Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device in 510(K) notification is a blue nitrile examination gloves powder free tested for use with Chemotherapy drugs and Fentanyl Citrate. The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I). The device is available in Small, Medium, Large and Extra Large sizes. The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

    AI/ML Overview

    The provided document is a 510(k) premarket notification for "Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate." This document outlines the acceptance criteria and the results of a study to demonstrate the device's performance.

    Here's the breakdown of the information requested based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document provides extensive tables for the performance of the gloves. Rather than recreating the entire tables found on pages 10-14, a summary of key acceptance criteria and reported performance is presented below. Please refer to pages 10-14 of the original document for the complete lists of chemotherapy drugs and specific results.

    Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-2019Length of the glovesMin 220 mm for Size Small & Min 230 mm for all other sizesSmall: 404 mm
    Medium: 405 mm
    Large: 405 mm
    X-Large: 406 mm (All exceed minimums)
    ASTM D6319-2019Width of the glovesSmall: 80+/-10 mm; Medium: 95+/-10 mm; Large: 110+/-10 mm; X-Large: 120+/-10 mmSmall: 84 mm
    Medium: 94 mm
    Large: 105 mm
    X-Large: 115 mm (All within +/- 10mm range)
    ASTM D6319-2019Thickness of the glovesPalm 0.05 mm min; Finger 0.05 mm min for all sizesPalm: 0.19mm (all sizes); Finger: 0.21mm (all sizes) (All exceed minimums)
    ASTM D6319-2019Physical properties - Tensile strength (Before Ageing)$\ge$ 14 Mpa for all sizesSmall: 22.77Mpa; Medium: 24.46Mpa; Large: 24.51Mpa; X-Large: 24.59Mpa (All exceed minimum)
    ASTM D6319-2019Physical properties - Tensile strength (After Ageing)$\ge$ 14 Mpa for all sizesSmall: 20.50Mpa; Medium: 21.81Mpa; Large: 21.95Mpa; X-Large: 22.05Mpa (All exceed minimum)
    ASTM D6319-2019Physical properties - Ultimate Elongation (Before Ageing)$\ge$ 500% for all sizesSmall: 885%; Medium: 888%; Large: 891%; X-Large: 892% (All exceed minimum)
    ASTM D6319-2019Physical properties - Ultimate Elongation (After Ageing)$\ge$ 400% for all sizesSmall: 868%; Medium: 870%; Large: 872%; X-Large: 875% (All exceed minimum)
    ASTM D5151-2019Detection of Holes in Medical GlovesAQL 2.5Passes AQL 1.5 (Better than acceptance criteria, indicating fewer defects)
    ASTM D6124-06Residual Powder on Medical Gloves2 Mg/Glove MaxSmall: 0.16 mg/glove; Medium: 0.16 mg/glove; Large: 0.16 mg/glove; X-Large: 0.16mg/glove (All meet or are below maximum)
    ASTM D6978-05 (2019)Permeation by Chemotherapy Drugs and Fentanyl CitrateSpecified minimum breakthrough detection times for each drug (e.g., >240 Minutes for most)Most chemotherapy drugs and Fentanyl Citrate showed >240 Minutes breakthrough detection time.
    Exceptions: Carmustine (3.3 mg/ml) - 35.0 Minutes; Thiotepa (10 mg/ml) - 64.9 Minutes. Warnings are provided for these two drugs. (See pages 2-3 and 12-13 for full list and values).
    ISO 10993-10Dermal Irritation (Rabbits)Non irritantUnder the condition of study not an irritant (Meets criteria)
    ISO 10993-10Skin Sensitization (Guinea Pig)Non sensitizerUnder the conditions of the study not a sensitizer (Meets criteria)
    ISO 10993-5:2009In vitro CytotoxicityNon cytotoxicUnder the conditions of the study cytotoxic. (Does NOT meet criteria for non-cytotoxic)
    ISO 10993-11:2017Acute Systemic ToxicityNot pose systemic toxicity concernUnder the conditions of study the device extracts do not pose acute systemic toxicity concern (Meets criteria)
    ISO 10993-11:2017 / USP 41Material-Mediated PyrogenicityNot pose material mediated pyrogenicity responseUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response (Meets criteria)

    Note on Cytotoxicity: The device itself is reported as cytotoxic under the tested conditions. This is a critical finding that does not meet a "Non cytotoxic" acceptance criterion. However, the FDA has still cleared the device. This implies that either the level of cytotoxicity observed is deemed acceptable for the intended use and risk-benefit profile of examination gloves, or mitigating factors/information were provided elsewhere (not in this summary). It's important to note this discrepancy.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document lists various ASTM and ISO standards for testing, which define the sample sizes required for each test (e.g., AQL levels for watertightness). However, the specific quantity/sample size used for each individual test conducted by the manufacturer is not explicitly stated in this summary document beyond what's implied by the standard (e.g., AQL 1.5 for watertightness performance is a characteristic, not a sample size).

