Search Results
Found 6 results
510(k) Data Aggregation
(184 days)
Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl
Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
The subject device in 510(K) notification is a blue nitrile examination gloves powder free tested for use with Chemotherapy drugs and Fentanyl Citrate. The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I). The device is available in Small, Medium, Large and Extra Large sizes. The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)
The provided document is a 510(k) premarket notification for "Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate." This document outlines the acceptance criteria and the results of a study to demonstrate the device's performance.
Here's the breakdown of the information requested based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document provides extensive tables for the performance of the gloves. Rather than recreating the entire tables found on pages 10-14, a summary of key acceptance criteria and reported performance is presented below. Please refer to pages 10-14 of the original document for the complete lists of chemotherapy drugs and specific results.
Test Method (Standard) | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D6319-2019 | Length of the gloves | Min 220 mm for Size Small & Min 230 mm for all other sizes | Small: 404 mm |
Medium: 405 mm | |||
Large: 405 mm | |||
X-Large: 406 mm (All exceed minimums) | |||
ASTM D6319-2019 | Width of the gloves | Small: 80+/-10 mm; Medium: 95+/-10 mm; Large: 110+/-10 mm; X-Large: 120+/-10 mm | Small: 84 mm |
Medium: 94 mm | |||
Large: 105 mm | |||
X-Large: 115 mm (All within +/- 10mm range) | |||
ASTM D6319-2019 | Thickness of the gloves | Palm 0.05 mm min; Finger 0.05 mm min for all sizes | Palm: 0.19mm (all sizes); Finger: 0.21mm (all sizes) (All exceed minimums) |
ASTM D6319-2019 | Physical properties - Tensile strength (Before Ageing) | $\ge$ 14 Mpa for all sizes | Small: 22.77Mpa; Medium: 24.46Mpa; Large: 24.51Mpa; X-Large: 24.59Mpa (All exceed minimum) |
ASTM D6319-2019 | Physical properties - Tensile strength (After Ageing) | $\ge$ 14 Mpa for all sizes | Small: 20.50Mpa; Medium: 21.81Mpa; Large: 21.95Mpa; X-Large: 22.05Mpa (All exceed minimum) |
ASTM D6319-2019 | Physical properties - Ultimate Elongation (Before Ageing) | $\ge$ 500% for all sizes | Small: 885%; Medium: 888%; Large: 891%; X-Large: 892% (All exceed minimum) |
ASTM D6319-2019 | Physical properties - Ultimate Elongation (After Ageing) | $\ge$ 400% for all sizes | Small: 868%; Medium: 870%; Large: 872%; X-Large: 875% (All exceed minimum) |
ASTM D5151-2019 | Detection of Holes in Medical Gloves | AQL 2.5 | Passes AQL 1.5 (Better than acceptance criteria, indicating fewer defects) |
ASTM D6124-06 | Residual Powder on Medical Gloves | 2 Mg/Glove Max | Small: 0.16 mg/glove; Medium: 0.16 mg/glove; Large: 0.16 mg/glove; X-Large: 0.16mg/glove (All meet or are below maximum) |
ASTM D6978-05 (2019) | Permeation by Chemotherapy Drugs and Fentanyl Citrate | Specified minimum breakthrough detection times for each drug (e.g., >240 Minutes for most) | Most chemotherapy drugs and Fentanyl Citrate showed >240 Minutes breakthrough detection time. |
Exceptions: Carmustine (3.3 mg/ml) - 35.0 Minutes; Thiotepa (10 mg/ml) - 64.9 Minutes. Warnings are provided for these two drugs. (See pages 2-3 and 12-13 for full list and values). | |||
ISO 10993-10 | Dermal Irritation (Rabbits) | Non irritant | Under the condition of study not an irritant (Meets criteria) |
ISO 10993-10 | Skin Sensitization (Guinea Pig) | Non sensitizer | Under the conditions of the study not a sensitizer (Meets criteria) |
ISO 10993-5:2009 | In vitro Cytotoxicity | Non cytotoxic | Under the conditions of the study cytotoxic. (Does NOT meet criteria for non-cytotoxic) |
ISO 10993-11:2017 | Acute Systemic Toxicity | Not pose systemic toxicity concern | Under the conditions of study the device extracts do not pose acute systemic toxicity concern (Meets criteria) |
ISO 10993-11:2017 / USP 41 | Material-Mediated Pyrogenicity | Not pose material mediated pyrogenicity response | Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response (Meets criteria) |
Note on Cytotoxicity: The device itself is reported as cytotoxic under the tested conditions. This is a critical finding that does not meet a "Non cytotoxic" acceptance criterion. However, the FDA has still cleared the device. This implies that either the level of cytotoxicity observed is deemed acceptable for the intended use and risk-benefit profile of examination gloves, or mitigating factors/information were provided elsewhere (not in this summary). It's important to note this discrepancy.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document lists various ASTM and ISO standards for testing, which define the sample sizes required for each test (e.g., AQL levels for watertightness). However, the specific quantity/sample size used for each individual test conducted by the manufacturer is not explicitly stated in this summary document beyond what's implied by the standard (e.g., AQL 1.5 for watertightness performance is a characteristic, not a sample size).
