AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lancet: Lancet is intended for capillary blood sampling.
Lancing Device: The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared.

Device Description

Lancet: The Lancet (use with lancing device) is a single use, sterile medical device, which is designed for use of micro blood sampling puncture to obtain capillary blood samples from the fingertip. The Lancet is composed three components: needle (made of stainless steel and silicone oil), main body and protective cap (both made of PE, EVA and calcium powder). Main body and protective cap is plastic part that enclosed the needle. The Lancet is intended to be single use and the needle tip is sterilized by Radiation. The shelf-life of the product is 5 years.
Lancing Device: Along with a lancet, the lancing device is used to collect a capillary blood sample. The body and the active parts of the lancing device are made of ABS, POM and PC Resin. And the spring is made of carbon steel. The service life of lancing device (HH-X-T, HH-XIII-T, HH-XV-T, HH-XVI-T, HH-XVIII-T, HH-XIX, HH-XXI-T, HH-XXII-T, HH-XXIII-T, HH-XXIV-T) is no more than 3000 times or 3 years. The service life of lancing device (HH-XVII-T) is no more than 5000 times or 5 years. The Lancing Device is provided non-sterile.

AI/ML Overview

The provided text describes the non-clinical testing performed to demonstrate that the proposed Lancet and Lancing devices meet acceptance criteria.

Here's the breakdown of the information requested:

1. A table of acceptance criteria and the reported device performance

ItemsAcceptance criteriaReported Device Performance
AppearanceThe surface shall be free of burr and no scratches.Meet the requirement
DimensionProduct dimensions shall be consistent with the drawings.Meet the requirement
Needle diameterThe diameter of the needle shall be consistent with the requirement.Meet the requirement
CleannessNo dust, no grease, no hair, no dirt.Meet the requirement
FirmnessNeedle should connect firmly with plastic handle.Meet the requirement
Resistance to corrosionCorrosion resistance of the needle of the lancet shall show no evidence of corrosion.Meet the requirement
Acidity or AlkalinityWhen determined with a laboratory pH meter and using a general purpose electrode, the pH value of an extract prepared shall be within one pH unit of that of the control fluid.Meet the requirement
Extractable MetalsWhen corrected for the metals content of the control fluid, contain not greater than a combined total of 5 mg/l of lead, tin, zinc and iron. The cadmium content of the extract shall, when corrected for the cadmium content of the control fluid, be lower than 0.1 mg/l.Meet the requirement
Puncture forceThe needle tip of the needle should have good puncture ability.Meet the requirement
LubricantVisual, should not be visible droplets.Meet the requirement
MatchingIt should be well assembled with the lancing devices, with no dislocation in the appearance coordination.Meet the requirement

Biocompatibility Testing:

ItemTest methodReported Device Performance
In Vitro CytotoxicityISO 10993-5: 2009No Cytotoxicity
Skin SensitizationISO 10993-10: 2010No Skin sensitization
Intracutaneous reactivityISO 10993-10: 2010No irritation
Acute Systemic ToxicityISO 10993-11: 2017No Acute Systemic Toxicity
PyrogenicityISO 10993-11: 2017no thermogenic reaction

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

For the simulated clinical use study, 500 device samples each for the Lancet and Lancing Device were used. The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective nature, but it is a "simulated clinical use study," implying it was conducted as part of the device's development/testing process. The manufacturer is Tianjin Huahong Technology Co., Ltd. in Tianjin, China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. The studies described are non-clinical bench testing and simulated clinical use for safety mechanisms, which typically do not involve expert-established ground truth in the same way imaging or diagnostic algorithms would.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/not provided for the types of non-clinical bench tests and simulated clinical use study described. Adjudication methods are more pertinent to studies involving human interpretation or subjective assessments.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. The submission explicitely states: "No clinical study is included in this submission." The device being described is a physical medical device (lancet and lancing device), not an AI-powered diagnostic or interpretive system that would typically undergo an MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a physical medical device (lancet and lancing device), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests (appearance, dimension, needle diameter, cleanness, firmness, corrosion resistance, acidity/alkalinity, extractable metals, puncture force, lubricant, matching) and biocompatibility tests, the "ground truth" is defined by the pre-established physical, chemical, and biological requirements and standards (e.g., drawings, ISO standards, specific chemical limits). For the simulated clinical use, the ground truth is whether the device "met the pre-established criteria" for its safety mechanism, as evaluated against FDA guidance and ISO 23908.