    • Data Provenance: The manufacturing location of the device owner, JR Engineering & Medical Technologies (M) Sdn. Bhd., is in Malaysia (Lot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1, Pusat Perindustrian 2, 44200 Rasa, Hulu Selangor, Selangor Darul Ehsan, Malaysia). The testing was conducted in accordance with international standards (ASTM, ISO), implying that the data originates from laboratory studies, likely conducted by third-party labs or the manufacturer's own qualified labs, but the specific country of the testing labs is not mentioned.
    • Retrospective or Prospective: The testing described is prospective laboratory performance testing of the device, rather than a retrospective analysis of existing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a medical glove, not an AI or imaging device requiring expert consensus for ground truth. The "ground truth" for this device's performance is established through objective, standardized laboratory testing (e.g., ASTM D6319 for physical properties, ASTM D6978 for chemical permeation, ISO 10993 for biocompatibility). These tests rely on chemical and physical measurements, not human expert interpretation of images or clinical outcomes. Therefore, the concept of "number of experts used to establish ground truth" with qualifications like "radiologist with 10 years of experience" is not applicable to this type of device. The "experts" are the technicians and scientists performing the standardized validation tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As this is not a study involving human interpretation of data where adjudication would be necessary (like in imaging diagnosis), there is no adjudication method described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical examination glove, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device. The tests performed are "standalone" in the sense that they assess the intrinsic material properties and barrier performance of the glove itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device is based on standardized, objective physical, chemical, and biological measurements obtained through laboratory testing, as defined by relevant ASTM and ISO standards. This includes:

    • Physical measurements (e.g., length, width, thickness, tensile strength, elongation).
    • Chemical analysis (e.g., residual powder, breakthrough detection time for drugs).
    • Biological testing (e.g., irritation, sensitization, cytotoxicity, systemic toxicity, pyrogenicity).

    8. The sample size for the training set

    Not applicable. This device is a manufactured product (glove), not an AI algorithm or model that requires a "training set" of data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222348
    Device Name
    JR MEDIC
    Date Cleared
    2022-10-27

    (85 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    JR MEDIC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Latex Examination Powder Free Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color (No color is added) and are powder free and Non-Sterile.

    AI/ML Overview

    The provided document describes the FDA 510(k) premarket notification for "Latex Examination Powder Free Gloves." This is a Class I medical device, which typically does not require a complex AI/algorithm-based study with human readers, ground truth establishment by experts, or MRMC studies. The approval is based on demonstrating substantial equivalence to a predicate device through non-clinical performance data against established industry standards.

    Therefore, many of the requested elements for describing AI acceptance criteria and study methodology (like sample sizes for test/training sets, expert ground truth, adjudication, MRMC studies, standalone algorithm performance, etc.) are not applicable to this type of device and submission.

    The document explicitly states "H. SUMMARY OF CLINICAL TESTING: Not applicable." This reinforces that the submission did not rely on clinical performance data involving human readers or sophisticated AI algorithms.

    However, I can extract the acceptance criteria and performance data as presented for these gloves.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the "Latex Examination Powder Free Gloves" are based on established ASTM (American Society for Testing and Materials) standards for rubber examination gloves. The table below summarizes these criteria and the reported performance of the device (K222348) compared to the predicate device (K192329).

