Search Results
Found 26 results
510(k) Data Aggregation
(248 days)
UPJ Occlusion Balloon Catheter
Used to temporarily occlude the ureteropelvic junction to prevents from entering the ureter during percutaneous lithotripsy, and for injection of contrast medium.
The UPJ Occlusion Balloon Catheter is composed of a balloon catheter, a removable inflation/injection adapter, a 1 mL syringe, and a wire guide assembly. The UPJ Occlusion Balloon Catheter is manufactured from a double lumen polyurethane radiopaque tubing with a working length of 75 centimeters. The smaller lumen is designed for balloon inflation, while the larger lumen fits over the wire guide assembly. The distal balloon is manufactured from silicone and has an approximate inflation volume of 4 mL. The inflation/injector adapter is removable, allowing for a cystoscope to be removed after the catheter is in place. The wire guide assembly is manufactured in 0.028- or 0.038-inch diameter stainless steel coils. The stainless-steel coils are coated with polytetrafluoroethylene and measure 80 centimeters. The wire guide assembly is designed to be placed through the small lumen of the balloon catheter. The UPJ Occlusion Balloon Catheter is supplied sterile, intended for one-time use only, and packaged in a peel-open pouch with a shelf-life of 3 years.
This document describes the Cook Incorporated UPJ Occlusion Balloon Catheter (K183323). This is a physical medical device, not an AI/ML powered device. As such, the information you requested about acceptance criteria and studies is not applicable in the context of an AI/ML device.
Here's a breakdown based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not provide a table with specific acceptance criteria and reported device performance in the manner typically seen for AI/ML models (e.g., sensitivity, specificity, AUC). Instead, it lists performance data in the form of engineering and material testing.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Biocompatibility | Met requirements |
Tensile Strength | Met requirements |
Kink Radius | Met requirements |
Leakage and Lumen Blockage | Met requirements |
Radiopacity | Met requirements |
Balloon Fatigue and Burst | Met requirements |
Dimensional Measurement and Compatibility | Met requirements |
2. Sample size used for the test set and the data provenance:
This information is not explicitly provided. The "testing" mentioned is for a physical device, and typically involves a certain number of manufactured units or material samples. The document does not specify quantities or data provenance (e.g., country of origin, retrospective/prospective) because these concepts are not applicable to the engineering tests described.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable. For physical devices, "ground truth" is established through engineering and material standards, not expert consensus on medical images or clinical data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable as the testing involves physical device properties, not diagnostic interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable as the device is a physical catheter, not an AI/ML diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable as the device is a physical catheter, not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this physical device is based on established engineering and material standards, and compliance with the device's design specifications and intended use. This is demonstrated through various physical and material performance tests.
8. The sample size for the training set:
This information is not applicable as the device is a physical catheter and does not involve an AI/ML training set.
9. How the ground truth for the training set was established:
This information is not applicable as the device is a physical catheter and does not involve an AI/ML training set.
Ask a specific question about this device
(278 days)
Scepter Mini Occlusion Balloon Catheter
The Scepter Mini Occlusion Balloon Catheter is intended:
For use in the peripheral and neuro vasculature where temporary occlusion is desired. The balloon catheter provides temporary vascular occlusion which is useful in selectively stopping or controlling blood flow. The balloon catheter also offers balloon assisted embolization of intracranial aneurysms.
For use in the peripheral vasculature for the delivery of diagnostic agents, such as contrast media, that have been approved or cleared for use in the peripheral vasculature and are compatible with the inner lumen of the Scepter Mini Occlusion Balloon Catheter.
For neurovascular use for the delivery of diagnostic agents, such as contrast media, and liquid embolic agents that have been approved or cleared for use in the neurovasculature and are compatible with the inner lumen of the Scepter Mini Occlusion Balloon Catheter.
Scepter Mini Occlusion Balloon Catheter is a dual co-axial lumen balloon catheter. The catheter is designed to track over a steerable guidewire. The outer lumen is used for the inflation of the balloon independent of guidewire position. The inner lumen can be used to deliver diagnostic agents or liquid embolics to distal locations in tortuous anatomy. Radiopaque marker bands are located at each end of the balloon to facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the microcatheter hub is used for the attachment of accessories. The catheter is packaged sterile for single use only.
The Scepter Mini has similar indications for use as the predicates, however, incorporates several minor design differences. The Scepter Mini has a slightly longer length and a slightly smaller diameter. The balloon of the Scepter Mini is slightly shorter. The distal tip of the Scepter Mini extends a shorter distance from the distal end of the balloon than that of the Scepter C and XC. For the Scepter Mini, the purge hole is covered by a semi-permeable membrane designed to allow air to escape while preventing liquids from passing. The predicate Scepter C and XC incorporate 3 radiopaque marker bands, while the design of the Scepter Mini allows for visualization under fluoroscopy with only 2 radiopaque marker bands (due to shorter distal tip segment). All Scepter catheters have a hydrophilic coating.
This document is a 510(k) Pre-market Notification for the Scepter Mini Occlusion Balloon Catheter, seeking to demonstrate its substantial equivalence to previously cleared predicate devices. The study detailed is a set of verification and validation tests, and a summary of an animal study, used to support this claim.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document provides a comprehensive table of "Verification Test Summary" where each test description implicitly defines the acceptance criteria by stating the desired outcome (e.g., "does not break," "does not leak," "pass") and then reports that the subject device "Pass"ed. For many tests, it explicitly states that the device "meets the same specification as predicates Scepter C/XC (K121785)."
