K Number
K221626

Validate with FDA (Live)

Date Cleared
2022-12-07

(184 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

Device Description

The subject device in 510(K) notification is a blue nitrile examination gloves powder free tested for use with Chemotherapy drugs and Fentanyl Citrate. The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I). The device is available in Small, Medium, Large and Extra Large sizes. The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

AI/ML Overview

The provided document is a 510(k) premarket notification for "Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate." This document outlines the acceptance criteria and the results of a study to demonstrate the device's performance.

Here's the breakdown of the information requested based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document provides extensive tables for the performance of the gloves. Rather than recreating the entire tables found on pages 10-14, a summary of key acceptance criteria and reported performance is presented below. Please refer to pages 10-14 of the original document for the complete lists of chemotherapy drugs and specific results.

Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance
ASTM D6319-2019Length of the glovesMin 220 mm for Size Small & Min 230 mm for all other sizesSmall: 404 mmMedium: 405 mmLarge: 405 mmX-Large: 406 mm (All exceed minimums)
ASTM D6319-2019Width of the glovesSmall: 80+/-10 mm; Medium: 95+/-10 mm; Large: 110+/-10 mm; X-Large: 120+/-10 mmSmall: 84 mmMedium: 94 mmLarge: 105 mmX-Large: 115 mm (All within +/- 10mm range)
ASTM D6319-2019Thickness of the glovesPalm 0.05 mm min; Finger 0.05 mm min for all sizesPalm: 0.19mm (all sizes); Finger: 0.21mm (all sizes) (All exceed minimums)
ASTM D6319-2019Physical properties - Tensile strength (Before Ageing)$\ge$ 14 Mpa for all sizesSmall: 22.77Mpa; Medium: 24.46Mpa; Large: 24.51Mpa; X-Large: 24.59Mpa (All exceed minimum)
ASTM D6319-2019Physical properties - Tensile strength (After Ageing)$\ge$ 14 Mpa for all sizesSmall: 20.50Mpa; Medium: 21.81Mpa; Large: 21.95Mpa; X-Large: 22.05Mpa (All exceed minimum)
ASTM D6319-2019Physical properties - Ultimate Elongation (Before Ageing)$\ge$ 500% for all sizesSmall: 885%; Medium: 888%; Large: 891%; X-Large: 892% (All exceed minimum)
ASTM D6319-2019Physical properties - Ultimate Elongation (After Ageing)$\ge$ 400% for all sizesSmall: 868%; Medium: 870%; Large: 872%; X-Large: 875% (All exceed minimum)
ASTM D5151-2019Detection of Holes in Medical GlovesAQL 2.5Passes AQL 1.5 (Better than acceptance criteria, indicating fewer defects)
ASTM D6124-06Residual Powder on Medical Gloves2 Mg/Glove MaxSmall: 0.16 mg/glove; Medium: 0.16 mg/glove; Large: 0.16 mg/glove; X-Large: 0.16mg/glove (All meet or are below maximum)
ASTM D6978-05 (2019)Permeation by Chemotherapy Drugs and Fentanyl CitrateSpecified minimum breakthrough detection times for each drug (e.g., >240 Minutes for most)Most chemotherapy drugs and Fentanyl Citrate showed >240 Minutes breakthrough detection time.Exceptions: Carmustine (3.3 mg/ml) - 35.0 Minutes; Thiotepa (10 mg/ml) - 64.9 Minutes. Warnings are provided for these two drugs. (See pages 2-3 and 12-13 for full list and values).
ISO 10993-10Dermal Irritation (Rabbits)Non irritantUnder the condition of study not an irritant (Meets criteria)
ISO 10993-10Skin Sensitization (Guinea Pig)Non sensitizerUnder the conditions of the study not a sensitizer (Meets criteria)
ISO 10993-5:2009In vitro CytotoxicityNon cytotoxicUnder the conditions of the study cytotoxic. (Does NOT meet criteria for non-cytotoxic)
ISO 10993-11:2017Acute Systemic ToxicityNot pose systemic toxicity concernUnder the conditions of study the device extracts do not pose acute systemic toxicity concern (Meets criteria)
ISO 10993-11:2017 / USP 41<151>Material-Mediated PyrogenicityNot pose material mediated pyrogenicity responseUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response (Meets criteria)

