Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K231218
    Date Cleared
    2024-01-19

    (266 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Distal Access Catheter is indicated for the introduction of interventional devices into the peripheral and neurovasculature.

    Device Description

    The Distal Access Catheter is a single lumen, flexible, variable stiffness composite catheter which has a luer hub on the proximal end. The catheter shaft has a hydrophilic coating to reduce friction during use. The Distal Access Catheter has a radiopaque marker on the distal tip that is visible under fluoroscopy. The Distal Access Catheter inner lumen can accommodate guidewires up to 0.038 inches in diameter to aid in placement of the catheter system. The catheter has a straight tip. The catheter is offered in various lengths to accommodate physician preferences and anatomical variations. The catheter is provided sterile, nonpyrogenic, and is intended for single use only.

    AI/ML Overview

    The provided text describes information for a Distal Access Catheter, specifically related to its 510(k) premarket notification for FDA clearance. It details the device's characteristics, intended use, and comparisons to a predicate device, along with non-clinical performance testing (bench testing, biocompatibility, sterility, and shelf-life).

    However, the prompt asks for acceptance criteria and a study that proves a device meets acceptance criteria, with specific questions related to AI/Machine Learning (ML) performance, such as:

    1. A table of acceptance criteria and the reported device performance.
    2. Sample size used for the test set and data provenance.
    3. Number of experts used to establish ground truth and their qualifications.
    4. Adjudication method for the test set.
    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done and its effect size.
    6. If a standalone (algorithm only) performance study was done.
    7. The type of ground truth used.
    8. The sample size for the training set.
    9. How the ground truth for the training set was established.

    The provided FDA 510(k) document is for a physical medical device (catheter) and does NOT describe any AI/ML components or studies. Therefore, it does not contain the information necessary to answer the questions about AI/ML acceptance criteria, test sets, ground truth establishment, or clinical studies involving AI.

    Based on the provided document, here is a summary of acceptance criteria and performance for the physical device:

    Acceptance Criteria and Device Performance (for the physical catheter)

    1. A table of acceptance criteria and the reported device performance

    The document lists various non-clinical (bench) tests and biocompatibility tests. The "Result" column effectively indicates whether the device met the pre-defined acceptance criteria for each test.