8. The sample size for the training set

Not applicable. This device is a physical medical device, not an AI or algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. No training set was used as this is not an AI/algorithm device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a larger font and the word "Administration" in a smaller font.

July 6, 2022

Tianjin Huahong Technology Co., Ltd. % Stuart Situ Director Landlink Healthcare Technology (Shanghai) Co., Ltd Room 1308, Baohua International Plaza, West Guangzhong Road 555, Jingan District Shanghai, 200072 China

Re: K220475

Trade/Device Name: Lancet (I, II, III, V, V); Lancing device (HH-X-T, HH-XIII-T, HH-XV-T, HH-XVI-T, HH-XVII-T, HH-XVIII-T, HH-XIX, HH-XXI-T, HH-XXII-T, HH-XXIII-T, HH-XXIV-T) Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: QRL, QRK Dated: June 6, 2022 Received: June 6, 2022

Dear Stuart Situ:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Colin Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220475

Device Name Lancet (I, II, III, III, V, VI); Lancing device (HH-X-T, HH-XIII-T, HH-XV-T, HH-XVI-T, HH-XVII-T, HH-XVII-T, HH-XIX, HH-XXI-T, HH-XXII-T, HH-XXIII-T, HH-XXIV-T)

Indications for Use (Describe) Lancet: Lancet is intended for capillary blood sampling.

Lancing Device:

The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) summary-K220475

| Submitter

Tianjin Huahong Technology Co., Ltd. A01, Plant B, No.278, Hangkong Road, Tianjin Pilot Free Trade Zone (Air Port Industrial Park), 300308 Tianjin, China

Establishment Registration Number: 3009498536

Contact person: Ms. Ying Yuan Quality Manager Tel.: +86-15302127671 E-mail: ying.yuan@hh-technology.com

Preparation date: Feb 15, 2022

II Proposed Device

Trade Name of Device:Lancet, Lancing device
Common name:Multiple Use Blood Lancet For Single Patient Use Only
Regulation Number:21 CFR 878.4850
Regulatory Class:Class II
Product code:QRL and QRK
Review PanelGeneral & Plastic Surgery

III Predicate Devices

510(k) Number:K192666
Trade name:Promosemed Blood Lancet
Classification:Class II
Product Code:FMK
ManufacturerPromisemed Hangzhou Meditech Co., Ltd.
510(k) Number:K113332
Trade name:On Call® Lancing Device
Classification:Class II
Product Code:FMK
ManufacturerACON Laboratories, Inc.

{4}------------------------------------------------

IV Device description

Lancet

The Lancet (use with lancing device) is a single use, sterile medical device, which is designed for use of micro blood sampling puncture to obtain capillary blood samples from the fingertip.

The Lancet is composed three components: needle (made of stainless steel and silicone oil), main body and protective cap (both made of PE, EVA and calcium powder). Main body and protective cap is plastic part that enclosed the needle. The Lancet is intended to be single use and the needle tip is sterilized by Radiation. The shelf-life of the product is 5 years.

Lancing Device

Along with a lancet, the lancing device is used to collect a capillary blood sample. The body and the active parts of the lancing device are made of ABS, POM and PC Resin. And the spring is made of carbon steel.

The service life of lancing device (HH-X-T, HH-XIII-T, HH-XV-T, HH-XVI-T, HH-XVIII-T, HH-XIX, HH-XXI-T, HH-XXII-T, HH-XXIII-T, HH-XXIV-T) is no more than 3000 times or 3 years. The service life of lancing device (HH-XVII-T) is no more than 5000 times or 5 years. The Lancing Device is provided non-sterile.

V Indication for use

Lancet

Lancet is intended for capillary blood sampling.

Lancing Device

The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared.

VI Comparison of technological characteristics with the predicate devices

The comparison and discussion between the subject device and the predicate devices are listed in below table 1&2:

