AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

JR MEDIC Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for "Blue Nitrile Examination Gloves Powder Free." It details the device's technical characteristics, non-clinical performance, and comparison to a predicate device to demonstrate substantial equivalence.

This document describes the testing of a medical device (examination gloves), which
Does not involve AI, machine learning, or complex algorithmic performance in its functionality or assessment. The tests performed are
physical, chemical, and biological compatibility tests of the material itself.

Therefore, the requested information regarding acceptance criteria and study proving AI/ML device performance (such as sample size for test/training sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, etc.) is not applicable to this document. The document describes the data and testing for a Class I medical device (gloves) and demonstrates its "substantial equivalence" to an existing predicate device, primarily through non-clinical bench testing and biocompatibility studies.

However, I can extract the acceptance criteria and reported performance for the physical and biological properties of the gloves, as provided in the document.


Acceptance Criteria and Reported Device Performance for Blue Nitrile Examination Gloves Powder Free

Since this is a physical medical device (gloves) and not an AI/ML-driven device, the criteria typically associated with AI/ML performance (e.g., AUC, sensitivity, specificity for a diagnostic algorithm) are not relevant here. Instead, the acceptance criteria relate to the physical properties, chemical composition, and biocompatibility of the gloves, demonstrating they meet established standards for medical examination gloves.

Here's the table of acceptance criteria and reported device performance based on the non-clinical testing summary:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicTest MethodPurposeAcceptance CriteriaReported Device Performance
Physical Properties
LengthASTM D6319-10 (Reapproved 2015)To determine the length of the glovesMin 230 mm for all sizesX-Small: 404 mmSmall: 404 mmMedium: 405 mmLarge: 404 mmX-Large: 404 mm
WidthASTM D6319-10 (Reapproved 2015)To determine the width of the glovesX-small: 70+/-10 mmSmall: 80+/-10 mmMedium: 95+/-10 mmLarge: 110+/-10 mmX-Large: 120+/-10 mmX-small: 74 mmSmall: 84 mmMedium: 94 mmLarge: 105 mmX-Large: 115 mm
ThicknessASTM D6319-10 (Reapproved 2015)To determine the thickness of the glovesPalm 0.05 mm minFinger 0.05 mm min for all sizesPalm:X-Small: 0.20mmSmall: 0.19mmMedium: 0.20mmLarge: 0.20mmX-Large: 0.20mmFinger:X-Small: 0.22mmSmall: 0.21mmMedium: 0.22mmLarge: 0.21mmX-Large: 0.22mm
Tensile StrengthASTM D6319-10 (Reapproved 2015)To determine the physical propertiesBefore Ageing: 14Mpa Min for all sizesAfter Ageing: 14Mpa Min for all sizesBefore Ageing:X-Small: 22.0MpaSmall: 23.0MpaMedium: 25.6MpaLarge: 24.0MpaX-Large: 24.5MpaAfter Ageing:X-Small: 18.5MpaSmall: 20.5MpaMedium: 22.0MpaLarge: 21.0MpaX-Large: 21.7Mpa
Ultimate ElongationASTM D6319-10 (Reapproved 2015)To determine the physical propertiesBefore Ageing: 500% Min for all sizesAfter Ageing: 400% Min for all sizesBefore Ageing:X-Small: 898%Small: 896%Medium: 868%Large: 899%X-Large: 874%After Ageing:X-Small: 872%Small: 861%Medium: 828%Large: 869%X-Large: 853%
Water Tightness (Holes)ASTM D5151-06 (Reapproved 2015)To determine the holes in the glovesAQL 2.5Gloves pass AQL 1.5 (This meets or exceeds the AQL 2.5 criteria, as a lower AQL indicates fewer allowable defects).
Residual PowderASTM D6124-06 (Reapproved 2017)To determine the residual powder in the gloves2 mg/glove MaxX-small: 0.16 mg/gloveSmall: 0.20 mg/gloveMedium: 0.21 mg/gloveLarge: 0.20 mg/gloveX-Large: 0.21 mg/glove
Biocompatibility-
Primary Skin IrritationISO 10993-10:2010(E)To determine the potential to produce dermal irritation in RabbitsUnder the condition of study not an irritantUnder the condition of study not an irritant
Dermal SensitizationISO 10993-10:2010(E)To determine skin sensitization potential in Guinea PigUnder the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
In vitro CytotoxicityISO 10993-5:2009(E)To evaluate in vitro cytotoxic potential in L-929 mouse fibroblasts cellsUnder the conditions of study non cytotoxicUnder the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern. (This indicates an initial result that was followed up to ensure safety).
Acute Systemic ToxicityISO 10993-11:2017(E)To determine acute systemic toxicity potential in swiss Albino miceUnder the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern. (This addresses the concern from the cytotoxicity test).
Material Mediated PyrogenicityISO 10993-11:2017(E) / USP 41 <151>To determine pyrogenic potential in New Zealand white RabbitsUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity responseUnder the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.

