Search Results
Found 14 results
510(k) Data Aggregation
(330 days)
Esperance Distal Access Catheter
The Esperance Distal Access Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
The Esperance® Distal Access Catheter is a single-use, single lumen, variable stiffness, composite catheter. The device includes 5F and 6F catheters with inner diameters of 0.055" and 0.071", respectively, designed with three different working lengths for both sizes: 115 cm, and 131 cm. The device is supplied with a peelable introducer and shaping mandrel. The distal tip of each catheter is visible under fluoroscopy and the distal shaft of each catheter is designed with an external hydrophilic coating to reduce friction during use. The proximal end of each catheter incorporates a strain relief and a standard Luer adapter to facilitate the attachment of accessories. Each catheter has a semi-rigid proximal shaft which transitions into a flexible distal shaft to facilitate the advancement of the catheter in tortuous anatomy.
The Esperance® Distal Access Catheter is a non-active, surgically invasive device intended for short term use within the vasculature.
The provided document is a 510(k) Substantial Equivalence Determination letter from the FDA for a medical device called the "Esperance Distal Access Catheter." It details non-clinical performance testing but does not include information about a study proving device performance against acceptance criteria in the context of an AI/ML-driven device or human-in-the-loop performance.
Specifically, the document states:
- "Animal data were not deemed necessary as substantial equivalence was established based upon successful completion of non-clinical bench testing."
- "A clinical study was not deemed necessary as substantial equivalence was established based upon successful completion of non-clinical bench testing."
Therefore, I cannot extract the requested information points related to a study proving the device meets acceptance criteria for an AI/ML device, as such a study was not performed or detailed in this submission. The tests listed are primarily engineering bench tests for a physical catheter, not AI model validation.
However, I can extract the acceptance criteria for the physical device's performance from the "NON-CLINICAL PERFORMANCE DATA" section.
Here's what I can provide based on the document:
1. A table of acceptance criteria and the reported device performance (for non-clinical bench tests):
Test | Acceptance Criteria (Implied / Stated) | Reported Device Performance |
---|---|---|
Visual Inspection* | Requirements were met. | The device met all pre-defined acceptance criteria. |
Dimensional Inspection* (ID, OD, Overall Length, Working Length, Coating Length, Distal Tip to Marker Band, Hub/Strain Relief Length) | Requirements were met. | The device met all pre-defined acceptance criteria. |
Simulated Use | Performs as intended, meets pre-defined acceptance criteria for: preparation/ease of assembly, introducer sheath interaction, introducer peel away, compatibility with guidewire/microcatheter, lubricity and durability of hydrophilic coating, and kink resistance. | Device performs as intended and met all pre-defined acceptance criteria under simulated use conditions. (Supplemented by additional testing beyond reference device, K211697) |
Physician Validation* | Performs as intended under simulated use conditions. | Device performs as intended under simulated use conditions. |
Delivery and Retrieval Forces | Meets pre-defined acceptance criteria in a vascular model under simulated use conditions. | Met all pre-defined acceptance criteria. (Supplemented by additional testing beyond reference device, K211697) |
Tip Stiffness* | Meets acceptance criteria when deflected on a universal testing machine. | Met acceptance criteria. |
Tip Shaping* | Meets pre-defined acceptance criteria when shaped with shaping mandrel and steam. | Met the pre-defined acceptance criteria. |
System Tensile* (Hub, Shaft, Tip) | Meets the minimum tensile requirement. | The device met all predefined acceptance criteria. |
Elongation to Failure* | Met all pre-defined acceptance criteria, obtained from shaft tensile testing data. | The device met all pre-defined acceptance criteria. |
Torque Strength* | Meets the predefined acceptance criteria in a vascular model. | Met the predefined acceptance criteria. |
Coating Integrity (after particulate testing) | Meets all pre-defined acceptance criteria (inspected pre- and post-insertion and retrieval through a vascular model). | The device coating integrity was inspected pre- and post-insertion and retrieval through a vascular model and met all pre-defined acceptance criteria. (Supplemented by additional testing beyond reference device, K211697) |
Coating Lubricity* | Meets all pre-defined acceptance criteria from frictional forces on a universal testing machine. | Met all pre-defined acceptance criteria. |
Catheter Burst, Leak (Liquid and Air) | Does not leak, burst, and is compatible with accessories per ISO 10555-1 and meets acceptance criteria. | Met acceptance criteria. (Supplemented by additional testing beyond reference device, K211697) |
Kink Resistance* | Meets acceptance criteria for resistance to kinking around bends with clinically relevant radii. | Met acceptance criteria. |
Particulate | Any particulate generated is comparable to cleared comparator devices and meets acceptance criteria. | The device met acceptance criteria. (Supplemented by additional testing beyond reference device, K211697) |
Corrosion Resistance* | Is corrosion resistant per ISO 10555-1. | The catheter is corrosion resistant per ISO 10555-1. |
Radiopacity* | Meets the pre-defined acceptance criteria for marker band visibility under fluoroscopy. | Met the pre-defined acceptance criteria. |
Luer Hub Testing* | Meets the pre-defined specifications per ISO 80369-7 and ISO 80369-20. | Met the pre-defined specifications. |
For the remaining points, the document explicitly states that clinical or animal studies were not deemed necessary beyond the non-clinical bench testing for this specific device (a physical catheter). Therefore, the following information cannot be provided from the given text:
- Sample sized used for the test set and the data provenance: Not applicable for an AI/ML context, as no such test set was described. Bench tests were performed on physical units, but specific sample sizes for each test are not listed.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/ML system is not relevant to this physical device's non-clinical bench testing. "Physician Validation" was mentioned, but no details on the number or qualifications of physicians, nor their role in establishing a ground truth for an AI system.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an algorithm.
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable for an AI/ML system. Ground truth in this context would refer to the reference standards for the physical tests (e.g., precise measurements, material specifications, functional performance during simulated use).
- The sample size for the training set: Not applicable, as there is no AI/ML training set.
- How the ground truth for the training set was established: Not applicable, as there is no AI/ML training set.
Ask a specific question about this device
(266 days)
Distal Access Catheter
The Distal Access Catheter is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
The Distal Access Catheter is a single lumen, flexible, variable stiffness composite catheter which has a luer hub on the proximal end. The catheter shaft has a hydrophilic coating to reduce friction during use. The Distal Access Catheter has a radiopaque marker on the distal tip that is visible under fluoroscopy. The Distal Access Catheter inner lumen can accommodate guidewires up to 0.038 inches in diameter to aid in placement of the catheter system. The catheter has a straight tip. The catheter is offered in various lengths to accommodate physician preferences and anatomical variations. The catheter is provided sterile, nonpyrogenic, and is intended for single use only.
The provided text describes information for a Distal Access Catheter, specifically related to its 510(k) premarket notification for FDA clearance. It details the device's characteristics, intended use, and comparisons to a predicate device, along with non-clinical performance testing (bench testing, biocompatibility, sterility, and shelf-life).
However, the prompt asks for acceptance criteria and a study that proves a device meets acceptance criteria, with specific questions related to AI/Machine Learning (ML) performance, such as:
- A table of acceptance criteria and the reported device performance.
- Sample size used for the test set and data provenance.
- Number of experts used to establish ground truth and their qualifications.
- Adjudication method for the test set.
- If a multi-reader, multi-case (MRMC) comparative effectiveness study was done and its effect size.
- If a standalone (algorithm only) performance study was done.
- The type of ground truth used.
- The sample size for the training set.
- How the ground truth for the training set was established.
The provided FDA 510(k) document is for a physical medical device (catheter) and does NOT describe any AI/ML components or studies. Therefore, it does not contain the information necessary to answer the questions about AI/ML acceptance criteria, test sets, ground truth establishment, or clinical studies involving AI.
Based on the provided document, here is a summary of acceptance criteria and performance for the physical device:
Acceptance Criteria and Device Performance (for the physical catheter)
1. A table of acceptance criteria and the reported device performance
The document lists various non-clinical (bench) tests and biocompatibility tests. The "Result" column effectively indicates whether the device met the pre-defined acceptance criteria for each test.
