Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K183679
    Manufacturer
    Date Cleared
    2019-04-24

    (117 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Occlusion Balloon Catheter is indicated for temporary occlusion of large vessels, including the superior vena cava, in applications including perioperative occlusion and emergency control of hemorrhage.

    Device Description

    The Occlusion Balloon Catheter is a multi-lumen catheter which has a compliant polyurethane balloon with a maximum diameter of 32mm at 60cc inflation volume. The balloon has a nominal length of 80mm. The device is constructed with a blended PEBA shaft and is available in two models; one model is compatible with 8Fr (or larger) introducer sheaths and one model is compatible with 10Fr (or larger) introducer sheaths. Both models provide the same balloon performance with respect to sizing and occlusion. The two models are intended for individual physician preference for patient-specific sheath selection. The device has an effective length of 90cm and is compatible with 0.035" diameter guidewires. Three (3) platinum-iridium radiopaque marker bands are placed on the shaft to facilitate balloon placement in the anatomy prior to inflation. The proximal end of the catheter has an integral PEBA bifurcation manifold with female luer ports to allow communication with the balloon inflation lumen and guidewire lumen. A PVC extension tube (with stopcock) is connected to the balloon inflation port to facilitate handling. The device is a single use, sterile device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an Occlusion Balloon Catheter. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria and reporting its performance.

    Therefore, many of the requested items (e.g., acceptance criteria table, sample size for test set, number of experts for ground truth, MRMC study, training set details) are not applicable and cannot be extracted from this document because the submission method relies on demonstrating equivalence through non-clinical testing and comparison to predicate devices, not through a clinical trial with specific performance metrics against acceptance criteria.

    Here's what can be extracted based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. The document does not provide specific quantitative acceptance criteria or reported performance metrics against such criteria in the context of a clinical study. It lists a series of non-clinical tests performed to ensure the device meets established performance criteria and will perform as intended for substantial equivalence.
    Acceptance CriteriaReported Device Performance
    Not provided for clinically relevant performance metrics. Non-clinical tests were performed to ensure substantial equivalence and intended performance.Not provided for clinically relevant performance metrics. The non-clinical tests included items like visual/dimensional inspections, vessel occlusion, balloon fatigue, kink resistance, burst/leak volume, tensile strength, torque strength, and shelf life.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No clinical test set or data provenance is detailed, as clinical evaluation was not required. The tests performed were non-clinical (mechanical, performance, biocompatibility).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No ground truth for a test set was established using experts, as no clinical study requiring such ground truth was conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No adjudication method for a test set is mentioned, as no clinical study requiring this was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. No MRMC study was done, nor does the device involve AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a physical medical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. No ground truth of this nature was used, as clinical studies involving such data were not performed. The "ground truth" for this submission was based on successful completion of non-clinical performance and mechanical tests to demonstrate substantial equivalence to the predicate device.

    8. The sample size for the training set

    • Not Applicable. No training set is mentioned, as this is a medical device submission based on substantial equivalence and non-clinical testing, not a machine learning model.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set or ground truth for it was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142692
    Manufacturer
    Date Cleared
    2015-05-31

    (251 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Embolx Occlusion Balloon Catheter is intended for use in the peripheral vasculature where temporary occlusion is desired and offers a vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Embolx Occlusion Catheter is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.

    Device Description

    The Embolx Occlusion Balloon Catheter is a coaxial dual lumen device that consists of an occlusion balloon at the distal end with two embedded radiopaque bands to allow for visualization and positioning of the device under fluoroscopic guidance. The proximal hub consists of two ports: one port for use by the guidewire and delivery of fluids and the second port for inflation and deflation of the balloon. The low profile balloon is manufactured of a compliant material that allows ease of insertion and withdrawal from the vasculature and conforms to the vessel wall. The balloon is inflated and deflated with a hand held syringe. The device is supplied sterile by EtO and is intended for single use. The occlusion catheter has an outside diameter of 2.9F proximally and 2.2F distally. The occlusion balloon on the distal end can be inflated up to 5mm in diameter and 11mm in length. The usable length of the device is 110cm. The device can withstand an infusion pressure up to 900 psi.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the Embolx Occlusion Balloon Catheter. It is a submission to the FDA seeking to demonstrate substantial equivalence to a legally marketed predicate device. Therefore, the information provided focuses on comparative testing rather than detailed novel performance criteria and studies as would be expected for a de novo submission or a device with new indications.