Since the device is a simple medical examination glove and not an AI/ML system, the following points regarding AI/ML study design are explicitly not applicable to the data provided in this document:

  1. Sample sizes used for the test set and the data provenance: Not applicable. Performance is based on physical and chemical tests on manufactured gloves, not a "test set" of patient data for AI.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves is defined by adherence to physical standards (ASTM) and biological safety (ISO, USP), not expert annotation of medical images/data.
  3. Adjudication method: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable. This is for performance assessment of diagnostic/AI tools with human readers.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. There is no algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of AI/ML. The "ground truth" here is compliance with established international standards for physical properties and biocompatibility of medical gloves.
  7. The sample size for the training set: Not applicable. There is no training set as there is no AI/ML model.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below it. The logo is simple and professional, and it is easily recognizable.

January 24, 2020

JR Engineering & Medical Technologies (M) SDN.BHD. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114 Aurora, Illinois 60504

Re: K192333

Trade/Device Name: Blue Nitrile Examination Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: December 11, 2019 Received: December 16, 2019

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192333

Device Name

Blue Nitrile Examination Gloves Powder Free

Indications for Use (Describe)

Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 201 Subpart D)
Over-The-Counter Use (21 CFR 201 Subpart C)

__ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR BLUE NITRILE EXAMINATION GLOVES POWDER FREE

510K SUMMARY as required by: 21CFR § 807.92(c) K192333

A. APPLICANT INFORMATION

510(k) Owner's NameJR Engineering & Medical Technologies (M) SDN.BHD.
AddressLot 8 & 10, Jalan Zurah 3, Lot 1 & 3, Jalan Zurah, 3A/1,Pusat Perindustrian 2, 44200 Rasa, Selangor DarulEhsan, Malaysia.
Phone+603-60572081
Fax+603-60572181
E-mailganeshjrmt@gmail.com
Contact PersonMr.Ganesan Subramaniam
DesignationManaging Director
Contact Number+6012 224 6677
Contact Emailganeshjrmt@gmail.com
Date Submitted5th Aug 2019

B. DEVICE IDENTIFICATION

Name of the deviceBlue Nitrile Examination Gloves Powder Free
Product proprietary or trade nameJR MEDIC
Common or usual nameExam Gloves
Classification namePatient Examination Gloves
Device ClassificationClass-1
Product CodeLZA
Regulation Number21 CFR 880.6250
Review PanelGeneral Hospital

C. Predicate Device

Predicate DeviceMCare Blue Nitrile Examination GlovesPowder-free
510( k) NumberK172930
Regulatory Class1
Product codeLZA

D. DESCRIPTION OF THE DEVICE:

JR MEDIC Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250).

{4}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR BLUE NITRILE EXAMINATION GLOVES POWDER FREE

The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.