Note on Cytotoxicity: The device itself is reported as cytotoxic under the tested conditions. This is a critical finding that does not meet a "Non cytotoxic" acceptance criterion. However, the FDA has still cleared the device. This implies that either the level of cytotoxicity observed is deemed acceptable for the intended use and risk-benefit profile of examination gloves, or mitigating factors/information were provided elsewhere (not in this summary). It's important to note this discrepancy.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document lists various ASTM and ISO standards for testing, which define the sample sizes required for each test (e.g., AQL levels for watertightness). However, the specific quantity/sample size used for each individual test conducted by the manufacturer is not explicitly stated in this summary document beyond what's implied by the standard (e.g., AQL 1.5 for watertightness performance is a characteristic, not a sample size).

  • Data Provenance: The manufacturing location of the device owner, JR Engineering & Medical Technologies (M) Sdn. Bhd., is in Malaysia (Lot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1, Pusat Perindustrian 2, 44200 Rasa, Hulu Selangor, Selangor Darul Ehsan, Malaysia). The testing was conducted in accordance with international standards (ASTM, ISO), implying that the data originates from laboratory studies, likely conducted by third-party labs or the manufacturer's own qualified labs, but the specific country of the testing labs is not mentioned.
  • Retrospective or Prospective: The testing described is prospective laboratory performance testing of the device, rather than a retrospective analysis of existing data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This product is a medical glove, not an AI or imaging device requiring expert consensus for ground truth. The "ground truth" for this device's performance is established through objective, standardized laboratory testing (e.g., ASTM D6319 for physical properties, ASTM D6978 for chemical permeation, ISO 10993 for biocompatibility). These tests rely on chemical and physical measurements, not human expert interpretation of images or clinical outcomes. Therefore, the concept of "number of experts used to establish ground truth" with qualifications like "radiologist with 10 years of experience" is not applicable to this type of device. The "experts" are the technicians and scientists performing the standardized validation tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As this is not a study involving human interpretation of data where adjudication would be necessary (like in imaging diagnosis), there is no adjudication method described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical examination glove, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device. The tests performed are "standalone" in the sense that they assess the intrinsic material properties and barrier performance of the glove itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for this device is based on standardized, objective physical, chemical, and biological measurements obtained through laboratory testing, as defined by relevant ASTM and ISO standards. This includes:

  • Physical measurements (e.g., length, width, thickness, tensile strength, elongation).
  • Chemical analysis (e.g., residual powder, breakthrough detection time for drugs).
  • Biological testing (e.g., irritation, sensitization, cytotoxicity, systemic toxicity, pyrogenicity).

8. The sample size for the training set

Not applicable. This device is a manufactured product (glove), not an AI algorithm or model that requires a "training set" of data.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

December 7, 2022

JR Engineering & Medical Technologies (M) SDN. BHD. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. Ste 114 Aurora, Illinois 60504

Re: K221626

Trade/Device Name: Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: November 10, 2022 Received: November 14, 2022

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221626

Device Name

Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate

Indications for Use (Describe)

Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

The tested chemotherapy drugs and their breakthrough detection times are as follows:

