Search Results
Found 7 results
510(k) Data Aggregation
(130 days)
BubbleView Single-Use Digital Flexible Cystoscope (C50-C, C50-CR, C50-CB, C50-CBR); ViewHub® Video Processor
The BubbleView™ Single-Use Digital Flexible Cystoscope is a sterile, single-use, flexible endoscope intended to be used for endoscopic access to and examination of the lower urinary tract. The BubbleView™ is intended to provide visualization via ViewHub® video processor and can be used with endoscopic accessories.
BubbleView™ is intended for use in a hospital environment or medical office environment. BubbleView™ is designed for use in adults.
ViewHub® is designed to display live imaging data from compatible MACROLUX™ endoscopes.
The Single-Use Digital Flexible Cystoscope System consists of BubbleView™ Single-Use Digital Flexible Cystoscope (consists of a handle with control lever, endoscope cable and access port for accessories, and a flexible body portion) and ViewHub® Video Processor with itsaccessories including power cables. The BubbleView™ is provided sterile (sterilized by EQ) and intended to be single-use. The ViewHub® is a reusable multi-patient use device. The built-in LED at the Distal tip of the BubbleView™ Single-Use Digital Flexible Cystoscope provides a light source, the lens module captures the light signal, then the CMOS module converts the light signal into an electrical signal: the endoscope cable connects the BubbleView™ to the ViewHub®, which provides power and processes video signal from the endoscope; the control lever on the handleconnected to the controllable portion by a wire rope controls the bending direction and angle of the controllable portion: the instrument channel delivers water and other instruments. The ViewHub® Video Processor is a video imaging system that receives video signals from the connected endoscope, controls the lightat the endoscope tip and outputs this signal including a graphical user interface to a connected external video monitor.
The provided text is a 510(k) summary for the BubbleView™ Single-Use Digital Flexible Cystoscope and ViewHub® Video Processor. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria for an AI/ML-driven device. As such, it does not contain the detailed information requested regarding acceptance criteria, study methodologies for AI/ML performance, ground truth establishment, or multi-reader multi-case studies.
The document primarily describes:
- The device and its intended use.
- A comparison to a legally marketed predicate device (Pusen Single Use Flexible Video Cystoscope System K222602) to establish substantial equivalence based on indications for use and technological characteristics.
- Non-clinical performance tests for mechanical and optical characteristics. These tests are to support the safety and effectiveness of the hardware (the cystoscope and video processor) itself, not an AI component.
Therefore, I cannot extract the information requested about acceptance criteria and study data for an AI/ML device from this specific FDA submission document. The questions you posed are highly relevant for AI/ML medical devices, but this document appears to be for a traditional medical imaging device without an explicitly stated AI/ML component.
Ask a specific question about this device
(392 days)
C50
C50 is intended for clinical practice of general dentistry, as an aid in the diagnosis of pit and fissure caries, as an aid to highlight dental plaque and gingival inflammations (restricted to gingival inflammations which lead to bleeding upon probing) and as intra-oral camera to visualize anatomical details invisible to the naked eye or with a mirror (thanks to magnification).
The C50 is an intraoral video camera equipped with LEDs to illuminate the inspection site. The C50 optics and its complementary metal oxide semiconductor (CMOS) sensor capture the natural fluorescence of the site under observation and convert the images into a video signal that is sent to the computer via a USB interface. The dental practitioner can use the images displayed on the computer screen as an aid in diagnosis. A USB Type-C (camera) to USB-A cable is provided to connect the C50 to a computer. The C50 provides the following functions: Aid in the detection of pit and fissure caries Information about patient dental hygiene Highlighting of dental plaque Highlighting of gingival inflammations (restricted to gingival inflammation that leads to bleeding upon probing) Showing the difference between "before" and "after" care (follow-up). In CARIO mode (blue mode), the camera assists the dental practitioner by highlighting potential carious areas in pits and fissures on the occlusal surface of the teeth. In DAYLIGHT mode (white mode), the camera enables visualization of anatomical details invisible to the naked eye or with a mirror (thanks to its magnification). DAYLIGHT+ mode (white mode) is a shortcut version of DAYLIGHT mode where the contrast has been pushed from 20 to 70. In PERIO mode (yellow mode), the camera helps the dental practitioner to visualize dental plaque, in addition to highlighting areas of gingival inflammation (restricted to gingival inflammation that leads to bleeding upon probing).
The C50 is an intraoral camera intended for use in general dentistry. Its indications for use include aid in the diagnosis of pit and fissure caries, highlighting dental plaque and gingival inflammations (restricted to those leading to bleeding upon probing), and functioning as an intra-oral camera for magnified visualization of anatomical details.
Here's the breakdown of acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes comparison testing and other verification and validation tests rather than specific quantitative acceptance criteria for clinical performance metrics (e.g., sensitivity, specificity for caries detection). Instead, the performance testing focuses on demonstrating substantial equivalence to the predicate device and compliance with relevant safety and technical standards.