Test Description | Acceptance Criteria (Implied / Explicit) | Reported Device Performance |
---|---|---|
Sterility | Meets sterility assurance level (SAL 10-6) per ISO 11135-1. EtO and ECH residual levels are ≤ 0.2 mg per device per ISO 10993-7. Bacterial endotoxin test results . | Pass |
Physical Attributes | Meets design specifications. | Pass |
Force at break | Does not break during use and meets same specification as predicates Scepter C/XC (K121785). | Pass |
Freedom from Leakage – Fluids (low pressure, long duration) | Does not leak fluids at low pressure/long duration and meets same specification as predicates Scepter C/XC (K121785). | Pass |
Freedom from Leakage – Air | Air does not leak into subject device meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Freedom from Leakage - Liquid (high pressure, short time) | Does not leak fluids at high pressure/short duration and meets same specification as predicates Scepter C/XC (K121785). | Pass |
Burst Pressure of Catheter | Does not burst statically below rated burst pressure meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Gauging Test | Catheter luer compatible to other standard luer fittings. (Data leveraged from predicate Scepter C (K110741, K121785) due to the same hub design). | Pass |
Separation Force | Catheter luer compatible to other standard luer fittings. (Data leveraged from predicate Scepter C (K110741, K121785) due to the same hub design). | Pass |
Unscrewing Torque | Catheter luer compatible to other standard luer fittings. (Data leveraged from predicate Scepter C (K110741, K121785) due to the same hub design). | Pass |
Ease of Assembly | Subject device luer mates together with other compatible fittings. | Pass |
Resistance to Overriding | Catheter luer mates with other compatible fittings. (Data leveraged from predicate Scepter C (K110741, K121785) due to the same hub design). | Pass |
Stress Cracking | Catheter hub does not leak. (Data leveraged from predicate Scepter C (K110741, K121785) due to the same hub design). | Pass |
Durability of Hydrophilic Coating | Hydrophilic coating does not flake off during use, is of the same material and meets same specification as predicates Scepter C/XC (K121785). | Pass |
Lubricity of Hydrophilic Coating | Hydrophilic coating is lubricious, of the same material and meets same specification as predicates Scepter C/XC (K121785). | Pass |
Simulated Use | Demonstrated equivalent performance during simulated use with similar ratings to predicates Scepter C/XC (K121785). | Pass |
Compatibility with device/agents: Embolic material, contrast media, dimethyl sulfoxide (DMSO). | Subject device is compatible with embolic material, contrast media, and DMSO, meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Dynamic Burst Pressure | Does not burst dynamically below rated burst pressure and meets same specification as predicates Scepter C/XC (K121785). | Pass |
Radio-Detectability | Radiopaque marker bands are visualized under fluoroscopy. | Pass |
Kink resistance | Does not kink during normal use meeting same specification as predicates Scepter C/XC. | Pass |
Non-pyrogenic | Bacterial endotoxins level is less than 2.15 EU/device. | Pass |
Simulated Shipping and Packaging Testing | No defects that compromise integrity of package, meets seal strength, creep to burst, and dye penetration specification. | Pass |
Catheter Flexural Fatigue | Meets same specification as predicates Scepter C/XC (K121785) for flexural fatigue, pressure integrity, and hoop stress. | Pass |
In Vivo Testing (Histopathology evaluation) | No significant differences between the Scepter C balloon catheter and the control catheter in categories of performance and histopathologic evaluation. No denudation, perforation, dissection, or clinicant injury to the target vasculature. | Pass |
Balloon Rated Burst Volume | Does not burst during use meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Balloon Compliance (rated volume) | Consistently inflates to the desired OD meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Balloon Inflation/Deflation Times | Inflates and deflates within an acceptable time range meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Balloon Fatigue Test | Does not burst before acceptable minimum cycle(s) meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Torque Test | Maintains acceptable torque during use meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Packaging and Shelf Life | Sterile barrier is maintained during shelf life of product. | Pass |
Insertion tool performance: Ease to enter RHV. | Ease to enter RHV rated 3 or higher in tested category meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Decay Test | Balloon maintains rated burst OD for a minimum of 30 min meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Surface Contamination | No Contamination (no uncured coating, particulate greater than 0.02 mm², sharp edges, and embedded particulate). | Pass |
Corrosion Resistance | Metallic components show no signs of corrosion. (Data leveraged from reference device Headway 17 (K083343)). | Pass |
Catheter Particle Testing | Less than 25 particles greater than 10 microns and less than 3 particles greater than 25 microns per 1mL meeting same specification as predicates Scepter C/XC (K121785). | Pass |
Cytotoxicity - Medium Eluate Method | Non-cytotoxic (between no cell lysis (grade 0) to slight reactivity (grade 1)). | Non-cytotoxic |
Sensitization: Maximization Test in Guinea Pigs | Non-sensitizer (no irritation present on test or negative control guinea pigs). | Non-sensitizer |
Intracutaneous Reactivity | Non-irritating (no evidence of irritation (score 0.0)). | Non-irritating |
Systemic Injection Test in Mice | Systemically non-toxic (no weight loss, mortality, or evidence of systemic toxicity). | Systemically non-toxic |
Rabbit Pyrogen Test | Nonpyrogenic (rise of rabbit temperatures during three hours of observation did not exceed 0.5 ℃). | Nonpyrogenic |
ASTM Blood Compatibility - Direct and Indirect Contact Hemolysis | Non-hemolytic (0.59% hemolysis in direct contact and 1.25% hemolysis in indirect contact). | Non-hemolytic |
Unactivated Partial Thromboplastin Time Test | No effect on clotting (average clotting time of the test article showed no significant difference from the control). | No effect on clotting |
Complement Activation | Non-activated (plasma exposed to test article for 90 minutes exhibited no statistically significant increase in SC5b-9). | Non-activated |
In Vitro Hemocompatibility Test - Human Blood, Direct Contact | No effect on hematological parameters (concentration of White Blood Cells (WBC) and Platelets in human blood exposed to the test article was not statistically significantly decreased). | No effect on hematological parameters |
Salmonella thvpimurium and Escherichia coli Reverse Mutation Assay | Non-mutagenic (test article extracts did not induce a statistically significant increase in the number of revertant colonies). | Non-mutagenic |
Mouse Lymphoma Mutagenesis Assay | Non-mutagenic (increased mutant frequency (IMF) of the cells exposed to the test article extracts was less than the Global Evaluation Factor (GEF) 126 x 10^-6). | Non-mutagenic |
Rodent Blood Micronucleus Assay | No clastogenic effect (test article did not result in a statistically significant increase in the percentage of reticulocytes containing micronuclei). | No clastogenic effect |
2. Sample Size Used for the Test Set and the Data Provenance:
- Verification Tests: The document doesn't explicitly state the sample sizes for each individual mechanical or physical verification test. However, it indicates these tests were performed on the "subject device" (Scepter Mini Occlusion Balloon Catheter). The provenance of this data is MicroVention, Inc.'s internal testing/laboratories. This data is likely prospective, generated specifically for this submission.