Test | Test Method Summary | Acceptance Criteria (Implied by Result) & Reported Device Performance |
---|---|---|
Bench Testing | ||
Dimensional Verification | Measured effective length, inner diameter (proximal/distal), max outer diameter. | Device met the dimensional specifications. (Implied acceptance: within specified dimensions) |
Visual Inspection | Inspected for structural or mechanical damage. | Device met the visual specifications. (Implied acceptance: no structural/mechanical damage) |
Simulated Use and Compatibility Testing | Simulated clinical use with compatible devices in a tortuous anatomical bench test model. | Device performed as intended under simulated use conditions. (Implied acceptance: performed as expected) |
Delivery and Retrieval | Measured maximum forces to deliver and retrieve device through tortuous anatomical bench model. | Device met the delivery and retrieval force specifications. (Implied acceptance: forces within specification) |
Tensile Strength | Tested peak tensile force of all bonding points using a universal tensile machine per ISO 10555-1. | Device met acceptance criteria determined with delivery and retrieval force testing. (Implied acceptance: passed) |
Torque Strength | Devices pre-conditioned and torqued to failure inside tortuous anatomical model with distal tip held fixed. | Device torque strength is similar to the predicate. (Implied acceptance: comparable to predicate) |
Burst Pressure | Evaluated per ISO 10555-1:2013 following pre-conditioning. | Device resistance to burst pressure is similar to the predicate. (Implied acceptance: comparable to predicate) |
Kink Resistance | Devices pre-conditioned and bonding points wrapped around varying size pin gauges until failure. | Device kink resistance is similar to the predicate. (Implied acceptance: comparable to predicate) |
Coating Integrity | Hydrophilic coating on catheter surface evaluated after particulate testing for defects. | Device coating integrity is similar to the predicate. (Implied acceptance: comparable to predicate) |
Corrosion Resistance | Evaluated per ISO 10555-1:2013. | Device met corrosion resistance specification. (Implied acceptance: passed) |
Air Leakage | Evaluated per ISO 10555-1:2013 following pre-conditioning. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. (Implied acceptance: passed) |
Liquid Leakage | Evaluated per ISO 10555-1:2013 following pre-conditioning. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. (Implied acceptance: passed) |
Radiopacity | Visibility of distal marker band and catheter body evaluated under X-ray. | Device radiopacity comparable to the predicate device. (Implied acceptance: comparable to predicate) |
Particulate Testing | Test articles tracked multiple times through tortuous anatomical model with ancillary devices; particulates collected. | Particulate generation of the subject device was comparable to the predicate device. (Implied acceptance: comparable to predicate) |
Luer Connector Testing | Evaluated per ISO 80369-7 and ISO 80369-20. | Device met the establish acceptance criteria. (Implied acceptance: passed) |
Tip Flexibility and Tip Buckling Testing | Catheter distal tip evaluated following simulated use testing. | Tip flexibility and tip buckling of the subject device are similar to the predicate device. (Implied acceptance: comparable to predicate) |
Biocompatibility | ||
Cytotoxicity | ISO 10993-5:2009 MTT Method. | The Distal Access Catheter, introducer sheath, and shaping mandrel were non-cytotoxic. (Implied acceptance: non-cytotoxic) |
Skin Sensitization | ISO 10993-10:2021 Guinea Pig Maximization Test. | The Distal Access Catheter, introducer sheath, and shaping mandrel were non-sensitizers. (Implied acceptance: non-sensitizing) |
Irritation | ISO 10993-23:2021 Intracutaneous Reactivity Test. | The Distal Access Catheter, introducer sheath, and shaping mandrel were non-irritant. (Implied acceptance: non-irritant) |
Acute Systemic Toxicity | ISO 10993-11:2017 Acute Systemic Toxicity Study in Mice. | No mortality or evidence of acute systemic toxicity with the Distal Access Catheter, introducer sheath, and shaping mandrel test articles. (Implied acceptance: passed) |
Material-Mediated Pyrogenicity | USP , ISO 10993-11:2017 Rabbit Pyrogen Study. | The Distal Access Catheter, introducer sheath, and shaping mandrel were non-pyrogenic. (Implied acceptance: non-pyrogenic) |
Hemocompatibility (Hemolysis) | ASTM F756-2017, ISO 10993-4:2017 ASTM Hemolysis Study (Direct and Indirect Contact). | The Distal Access Catheter was non-hemolytic (direct and indirect contact). The introducer sheath and shaping mandrel were non-hemolytic (indirect contact). (Implied acceptance: non-hemolytic) |
Hemocompatibility (Complement Activation) | ISO 10993-4:2017 Complement Activation, SC5b-9. | The Distal Access Catheter results were similar to the negative control group. (Implied acceptance: no abnormal activation) |
Hemocompatibility (Partial Thromboplastin | ISO 10993-4:2017, ASTM F2382-2018 Partial Thromboplastin Time Study. | The Distal Access Catheter results were similar to the negative control group. (Implied acceptance: no abnormal effect on clotting) |
Hemocompatibility (Thromboresistance) | ISO 10993-4:2017 In Vivo Vein Thromboresistance Study. | The Distal Access Catheter results were similar to the control and had no evidence of thrombosis. (Implied acceptance: no thrombosis) |
Genotoxicity (Bacterial Reverse Mutation) | ISO 10993-3:2014 Bacterial Reverse Mutation Test. | No backward mutation in Salmonella typhimurium with the Distal Access Catheter test article. (Implied acceptance: non-mutagenic) |
Genotoxicity (Mouse Lymphoma TK Assay) | ISO 10993-3:2014 Mouse Lymphoma TK Assay. | The Distal Access Catheter was non-mutagenic to mouse lymphoma cells (L5178Y TK+/--3.7.2C). (Implied acceptance: non-mutagenic) |
Genotoxicity (Chromosome Aberration Test) | ISO 10993-3:2014 In vitro Mammalian Chromosome Aberration Test. | The Distal Access Catheter results were similar to the negative control group. (Implied acceptance: no abnormal chromosomal aberration) |
Sterility | Sterilization process validated per ISO 11135:2014. | Sterility Assurance Level (SAL) of 10^-4 achieved. EO and ECH residuals below ISO 10993-7:2008 limits. Bacterial Endotoxin Levels ). |
Shelf-life | Baseline and accelerated shelf-life testing. | Demonstrated the device will perform as intended to support the proposed 1-year shelf-life. (Implied acceptance: meets 1-year shelf-life stability) |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- The document describes bench testing and biocompatibility testing only. It does not mention any clinical data, patient data, retrospective, or prospective studies.
- The sample sizes for these tests are not explicitly stated in numerical values (e.g., "n=X units were tested"), but implied by "test articles," "devices," or descriptions like "all bonding points."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- This question is irrelevant for the provided document as it concerns a physical catheter, not an AI/ML product requiring "ground truth" from experts interpreting medical images or data. The "ground truth" for this device's performance is established by objective engineering and biocompatibility test standards and measurements.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- This question is irrelevant for the provided document as it concerns a physical catheter, not an AI/ML product requiring clinical adjudication of interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC or any clinical effectiveness studies were performed or described. The document explicitly states: "No clinical studies were deemed necessary to demonstrate substantial equivalence."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the physical device, the "ground truth" is defined by established engineering standards (e.g., ISO standards), physical measurements, chemical analyses, and biological assays. It's not based on expert clinical consensus or pathology in the context of diagnostic interpretation.
8. The sample size for the training set
- Not applicable. This is a physical device, not an AI/ML product that undergoes "training."
9. How the ground truth for the training set was established
- Not applicable. This is a physical device, not an AI/ML product.
Ask a specific question about this device
(85 days)
Distal Access Catheter
The Distal Access Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The Distal Access Catheter is also indicated for use as a conduit for retrieval devices.
The Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The catheter comes in Straight (with no shaping of the catheter tip) and Curved (curved catheter tip) models based on the distal end. The catheter shaft has a hydrophilic coating on the distal end of the shaft to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end.
This document is a 510(k) summary for a medical device called the "Distal Access Catheter." It describes non-clinical performance testing and biocompatibility testing conducted to demonstrate that the device is substantially equivalent to a predicate device.
1. Table of acceptance criteria and the reported device performance:
Test | Acceptance Criteria (Test Method / Applicable Standard) | Reported Device Performance |
---|---|---|
Dimensional Verification | Verify dimensions using specified measurement tool. | Pass |
Tip Configuration | Visually inspect distal tip at 10x magnification. | Pass |
Surface Integrity | Inspect catheter surface. ISO 10555-1:2013 | Pass |
Tip Buckling | Use a testing apparatus to measure the maximum force required to cause a test sample to buckle at 5 mm, 10 mm and 20 mm from distal tip. | Pass |
Luer Connector Tests | ISO 80369-7: 2021 | Pass |
Flexural Fatigue | After ten runs of the Distal Access Catheter with the accessories in the simulated use model, remove the test sample and inspect the sample for kinks or damage. | Pass |
Trackability | Test the peak tracking force and the compatibility with the ancillary devices during insertion and retraction of the catheter in the simulated use model. | Pass |
Catheter Lubricity and Durability | Use friction tester to measure the frictional force of the test samples when pulled between two clamped pads. Record the peak frictional force over 5 cycles. | Pass |
Particulate and Coating Integrity Test | Measure the size and number of particulates generated during simulated use and conduct visual inspection at 50x and 200x magnification of the outer surface of the catheter after simulated use. | Pass |
Tensile Strength Test | ISO 10555-1:2013 | Pass |
Air Leakage | ISO 10555-1:2013 | Pass |
Liquid Leakage | ISO 10555-1:2013 | Pass |
Torque Resistance | Fix the distal tip of the Distal Access Catheter and rotate the proximal end until the Distal Access Catheter shows kink or other signs of failure. | Pass |
Kink Resistance Test | Visually inspect whether the Distal Access Catheter kinks or not when wrapped around mandrels with diameters 1 mm to 10 mm (in 1 mm increments). | Pass |
Catheter Tip and Lumen Integrity | After simulated use, aspirate test sample for 10 sec using a 60 cc syringe during retrieval device withdrawal. Visually inspect the test sample to verify absence of tip or lumen collapse. | Pass |
Corrosion Resistance Test | ISO 10555-1:2013 | Pass |
Tip Flexibility Test | Use a mechanical force meter to push the tip from left to right at 5 mm, 10 mm, and 20 mm from the distal tip. | Pass |
Radiopacity | ASTM F640-12 | Pass |
Static Burst Pressure | ISO 10555-1:2013 | Pass |
Compatibility Test | The compatibility of the Distal Access Catheter with a marketed introducer sheath, guide catheter, guidewire, micro guidewire, microcatheter, and stent-retriever is evaluated. | Pass |
Flow Rate | Use a flow rate tester to measure the flow rate of the Distal Access Catheter. | Pass |
Biocompatibility Tests | ||
Cytotoxicity | Viability > 70% (ISO 10993-5: 2009, MTT Method) | Pass |
Skin Sensitization | Magnusson & Kligman grades ) | Below 2.15 EU/device |
Shelf-life | Device performs as intended to support the proposed 2-year shelf-life. | Demonstrates 2-year shelf-life |
2. Sample size used for the test set and the data provenance:
The document describes non-clinical (bench and in vitro) and biocompatibility testing. It does not mention a "test set" in the context of patient data or clinical images. Therefore, questions regarding data provenance (country of origin, retrospective/prospective) are not applicable here.