- Data Provenance: The manufacturing location of the device owner, JR Engineering & Medical Technologies (M) Sdn. Bhd., is in Malaysia (Lot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1, Pusat Perindustrian 2, 44200 Rasa, Hulu Selangor, Selangor Darul Ehsan, Malaysia). The testing was conducted in accordance with international standards (ASTM, ISO), implying that the data originates from laboratory studies, likely conducted by third-party labs or the manufacturer's own qualified labs, but the specific country of the testing labs is not mentioned.
- Retrospective or Prospective: The testing described is prospective laboratory performance testing of the device, rather than a retrospective analysis of existing data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This product is a medical glove, not an AI or imaging device requiring expert consensus for ground truth. The "ground truth" for this device's performance is established through objective, standardized laboratory testing (e.g., ASTM D6319 for physical properties, ASTM D6978 for chemical permeation, ISO 10993 for biocompatibility). These tests rely on chemical and physical measurements, not human expert interpretation of images or clinical outcomes. Therefore, the concept of "number of experts used to establish ground truth" with qualifications like "radiologist with 10 years of experience" is not applicable to this type of device. The "experts" are the technicians and scientists performing the standardized validation tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As this is not a study involving human interpretation of data where adjudication would be necessary (like in imaging diagnosis), there is no adjudication method described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical examination glove, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm or AI device. The tests performed are "standalone" in the sense that they assess the intrinsic material properties and barrier performance of the glove itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for this device is based on standardized, objective physical, chemical, and biological measurements obtained through laboratory testing, as defined by relevant ASTM and ISO standards. This includes:
- Physical measurements (e.g., length, width, thickness, tensile strength, elongation).
- Chemical analysis (e.g., residual powder, breakthrough detection time for drugs).
- Biological testing (e.g., irritation, sensitization, cytotoxicity, systemic toxicity, pyrogenicity).
8. The sample size for the training set
Not applicable. This device is a manufactured product (glove), not an AI algorithm or model that requires a "training set" of data.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(85 days)
JR MEDIC
Latex Examination Powder Free Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color (No color is added) and are powder free and Non-Sterile.
The provided document describes the FDA 510(k) premarket notification for "Latex Examination Powder Free Gloves." This is a Class I medical device, which typically does not require a complex AI/algorithm-based study with human readers, ground truth establishment by experts, or MRMC studies. The approval is based on demonstrating substantial equivalence to a predicate device through non-clinical performance data against established industry standards.
Therefore, many of the requested elements for describing AI acceptance criteria and study methodology (like sample sizes for test/training sets, expert ground truth, adjudication, MRMC studies, standalone algorithm performance, etc.) are not applicable to this type of device and submission.
The document explicitly states "H. SUMMARY OF CLINICAL TESTING: Not applicable." This reinforces that the submission did not rely on clinical performance data involving human readers or sophisticated AI algorithms.
However, I can extract the acceptance criteria and performance data as presented for these gloves.
Acceptance Criteria and Reported Device Performance
The acceptance criteria for the "Latex Examination Powder Free Gloves" are based on established ASTM (American Society for Testing and Materials) standards for rubber examination gloves. The table below summarizes these criteria and the reported performance of the device (K222348) compared to the predicate device (K192329).