Tested Chemotherapy Drug Name & ConcentrationMinimum Breakthrough Detection Time
Bendamustine (5 mg/ml)>240 Minutes
Bleomycin (15 mg/ml)>240 Minutes
Busulfan (6 mg/ml)>240 Minutes
Carfilzomib (2 mg/ml)>240 Minutes
Cetuximab (2 mg/ml)>240 Minutes
Cladribine (1 mg/ml)>240 Minutes
Cytarabine (100 mg/ml)>240 Minutes
Daunorubicin HCl (5 mg/ml)>240 Minutes
Decitabine (5 mg/ml)>240 Minutes
Docetaxel (10 mg/ml)>240 Minutes
Epirubicin HCl (2 mg/ml)>240 Minutes
Fludarabine (25 mg/ml)>240 Minutes
Fulvestrant (50 mg/ml)>240 Minutes
Gemcitabine (38 mg/ml)>240 Minutes
Idarubicin HCl (1 mg/ml)>240 Minutes
Irinotecan (20 mg/ml)>240 Minutes
Mechlorethamine (1 mg/ml)>240 Minutes
Melphalan (5 mg/ml)>240 Minutes
Oxaliplatin (5 mg/ml)>240 Minutes
Paraplatin (10 mg/ml)>240 Minutes
Pemetrexed (25 mg/ml)>240 Minutes
Raltitrexed (0.5 mg/ml)>240 Minutes
Rituximab (10 mg/ml)>240 Minutes
Topotecan (1 mg/ml)>240 Minutes
Triclosan (2 mg/ml)>240 Minutes
Trisenox (1 mg/ml)>240 Minutes
Velcade (Bortezomib) (1 mg/ml)>240 Minutes
Vidaza (Azacytidine) (25 mg/ml)>240 Minutes
Vinblastine (1 mg/ml)>240 Minutes
Vinorelbine (10 mg/ml)>240 Minutes
Carmustine (BCNU) (3.3 mg/ml)35.0 Minutes
Carboplatin (10 mg/ml)>240 Minutes

{3}------------------------------------------------

Cisplatin (1 mg/ml)>240 Minutes
Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
Dacarbazine (10.0 mg/ml)>240 Minutes
Doxorubicin HCl (2 mg/ml)>240 Minutes
Etoposide (20 mg/ml)>240 Minutes
Fluorouracil (50 mg/ml)>240 Minutes
Ifosfamide (50 mg/ml)>240 Minutes
Methotrexate (25 mg/ml)>240 Minutes
Mitomycin C (0.5 mg/ml)>240 Minutes
Mitoxantrone (2 mg/ml)>240 Minutes
Paclitaxel (6 mg/ml)>240 Minutes
Thiotepa (10 mg/ml)64.9 Minutes
Vincristine Sulfate (1 mg/ml)>240 Minutes

The following hazardous drugs (opioid) and concentration had no breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2ml)

Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 35.0 Minutes Thiotepa (10 mg/ml) 64.9 Minutes

Warning: Do not use with Carmustine or Thiotepa.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY K221626 As required by: 21CFR§807.92(c)

A. APPLICANTINFORMATION

510(K) Owner's NameJR Engineering & Medical Technologies (M) Sdn. Bhd.
AddressLot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1,Pusat Perindustrian 2, 44200 Rasa, Hulu Selangor, SelangorDarul Ehsan, Malaysia.
Phone+603-60572081
Fax+603-60572181
E-mailganeshjrmt@gmail.com
Contact PersonMr. Ganesan Subramaniam
DesignationManaging Director
Contact Number+6012 224 6677
Contact Emailganeshjrmt@gmail.com
Date Submitted10th November 2022

B. DEVICE IDENTIFICATION

Name of the deviceJr Medic Blue Nitrile Examination Gloves Powder Freetested for use with Chemotherapy drugs and FentanylCitrate
Product proprietary or trade nameJR MEDIC
Common or usual nameExam Gloves
Classification namePatient Examination Gloves, Specialty &Fentanyl and other opioid protection glove
Device ClassificationClass-1
Product CodeLZA, LZC, QDO, OPJ
Regulation Number21 CFR 880.6250
Review PanelGeneral Hospital

C. PREDICATE DEVICE

Predicate DeviceHalyard Lavender Nitrile, Powder-free Exam GlovesTested for use with Chemotherapy Drugs and FentanylCitrate
510(K) NumberK202622
Regulatory Class1
Product codeLZC, QDO

D. DESCRIPTION OF THEDEVICE:

The subject device in 510(K) notification is a blue nitrile examination gloves powder free tested for use with Chemotherapy drugs and Fentanyl Citrate.

The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I). The device is

{5}------------------------------------------------

available in Small, Medium, Large and Extra Large sizes.