    Based on the provided text, here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative, pass/fail manner for many of the tests listed. Instead, it states that "all testing was conducted on sterilized and aged test articles" and that "Substantial equivalence was demonstrated through the following non-clinical testing." This implies that the device met whatever internal or recognized standard criteria existed for each test to demonstrate equivalence.

    Here's a table based on the "Summary of Supporting Data" with reported performance being that the device "demonstrates substantial equivalence" or "is as safe, as effective, and performs as well as or better than the predicate device."

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    CytotoxicityNo cytotoxic effectsSatisfactory (implied, as part of demonstrating substantial equivalence)
    SensitizationNo sensitizationSatisfactory (implied, as part of demonstrating substantial equivalence)
    IrritationNo irritationSatisfactory (implied, as part of demonstrating substantial equivalence)
    Acute Systemic ToxicityNo acute systemic toxicitySatisfactory (implied, as part of demonstrating substantial equivalence)
    Complement Activation (C3a and Sc5b-9)Acceptable levels of complement activationSatisfactory (implied, as part of demonstrating substantial equivalence)
    Hemocompatibility (ASTM Hemolysis)Acceptable levels of hemolysis (direct and indirect)Satisfactory (implied, as part of demonstrating substantial equivalence)
    Hemocompatibility (In-vivo Thrombogenicity)Acceptable levels of thrombogenicity in a canine modelSatisfactory (implied, as part of demonstrating substantial equivalence)
    Hemocompatibility (PTT & PT)No significant interference with coagulation timesSatisfactory (implied, as part of demonstrating substantial equivalence)
    Pyrogenicity (LAL)No pyrogenic responseSatisfactory (implied, as part of demonstrating substantial equivalence)
    Material-mediated PyrogenicityNo material-mediated pyrogenic responseSatisfactory (implied, as part of demonstrating substantial equivalence)
    Sterilization ValidationValidation of sterilization process effectivenessSatisfactory (implied, as part of demonstrating substantial equivalence)
    Packaging Seal IntegrityMaintenance of sterile barrier and seal integritySatisfactory (implied, as part of demonstrating substantial equivalence)
    Dye PenetrationNo dye penetration into packagingSatisfactory (implied, as part of demonstrating substantial equivalence)
    Transit TestingPackaging integrity maintained after transit simulationSatisfactory (implied, as part of demonstrating substantial equivalence)
    Dimensional VerificationDevice dimensions meet specifications (e.g., 2.9F proximally, 2.2F distally, 5mm diameter, 11mm length for balloon, 110cm usable length)Verified to specifications (implied, as part of demonstrating substantial equivalence)
    Balloon Prep, Deployment, & RetractionSuccessful preparation, deployment, and retraction in simulated useSatisfactory (implied, as part of demonstrating substantial equivalence)
    Balloon Rated Burst VolumeBalloon withstands specified burst pressureSatisfactory (implied, as part of demonstrating substantial equivalence)
    Balloon FatigueBalloon maintains integrity over repeated inflation/deflation cyclesSatisfactory (implied, as part of demonstrating substantial equivalence)
    Balloon ComplianceBalloon inflates and conforms as expectedSatisfactory (implied, as part of demonstrating substantial equivalence)
    Balloon Inflation/Deflation TimeInflation/deflation within acceptable timeframeSatisfactory (implied, as part of demonstrating substantial equivalence)
    Balloon Position TestRadiopaque markers visible for accurate positioningSatisfactory (implied, as part of demonstrating substantial equivalence)
    Catheter Bond Strength (tensile)Bonds withstand required tensile forcesSatisfactory (implied, as part of demonstrating substantial equivalence)
    Tip Pull Strength (tensile)Tip withstands required tensile forcesSatisfactory (implied, as part of demonstrating substantial equivalence)
    Flexibility & Kink TestDevice demonstrates adequate flexibility and resistance to kinkingSatisfactory (implied, as part of demonstrating substantial equivalence)
    Torque StrengthDevice transmits torque adequately without damageSatisfactory (implied, as part of demonstrating substantial equivalence)
    RadiopacityDevice/markers are visible under fluoroscopySatisfactory (implied, as part of demonstrating substantial equivalence)
    Coating IntegrityCoating remains intactSatisfactory (implied, as part of demonstrating substantial equivalence)
    Catheter Body Burst TestCatheter body withstands specified burst pressureSatisfactory (implied, as part of demonstrating substantial equivalence)
    Catheter Body Leakage TestCatheter body exhibits no leakageSatisfactory (implied, as part of demonstrating substantial equivalence)
    Contrast Media Flow RateContrast media flows at an acceptable rateSatisfactory (implied, as part of demonstrating substantial equivalence)
    Corrosion ResistanceDevice materials exhibit adequate corrosion resistanceSatisfactory (implied, as part of demonstrating substantial equivalence)