E.INTENDED USE OF THE DEVICE:

JR MEDIC Blue Nitrile Examination Gloves Powder Free is disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

CHARACTERSTICSSTANDARDSDEVICE PERFORMANCERemarks
PREDICATESUBJECT
510(k) NumberK172930K192333----
Name of deviceMCare Blue NitrileExamination GlovesPowder FreeJR MEDIC BlueNitrile ExaminationGloves Powder Free----
DimensionsASTMD6319-10(Reapproved 2015)Length Min 230 mWidth Min 95+/-10mm(for mediumsize)Length Min 230 mmWidth Min 95+/-10mm(for medium size)Same
Physical PropertiesASTMD6319-10(Reapproved 2015)Before AgingTensile Strengthmin 14 MpaUltimate ElongationMin 500%After AgingTensile Strengthmin 14 MpaUltimate ElongationMin 400%Before AgingTensile Strengthmin 14 MpaUltimate ElongationMin 500%After AgingTensile Strengthmin 14 MpaUltimate ElongationMin 400%Same
ThicknessASTMD6319-10(Reapproved 2015)Palm min 0.05 mmFinger min 0.05 mmPalm min 0.05 mmFinger min 0.05 mmSame
Powder FreeASTMD6319-10≤2 mg/glove≤2 mg/gloveSimilar
BiocompatibilityPrimary SkinIrritation-ISO 10993-10:2010(E)Under the conditionof study not anirritantUnder the conditionof study not anirritantSame
Dermal Sensitization-ISO 10993-10:2010(E)Under the conditionsof the study not asensitizerUnder the conditionsof the study not asensitizerSame
In vitro cytotoxicityISO10993-5 :2009(E)Data Not availableUnder the conditionsof the study,cytotoxic.Additional testingwas performed todetermine if this wasa systemic toxicityconcern----

{5}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR BLUE NITRILE EXAMINATION GLOVES POWDER FREE

POWDER FREE
Acute SystemicToxicity Test ISO10993-11:2017(E)Data Not availableUnder the conditionof study the deviceextracts do not pose asystemic toxicityconcern-----
Material MediatedPyrogenicity ISO10993-11:2017(E) /USP 41<151>Data Not availableUnder the conditionsof the study, thedevice did notdemonstrate amaterial mediatedpyrogenicityresponse.-----
CHARACTERSTICSSTANDARDSDEVICE PERFORMANCERemarks
PREDICATECURRENT
Water Tight (1000 ml)ASTM D5151-06PassesPasses AQL-2.5Similar
Intended useMCare Blue NitrileExaminationGlovesPowderfree is disposabledevicesintendedformedicalpurpose that arewon ontheexaminer's hand topreventcontaminationbetweenpatientand examiner.JR MEDIC Blue NitrileExaminationGlovesPowder Free is disposabledevices intendedformedical purpose that arewon on the examiner'shand topreventcontaminationbetweenpatient and examiner.Similar
MaterialASTMD6910-10(Reapproved2015)NitrileNitrileSame
Color-BlueBlueSame
Texture-Finger TextureFinger textureSame
SizeASTMD6319-10(Reapproved2015)Extra Small, Small,Medium, Large,Extra LargeExtra Small, Small,Medium, Large, ExtraLargeSame
Single UseMedical GloveGuidance Manual- LabelingSingle UseSingle UseSame
Manufacturer(s)-Mercator Medical(Thailand) LTDJR Engineering &Medical Technologies(M) SDN.BHD.Malaysia-----

There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standard D6913-10.