1. Table of Acceptance Criteria and Reported Device Performance
Characteristics / Test Method | Standard / Purpose | Acceptance Criteria (K192329 Predicate) | Reported Device Performance (K222348 - Current Device) | Comparison |
---|---|---|---|---|
Dimensions - Length (ASTM D3578-2019) | To determine the length of the gloves | > 230 mm (all sizes) | > 230 mm (Actual values: X-Small 305 mm, Small 306 mm, Medium 307 mm, Large 308 mm, X-Large 310 mm) | Similar |
Dimensions - Width (ASTM D3578-2019) | To determine the width of the gloves | Min 95+/-10 mm (for medium size) | X-Small 76 mm, Small 85 mm, Medium 96 mm, Large 106 mm, X-Large 116 mm | Similar |
Physical Properties - Tensile Strength (ASTM D3578-2019) | To Determine the physical properties | Before Ageing: > 18 Mpa (all sizes) | ||
After Ageing: > 14 Mpa (all sizes) | Before Ageing: X-Small 22.07 Mpa, Small 22.15 Mpa, Medium 22.22 Mpa, Large 22.30 Mpa, X-Large 22.32 Mpa | |||
After Ageing: X-Small 18.49 Mpa, Small 18.56 Mpa, Medium 18.67 Mpa, Large 18.74 Mpa, X-Large 18.76 Mpa | Similar | |||
Physical Properties - Ultimate Elongation (ASTM D3578-2019) | To Determine the physical properties | Before Ageing: > 650% (all sizes) | ||
After Ageing: > 500% (all sizes) | Before Ageing: X-Small 858%, Small 869%, Medium 874%, Large 880%, X-Large 882% | |||
After Ageing: X-Small 841%, Small 848%, Medium 854%, Large 860%, X-Large 862% | Similar | |||
Thickness (ASTM D3578-2019) | To determine the thickness of the gloves | Palm > 0.08 mm, Finger > 0.08 mm (all sizes) | Palm: 0.16mm (all sizes) | |
Finger: 0.22mm (all sizes) | Similar | |||
Powder Free Residue (ASTM D6214) | To determine the residual powder in the gloves | ≤ 2 mg/glove | X-Small 0.21 mg/glove, Small 0.21 mg/glove, Medium 0.22 mg/glove, Large 0.22 mg/glove, X-Large 0.22 mg/glove | Similar |
Protein Content (ASTM D5712) | To determine the extractable protein in the gloves | Max 200 µg/dm² | 43.19 µg/dm² (reported as 43.65 µg/dm² for all sizes in detailed results) | Similar |
Biocompatibility - Primary Skin Irritation (ISO 10993-10:2010) | Not specified in detail, but expected to ensure safe contact with skin. | Under the condition of study, not an irritant | Under the condition of study, not an irritant | Same |
Biocompatibility - Dermal Sensitization (ISO 10993-10:2010) | Not specified in detail, but expected to ensure safe contact with skin. | Under the conditions of the study, not a sensitizer | Under the conditions of the study, not a sensitizer | Same |
Biocompatibility - In vitro cytotoxicity (ISO 10993-5:2009) | Not specified in detail, but expected to ensure no toxic effects on cells. | Under the conditions of the study, cytotoxic (This might be a typo and should be "not cytotoxic", or it implies a specific acceptable level of cytotoxicity.) | Under the conditions of the study, cytotoxic (As above, likely implies within acceptable limits for the device type, or a typo). | Same |
Biocompatibility - Acute Systemic Toxicity Test (ISO 10993-11:2017) | Not specified in detail, but expected to ensure no systemic toxic effects. | Device extracts do not pose a systemic toxicity concern | Device extracts do not pose a systemic toxicity concern | Same |
Water Tightness (ASTM D5151-2019) | To determine the holes in the gloves | AQL 2.5 | Passes AQL 1.5 (AQL 1.5 is a tighter requirement than AQL 2.5, indicating better performance) | Same |
Material-mediated Pyrogenicity (ISO 10993-11:2017 / USP 41) | To ensure the material does not cause fever. | Under the conditions of the study, non-pyrogenic | No data available (for the current device in this specific row, but typically expected for such devices. May be covered by general biocompatibility; the predicate was non-pyrogenic). | ---- |
Intended use/Indication for use | JR MEDIC Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner. | JR MEDIC Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner. | Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner. | Similar |
Material | - | Natural Latex | Natural Latex | Same |
Color | - | Blue | Natural (No color is added) | Different |
Texture | - | Finger Texture | Finger texture | Same |
Size | ASTM D3578-2019 | Small, Medium, Large & X Large | X Small, Small, Medium, Large, X-Large (Added X-Small size) | Similar |
Single Use | Medical Glove Guidance Manual - Labeling | Single Use | Single Use | Same |
Sterile/non sterile | - | Non sterile | Non sterile | Same |
Powder/Powder free | - | Powder free | Powder free | Same |
Study Details (Applicable to this Device Type)
Given this is a 510(k) for examination gloves, the "study" is a series of non-clinical, laboratory tests performed according to specific ASTM and ISO standards, rather than a clinical trial or an AI algorithm performance study.
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample sizes for each specific test (e.g., how many gloves were tested for length, width, or tensile strength). However, it implies that standard testing methodologies as defined by ASTM and ISO were followed. These standards usually specify the number of samples required for each test. The data provenance is from JR Engineering & Medical Technologies (M) SDN.BHD., Malaysia, indicating the tests were conducted by the manufacturer or a contracted lab in Malaysia. The data is non-clinical performance data, reported as part of a premarket notification.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. The "ground truth" for this device is physical and chemical adherence to specific measurable standards (e.g., a glove is X mm long, has Y tensile strength), not a diagnostic interpretation requiring expert medical opinion. The standards themselves, like ASTM D3578-2019, serve as the "ground truth" for performance.