For the non-clinical and biocompatibility tests, the sample sizes are not explicitly stated for each individual test. The document generally states "test sample" or refers to the number of cycles/animals without specifying the exact number for statistical power. The tests are performed on the device itself or its extracted components, not on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This is a submission for a physical medical device (catheter) based on non-clinical and biocompatibility testing, not an AI/software device that requires expert-established ground truth for image interpretation or diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. As noted above, there is no "test set" in the context of clinical data for expert adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI/software device, so MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. This is not an AI/software device. The performance reviewed here is the physical device's standalone performance under various bench and in vitro conditions.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for the non-clinical and biocompatibility tests are defined by the established industry standards and test methods (e.g., ISO 10555-1:2013, ISO 10993 series, ASTM F640-12). The acceptance criteria outlined within these standards serve as the "truth" against which the device's performance is measured. For example, for cytotoxicity, the ground truth is "Viability > 70%."
8. The sample size for the training set:
Not applicable. There is no training set as this is not an AI/software device.
9. How the ground truth for the training set was established:
Not applicable. There is no training set for this type of device.
Ask a specific question about this device
(357 days)
Q Distal Access Catheter
The Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.
The Q Distal Access Catheter is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The proximal portion of the device is a stainless-steel control wire. The distal portion of the device is a coiled/braided catheter shaft with hydrophilic coating along the entire length to reduce friction during use. The Q Distal Access Catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The pin vise may be used to advance the catheter. The syringe and flush tool components may aid in the flushing of the Q Distal Access Catheter.
The medical device described, the Q Distal Access Catheter, is not an AI/ML powered device, therefore, the requested information regarding AI/ML related aspects (such as effect size, standalone performance, training set details, and expert qualifications for ground truth) is not applicable. The provided text details the non-clinical performance testing for a physical medical catheter.
Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Q Distal Access Catheter are derived from various non-clinical performance tests, including bench testing and biocompatibility testing. The "Results" column from the provided tables serves as the reported device performance, indicating that the device "Pass"ed each criterion.
Test Category | Specific Test | Acceptance Criterion (Test Method Summary) | Reported Device Performance (Result) |
---|---|---|---|
Bench Testing | Label Integrity | Device labels were evaluated to ensure labels are intact and legible after environmental conditioning, distribution simulation and aging. | Pass |
Bubble Leak | Device packaging was evaluated to detect gross leaks in packaging. | Pass | |
Pouch Seal Strength | Pouch was evaluated for the mechanical strength of the seal. | Pass | |
Coating Particulate | The number and size of particles generated during simulated use were measured. | Pass | |
Dimensional Verification | Dimensional specifications were verified to ensure device meets all the requirements. | Pass | |
Tip Inspection | Catheter tip was evaluated to ensure the complete retrieval of tip during its removal. | Pass | |
Surface Integrity | Device visually inspected to ensure surface is free of defects that can cause tissue trauma. | Pass | |
Heat Shrink Inspection | To ensure heat shrink identifier remains intact and legible after environmental conditioning, distribution simulation and aging. | Pass | |
Coating Uniformity | To evaluate the devices for uniform coating. | Pass | |
Simulate Use/Compatibility | The catheter performance was evaluated when used in conjunction with other devices used in standard procedures without sustaining damage or kinks and without causing damage to the other devices. | Pass | |
Coating Adhesion | To visually categorize the integrity of coating on the device. | Pass | |
Push/Track | To ensure device is able to be tracked without kink under normal conditions in tortuous anatomy. | Pass | |
Kink Resistance | The distal shaft of the device was evaluated for kink resistance when subjected to a bend radius of 1.0 cm. | Pass | |
Liquid Leak Under Pressure | To ensure device does not exhibit leakage of liquid under a minimum pressure of 44 psi for a minimum of 30 seconds. | Pass | |
Static Burst | To ensure device withstands a minimum pressure of 100 psi prior to burst. | Pass | |
Tensile Strength - Push Wire | To ensure device meets the required tensile strength between the push wire and catheter body: 15 N pull force minimum. | Pass | |
Tensile Strength - Distal Section | To ensure device meets the required tensile strength of the distal section of the shaft: 5N pull force minimum. | Pass | |
Biocompatibility | Cytotoxicity | Biological reactivity Grade 0 on the test article after 48 hrs. incubation with mouse fibroblast. The response on the positive and negative control article extracts confirmed the suitability of the tests. The test article (TA) is considered non-cytotoxic. | Pass |
Hemolysis (Direct and Indirect) | The test article in direct contact with blood had a hemolytic index of 0.3%, and the test article extract had a hemolytic index of 0.2%. Both the test article in direct contact with blood and the test article extract were non-hemolytic. | Pass | |
Sterilization | Ethylene Oxide (EO) Residuals | In accordance with ISO 10993-7. | Pass (Implied by Conclusion) |
Limulus Amoebocyte Lysate (LAL) | In accordance with USP 24 chapter 85 and FDA guidance. | Pass (Implied by Conclusion) | |
Bioburden Testing | In accordance with relevant standards (not explicitly listed but implied alongside EO residuals and LAL). | Pass (Implied by Conclusion) | |
Shelf-life | Shelf-life Verification (3 years) | The device will be labeled with an expiration date of 3 years from the date of sterilization. | Pass (Implied by Conclusion) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific bench or biocompatibility test. The evaluation involved "design verification testing" and "additional biocompatibility testing." The provenance of the data is from internal testing conducted by MIVI Neuroscience, Inc., as part of their 510(k) submission. This is considered retrospective for the purpose of the submission, as the tests were performed to demonstrate compliance of an already designed and manufactured device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This question is not applicable as the Q Distal Access Catheter is a physical medical device, not an AI/ML powered device that relies on expert interpretation for "ground truth" establishment in its performance evaluation. The "ground truth" in this context refers to established engineering specifications, material science standards, and biocompatibility guidelines.
4. Adjudication Method for the Test Set
This question is not applicable for a physical medical device. The "adjudication method" concept is typically used for subjective assessments in clinical or AI/ML studies where human interpretation of ambiguous data is involved. For engineering and scientific testing of a physical device, results are typically objective and determined by established protocols and measurement techniques.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not involve AI assistance for human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not have a standalone algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the Q Distal Access Catheter's performance evaluation is based on a combination of:
- Design Specifications: Predefined engineering requirements and dimensional tolerances for the catheter.
- Established Industry Standards: Such as ISO 11135:2014 for sterilization, ISO 10993-7 for biocompatibility residuals, ASTM F756 for hemolysis, and USP 24 chapter 85 for bacterial endotoxins.
- Visual and Mechanical Inspection Criteria: Standardized methods to assess physical integrity, coating uniformity, and functional performance.
- Physiological Response Criteria: For biocompatibility, established thresholds for cytotoxicity and hemolysis.
8. The Sample Size for the Training Set
This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not involve an AI/ML training set.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable. The Q Distal Access Catheter is a physical medical device and does not involve an AI/ML training set.
Ask a specific question about this device
(250 days)
Distal Access Catheter
The Distal Access Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. It can be used to facilitate introduction of diagnostic agents or therapeutic devices. It is not intended for use in coronary arteries.
The Distal Access Catheter consists of a catheter and accessories (hemostatic valve, introducer sheath, and a shaping mandrel). The catheter is a single-lumen, flexible catheter designed with coil and braid reinforcement. The distal segment is steamshapeable, and a hydrophilic coating is applied for navigation of the catheter through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy.
The catheter body is constructed with a stainless-steel coil (less 2 cm of the catheter length) over the inner lumen liner comprised of polytetrafluoroethylene (PTFE). To provide additional shaft support, a stainless-steel wire braiding has been added over the stainless-steel coil from the proximal end to distal end. A platinum/iridium alloy radiopaque marker band is located at the distal tip of the catheter. An outer layer of varying durometers and lengths of polyamide (PA), polyether block amide (Pebax) and polyurethane (PU) covers the entire catheter body from proximal to distal end, respectively.
A hub (PC) is attached to the proximal end of the catheter. A strain relief made from polyether block amide (Pebax) is placed at the proximal end of the catheter and distal end of the hub. The hub-strain relief provides for the kink resistance for the proximal end. A luer fitting on the catheter hub is used for the attachment of accessories.
The outer surface of the catheter (distal 60 cm) is coated with a hydrophilic coating to reduce friction during navigation in the vasculature.
A shaping mandrel (stainless steel, 80 mm in length) is provided with the catheter to be used by the physician for tip shaping. An introducer sheath (PTFE) is included to facilitate the introduction of the catheter into guide catheters during clinical use. A hemostatic valve is used to connect to the proximal section of the catheter.
The Distal Access Catheter is provided sterile and for single use only. The catheter is placed in a dispenser tube (HDPE) and is placed on a packaging card (HDPE) that is provided in a sterile barrier PET/PE film and Tyvek pouch and placed in a carton box.