The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

E. INTENDED USE OF THE DEVICE/INDICATIONS FOR USE:

Ir Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

Tested Chemotherapy Drug Name &Minimum Breakthrough Detection
ConcentrationTime (Minutes)
Bendamustine (5 mg/ml)>240 Minutes
Bleomycin (15 mg/ml)>240 Minutes
Busulfan (6 mg/ml)>240 Minutes
Carfilzomib (2 mg/ml)>240 Minutes
Cetuximab (2 mg/ml)>240 Minutes
Cladribine (1 mg/ml)>240 Minutes
Cytarabine (100 mg/ml)>240 Minutes
Daunorubicin HCl (5 mg/ml)>240 Minutes
Decitabine (5 mg/ml)>240 Minutes
Docetaxel (10 mg/ml)>240 Minutes
Epirubicin HCl (2 mg/ml)>240 Minutes
Fludarabine (25 mg/ml)>240 Minutes
Fulvestrant (50 mg/ml)>240 Minutes
Gemcitabine (38 mg/ml)>240 Minutes
Idarubicin HCl (1 mg/ml)>240 Minutes
Irinotecan (20 mg/ml)>240 Minutes
Mechlorethamine (1 mg/ml)>240 Minutes
Melphalan (5 mg/ml)>240 Minutes
Oxaliplatin (5 mg/ml)>240 Minutes
Paraplatin (10 mg/ml)>240 Minutes
Pemetrexed (25 mg/ml)>240 Minutes
Raltitrexed (0.5 mg/ml)>240 Minutes
Rituximab (10 mg/ml)>240 Minutes
Topotecan (1 mg/ml)>240 Minutes
Triclosan (2 mg/ml)>240 Minutes
Trisenox (1 mg/ml)>240 Minutes
Velcade (Bortezomib) (1 mg/ml)>240 Minutes
Vidaza (Azacytidine) (25 mg/ml)>240 Minutes

The tested chemotherapy drugs and their breakthrough detection times are as follows:

{6}------------------------------------------------

Vinblastine (1 mg/ml)>240 Minutes
Vinorelbine (10 mg/ml)>240 Minutes
Carmustine (BCNU) (3.3 mg/ml)35.0 Minutes
Carboplatin (10 mg/ml)>240 Minutes
Cisplatin (1 mg/ml)>240 Minutes
Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
Dacarbazine (10.0 mg/ml)>240 Minutes
Doxorubicin HCl (2 mg/ml)>240 Minutes
Etoposide (20 mg/ml)>240 Minutes
Fluorouracil (50 mg/ml)>240 Minutes
Ifosfamide (50 mg/ml)>240 Minutes
Methotrexate (25 mg/ml)>240 Minutes
Mitomycin C (0.5 mg/ml)>240 Minutes
Mitoxantrone (2 mg/ml)>240 Minutes
Paclitaxel (6 mg/ml)>240 Minutes
Thiotepa (10 mg/ml)64.9 Minutes
Vincristine Sulfate (1 mg/ml)>240 Minutes

The following hazardous drugs (opioid) and concentration had no breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2ml)

Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 35.0 Minutes Thiotepa (10 mg/ml) 64.9 Minutes

Warning: Do not use with Carmustine (BCNU) & Thiotepa

F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

CHARACTERSTICSSTANDARDSDEVICE PERFORMANCEREMARKS
PREDICATEPROPOSED DEVICE
510(K) Number-K202622K221626
Name of deviceHalyard LavenderNitrile, Powder-free Exam GlovesTested for usewithChemotherapyDrugs andFentanyl CitrateJr MedicBlue NitrileExamination GlovesPowder Free testedfor use withChemotherapy drugsand Fentanyl CitrateSimilar
Product Code-LZA, LZC, QDOLZA, LZC, QDO, OPJSimilar
Intended use /Indications for UseThe HalyardLavender Nitrile,Powder-free ExamGlovesTested for usewithChemotherapyDrugs andJr Medic Blue NitrileExamination GlovesPowder Free testedfor use withChemotherapy drugsand Fentanyl Citrateis a disposable deviceintended for medicalSimilar

