional through 500 hours | ||
of expected service life. | The Riptide™ Aspiration Pump | |
met the acceptance criteria for | ||
Durability. | ||
Maximum Vacuum | The vacuum control valve of | |
the Riptide™ Aspiration Pump | ||
is set to maximum vacuum and | The Riptide™ Aspiration Pump | |
met the acceptance criteria for | ||
Maximum Vacuum. |
14
Riptide™ Aspiration Pump | ||
---|---|---|
Test | Test Method Summary | Results |
recorded. | ||
Flow Capacity | The Riptide™ Aspiration Pump | |
was ran at specified pressure | ||
prior to recording the air flow. | The Riptide™ Aspiration Pump | |
met the acceptance criteria for | ||
Flow Capacity. | ||
Electrical Safety | Electrical safety for the | |
Riptide™ Aspiration Pump is | ||
performed per IEC 60601-1. | The Riptide™ Aspiration Pump | |
met the acceptance criteria for | ||
Electrical Safety | ||
Electromagnetic Compatibility | Electromagnetic compatibility | |
for the Riptide™ Aspiration | ||
Pump is performed per IEC | ||
60601-1-2. | The Riptide™ Aspiration Pump | |
met the acceptance criteria for | ||
Electromagnetic Compatibility. |
The following non-clinical bench testing was performed for the Riptide™ Canister with Intermediate Tubing:
Riptide™ Collection Canister with Intermediate Tubing | ||
---|---|---|
Test | Test Method Summary | Results |
Packaging | ||
Packaging Integrity | Packaging Integrity for the | |
Riptide™ Collection Canister | ||
with Intermediate Tubing is | ||
performed per ISTA 2A and | ||
ATSM D4169. | The Riptide™ Collection | |
Canister met the acceptance | ||
criteria for Packaging Integrity. | ||
Packaging Labeling | Packaging Labeling for the | |
Riptide™ Collection Canister | ||
with Intermediate Tubing is | ||
performed to ensure the | ||
labeling remains affixed and | ||
legible over the labeled shelf | ||
life of three (3) months. | The Riptide™ Collection | |
Canister met the acceptance | ||
criteria for Packaging Labeling. | ||
Performance | ||
Volume Capacity | The Riptide™ Collection | |
Canister is visually inspected to | ||
ensure graduations up to | ||
1200mL. | The Riptide™ Collection | |
Canister met the acceptance | ||
criteria for Volume Capacity. | ||
Volume Marking | The Riptide™ Collection | |
Canister is visually inspected to | ||
ensure graduations in 100mL. | The Riptide™ Collection | |
Canister met the acceptance | ||
criteria for Volume Marking. | ||
Canister Strength | The Riptide™ Collection | |
Canister is tested for canister | ||
strength per ISO 10079-3. | The Riptide™ Collection | |
Canister met the acceptance | ||
criteria for Canister Strength. | ||
Overfill Protection | The Riptide™ Collection | |
Canister is tested for overfill | The Riptide™ Collection | |
Canister met the acceptance |
15
Riptide™ Collection Canister with Intermediate Tubing | ||
---|---|---|
Test | Test Method Summary | Results |
Dimensional Inspection | protection per ISO 10079-3. | |
The overall length of the | ||
Intermediate Tubing is | ||
recorded. | criteria for Overfill Protection. | |
The Intermediate Tubing met | ||
the acceptance criteria for | ||
Dimensional Inspection. | ||
Degree of Collapse | Degree of Collapse for the | |
Intermediate Tubing is | ||
performed per ISO 10079-3. | The Intermediate Tubing met | |
the acceptance criteria for | ||
Degree of Collapse. | ||
Port Identification | The Riptide™ Collection | |
Canister is visually inspected to | ||
ensure the connection ports are | ||
correctly labeled. | The Riptide™ Collection | |
Canister met the acceptance | ||
criteria for Port Identification. | ||
Compatibility | The Intermediate Tubing is | |
connected to the vacuum inlet | ||
port of the Riptide™ Aspiration | ||
Pump. | The Intermediate Tubing met | |
the acceptance criteria for | ||
Compatibility. |
The following non-clinical bench testing was performed for the Riptide™ Aspiration System:
Riptide™ Aspiration System | ||
---|---|---|
Test | Test Method Summary | Results |
Simulated Clot Retrieval | The Arc™ Catheter is navigated through a tortuous model to the intended treatment location for simulated clot retrieval. | The Riptide™ Aspiration System met the acceptance criteria for Simulated Clot Retrieval. |
Vacuum Pressure | The vacuum pressure experienced at the distal tip of the Arc™ Catheter is recorded. | The Riptide™ Aspiration System met the acceptance criteria for Vacuum Pressure. |
Flow Rate | The flow rate through the Arc™ Catheter is recorded under full-flow conditions. | The Riptide™ Aspiration System met the acceptance criteria for Flow Rate. |
Lumen Collapse | The Arc™ Catheter is navigated through a tortuous model and subjected to maximum vacuum pressure generated by the Riptide™ Aspiration Pump at which point the catheter is inspected for any collapse or damage. | The Arc™ Catheter is resistant to Lumen Collapse. |
Usability | The Penumbra System® and Penumbra Pump MAX™ and the Riptide™ Aspiration System were evaluated for navigability, flexibility, the | The Riptide™ Aspiration System met the acceptance criteria for Usability. |
16
Riptide™ Aspiration System | ||
---|---|---|
Test | Test Method Summary | Results |
ability to engage and apply | ||
aspiration, and for the ability to | ||
safely retrieve clot by end | ||
users. |
The non-clinical bench testing, as described above, indicates that the Riptide™ Aspiration System is substantially equivalent to the Penumbra System® and Penumbra Pump MAXTM.
Performance Data - Animal:
Non-clinical animal testing was performed to evaluate the safety, efficacy, and usability of the Riptide™ Aspiration System in comparison to the Penumbra System® and Penumbra Pump MAX™ at acute and chronic time points in a porcine model, both in the presence and absence of simulated clot. Non-clinical animal testing was performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies.
The non-clinical animal testing, as described above, indicates that the Riptide™ Aspiration System is substantially equivalent to the Penumbra System® and Penumbra Pump MAXTM.
Performance Data - Clinical:
Not Applicable. A determination of substantial equivalence is based upon non-clinical bench and animal testing.
Conclusion:
The design modifications incorporated into the Riptide™ Aspiration System do not alter the fundamental scientific technology or intended use.
Non-clinical bench and animal testing supports a determination that the subject RiptideTM Aspiration System is substantially equivalent to the predicate Penumbra System® and Penumbra Pump MAX™.