    1. Table of Acceptance Criteria and Reported Device Performance

    Characteristics / Test MethodStandard / PurposeAcceptance Criteria (K192329 Predicate)Reported Device Performance (K222348 - Current Device)Comparison
    Dimensions - Length (ASTM D3578-2019)To determine the length of the gloves> 230 mm (all sizes)> 230 mm (Actual values: X-Small 305 mm, Small 306 mm, Medium 307 mm, Large 308 mm, X-Large 310 mm)Similar
    Dimensions - Width (ASTM D3578-2019)To determine the width of the glovesMin 95+/-10 mm (for medium size)X-Small 76 mm, Small 85 mm, Medium 96 mm, Large 106 mm, X-Large 116 mmSimilar
    Physical Properties - Tensile Strength (ASTM D3578-2019)To Determine the physical propertiesBefore Ageing: > 18 Mpa (all sizes)
    After Ageing: > 14 Mpa (all sizes)Before Ageing: X-Small 22.07 Mpa, Small 22.15 Mpa, Medium 22.22 Mpa, Large 22.30 Mpa, X-Large 22.32 Mpa
    After Ageing: X-Small 18.49 Mpa, Small 18.56 Mpa, Medium 18.67 Mpa, Large 18.74 Mpa, X-Large 18.76 MpaSimilar
    Physical Properties - Ultimate Elongation (ASTM D3578-2019)To Determine the physical propertiesBefore Ageing: > 650% (all sizes)
    After Ageing: > 500% (all sizes)Before Ageing: X-Small 858%, Small 869%, Medium 874%, Large 880%, X-Large 882%
    After Ageing: X-Small 841%, Small 848%, Medium 854%, Large 860%, X-Large 862%Similar
    Thickness (ASTM D3578-2019)To determine the thickness of the glovesPalm > 0.08 mm, Finger > 0.08 mm (all sizes)Palm: 0.16mm (all sizes)
    Finger: 0.22mm (all sizes)Similar
    Powder Free Residue (ASTM D6214)To determine the residual powder in the gloves≤ 2 mg/gloveX-Small 0.21 mg/glove, Small 0.21 mg/glove, Medium 0.22 mg/glove, Large 0.22 mg/glove, X-Large 0.22 mg/gloveSimilar
    Protein Content (ASTM D5712)To determine the extractable protein in the glovesMax 200 µg/dm²43.19 µg/dm² (reported as 43.65 µg/dm² for all sizes in detailed results)Similar
    Biocompatibility - Primary Skin Irritation (ISO 10993-10:2010)Not specified in detail, but expected to ensure safe contact with skin.Under the condition of study, not an irritantUnder the condition of study, not an irritantSame
    Biocompatibility - Dermal Sensitization (ISO 10993-10:2010)Not specified in detail, but expected to ensure safe contact with skin.Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizerSame
    Biocompatibility - In vitro cytotoxicity (ISO 10993-5:2009)Not specified in detail, but expected to ensure no toxic effects on cells.Under the conditions of the study, cytotoxic (This might be a typo and should be "not cytotoxic", or it implies a specific acceptable level of cytotoxicity.)Under the conditions of the study, cytotoxic (As above, likely implies within acceptable limits for the device type, or a typo).Same
    Biocompatibility - Acute Systemic Toxicity Test (ISO 10993-11:2017)Not specified in detail, but expected to ensure no systemic toxic effects.Device extracts do not pose a systemic toxicity concernDevice extracts do not pose a systemic toxicity concernSame
    Water Tightness (ASTM D5151-2019)To determine the holes in the glovesAQL 2.5Passes AQL 1.5 (AQL 1.5 is a tighter requirement than AQL 2.5, indicating better performance)Same
    Material-mediated Pyrogenicity (ISO 10993-11:2017 / USP 41)To ensure the material does not cause fever.Under the conditions of the study, non-pyrogenicNo data available (for the current device in this specific row, but typically expected for such devices. May be covered by general biocompatibility; the predicate was non-pyrogenic).----
    Intended use/Indication for useJR MEDIC Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.JR MEDIC Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.Similar
    Material-Natural LatexNatural LatexSame
    Color-BlueNatural (No color is added)Different
    Texture-Finger TextureFinger textureSame
    SizeASTM D3578-2019Small, Medium, Large & X LargeX Small, Small, Medium, Large, X-Large (Added X-Small size)Similar
    Single UseMedical Glove Guidance Manual - LabelingSingle UseSingle UseSame
    Sterile/non sterile-Non sterileNon sterileSame
    Powder/Powder free-Powder freePowder freeSame

    Study Details (Applicable to this Device Type)

    Given this is a 510(k) for examination gloves, the "study" is a series of non-clinical, laboratory tests performed according to specific ASTM and ISO standards, rather than a clinical trial or an AI algorithm performance study.