This document describes the regulatory clearance for a medical device called a "Distal Access Catheter," not an AI/ML device. Therefore, the questions related to AI/ML device performance, such as sample size for test/training sets, expert ground truth, MRMC studies, and human-in-the-loop performance, are not applicable.
The acceptance criteria and the study that proves the device meets them are based on bench testing (physical and mechanical properties) and biocompatibility testing.
1. Table of Acceptance Criteria and Reported Device Performance
Test Type | Acceptance Criteria (Specification) | Reported Device Performance (Conclusion) |
---|---|---|
Bench Testing | ||
Surface Inspection | External surface free from extraneous matter, process, and surface defects that could cause trauma to vessels. | The surface integrity is suitable for intended clinical use. |
Surface Contamination | Free from surface contaminants from uncured coating surface particulates > 0.02 mm², embedded particulates. Distal tip smooth and tapered. PTFE inner layer not delaminated. | The surface integrity is suitable for intended clinical use. |
Dimensional | Meet specified dimensional requirements (catheter OD, ID, effective length, distal tip length, total accessories length). | The device met the dimensional and physical specifications. |
Distal Tip | Smooth, rounded, tapered, or similarly finished to minimize trauma to vessels. | Distal tip is suitable for intended clinical use. |
Radiodetectability | Tip of the catheter visible under fluoroscopy. | Device radiopacity is suitable for intended clinical use. |
Corrosion Resistance | Metallic components show no signs of corrosion. | Corrosion resistance is suitable for intended clinical use and met requirements of ISO 10555-1. |
Peak Tensile Force | 6F Catheter: ≥15N for distal section and hub/catheter junction. 5F Catheter: ≥10N for distal section and hub/catheter junction. | Peak tensile force is suitable for intended clinical use and met requirements of ISO 10555-1. |
Fluid Leakage | No liquid leakage from hub and catheter shaft at 46 psi for 30 seconds. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. |
Air Leakage | No air leakage at hub into syringe when negative pressure was applied for 15 seconds. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. |
Gauging | Plane of maximum diameter at opening of female conical fitting lies between two limit planes of the gauge. | The device hub met the requirements of ISO 80369-7. |
Separation Force | Mating parts separation force > 35 N. | The device hub met the requirements of ISO 80369-7. |
Unscrewing Torque | Test article luer remains attached after applying unscrewing torque not less than 0.02 N·m for a minimum of 10 seconds. | The device hub met the requirements of ISO 80369-7. |
Stress Cracking | No stress cracks on the test article hub. | The device hub met the requirements of ISO 80369-7. |
Ease of Assembly | Components fit together securely with no resistance between test article luer and reference fitting. | The device hub met the requirements of ISO 80369-7. |
Resistance to Overriding | Test article luer does not override reference fitting threads. | The device hub met the requirements of ISO 80369-7. |
Particulate | Amount of particulate matter during simulated use determined and compared to predicate device. | The amount and size of particles from the subject device was comparable to the predicate device. |
Static Burst Pressure | Evaluate burst pressure under static conditions per ISO 10555-1:2013(E) Annex F. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. |
Lubricity and Durability | Frictional force equivalent to predicate device and less than 0.3N. No coating cracking or separation. | Frictional force is suitable for intended clinical use, no coating cracking or separation, and is equivalent to the predicate device. |
Equipment Interface | Compatible with 0.035" guidewire, ≥ 0.088" ID guide catheter/introducer sheath, ≤ 0.027" OD microcatheters, and common hemostatic valve. | The device is compatible with the accessories as specified. |
Tip Shapeability | Distal tip steam shapeable and equivalent to predicate devices. | Shapeability of the distal tip after steam shaping is equivalent to predicate device. |
Kink Resistance | No kinks when wrapped around pin gauges of clinical use relevant radii. No kinks during simulated use. Equivalent to predicate devices. | The device is resistant to kinking around relevant radii turns. |
Simulated Use | Catheter reaches target locations in tortuous vessel model; delivered and retracted smoothly with 0.035" guidewire; no device damage or defects after simulated use. | Device performs as intended under simulated use conditions. |
Torque Response | Torque response no worse than the predicate device. | Device torque response is equivalent to the predicate device. |
Torque Strength | No catheter breakage after 50 rotations. | Device torque strength is equivalent to the predicate device. |
Pushability/Retractability | Pushability/retractability in tortuous vessel model no worse than the predicate device. | Device pushability/retractability is equivalent to the predicate device. |
Catheter Stiffness | Catheter stiffness equivalent to the predicate devices. | The catheter stiffness is equivalent to the predicate device. |
Catheter Flexural Fatigue | No flexural fatigue following repeated bending or hoop stress. | The catheter flexural fatigue is equivalent to the predicate device. |
Dynamic Burst | Catheter does not burst under dynamic pressure of 300 psi for 30 seconds. | The device met the test acceptance criteria. |
Flow Rate | Catheter withstands manual injection of contrast media and saline at clinically relevant flow rates. | The device can withstand flow rates suitable for intended clinical use. |
Ethylene Oxide Residue | Residual amount of ethylene oxide in single package ≤ 10µg/g. | Ethylene Oxide Residue met the acceptance criteria per ISO 10993-7:2008. |
Sterility | Product shall be sterile. | Sterility of the catheter met the acceptance criteria per ISO 11135:2014. |
Bacterial Endotoxins | Endotoxin content ≤ 2.15 EU/kit. | Bacterial endotoxins met the acceptance criteria per USP . |
Biocompatibility Testing | ||
Cytotoxicity Test (ISO 10993-5) | Cell viability of 100% test article extract shall be acceptable (not cytotoxic). | The cell viability of 100% test article extract was 95.5%. (Conclusion: Non-cytotoxic) |
Skin Sensitization Test (ISO 10993-10) | No significant evidence of skin sensitization. | The positive rate of the test article was 0%. (Conclusion: No significant evidence of skin sensitization) |
Intracutaneous Reactivity Test (ISO 10993-10) | Non-irritant. Final test article score acceptable. | The final test article score was calculated to be 0. (Conclusion: Non-irritant) |
Acute Systemic Toxicity Test (ISO 10993-11) | No significant evidence of systemic toxicity. Body weight data acceptable and equivalent between groups. | Body weight data were acceptable and equivalent between the corresponding test and control treatment groups. (Conclusion: No significant evidence of systemic toxicity) |
Hemolytic Properties Test (ASTM F756) | No influence on hemolytic properties. | The hemolysis index was 0.44% (direct contact) and 0.00 (indirect contact). (Conclusion: No influence on hemolytic properties) |
Partial Thromboplastin Time (PTT) Test (ISO 10993-4) | No effect on PTT—no significant differences between sample group and negative group. | No significant differences between the sample group and the negative group. (Conclusion: No effect on PTT) |
In Vivo Thrombogenicity Test (ISO 10993-4) | Equivalent to the control article—no significant differences between test and control articles. | No significant differences between the test and control articles. (Conclusion: Equivalent to the control article) |
Pyrogen Test (ISO 10993-11) | Non-pyrogenic—no rabbit shows an individual rise in temperature of 0.5 °C or more. | No rabbit shows an individual rise in temperature of 0.5 °C or more. (Conclusion: Non-pyrogenic) |
Complement Activity (C3a, SC5b-9) Test (ISO 10993-4) | Equivalent to the negative control group—no significant difference between sample group and negative control group. | No significant difference between the sample group and negative control group. (Conclusion: Equivalent to the negative control group) |
Study Proving Device Meets Acceptance Criteria:
The device met its acceptance criteria through a series of bench tests and biocompatibility evaluations.
Bench Test Study:
- Study Design: The bench testing of the Distal Access Catheter was performed using the applicable sections of the ISO 10555-1 international standard for sterile, single-use intravascular catheters. This standard specifies requirements for sterile, single-use intravascular catheters. The tests covered various physical and mechanical properties crucial for safe and effective device function.
- Methodology: The document lists 24 distinct bench tests (e.g., Surface Inspection, Dimensional Verification, Corrosion Resistance, Peak Tensile Force, Fluid Leakage, Kink Resistance, Simulated Use, Dynamic Burst, Flow Rate, etc.). Each test had specific acceptance criteria derived from the ISO standard or established engineering specifications for the device. The results consistently concluded that the device was "suitable for intended clinical use," "met requirements," or was "equivalent to the predicate device."
- Conclusion: The bench testing demonstrated that "the in vitro behavior of the device is well characterized within design specifications."
Biocompatibility Study:
- Study Design: The biocompatibility evaluation was conducted in accordance with FDA's biocompatibility guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"."
- Methodology: The biological tests were conducted in compliance with the Good Laboratory Practice (GLP) Regulation 21 CFR 58. These tests included Cytotoxicity, Skin Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Hemolytic Properties, Partial Thromboplastin Time (PTT), In Vivo Thrombogenicity, Pyrogen, and Complement Activity tests. Each test had specific criteria for acceptable biological response. The conclusions for these tests consistently indicated that the device was non-cytotoxic, non-sensitizing, non-irritant, non-systemically toxic, had no effect on hemolytic properties or PTT, was equivalent to control/predicate for thrombogenicity and complement activity, and was non-pyrogenic.
- Conclusion: The biocompatibility testing confirmed the device material safety for its intended biological contact.
The following information is not applicable (N/A) as this is a traditional medical device (catheter) and not an AI/ML powered device.