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "non-clinical testing" which typically refers to bench testing, in-vitro (e.g., cytotoxicity, hemolysis), and animal studies (in-vivo thrombogenicity in canine).

    • Sample sizes: Specific sample sizes for each test are not provided in this summary.
    • Data provenance: Not specified. Given it's a 510(k) for a US market, the testing likely adheres to US or international standards (e.g., ISO for biocompatibility). The in-vivo thrombogenicity study was done using canine models, which is an animal model. All tests appear to be prospective in nature, as they were conducted to support the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is a medical device submission, not an AI software submission. Therefore, the concept of "ground truth" as established by human experts (e.g., radiologists) for a test set of images or data relevant to an AI algorithm does not apply here. The "truth" for these tests is based on objective laboratory measurements, physical testing parameters, and established biocompatibility standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is also not applicable as it pertains to expert consensus for data labeling (ground truth) relevant to AI/diagnostic software. The testing outlined for this medical device involves physical, mechanical, and biological evaluations, not expert adjudication of clinical cases.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This is a submission for a physical medical device (an occlusion balloon catheter), not an AI-powered diagnostic tool, and therefore no MRMC study or assessment of human reader improvement with AI assistance would have been performed or relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. No algorithm performance is being evaluated for this physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    As explained in point 3, the concept of "ground truth" for expert consensus on images or outcomes data for an AI algorithm is not applicable. The "truth" in this context is defined by:

    • Physical measurements: For dimensional verification, tensile strength, burst pressures, flow rates.
    • Chemical/Biological assays: For cytotoxicity, sensitization, irritation, acute systemic toxicity, complement activation, LAL (pyrogenicity), PTT/PT.
    • In-vivo observations/histopathology: For in-vivo thrombogenicity in canine models.
    • Industry/Regulatory Standards: Adherence to recognized standards for sterility, packaging, and biocompatibility.

    8. The sample size for the training set

    This question is not applicable. This is a physical medical device, not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable. No training set or associated ground truth for an AI algorithm is involved in this medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132990
    Date Cleared
    2013-10-22

    (28 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Occlusion Balloon Catheters are indicated for temporary vessel occlusion in applications including arteriography, preoperative occlusion, emergency control of hemorrhage, chemotherapeutic drug infusion and renal opacification procedures.

    The Occlusion Balloon Catheter product line consists of two specific designs - Standard Occlusions Balloon Catheters and Berenstein™ Occlusion Balloon Catheter.

    Only the Berenstein Occlusion Balloon Catheter has been designed for coaxial delivery of small catheters or embolic agents.

    Device Description

    The Occlusion Balloon Catheters are a compliant latex balloon mounted on the distal tip of a dual lumen, radiopaque catheter shaft to which two luer fittings are attached proximally. In addition to the balloon inflation lumen, the central lumen is used to pass the catheter over the guidewire as well as infusion of contrast medium, and in the case of the Berenstein™ Occlusion Balloon Catheter, coaxial delivery of small catheters or embolic agents.