    2. Sample Size Used for the Test Set and Data Provenance:
    The document does not explicitly state the sample sizes for each specific test (e.g., how many gloves were tested for length, width, or tensile strength). However, it implies that standard testing methodologies as defined by ASTM and ISO were followed. These standards usually specify the number of samples required for each test. The data provenance is from JR Engineering & Medical Technologies (M) SDN.BHD., Malaysia, indicating the tests were conducted by the manufacturer or a contracted lab in Malaysia. The data is non-clinical performance data, reported as part of a premarket notification.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
    Not applicable. The "ground truth" for this device is physical and chemical adherence to specific measurable standards (e.g., a glove is X mm long, has Y tensile strength), not a diagnostic interpretation requiring expert medical opinion. The standards themselves, like ASTM D3578-2019, serve as the "ground truth" for performance.

    4. Adjudication Method for the Test Set:
    Not applicable. This is not a human-interpreted diagnostic task requiring adjudication. The tests are objective measurements against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
    No, this was not done. MRMC studies are typically for diagnostic imaging devices or algorithms where the performance of human readers (with and without AI assistance) on clinical cases is evaluated. This device is an examination glove.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    No, this is not an AI algorithm. Its performance is measured through physical and chemical properties.

    7. The Type of Ground Truth Used:
    The ground truth used is objective engineering and material science standards (e.g., ASTM D3578-2019, ASTM D5151-2019, ASTM D6124-06, ASTM D5712-95, ISO 10993 series). These are internationally recognized standards for manufacturing and safety.

    8. The Sample Size for the Training Set:
    Not applicable. This is not an AI device trained on a dataset. The "training" for such a device would be the manufacturing process development and quality control to ensure consistency.

    9. How the Ground Truth for the Training Set was Established:
    Not applicable, as it's not an AI training set. The "ground truth" for manufacturing would be the specifications and quality control procedures adherence to ensure gloves meet the physical and chemical standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221169
    Date Cleared
    2022-10-25

    (186 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    JR MEDIC Nitrile Examination Gloves Powder Free- Black

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Medical Examination Gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Nitrile Examination Gloves Powder Free-Black are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are black in color and are powder free and non-sterile.

    AI/ML Overview

    The provided text describes the regulatory clearance for "JR MEDIC Nitrile Examination Gloves Powder Free- Black" and includes performance data compared to a predicate device. However, it does not describe an AI/ML powered device or a study that evaluates its performance for diagnostic or predictive purposes. This information pertains to medical gloves, which are physical products with established performance standards.

    Therefore, many of the requested categories related to AI/ML device testing (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, training set sample size, ground truth for training set) are not applicable to this document.

    I will provide the available information from the text regarding the acceptance criteria and performance of the medical glove.


    Acceptance Criteria and Device Performance for JR MEDIC Nitrile Examination Gloves Powder Free-Black

    The device is a medical glove, and its performance is evaluated against established ASTM standards for physical properties and biocompatibility.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicStandard/Test MethodAcceptance CriteriaReported Device Performance (JR MEDIC Nitrile Examination Gloves Powder Free-Black)
    DimensionsASTM D6319-2019Length: Min 230 mm (all sizes)X-Small: 483 mm, Small: 484 mm, Medium: 485 mm, Large: 486 mm, X-Large: 486 mm (All meet criteria)
    Width: X-Small: 70+/-10 mm, Small: 80+/-10 mm, Medium: 95+/-10 mm, Large: 110+/-10 mm, X-Large: 120+/-10 mmX-small: 74 mm, Small: 84 mm, Medium: 94 mm, Large: 104 mm, X-Large: 115 mm (All meet criteria)
    Physical Properties (Before Aging)ASTM D6319-2019Tensile Strength: Min 14 Mpa (all sizes)X-Small: 18.44Mpa, Small: 18.49Mpa, Medium: 18.56Mpa, Large: 18.62Mpa, X-Large: 18.64Mpa (All meet criteria)
    Ultimate Elongation: Min 500% (all sizes)X-Small: 683%, Small: 687%, Medium: 692%, Large: 695%, X-Large: 698% (All meet criteria)
    Physical Properties (After Aging)ASTM D6319-2019Tensile Strength: Min 14 Mpa (all sizes)X-Small: 17.49Mpa, Small: 17.53Mpa, Medium: 17.65Mpa, Large: 17.71Mpa, X-Large: 17.76Mpa (All meet criteria)
    Ultimate Elongation: Min 400% (all sizes)X-Small: 656%, Small: 660%, Medium: 663%, Large: 666%, X-Large: 669% (All meet criteria)
    ThicknessASTM D6319-2019Palm: 0.05 mm min (all sizes), Finger: 0.05 mm min (all sizes)Palm: 0.19mm (all sizes), Finger: 0.21mm (all sizes) (All meet criteria)
    Powder ResidueASTM D6124-06 (Reapproved 2017)≤2 mg/glove0.16 mg/glove (for all sizes X-Small, Small, Medium, Large, X-Large) (Meets criteria)
    Water TightASTM D5151-2019Passes AQL-1.5Passes AQL 1.5 (Meets criteria)
    Biocompatibility - Primary Skin IrritationISO 10993-10:2010(E)Under the condition of study not an irritantUnder the condition of study not an irritant (Meets criteria)
    Biocompatibility - Dermal SensitizationISO 10993-10:2010(E)Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer (Meets criteria)
    Biocompatibility - In vitro CytotoxicityISO 10993-5:2009(E)Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic (Note: This is a discrepancy. The acceptance criteria states "non cytotoxic," but the result is "cytotoxic." Further investigation would be needed on why this was deemed acceptable for clearance.)
    Biocompatibility - Acute Systemic ToxicityISO 10993-11:2017(E)Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern (Meets criteria)
    Biocompatibility - Material Mediated PyrogenicityISO 10993-11:2017(E) / USP 41Under the conditions of the study non pyrogenicNo data available in the table, but the predicate device met this criteria. In the summary table on page 4, it states "No data available" for the subject device but lists "Same" in remarks implying it aligns with the predicate which was "non pyrogenic".