- Sample size used for the test set and the data provenance: N/A (for AI/ML test sets). These are physical and biological tests, not data-driven evaluations.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (for AI/ML expert annotation). Ground truth for physical/biological tests is based on objective measurements against established standards.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: N/A (for AI/ML expert adjudication).
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (not an AI-assisted device).
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A (not an AI algorithm).
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): For this device, the "ground truth" or acceptance standard is derived from established international medical device standards (e.g., ISO 10555-1, ISO 80369-7, ISO 10993 series, USP ) and internal design specifications, verified through objective bench and biological testing.
- The sample size for the training set: N/A (not an AI/ML device; no training set in this context).
- How the ground truth for the training set was established: N/A (not an AI/ML device).
Ask a specific question about this device
(136 days)
MIVI Q Distal Access Catheter
The MIVI Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.
The MIVI Q Distal Access Catheter (Q Catheter) is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. The proximal portion of the catheter is a stainless-steel control (push) wire. The Q Catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The Q Docking Station may be used to facilitate insertion and extraction of the Q Catheter through a hemostasis valve attached to the 8F guide catheter/6F guide sheath. The pin vise may be used to advance the catheter.
Here's an analysis of the provided text regarding the acceptance criteria and study for the MIVI Q Distal Access Catheter:
The document provided is a 510(k) Summary for the MIVI Q Distal Access Catheter, which is a premarket notification to the FDA for a medical device seeking substantial equivalence to a predicate device. This type of submission focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate, rather than proving novel effectiveness.
Given the nature of a 510(k) submission for a catheter, the "acceptance criteria" and "device performance" in this context refer to the ability of the device to meet specified engineering and biocompatibility standards, and to perform its intended function without significant differences from the predicate. There is no clinical study in the traditional sense (e.g., patient trials with efficacy endpoints) described for this device, as it's a substantially equivalent determination based on non-clinical testing.
Here's the information extracted and organized according to your request:
Acceptance Criteria and Device Performance for MIVI Q Distal Access Catheter
1. Table of Acceptance Criteria and Reported Device Performance
For Bench Testing (Non-Clinical Performance):
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Dimensional Verification | The device must meet dimensional specifications. | Pass |
System Introduction (Simulated Use) | The device must function as intended. | Pass |
Liquid Leakage under Pressure | The device must hold a hydrostatic pressure. | Pass |
Tensile Strength | Tensile strength pull force minimum must be met. | Pass |
For Biocompatibility Testing:
Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Cytotoxicity | Non-toxic to cells (determined by specific extract and cell line). | PASS |
Irritation or Intracutaneous Reactivity | No significant irritation or reactivity in vivo (animal model). | PASS |
Sensitization | No significant sensitization reaction (animal model). | PASS |
Acute Systemic Toxicity | No acute systemic toxicity (animal model). | PASS |
Pyrogenicity (Material-mediated) | Non-pyrogenic (no fever-inducing substances in vivo). | PASS |
Hemolysis | No significant hemolysis (destruction of red blood cells) in vitro. | PASS |
Unactivated Partial Thromboplastin Time | No significant impact on blood clotting time (in vitro). | PASS |
Complement Activation | No significant complement activation (immune response) in vitro. | PASS |
For Sterilization and Shelf Life:
Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Sterilization | Confirmed sterility of the final packaging configuration and method. | Met specification |
Shelf Life and Expiration Dating | Maintain performance and integrity for the specified period. | 3 years (same as predicate) |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not specify the exact sample sizes used for each individual non-clinical (bench and biocompatibility) test. Such details are typically found in the full test reports, not in the summary document.
Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting (in vitro or in animal models), not on human subjects. Therefore, there is no country of origin for human patient data. The provenance of the data is the laboratory where the tests were performed. The studies are prospective in the sense that they were designed and executed to test the characteristics of the MIVI Q Distal Access Catheter.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This question is not applicable to the provided document. The "ground truth" in this context is established by objective measurements and standardized test methods (e.g., ISO standards for biocompatibility, engineering specifications for bench tests), not by expert consensus on clinical data. No human "experts" were used to establish ground truth for this type of non-clinical testing.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies, particularly for interpreting imaging or clinical outcomes where there might be inter-reader variability. For non-clinical bench and biocompatibility tests, results are typically objective (e.g., a tensile strength value, a cytotoxicity rating) and do not require expert adjudication in this manner.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The MIVI Q Distal Access Catheter is a physical medical device (a catheter), not an AI algorithm, and therefore does not involve human readers interpreting output from an AI system. The submission is for substantial equivalence of a physical device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. As explained above, this device is a physical catheter, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical tests described:
- Bench Testing: Ground truth is defined by engineering specifications and objective physical measurements (e.g., specified dimensions, force thresholds, pressure integrity).
- Biocompatibility Testing: Ground truth is established by standardized biological assays and observation in animal models and in vitro systems, as per recognized international standards (e.g., ISO 10993 series). The results are objective measurements against defined pass/fail criteria.
8. The sample size for the training set
This is not applicable. There is no AI algorithm involved that requires a training set. The "study" here consists of non-clinical performance and biocompatibility testing of the physical catheter.
9. How the ground truth for the training set was established
This is not applicable, as there is no AI algorithm or training set.
Ask a specific question about this device
(202 days)
Q3 Distal Access Catheter, Q4 Distal Access Catheter, Q5 Distal Access Catheter, Q6Distal Access Catheter
The Q Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.
The MIVI Q Catheter is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. The proximal portion of the catheter is a stainless-steel control (push) wire. The Q catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The catheter contains a pin vise threaded on the control wire, which may be used to advance the catheter.
Acceptance Criteria and Study for MIVI Q Distal Access Catheter
This document describes the acceptance criteria and study proving the MIVI Q Distal Access Catheter meets these criteria, based on the provided FDA 510(k) summary (K192558).
It's important to note that this device is a medical catheter and not an AI/ML powered device, so many of the requested categories (e.g., sample size for test/training sets, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, data provenance) are not directly applicable in the context of this traditional medical device submission. The study focuses on bench testing (non-clinical performance testing) to demonstrate substantial equivalence to a predicate device, rather than clinical trials or AI model validation.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Test) | Criterion (Pass/Fail) | Reported Device Performance (Result) |
---|---|---|
Dimensional Verification | Device meets specified dimensional requirements. | Pass |
Corrosion Resistance | Exposed metallic components show no signs of corrosion. | Pass |
Tensile Strength – Pin Vise | Adequate device tensile strength. | Pass |
System Compatibility / Simulated Use | Successful simulated use testing with "worst case" dimensional stack up. | Pass |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated for each test. For medical devices undergoing bench testing, samples are typically chosen based on statistical methods to ensure representativeness and confidence in the results, though the exact numbers are not detailed in this summary.
- Data Provenance: This is bench testing conducted by the manufacturer, MIVI Neuroscience, Inc., at their facilities. It is inherently prospective for the specific design verification of the modified device. Country of origin for the data is implicitly the United States, given the company's location (Eden Prairie, Minnesota).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. For this type of medical device submission focused on physical properties and performance, "ground truth" is established by engineering specifications, validated test methods, and industry standards, not by expert medical consensus on human-generated data. The "experts" would be the engineers and quality control personnel designing and executing the tests.
4. Adjudication Method for the Test Set
Not applicable. The tests are objective and quantitative (e.g., measuring dimensions, observing corrosion, measuring tensile strength). There is no "adjudication" in the sense of resolving disagreements between human observers. The outcomes are determined by adherence to pre-defined engineering and performance specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML-powered diagnostic devices where human readers (e.g., radiologists) interpret cases with and without AI assistance. The MIVI Q Distal Access Catheter is a physical medical device (catheter) with no AI/ML components or human diagnostic interpretation associated with its direct function.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
No, this is not applicable. The MIVI Q Distal Access Catheter is a physical instrument, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the performance testing is based on:
- Engineering Specifications: Pre-defined dimensional requirements, material properties, and functional performance targets derived from the device's intended use and comparison to the predicate device.
- Industry Standards: Relevant ISO or ASTM standards governing medical device testing (though not explicitly listed in this summary, they are implicitly followed in such submissions).
- Predicate Device Performance: The established performance and safety profile of the legally marketed predicate device (MIVI Mi-EXT Extension Catheter, now branded as the MIVI Q Distal Access Catheter K163233) serves as a benchmark for substantial equivalence.
8. The Sample Size for the Training Set
Not applicable. This device does not involve machine learning or a "training set."
9. How the Ground Truth for the Training Set was Established
Not applicable. There is no training set for this device.
Ask a specific question about this device
(89 days)
AXS Catalyst Distal Access Catheter 068 x 115cm, AXS Catalyst Distal Access Catheter 068 x 125cm, AXS
Catalyst Distal Access Catheter 068 x 132cm
The AXS Catalyst Distal Access Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The AXS Catalyst Distal Access Catheter is also indicated for use as a conduit for retrieval devices.
The AXS Catalyst™ Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The catheter shaft has a hydrophilic coating to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostasis Valve (RHV), Tuohy Borst Valve with Sideport, and two peel away introducer sheaths. The RHV and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst.
The provided document describes the acceptance criteria and the studies conducted to demonstrate the substantial equivalence of the AXS Catalyst 7 Distal Access Catheter to its predicate devices, rather than proving it meets specific acceptance criteria for a new, non-predicate device. The entire submission focuses on establishing substantial equivalence for 510(k) clearance.