- Animal Study (In Vivo Testing): The animal study used for the "In Vivo Testing" was performed using the predicate device Scepter C in an "acute swine animal model." The exact number of swine used is not explicitly stated in this summary, but it implies multiple animals ("compared with a commercially equivalent device"). The data provenance is from this predicate device study. This would be prospective data for the predicate, but retrospective in its application to the subject device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
- The document does not mention expert involvement in establishing ground truth for the mechanical, physical, or biocompatibility tests. These are objective, quantitative tests with defined pass/fail criteria.
- For the animal study, the "histopathologic evaluation" implies assessment by a qualified pathologist. However, the exact number of experts or their specific qualifications (e.g., "veterinary pathologist with X years of experience") are not specified in this summary.
4. Adjudication Method for the Test Set:
- For the technical verification tests, the results are objective (numerical values meeting specifications, or physical observations like "no leak," "no break"). An adjudication method is not applicable in the sense of expert consensus on ambiguous findings, as the tests produce clear pass/fail outcomes against predefined criteria.
- For the animal study's histopathologic evaluation, it's not specified if multiple pathologists reviewed slides or if an adjudication method was used for discrepancies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was performed. This submission is for a physical medical device (balloon catheter), not an artificial intelligence (AI) or software as a medical device (SaMD) that typically relies on human reader performance. Therefore, there is no AI component, and no effect size regarding human reader improvement with AI assistance is relevant or reported.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Does not apply. This is a physical medical device; there is no AI algorithm to evaluate.
7. The Type of Ground Truth Used:
- Verification Tests (Mechanical, Physical, Chemical): The ground truth for these tests is based on engineering specifications, industry standards (e.g., ISO), and regulatory requirements (e.g., USP). These are objective, measurable outcomes.
- Biocompatibility Tests: Ground truth is established based on pre-defined biological responses (e.g., lack of cytotoxicity, non-irritating, non-pyrogenic) as per ISO 10993-1:2009 and FDA guidelines.
- Animal Study (In Vivo Testing): The ground truth for the efficacy and safety during the in-vivo performance characteristics and histopathology evaluation is based on direct observation in the animal model and subsequent pathological examination of tissues for injury or adverse events. This could be considered pathology and in-vivo observational data.
8. The Sample Size for the Training Set:
- Does not apply. This is a physical medical device, not a machine learning model; therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established:
- Does not apply. There is no training set for this type of device submission.
Ask a specific question about this device
(117 days)
Occlusion Balloon Catheter
The Occlusion Balloon Catheter is indicated for temporary occlusion of large vessels, including the superior vena cava, in applications including perioperative occlusion and emergency control of hemorrhage.
The Occlusion Balloon Catheter is a multi-lumen catheter which has a compliant polyurethane balloon with a maximum diameter of 32mm at 60cc inflation volume. The balloon has a nominal length of 80mm. The device is constructed with a blended PEBA shaft and is available in two models; one model is compatible with 8Fr (or larger) introducer sheaths and one model is compatible with 10Fr (or larger) introducer sheaths. Both models provide the same balloon performance with respect to sizing and occlusion. The two models are intended for individual physician preference for patient-specific sheath selection. The device has an effective length of 90cm and is compatible with 0.035" diameter guidewires. Three (3) platinum-iridium radiopaque marker bands are placed on the shaft to facilitate balloon placement in the anatomy prior to inflation. The proximal end of the catheter has an integral PEBA bifurcation manifold with female luer ports to allow communication with the balloon inflation lumen and guidewire lumen. A PVC extension tube (with stopcock) is connected to the balloon inflation port to facilitate handling. The device is a single use, sterile device.
The provided text describes a 510(k) premarket notification for an Occlusion Balloon Catheter. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria and reporting its performance.
Therefore, many of the requested items (e.g., acceptance criteria table, sample size for test set, number of experts for ground truth, MRMC study, training set details) are not applicable and cannot be extracted from this document because the submission method relies on demonstrating equivalence through non-clinical testing and comparison to predicate devices, not through a clinical trial with specific performance metrics against acceptance criteria.
Here's what can be extracted based on the provided text:
1. A table of acceptance criteria and the reported device performance
- Not Applicable. The document does not provide specific quantitative acceptance criteria or reported performance metrics against such criteria in the context of a clinical study. It lists a series of non-clinical tests performed to ensure the device meets established performance criteria and will perform as intended for substantial equivalence.
Acceptance Criteria | Reported Device Performance |
---|---|
Not provided for clinically relevant performance metrics. Non-clinical tests were performed to ensure substantial equivalence and intended performance. | Not provided for clinically relevant performance metrics. The non-clinical tests included items like visual/dimensional inspections, vessel occlusion, balloon fatigue, kink resistance, burst/leak volume, tensile strength, torque strength, and shelf life. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. No clinical test set or data provenance is detailed, as clinical evaluation was not required. The tests performed were non-clinical (mechanical, performance, biocompatibility).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. No ground truth for a test set was established using experts, as no clinical study requiring such ground truth was conducted.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. No adjudication method for a test set is mentioned, as no clinical study requiring this was conducted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. No MRMC study was done, nor does the device involve AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This device is a physical medical catheter, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable. No ground truth of this nature was used, as clinical studies involving such data were not performed. The "ground truth" for this submission was based on successful completion of non-clinical performance and mechanical tests to demonstrate substantial equivalence to the predicate device.
8. The sample size for the training set
- Not Applicable. No training set is mentioned, as this is a medical device submission based on substantial equivalence and non-clinical testing, not a machine learning model.
9. How the ground truth for the training set was established
- Not Applicable. No training set or ground truth for it was established.
Ask a specific question about this device
(159 days)
GORE Molding and Occlusion Balloon Catheter
The GORE® Molding and Occlusion Balloon Catheter is intended for temporary occlusion of large vessels or to assist in the expansion of selfexpanding endovascular prostheses (stent grafts).