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document does not explicitly state the sample sizes for each test. However, tests like ASTM D6319-2019 and D5151-2019 involve specific sampling plans (e.g., AQL for water tightness) where a number of gloves are sampled from a lot.
    • The data provenance is from Malyasia, as the applicant (JR Engineering & Medical Technologies (M) Sdn. Bhd.) is located there. The studies are non-clinical tests, likely conducted in a laboratory setting to evaluate the product's physical and biological properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This is for a medical glove, not an AI/ML diagnostic device. Ground truth is established by objective measurements against industry standards (ASTM, ISO).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This is for a medical glove, not an AI/ML diagnostic device requiring expert review and adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is for a medical glove, not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is for a medical glove, not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the performance evaluation of the medical glove is based on established international and national standards (such as ASTM D6319-2019, ASTM D5151-2019, ASTM D6124-06, ISO 10993 series). These standards define objective pass/fail criteria for physical properties and biocompatibility.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device requiring a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222349
    Date Cleared
    2022-10-21

    (79 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple) are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are available in Orange & Purple color and are powder free and are provided non-sterile. These gloves have a shelf life for 3 years.

    AI/ML Overview

    This document describes the premarket notification (510k) for JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple). This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness. Therefore, the information provided focuses on comparative testing against established standards and a predicate device, rather than a clinical study with human subjects, AI algorithms, or ground truth established by experts.

    Here's an analysis of the provided text in relation to your questions:

    1. A table of acceptance criteria and the reported device performance

    The document provides extensive tables detailing the acceptance criteria (standards) and the device's performance for both the orange and purple gloves. Below is a consolidated table with key physical, chemical, and biological performance characteristics.

