K Number
K142692
Manufacturer
Date Cleared
2015-05-31

(251 days)

Product Code
Regulation Number
870.4450
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Embolx Occlusion Balloon Catheter is intended for use in the peripheral vasculature where temporary occlusion is desired and offers a vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Embolx Occlusion Catheter is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.

Device Description

The Embolx Occlusion Balloon Catheter is a coaxial dual lumen device that consists of an occlusion balloon at the distal end with two embedded radiopaque bands to allow for visualization and positioning of the device under fluoroscopic guidance. The proximal hub consists of two ports: one port for use by the guidewire and delivery of fluids and the second port for inflation and deflation of the balloon. The low profile balloon is manufactured of a compliant material that allows ease of insertion and withdrawal from the vasculature and conforms to the vessel wall. The balloon is inflated and deflated with a hand held syringe. The device is supplied sterile by EtO and is intended for single use. The occlusion catheter has an outside diameter of 2.9F proximally and 2.2F distally. The occlusion balloon on the distal end can be inflated up to 5mm in diameter and 11mm in length. The usable length of the device is 110cm. The device can withstand an infusion pressure up to 900 psi.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called the Embolx Occlusion Balloon Catheter. It is a submission to the FDA seeking to demonstrate substantial equivalence to a legally marketed predicate device. Therefore, the information provided focuses on comparative testing rather than detailed novel performance criteria and studies as would be expected for a de novo submission or a device with new indications.

Based on the provided text, here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria in a quantitative, pass/fail manner for many of the tests listed. Instead, it states that "all testing was conducted on sterilized and aged test articles" and that "Substantial equivalence was demonstrated through the following non-clinical testing." This implies that the device met whatever internal or recognized standard criteria existed for each test to demonstrate equivalence.

Here's a table based on the "Summary of Supporting Data" with reported performance being that the device "demonstrates substantial equivalence" or "is as safe, as effective, and performs as well as or better than the predicate device."

Test CategoryAcceptance Criteria (Implied)Reported Device Performance
CytotoxicityNo cytotoxic effectsSatisfactory (implied, as part of demonstrating substantial equivalence)
SensitizationNo sensitizationSatisfactory (implied, as part of demonstrating substantial equivalence)
IrritationNo irritationSatisfactory (implied, as part of demonstrating substantial equivalence)
Acute Systemic ToxicityNo acute systemic toxicitySatisfactory (implied, as part of demonstrating substantial equivalence)
Complement Activation (C3a and Sc5b-9)Acceptable levels of complement activationSatisfactory (implied, as part of demonstrating substantial equivalence)
Hemocompatibility (ASTM Hemolysis)Acceptable levels of hemolysis (direct and indirect)Satisfactory (implied, as part of demonstrating substantial equivalence)
Hemocompatibility (In-vivo Thrombogenicity)Acceptable levels of thrombogenicity in a canine modelSatisfactory (implied, as part of demonstrating substantial equivalence)
Hemocompatibility (PTT & PT)No significant interference with coagulation timesSatisfactory (implied, as part of demonstrating substantial equivalence)
Pyrogenicity (LAL)No pyrogenic responseSatisfactory (implied, as part of demonstrating substantial equivalence)
Material-mediated PyrogenicityNo material-mediated pyrogenic responseSatisfactory (implied, as part of demonstrating substantial equivalence)
Sterilization ValidationValidation of sterilization process effectivenessSatisfactory (implied, as part of demonstrating substantial equivalence)
Packaging Seal IntegrityMaintenance of sterile barrier and seal integritySatisfactory (implied, as part of demonstrating substantial equivalence)
Dye PenetrationNo dye penetration into packagingSatisfactory (implied, as part of demonstrating substantial equivalence)
Transit TestingPackaging integrity maintained after transit simulationSatisfactory (implied, as part of demonstrating substantial equivalence)
Dimensional VerificationDevice dimensions meet specifications (e.g., 2.9F proximally, 2.2F distally, 5mm diameter, 11mm length for balloon, 110cm usable length)Verified to specifications (implied, as part of demonstrating substantial equivalence)
Balloon Prep, Deployment, & RetractionSuccessful preparation, deployment, and retraction in simulated useSatisfactory (implied, as part of demonstrating substantial equivalence)
Balloon Rated Burst VolumeBalloon withstands specified burst pressureSatisfactory (implied, as part of demonstrating substantial equivalence)
Balloon FatigueBalloon maintains integrity over repeated inflation/deflation cyclesSatisfactory (implied, as part of demonstrating substantial equivalence)
Balloon ComplianceBalloon inflates and conforms as expectedSatisfactory (implied, as part of demonstrating substantial equivalence)
Balloon Inflation/Deflation TimeInflation/deflation within acceptable timeframeSatisfactory (implied, as part of demonstrating substantial equivalence)
Balloon Position TestRadiopaque markers visible for accurate positioningSatisfactory (implied, as part of demonstrating substantial equivalence)
Catheter Bond Strength (tensile)Bonds withstand required tensile forcesSatisfactory (implied, as part of demonstrating substantial equivalence)
Tip Pull Strength (tensile)Tip withstands required tensile forcesSatisfactory (implied, as part of demonstrating substantial equivalence)
Flexibility & Kink TestDevice demonstrates adequate flexibility and resistance to kinkingSatisfactory (implied, as part of demonstrating substantial equivalence)
Torque StrengthDevice transmits torque adequately without damageSatisfactory (implied, as part of demonstrating substantial equivalence)
RadiopacityDevice/markers are visible under fluoroscopySatisfactory (implied, as part of demonstrating substantial equivalence)
Coating IntegrityCoating remains intactSatisfactory (implied, as part of demonstrating substantial equivalence)
Catheter Body Burst TestCatheter body withstands specified burst pressureSatisfactory (implied, as part of demonstrating substantial equivalence)
Catheter Body Leakage TestCatheter body exhibits no leakageSatisfactory (implied, as part of demonstrating substantial equivalence)
Contrast Media Flow RateContrast media flows at an acceptable rateSatisfactory (implied, as part of demonstrating substantial equivalence)
Corrosion ResistanceDevice materials exhibit adequate corrosion resistanceSatisfactory (implied, as part of demonstrating substantial equivalence)