{7}------------------------------------------------

Fentanyl Citrateare disposablepurpose that is wornon the examiner's
-------------------------------------------------------------------------------------
CHARACTERSTICSSTANDARDSDEVICE PERFORMANCEREMARKS
PREDICATEK202622PROPOSEDDEVICE
device intendedfor medicalpurpose that isworn on theexaminer's handto preventcontaminationbetween patientand examiner.The gloves weretested for usewith thechemotherapydrugs andFentanyl Citrateas per ASTM D6978-05hand to preventcontaminationbetween patientand examiner.Additionally, thegloves were testedfor use withchemotherapydrugs andFentanyl Citrate inaccordance withASTM D6978-05(2019) StandardPractice forAssessment ofMedical Glove toPermeation byChemotherapyDrugs.
Regulation Number-21 CFR 880.625021 CFR 880.6250Same
Material-NitrileNitrileSame
Color-LavenderBlueDifferent
SizeASTM D6319-2019Extra Small,Small, Medium,Large, Extra LargeSmall, Medium,Large, Extra LargeSimilar
Single UseMedical GloveGuidanceManual- LabelingSingle UseSingle UseSame
Sterile/non sterile-Non sterileNon sterileSame
DimensionsASTM D6319-2019Length:295-325 mmWidthXS- 60-80mmS- 70-90mmM- 85-105mmL- 100-120mmXL- 110-130mmLength > 230 mmWidth Min 95+/-10 mm(formedium size)Similar
CHARACTERSTICSSTANDARDSDEVICE PERFORMANCEREMARKS
PREDICATE DEVICEK202622PROPOSED DEVICEK221626
PhysicalPropertiesASTM D6319-2019Before AgeingTensile Strength$\ge$ 14 MpaUltimate Elongation$\ge$ 500%After AgeingTensile Strength$\ge$ 14 MpaUltimate Elongation$\ge$ 400%Before AgeingTensile Strength> 14 MpaUltimate Elongation>500%After AgeingTensile Strength>14 MpaUltimate Elongation> 400%Similar
ThicknessASTM D6319-2019Palm0.10-0.16 mmFinger0.10-0.19 mmPalm >0.05 mmFinger > 0.05 mmSimilar
Powder FreeResidueASTM D6319-2019$\le$ 2 mg/glove$\le$ 2 mg/gloveSimilar
Watertight(1000 ml)ASTM D5151-2019Passes AQL-2.5Passes AQL-1.5Similar
Label andLabelingFDA LabelrequirementsMeets FDA'srequirementsMeets FDA'srequirementsSame
Bio-compatibilityPrimary SkinIrritation-ISO 10993-10:2010(E)Under the condition ofstudy the deviceextracts were not foundto cause a systemicresponse in the animalmodel.Under the conditionof study not anirritantSimilar
Dermal Sensitization-ISO 10993-10:2010(E)No Data AvailableUnder the conditionsof the study not asensitizer----
In vitrocytotoxicityISO10993-5 :2009(E)No Data AvailableUnder theconditions of thestudy, cytotoxic----
MaterialMediatedPyrogenicityISO 10993-11:2017(E) /USP 41<151>No Data AvailableUnder theconditions ofthe study nomaterial-mediatedpyrogenic----
Acute SystemicToxicity Test ISO10993-11:2017(E)Under the conditions ofthe study, the testarticle was considerednon toxicUnder the conditionof study does notinduce any systemictoxic concernSimilar