Here's an analysis of the provided information, framed around the requested categories:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly provide a table of "acceptance criteria" alongside specific numerical "reported device performance." Instead, for each test, it outlines the "Purpose" (which implicitly defines the acceptance criteria) and a "Conclusion" stating whether the device "meets acceptance criteria" or was "acceptable."
Here's a summary derived from the "Performance Data – Bench Testing" section:
Test | Acceptance Criteria (Derived from Purpose) | Reported Device Performance/Conclusion |
---|---|---|
Dimensional Verification | Verified dimensions using specified measurement tool. | Dimensional verification meets acceptance criteria. |
Tip Configuration | Catheter tip is smooth, rounded, tapered, or similarly finished to minimize trauma. | Tip configuration meets acceptance criteria. |
Surface Integrity | External surface free from extraneous matter, defects, and lubricant drops. | Surface integrity meets acceptance criteria. |
Tip Buckling | Withstand a maximum force before buckling. | Tip buckling meets acceptance criteria. |
Catheter Lubricity & Durability | Lubricity and durability of the coating on the outer shaft. | Coating lubricity and durability meets acceptance criteria. |
Particulate Characterization | Acceptable levels of particulates generated during simulated use (≥10µm, ≥25µm, ≥50µm). | Particulate generation was acceptable. |
Coating Integrity | Visual absence of coating anomalies, defects, or artifacts pre and post-simulated use. | Coating integrity was acceptable. |
Trackability | Acceptable track advance force of catheter over microcatheter. | Track advance force meets acceptance criteria. |
Tensile Strength | Withstand tensile force before failure of fused joints, shaft junctions, and marker band. | Tensile strength meets acceptance criteria. |
Liquid Leak Resistance | Meets freedom from leakage-liquid leak requirement (EN ISO 10555-1, EN 1707). | Liquid leak resistance of catheter meets acceptance criteria. |
Air Leak Resistance | Meets freedom from leakage-air aspiration requirement (EN ISO 10555-1, EN 1707). | Air leak resistance of catheter meets acceptance criteria. |
Catheter Torsional Bond Strength | Withstand a specified number of rotations before failure when torque is applied. | Catheter torsional bond strength meets acceptance criteria. |
Flexural Fatigue | Absence of kinks or damage after repeated advancement and withdrawal through a model. | Flexural fatigue meets acceptance criteria. |
Catheter Kink Radius | Acceptable kink radius at distal and mid-shaft joint sections. | Catheter kink radius meets acceptance criteria. |
Catheter Tip and Lumen Integrity (Adjunctive Aspiration) | Ability to deliver and withdraw retrieval device 3 times without negative impact on function/integrity. | Catheter tip and lumen integrity during adjunctive aspiration meets acceptance criteria. |
Chemical Compatibility | Visual and dimensional integrity after exposure to saline and non-ionic contrast. | Chemical compatibility meets acceptance criteria. |
Hub Gauging | Meets gauging requirement (EN 1707). | Hub gauging meets acceptance criteria. |
In-vitro Simulated Use Study | Durability and kink resistance of devices, successful interventional device delivery. | All test samples meet acceptance criteria. (Evaluated subject and primary predicate devices in a tortuous model). |
Biocompatibility Testing (Table 3):
Test Performed | Acceptance Criteria | Reported Device Performance/Conclusion |
---|---|---|
MEM Elution Cytotoxicity | No biological activity (Grade | Non-volatile residue: ≤ 10 mg; residue on ignition: ≤ 1 mg; heavy metal: ≤ 1 ppm; buffering capacity: ≤ 1.0 ml. |
Ask a specific question about this device
(89 days)
AXS Catalyst 7 Distal Access Catheter (068 x 115cm, 068 x 125cm, 068 x 132cm)
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
The AXS Catalyst 7 Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in the removal of thrombus from the neuro vasculature using continuous aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostasis Valve (RHV), Tuohy Borst Valve with Sideport, and two peel-away introducer sheaths. The RHV and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst.
When used as part of the aspiration system, the AXS Catalyst Distal Access Catheter requires a minimum vacuum pressure of -68 kPa [-20.08 in Hg] from an external device (Medela Dominant Flex Pump). Bench and Animal testing utilizing the Medela Dominant Flex Pump and Suction Jar, AXS Universal Aspiration Tubing, and AXS Universal Liner Set were conducted to successfully demonstrate aspiration.
The provided document is a 510(k) summary for the AXS Universal Aspiration System (AXS Catalyst 7 Distal Access Catheter). It details the device's characteristics, comparison to predicate devices, and a summary of performance data (bench, animal, and clinical) to demonstrate substantial equivalence.
Based on the document, here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document presents its "acceptance criteria" implicitly through the "Conclusions" column in Table 2 (Performance Data - Bench Testing), Table 3 (Performance Data - Animal Testing), and Table 4 (Biocompatibility Studies). The acceptance criteria are "meets acceptance criteria," "acceptable," "equivalent," or "PASS."
Test Type | Acceptance Criteria (Implicit from "Conclusions") | Reported Device Performance ("Conclusions" column) |
---|---|---|
Design Verification Bench Testing | ||
Dimensional Verification | Meets acceptance criteria | Dimensional verification meets acceptance criteria. |
Tip Configuration | Meets acceptance criteria | Tip configuration meets acceptance criteria. |
Surface Integrity | Meets acceptance criteria | Surface integrity meets acceptance criteria. |
Tip Buckling | Meets acceptance criteria | Tip buckling meets acceptance criteria. |
Catheter lubricity and durability | Meets acceptance criteria | Coating lubricity and durability meets acceptance criteria. |
Particulate characterization | Acceptable | Particulate generation was acceptable. |
Coating integrity | Acceptable | Coating integrity was acceptable. |
Trackability | Meets acceptance criteria | Track advance force meets acceptance criteria. |
Tensile Strength | Meets acceptance criteria | Tensile strength meets acceptance criteria. |
Liquid Leak Resistance | Meets acceptance criteria | Liquid leak resistance of catheter meets acceptance criteria. |
Air Leak Resistance | Meets acceptance criteria | Air leak resistance of catheter meets acceptance criteria. |
Catheter Torsional Bond Strength | Meets acceptance criteria | Catheter torsional bond strength meets acceptance criteria. |
Flexural Fatigue | Meets acceptance criteria | Flexural fatigue meets acceptance criteria. |
Catheter Kink Radius | Meets acceptance criteria | Catheter kink radius meets acceptance criteria. |
Catheter Tip and Lumen Integrity (Direct Aspiration) | Meets acceptance criteria | Catheter tip and lumen integrity during direct aspiration meets acceptance criteria. |
Chemical Compatibility | Meets acceptance criteria | Chemical compatibility meets acceptance criteria. |
Hub Gauging | Meets acceptance criteria | Hub gauging meets acceptance criteria. |
Aspiration Flow Rate Bench Testing | ||
Aspiration Flow Rate | Meets acceptance criteria | All test samples meet acceptance criteria. |
Design Validation Bench Testing | ||
In-vitro Usability Study | Meets acceptance criteria | All test samples meet acceptance criteria. |
Animal Study | Equivalent to predicate in safety | Subject Device is equivalent to the Additional Predicate Device in safety. |
Biocompatibility Studies | PASS | MEM Elution Cytotoxicity: PASS (No cytotoxicity or cell lysis) |
Hemolysis Extract/Direct Contact Method: PASS (Non-hemolytic) | ||
USP Physiochemical : PASS | ||
Chemistry (Heptane) Analysis: PASS (Below level of detection) | ||
FTIR (ISO 10993-18): PASS | ||
Natural Rubber Latex ELISA Inhibition Assay for Antigenic Protein ASTM D6499-12: PASS (Below level of detection) |
2. Sample sizes used for the test set and the data provenance
The document does not explicitly state the numerical sample sizes (e.g., number of catheters tested) for each bench test. It mentions qualitative assessments like "Visually inspect" or "Prepare sample for test." For the animal study, it mentions "porcine test subjects" but does not give the exact number.
- Provenance: The studies were conducted by Stryker Neurovascular ("Submitter Name, Address, and Content" section) and likely took place at their facilities or certified labs. The animal study was conducted "in compliance with applicable requirements in the GLP regulation (21 CFR Part 58)," which specifies good laboratory practices, but not geographic provenance. The clinical data was not from a study conducted by the submitter but rather a "review was conducted considering published clinical study articles that featured the Predicate devices." This implies that the clinical data is retrospective from existing literature, and its origin would vary based on those publications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The studies described are primarily bench and animal tests, along with a review of published clinical articles. There's no mention of expert consensus or human readers being used to establish ground truth for a test set in the context of device performance, as one might find for an AI/imaging device. The "In-vitro Usability Study" mentions "multiple physician users" but doesn't specify their number or qualifications, nor is it described as establishing "ground truth" for the device's technical performance, but rather "evaluat[ing] aspiration integrity, ability to restore flow and the durability and kink resistance."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable/not provided. Adjudication methods are typically used in studies involving human readers or expert panels evaluating diagnostic outputs (like images), especially in the context of establishing ground truth where there might be disagreement. The studies described here are primarily engineering (bench) and animal model tests, where outcomes are determined by physical measurements or biological responses, not human interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a "Percutaneous Catheter" (specifically, a distal access catheter for stroke treatment), not an AI-based diagnostic or assistive software. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance would not be relevant to this device's submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical medical instrument, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the bench and animal tests, the "ground truth" is established through physical measurements, established engineering test methods (e.g., EN ISO 10555-1, EN 1707, EN ISO 14971), and biological evaluations (histopathology in the animal study). For the biocompatibility tests, it's based on established ISO standards (EN ISO 10993-1 and sub-parts) and analytical chemistry results. There is no mention of expert consensus or pathology reports serving as "ground truth" in the way it might for a diagnostic imaging AI. The "clinical performance" relies on a review of published clinical outcomes of predicate devices, leveraging their established safety and efficacy rather than generating new outcomes data for the subject device.