4. Adjudication Method for the Test Set:
Not applicable. This is not a human-interpreted diagnostic task requiring adjudication. The tests are objective measurements against defined criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, this was not done. MRMC studies are typically for diagnostic imaging devices or algorithms where the performance of human readers (with and without AI assistance) on clinical cases is evaluated. This device is an examination glove.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
No, this is not an AI algorithm. Its performance is measured through physical and chemical properties.
7. The Type of Ground Truth Used:
The ground truth used is objective engineering and material science standards (e.g., ASTM D3578-2019, ASTM D5151-2019, ASTM D6124-06, ASTM D5712-95, ISO 10993 series). These are internationally recognized standards for manufacturing and safety.
8. The Sample Size for the Training Set:
Not applicable. This is not an AI device trained on a dataset. The "training" for such a device would be the manufacturing process development and quality control to ensure consistency.
9. How the Ground Truth for the Training Set was Established:
Not applicable, as it's not an AI training set. The "ground truth" for manufacturing would be the specifications and quality control procedures adherence to ensure gloves meet the physical and chemical standards.
Ask a specific question about this device
(186 days)
JR MEDIC Nitrile Examination Gloves Powder Free- Black
Nitrile Medical Examination Gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Nitrile Examination Gloves Powder Free-Black are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are black in color and are powder free and non-sterile.
The provided text describes the regulatory clearance for "JR MEDIC Nitrile Examination Gloves Powder Free- Black" and includes performance data compared to a predicate device. However, it does not describe an AI/ML powered device or a study that evaluates its performance for diagnostic or predictive purposes. This information pertains to medical gloves, which are physical products with established performance standards.
Therefore, many of the requested categories related to AI/ML device testing (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, training set sample size, ground truth for training set) are not applicable to this document.
I will provide the available information from the text regarding the acceptance criteria and performance of the medical glove.
Acceptance Criteria and Device Performance for JR MEDIC Nitrile Examination Gloves Powder Free-Black
The device is a medical glove, and its performance is evaluated against established ASTM standards for physical properties and biocompatibility.
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Standard/Test Method | Acceptance Criteria | Reported Device Performance (JR MEDIC Nitrile Examination Gloves Powder Free-Black) |
---|---|---|---|
Dimensions | ASTM D6319-2019 | Length: Min 230 mm (all sizes) | X-Small: 483 mm, Small: 484 mm, Medium: 485 mm, Large: 486 mm, X-Large: 486 mm (All meet criteria) |
Width: X-Small: 70+/-10 mm, Small: 80+/-10 mm, Medium: 95+/-10 mm, Large: 110+/-10 mm, X-Large: 120+/-10 mm | X-small: 74 mm, Small: 84 mm, Medium: 94 mm, Large: 104 mm, X-Large: 115 mm (All meet criteria) | ||
Physical Properties (Before Aging) | ASTM D6319-2019 | Tensile Strength: Min 14 Mpa (all sizes) | X-Small: 18.44Mpa, Small: 18.49Mpa, Medium: 18.56Mpa, Large: 18.62Mpa, X-Large: 18.64Mpa (All meet criteria) |
Ultimate Elongation: Min 500% (all sizes) | X-Small: 683%, Small: 687%, Medium: 692%, Large: 695%, X-Large: 698% (All meet criteria) | ||
Physical Properties (After Aging) | ASTM D6319-2019 | Tensile Strength: Min 14 Mpa (all sizes) | X-Small: 17.49Mpa, Small: 17.53Mpa, Medium: 17.65Mpa, Large: 17.71Mpa, X-Large: 17.76Mpa (All meet criteria) |
Ultimate Elongation: Min 400% (all sizes) | X-Small: 656%, Small: 660%, Medium: 663%, Large: 666%, X-Large: 669% (All meet criteria) | ||
Thickness | ASTM D6319-2019 | Palm: 0.05 mm min (all sizes), Finger: 0.05 mm min (all sizes) | Palm: 0.19mm (all sizes), Finger: 0.21mm (all sizes) (All meet criteria) |
Powder Residue | ASTM D6124-06 (Reapproved 2017) | ≤2 mg/glove | 0.16 mg/glove (for all sizes X-Small, Small, Medium, Large, X-Large) (Meets criteria) |
Water Tight | ASTM D5151-2019 | Passes AQL-1.5 | Passes AQL 1.5 (Meets criteria) |
Biocompatibility - Primary Skin Irritation | ISO 10993-10:2010(E) | Under the condition of study not an irritant | Under the condition of study not an irritant (Meets criteria) |
Biocompatibility - Dermal Sensitization | ISO 10993-10:2010(E) | Under the conditions of the study not a sensitizer | Under the conditions of the study not a sensitizer (Meets criteria) |
Biocompatibility - In vitro Cytotoxicity | ISO 10993-5:2009(E) | Under the conditions of study non cytotoxic | Under the conditions of the study cytotoxic (Note: This is a discrepancy. The acceptance criteria states "non cytotoxic," but the result is "cytotoxic." Further investigation would be needed on why this was deemed acceptable for clearance.) |
Biocompatibility - Acute Systemic Toxicity | ISO 10993-11:2017(E) | Under the conditions of study the device extracts do not pose a systemic toxicity concern | Under the conditions of study the device extracts do not pose a systemic toxicity concern (Meets criteria) |
Biocompatibility - Material Mediated Pyrogenicity | ISO 10993-11:2017(E) / USP 41 | Under the conditions of the study non pyrogenic | No data available in the table, but the predicate device met this criteria. In the summary table on page 4, it states "No data available" for the subject device but lists "Same" in remarks implying it aligns with the predicate which was "non pyrogenic". |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- The document does not explicitly state the sample sizes for each test. However, tests like ASTM D6319-2019 and D5151-2019 involve specific sampling plans (e.g., AQL for water tightness) where a number of gloves are sampled from a lot.