The GORE® Molding and Occlusion Balloon Catheter is a sterile (EtO), single use, single-lobed polyurethane balloon catheter. The compliant polyurethane balloon is mounted on the leading end of a 3 lumen catheter shaft (two inflation lumens and one quidewire lumen). Termination of the leading end of the catheter is an atraumatic catheter leading tip for smooth transition from the quidewire to catheter transition. Radiopaque markers (approximately 40 mm apart) indicate the proximal and distal end of the balloon which aid in proper balloon placement under fluoroscopy. Both of the (2) inflation lumens are in communication with each end of the balloon to facilitate balloon catheter preparation, inflation and deflation. At the trailing end of the balloon catheter is a dual port (balloon inflation and quidewire) yarm. The inflation port of the y-arm is in communication with both of the balloon inflation lumens and is affixed with a luer lock and three way stopcock via an extension tube. The quidewire lumen of the v-arm allows introduction of a 0.035" (0.89 mm) diameter guidewire for over-the-wire access. The trailing end of the quidewire lumen is affixed with a flushing / guidewire port with luer lock used for flushing the guidewire lumen. The balloon catheter proximal y-arm is provided within a housing which contains markings of the balloon length and inflation range diameter. The balloon can be inflated to diameters of 10 mm to a maximum inflation diameter of 37 mm. The balloon catheter profile is 10 Fr introducer sheath compatible.
Here's an analysis of the provided text regarding the acceptance criteria and study for the GORE Molding and Occlusion Balloon Catheter:
1. Table of Acceptance Criteria and Reported Device Performance:
The document describes a series of "Design Verification Testing" activities as the performance data. While specific pass/fail criteria (e.g., a numerical range for "Inflation Time") are not explicitly listed in this summary, the document states that these tests "demonstrated the subject device performed as intended and was substantially equivalent to the predicate device." This implies that the device met internal specifications and expectations derived from the predicate device's performance.
Acceptance Criterion (Test Name) | Reported Device Performance |
---|---|
Device Profile / Sheath Compatibility | Performed as intended, substantially equivalent to predicate device. |
Guidewire Compatibility | Performed as intended, substantially equivalent to predicate device. |
Repeat Inflation / Deflation | Performed as intended, substantially equivalent to predicate device. |
Balloon & Marker Band Position & Length | Performed as intended, substantially equivalent to predicate device. |
Inflation Time | Performed as intended, substantially equivalent to predicate device. |
Catheter Effective Length | Performed as intended, substantially equivalent to predicate device. |
Deflation Time | Performed as intended, substantially equivalent to predicate device. |
Visual Inspection Mechanical Defects | Performed as intended, substantially equivalent to predicate device. |
Occlusion | Performed as intended, substantially equivalent to predicate device. |
Tip Durability | Performed as intended, substantially equivalent to predicate device. |
Balloon Inflation Diameter | Performed as intended, substantially equivalent to predicate device. |
Tensile Strength | Performed as intended, substantially equivalent to predicate device. |
Burst Volume | Performed as intended, substantially equivalent to predicate device. |
Shelf Life Testing | Performed as intended, substantially equivalent to predicate device. |
Leakage | Performed as intended, substantially equivalent to predicate device. |
Packaging Validation Testing | Performed as intended, substantially equivalent to predicate device. |
Aspiration | Performed as intended, substantially equivalent to predicate device. |
Sterilization Validation Testing | Performed as intended, substantially equivalent to predicate device. |
Accessory Compatibility | Performed as intended, substantially equivalent to predicate device. |
Biocompatibility | Performed as intended, substantially equivalent to predicate device. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not specify the exact sample sizes for each of the "Design Verification Testing" activities. It only references "bench study" data.
- Sample Size: Not specified.
- Data Provenance: The studies were "bench studies," which are laboratory-based tests. The country of origin is not explicitly stated, but the company (W.L. Gore & Associates) is based in the U.S. (Flagstaff, Arizona), suggesting the studies were likely conducted in the U.S. These are retrospective tests performed to verify design specifications.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. This device is a physical medical device, and the "ground truth" for its performance is established through objective engineering and materials testing, not through expert human interpretation of data like in an AI/imaging device.
4. Adjudication Method for the Test Set:
Not applicable, as the tests are objective bench tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:
Not applicable. This is a physical medical device, not an AI or imaging diagnostic tool that would involve human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Not applicable. This is a physical medical device.
7. The Type of Ground Truth Used:
The ground truth used for this device's performance evaluation is based on objective engineering specifications, material properties, and functional requirements for a balloon catheter intended for occlusion and stent graft expansion. These are derived from established medical device standards and comparison to a predicate device. For example, "Burst Volume" would have a defined expected value, and the tested device's performance would be compared against that.
8. The Sample Size for the Training Set:
Not applicable. There is no concept of a "training set" for physical medical device bench testing in the same way there would be for an AI algorithm. The design and manufacturing processes are refined based on general engineering principles and iterative development, not a specific training dataset.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set in the context of this device. The "ground truth" for the device's design and manufacturing is derived from industry standards, regulatory requirements, scientific principles, and iterative design and testing.
Ask a specific question about this device
(251 days)
Occlusion Balloon Catheter
The Embolx Occlusion Balloon Catheter is intended for use in the peripheral vasculature where temporary occlusion is desired and offers a vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Embolx Occlusion Catheter is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.
The Embolx Occlusion Balloon Catheter is a coaxial dual lumen device that consists of an occlusion balloon at the distal end with two embedded radiopaque bands to allow for visualization and positioning of the device under fluoroscopic guidance. The proximal hub consists of two ports: one port for use by the guidewire and delivery of fluids and the second port for inflation and deflation of the balloon. The low profile balloon is manufactured of a compliant material that allows ease of insertion and withdrawal from the vasculature and conforms to the vessel wall. The balloon is inflated and deflated with a hand held syringe. The device is supplied sterile by EtO and is intended for single use. The occlusion catheter has an outside diameter of 2.9F proximally and 2.2F distally. The occlusion balloon on the distal end can be inflated up to 5mm in diameter and 11mm in length. The usable length of the device is 110cm. The device can withstand an infusion pressure up to 900 psi.
This document is a 510(k) premarket notification for a medical device called the Embolx Occlusion Balloon Catheter. It is a submission to the FDA seeking to demonstrate substantial equivalence to a legally marketed predicate device. Therefore, the information provided focuses on comparative testing rather than detailed novel performance criteria and studies as would be expected for a de novo submission or a device with new indications.