    TestTest Method SummaryAcceptance Criteria (Implied by Result) & Reported Device Performance
    Bench Testing
    Dimensional VerificationMeasured effective length, inner diameter (proximal/distal), max outer diameter.Device met the dimensional specifications. (Implied acceptance: within specified dimensions)
    Visual InspectionInspected for structural or mechanical damage.Device met the visual specifications. (Implied acceptance: no structural/mechanical damage)
    Simulated Use and Compatibility TestingSimulated clinical use with compatible devices in a tortuous anatomical bench test model.Device performed as intended under simulated use conditions. (Implied acceptance: performed as expected)
    Delivery and RetrievalMeasured maximum forces to deliver and retrieve device through tortuous anatomical bench model.Device met the delivery and retrieval force specifications. (Implied acceptance: forces within specification)
    Tensile StrengthTested peak tensile force of all bonding points using a universal tensile machine per ISO 10555-1.Device met acceptance criteria determined with delivery and retrieval force testing. (Implied acceptance: passed)
    Torque StrengthDevices pre-conditioned and torqued to failure inside tortuous anatomical model with distal tip held fixed.Device torque strength is similar to the predicate. (Implied acceptance: comparable to predicate)
    Burst PressureEvaluated per ISO 10555-1:2013 following pre-conditioning.Device resistance to burst pressure is similar to the predicate. (Implied acceptance: comparable to predicate)
    Kink ResistanceDevices pre-conditioned and bonding points wrapped around varying size pin gauges until failure.Device kink resistance is similar to the predicate. (Implied acceptance: comparable to predicate)
    Coating IntegrityHydrophilic coating on catheter surface evaluated after particulate testing for defects.Device coating integrity is similar to the predicate. (Implied acceptance: comparable to predicate)
    Corrosion ResistanceEvaluated per ISO 10555-1:2013.Device met corrosion resistance specification. (Implied acceptance: passed)
    Air LeakageEvaluated per ISO 10555-1:2013 following pre-conditioning.Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. (Implied acceptance: passed)
    Liquid LeakageEvaluated per ISO 10555-1:2013 following pre-conditioning.Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. (Implied acceptance: passed)
    RadiopacityVisibility of distal marker band and catheter body evaluated under X-ray.Device radiopacity comparable to the predicate device. (Implied acceptance: comparable to predicate)
    Particulate TestingTest articles tracked multiple times through tortuous anatomical model with ancillary devices; particulates collected.Particulate generation of the subject device was comparable to the predicate device. (Implied acceptance: comparable to predicate)
    Luer Connector TestingEvaluated per ISO 80369-7 and ISO 80369-20.Device met the establish acceptance criteria. (Implied acceptance: passed)
    Tip Flexibility and Tip Buckling TestingCatheter distal tip evaluated following simulated use testing.Tip flexibility and tip buckling of the subject device are similar to the predicate device. (Implied acceptance: comparable to predicate)
    Biocompatibility
    CytotoxicityISO 10993-5:2009 MTT Method.The Distal Access Catheter, introducer sheath, and shaping mandrel were non-cytotoxic. (Implied acceptance: non-cytotoxic)
    Skin SensitizationISO 10993-10:2021 Guinea Pig Maximization Test.The Distal Access Catheter, introducer sheath, and shaping mandrel were non-sensitizers. (Implied acceptance: non-sensitizing)
    IrritationISO 10993-23:2021 Intracutaneous Reactivity Test.The Distal Access Catheter, introducer sheath, and shaping mandrel were non-irritant. (Implied acceptance: non-irritant)
    Acute Systemic ToxicityISO 10993-11:2017 Acute Systemic Toxicity Study in Mice.No mortality or evidence of acute systemic toxicity with the Distal Access Catheter, introducer sheath, and shaping mandrel test articles. (Implied acceptance: passed)