    AI/ML Overview

    I am sorry, but I cannot fulfill your request to describe the acceptance criteria and study for the provided text. The document describes a medical device, specifically an Occlusion Balloon Catheter, and its 510(k) submission to the FDA. However, the provided text does not contain explicit numerical acceptance criteria or detailed study results that would allow me to construct the specified table and answer all your questions.

    The "Performance Data" section and the "Biocompatibility" and "Bench" testing sections generally state that testing was performed and that the results provide "reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." It also concludes that the device is "substantially equivalent" to a predicate device.

    Key Missing Information:

    • Specific, quantifiable acceptance criteria: For example, what was the minimum tensile strength required for the proximal bond, or what was the maximum allowable balloon deflation time?
    • Reported device performance values: The document states what tests were done (e.g., "Proximal Bond Tensile"), but not the results (e.g., "Proximal Bond Tensile = X N").
    • Sample sizes for test sets (for specific tests): While the tests are listed, the number of units tested for each specific bench test is not provided.
    • Data provenance, expert qualifications, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth details: These types of information are typically associated with performance studies for AI/ML devices or diagnostic accuracy studies, which are not described in this document for an occlusion balloon catheter. This device appears to be a physical medical instrument, not an AI/ML diagnostic tool.

    Therefore, I cannot generate the table or answer questions 2, 3, 4, 5, 6, 7, 8, or 9 as the necessary information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101877
    Manufacturer
    Date Cleared
    2010-08-06

    (31 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Occlusion Balloon Catheter is intended for temporary occlusion of large vessels.

    Device Description

    The Occlusion Balloon Catheter is a triple lumen balloon catheter designed to temporarily occlude large vessels. The "Infusion" lumen is used to infuse contrast material through the catheter. The "Distal .035" lumen extends the length of the catheter and is used for placement over wire guides. The "Balloon" lumen is used to inflate and deflate the balloon. The Occlusion Balloon Catheter is available in one configuration. The catheter has two radiopaque marker bands on the shaft, enclosed within the balloon bonds, to help identify the location of the balloon under fluoroscopy. The balloon has a variable diameter up to a maximum of 40 mm when inflated and a catheter length of 75 cm.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Occlusion Balloon Catheter (K101877)

    This document describes the acceptance criteria and the study that proves the Occlusion Balloon Catheter meets these criteria, based on the provided 510(k) submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) submission for the Occlusion Balloon Catheter (K101877) focuses on establishing substantial equivalence to a predicate device (CODA Balloon Catheter, K032869) through design and performance testing. The document lists several tests conducted but does not explicitly state numerical acceptance criteria or specific quantitative performance values for each test within the document. Instead, it concludes that the "results of these tests provide reasonable assurance that the device has been designed and tested to ensure conformance to the requirements for its intended use."

    However, based on the nature of the tests listed and general device safety and effectiveness principles, we can infer the common acceptance criteria for such a device and assume the reported performance met these underlying standards.

    Test CategoryImplicit Acceptance Criteria (Inferred)Reported Device Performance
    Burst Pressure TestingThe balloon catheter must withstand specified maximum internal pressures without bursting, maintaining integrity during clinical use."The proposed Occlusion Balloon Catheter was subjected to... Burst Pressure Testing." (Implied: device met predefined non-bursting pressure limits)
    Bond Strength TestingAll catheter bonds (e.g., balloon to shaft) must withstand specified forces without delamination or failure."The proposed Occlusion Balloon Catheter was subjected to... Bond Strength Testing." (Implied: bonds met predefined strength requirements)
    Tensile TestingThe catheter shaft and its components must withstand specified tensile forces without breaking or permanent deformation."The proposed Occlusion Balloon Catheter was subjected to... Tensile Testing." (Implied: device met predefined tensile strength requirements)
    Fatigue TestingThe device must withstand repeated inflation/deflation cycles or other relevant mechanical stresses without failure, demonstrating durability."The proposed Occlusion Balloon Catheter was subjected to... Fatigue Testing." (Implied: device demonstrated durability over specified cycles)
    Sheath Compatibility TestingThe catheter must be able to be introduced and withdrawn smoothly through compatible introducer sheaths without damage to the catheter or sheath."The proposed Occlusion Balloon Catheter was subjected to... Sheath Compatibility Testing." (Implied: demonstrated compatibility and smooth passage)
    Inflation/Deflation TimeThe balloon must inflate and deflate within clinically acceptable timeframes for efficient temporary occlusion."The proposed Occlusion Balloon Catheter was subjected to... Inflation/Deflation Time." (Implied: met predefined time limits for inflation/deflation)
    Occlusion TestingThe inflated balloon must effectively occlude the intended vessel lumen, preventing flow past the balloon."The proposed Occlusion Balloon Catheter was subjected to... Occlusion Testing." (Implied: successfully demonstrated effective vessel occlusion)
    Accelerated Aged TestingThe device must maintain its functional and mechanical properties after simulating a specified shelf-life period."The proposed Occlusion Balloon Catheter was subjected to... Accelerated Aged Testing." (Implied: demonstrated stability over simulated shelf-life)
    Biocompatibility TestingThe device materials must be biocompatible, not eliciting adverse biological responses (e.g., cytotoxicity, sensitization, irritation)."The proposed Occlusion Balloon Catheter was subjected to... Biocompatibility Testing." (Implied: materials met ISO 10993 standards for biocompatibility)