{6}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR BLUE NITRILE EXAMINATION GLOVES POWDER FREE

G. Non-Clinical Testing Summary

PERFORMANCE DATA

Test MethodPurposeAcceptance CriteriaResult
ASTM D6319-10 (Reapproved2015) Standard Specificationfor Nitrile Examination Glovesfor Medical ApplicationTo determine thelength of the glovesMin 230 mm for all sizesX-Small:-404 mmSmall:-404 mmMedium:-405 mmLarge:-404 mmX-Large:-404 mm
ASTM D6319-10 (Reapproved2015) Standard Specificationfor Nitrile Examination Glovesfor Medical ApplicationTo determine thewidth of the glovesX-small:- 70+/-10 mmSmall:-80+/-10 mmMedium:-95+/-10 mmLarge:-110+/-10 mmX-Large:-120+/-10 mmX-small:-74 mmSmall:-84 mmMedium:-94 mmLarge:-105 mmX-Large:-115 mm
Test MethodPurposeAcceptance CriteriaResult
ASTM D6319-10 (Reapproved2015) Standard Specificationfor Nitrile Examination Glovesfor Medical ApplicationTo determine thethickness of theglovesPalm 0.05 mm minFinger 0.05 mm minfor all sizesSize                Palm                FingerX-Small                0.20mm                0.22mmSmall                    0.19mm                0.21mmMedium                0.20mm                0.22mmLarge                    0.20mm                0.21mmX-Large                0.20mm                0.22mm
To Determine thephysical properties-Tensile strengthBefore AgeingTensile Strength14Mpa Min for allsizesAfter AgeingTensile Strength14Mpa Min for allsizesSize                Before                Afterageing                ageingX-Small                22.0Mpa                18.5MpaSmall                    23.0Mpa                20.5MpaMedium                25.6Mpa                22.0 MpaLarge                    24.0Mpa                21.0MpaX-Large                24.5Mpa                21.7Mpa
ASTM D6319-10 (Reapproved2015) Standard Specificationfor Nitrile Examination Glovesfor Medical ApplicationTo Determine thephysical properties-Ultimate ElongationBefore AgeingUltimate Elongation500% Min for allsizesAfter AgeingUltimateElongation 400%Min for all sizesSize                Before                Afterageing                ageingX-Small                898%                    872%Small                    896%                    861%Medium                868%                    828%Large                    899%                    869%X-Large                874%                    853%

{7}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR BLUE NITRILE EXAMINATION GLOVES POWDER FREE

Test MethodPurposeAcceptance CriteriaResult
ASTM D5151-06 (Reapproved2015) Standard Test Methodfor Detection of Holes inMedical GlovesTo determine theholes in the glovesAQL 2.5Gloves Passes AQL 1.5
ASTM D6124-06 (Reapproved2017) Standard Test Methodfor Residual Powder onMedical GlovesTo determine theresidual powder in thegloves2 Mg/Glove MaxSizeX-small 0.16 mg/gloveSmall 0.20 mg/gloveMedium 0.21 mg/gloveLarge 0.20 mg/gloveX-Large 0.21 mg/glove

BIO-COMPATIBILITY DATA

Test MethodPurposeAcceptance CriteriaResult
ISO10993-10 BiologicalEvaluation of Medical DevicesTest for Irritation and SkinSensitization. Test done forirritation.To determine the potential ofthe material under test toproduce dermal irritation inRabbitsUnder the conditionof study not anirritantUnder the condition ofstudy not an irritant
ISO 10993-10 BiologicalEvaluation of Medical DevicesTest for Irritation and SkinSensitization. Test done Skinsensitization.To determine the skinsensitization potential of thematerial both in terms ofinduction and elicitation inGuinea Pig.Under theconditions of thestudy not asensitizerUnder the conditions ofthe study not asensitizer
ISO 10993-5:2009 biologicalevaluation of medical devices -part 5, tests for in vitrocytotoxicity.To evaluate the in vitrocytotoxic potential of the testitem (both inner and outersurface) Extractsin L-929mouse fibroblasts cells usingelution method.Under theconditions of studynon cytotoxicUnder the conditions ofthe study, cytotoxic.Additional testing wasperformed to determineif this was a systemictoxicity concern.
ISO 10993-11:2017 biologicalevaluation of medical devices -part 11, tests for systemictoxicity.To determine the acutesystemic toxicity potential ofthe test item extracts (bothinside and outer surfaces) inswiss Albino mice.Under theconditions of studythe device extractsdo not pose asystemic toxicityconcernUnder the conditionsof study the deviceextracts do not pose asystemic toxicityconcern
USP 41<151> Pyrogen Test.To determine the pyrogenicpotential of the test itemextract following intravenousinjection in New Zealandwhite Rabbits.Under theconditions of thestudy, the devicedid notdemonstrate amaterial mediatedpyrogenicityresponse.Under the conditions ofthe study, the device didnot demonstrate amaterial mediatedpyrogenicity response.

{8}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR BLUE NITRILE EXAMINATION GLOVES POWDER FREE

H. Clinical Testing Summary

Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

I. CONCLUSION

The conclusions drawn from the nonclinical test demonstrate that the subject device in 510(K) submission K192333, Blue Nitrile Examination Gloves Powder Free is as safe, as effective, and performs as well as or better than the legally marketed predicate device K172930.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.