Search Results
Found 13 results
510(k) Data Aggregation
(46 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) are disposable devices which made of PVC material ,intended for medical purpose that worn on examiner's hand or finger to prevent contamination between patient and examiner and they meets all of the requirements of ASTM standard D 5250-06 e1.
Here's a breakdown of the acceptance criteria and the study information for the device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 e1 | Meets |
Physical Properties | ASTM standard D 5250-06 e1 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 e1 and D6124-06 | Meets ( |
Ask a specific question about this device
(169 days)
This glove is disposable and intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Powder Free Nitrile Examination Gloves (Blue), non-sterile.
Here's an analysis of the provided information regarding the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
CHARACTERISTICS | Acceptance Criteria (Standards) | Reported Device Performance |
---|---|---|
Dimensions | ASTM D 6319-00a(2005) | Meets |
Physical Properties | ASTM D 6319-00a(2005) | Meets |
Freedom from Holes | ASTM D 6319-00a(2005) | |
ASTM D 5151-06 | Meets | |
Meets AQL 1.5 | ||
Powder Free Residual | ASTM 6124-06 | Less than 2 mg per glove |
Biocompatibility: Primary Skin Irritation Test in Rabbits | Consumer Product Safety Commission, Title 16, Chapter II, Part 1500.41 & 1500:3(C)(4) | Passes |
Biocompatibility: Guinea Pig Sensitization Test | ISO 10993-10: 2002(E), Dermal Sensitization Assay-Closed Patch Test | Passes |
Regarding the study proving the device meets the acceptance criteria, the provided document is a 510(k) summary for a medical device (Powder-Free Nitrile Examination Gloves). As such, it focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive clinical study report for an AI/imaging device. Therefore, many of the requested points below (sample size, experts, adjudication, MRMC, standalone performance, ground truth establishment) are not applicable in the context of this type of submission for examination gloves.
However, based on the information provided, we can infer the following about the testing and evidence:
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated. For physical property testing of gloves, sample sizes are typically defined by the specific ASTM standards (e.g., a certain number of gloves per batch for dimension, tensile strength, and freedom from holes testing).
- Data Provenance: Not explicitly stated, but based on the nature of manufacturing and testing for medical devices, it would be retrospective data collected during the production and quality control processes of the manufacturer (THAI HUA HOLDING COMPANY LIMITED). The country of origin of the data would likely be Thailand, where the manufacturer is located, or potentially the location of the testing laboratories.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
- Not Applicable. For physical and biocompatibility testing of gloves, "experts" in the sense of clinician reviewers establishing ground truth for disease detection are not relevant. The "ground truth" is objective measurement against specified standards. Testing would be performed by lab technicians or engineers trained in the relevant ASTM and ISO standards.
4. Adjudication Method for the Test Set:
- Not Applicable. As the "ground truth" relies on objective measurements against standards, an adjudication method for different human interpretations is not needed. Test results either meet the numerical criteria or they don't.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This device is an examination glove, not an AI or imaging device that would involve human readers or AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not Applicable. See point 5.
7. The type of ground truth used:
- Objective Measurements/Standards: The ground truth for this device's performance is established by the specified ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards, and Consumer Product Safety Commission regulations. For example:
- Dimensions and Physical Properties: Direct measurements of glove size, tensile strength, elongation, etc., compared against the numerical ranges defined in ASTM D 6319-00a(2005).
- Freedom from Holes: Water leak testing as per ASTM D 5151-06, yielding a quantified Acceptable Quality Level (AQL).
- Powder Free Residual: Measurement of residual powder, compared against the limit in ASTM 6124-06.
- Biocompatibility: Results from standardized animal (rabbit, guinea pig) tests for irritation and sensitization, evaluated against established toxicology protocols.
8. The Sample Size for the Training Set:
- Not Applicable. This is not an AI/machine learning device that requires a "training set." The testing methods are based on established engineering and materials science principles.
9. How the Ground Truth for the Training Set was Established:
- Not Applicable. See point 8.
Ask a specific question about this device
(127 days)
This glove is disposable and intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Powdered Latex Examination Gloves. non-sterile.