- The data provenance is from Malyasia, as the applicant (JR Engineering & Medical Technologies (M) Sdn. Bhd.) is located there. The studies are non-clinical tests, likely conducted in a laboratory setting to evaluate the product's physical and biological properties.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This is for a medical glove, not an AI/ML diagnostic device. Ground truth is established by objective measurements against industry standards (ASTM, ISO).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This is for a medical glove, not an AI/ML diagnostic device requiring expert review and adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is for a medical glove, not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is for a medical glove, not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the performance evaluation of the medical glove is based on established international and national standards (such as ASTM D6319-2019, ASTM D5151-2019, ASTM D6124-06, ISO 10993 series). These standards define objective pass/fail criteria for physical properties and biocompatibility.
8. The sample size for the training set
- Not applicable. This is not an AI/ML device requiring a training set.
9. How the ground truth for the training set was established
- Not applicable. This is not an AI/ML device requiring a training set.
Ask a specific question about this device
(79 days)
JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple)
Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple) are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are available in Orange & Purple color and are powder free and are provided non-sterile. These gloves have a shelf life for 3 years.
This document describes the premarket notification (510k) for JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple). This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness. Therefore, the information provided focuses on comparative testing against established standards and a predicate device, rather than a clinical study with human subjects, AI algorithms, or ground truth established by experts.
Here's an analysis of the provided text in relation to your questions:
1. A table of acceptance criteria and the reported device performance
The document provides extensive tables detailing the acceptance criteria (standards) and the device's performance for both the orange and purple gloves. Below is a consolidated table with key physical, chemical, and biological performance characteristics.
Characteristics | Standards / Purpose | Acceptance Criteria | Reported Device Performance (Orange Gloves) | Reported Device Performance (Purple Gloves) |
---|---|---|---|---|
Physical Dimensions | ||||
Length | ASTM D6319-2019 (To determine length) | Min 220mm (XS, S), Min 230mm (M, L, XL) | XS: 236mm, S: 236mm, M: 238mm, L: 238mm, XL: 240mm | XS: 245mm, S: 245mm, M: 246mm, L: 248mm, XL: 248mm |
Width | ASTM D6319-2019 (To determine width) | XS: 70+/-10mm, S: 80+/-10mm, M: 95+/-10mm, L: 110+/-10mm, XL: 120+/-10mm | XS: 74mm, S: 84mm, M: 94mm, L: 104mm, XL: 114mm | XS: 76mm, S: 85mm, M: 95mm, L: 104mm, XL: 115mm |
Thickness (Palm) | ASTM D6319-2019 (To determine thickness) | Palm 0.05 mm min for all sizes | XS, S, M, L, XL: 0.18mm | XS, S, M, L, XL: 0.18mm |
Thickness (Finger) | ASTM D6319-2019 (To determine thickness) | Finger 0.05 mm min for all sizes | XS, S, M, L, XL: 0.20mm | XS, S, M, L, XL: 0.21mm |
Physical Properties | ||||
Tensile Strength (Before Aging) | ASTM D6319-2019 (To determine tensile strength) | 14 Mpa Min for all sizes | XS: 22.77 Mpa, S: 22.80 Mpa, M: 24.46 Mpa, L: 24.51 Mpa, XL: 24.59 Mpa | XS: 22.78 Mpa, S: 22.82 Mpa, M: 24.47 Mpa, L: 24.52 Mpa, XL: 24.58 Mpa |
Tensile Strength (After Aging) | ASTM D6319-2019 (To determine tensile strength) | 14 Mpa Min for all sizes | XS: 20.50 Mpa, S: 20.69 Mpa, M: 21.28 Mpa, L: 21.34 Mpa, XL: 21.36 Mpa | XS: 20.48 Mpa, S: 20.67 Mpa, M: 21.30 Mpa, L: 21.35 Mpa, XL: 21.36 Mpa |
Ultimate Elongation (Before Aging) | ASTM D6319-2019 (To determine ultimate elongation) | 500% Min for all sizes | XS: 885%, S: 886%, M: 888%, L: 891%, XL: 892% | XS: 884%, S: 885%, M: 887%, L: 892%, XL: 893% |
