K Number
K222055
Date Cleared
2022-11-15

(126 days)

Product Code
Regulation Number
878.4850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is intended for capillary blood sampling.

Device Description

Disposable Sterile Lancet is Class II Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature bearing the product code FMK (21CFR878.4850). The lancet is for single use, disposable sterile devices, which is designed to collect capillary blood samples. The Disposable Sterile Lancet consist of protective cap, housing, bottom, fire button, connection ring, springs (ejection spring and fire spring), and core of lancet (including needle). The sterile part of the lancet is the needle tip. The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf life of the product is 5 years.

AI/ML Overview

The provided text is a 510(k) summary for a Disposable Sterile Lancet. It describes the device, its intended use, comparison to a predicate device, and the non-clinical testing performed to establish substantial equivalence.

Based on the information provided, here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them:

Device: Disposable Sterile Lancet
Intended Use: Capillary blood sampling.

1. Table of Acceptance Criteria and the Reported Device Performance

Test ItemSpecific CharacteristicAcceptance CriteriaReported Performance (Results)
AppearanceProduct AppearanceNo obvious foreign body, fracture, unformed defects etc.Meet the requirements
Needle-tipNo obvious burrs, curved hooks, oil stains etc. Lancet should be well assembled, no obvious dislocation at the fitting of the shell.Meet the requirements
DimensionsDiameter of needle coreProduct dimensions shall be consistent with the drawings.Meet the requirements
Length of exposed needle-tipThe length of exposed needle-tip shall be consistent with the drawings.Meet the requirements
PerformancePuncture DepthUsing A4 printing paper for simulated testing and meet the requirements.Meet the requirements
Launch performanceThe operation of Fire should be smooth, after launching, the tip of the needle should be automatically retracted into the housing.Meet the requirements
FirmnessNeedle should connect firmly with plastic handle.Meet the requirements
SterilitySterilityThe sterility for the device shall be at SAL (Sterility Assurance Level) of 10^-6.Meet the requirements
BiocompatibilityCytotoxicityUndetermined to be non-cytotoxic.Pass
IrritationUndetermined to be non-irritating.Pass
SensitizationUndetermined to be non-sensitizing.Pass
Acute Systemic ToxicityNot an acute systemic cytotoxic potential.Pass
PyrogenicityNo pyrogenicity reaction.Pass

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the sample sizes for each non-clinical test conducted. It generally refers to "tests" and "results" without specifying the number of units tested.
The data provenance is the manufacturer, Tianjin Rilifine Medical Device Co., Ltd., based in China. The study is a non-clinical study (laboratory testing) rather than a clinical study involving human patients, and therefore, the terms "retrospective" or "prospective" are not directly applicable in the same way they would be for patient data. It is a series of tests performed on the manufactured device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not mention the use of experts to establish "ground truth" for the non-clinical tests. These tests are based on objective measurements and established standards (e.g., ISO standards) rather than subjective expert consensus. For example, sterility is tested to a SAL of 10^-6, not determined by expert opinion.

4. Adjudication Method for the Test Set

Not applicable. As these are non-clinical, objective tests against pre-defined acceptance criteria (often tied to ISO standards), there is no mention of an adjudication method by multiple experts. The results are binary (meets/does not meet requirements, pass/fail).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI/image analysis devices where human readers interpret diagnostic images, and the AI assists in that interpretation. The Disposable Sterile Lancet is a physical medical device for blood sampling, and its performance is evaluated through physical and biological testing, not by human reader interpretation of data in a comparative effectiveness study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithmic device. The performance evaluated is the physical function and material properties of the lancet itself.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established by objective measurements against predefined specifications and recognized international standards. For example:

  • Dimensions: Comparison to engineering drawings.
  • Sterility: Adherence to ISO 11137 standards for a SAL of 10^-6.
  • Biocompatibility: Adherence to ISO 10993 standards (e.g., non-cytotoxic, non-irritating, non-sensitizing).
  • Functional Performance (Puncture Depth, Launch Performance, Firmness): Measured against internal design specifications and simulated real-world use (e.g., A4 paper for puncture depth, observation of retraction).