Acceptance Criteria Category | Specific Test/Performance Aspect | Reported Device Performance |
---|---|---|
Functional Equivalence | Clinical Modes (CARIO, DAYLIGHT, PERIO) | Demonstrated to be identical to the predicate device. |
New Mode (DAYLIGHT+) | Functions similarly to DAYLIGHT mode but offers higher contrast. | |
Image Quality (sharpness, size, resolution, focus position) | Comparison testing showed substantial equivalence to the predicate device, even with the Full HD resolution enhancement of the C50. | |
Safety and Performance | Temperature elevation in target tissue | Comparison testing showed substantial equivalence to the predicate device. |
Mechanical performance of reprocessed C50TIPS (maintains position, blocks external light) | Testing ensured the device maintains its position and completely blocks external light. | |
Fit of third-party protective sheath with C50 and C50TIPS | Mechanical testing confirmed appropriate fit. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | |
Electrical Safety (IEC 60601-1 and 60601-1-2) | Testing was conducted, and the device was found to be consistent with appropriate sections of the standards. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | |
EMC (IEC 60601-1-2, IEC/TR 60601-4-2) | Testing was conducted, and the device was found to be consistent with appropriate sections of the standards. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | |
Photobiological Safety (IEC 62471) | Testing was conducted, and the device was found to be consistent with the appropriate section of the standard. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | |
Software Validation (IEC 62304) | Validation was performed on the C50's firmware, consistent with the appropriate section of the standard. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | |
Reprocessing Validation (FDA Guidance Document "Reprocessing Medical Devices in Health Care Settings") | Validation was conducted on the C50, C50TIPS, and USB cable in accordance with FDA guidance. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | |
Biocompatibility (ISO 10993-1, FDA Guidance Document "Use of International Standard ISO 10993-1") (C50TIPS) | Chemical characterization, cytotoxicity, and acute systemic toxicity tests were considered, and acceptance criteria were met for all. The patient contact materials are identical in formulation, manufacturing, processing, and geometry to the predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a test set in the traditional sense of a clinical diagnostic study with patient data used for accuracy evaluation. Instead, it refers to "comparison testing of image quality" and "mechanical performance testing" conducted on the device and its accessories (C50, C50TIPS, protective sheath, USB cable).
- Sample Size: Not specified in terms of number of patients or cases. The testing appears to have been performed on the devices themselves and associated accessories.
- Data Provenance: Not applicable in the context of device performance testing. The "data" generated comes from direct laboratory and safety testing of the device hardware and software. No patient or human subject data is mentioned for performance evaluation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. As described above, the testing performed was focused on device characteristics and substantial equivalence to a predicate, not on a diagnostic accuracy study requiring expert ground truth establishment for patient cases.
4. Adjudication Method for the Test Set
Not applicable, as no clinical diagnostic test set with multiple reader interpretations requiring adjudication is described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
No MRMC comparative effectiveness study is described. The C50 is an intraoral camera and not an AI-powered diagnostic algorithm that assists human readers in interpretation. It provides visual information to the dental practitioner.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
The C50 is a hardware device (intraoral camera with different illumination modes) that provides visual information. It is not an algorithm, and therefore, a standalone algorithm performance study is not applicable. The device's "diagnosis aid" function relies on the visual information it provides to the human practitioner.
7. The Type of Ground Truth Used
The ground truth used for relevant testing was based on:
- Comparison to a predicate device (SOPROCARE): Many aspects of the C50's performance (image quality, temperature elevation, functional modes) were evaluated against the known performance of the predicate device to establish substantial equivalence.
- Compliance with established standards: Electrical safety, EMC, photobiological safety, software validation, reprocessing validation, and biocompatibility were evaluated against recognized international and FDA standards.
- Engineering specifications/functional requirements: Mechanical tests ensured specific physical functionalities like the fit of accessories and light blocking.
8. The Sample Size for the Training Set
Not applicable. The C50 is a medical device, not an AI/machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(243 days)
Capenergy C Equipment RF System - C25, C50, C100, C200, C300, C400, C500
The Capenergy C equipments RF System - C25, C50, C100, C200, C300, C400, C500 are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation The massage device provided is intended to provide a temporary reduction in the appearance of cellulite.
Capenergy C equipments RF System - C25, C50, C100, C200, C300, C400, C500 are indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
Capenergy C equipment's RF System - C25, C50, C100, C200, C300, C400, C500, including the hand piece, is a system generating RF energy with integral temperature and impedance feedback mechanism for procedures that require elevating tissue temperature. The Capenergy System constantly monitors the temperature and impedance of the target treatment tissue, automatically adjusting energy delivery to maintain effective and safe tissue heating.
Capenergy C equipment's RF System - C25, C50, C100, C200, C300, C400, C500 consists of an AC/DC power supply unit, RF generator, controller and user interface. The RF hand piece is connected to the console via a cable and a switch activates the energy delivery to the hand piece. The hand piece is comprised of conductive and capacitive electrodes.