    Material-Mediated PyrogenicityUSP <151>, ISO 10993-11:2017 Rabbit Pyrogen Study.The Distal Access Catheter, introducer sheath, and shaping mandrel were non-pyrogenic. (Implied acceptance: non-pyrogenic)
    Hemocompatibility (Hemolysis)ASTM F756-2017, ISO 10993-4:2017 ASTM Hemolysis Study (Direct and Indirect Contact).The Distal Access Catheter was non-hemolytic (direct and indirect contact). The introducer sheath and shaping mandrel were non-hemolytic (indirect contact). (Implied acceptance: non-hemolytic)
    Hemocompatibility (Complement Activation)ISO 10993-4:2017 Complement Activation, SC5b-9.The Distal Access Catheter results were similar to the negative control group. (Implied acceptance: no abnormal activation)
    Hemocompatibility (Partial ThromboplastinISO 10993-4:2017, ASTM F2382-2018 Partial Thromboplastin Time Study.The Distal Access Catheter results were similar to the negative control group. (Implied acceptance: no abnormal effect on clotting)
    Hemocompatibility (Thromboresistance)ISO 10993-4:2017 In Vivo Vein Thromboresistance Study.The Distal Access Catheter results were similar to the control and had no evidence of thrombosis. (Implied acceptance: no thrombosis)
    Genotoxicity (Bacterial Reverse Mutation)ISO 10993-3:2014 Bacterial Reverse Mutation Test.No backward mutation in Salmonella typhimurium with the Distal Access Catheter test article. (Implied acceptance: non-mutagenic)
    Genotoxicity (Mouse Lymphoma TK Assay)ISO 10993-3:2014 Mouse Lymphoma TK Assay.The Distal Access Catheter was non-mutagenic to mouse lymphoma cells (L5178Y TK+/--3.7.2C). (Implied acceptance: non-mutagenic)
    Genotoxicity (Chromosome Aberration Test)ISO 10993-3:2014 In vitro Mammalian Chromosome Aberration Test.The Distal Access Catheter results were similar to the negative control group. (Implied acceptance: no abnormal chromosomal aberration)
    SterilitySterilization process validated per ISO 11135:2014.Sterility Assurance Level (SAL) of 10^-4 achieved. EO and ECH residuals below ISO 10993-7:2008 limits. Bacterial Endotoxin Levels < 2.15 EU/device (USP <85>).
    Shelf-lifeBaseline and accelerated shelf-life testing.Demonstrated the device will perform as intended to support the proposed 1-year shelf-life. (Implied acceptance: meets 1-year shelf-life stability)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document describes bench testing and biocompatibility testing only. It does not mention any clinical data, patient data, retrospective, or prospective studies.
    • The sample sizes for these tests are not explicitly stated in numerical values (e.g., "n=X units were tested"), but implied by "test articles," "devices," or descriptions like "all bonding points."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This question is irrelevant for the provided document as it concerns a physical catheter, not an AI/ML product requiring "ground truth" from experts interpreting medical images or data. The "ground truth" for this device's performance is established by objective engineering and biocompatibility test standards and measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This question is irrelevant for the provided document as it concerns a physical catheter, not an AI/ML product requiring clinical adjudication of interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC or any clinical effectiveness studies were performed or described. The document explicitly states: "No clinical studies were deemed necessary to demonstrate substantial equivalence."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the physical device, the "ground truth" is defined by established engineering standards (e.g., ISO standards), physical measurements, chemical analyses, and biological assays. It's not based on expert clinical consensus or pathology in the context of diagnostic interpretation.