Table 1 General Comparison of Lancet
ItemProposed devicePredicate deviceDiscussion
(K220475)(K192666)
ProductnameLancetPromosemed Blood LancetSame
ProductCodeQRL and QRKFMKDifferent¹
RegulationNo.21 CFR § 878.485021 CFR § 878.4850Same
ClassIIIISame
Prescription/over-the-counter useOver-The-Counter UseOver-The-Counter UseSame
Indication foruseIt is intended for capillaryblood samplingIt is intended for capillaryblood samplingSame
ApplicableuserHealthcare professional orlay personHealthcare professional orlay personSame
ReusedurabilitySingle useSingle useSame
Sterilizationmethod andSALSterilized by RadiationSAL=10⁻⁶Sterilized by RadiationSAL=10⁻⁶Same
Manufacturing aspectsFor the Lancet, stainlesssteel needle is fed into aninjection molding machine toover-mold plastic material(polyethylene (PE) andEthylene Vinyl Acetate (EVA)and calcium powder) forminga body and cap,encapsulating the stainlesssteel needles.For the Promisemed BloodLancet, stainless steelneedle is fed into aninjection molding machineto over-mold plasticmaterial (polyethylene)forming a body and cap,encapsulating the stainlesssteel needles.Same
Design andFunctionalityaspectsThe Lancet comprises astainless steel needleencapsulated with a plasticbody and cap, the cap istwisted off to expose theThe Promisemed BloodLancet comprises astainless steel needleencapsulated with a plasticbody and cap the cap isSame
needle for usetwisted off to expose the needle for use
Needle length range3.2±0.3mm (Model: IA、IB、IC、ID、IE、IK、IL、IM、IIA、IIB、III、VI)2.1±0.3mm (Model: V)unknownDifferent²
Gauge range1.50±0.02mm (16G)1.20±0.01mm (18G)1.07±0.01mm (19G)0.91±0.01mm (20G)0.82±0.01mm (21G)0.72±0.01mm (22G)0.64±0.01mm (23G)0.57±0.01mm (24G)0.51±0.01mm (25G)0.46±0.01mm (26G)0.41±0.01mm (27G)0.36±0.01mm (28G)0.34±0.01mm (29G)0.31±0.01mm (30G)0.26±0.01mm (31G)0.24±0.01mm (32G)0.21±0.01mm (33G)0.19±0.01mm (34G)unknown
Shelf-life5 years5 yearsSame
Materials of parts in contact with human bodyThe Lancet has a needle that is made of stainless steel and silicone oil. The body and cap are made of polyethylene (PE) and Ethylene Vinyl Acetate (EVA)The Promisemed Blood Lancet has a needle that is made of stainless steel and a body and a cap that are made of polyethylene (PE)Similar²
and calcium powder.
BiocompatibilityConforms to therequirements of ISO 10993series standards.Conforms to therequirements of ISO 10993series standards.Same
Label/LabelingComplied with 21 CFR part801Complied with 21 CFR part801Same

Table 1 General Comparison of Lancet

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

1 The product code is different because FDA issued the final order about Reclassification of Blood Lancets on 11/22/2021. Three new codes have been added of Blood Lancets. The product code QRL (Multiple Use Blood Lancet For Single Patient Use Only) and QRK (Single Use Only Blood Lancet Without An Integral Sharps Injury Prevention Feature) are closer to the purpose of our device.

2 The component and material of proposed devices is different from the predicate devices. However, all the materials are known biocompatible materials that have been used in lancets or other similar medical devices. After the comparison test of the I-type needle of our device and the predicate device, the result shows that the length of the exposed tip of this product (3.2±0.3mm) is similar to that of the predicate device, see the comparison test report (No.: VC-22032202). Model V (the length of the exposed tip is 2.1±0.3mm) can only be used in conjunction with the company's HH-XVIII-T lancing pen. After the performance test and the matching test, the product can be guaranteed to be safe and effective.

3 The needle length range or gauge range of the predicate device are unknown. There may be some differences between predicate device and proposed device. But the durability and strength of the needles were tested in a series of penetration tests using the proposed device, which shows positive result in these testing. Therefore, the differences do not raise concerns related to the device safety and effectiveness.

ItemProposed device(K220475)Predicate device(K113332)Discussion
Product nameLancing deviceGenteel Lancing deviceSame
Product CodeQRLFMKDifferent1
Regulation No.21 CFR § 878.485021 CFR § 878.4850Same