    Overall Reported Performance: The document states: "The results of these tests provide reasonable assurance that the device has been designed and tested to ensure conformance to the requirements for its intended use." This indicates that all acceptance criteria were met.

    2. Sample Size for Test Set and Data Provenance

    The provided 510(k) submission does not explicitly state the sample sizes used for each of the performance tests (Burst Pressure, Bond Strength, Tensile, Fatigue, Sheath Compatibility, Inflation/Deflation Time, Occlusion, Accelerated Aged, and Biocompatibility Testing).

    The data provenance is from Cook Incorporated, USA, as the submitter is located in Bloomington, IN, USA. The studies are described as prospective device performance tests conducted on the manufactured device.

    3. Number of Experts and Qualifications for Ground Truth

    This type of device submission (mechanical medical device) does not typically involve human expert adjudication for establishing "ground truth" in the same way an AI/ML-based diagnostic device would.

    For mechanical performance testing, the "ground truth" is based on engineering specifications and scientific principles, verified through objective measurements and validated test methods. The "experts" involved would be the design engineers, quality control personnel, and test laboratory technicians conducting and interpreting the results according to established international and internal standards (e.g., ISO, ASTM, internal Cook specifications). Their qualifications would include engineering degrees, relevant certifications, and experience in medical device testing.

    4. Adjudication Method for Test Set

    As explained in point 3, no human adjudication method (like 2+1 or 3+1) was used or is applicable for this type of mechanical device performance testing. The "adjudication" is inherent in the objective measurement and comparison of test results against predefined engineering specifications and criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This type of study is relevant for diagnostic devices (especially imaging-based AI) where human readers interpret data, and the AI's impact on their performance is evaluated. The Occlusion Balloon Catheter is a mechanical therapeutic device, not a diagnostic one.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. The Occlusion Balloon Catheter is a physical medical device, not an algorithm. Therefore, "standalone" algorithm performance is not a relevant concept for this submission.

    7. Type of Ground Truth Used

    The ground truth used for the device's performance testing is based on:

    • Engineering Specifications: Predefined performance metrics (e.g., minimum burst pressure, maximum inflation time, required tensile strength) derived from design requirements, risk analysis, and relevant standards.
    • Physical Measurements: Objective data obtained from laboratory testing using calibrated equipment.
    • Biocompatibility Standards: Compliance with recognized international standards for biological evaluation of medical devices (e.g., ISO 10993 series).

    8. Sample Size for the Training Set

    Not applicable. This submission is for a physical medical device, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The "training" in the context of device development refers to iterative design, prototyping, and testing phases to refine the device, but not in the sense of a machine learning algorithm's training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As stated in point 8, there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081052
    Manufacturer
    Date Cleared
    2008-06-18

    (65 days)

    Product Code
    Regulation Number
    870.1240
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Occlusion Balloon Catheter, model Venos™, is intended for a temporary occlusion of coronary sinus during a venogram, or infusion of contrast media or drug, or for possible introduction of devices into the coronary venous system.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device (Occlusion Balloon Catheter, Model Venos™). It does not contain information about acceptance criteria and a study proving the device meets those criteria.