    8. The sample size for the training set

    • Not applicable. This is a physical device, not an AI/ML product that undergoes "training."

    9. How the ground truth for the training set was established

    • Not applicable. This is a physical device, not an AI/ML product.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233205
    Date Cleared
    2023-12-22

    (85 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Distal Access Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The Distal Access Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The catheter comes in Straight (with no shaping of the catheter tip) and Curved (curved catheter tip) models based on the distal end. The catheter shaft has a hydrophilic coating on the distal end of the shaft to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end.

    AI/ML Overview

    This document is a 510(k) summary for a medical device called the "Distal Access Catheter." It describes non-clinical performance testing and biocompatibility testing conducted to demonstrate that the device is substantially equivalent to a predicate device.

    1. Table of acceptance criteria and the reported device performance:

    TestAcceptance Criteria (Test Method / Applicable Standard)Reported Device Performance
    Dimensional VerificationVerify dimensions using specified measurement tool.Pass
    Tip ConfigurationVisually inspect distal tip at 10x magnification.Pass
    Surface IntegrityInspect catheter surface. ISO 10555-1:2013Pass
    Tip BucklingUse a testing apparatus to measure the maximum force required to cause a test sample to buckle at 5 mm, 10 mm and 20 mm from distal tip.Pass
    Luer Connector TestsISO 80369-7: 2021Pass
    Flexural FatigueAfter ten runs of the Distal Access Catheter with the accessories in the simulated use model, remove the test sample and inspect the sample for kinks or damage.Pass
    TrackabilityTest the peak tracking force and the compatibility with the ancillary devices during insertion and retraction of the catheter in the simulated use model.Pass
    Catheter Lubricity and DurabilityUse friction tester to measure the frictional force of the test samples when pulled between two clamped pads. Record the peak frictional force over 5 cycles.Pass
    Particulate and Coating Integrity TestMeasure the size and number of particulates generated during simulated use and conduct visual inspection at 50x and 200x magnification of the outer surface of the catheter after simulated use.Pass
    Tensile Strength TestISO 10555-1:2013Pass
    Air LeakageISO 10555-1:2013Pass
    Liquid LeakageISO 10555-1:2013Pass
    Torque ResistanceFix the distal tip of the Distal Access Catheter and rotate the proximal end until the Distal Access Catheter shows kink or other signs of failure.Pass
    Kink Resistance TestVisually inspect whether the Distal Access Catheter kinks or not when wrapped around mandrels with diameters 1 mm to 10 mm (in 1 mm increments).Pass
    Catheter Tip and Lumen IntegrityAfter simulated use, aspirate test sample for 10 sec using a 60 cc syringe during retrieval device withdrawal. Visually inspect the test sample to verify absence of tip or lumen collapse.Pass
    Corrosion Resistance TestISO 10555-1:2013Pass
    Tip Flexibility TestUse a mechanical force meter to push the tip from left to right at 5 mm, 10 mm, and 20 mm from the distal tip.Pass
    RadiopacityASTM F640-12Pass
    Static Burst PressureISO 10555-1:2013Pass
    Compatibility TestThe compatibility of the Distal Access Catheter with a marketed introducer sheath, guide catheter, guidewire, micro guidewire, microcatheter, and stent-retriever is evaluated.Pass
    Flow RateUse a flow rate tester to measure the flow rate of the Distal Access Catheter.Pass
    Biocompatibility Tests
    CytotoxicityViability > 70% (ISO 10993-5: 2009, MTT Method)Pass
    Skin SensitizationMagnusson & Kligman grades < 1 (ISO 10993-10: 2010, Guinea Pig Maximization Test (GPMT))Pass
    Intracutaneous Reactivity TestScore < 1.0 (ISO 10993-10: 2010, Intracutaneous Reactivity Study (Dermal))Pass
    Acute ToxicityNo acute toxicity (ISO 10993-11: 2017, Acute Toxicity in Mice)Pass
    Rabbit PyrogenNone of the animals showed a significantly greater reaction than the corresponding control animals. (ISO 10993-11: 2017, Rabbit Pyrogen Study, Material Mediated)Pass
    Complement ActivationThe thrombus formation score shall not be higher than control. (ISO 10993-4: 2017, In Vivo Thromboresistance Study in the Dog NAVI Mode)Pass
    In Vivo ThromboresistanceThe thrombus formation score shall not be higher than control. (ISO 10993-4: 2017, In Vivo Thromboresistance Study in the Dog NAVI Mode)Pass
    ASTM HemolysisHemolytic Index 0 - < 2% Non-Hemolytic (ASTM F756-2017, ASTM Hemolysis Study Direct Contact and Indirect Contact)Pass
    Bacterial Reverse MutationNo mutation effects (ISO 10993-3: 2014, Bacterial Reverse Mutation Test)Pass
    In Vitro Mammalian Cell MutationNo mutation effect (ISO 10993-3: 2014, Thymidine Kinase Gene)Pass
    Partial Thromboplastin Time StudyNo statistically significant difference between test article and control article or the negative reference control. (ISO 10993-4: 2017, ASTM F2382-2018, Partial Thromboplastin Time Study)Pass
    Sterilization & Shelf Life
    Sterilization Assurance Level (SAL)10⁻⁶ (ISO 11135:2014)Achieved
    EO and ECH ResidualsBelow limits specified in ISO 10993-7:2008Below specified limits
    Bacterial Endotoxin LevelsBelow 2.15 EU/device (USP <85>)Below 2.15 EU/device
    Shelf-lifeDevice performs as intended to support the proposed 2-year shelf-life.Demonstrates 2-year shelf-life

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical (bench and in vitro) and biocompatibility testing. It does not mention a "test set" in the context of patient data or clinical images. Therefore, questions regarding data provenance (country of origin, retrospective/prospective) are not applicable here.

    For the non-clinical and biocompatibility tests, the sample sizes are not explicitly stated for each individual test. The document generally states "test sample" or refers to the number of cycles/animals without specifying the exact number for statistical power. The tests are performed on the device itself or its extracted components, not on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This is a submission for a physical medical device (catheter) based on non-clinical and biocompatibility testing, not an AI/software device that requires expert-established ground truth for image interpretation or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As noted above, there is no "test set" in the context of clinical data for expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI/software device, so MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is not an AI/software device. The performance reviewed here is the physical device's standalone performance under various bench and in vitro conditions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for the non-clinical and biocompatibility tests are defined by the established industry standards and test methods (e.g., ISO 10555-1:2013, ISO 10993 series, ASTM F640-12). The acceptance criteria outlined within these standards serve as the "truth" against which the device's performance is measured. For example, for cytotoxicity, the ground truth is "Viability > 70%."