{8}------------------------------------------------

{9}------------------------------------------------

Chemotherapy Drugs tested with Minimum Breakthrough Detection Time as tested per ASTM
D6978-05 (2019)
Bendamustine (5 mg/ml)>240 Minutes>240 MinutesSame
Bleomycin (15 mg/ml)>240 Minutes>240 MinutesSame
Busulfan (6 mg/ml)>240 Minutes>240 MinutesSame
Carfilzomib (2 mg/ml)>240 Minutes>240 MinutesSame
Cetuximab (2 mg/ml)>240 Minutes>240 MinutesSame
Cladribine (1 mg/ml)>240 Minutes>240 MinutesSame
Cytarabine (100 mg/ml)>240 Minutes>240 MinutesSame
Daunorubicin HCl (5 mg/ml)>240 Minutes>240 MinutesSame
Decitabine (5 mg/ml)>240 Minutes>240 MinutesSame
Docetaxel (10 mg/ml)>240 Minutes>240 MinutesSame
Epirubicin HCl (2 mg/ml)>240 Minutes>240 MinutesSame
Fludarabine (25 mg/ml)>240 Minutes>240 MinutesSame
Fulvestrant (50 mg/ml)>240 Minutes>240 MinutesSame
Gemcitabine (38 mg/ml)>240 Minutes>240 MinutesSame
Idarubicin HCl (1 mg/ml)>240 Minutes>240 MinutesSame
Irinotecan (20 mg/ml)>240 Minutes>240 MinutesSame
Mechlorethamine (1 mg/ml)>240 Minutes>240 MinutesSame
Melphalan (5 mg/ml)>240 Minutes>240 MinutesSame
Oxaliplatin (5 mg/ml)>240 Minutes>240 MinutesSame
Paraplatin (10 mg/ml)Not Tested>240 MinutesDifferent
Pemetrexed (25 mg/ml)>240 Minutes>240 MinutesSame
Raltitrexed (0.5 mg/ml)>240 Minutes>240 MinutesSame
Rituximab (10 mg/ml)>240 Minutes>240 MinutesSame
Topotecan (1 mg/ml)>240 Minutes>240 MinutesSame
Triclosan (2 mg/ml)>240 Minutes>240 MinutesSame
Trisenox (1 mg/ml)>240 Minutes>240 MinutesSame
Velcade (Bortezomib) (1 mg/ml)>240 Minutes>240 MinutesSame
Vidaza (Azacytidine) (25 mg/ml)>240 Minutes>240 MinutesSame
Vinblastine (1 mg/ml)>240 Minutes>240 MinutesSame
Vinorelbine (10 mg/ml)>240 Minutes>240 MinutesSame
Capecitabine (26 mg/ml)>240 MinutesNot TestedDifferent
Carboplatin (10 mg/ml)>240 Minutes>240 MinutesSame
Cisplatin (1 mg/ml)>240 Minutes>240 MinutesSame
Cyclophosphamide (20 mg/ml)>240 Minutes>240 MinutesSame
Dacarbazine (10 mg/ml)>240 Minutes>240 MinutesSame
Dactinomycin (0.5 mg/ml)>240 MinutesNot TestedDifferent
Doxorubicin HCl (5 mg/ml)>240 Minutes>240 MinutesSame
Etoposide (20 mg/ml)>240 Minutes>240 MinutesSame
5-Fluorouracil (50 mg/ml)>240 Minutes>240 MinutesSame

{10}------------------------------------------------

Ifosfamide (50 mg/ml)>240 Minutes>240 MinutesSame
Leuprolide Acetate Salt ( 5mg/ml)>240 MinutesNot TestedDifferent
Methotrexate (25 mg/ml)>240 Minutes>240 MinutesSame
Mitomycin C (0.5 mg/ml)>240 Minutes>240 MinutesSame
Mitoxantrone (2 mg/ml)>240 Minutes>240 MinutesSame
Paclitaxel (6 mg/ml)>240 Minutes>240 MinutesSame
Temsirolimus (25 mg/ml)>240 MinutesNot TestedDifferent
Vincristine (1 mg/ml)>240 Minutes>240 MinutesSame
Zoledronic Acid (0.8 mg/ml)>240 MinutesNot TestedDifferent
Carmustine (3.3 mg/ml)0.3 Minutes35.0 MinutesDifferent
Thiotepa (10 mg/ml)30.9 Minutes64.9 MinutesDifferent

Hazardous Drugs (opioid) tested with Minimum Breakthrough Detection Time as tested per ASTM D6978-05 (2019)

Fentanyl Citrate (100 mg/ml)>240 Minutes>240 MinutesSame
----------------------------------------------------------------

There are no significant differences between the proposed device and the predicate device and are identical in terms of intended use, materials, design, manufacturing methods & resistance of medical gloves to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTM D6978- 05(2019).