    CharacteristicsStandards / PurposeAcceptance CriteriaReported Device Performance (Orange Gloves)Reported Device Performance (Purple Gloves)
    Physical Dimensions
    LengthASTM D6319-2019 (To determine length)Min 220mm (XS, S), Min 230mm (M, L, XL)XS: 236mm, S: 236mm, M: 238mm, L: 238mm, XL: 240mmXS: 245mm, S: 245mm, M: 246mm, L: 248mm, XL: 248mm
    WidthASTM D6319-2019 (To determine width)XS: 70+/-10mm, S: 80+/-10mm, M: 95+/-10mm, L: 110+/-10mm, XL: 120+/-10mmXS: 74mm, S: 84mm, M: 94mm, L: 104mm, XL: 114mmXS: 76mm, S: 85mm, M: 95mm, L: 104mm, XL: 115mm
    Thickness (Palm)ASTM D6319-2019 (To determine thickness)Palm 0.05 mm min for all sizesXS, S, M, L, XL: 0.18mmXS, S, M, L, XL: 0.18mm
    Thickness (Finger)ASTM D6319-2019 (To determine thickness)Finger 0.05 mm min for all sizesXS, S, M, L, XL: 0.20mmXS, S, M, L, XL: 0.21mm
    Physical Properties
    Tensile Strength (Before Aging)ASTM D6319-2019 (To determine tensile strength)14 Mpa Min for all sizesXS: 22.77 Mpa, S: 22.80 Mpa, M: 24.46 Mpa, L: 24.51 Mpa, XL: 24.59 MpaXS: 22.78 Mpa, S: 22.82 Mpa, M: 24.47 Mpa, L: 24.52 Mpa, XL: 24.58 Mpa
    Tensile Strength (After Aging)ASTM D6319-2019 (To determine tensile strength)14 Mpa Min for all sizesXS: 20.50 Mpa, S: 20.69 Mpa, M: 21.28 Mpa, L: 21.34 Mpa, XL: 21.36 MpaXS: 20.48 Mpa, S: 20.67 Mpa, M: 21.30 Mpa, L: 21.35 Mpa, XL: 21.36 Mpa
    Ultimate Elongation (Before Aging)ASTM D6319-2019 (To determine ultimate elongation)500% Min for all sizesXS: 885%, S: 886%, M: 888%, L: 891%, XL: 892%XS: 884%, S: 885%, M: 887%, L: 892%, XL: 893%
    Ultimate Elongation (After Aging)ASTM D6319-2019 (To determine ultimate elongation)400% Min for all sizesXS: 760%, S: 764%, M: 767%, L: 769%, XL: 772%XS: 759%, S: 763%, M: 766%, L: 768%, XL: 772%
    Watertight (Holes)ASTM D5151-2019 (To determine holes)AQL 2.5 (for subject device K222349), AQL 1.5 (for predicate K192333) Note: AQL 2.5 is a less stringent acceptance criteria than AQL 1.5, which is noted as "Similar" in comparison analysisPasses AQL 2.5Passes AQL 2.5
    Powder ResidueASTM D6124-06 (Reapproved 2017) (To determine residual powder)≤2 mg/glove MaxXS, S, M, L, XL: 0.16 mg/gloveXS, S, M, L, XL: 0.16 mg/glove
    Biocompatibility
    Primary Skin IrritationISO 10993-10 (To determine dermal irritation in Rabbits)Under the condition of study not an irritantUnder the condition of study not an irritantUnder the condition of study not an irritant
    Dermal SensitizationISO 10993-10 (To determine skin sensitization potential in Guinea Pig)Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    In vitro CytotoxicityISO 10993-5:2009 (To evaluate cytotoxic potential)Under the conditions of study non cytotoxicUnder the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern.Under the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern.
    Acute Systemic ToxicityISO 10993-11:2017 (To determine acute systemic toxicity potential)Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each specific test. However, the tests are adherence to ASTM and ISO standards, which define the sample sizes and methodology for such testing. For example, ASTM D5151-2019 (Watertightness) involves sampling a number of gloves from a lot to determine AQL (Acceptance Quality Limit). The testing presumably took place in Malaysia, where the manufacturer is located. The nature of these tests (e.g., physical property measurements, biocompatibility) suggests they are prospective, as new gloves are being manufactured and tested to ensure they meet specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a medical glove, not a diagnostic AI device requiring expert interpretation of medical images or data. The "ground truth" for the test set is established by the specifications defined in the relevant ASTM and ISO standards for physical, chemical, and biological properties of examination gloves. Therefore, there are no "experts" in the sense of radiologists or clinicians establishing ground truth for the test set. The validation relies on standard laboratory testing procedures performed by qualified technicians/laboratories.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human readers or interpretation of complex data where adjudication would be necessary. The testing involves objective measurements against predefined criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device or a diagnostic tool that involves human readers. It is a physical medical device (examination gloves).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is based on standardized test methods and specifications (ASTM and ISO standards). For example:

    • Physical Property Ground Truth: Defined by acceptable ranges for length, width, thickness, tensile strength, and elongation as per ASTM D6319-2019.
    • Watertightness Ground Truth: Defined by an Acceptance Quality Limit (AQL) as per ASTM D5151-2019.
    • Chemical Property Ground Truth: Defined by maximum allowable powder residue as per ASTM D6124-06.
    • Biocompatibility Ground Truth: Defined by the absence of irritation, sensitization, or systemic toxicity as per ISO 10993 series, and a determination regarding cytotoxicity.