    8. The sample size for the training set:

    Not applicable. There is no training set as this is not an AI/software device.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213065
    Date Cleared
    2022-05-31

    (250 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Distal Access Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. It can be used to facilitate introduction of diagnostic agents or therapeutic devices. It is not intended for use in coronary arteries.

    Device Description

    The Distal Access Catheter consists of a catheter and accessories (hemostatic valve, introducer sheath, and a shaping mandrel). The catheter is a single-lumen, flexible catheter designed with coil and braid reinforcement. The distal segment is steamshapeable, and a hydrophilic coating is applied for navigation of the catheter through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy.

    The catheter body is constructed with a stainless-steel coil (less 2 cm of the catheter length) over the inner lumen liner comprised of polytetrafluoroethylene (PTFE). To provide additional shaft support, a stainless-steel wire braiding has been added over the stainless-steel coil from the proximal end to distal end. A platinum/iridium alloy radiopaque marker band is located at the distal tip of the catheter. An outer layer of varying durometers and lengths of polyamide (PA), polyether block amide (Pebax) and polyurethane (PU) covers the entire catheter body from proximal to distal end, respectively.

    A hub (PC) is attached to the proximal end of the catheter. A strain relief made from polyether block amide (Pebax) is placed at the proximal end of the catheter and distal end of the hub. The hub-strain relief provides for the kink resistance for the proximal end. A luer fitting on the catheter hub is used for the attachment of accessories.

    The outer surface of the catheter (distal 60 cm) is coated with a hydrophilic coating to reduce friction during navigation in the vasculature.

    A shaping mandrel (stainless steel, 80 mm in length) is provided with the catheter to be used by the physician for tip shaping. An introducer sheath (PTFE) is included to facilitate the introduction of the catheter into guide catheters during clinical use. A hemostatic valve is used to connect to the proximal section of the catheter.

    The Distal Access Catheter is provided sterile and for single use only. The catheter is placed in a dispenser tube (HDPE) and is placed on a packaging card (HDPE) that is provided in a sterile barrier PET/PE film and Tyvek pouch and placed in a carton box.

    AI/ML Overview

    This document describes the regulatory clearance for a medical device called a "Distal Access Catheter," not an AI/ML device. Therefore, the questions related to AI/ML device performance, such as sample size for test/training sets, expert ground truth, MRMC studies, and human-in-the-loop performance, are not applicable.