Ultimate Elongation (After Aging) | ASTM D6319-2019 (To determine ultimate elongation) | 400% Min for all sizes | XS: 760%, S: 764%, M: 767%, L: 769%, XL: 772% | XS: 759%, S: 763%, M: 766%, L: 768%, XL: 772% |
Watertight (Holes) | ASTM D5151-2019 (To determine holes) | AQL 2.5 (for subject device K222349), AQL 1.5 (for predicate K192333) Note: AQL 2.5 is a less stringent acceptance criteria than AQL 1.5, which is noted as "Similar" in comparison analysis | Passes AQL 2.5 | Passes AQL 2.5 |
Powder Residue | ASTM D6124-06 (Reapproved 2017) (To determine residual powder) | ≤2 mg/glove Max | XS, S, M, L, XL: 0.16 mg/glove | XS, S, M, L, XL: 0.16 mg/glove |
Biocompatibility | ||||
Primary Skin Irritation | ISO 10993-10 (To determine dermal irritation in Rabbits) | Under the condition of study not an irritant | Under the condition of study not an irritant | Under the condition of study not an irritant |
Dermal Sensitization | ISO 10993-10 (To determine skin sensitization potential in Guinea Pig) | Under the conditions of the study not a sensitizer | Under the conditions of the study not a sensitizer | Under the conditions of the study not a sensitizer |
In vitro Cytotoxicity | ISO 10993-5:2009 (To evaluate cytotoxic potential) | Under the conditions of study non cytotoxic | Under the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern. | Under the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern. |
Acute Systemic Toxicity | ISO 10993-11:2017 (To determine acute systemic toxicity potential) | Under the conditions of study the device extracts do not pose a systemic toxicity concern | Under the conditions of study the device extracts do not pose a systemic toxicity concern | Under the conditions of study the device extracts do not pose a systemic toxicity concern |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes used for each specific test. However, the tests are adherence to ASTM and ISO standards, which define the sample sizes and methodology for such testing. For example, ASTM D5151-2019 (Watertightness) involves sampling a number of gloves from a lot to determine AQL (Acceptance Quality Limit). The testing presumably took place in Malaysia, where the manufacturer is located. The nature of these tests (e.g., physical property measurements, biocompatibility) suggests they are prospective, as new gloves are being manufactured and tested to ensure they meet specifications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This product is a medical glove, not a diagnostic AI device requiring expert interpretation of medical images or data. The "ground truth" for the test set is established by the specifications defined in the relevant ASTM and ISO standards for physical, chemical, and biological properties of examination gloves. Therefore, there are no "experts" in the sense of radiologists or clinicians establishing ground truth for the test set. The validation relies on standard laboratory testing procedures performed by qualified technicians/laboratories.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a study involving human readers or interpretation of complex data where adjudication would be necessary. The testing involves objective measurements against predefined criteria.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/software device or a diagnostic tool that involves human readers. It is a physical medical device (examination gloves).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" in this context is based on standardized test methods and specifications (ASTM and ISO standards). For example:
- Physical Property Ground Truth: Defined by acceptable ranges for length, width, thickness, tensile strength, and elongation as per ASTM D6319-2019.
- Watertightness Ground Truth: Defined by an Acceptance Quality Limit (AQL) as per ASTM D5151-2019.
- Chemical Property Ground Truth: Defined by maximum allowable powder residue as per ASTM D6124-06.
- Biocompatibility Ground Truth: Defined by the absence of irritation, sensitization, or systemic toxicity as per ISO 10993 series, and a determination regarding cytotoxicity.
8. The sample size for the training set
Not applicable. As this is a traditional medical device (gloves) and not an AI/machine learning model, there is no "training set."
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device submission.