Based on the provided text, here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative, pass/fail manner for many of the tests listed. Instead, it states that "all testing was conducted on sterilized and aged test articles" and that "Substantial equivalence was demonstrated through the following non-clinical testing." This implies that the device met whatever internal or recognized standard criteria existed for each test to demonstrate equivalence.
Here's a table based on the "Summary of Supporting Data" with reported performance being that the device "demonstrates substantial equivalence" or "is as safe, as effective, and performs as well as or better than the predicate device."
Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Cytotoxicity | No cytotoxic effects | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Sensitization | No sensitization | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Irritation | No irritation | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Acute Systemic Toxicity | No acute systemic toxicity | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Complement Activation (C3a and Sc5b-9) | Acceptable levels of complement activation | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Hemocompatibility (ASTM Hemolysis) | Acceptable levels of hemolysis (direct and indirect) | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Hemocompatibility (In-vivo Thrombogenicity) | Acceptable levels of thrombogenicity in a canine model | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Hemocompatibility (PTT & PT) | No significant interference with coagulation times | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Pyrogenicity (LAL) | No pyrogenic response | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Material-mediated Pyrogenicity | No material-mediated pyrogenic response | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Sterilization Validation | Validation of sterilization process effectiveness | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Packaging Seal Integrity | Maintenance of sterile barrier and seal integrity | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Dye Penetration | No dye penetration into packaging | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Transit Testing | Packaging integrity maintained after transit simulation | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Dimensional Verification | Device dimensions meet specifications (e.g., 2.9F proximally, 2.2F distally, 5mm diameter, 11mm length for balloon, 110cm usable length) | Verified to specifications (implied, as part of demonstrating substantial equivalence) |
Balloon Prep, Deployment, & Retraction | Successful preparation, deployment, and retraction in simulated use | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Balloon Rated Burst Volume | Balloon withstands specified burst pressure | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Balloon Fatigue | Balloon maintains integrity over repeated inflation/deflation cycles | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Balloon Compliance | Balloon inflates and conforms as expected | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Balloon Inflation/Deflation Time | Inflation/deflation within acceptable timeframe | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Balloon Position Test | Radiopaque markers visible for accurate positioning | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Catheter Bond Strength (tensile) | Bonds withstand required tensile forces | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Tip Pull Strength (tensile) | Tip withstands required tensile forces | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Flexibility & Kink Test | Device demonstrates adequate flexibility and resistance to kinking | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Torque Strength | Device transmits torque adequately without damage | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Radiopacity | Device/markers are visible under fluoroscopy | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Coating Integrity | Coating remains intact | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Catheter Body Burst Test | Catheter body withstands specified burst pressure | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Catheter Body Leakage Test | Catheter body exhibits no leakage | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Contrast Media Flow Rate | Contrast media flows at an acceptable rate | Satisfactory (implied, as part of demonstrating substantial equivalence) |
Corrosion Resistance | Device materials exhibit adequate corrosion resistance | Satisfactory (implied, as part of demonstrating substantial equivalence) |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "non-clinical testing" which typically refers to bench testing, in-vitro (e.g., cytotoxicity, hemolysis), and animal studies (in-vivo thrombogenicity in canine).
- Sample sizes: Specific sample sizes for each test are not provided in this summary.
- Data provenance: Not specified. Given it's a 510(k) for a US market, the testing likely adheres to US or international standards (e.g., ISO for biocompatibility). The in-vivo thrombogenicity study was done using canine models, which is an animal model. All tests appear to be prospective in nature, as they were conducted to support the submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is a medical device submission, not an AI software submission. Therefore, the concept of "ground truth" as established by human experts (e.g., radiologists) for a test set of images or data relevant to an AI algorithm does not apply here. The "truth" for these tests is based on objective laboratory measurements, physical testing parameters, and established biocompatibility standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is also not applicable as it pertains to expert consensus for data labeling (ground truth) relevant to AI/diagnostic software. The testing outlined for this medical device involves physical, mechanical, and biological evaluations, not expert adjudication of clinical cases.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. This is a submission for a physical medical device (an occlusion balloon catheter), not an AI-powered diagnostic tool, and therefore no MRMC study or assessment of human reader improvement with AI assistance would have been performed or relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. No algorithm performance is being evaluated for this physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
As explained in point 3, the concept of "ground truth" for expert consensus on images or outcomes data for an AI algorithm is not applicable. The "truth" in this context is defined by:
- Physical measurements: For dimensional verification, tensile strength, burst pressures, flow rates.
- Chemical/Biological assays: For cytotoxicity, sensitization, irritation, acute systemic toxicity, complement activation, LAL (pyrogenicity), PTT/PT.
- In-vivo observations/histopathology: For in-vivo thrombogenicity in canine models.
- Industry/Regulatory Standards: Adherence to recognized standards for sterility, packaging, and biocompatibility.
8. The sample size for the training set
This question is not applicable. This is a physical medical device, not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established
This question is not applicable. No training set or associated ground truth for an AI algorithm is involved in this medical device submission.
Ask a specific question about this device
(169 days)
EQUALIZER OCCLUSION BALLOON CATHETER
Equalizer™ Occlusion Balloon Catheter is indicated for use for temporary vessel occlusion in applications including arteriography, preoperative occlusion, emergency control of hemorrhage, chemotherapeutic drug infusion and renal opacification procedures.
Any use for procedures other than those indicated in the instructions is not recommended.
The Equalizer™ Occlusion Balloon Catheter is constructed of a soft compliant latex balloon mounted near the tip of a dual-lumen nylon catheter shaft. Radiopaque markers are placed adjacent to the balloon to provide visual reference points for balloon positioning within the vessel. The catheter shaft is radiopaque, maximizing fluoroscopic visibility. Proximal to the bifurcation, the two lumens of the catheter are marked to differentiate their use. The tubing marked 'BALLOON' is the balloon inflation lumen. The tubing marked 'DISTAL' is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire. The lumen can also be used for the infusion of contrast media or therapeutic drugs. Each lumen ends in a luer fitting hub for attachment to ancillary devices.