The provided text describes a 510(k) premarket notification for "Powdered Latex Examination Gloves" by THAI HUA HOLDING CO. LTD. It outlines the device's characteristics, intended use, and conformance to standards.
Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE |
---|---|---|
Dimensions | ASTM D 3578-05 | Meets |
Physical Properties | ASTM D 3578-05 | Meets |
Freedom from Holes | ASTM D 3578-05 | |
ASTM D 5151-06 | Meets | |
AQL 1.5 | ||
Residual Powder | ASTM 6124-06 | Less than 10 mg per dm² |
Protein Content | ASTM D 5712-99 | Less than 200 µg per dm² |
Biocompatibility (Irritation) | Primary Skin Irritation in Rabbits | |
CPSC Title 16, Chapter II, Part 1500:41 | Passes | |
Biocompatibility (Sensitization) | Guinea Pig Sensitizations | |
ISO 10993-10: 2002(E) | Passes |
The "DEVICE PERFORMANCE" column indicates that the gloves meet or pass the specified criteria, demonstrating compliance with the referenced ASTM, CPSC, and ISO standards.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size used for the test set (e.g., how many gloves were tested for dimensions, holes, etc.).
It also does not specify the data provenance (e.g., country of origin of the data, retrospective or prospective). The testing appears to be conducted by the manufacturer to demonstrate compliance with international standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable and therefore not provided in the document. The "ground truth" for this type of device (examination gloves) is established by adherence to predefined objective physical and chemical standards (e.g., ASTM, ISO), not by expert opinion or interpretation in a medical imaging or diagnostic context.
4. Adjudication Method for the Test Set
This information is not applicable and therefore not provided in the document. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation (e.g., medical image reading) where disagreements between experts need to be resolved. For objective physical and chemical tests, the results are quantifiable and do not require adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, and the impact of an AI algorithm on their performance is being evaluated. For medical gloves, the performance is assessed against objective technical standards, not through human reader interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This information is not applicable. The device is a physical product (examination gloves), not an algorithm or AI system. Therefore, the concept of "standalone algorithm performance" does not apply.
7. The Type of Ground Truth Used
The "ground truth" for this device is based on objective, verifiable physical and chemical standards as defined by:
- ASTM D 3578-05: Standard Specification for Rubber Examination Gloves (for dimensions, physical properties, freedom from holes).
- ASTM D 5151-06: Standard Test Method for Detection of Holes in Medical Gloves.
- ASTM 6124-06: Standard Test Method for Residual Powder on Medical Gloves.
- ASTM D 5712-99: Standard Test Method for the Measurement of Latex Allergens in Natural Rubber and Latex Products.
- CPSC Title 16, Chapter II, Part 1500:41: For Primary Skin Irritation in Rabbits.
- ISO 10993-10: 2002(E): Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (specifically for Guinea Pig Sensitizations).
These standards provide the definitive criteria against which the device's performance is measured.
8. The Sample Size for the Training Set
This information is not applicable. The device is a physical product, not an AI/ML model that requires a training set. The manufacturing process of gloves does not involve machine learning training sets.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable as there is no training set for this type of physical device.
Ask a specific question about this device
(150 days)
A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner against potentially infectious materials.
Powder-Free Vinyl Examination Glove
The provided documentation describes the acceptance criteria and a study demonstrating that the Cartagloves S.A. Powder-Free Vinyl Examination Glove meets these criteria, supporting its substantial equivalence claim.