8. The sample size for the training set
This is not applicable/not provided. The device is a physical medical device, not a machine learning model. Therefore, there is no "training set."
9. How the ground truth for the training set was established
This is not applicable. As there is no training set for a physical device, there's no ground truth established for one.
Ask a specific question about this device
(90 days)
AXS Catalyst Distal Access Catheter 058 x 115cm; AXS Catalyst Distal Access Catheter 058 x 132cm; AXS
Catalyst Distal Access Catheter 060 x 132cm
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. The AXS Universal Aspiration System is designed for use within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries. The AXS Universal Aspiration System is composed of the following components:
- AXS Catalyst™ Distal Access Catheter ●
- AXS Universal Aspiration Tubing
- . Medela Dominant Flex Pump
- AXS Universal Liner Set
The AXS Catalyst™ Distal Access Catheter is a sterile, single lumen, variable stiffness catheter. The catheter shaft has a hydrophilic coating to reduce friction during use, includes a radiopaque marker on the distal end for angiographic visualization, and includes a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostastic Valve (RHV), Tuohy Borst Valve with Sideport, and Peel Away Introducer. The Rotating Hemostastic Valve and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst. The AXS Catalyst Distal Access Catheter is the only component of the AXS Universal Aspiration System that is used intravascularly.
The AXS Universal Aspiration Tubing serves as a conduit to supply vacuum from the Medela Dominant Flex Pump to the distal tip of the AXS Catalyst Distal Access Catheter. The AXS Universal Aspiration Tubing provides a connection between the sterile and nonsterile environments. The proximal end of the AXS Universal Aspiration Tubing is connected to the AXS Universal Liner Set (outside of the sterile environment) while the distal end of the AXS Universal Aspiration Tubing is connected to the AXS Catalyst Distal Access Catheter (inside the sterile environment). The AXS Universal Liner Set is connected to the Medela Dominant Flex Pump (also outside of the sterile environment).
The Medela Dominant Flex Pump is designed to generate vacuum for the AXS Universal Aspiration System. When used as part of the AXS Universal Aspiration System, the AXS Catalyst Distal Access Catheter requires a minimum vacuum pressure of -68 kPa [-20.08 in Hg] from the Medela Dominant Flex Pump. The Medela Dominant Flex Pump is reusable, non-sterile, and intended to be utilized outside of the sterile environment.
The AXS Universal Liner Set is provided non-sterile and consists of an individually packaged canister liner and a ClotFinder specimen cup. The AXS Universal Liner Set is offered with and without a desiccant. The AXS Universal Liner Set is single-use and the repository for aspirated material.
This document is a 510(k) Premarket Notification from the U.S. Food and Drug Administration (FDA) regarding the AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System. This type of document is a submission to the FDA to demonstrate that the device to be marketed is substantially equivalent to a legally marketed predicate device.
The document does not describe a study that proves the device meets specific acceptance criteria related to AI/algorithm performance. Instead, it details bench testing and animal studies to demonstrate the substantial equivalence of the mechanical device (catheter and aspiration system) to existing predicate devices.
Therefore, many of the requested information points regarding AI/algorithm performance, such as sample size for test sets, data provenance, number of experts for ground truth, MRMC studies, or standalone algorithm performance, are not applicable to this document. The document focuses on the physical and functional characteristics of a medical device intended for mechanical thrombectomy in stroke patients.
Nonetheless, I will extract and present the information that is available in the document, framed within the context of the device's mechanical performance and substantial equivalence.
Device Acceptance Criteria and Performance (Based on Mechanical/Physical Testing)
The acceptance criteria for this device are demonstrated through various bench tests and animal studies, showing that the AXS Catalyst Distal Access Catheter, when used as part of the AXS Universal Aspiration System, performs similarly to its predicate devices concerning its physical properties, functionality, and safety. There is no AI component in this device.
1. A table of acceptance criteria and the reported device performance
Since this is a 510(k) submission for a mechanical device, formal "acceptance criteria" are not listed in terms of numerical thresholds for clinical outcomes in the same way they would be for an AI device. Instead, the "acceptance criteria" are implicit in the conclusion that "All test samples met acceptance criteria" or "met established criteria," demonstrating parity or superiority to predicate devices in specific bench and animal tests. The device's performance is reported as meeting these (unspecified) criteria.
Test Category | Test Name | Test Method Summary | Reported Device Performance/Conclusion |
---|---|---|---|
Design Validation (Bench) | In-vitro Simulated Use (Direct Aspiration) | Purpose: To evaluate the performance of the Subject Device to Primary Predicate when aspirating clot in tortuous anatomical model. Method: Simulated use testing uses a physiological neurovascular model where clot is aspirated following the written protocol. | All test samples met acceptance criteria. |
In-vitro Usability Study | Purpose: Multiple User evaluation of the clot retrieval, durability and kink resistance of the Subject and Primary Predicate Devices in a tortuous anatomical model. Method: Users performed a direct aspiration of a clot procedure, tracked the devices to the site of the occlusion using a neurovascular model that replicated the tortuosity, diameter and location of the arteries in the neurovasculature. | All test samples met acceptance criteria. | |
Design Verification (Bench) | Aspiration Flow Rate | Purpose: To determine the aspiration flow rate and lumen integrity of the system when no occlusion is present. Method: The volume of water an aspiration system (catheter + associated pump and accessories) could aspirate in 20s was measured. This volume in mL was divided by 20 to give the flow rate in mL/s. | All test samples met acceptance criteria. |
Tip Buckling | Purpose: To determine the maximum force a catheter tip could withstand before buckling. Method: The test catheter was soaked in 37°C water prior to testing. A mandrel was then inserted into the distal end of the catheter and the catheter with mandrel was placed into the test fixture on a tensile test machine. The test catheter was compressed against a load cell until the distal tip buckled. The compression force was recorded. | All test samples met acceptance criteria. | |
Track Force | Purpose: To determine the maximum force a test catheter exerts on the tortuous vessel model as it tracks to the M2 for both advancing and retracting. Method: The test catheter was soaked in 37°C water prior to testing. The catheter was tracked to the M2 through the smallest ID compatible guide in the neurovascular challenge path model and the force was recorded. The catheter was then retracted through the same model path and the force was recorded. | All test samples met acceptance criteria. | |
Catheter Tip and Lumen Patency (Direct Aspiration) | Purpose: To test resistance to tip and lumen collapse during direct aspiration and test tip integrity to tears and missing material. Method: Prepare test sample and simulated use model. Insert plug in catheter tip. Place test sample in the model to a specified location following procedural instructions outlined in the Instructions for Use. Aspirate test sample using 60cc syringe and aspiration pump. Visually inspect test sample to verify indication of no tip or lumen collapse. | All test samples met acceptance criteria. | |
Previously Cleared Bench Tests | Dimensional Verification | Purpose: To describe the procedure and technique of making dimensional measurements using various measurement equipment. Method: Verify dimensions using specified measurement tool. Record measurements. | Dimensional verification met acceptance criteria and was cleared as part of K151667. |
Tip Configuration | Purpose: To verify that the catheter tip is smooth, rounded, tapered or similarly finished in order to minimize trauma to vessels during use per EN ISO 10555-1. Method: Visually inspect distal tip at 10X magnification to verify distal tip end is smooth, rounded, tapered or similarly finished. Record results. | Tip configuration met acceptance criteria and was cleared as part of K151667. | |
Surface Integrity | Purpose: To determine if external surface of the catheter is free from extraneous matter, process and surface defects, and does not have drops of lubricant fluids. Method: Visually inspect external surface of catheter for extraneous matter, process and surface defects, and drops of lubricant fluids. Record results. | Surface integrity met acceptance criteria and was cleared as part of K151667. | |
Tip Buckling | Purpose: To measure the maximum force required to cause a test sample to buckle. Method: Prepare sample for test. Use buckling tester to measure the maximum force required to cause a test sample to buckle. Record results. | Tip buckling met acceptance criteria and was cleared as part of K151667. | |
Catheter lubricity and durability | Purpose: To determine the lubricity and durability of the coating on the catheter outer shaft. Method: Prepare sample for test. Use friction tester to measure the frictional force of the device sample when pulled between two clamped pads. Record the peak frictional force after 5 cycles. | Coating lubricity and durability met acceptance criteria and was cleared as part of K151667. | |
Trackability | Purpose: To measure track advance force of catheter over microcatheter. Method: A neurovascular model is placed in a re-circulating water bath at 37°C to simulate human arterial circulation. The sample is inserted through model over a microcatheter and attached to a tensile tester. Advance catheter through model and determine peak tracking force. Record results. | Track advance force met acceptance criteria and was cleared as part of K151667. | |
Tensile Strength | Purpose: To determine tensile force tensile force required to induce failure of fused joints, shaft junctions, and marker band for non-hydratable catheters based on EN ISO 10555-1. Method: Identify joint and prepare sample for test. Use tensile tester to determine applied peak tensile force. Record results. | Tensile strength met acceptance criteria and was cleared as part of K151667. | |