    Therefore, I cannot provide the requested information from this document. The letter only states that the device is substantially equivalent to a legally marketed predicate device, allowing it to be marketed. It does not detail the specific performance metrics or studies used to demonstrate that equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070126
    Manufacturer
    Date Cleared
    2008-02-08

    (388 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SentreHeart Occlusion Balloon Catheter is intended for temporary occlusion of large vessels in applications such as arteriography, preoperative occlusion, and emergency controlled hemorrhage procedures.

    Device Description

    The Occlusion Balloon Catheter consists of a catheter shaft with two independent lumens upon which an expandable balloon material is bonded. The "Guide Wire" lumen extends the length of the catheter and is used for placement of guide wires and injection of contrast for angiographic visualization. The "Balloon" lumen is used to inflate and deflate the balloon. Radiopaque markers at the location of the balloon provide fluoroscopic visualization of balloon placement. The outer diameter of the catheter shaft is coated with a hydrophilic polymer that reduces friction during manipulation in the vessel.

    AI/ML Overview

    The provided text describes a 510(k) summary for the SentreHeart Occlusion Balloon Catheter. However, it does not contain specific details about acceptance criteria for performance, device performance data through a study, sample sizes for testing, ground truth establishment, or human-in-the-loop studies.

    Here's a breakdown of what can be extracted and what information is missing based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    (Not specified in the document)"Functional testing was conducted to support the claim of substantial equivalence and to demonstrate the Occlusion Balloon Catheter is safe and effective for its intended use." (No specific performance metrics or thresholds are provided)
    Biocompatibility"Results of testing demonstrate the Occlusion Balloon Catheter is biocompatible." (No specific tests or criteria detailed)

    Missing Information: The document states that functional testing and biocompatibility testing were performed but does not quantify the acceptance criteria or the results against those criteria. It relies on a general statement of "safe and effective" and "biocompatible."

    2. Sample size used for the test set and the data provenance

    Missing Information: The document does not mention any specific test set, its sample size, or data provenance (e.g., country of origin, retrospective/prospective). The "functional testing" is broadly mentioned without any details.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Missing Information: Since no specific test set or clinical study is detailed, there is no mention of experts or how ground truth would have been established. This device is a catheter, not an AI/diagnostic imaging device where expert interpretation is typically a primary ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Missing Information: No information provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Missing Information: This is not applicable to a device like an "Occlusion Balloon Catheter." MRMC studies are typically for diagnostic imaging devices where human interpretation is involved. This is a medical device for temporary occlusion, not a diagnostic tool assisted by AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Missing Information: This is not applicable. The device is a physical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Missing Information: For a medical device like this, "ground truth" would typically relate to the physical and mechanical properties of the device (e.g., balloon inflation pressure, burst pressure, material strength, biocompatibility results) and its ability to achieve its intended function (occlusion) in relevant models (in vitro, animal, or human studies, though clinical efficacy data isn't detailed here for the 510(k)). The document only generally refers to "functional testing."

    8. The sample size for the training set

    Missing Information: Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Missing Information: Not applicable.

    Summary of the Study Mentioned in the Document:

    The document briefly mentions:

    • "Functional testing": Conducted to support substantial equivalence and demonstrate safety and effectiveness.
    • "Biocompatibility testing": Conducted for the materials used in the catheter.

    Key takeaway: This 510(k) summary is for a medical device (a catheter), not a software or AI-based diagnostic tool. Therefore, many of the requested categories (like ground truth from experts, MRMC studies, training sets, algorithms) are not applicable. The document itself provides a very high-level summary of testing without specific quantitative performance data or detailed study designs. It relies on the concept of "substantial equivalence" to predicate devices rather than novel, extensive performance studies with detailed acceptance criteria for statistical clinical endpoints.

    Ask a Question

    Ask a specific question about this device

    K Number
    K960832
    Date Cleared
    1996-05-29

    (90 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1