1. Acceptance Criteria and Reported Device Performance
Parameter | Acceptance Criteria (Standard) | Reported Device Performance (Applicant Device Specification) |
---|---|---|
Dimensions | Meets ASTM D 5250-06 | |
Width (Small) | 85 mm | 85 +/- 5 mm |
Width (Medium) | 95 mm | 95 +/- 5 mm |
Width (Large) | 105 mm | 105 +/- 5 mm |
Width (X-Large) | 115 mm | 115 +/- 5 mm |
Length (all sizes) | 230 mm minimum | 250 +/- 10 mm |
Thickness (Finger) | 0.05 mm minimum | 0.05 mm minimum |
Thickness (Palm) | 0.08 mm minimum | 0.1 mm minimum |
Physical Properties | Meets ASTM D 5250-06 | |
Tensile Strength (Before Aging) | 9 MPa minimum | 9 MPa minimum |
Tensile Strength (After Aging) | 9 MPa minimum | 9 MPa minimum |
Ultimate Elongation (Before Aging) | 300% minimum | 300% minimum |
Ultimate Elongation (After Aging) | 300% minimum | 300% minimum |
Freedom from Pinholes | Meets ASTM D 5250-06 and ASTM D 5151-99 | AQL 2.5, Level I (Water Leak Test) |
Powder Free | Meets ASTM D 6124-06 and ASTM D 5250-06 | Adheres to standard requirements |
Biocompatibility | Based on ISO 10993 | |
Cytotoxicity - Agar Diffusion | N/A (Standard for "Passed") | Passed |
Primary Skin Irritation | N/A (Standard for "Passed") | Passed |
Repeated Patch Dermal Sensitization | N/A (Standard for "Passed") | Passed |
2. Sample size used for the test set and the data provenance
The documentation does not explicitly state the specific "sample size" for each test (e.g., number of gloves tested for dimensions, physical properties, or pinholes). However, it implies that the testing was conducted according to the sampling plans outlined within ASTM D5250, ASTM D5151, and ASTM D6124, which typically include specific AQL (Acceptable Quality Level) and inspection levels for quality control.
The data provenance is not explicitly stated in terms of country of origin but is from the applicant, Cartagloves S.A., located in Cartagena, Colombia. The testing appears to be internal to the manufacturer or conducted by a designated lab on their behalf. The data is presented as retrospective performance data collected to support the 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to this device. The device is a patient examination glove, and its performance is evaluated against established physical, chemical, and biocompatibility standards (ASTM and ISO), not subjective expert interpretation of images or clinical data. There is no concept of "ground truth" established by experts in the context of these types of engineering and material performance tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable to this device. Adjudication methods like 2+1 or 3+1 are used in studies where subjective interpretation (e.g., medical image reading) requires consensus among multiple evaluators. For the performance testing of examination gloves, the measurements are objective (e.g., tensile strength, dimensions, pinhole count) and determined by standardized test procedures, not expert adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable to this device. An MRMC study with AI assistance is relevant for diagnostic devices that analyze medical images or data. The Cartagloves S.A. device is an examination glove, which is a physical barrier device, and does not involve human readers or AI in its intended use or performance evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable to this device. "Standalone performance" typically refers to the performance of an AI algorithm or an automated diagnostic system operating without human interaction. This is not relevant for a physical medical device like an examination glove.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance of the examination glove is defined by adherence to published and accepted industry standards:
- ASTM D 5250-06 ("Standard Specification for Poly(vinyl chloride) Gloves for Medical Application")
- ASTM D 5151-99 (Standard Test Method for Detection of Holes in Medical Gloves)
- ASTM D 6124-06 (Standard Test Method for Residual Powder on Medical Gloves)
- ISO 10993 (Biological evaluation of medical devices)
The device's performance against these quantifiable criteria (e.g., specific dimensions, tensile strength values, AQL levels for pinholes) constitutes the "ground truth" for its safety and effectiveness relative to the specified standards.
8. The sample size for the training set
This question is not applicable to this device. "Training set" refers to data used to train machine learning algorithms. The Cartagloves S.A. device is an examination glove, and its manufacturing and testing do not involve machine learning or AI.
9. How the ground truth for the training set was established
This question is not applicable to this device, as there is no "training set" in the context of examination glove manufacturing and testing.
Ask a specific question about this device
(69 days)
A medical glove is worn on the hand of healthcare and similar personnel to prevent contamination between health care personnel and the patient.
Not Found
This document is a 510(k) premarket notification for patient examination gloves. It approves the device as substantially equivalent to a predicate device. The information provided does not contain details about specific acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML powered device, which is implied by the detailed questions.