G. NON-CLINICAL TESTING SUMMARY PERFORMANCE DATA

Test MethodPurposeAcceptance CriteriaResult
ASTM D6319-2019Standard Specification forNitrile Examination Glovesfor Medical ApplicationTo determine the lengthof the glovesMin 220 mm for SizeSmall & Min 230 mmfor all other sizesSmall:- 404 mmMedium:- 405 mmLarge:- 405 mmX-Large:- 406 mm

{11}------------------------------------------------

Test MethodPurposeAcceptance CriteriaResult
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo determine thewidth of the glovesSmall:- $80+/-10$ mmMedium:- $95+/-10$ mmLarge:- $110+/-10$ mmX-Large:- $120+/10$ mmSmall:- 84 mmMedium:- 94 mmLarge:- 105 mmX-Large:- 115 mm
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo determine thethickness of theglovesPalm 0.05 mm minFinger 0.05 mm minfor all sizesSize Palm Finger Small 0.19mm 0.21mm Medium 0.19mm 0.21mm Large 0.19mm 0.21mm X-Large 0.19mm 0.21mm
To Determine thephysical properties-Tensile strengthBefore AgeingTensile Strength14Mpa Min for allsizesSize Beforeageing Afterageing Small 22.77Mpa 20.50Mpa Medium 24.46Mpa 21.81Mpa Large 24.51Mpa 21.95Mpa X-Large 24.59Mpa 22.05Mpa
After Ageing TensileStrength 14Mpa Minfor all sizes
ASTM D6319-2019 StandardSpecification for NitrileExamination Gloves forMedical ApplicationTo Determine thephysical properties-Ultimate ElongationBefore AgeingUltimate Elongation500% Min for allsizesSize Beforeageing Afterageing Small 885% 868% Medium 888% 870% Large 891% 872% X-Large 892% 875%
After AgeingUltimate Elongation400% Min for allsizes
Test MethodPurposeAcceptanceCriteriaResult
ASTM D5151-2019 StandardTest Method for Detection ofHoles in Medical GlovesTo determine theholes in the glovesAQL 2.5Gloves Passes AQL 1.5
ASTM D6124-06 (Reapproved2017) Standard Test Methodfor Residual Powder onMedical GlovesTo determine theresidual powder inthe gloves2 Mg/Glove MaxSizeSmall 0.16 mg/gloveMedium 0.16 mg/gloveLarge 0.16 mg/gloveX-Large 0.16mg/glove