The provided K140273 510(k) Summary describes a medical device, the Equalizer™ Occlusion Balloon Catheter, and its performance evaluation. However, it does not involve an AI/ML device, and therefore several of the requested categories are not applicable. The study primarily relies on bench testing and biocompatibility testing rather than clinical performance data related to diagnostic accuracy or AI model performance.
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Biocompatibility | ||
Cytotoxicity | Not explicitly stated but implied to meet ISO 10993 standards for medical devices to not induce cell toxicity. | "Biocompatibility testing were performed to support a determination of substantial equivalence." (Implied to meet standards) |
Sensitization | Not explicitly stated but implied to meet ISO 10993 standards for medical devices to not induce sensitization. | "Biocompatibility testing were performed to support a determination of substantial equivalence." (Implied to meet standards) |
Intracutaneous Reactivity | Not explicitly stated but implied to meet ISO 10993 standards for medical devices to not induce intracutaneous reactivity. | "Biocompatibility testing were performed to support a determination of substantial equivalence." (Implied to meet standards) |
Acute Systemic Toxicity | Not explicitly stated but implied to meet ISO 10993 standards for medical devices to not induce acute systemic toxicity. | "Biocompatibility testing were performed to support a determination of substantial equivalence." (Implied to meet standards) |
Materials Mediated Pyrogenicity | Not explicitly stated but implied to meet ISO 10993 standards for medical devices to not induce pyrogenicity. | "Biocompatibility testing were performed to support a determination of substantial equivalence." (Implied to meet standards) |
Hemocompatibility (Direct Hemolysis, PTT, Complement Activation, In Vitro Hemocompatibility) | Not explicitly stated but implied to meet ISO 10993 standards for medical devices for blood-contacting applications. | "Biocompatibility testing were performed to support a determination of substantial equivalence." (Implied to meet standards) |
USP Physicochemical Tests for Plastics | Not explicitly stated but implied to meet USP standards for plastics used in medical devices. | "Biocompatibility testing were performed to support a determination of substantial equivalence." (Implied to meet standards) |
Bench Performance | ||
Deflated Balloon Profile | Not explicitly stated but implied to be comparable to or within acceptable limits of the predicate device for proper insertion and navigation. | "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." (Implied to meet criteria) |
Proximal Bond Tensile | Not explicitly stated but implied to be sufficient for the intended use and comparable to the predicate device to prevent detachment. | "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." (Implied to meet criteria) |
Inflated Balloon O.D. | Not explicitly stated but implied to be within specified dimensions for effective vessel occlusion and comparable to the predicate device. | "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." (Implied to meet criteria) |
Balloon Deflation Time | Not explicitly stated but implied to be rapid enough for safe and effective use, comparable to the predicate device. | "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." (Implied to meet criteria) |
Multiple Inflation, Challenge | Not explicitly stated but implied to withstand multiple inflations without loss of integrity or function, comparable to the predicate device. | "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." (Implied to meet criteria) |
Balloon Burst, Challenge | Not explicitly stated but implied to withstand pressures beyond normal operating conditions without premature burst, comparable to the predicate device. | "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." (Implied to meet criteria) |
Sheath Compatibility | Not explicitly stated but implied to be compatible with intended sheaths, allowing for smooth insertion and removal. | "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." (Implied to meet criteria) |
Notes on Acceptance Criteria: The document primarily states that testing was performed to "support a determination of substantial equivalence" and provide "reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." Specific quantitative acceptance criteria are not detailed in this summary for each test, but they would typically reference established industry standards (e.g., ISO for biocompatibility) or internal specifications benchmarked against the predicate device (Equalizer™ Occlusion Balloon Catheter K021721).
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The study described is bench testing and biocompatibility testing, not a clinical study involving human data. Therefore:
- Sample Size for Test Set: Not specified in terms of human subjects or clinical cases. For bench tests, it would refer to the number of devices or components tested. For biocompatibility, it refers to the samples tested in vitro or in animal models (though details are not provided).
- Data Provenance: Not applicable as it's not clinical data. The tests were performed in a laboratory setting, likely at the manufacturer's facilities or a contract research organization.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This device is a physical medical device (catheter), and its performance evaluation involves engineering and biocompatibility testing, not interpretation of medical images or diagnostic outputs for which ground truth would be established by medical experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are used in clinical studies or studies involving human readers/interpreters to resolve discrepancies in ground truth, which is not relevant for bench and biocompatibility testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device and no MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The concept of "ground truth" as typically applied to diagnostic AI/ML devices is not directly applicable here. Instead, the "ground truth" for this device's performance can be considered:
- Biocompatibility Standards: Established international standards (e.g., ISO 10993 series) define acceptable biological responses.
- Engineering Specifications: Predetermined design specifications and performance requirements (e.g., burst pressure, inflation time, tensile strength) established by the manufacturer and benchmarked against the predicate device.
- Predicate Device Performance: The performance characteristics of the legally marketed predicate device (Equalizer™ Occlusion Balloon Catheter K021721) served as the reference for establishing substantial equivalence.
8. The sample size for the training set
Not applicable. There is no AI/ML component; therefore, no training set is relevant.
9. How the ground truth for the training set was established
Not applicable. There is no AI/ML component; therefore, no ground truth for a training set was established.
Ask a specific question about this device
(30 days)
OCCLUDER OCCLUSION BALLOON CATHETER
The Occluder™ Occlusion Balloon Catheters are indicated for use for temporary Ureteral occlusion and applications including, renal opacification, dislodgment of calculi and preventative calculi migration.
Occluder Occlusion Balloon Catheters are designed for use for temporary occlusion of the ureter and applications including renal opacification, dislodgement of calculi and preventative calculi migration. The devices are provided sterile and are intended for single use. The Occluder Occlusion Balloon Catheters are constructed of a soft compliant latex balloon mounted on the tip of a catheter shaft. Catheter shafts are radiopaque, maximizing fluoroscopic visibility. The Occluder Occlusion Balloon Catheters have two lumens that are marked and color-coded. The balloon tubing, marked BALLOON is a balloon inflation lumen. The tubing marked DISTAL, is the essential lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire. This lumen can also be used for infusion of contrast medium.