Instead, this document pertains to a physical medical device (examination gloves) and its approval process by the FDA based on substantial equivalence. Therefore, I cannot directly answer your questions about acceptance criteria, study details, sample sizes, ground truth establishment, or multi-reader multi-case studies, as they are not applicable to the content of this document.
The document indicates:
- Device Name: Omni Flex Nitrile Examination Glove, Powder Free, Turquoise
- Intended Use: "A medical glove is worn on the hand of healthcare and similar personnel to prevent contamination between health care personnel and the patient."
- Regulatory Class: I
- Product Code: LZA
- FDA Action: Substantial Equivalence determination based on comparison to pre-amendment devices, subject to general controls (annual registration, listing, GMP, labeling, prohibitions against misbranding/adulteration).
To answer your specific questions in the context of an AI/ML powered medical device, I would need a different type of document that describes a performance study for such a device.
Ask a specific question about this device
(69 days)
A medical glove is worn on the hand of healthcare and similar personnel to prevent contamination between health care personnel and the patient.
Patient Examination Glove, Nitrile, Powdered, Turquoise
This submission is for a medical device called an "Omni Flex Nitrile Examination Glove, Powdered, Turquoise." The provided documents are a 510(k) clearance letter and an "Indications For Use Statement." These documents do not contain information about acceptance criteria or performance studies as typically required for clinical software or diagnostic devices.
Instead, they address the regulatory clearance of a physical medical device (gloves) based on substantial equivalence to a predicate device already on the market. Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, ground truth, and expert involvement is not applicable or available in these documents.
Here's an attempt to address the points based only on the provided text, recognizing its limitations for this type of query:
1. A table of acceptance criteria and the reported device performance
This information is not provided in the given documents. The 510(k) clearance process for general-purpose medical gloves typically involves demonstrating substantial equivalence to a legally marketed predicate device through performance criteria such as:
- Physical Properties: Tensile strength, elongation at break, force at break (often specified in ASTM D6319 or similar standards).
- Barrier Integrity: Water leak test (AQL, Acceptable Quality Level, often 2.5 or 4.0 for medical gloves).
- Biocompatibility: Tests for skin irritation and sensitization.
- Powder Residue (if powdered).
However, the specific quantitative acceptance criteria and the Omni Flex Nitrile Examination Glove's performance against these criteria are not detailed in the provided regulatory letter or indications for use. The letter simply states that the FDA has determined the device is "substantially equivalent" to predicate devices.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the given documents. For a 510(k) submission for gloves, the "test set" would refer to the samples of gloves tested for physical and barrier properties. The letter indicates the manufacturer is OmniGrace (Thailand) LTD., located in Thailand, suggesting the testing data would likely originate from Thailand or a designated testing facility. The nature of these tests is usually prospective, as manufactured batches are tested.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable to this type of device and submission. "Ground truth" in the context of diagnostic devices often refers to an expert-validated diagnosis or pathology result. For medical gloves, the "ground truth" is typically defined by adherence to established material standards and performance specifications, not expert consensus on an image or diagnosis. Quality control personnel and material science engineers would be involved in testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable to this type of device and submission. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations (e.g., in reading medical images), which is not relevant to testing the physical properties of gloves.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable to this type of device and submission. MRMC studies are used for evaluating the performance of diagnostic imaging devices or AI algorithms that assist human readers. Medical gloves are barrier devices, not diagnostic or AI-assisted tools.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable to this type of device and submission. This refers to the evaluation of AI algorithms, which is not relevant to medical gloves.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
As mentioned in point 3, the concept of "ground truth" in the diagnostic or AI sense does not apply here. For medical gloves, the "truth" is established by measured physical properties (e.g., tensile strength, barrier integrity) meeting predefined standards (e.g., ASTM standards, AQL limits) and biocompatibility test results.
8. The sample size for the training set
This is not applicable to this type of device and submission. Training sets are used in machine learning.
9. How the ground truth for the training set was established
This is not applicable to this type of device and submission.
Ask a specific question about this device
(28 days)
Ask a specific question about this device
(140 days)
Ask a specific question about this device
(114 days)
Ask a specific question about this device
(283 days)
Ask a specific question about this device
Page 1 of 2