8. The Sample Size for the Training Set

Not applicable. This device does not involve a "training set" in the context of machine learning or AI models. The testing described is for a physical medical device.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set was used.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

11/15/2022

Tianjin Rilifine Medical Device Co., Ltd. Qiusheng Jiang Manager No 32, Jingguan Road, Yixingbu, Beichen District Tianiin. 300402 China

Re: K222055

Trade/Device Name: Disposable Sterile Lancet Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: October 20, 2022 Received: October 21, 2022

Dear Qiusheng Jiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222055

Device Name Disposable Sterile Lancet

Indications for Use (Describe) It is intended for capillary blood sampling.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K222055

510(K) SUMMARY

(In accordance with 21 CFR 807.92)

1.0 Submitter's Information

Name:Tianjin Rilifine Medical Device Co., Ltd.
Address:No 32, Jingguan Road, Yixingbu, Beichen DistrictTianjin, 300402, CHINA
Phone Number:86-22-23233999
Contact:Mr. Qiusheng Jiang
Date of Preparation:July 5, 2022

2.0 Device Information

Device Name:Disposable Sterile Lancet
Common Name:Rilifine Safety Lancet
Classification Name:Single Use Only Blood Lancet With An Integral SharpsInjury Prevention Feature

3.0 Classification

Product Code:FMK
Regulation Number:21 CFR 878.4850
Classification:II
Review Panel:General & Plastic Surgery

4.0 Predicate Device Information

Manufacturer:Promisemed Hangzhou Meditech Co., Ltd.
Device:Promisemed Blood Lancet, VeriFine Safety Lancet,VeriFine Mini-Safety Lancet
510(k) Number:K192666
ClassificationI
Product CodeFMK

5.0 Intended Use

It is intended for capillary blood sampling.

6.0 Device Description

Disposable Sterile Lancet is Class II Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature bearing the product code FMK (21CFR878.4850).

{4}------------------------------------------------

The lancet is for single use, disposable sterile devices, which is designed to collect capillary blood samples.

The intended users include healthcare personnel, patients and lay users.

The Disposable Sterile Lancet consist of protective cap, housing, bottom, fire button, connection ring, springs (ejection spring and fire spring), and core of lancet (including needle).

The sterile part of the lancet is the needle tip. The sterile barrier is the needle sleeve and sterilized to a SAL of 10° by radiation sterilization. It is intended for single use only. The shelf life of the product is 5 years.

7.0 Comparison of Technological Characteristics with Predicate Device

ItemSubject DevicePredicate Device(K192666)ComparisonResult
ManufacturerTianjin Rilifine MedicalDevice Co., Ltd.Promisemed HangzhouMeditech Co., Ltd.--
510K number--K192666--
Product nameDisposable Sterile LancetPromisemed BloodLancet,VeriFine Safety Lancet,VeriFine Mini-SafetyLancet--
ClassificationIIISee Note 1
Product CodeFMKFMKSame
RegulationNumber21 CFR 878.485021 CFR 878.4800See Note 1
Intended UseIt is intended forcapillary bloodsampling.It is intended forcapillary bloodsampling.Same
Single UseSingle UseSingle UseSame
FeatureSafety protectionSafety protectionSame
Principles ofoperationPressure activatedPressure activatedSame
Sterilizationmethod and SALSterilized by radiationSAL=10-6Sterilized by radiationSAL=10-6Same
Shelf life5 years5 yearsSame

The following table is the summary of the technological characteristics, biocompatibility of the proposed subject device and predicate device.