This document describes the Boston Scientific Occluder™ Occlusion Balloon Catheter, which is cleared through the 510(k) pathway as substantially equivalent to a predicate device. As such, the study focuses on demonstrating this substantial equivalence rather than establishing new acceptance criteria or conducting a multi-reader multi-case comparative effectiveness study in the way one might for an AI/ML diagnostic device.
Here's an analysis of the provided text in the context of your request:
1. Table of Acceptance Criteria and Reported Device Performance
Criterion Type | Criterion | Reported Performance (from provided text) |
---|---|---|
Substantial Equivalence | The device must demonstrate substantial equivalence to a legally marketed predicate device (Van-Tec Occlusion Balloon Catheter, K841941/A) in terms of intended use, technological characteristics, types of materials, and performance characteristics, demonstrating it is as safe, as effective, and performs as well as the predicate device. | "A direct comparison of key characteristics has been performed and demonstrates that the proposed Occluder Occlusion Balloon Catheter is substantially equivalent to the predicate device in terms of intended use. The proposed device is substantially equivalent to the reference devices in terms of technological characteristics, types of materials and performance characteristics. The proposed Occluder Occlusion Balloon Catheter is as safe, as effective, and performs as well as the predicate device." "The results of the performance testing demonstrate equivalence of the Occluder Catheter to the predicate device. The Occluder Occlusion Balloon Catheter is considered safe and effective for its intended use." |
Performance Testing | Performance testing (bench evaluation) on samples (aged at T=0 and 7-months Accelerated Aging) to support balloon material and shaft material changes. | "Boston Scientific has conducted performance testing with samples aged at T=0 and 7-months Accelerated Aging in support of the balloon material and shaft material changes. The results of the performance testing demonstrate equivalence of the Occluder Catheter to the predicate device." |
Intended Use | The device's intended use must align with temporary ureteral occlusion and applications including renal opacification, dislodgment of calculi, and preventative calculi migration. | "The Occluder™ Occlusion Balloon Catheters are indicated for use for temporary Ureteral occlusion and applications including, renal opacification, dislodgment of calculi and preventative calculi migration." (This is both the stated indication and the demonstration of alignment with the predicate's implied use). |
Technological Characteristics | The device must have similar technological characteristics and fundamental occlusion balloon catheter design to the predicate device. Minor differences (e.g., packaging) must not raise new questions of safety or effectiveness. | "The Occluder Occlusion Balloon Catheters have the same technological characteristics and fundamental occlusion balloon catheter design as the predicate device. The proposed Occluder Occlusion Balloon Catheters are packed using a thermoformed, multi-product tray and supplied with a stopcock and syringe. A tray lid is applied to help secure the device inside the tray cavities. The lidded tray is then placed into a heal-sealed poly/Tyvek pouch. The pouch is labeled and placed into a labeled shelf carton along with a DFU." |
Material Changes | Any balloon and shaft material changes must be supported by performance testing demonstrating equivalence. | "Boston Scientific has conducted performance testing with samples aged at T=0 and 7-months Accelerated Aging in support of the balloon material and shaft material changes. The results of the performance testing demonstrate equivalence of the Occluder Catheter to the predicate device." |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in the context of clinical data or patient samples. The performance testing mentioned is a bench evaluation.
- Sample Size: The document mentions "samples aged at T=0 and 7-months Accelerated Aging." The specific number of samples tested is not provided.
- Data Provenance: The data is from "Boston Scientific" and is generated through "performance testing (Bench Evaluation)." There is no mention of country of origin of data or whether it is retrospective or prospective, as it does not involve human subjects or historical patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this submission. The "ground truth" for a substantial equivalence claim based on bench testing does not involve expert consensus on clinical findings. Instead, the "truth" is established by direct measurement against engineering specifications and comparison to the predicate device's known performance/specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable. Adjudication methods are typically used in clinical studies or when interpreting complex images/data to establish ground truth for diagnostic devices, which is not the case here.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices involving human interpretation, not for a medical device like an occlusion balloon catheter, especially when seeking 510(k) clearance based on substantial equivalence.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No, a standalone algorithm-only performance study was not done. This device is a physical medical instrument, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this submission is based on engineering specifications and direct measurements/observations from bench testing. The goal was to show that the new device's physical and functional characteristics (e.g., balloon inflation, shaft properties, radiopacity) meet predetermined criteria and are equivalent to the predicate device. This essentially means the "ground truth" is adherence to design specifications and performance metrics established from the predicate.
8. The sample size for the training set
This information is not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set.
Ask a specific question about this device
(28 days)
OCCLUSION BALLOON CATHETER
Occlusion Balloon Catheters are indicated for temporary vessel occlusion in applications including arteriography, preoperative occlusion, emergency control of hemorrhage, chemotherapeutic drug infusion and renal opacification procedures.
The Occlusion Balloon Catheter product line consists of two specific designs - Standard Occlusions Balloon Catheters and Berenstein™ Occlusion Balloon Catheter.
Only the Berenstein Occlusion Balloon Catheter has been designed for coaxial delivery of small catheters or embolic agents.
The Occlusion Balloon Catheters are a compliant latex balloon mounted on the distal tip of a dual lumen, radiopaque catheter shaft to which two luer fittings are attached proximally. In addition to the balloon inflation lumen, the central lumen is used to pass the catheter over the guidewire as well as infusion of contrast medium, and in the case of the Berenstein™ Occlusion Balloon Catheter, coaxial delivery of small catheters or embolic agents.
I am sorry, but I cannot fulfill your request to describe the acceptance criteria and study for the provided text. The document describes a medical device, specifically an Occlusion Balloon Catheter, and its 510(k) submission to the FDA. However, the provided text does not contain explicit numerical acceptance criteria or detailed study results that would allow me to construct the specified table and answer all your questions.
The "Performance Data" section and the "Biocompatibility" and "Bench" testing sections generally state that testing was performed and that the results provide "reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." It also concludes that the device is "substantially equivalent" to a predicate device.
Key Missing Information:
- Specific, quantifiable acceptance criteria: For example, what was the minimum tensile strength required for the proximal bond, or what was the maximum allowable balloon deflation time?
- Reported device performance values: The document states what tests were done (e.g., "Proximal Bond Tensile"), but not the results (e.g., "Proximal Bond Tensile = X N").