Table 2 General Comparison of Lancing device

{8}------------------------------------------------

ClassIIIISame
Prescription/over-the-counter useOver-The-Counter UseOver-The-Counter UseSame
Indication foruseThe Lancing Device is usedwith lancets to draw capillaryblood from the fingertip, fortesting utilizing smallamounts of blood. TheLancing Device is intendedto be used by a singlepatient and should not beshared.The On Call® LancingDevice is used with OnCall® disposable sterilelancets to draw capillaryblood from the fingertip,palm (at the base of thethumb) or forearm, forblood glucose testing orother testing utilizing smallamounts of blood. The OnCall® Lancing Device isintended to be used by asingle patient and shouldnot be shared.Same
Puncturedevice toobtain microbloodsamplesYesYesSame
Lancetretractedafter use topreventsharp injureYesYesSame
Devicepenetrationdepth range0.85mm(±0.30mm)~2.20mm(±0.30mm) (Model: HH-X-T);0.60mm(±0.30mm)~2.00mm(±0.30mm) (Model HH-XIII-T、HH-XXII-T);0.50mm(±0.30mm)~1.70mm(±0.30mm) (Model HH-XV-T);0.60mm(±0.30mm)~1.95mm(±unknownDifferent4
0.30mm) (Model HH-XVI-T、HH-XXI-T、HH-XXIII-T、HH-XXIV-T);
0.60mm(±0.30mm)~1.80mm(±0.30mm) (Model HH-XVII-T、HH-XIX);
0.60mm(±0.30mm)~2.10mm(±0.30mm) (Model HH-XVIII-T)
Mechanicalloading andfiring functionCocking barrel with releasingbuttonCocking barrel withreleasing buttonSame
MaterialsABS, POM and PC ResinABSSimilar5
ReusedurabilityReusableSingle Patient Use OnlyReusableSingle Patient Use OnlySame
Sterilizationmethod andSALNANASame
Label/LabelingComplied with 21 CFR part801Complied with 21 CFR part801Same

{9}------------------------------------------------

4 The Device penetration depth range of the predicate device is unknown. Krleza et al.[1] recommended that if the recommended puncture site is finger, the recommended incision depth is up to 1.5mm for child aged 6 months to 8 years, and up to 2.4mm for child older than 8 years and adults; if the recommended puncture site is heel, the recommended incision depth is up to 0.85mm for premature neonates(up to 3kg), and up to 2.0mm for infants under 6 months of age. The puncture depth of the product ranges from 0.50mm ( ± 0.30mm) to 2.2mm ( ± 0.30mm) covers the requirements of recommended incision depth. Therefore, this differences in the subject and predicate device do not raise concerns related to device safety and effectiveness.

5 The raw materials of proposed devices may different from the predicate devices. However, all the materials are medical grade materials that have been used in lancets or other similar medical devices.

{10}------------------------------------------------

VII Non-Clinical Testing

The bench testing performed verifies that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device in terms of critical performance characteristics. These tests are as follow.

ItemsAcceptance criteriaResults
AppearanceThe surface shall be free of burr and no scratches.Meet the requirement
DimensionProduct dimensions shall be consistent to the drawingsMeet the requirement
Needle diameterThe diameter of the needle shall be consistent to the requirementMeet the requirement
CleannessNo dust, no grease, no hair, no dirtMeet the requirement
FirmnessNeedle should connect firmly with plastic handle:Meet the requirement
Resistance to corrosionCorrosion resistance of needle of lancet shall show no evidence of corrosion.Meet the requirement
Acidity or AlkalinityWhen determined with a laboratory pH meter and using a general purpose electrode, the pH value of an extract prepared shall be within one pH unit of that of the control fluid.Meet the requirement
Extractable MetalsWhen corrected for the metals content of the control fluid, contain not greater than a combined total of 5 mg/l of lead, tin, zinc and iron. The cadmium content of the extract shall, when corrected for the cadmium content of the control fluid, be lower than 0.1 mg/l.Meet the requirement
Puncture forceThe needle tip of the needle should have good puncture ability.Meet the requirement
LubricantVisual, should not be visible droplets.Meet the requirement
MatchingIt should be well assembled with the lancing devices, with no dislocation inMeet the requirement
the appearance coordination.

{11}------------------------------------------------

Biocompatibility Testing:

The biocompatibility evaluations were conducted in accordance with the 2020 FDA Guidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" . The tests include the following tests:

ItemTest methodTest results
In Vitro CytotoxicityISO 10993-5: 2009No Cytotoxicity
Skin SensitizationISO 10993-10: 2010No Skin sensitization
Intracutaneous reactivityISO 10993-10: 2010No irritation
Acute Systemic ToxicityISO 10993-11: 2017No Acute Systemic Toxicity
PyrogenicityISO 10993-11: 2017no thermogenic reaction

Simulated Clinical Use

A simulated clinical use study was performed on 500 device samples each for the Lancet and Lancing Device according to FDA Guidance, Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 and ISO 23908 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.

VIII Clinical Testing

No clinical study is included in this submission.

IX Conclusion

The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.