Ask a specific question about this device
(131 days)
JR Medic
JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
The tested chemotherapy drugs and their breakthrough detection times are as follows:
Tested Chemotherapy Drug Name & Concentration | Minimum Breakthrough Detection Time |
---|---|
Carmustine (BCNU) (3.3 mg/ml) | 23.8 Minutes |
Carboplatin (10 mg/ml) | >240 Minutes |
Cisplatin (1 mg/ml) | >240 Minutes |
Cyclophosphamide (Cytoxan) (20 mg/ml) | >240 Minutes |
Dacarbazine (10.0 mg/ml) | >240 Minutes |
Doxorubicin HCl (2 mg/ml) | >240 Minutes |
Etoposide (20 mg/ml) | >240 Minutes |
Fluorouracil (50 mg/ml) | >240 Minutes |
Ifosfamide (50 mg/ml) | >240 Minutes |
Methotrexate (25 mg/ml) | >240 Minutes |
Mitomycin C (0.5 mg/ml) | >240 Minutes |
Mitoxantrone (2 mg/ml) | >240 Minutes |
Paclitaxel (6 mg/ml) | >240 Minutes |
Thiotepa (10 mg/ml) | 24.1 Minutes |
Vincristine Sulfate (1 mg/ml) | >240 Minutes |
Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 23.8 Minutes Thiotepa (10 mg/ml) 24.1 Minutes
Warning: Do not use with Carmustine or Thiotepa.
The subject device in 510(K) notification K221157 is JR Medic Blue Latex Examination Powder Free Gloves tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from Natural Rubber Latex, blue color, powder free and non-sterile (as per 21CFR 880.6250, class I).
The subject device meets all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019).
The provided document describes the safety and performance testing of "JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs" to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets general acceptance criteria for a new medical device technology like an AI algorithm.
Therefore, the requested information elements related to AI/algorithm performance, such as sample size for test set with provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and how ground truth was established for the training set are not applicable to this document.
The document focuses on the physical, chemical, and biological properties of examining gloves and their resistance to chemotherapy drugs.
However, I can extract the acceptance criteria and reported device performance from the provided "G. SUMMARY OF NON-CLINICAL PERFORMANCE DATA" section, focusing on the glove's performance criteria.
1. Table of Acceptance Criteria and the Reported Device Performance
Test Method | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To determine the length of the gloves | Min 230 mm for all sizes | Small: 304 mm, Medium: 304 mm, Large: 305 mm, X-Large: 305 mm |
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To determine the width of the gloves | Small: 80+/-10mm, Medium: 95+/-10 mm, Large: 111+/-10 mm, X-Large: 115+/-10 mm | Small: 84 mm, Medium: 94 mm, Large: 105 mm, X-Large: 114 mm |
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To determine the thickness of the gloves | Palm: 0.08 mm min, Finger: 0.08 mm min for all sizes | Palm: 0.31mm (all sizes), Finger: 0.38mm (all sizes) |
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To Determine the physical properties - Tensile strength (Before Ageing) | 18Mpa Min for all sizes | Small: 33.16 Mpa, Medium: 32.93 Mpa, Large: 32.43 Mpa, X-Large: 32.36 Mpa |
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To Determine the physical properties - Tensile strength (After Ageing) | 14Mpa Min for all sizes | Small: 30.78 Mpa, Medium: 30.73 Mpa, Large: 30.23 Mpa, X-Large: 30.16 Mpa |
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To Determine the physical properties - Ultimate Elongation (Before Ageing) | 650% Min for all sizes | Small: 1320%, Medium: 1288%, Large: 1376%, X-Large: 1200% |
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To Determine the physical properties - Ultimate Elongation (After Ageing) | 500% Min for all sizes | Small: 1066%, Medium: 1105%, Large: 1225%, X-Large: 1041% |
ASTM D3578-2019 Standard Specification for Rubber Examination Gloves | To Determine the physical properties - Stress at 500% Elongation (Before Ageing) | 5.5 Mpa Max for all sizes | Small: 5.1 Mpa, Medium: 5.2 Mpa, Large: 5.2 Mpa, X-Large: 5.2 Mpa |
ASTM D5151-2019 Standard Test Method for Detection of Holes in Medical Gloves | To determine the holes in the gloves | AQL 2.5 | Gloves Passes AQL 1.5 |
ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder | To determine the residual powder in the gloves | 2 Mg/Glove Max | Small: 0.21 mg/glove, Medium: 0.22 mg/glove, Large: 0.22 mg/glove, X-Large: 0.22 mg/glove |
ASTM D 5712-95 (Reapproved 2010) Standard Test Method for the Analysis of Protein | To determine the extractable protein in the gloves. | 200 µg/ dm² Max for all sizes | Small: 43.12 µg/ dm², Medium: 41.16 µg/ dm², Large: 42.06 µg/ dm², X-Large: 46.24 µg/ dm² |
ISO 10993-10 Biological Evaluation of Medical Devices: Irritation Test | To determine the potential of the material under test to produce dermal irritation. | Under the condition of study not an irritant | Under the condition of study not an irritant |