    8. The sample size for the training set

    Not applicable. As this is a traditional medical device (gloves) and not an AI/machine learning model, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221157
    Device Name
    JR Medic
    Date Cleared
    2022-08-30

    (131 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    JR Medic

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Tested Chemotherapy Drug Name & ConcentrationMinimum Breakthrough Detection Time
    Carmustine (BCNU) (3.3 mg/ml)23.8 Minutes
    Carboplatin (10 mg/ml)>240 Minutes
    Cisplatin (1 mg/ml)>240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
    Dacarbazine (10.0 mg/ml)>240 Minutes
    Doxorubicin HCl (2 mg/ml)>240 Minutes
    Etoposide (20 mg/ml)>240 Minutes
    Fluorouracil (50 mg/ml)>240 Minutes
    Ifosfamide (50 mg/ml)>240 Minutes
    Methotrexate (25 mg/ml)>240 Minutes
    Mitomycin C (0.5 mg/ml)>240 Minutes
    Mitoxantrone (2 mg/ml)>240 Minutes
    Paclitaxel (6 mg/ml)>240 Minutes
    Thiotepa (10 mg/ml)24.1 Minutes
    Vincristine Sulfate (1 mg/ml)>240 Minutes

    Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 23.8 Minutes Thiotepa (10 mg/ml) 24.1 Minutes

    Warning: Do not use with Carmustine or Thiotepa.

    Device Description

    The subject device in 510(K) notification K221157 is JR Medic Blue Latex Examination Powder Free Gloves tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from Natural Rubber Latex, blue color, powder free and non-sterile (as per 21CFR 880.6250, class I).

    The subject device meets all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019).

    AI/ML Overview

    The provided document describes the safety and performance testing of "JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs" to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets general acceptance criteria for a new medical device technology like an AI algorithm.

    Therefore, the requested information elements related to AI/algorithm performance, such as sample size for test set with provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and how ground truth was established for the training set are not applicable to this document.

    The document focuses on the physical, chemical, and biological properties of examining gloves and their resistance to chemotherapy drugs.

    However, I can extract the acceptance criteria and reported device performance from the provided "G. SUMMARY OF NON-CLINICAL PERFORMANCE DATA" section, focusing on the glove's performance criteria.


    1. Table of Acceptance Criteria and the Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the length of the glovesMin 230 mm for all sizesSmall: 304 mm, Medium: 304 mm, Large: 305 mm, X-Large: 305 mm
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the width of the glovesSmall: 80+/-10mm, Medium: 95+/-10 mm, Large: 111+/-10 mm, X-Large: 115+/-10 mmSmall: 84 mm, Medium: 94 mm, Large: 105 mm, X-Large: 114 mm
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the thickness of the glovesPalm: 0.08 mm min, Finger: 0.08 mm min for all sizesPalm: 0.31mm (all sizes), Finger: 0.38mm (all sizes)
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Tensile strength (Before Ageing)18Mpa Min for all sizesSmall: 33.16 Mpa, Medium: 32.93 Mpa, Large: 32.43 Mpa, X-Large: 32.36 Mpa
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Tensile strength (After Ageing)14Mpa Min for all sizesSmall: 30.78 Mpa, Medium: 30.73 Mpa, Large: 30.23 Mpa, X-Large: 30.16 Mpa
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Ultimate Elongation (Before Ageing)650% Min for all sizesSmall: 1320%, Medium: 1288%, Large: 1376%, X-Large: 1200%
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Ultimate Elongation (After Ageing)500% Min for all sizesSmall: 1066%, Medium: 1105%, Large: 1225%, X-Large: 1041%
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Stress at 500% Elongation (Before Ageing)5.5 Mpa Max for all sizesSmall: 5.1 Mpa, Medium: 5.2 Mpa, Large: 5.2 Mpa, X-Large: 5.2 Mpa
    ASTM D5151-2019 Standard Test Method for Detection of Holes in Medical GlovesTo determine the holes in the glovesAQL 2.5Gloves Passes AQL 1.5
    ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual PowderTo determine the residual powder in the gloves2 Mg/Glove MaxSmall: 0.21 mg/glove, Medium: 0.22 mg/glove, Large: 0.22 mg/glove, X-Large: 0.22 mg/glove
    ASTM D 5712-95 (Reapproved 2010) Standard Test Method for the Analysis of ProteinTo determine the extractable protein in the gloves.200 µg/ dm² Max for all sizesSmall: 43.12 µg/ dm², Medium: 41.16 µg/ dm², Large: 42.06 µg/ dm², X-Large: 46.24 µg/ dm²
    ISO 10993-10 Biological Evaluation of Medical Devices: Irritation TestTo determine the potential of the material under test to produce dermal irritation.Under the condition of study not an irritantUnder the condition of study not an irritant
    ISO 10993-10 Biological Evaluation of Medical Devices: Skin Sensitization TestTo determine the skin sensitization potential of the material.Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    ISO 10993-5:2009 Biological Evaluation of Medical Devices: Cytotoxicity TestTo evaluate the in vitro cytotoxic potential of the test item.Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic.
    Acute systemic toxicityTo determine the acute systemic toxicity potential of the test item.No specific acceptance criteria stated in table, but implication is to show no systemic toxicityUnder the conditions of the study, no systemic toxicity
    Material Mediated Pyrogenicity ISO 10993-11:2017(E) / USP 41To determine the pyrogenic potential of the test item extract.Under the conditions of the study non pyrogenicUnder the conditions of the study non pyrogenic
    ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy DrugsTo assess breakthrough time for various chemotherapy drugs.Varies per drug, generally aiming for longer breakthrough times. Specific minimums were not explicitly stated as "acceptance criteria" but rather as "minimum breakthrough detection time" in the Indications for Use.Carmustine (BCNU) (3.3 mg/ml): 23.8 Minutes
    Carboplatin (10 mg/ml): >240 Minutes
    Cisplatin (1 mg/ml): >240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml): >240 Minutes
    Dacarbazine (10.0 mg/ml): >240 Minutes
    Doxorubicin HCl (2 mg/ml): >240 Minutes
    Etoposide (20 mg/ml): >240 Minutes
    Fluorouracil (50 mg/ml): >240 Minutes
    Ifosfamide (50 mg/ml): >240 Minutes
    Methotrexate (25 mg/ml): >240 Minutes
    Mitomycin C (0.5 mg/ml): >240 Minutes
    Mitoxantrone (2 mg/ml): >240 Minutes
    Paclitaxel (6 mg/ml): >240 Minutes
    Thiotepa (10 mg/ml): 24.1 Minutes
    Vincristine Sulfate (1 mg/ml): >240 Minutes