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "non-clinical testing" which typically refers to bench testing, in-vitro (e.g., cytotoxicity, hemolysis), and animal studies (in-vivo thrombogenicity in canine).

  • Sample sizes: Specific sample sizes for each test are not provided in this summary.
  • Data provenance: Not specified. Given it's a 510(k) for a US market, the testing likely adheres to US or international standards (e.g., ISO for biocompatibility). The in-vivo thrombogenicity study was done using canine models, which is an animal model. All tests appear to be prospective in nature, as they were conducted to support the submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is a medical device submission, not an AI software submission. Therefore, the concept of "ground truth" as established by human experts (e.g., radiologists) for a test set of images or data relevant to an AI algorithm does not apply here. The "truth" for these tests is based on objective laboratory measurements, physical testing parameters, and established biocompatibility standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is also not applicable as it pertains to expert consensus for data labeling (ground truth) relevant to AI/diagnostic software. The testing outlined for this medical device involves physical, mechanical, and biological evaluations, not expert adjudication of clinical cases.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. This is a submission for a physical medical device (an occlusion balloon catheter), not an AI-powered diagnostic tool, and therefore no MRMC study or assessment of human reader improvement with AI assistance would have been performed or relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. No algorithm performance is being evaluated for this physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

As explained in point 3, the concept of "ground truth" for expert consensus on images or outcomes data for an AI algorithm is not applicable. The "truth" in this context is defined by:

  • Physical measurements: For dimensional verification, tensile strength, burst pressures, flow rates.
  • Chemical/Biological assays: For cytotoxicity, sensitization, irritation, acute systemic toxicity, complement activation, LAL (pyrogenicity), PTT/PT.
  • In-vivo observations/histopathology: For in-vivo thrombogenicity in canine models.
  • Industry/Regulatory Standards: Adherence to recognized standards for sterility, packaging, and biocompatibility.

8. The sample size for the training set

This question is not applicable. This is a physical medical device, not an AI algorithm that requires a training set.

9. How the ground truth for the training set was established

This question is not applicable. No training set or associated ground truth for an AI algorithm is involved in this medical device submission.

§ 870.4450 Vascular clamp.

(a)
Identification. A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.(b)
Classification. Class II (performance standards).