    The acceptance criteria and the study that proves the device meets them are based on bench testing (physical and mechanical properties) and biocompatibility testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria (Specification)Reported Device Performance (Conclusion)
    Bench Testing
    Surface InspectionExternal surface free from extraneous matter, process, and surface defects that could cause trauma to vessels.The surface integrity is suitable for intended clinical use.
    Surface ContaminationFree from surface contaminants from uncured coating surface particulates > 0.02 mm², embedded particulates. Distal tip smooth and tapered. PTFE inner layer not delaminated.The surface integrity is suitable for intended clinical use.
    DimensionalMeet specified dimensional requirements (catheter OD, ID, effective length, distal tip length, total accessories length).The device met the dimensional and physical specifications.
    Distal TipSmooth, rounded, tapered, or similarly finished to minimize trauma to vessels.Distal tip is suitable for intended clinical use.
    RadiodetectabilityTip of the catheter visible under fluoroscopy.Device radiopacity is suitable for intended clinical use.
    Corrosion ResistanceMetallic components show no signs of corrosion.Corrosion resistance is suitable for intended clinical use and met requirements of ISO 10555-1.
    Peak Tensile Force6F Catheter: ≥15N for distal section and hub/catheter junction. 5F Catheter: ≥10N for distal section and hub/catheter junction.Peak tensile force is suitable for intended clinical use and met requirements of ISO 10555-1.
    Fluid LeakageNo liquid leakage from hub and catheter shaft at 46 psi for 30 seconds.Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1.
    Air LeakageNo air leakage at hub into syringe when negative pressure was applied for 15 seconds.Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1.
    GaugingPlane of maximum diameter at opening of female conical fitting lies between two limit planes of the gauge.The device hub met the requirements of ISO 80369-7.
    Separation ForceMating parts separation force > 35 N.The device hub met the requirements of ISO 80369-7.
    Unscrewing TorqueTest article luer remains attached after applying unscrewing torque not less than 0.02 N·m for a minimum of 10 seconds.The device hub met the requirements of ISO 80369-7.
    Stress CrackingNo stress cracks on the test article hub.The device hub met the requirements of ISO 80369-7.
    Ease of AssemblyComponents fit together securely with no resistance between test article luer and reference fitting.The device hub met the requirements of ISO 80369-7.
    Resistance to OverridingTest article luer does not override reference fitting threads.The device hub met the requirements of ISO 80369-7.
    ParticulateAmount of particulate matter during simulated use determined and compared to predicate device.The amount and size of particles from the subject device was comparable to the predicate device.
    Static Burst PressureEvaluate burst pressure under static conditions per ISO 10555-1:2013(E) Annex F.Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1.
    Lubricity and DurabilityFrictional force equivalent to predicate device and less than 0.3N. No coating cracking or separation.Frictional force is suitable for intended clinical use, no coating cracking or separation, and is equivalent to the predicate device.
    Equipment InterfaceCompatible with 0.035" guidewire, ≥ 0.088" ID guide catheter/introducer sheath, ≤ 0.027" OD microcatheters, and common hemostatic valve.The device is compatible with the accessories as specified.
    Tip ShapeabilityDistal tip steam shapeable and equivalent to predicate devices.Shapeability of the distal tip after steam shaping is equivalent to predicate device.
    Kink ResistanceNo kinks when wrapped around pin gauges of clinical use relevant radii. No kinks during simulated use. Equivalent to predicate devices.The device is resistant to kinking around relevant radii turns.
    Simulated UseCatheter reaches target locations in tortuous vessel model; delivered and retracted smoothly with 0.035" guidewire; no device damage or defects after simulated use.Device performs as intended under simulated use conditions.
    Torque ResponseTorque response no worse than the predicate device.Device torque response is equivalent to the predicate device.
    Torque StrengthNo catheter breakage after 50 rotations.Device torque strength is equivalent to the predicate device.
    Pushability/RetractabilityPushability/retractability in tortuous vessel model no worse than the predicate device.Device pushability/retractability is equivalent to the predicate device.
    Catheter StiffnessCatheter stiffness equivalent to the predicate devices.The catheter stiffness is equivalent to the predicate device.
    Catheter Flexural FatigueNo flexural fatigue following repeated bending or hoop stress.The catheter flexural fatigue is equivalent to the predicate device.
    Dynamic BurstCatheter does not burst under dynamic pressure of 300 psi for 30 seconds.The device met the test acceptance criteria.
    Flow RateCatheter withstands manual injection of contrast media and saline at clinically relevant flow rates.The device can withstand flow rates suitable for intended clinical use.
    Ethylene Oxide ResidueResidual amount of ethylene oxide in single package ≤ 10µg/g.Ethylene Oxide Residue met the acceptance criteria per ISO 10993-7:2008.
    SterilityProduct shall be sterile.Sterility of the catheter met the acceptance criteria per ISO 11135:2014.
    Bacterial EndotoxinsEndotoxin content ≤ 2.15 EU/kit.Bacterial endotoxins met the acceptance criteria per USP <85/>.
    Biocompatibility Testing
    Cytotoxicity Test (ISO 10993-5)Cell viability of 100% test article extract shall be acceptable (not cytotoxic).The cell viability of 100% test article extract was 95.5%. (Conclusion: Non-cytotoxic)
    Skin Sensitization Test (ISO 10993-10)No significant evidence of skin sensitization.The positive rate of the test article was 0%. (Conclusion: No significant evidence of skin sensitization)
    Intracutaneous Reactivity Test (ISO 10993-10)Non-irritant. Final test article score acceptable.The final test article score was calculated to be 0. (Conclusion: Non-irritant)
    Acute Systemic Toxicity Test (ISO 10993-11)No significant evidence of systemic toxicity. Body weight data acceptable and equivalent between groups.Body weight data were acceptable and equivalent between the corresponding test and control treatment groups. (Conclusion: No significant evidence of systemic toxicity)
    Hemolytic Properties Test (ASTM F756)No influence on hemolytic properties.The hemolysis index was 0.44% (direct contact) and 0.00 (indirect contact). (Conclusion: No influence on hemolytic properties)
    Partial Thromboplastin Time (PTT) Test (ISO 10993-4)No effect on PTT—no significant differences between sample group and negative group.No significant differences between the sample group and the negative group. (Conclusion: No effect on PTT)
    In Vivo Thrombogenicity Test (ISO 10993-4)Equivalent to the control article—no significant differences between test and control articles.No significant differences between the test and control articles. (Conclusion: Equivalent to the control article)
    Pyrogen Test (ISO 10993-11)Non-pyrogenic—no rabbit shows an individual rise in temperature of 0.5 °C or more.No rabbit shows an individual rise in temperature of 0.5 °C or more. (Conclusion: Non-pyrogenic)
    Complement Activity (C3a, SC5b-9) Test (ISO 10993-4)Equivalent to the negative control group—no significant difference between sample group and negative control group.No significant difference between the sample group and negative control group. (Conclusion: Equivalent to the negative control group)