{5}------------------------------------------------

ItemSubject DevicePredicate Device(K192666)ComparisonResult
Bio-compatibilityMaterials of partsin contact withhuman bodyNeedle: stainless steelOther parts: plasticmaterials, such as ABS, PPetc.Needle: stainless steelOther parts: plasticmaterialsSimilar, Seenote 2
Needle length0.7mm, 0.8mm, 0.9mm,1.0mm, 1.1mm, 1.2mm,1.3mm, 1.4mm, 1.5mm,1.6mm, 1.7mm, 1.8mm,1.9mm, 2.0mm, 2.1mm,2.2mm, 2.3mm, 2.4mm,2.5mm, 2.6mm, 2.7mm,2.8mm, 2.9mm, and3.0mm1.2mm, 1.4mm, 1.6mm,1.8mm, 2.0mm, 2.2mm,2.4mm, 2.6mm, 2.8mmSimilar, seenote 3
PrimarySkinIrritationUnder the condition ofstudy not an irritantUnder the condition ofstudy not an irritantSame
DermalSensitizationUnder the conditions of thestudy not a sensitizerUnder the conditions of thestudy not a sensitizerSame
CytotoxicityUnder the conditions of thestudy, no cytotoxic potentialUnder the conditions of thestudy, no cytotoxic potentialSame

Note 1: The classification and regulation number are different because FDA issued the final order about reclassification of Blood Lancet Nov 22, 2021.

Note 2: The raw materials of the proposed devices may be different from the predicate device. However, all the materials are known biocompatible materials that have been used in lancets or other similar medical devices.

Note 3: The length for the proposed device is wider than the predicate device K192666. Concerning the differences to the predicate devices, testing is done to the representative length of the proposed device, which are the extreme conditions for the device, that is length of 0.7mm (to review whether the length is long enough to get the required penetration), and length of 3.0mm (to review whether the length is too long to affect the retraction of the needle-tip after the product is activated). And based on the testing report, it is showed that the performance and safety feature for the products are not affected.

There are no significant differences between the two products and are identical in terms of intended use, design and technological characteristics.

{6}------------------------------------------------

8.0 Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 11137-1 First edition 2006-04-15, Sterilization of health care products -. Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)]
  • ISO 11137-2 Third edition 2013-06-01, Sterilization of health care products -Radiation - Part 2: Establishing the sterilization dose
  • ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
  • ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
  • ISO 10993-23 First edition 2021-01,Biological evaluation of medical devices -● Part 23: Tests for irritation

Performance testing verified that the proposed device is as safe, as effective and performs as well as the legally marketed predicate device in terms of critical performance characteristics as follows:

ItemsAcceptance CriteriaResults
AppearanceProduct AppearanceNo obvious foreign body, fracture, unformed defects etc.Meet the requirements
Needle-tipNo obvious burrs, curved hooks, oil stains etc. Lancet should be well assembled, no obvious dislocation at the fitting of the shellMeet the requirements
DimensionsDiameter of needle coreProduct dimensions shall be consistent to the drawingsMeet the requirements
Length of exposed needle-tipThe length of exposed needle-tip shall be consistent to the drawings.Meet the requirements
PerformancePuncture DepthUsing A4 printing paper for simulated testing and meet the requirementsMeet the requirements
Launch performanceThe operation of Fire should be smooth, after launching,Meet the requirements

{7}------------------------------------------------

the tip of the needle shouldbe automatically retractedinto the housing
FirmnessNeedle should connectfirmly with plastic handleMeet the requirements
SterilityThe sterility for the deviceshall be at SAL of 10-6Meet the requirements

Biocompatibility testing as per ISO 10993 standards:

ItemSubject deviceResult
CytotoxicityUnder the conditions of the study, the subject device extract was determined to be non-cytotoxic.Pass
IrritationUnder the conditions of the study, the subject device non-polar and polar extracts were determined to be non-irritating.Pass
SensitizationUnder the conditions of the study, the subject device non-polar and polar extracts were determined to be non-sensitizing.Pass
Acute systemic toxicityUnder the condition of this study, not an acute systemic cytotoxic potentialPass
PyrogenicityUnder the condition of this study, no pyrogenicity reactionPass

9.0 Clinical Test

No clinical study is included in this submission.

10. Comparison to the Predicate Device and Conclusion

The conclusion drawn from the nonclinical tests demonstrate that the subject device Disposable Sterile Lancet has the same indication for use and has similar design features and technological characteristic as the predicate device, and the proposed device is as safe, as effective and performs as well as the legally marketed predicate device K192666.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.