Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K222246
    Date Cleared
    2022-11-21

    (118 days)

    Product Code
    Regulation Number
    878.4850
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Disposable Sterile Lancet

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Sterile Lancet: The lancet is intended for capillary blood sampling.

    Lancing Device:

    The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is single patient use only and should not be shared with anyone else, even a family member.

    Device Description

    Disposable Sterile Lancet is Class II Single Use Only Blood Lancet Without An Integral Sharps Injury Prevention Feature bearing the product code ORK (21CFR 878.4850), and The Disposable Sterile Lancet and lancing device is Class II Multiple Use Blood Lancet For Single Patient Use Only bearing the product code QRL (21CFR878.4850).

    The lancet is for single use, disposable sterile devices, which is designed to collect capillary blood samples. And the lancing device is intended for capillary blood sampling with Disposable Sterile Lancet. The lancet can be used independently, or used together with lancing device.

    The intended users include healthcare personnel, patients and lay users.

    The lancet is provided sterile. The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf life of the lancet is 5 years.

    AI/ML Overview

    This document describes a 510(k) submission for a Disposable Sterile Lancet and Lancing Device. The submission aims to demonstrate substantial equivalence to a predicate device (K220475).

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    ItemsAcceptance CriteriaReported Device Performance
    Appearance
    Product AppearanceNo obvious foreign body, fracture, unformed defects etc.Meet the requirements
    Needle-tipNo obvious burrs, curved hooks, oil stains etc. Lancet should be well assembled, no obvious dislocation at the fitting of the shellMeet the requirements
    Dimensions
    Diameter of needle coreProduct dimensions shall be consistent to the drawingsMeet the requirements
    Length of exposed needle-tipThe length of exposed needle-tip shall be consistent to the drawings.Meet the requirements
    Performance
    Puncture Force17-20G≤4N
    21-25G
    Ask a Question

    Ask a specific question about this device

    K Number
    K222055
    Date Cleared
    2022-11-15

    (126 days)

    Product Code
    Regulation Number
    878.4850
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Disposable Sterile Lancet

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is intended for capillary blood sampling.

    Device Description

    Disposable Sterile Lancet is Class II Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature bearing the product code FMK (21CFR878.4850). The lancet is for single use, disposable sterile devices, which is designed to collect capillary blood samples. The Disposable Sterile Lancet consist of protective cap, housing, bottom, fire button, connection ring, springs (ejection spring and fire spring), and core of lancet (including needle). The sterile part of the lancet is the needle tip. The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf life of the product is 5 years.

    AI/ML Overview

    The provided text is a 510(k) summary for a Disposable Sterile Lancet. It describes the device, its intended use, comparison to a predicate device, and the non-clinical testing performed to establish substantial equivalence.

    Based on the information provided, here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them:

    Device: Disposable Sterile Lancet
    Intended Use: Capillary blood sampling.

    1. Table of Acceptance Criteria and the Reported Device Performance

    Test ItemSpecific CharacteristicAcceptance CriteriaReported Performance (Results)
    AppearanceProduct AppearanceNo obvious foreign body, fracture, unformed defects etc.Meet the requirements
    Needle-tipNo obvious burrs, curved hooks, oil stains etc. Lancet should be well assembled, no obvious dislocation at the fitting of the shell.Meet the requirements
    DimensionsDiameter of needle coreProduct dimensions shall be consistent with the drawings.Meet the requirements
    Length of exposed needle-tipThe length of exposed needle-tip shall be consistent with the drawings.Meet the requirements
    PerformancePuncture DepthUsing A4 printing paper for simulated testing and meet the requirements.Meet the requirements
    Launch performanceThe operation of Fire should be smooth, after launching, the tip of the needle should be automatically retracted into the housing.Meet the requirements
    FirmnessNeedle should connect firmly with plastic handle.Meet the requirements
    SterilitySterilityThe sterility for the device shall be at SAL (Sterility Assurance Level) of 10^-6.Meet the requirements
    BiocompatibilityCytotoxicityUndetermined to be non-cytotoxic.Pass
    IrritationUndetermined to be non-irritating.Pass
    SensitizationUndetermined to be non-sensitizing.Pass
    Acute Systemic ToxicityNot an acute systemic cytotoxic potential.Pass
    PyrogenicityNo pyrogenicity reaction.Pass

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes for each non-clinical test conducted. It generally refers to "tests" and "results" without specifying the number of units tested.
    The data provenance is the manufacturer, Tianjin Rilifine Medical Device Co., Ltd., based in China. The study is a non-clinical study (laboratory testing) rather than a clinical study involving human patients, and therefore, the terms "retrospective" or "prospective" are not directly applicable in the same way they would be for patient data. It is a series of tests performed on the manufactured device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not mention the use of experts to establish "ground truth" for the non-clinical tests. These tests are based on objective measurements and established standards (e.g., ISO standards) rather than subjective expert consensus. For example, sterility is tested to a SAL of 10^-6, not determined by expert opinion.

    4. Adjudication Method for the Test Set

    Not applicable. As these are non-clinical, objective tests against pre-defined acceptance criteria (often tied to ISO standards), there is no mention of an adjudication method by multiple experts. The results are binary (meets/does not meet requirements, pass/fail).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI/image analysis devices where human readers interpret diagnostic images, and the AI assists in that interpretation. The Disposable Sterile Lancet is a physical medical device for blood sampling, and its performance is evaluated through physical and biological testing, not by human reader interpretation of data in a comparative effectiveness study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithmic device. The performance evaluated is the physical function and material properties of the lancet itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by objective measurements against predefined specifications and recognized international standards. For example:

    • Dimensions: Comparison to engineering drawings.
    • Sterility: Adherence to ISO 11137 standards for a SAL of 10^-6.
    • Biocompatibility: Adherence to ISO 10993 standards (e.g., non-cytotoxic, non-irritating, non-sensitizing).
    • Functional Performance (Puncture Depth, Launch Performance, Firmness): Measured against internal design specifications and simulated real-world use (e.g., A4 paper for puncture depth, observation of retraction).

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI models. The testing described is for a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1