ISO 10993-10 Biological Evaluation of Medical Devices: Skin Sensitization Test | To determine the skin sensitization potential of the material. | Under the conditions of the study not a sensitizer | Under the conditions of the study not a sensitizer |
ISO 10993-5:2009 Biological Evaluation of Medical Devices: Cytotoxicity Test | To evaluate the in vitro cytotoxic potential of the test item. | Under the conditions of study non cytotoxic | Under the conditions of the study cytotoxic. |
Acute systemic toxicity | To determine the acute systemic toxicity potential of the test item. | No specific acceptance criteria stated in table, but implication is to show no systemic toxicity | Under the conditions of the study, no systemic toxicity |
Material Mediated Pyrogenicity ISO 10993-11:2017(E) / USP 41 | To determine the pyrogenic potential of the test item extract. | Under the conditions of the study non pyrogenic | Under the conditions of the study non pyrogenic |
ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs | To assess breakthrough time for various chemotherapy drugs. | Varies per drug, generally aiming for longer breakthrough times. Specific minimums were not explicitly stated as "acceptance criteria" but rather as "minimum breakthrough detection time" in the Indications for Use. | Carmustine (BCNU) (3.3 mg/ml): 23.8 Minutes |
Carboplatin (10 mg/ml): >240 Minutes | |||
Cisplatin (1 mg/ml): >240 Minutes | |||
Cyclophosphamide (Cytoxan) (20 mg/ml): >240 Minutes | |||
Dacarbazine (10.0 mg/ml): >240 Minutes | |||
Doxorubicin HCl (2 mg/ml): >240 Minutes | |||
Etoposide (20 mg/ml): >240 Minutes | |||
Fluorouracil (50 mg/ml): >240 Minutes | |||
Ifosfamide (50 mg/ml): >240 Minutes | |||
Methotrexate (25 mg/ml): >240 Minutes | |||
Mitomycin C (0.5 mg/ml): >240 Minutes | |||
Mitoxantrone (2 mg/ml): >240 Minutes | |||
Paclitaxel (6 mg/ml): >240 Minutes | |||
Thiotepa (10 mg/ml): 24.1 Minutes | |||
Vincristine Sulfate (1 mg/ml): >240 Minutes |
Non-Applicable Information for this Document:
The following requested information is not applicable to the provided document as it describes the FDA 510(k) clearance for examination gloves and their resistance to chemotherapy drugs, which is a physical device, not an AI/algorithm-based diagnostic or therapeutic tool.
- Sample size used for the test set and the data provenance: Not applicable - tests are on gloves, not a data set for an algorithm.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable - "ground truth" concept does not apply in this context; instead, it's about objective physical/chemical measurements and biological responses.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(94 days)
Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs
JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examination between patient and examiner.
Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
The tested chemotherapy drugs and their breakthrough detection times are as follows:
Tested Chemotherapy Drug Name & Concentration | Minimum Breakthrough Detection Time |
---|---|
Carmustine (BCNU) (3.3 mg/ml) | 35.0 Minutes |
Carboplatin (10 mg/ml) | >240 Minutes |
Cisplatin (1 mg/ml) | >240 Minutes |
Cyclophosphamide (Cytoxan) (20 mg/ml) | >240 Minutes |
Dacarbazine (10.0 mg/ml) | >240 Minutes |
Doxorubicin HCl (2 mg/ml) | >240 Minutes |
Etoposide (20 mg/ml) | >240 Minutes |
Fluorouracil (50 mg/ml) | >240 Minutes |
Ifosfamide (50 mg/ml) | >240 Minutes |
Methotrexate (25 mg/ml) | >240 Minutes |
Mitomycin C (0.5 mg/ml) | >240 Minutes |
Mitoxantrone (2 mg/ml) | >240 Minutes |
Paclitaxel (6 mg/ml) | >240 Minutes |
Thiotepa (10 mg/ml) | 64.9 Minutes |
Vincristine Sulfate (1 mg/ml) | >240 Minutes |
The subject device in 510(K) notification is JR Medic blue nitrile examination gloves powder free tested for use with Chemotherapy drugs.
The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I).
The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)
I am sorry, but the provided text does not describe a study involving an AI device or algorithm, and therefore all the requested information about acceptance criteria, study design, sample sizes, ground truth, expert involvement, and MRMC studies cannot be extracted.
The document is a 510(k) premarket notification for "JR Medic Blue Nitrile Examination Powder Free Gloves tested for use with chemotherapy drugs." It details the non-clinical testing performed on these medical gloves to demonstrate their safety and effectiveness, specifically their resistance to permeation by chemotherapy drugs, in accordance with established ASTM (American Society for Testing and Materials) standards.
The document focuses on the physical, chemical, and biocompatibility properties of the gloves, not on the performance of a software algorithm or AI.
Ask a specific question about this device
Page 1 of 1