- Sample sizes for test sets (for specific tests): While the tests are listed, the number of units tested for each specific bench test is not provided.
- Data provenance, expert qualifications, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth details: These types of information are typically associated with performance studies for AI/ML devices or diagnostic accuracy studies, which are not described in this document for an occlusion balloon catheter. This device appears to be a physical medical instrument, not an AI/ML diagnostic tool.
Therefore, I cannot generate the table or answer questions 2, 3, 4, 5, 6, 7, 8, or 9 as the necessary information is not present in the provided text.
Ask a specific question about this device
(141 days)
TRANSFORM OCCLUSION BALLOON CATHETER (COMPLIANT AND SUPER COMPLIANT
The Stryker Neurovascular TransForm™ Occlusion Balloon Catheters (Compliant and Super Compliant) are indicated for use in the neuro and peripheral vasculature to temporarily stop or control blood flow and for balloon assisted embolization of intracranial aneurysms.
Stryker Neurovascular's TransForm Occlusion Balloon Catheters (Compliant and Super Compliant) are compliant, variable stiffness reinforced balloon catheters. The outer surface of the catheter's distal segment is coated with a lubricious hydrophilic coating designed to reduce friction. Each balloon catheter has two radiopaque markers to facilitate fluoroscopic visualization. The proximal end of the balloon catheter incorporates a strain relief and a standard luer fitting to facilitate the attachment of accessories.
Here's an analysis of the provided 510(k) summary for the TransForm™ Occlusion Balloon Catheter, detailing acceptance criteria and proof of their fulfillment.
1. Table of Acceptance Criteria and Reported Device Performance:
The provided 510(k) summary focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data (bench testing, shelf life, packaging, and biocompatibility). The "acceptance criteria" for this type of submission are typically that the new device meets the performance requirements for its intended use and is proven safe and effective for that use, typically demonstrated by showing it performs comparably or better than the predicate for relevant attributes.
Since specific numerical acceptance criteria (e.g., "Deflation time must be
Ask a specific question about this device
(53 days)
SCEPTER C OCCLUSION BALLOON CATHETER
For use in the blood vessels of the peripheral and neurovasculature where temporary occlusion is desired. These catheters offer a vessel selective technique of temporary vascular occlusion which is useful in selectively stopping or controlling blood flow and for balloon assisted embolization of intracranial aneurysms.
For use in the peripheral vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as embolization materials.
For neurovascular use for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as embolization materials, that have been approved or cleared for use in the neurovasculature and are compatible with the inner diameter of the Scepter C/XC Balloon Catheter.
The Scepter C and XC Occlusion Balloon Catheter is a dual coaxial lumen catheter with a nondetachable low inflation pressure compliant balloon attached to the distal end of the catheter. The catheter is designed to track over a steerable guidewire. The inner lumen can be used for infusion/delivery of diagnostic and therapeutic agents. The outer lumen is used for the inflation and deflation of the balloon independent of guidewire position. Radiopaque marker bands are located at ends of the balloon and distal tip of the catheter to facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the microcatheter hub is used for the attachment of accessories. The catheters are packaged sterile for single use only.
The Scepter C and XC Occlusion Balloon Catheter is intended for temporary occlusion in peripheral and neurovasculature, controlling blood flow for balloon-assisted embolization of intracranial aneurysms, and infusing diagnostic and therapeutic agents in peripheral and neurovascular applications.
Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
The document describes pre-clinical testing performed on the Scepter C and XC Occlusion Balloon Catheter. The acceptance criterion for each test is implicitly "Pass," indicating that the device must meet the specified performance standards for each category.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Visual Inspection must pass | Pass |
Tensile strength must pass | Pass |
Leakage (liquid and air) must pass | Pass |
Static and dynamic burst pressure must pass | Pass |
Simulated use must pass | Pass |
Catheter flexural fatigue must pass | Pass |
Compatibility with diagnostic and therapeutic agents must pass | Pass |
Delivery of embolization materials (i.e. Onyx®) must pass | Pass |
Balloon testing (burst, compliance, deflation time, fatigue) must pass | Pass |
DMSO Compatibility must pass | Pass |
Biocompatibility testing (Cytotoxicity, Sensitization/Irritation, Hemocompatibility, Systemic Toxicity) must pass | Pass |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample sizes used for each pre-clinical test. The testing is referred to as "Pre-clinical Testing," which typically involves laboratory-based assessments rather than human or animal studies with large sample sizes.
- Sample Size for Test Set: Not explicitly stated.
- Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. Given it's pre-clinical testing for a 510(k) submission, it's virtually certain to be prospective, laboratory-based testing conducted by the manufacturer, MicroVention, Inc., based in Tustin, California, U.S.A.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. For pre-clinical engineering and biocompatibility tests, ground truth is established by standardized test methods and criteria, not by expert consensus on clinical images or patient outcomes. The "ground truth" is defined by the physical or chemical properties being measured against established regulatory standards (e.g., ISO standards for biocompatibility).
4. Adjudication Method for the Test Set:
Not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical trials or image interpretation studies where there's variability in expert assessment. For pre-clinical engineering tests, results are typically objective measurements against defined pass/fail criteria.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic devices, especially those involving human interpretation of medical images (e.g., radiologists reading scans with and without AI assistance). The Scepter C and XC Occlusion Balloon Catheter is an interventional device, not a diagnostic one.
6. Standalone (Algorithm Only) Performance Study:
No, a standalone (algorithm only) performance study was not done. This type of study is specifically for evaluating the performance of AI algorithms without human intervention. The Scepter C and XC Occlusion Balloon Catheter is a physical medical device, not an AI algorithm.
7. Type of Ground Truth Used:
The ground truth used for these pre-clinical tests is based on engineering specifications, standardized test methods, and regulatory requirements (e.g., ISO 10993 for biocompatibility). For example, "Tensile strength Pass" implies the device met a pre-defined tensile strength threshold established by engineering design and industry standards; "Cytotoxicity (ISO 10993-5) Pass" means the device met the criteria outlined in that specific ISO standard for cytotoxicity.
8. Sample Size for the Training Set:
Not applicable. The Scepter C and XC Occlusion Balloon Catheter is a physical medical device, not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 3