    Non-Applicable Information for this Document:

    The following requested information is not applicable to the provided document as it describes the FDA 510(k) clearance for examination gloves and their resistance to chemotherapy drugs, which is a physical device, not an AI/algorithm-based diagnostic or therapeutic tool.

    1. Sample size used for the test set and the data provenance: Not applicable - tests are on gloves, not a data set for an algorithm.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable - "ground truth" concept does not apply in this context; instead, it's about objective physical/chemical measurements and biological responses.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examination between patient and examiner.

    Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Tested Chemotherapy Drug Name & ConcentrationMinimum Breakthrough Detection Time
    Carmustine (BCNU) (3.3 mg/ml)35.0 Minutes
    Carboplatin (10 mg/ml)>240 Minutes
    Cisplatin (1 mg/ml)>240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
    Dacarbazine (10.0 mg/ml)>240 Minutes
    Doxorubicin HCl (2 mg/ml)>240 Minutes
    Etoposide (20 mg/ml)>240 Minutes
    Fluorouracil (50 mg/ml)>240 Minutes
    Ifosfamide (50 mg/ml)>240 Minutes
    Methotrexate (25 mg/ml)>240 Minutes
    Mitomycin C (0.5 mg/ml)>240 Minutes
    Mitoxantrone (2 mg/ml)>240 Minutes
    Paclitaxel (6 mg/ml)>240 Minutes
    Thiotepa (10 mg/ml)64.9 Minutes
    Vincristine Sulfate (1 mg/ml)>240 Minutes
    Device Description

    The subject device in 510(K) notification is JR Medic blue nitrile examination gloves powder free tested for use with Chemotherapy drugs.

    The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I).

    The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

    AI/ML Overview

    I am sorry, but the provided text does not describe a study involving an AI device or algorithm, and therefore all the requested information about acceptance criteria, study design, sample sizes, ground truth, expert involvement, and MRMC studies cannot be extracted.

    The document is a 510(k) premarket notification for "JR Medic Blue Nitrile Examination Powder Free Gloves tested for use with chemotherapy drugs." It details the non-clinical testing performed on these medical gloves to demonstrate their safety and effectiveness, specifically their resistance to permeation by chemotherapy drugs, in accordance with established ASTM (American Society for Testing and Materials) standards.

    The document focuses on the physical, chemical, and biocompatibility properties of the gloves, not on the performance of a software algorithm or AI.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1