{12}------------------------------------------------

PERMEATION TEST RESULT AS PER ASTM D6978

Test Chemotherapy DrugsMinimumBreakthroughDetection Time(Minutes)Avg. SteadyStatePerm. Rate(µg/cm²/Minute)Other Observations
Bendamustine (5 mg/ml)>240 MinutesNASlight swelling & no degradation
Bleomycin (15 mg/ml)>240 MinutesNASlight swelling & no degradation
Busulfan (6 mg/ml)>240 MinutesNASlight swelling & no degradation
Carfilzomib (2 mg/ml)>240 MinutesNASlight swelling & no degradation
Cetuximab (2 mg/ml)>240 MinutesNASlight swelling & no degradation
Cladribine (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Cytarabine (100 mg/ml)>240 MinutesNASlight swelling & no degradation
Daunorubicin HCl (5 mg/ml)>240 MinutesNASlight swelling & no degradation
Decitabine (5 mg/ml)>240 MinutesNASlight swelling & no degradation
Docetaxel (10 mg/ml)>240 MinutesNASlight swelling & no degradation
Epirubicin HCl (2 mg/ml)>240 MinutesNASlight swelling & no degradation
Fludarabine (25 mg/ml)>240 MinutesNASlight swelling & no degradation
Fulvestrant (50 mg/ml)>240 MinutesNASlight swelling & no degradation
Gemcitabine (38 mg/ml)>240 MinutesNASlight swelling & no degradation
Idarubicin HCl (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Irinotecan (20 mg/ml)>240 MinutesNASlight swelling & no degradation
Mechlorethamine (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Melphalan (5 mg/ml)>240 MinutesNASlight swelling & no degradation
Oxaliplatin (5 mg/ml)>240 MinutesNASlight swelling & no degradation
Paraplatin (10 mg/ml)Not TestedNASlight swelling & no degradation
Pemetrexed (25 mg/ml)>240 MinutesNASlight swelling & no degradation
Raltitrexed (0.5 mg/ml)>240 MinutesNASlight swelling & no degradation
Rituximab (10 mg/ml)>240 MinutesNASlight swelling & no degradation
Topotecan (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Triclosan (2 mg/ml)>240 MinutesNASlight swelling & no degradation
Trisenox (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Velcade (Bortezomib)(1 mg/ml)>240 MinutesNASlight swelling & no degradation
Vidaza (Azacytidine)(25 mg/ml)>240 MinutesNASlight swelling & no degradation
Vinblastine (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Vinorelbine (10 mg/ml)>240 MinutesNASlight swelling & no degradation
Carboplatin (10 mg/ml)>240 MinutesNASlight swelling & no degradation
Cisplatin (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Cyclophosphamide (20 mg/ml)>240 MinutesNASlight swelling & no degradation
Dacarbazine (10 mg/ml)>240 MinutesNASlight swelling & no degradation
Doxorubicin HCl (5 mg/ml)>240 MinutesNASlight swelling & no degradation
Etoposide (20 mg/ml)>240 MinutesNASlight swelling & no degradation
5-Fluorouracil (50 mg/ml)>240 MinutesNASlight swelling & no degradation
Ifosfamide (50 mg/ml)>240 MinutesNASlight swelling & no degradation
Test Chemotherapy DrugsMinimumBreakthroughDetection Time(Minutes)Avg. SteadyStatePerm. Rate(µg/cm²/Minute)Other Observations
Methotrexate (25 mg/ml)>240 MinutesNASlight swelling & no degradation
Mitomycin C (0.5 mg/ml)>240 MinutesNASlight swelling & no degradation
Mitoxantrone (2 mg/ml)>240 MinutesNASlight swelling & no degradation
Paclitaxel (6 mg/ml)>240 MinutesNASlight swelling & no degradation
Vincristine (1 mg/ml)>240 MinutesNASlight swelling & no degradation
Carmustine (3.3 mg/ml)35.0 MinutesNASlight swelling & no degradation
Thiotepa (10 mg/ml)64.9 MinutesNASlight swelling & no degradation

{13}------------------------------------------------

H. NON- CLINICAL TESTING BIO-COMPATABILITY DATA

Test MethodPurposeAcceptanceCriteriaResult
ISO 10993-10 BiologicalEvaluation of Medical DevicesTest for Irritation and SkinSensitization. Test done forirritation.To determine the potential ofthe material under test toproduce dermal irritation inRabbitsNon irritantUnder the condition ofstudy not an irritant
ISO 10993-10 BiologicalEvaluation of Medical DevicesTest for Irritation and SkinSensitization. Test done Skinsensitization.To determine the skinsensitization potential of thematerial both in terms ofinduction and elicitation inGuinea Pig.Non sensitizerUnder the conditions ofthe study not asensitizer
ISO 10993-5:2009 biologicalevaluation of medical devices -part 5, tests for in vitrocytotoxicity.To evaluate the in vitrocytotoxic potential of the testitem (both inner and outersurface) Extracts in L-929mouse fibroblasts cells usingelution method.Non cytotoxicUnder the conditions ofthe study cytotoxic.
ISO 10993-11:2017 biologicalevaluation of medical devices -part 11, tests for systemictoxicity.To determine the acutesystemic toxicity potential ofthe test item extracts (bothinside and outer surfaces) inSwiss Albino mice.Not pose systemictoxicity concernUnder the conditionsof study the deviceextracts do not poseacute systemictoxicity concern
MaterialMediatedPyrogenicity ISO 10993-11:2017(E) / USP 41<151>To determine the pyrogenicpotential of the test itemextract following intravenousinjection in New Zealandwhite Rabbits.Not pose materialmediatedpyrogenicityresponseUnder the conditions ofthe study, the device didnot demonstrate amaterial mediatedpyrogenicity response.

{14}------------------------------------------------

I. Clinical Testing Summary

No clinical information is included in this submission

J. CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.