    Study Proving Device Meets Acceptance Criteria:

    The device met its acceptance criteria through a series of bench tests and biocompatibility evaluations.

    Bench Test Study:

    • Study Design: The bench testing of the Distal Access Catheter was performed using the applicable sections of the ISO 10555-1 international standard for sterile, single-use intravascular catheters. This standard specifies requirements for sterile, single-use intravascular catheters. The tests covered various physical and mechanical properties crucial for safe and effective device function.
    • Methodology: The document lists 24 distinct bench tests (e.g., Surface Inspection, Dimensional Verification, Corrosion Resistance, Peak Tensile Force, Fluid Leakage, Kink Resistance, Simulated Use, Dynamic Burst, Flow Rate, etc.). Each test had specific acceptance criteria derived from the ISO standard or established engineering specifications for the device. The results consistently concluded that the device was "suitable for intended clinical use," "met requirements," or was "equivalent to the predicate device."
    • Conclusion: The bench testing demonstrated that "the in vitro behavior of the device is well characterized within design specifications."

    Biocompatibility Study:

    • Study Design: The biocompatibility evaluation was conducted in accordance with FDA's biocompatibility guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"."
    • Methodology: The biological tests were conducted in compliance with the Good Laboratory Practice (GLP) Regulation 21 CFR 58. These tests included Cytotoxicity, Skin Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Hemolytic Properties, Partial Thromboplastin Time (PTT), In Vivo Thrombogenicity, Pyrogen, and Complement Activity tests. Each test had specific criteria for acceptable biological response. The conclusions for these tests consistently indicated that the device was non-cytotoxic, non-sensitizing, non-irritant, non-systemically toxic, had no effect on hemolytic properties or PTT, was equivalent to control/predicate for thrombogenicity and complement activity, and was non-pyrogenic.
    • Conclusion: The biocompatibility testing confirmed the device material safety for its intended biological contact.

    The following information is not applicable (N/A) as this is a traditional medical device (catheter) and not an AI/ML powered device.

    1. Sample size used for the test set and the data provenance: N/A (for AI/ML test sets). These are physical and biological tests, not data-driven evaluations.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (for AI/ML expert annotation). Ground truth for physical/biological tests is based on objective measurements against established standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: N/A (for AI/ML expert adjudication).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (not an AI-assisted device).
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A (not an AI algorithm).
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For this device, the "ground truth" or acceptance standard is derived from established international medical device standards (e.g., ISO 10555-1, ISO 80369-7, ISO 10993 series, USP <85/>) and internal design specifications, verified through objective bench and biological testing.
    7. The sample size for the training set: N/A (not an AI/ML device; no training set in this context).
    8. How the ground truth for the training set was established: N/A (not an AI/ML device).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1