Liquid Leak Resistance | Purpose: 1) To determine whether catheter meets the freedom from leakage-liquid leak requirement 4.7.1 of EN ISO 10555-1. 2) To determine if catheter hub meets the liquid leakage requirement 4.2.1 of EN 1707. Method: Connect test hub sample to fixture and flush with water to expel air. Occlude distal tip. Apply pressure of 300kPa minimum and maintain pressure for 30s. Visually inspect catheter/hub joint and catheter shaft for leaks. Record results. | Liquid leak resistance of catheter met acceptance criteria and was cleared as part of K151667. | |
Air Leak Resistance | Purpose: 1) To determine whether catheter meets the freedom from leakage-air aspiration requirement of 4.7.2 of EN ISO 10555-1. 2) To determine if catheter hub meets the air leakage requirement 4.2.2 of EN 1707. Method: Connect test hub sample to a partially filled syringe. With the nozzle of the syringe pointing down towards the ground, withdraw the plunger to the 10cc mark. Hold for 15 seconds and examine the water in the syringe for the formation of air bubbles. Record results. | Air leak resistance of catheter met acceptance criteria and was cleared as part of K151667. | |
Catheter Torsional Bond Strength | Purpose: To measure the strength of a catheter shaft when torque is applied. Torque strength is defined as number of rotations before failure occurs. Method: Prepare test sample and insert into torsional bond strength test fixture with tortuous path model. Apply torque to catheter shaft and observe number of 360-degree rotations before failure occurs. Record results. | Catheter torsional bond strength met acceptance criteria and was cleared as part of K151667. | |
Flexural Fatigue | Purpose: To determine the flexural fatigue on the catheter shaft. Method: Prepare test sample. Advance entire assembly of guide wire, microcatheter, and test sample into test model and track it through test model. While holding the guide wire, microcatheter, and test sample, pull the whole assembly pack proximally until it exits the models. Repeat for nine more runs. After run number ten, remove guide wire and microcatheter out of test sample and inspect for kink or damage. Record results. | Flexural fatigue met acceptance criteria and was cleared as part of K151667. | |
Catheter Kink Radius | Purpose: To measure the kink radius of a catheter at its distal and specific mid-shaft joint section. Method: Prepare test sample. Thread test sample through fixture loop and lock down test sample. Pull both ends of test sample until test sample kinks. Calculate kink radius using measurement of 2nd to final loop OD and sample OD. Record results. | Catheter kink radius met acceptance criteria and was cleared as part of K151667. | |
Chemical Compatibility | Purpose: To determine visual and dimensional integrity of catheter following exposure to saline, non-ionic and ionic contrast liquids. Method: Prepare sample for test. Flush sample with appropriate chemical. Measure ID and OD using RAM optical measurement system. Insert mandrel through sample to verify inner lumen integrity. Repeat with second mandrel and record results. Visually inspect distal end of sample for any chemical effects on the shaft, inner lumen and cross-sectional areas. Record results. | Chemical compatibility met acceptance criteria and was cleared as part of K151667. | |
Hub Gauging | Purpose: To determine if catheter hub meets gauging requirement 4.1 of EN 1707. Method: Using the appropriate gauge, the gauge was applied to the conical fitting with a total axial force of 5N without the use of torque. The axial load was then removed and the sample inspected. | Hub gauging met acceptance criteria and was cleared as part of K151667. | |
Animal Study | In-vivo Efficacy and Safety Evaluation (Arm 1) | Purpose: To assess vessel revascularization and adverse events, if any, associated with a mechanical thrombectomy procedure performed via direct aspiration using the Subject Device compared to the Primary Predicate Device. Method: Porcine test subjects were exposed to aspiration treatment using the AXS Universal Aspiration System and the predicate Penumbra System after a vascular occlusion was artificially induced. Vascular response was assessed by contrast angiography and histopathology. | Subject Device is equivalent to the Primary Predicate in efficacy and safety. |
In-vivo Efficacy and Safety Evaluation (Arm 2) | Purpose: To assess the vascular safety profile of treatment by aspiration and navigation of the Subject Device compared to the Primary Predicate Device. Method: Porcine test subjects were exposed to aspiration treatment using the AXS Universal Aspiration System and the predicate Penumbra System under worst-case aspiration force and treatment duration conditions. Vascular response was assessed by contrast angiography and histopathology. | Subject Device is equivalent to the Primary Predicate in safety. | |
In-vivo Vascular Response (Direct Aspiration) | Purpose: To assess the vascular response of direct aspiration through the Subject Device compared to Primary Predicate device. Method: A porcine model was used to evaluate acute and chronic vessel damage after direct aspiration in vessels sized appropriately to simulate the human M2. Vascular response was assessed by contrast angiography and histopathology. | Subject Device is equivalent to the Primary Predicate in safety. | |
Biocompatibility | MEM Elution Cytotoxicity | No biological activity (Grade 0) was observed in the L929 mammalian cells at 48 hours post exposure to the test article extract. The observed cellular response obtained from the positive control article extract (Grade 4) and negative control article extract (Grade 0) confirmed the suitability of the test system. | PASS (No cytotoxicity or cell lysis) |
Guinea Pig Maximization Sensitization | The USP 0.9% Sodium Chloride for Injection (NaCl) and Cottonseed Oil (CSO) extracts of the test article elicited no reaction at the challenge (0% sensitization), following an induction phase. | PASS (No evidence of sensitization) | |
Intracutaneous Reactivity | The test article sites did not show a significantly greater biological reaction than the sites injected with the control article. The difference of the overall mean score between the test article and the control article was 0.0. | PASS (Non-irritant) | |
Acute Systemic Injection | The 0.9% Sodium Chloride for Injection (NaCl) and Cottonseed Oil (CSO) extracts of the test article did not induce a significantly greater biological reaction than the control extracts, when tested in Swiss Albino mice. | PASS (No mortality or evidence of systemic toxicity) | |
Rabbit Pyrogen | No rabbit injected with the test article extract showed an individual rise in temperature of 0.5℃ or more. | PASS (Non-pyrogenic) | |
Hemolysis Extract/Direct Contact Method | The test article exhibited 0.17% hemolysis above the level of hemolysis exhibited by the negative control via the direct method and 0.12% hemolysis above the level of hemolysis exhibited by the negative control via the indirect method. | PASS (Non-hemolytic) | |
In Vitro Hemocompatibility | Results comparable to the Negative Control. The test article results for WBC, RBC, platelets, hematocrit and hemoglobin were: Group 1 89% - 98% Group 2 97% - 103% Group 3 100% - 105% Group 4 98% - 105% | PASS | |
Complement Activation (SC5b-9) | Concentration of SC5b-9 in the test articles was not statistically higher than the negative control. The test articles are not considered to be potential activators of the complement system. | PASS | |
Complement Activation (C3a) | Concentration of C3a in the test articles was not statistically higher than the negative control. The test articles are not considered to be a potential activator of the complement system. | PASS | |
Partial Thromboplastin (PTT) | The test sample and the predicate sample demonstrated a shortened clotting time when compared to the negative control. However, the test sample demonstrated a similar clotting time when compared to the predicate sample. | PASS (Results were comparable to the Negative Control; minimal activators) |
2. Sample size used for the test set and the data provenance:
- Bench Testing: The document states that "All test samples met acceptance criteria" for various bench tests, but the specific numerical sample sizes for each test (e.g., number of catheters tested for aspiration flow rate, tip buckling, etc.) are not explicitly provided in this summary. The tests were likely conducted in a controlled laboratory environment.
- Animal Study: "Porcine test subjects" were used. The exact number of animals is not specified in this summary. The studies were conducted in compliance with GLP regulation (21 CFR Part 58), typically implying rigorous, controlled experimental settings. Data provenance would be from these controlled in-vivo studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable as the device is mechanical and its evaluation does not involve image interpretation or AI-driven diagnostic tasks requiring expert ground truth for a test set in the traditional sense. The "ground truth" for mechanical properties is established by physical measurement, and for biological response by expert pathological/angiographic assessment. The "usability study" mentions "Multiple User evaluation," but does not detail their qualifications or roles in "ground truth" establishment beyond performing the procedure.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This information is not applicable. Adjudication methods are relevant for subjective assessments, typically in AI or interpretive diagnostic studies. The testing here is largely objective mechanical/biological evaluation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No, an MRMC study was not done. This type of study is typically performed for diagnostic imaging devices or AI systems where human readers interpret cases. This submission is for a mechanical medical device (catheter and aspiration system).
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- No. This is not an algorithm-based device. "Standalone" performance would refer to the mechanical operation of the device in a bench or animal model without direct human intervention in the data results (e.g., measuring flow rate via automated systems), which is implied in the bench testing.
7. The type of ground truth used:
- Bench Testing: Physical measurements, engineered models (e.g., tortuous anatomical models, neurovascular challenge path model), and established laboratory testing standards.
- Animal Study: Vascular response assessed by contrast angiography and histopathology. These are considered objective, often expert-interpreted, assessments of the biological effects of the device.
8. The sample size for the training set:
- Not applicable. This document describes the testing and validation of a mechanical medical device, not an AI/machine learning model. Therefore, there is no "training set" in the context of data for model development.
9. How the ground truth for the training set was established:
- Not applicable. As there is no AI/ML training set, this question is irrelevant to the provided document.
Ask a specific question about this device
Page 1 of 2