The provided document, a 510(k) Premarket Notification Submission for the Capenergy C Equipment RF System, does not describe the acceptance criteria and a study proving the device meets those criteria in the typical format of a clinical study assessing diagnostic accuracy or comparative effectiveness.
Instead, this document focuses on demonstrating substantial equivalence to existing predicate devices based on non-clinical performance testing and conformance to voluntary standards. The "acceptance criteria" here are primarily functional specifications and regulatory compliance rather than statistical performance metrics from a human-in-the-loop or standalone study.
Here's a breakdown of the information that can be extracted, and where the requested information is not applicable or not provided in this specific document:
1. Table of Acceptance Criteria and Reported Device Performance
Criterion / Performance Aspect | Acceptance Criteria (or Standard) | Reported Device Performance |
---|---|---|
Functional Performance: | ||
Ability to Reach Therapeutic Temperature | Not explicitly stated as an AC, but implied by intended use and predicate comparison (40-45°C) | "The device demonstrated ability to reach and maintain therapeutic temperature (40-45°C) on the surface of human skin for at least 10 minutes." |
RF Power Output | 45 W +/-10% for a charge of 06-j530 ohms to 1 MHz (from predicate) | Subject device matched predicate: 45 W +/-10% for a charge of 06-j530 ohms to 1 MHz |
Internal Cut-Off Temperature | 40-45°C (from predicate) | Subject device matched predicate: 40-45°C |
Treatment Time | 15-660 sec (from predicate) | Subject device matched predicate: 15-660 sec |
RF Frequency | Fixed values, 0.8MHz, 1.0 MHz, 1.2 MHz +/-25% (from predicate) | Subject device offered additional frequency: Fixed values are established for the treatment time, percentage of power output and working frequency. Four available options: 0.448 MHz +/-25%, 0,8MHz +/-25%, 1,0 MHz +/-25%, 1,2 MHz +/-25% |
Waveform | Sinusoidal (from predicate) | Subject device matched predicate: Sinusoidal |
Safety and Compliance: | ||
Electrical Safety | Conformance to IEC 60601-1:2005 +/A1:2012, IEC 60601-2-2:2017 | "Functional laboratory testing... showed correct operation of the device," and "Electrical safety" testing performed. (Specific test results not detailed, but conformance is asserted) |
Electromagnetic Compatibility (EMC) | Conformance to IEC 60601-1-2:2015 | "Functional laboratory testing... showed correct operation of the device," and "Electromagnetic compatibility" testing performed. (Specific test results not detailed, but conformance is asserted) |
Software Verification & Validation | Conformance to IEC 62304:2006 and FDA guidance | "Software verification and validation was conducted to IEC 62304: 2006... and FDA guidance... The results of this testing conclude the software has met these requirements." |
Biocompatibility (Patient-Contacting) | Conformance to ISO 10993-1:2009 | "Patient contacting materials have been evaluated according to the requirements of ISO 10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation and confirmed to be biocompatible for their intended use." |
Safety Class Protection | Class I – Type BF (from predicate) | Subject device matched predicate: Class I – Type BF |
2. Sample Size Used for the Test Set and the Data Provenance
This document primarily describes non-clinical, bench testing, and compliance with standards. There is no mention of a clinical "test set" in the context of patient data or clinical images for evaluating diagnostic performance. The "test set" implicitly refers to the physical devices themselves and their components undergoing various performance and safety assessments.
- Sample Size: Not applicable in the context of clinical data. For bench testing, typically multiple units or components are tested, but the exact count is not specified.
- Data Provenance: Not applicable in the context of clinical data. The testing is reported as "Functional laboratory testing performed in foreseeable operating conditions." This suggests internal laboratory testing rather than external clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This is not applicable to this type of submission. Ground truth, in the context of diagnostic or AI-based devices, refers to a definitive diagnosis or finding against which the device's performance is measured (e.g., pathology report, expert consensus on images). This submission pertains to an energy-based therapeutic device and its safety/functional performance, not a diagnostic AI system requiring expert-adjudicated ground truth.
4. Adjudication Method for the Test Set
This is not applicable for the same reasons as point 3. There is no diagnostic "test set" of cases requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The Capenergy C equipment RF System is a therapeutic device intended for "topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions" and "electrocoagulation and hemostasis." It is not an AI-assisted diagnostic device, and therefore, no MRMC study or assessment of human reader improvement with AI assistance would be relevant or performed for this product.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. As a therapeutic RF energy system, the device's primary function is to deliver energy for heating and coagulation. It's not an "algorithm-only" or "standalone" diagnostic system. Its performance relates to its physical outputs (RF power, temperature control, etc.), which were evaluated through bench testing.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This is not applicable in the sense of clinical ground truth for diagnostic purposes. The "ground truth" for this device's performance relies on:
- Engineering specifications and standards: Conformance to electrical safety, EMC, software, and biocompatibility standards (e.g., IEC 60601 series, IEC 62304, ISO 10993-1).
- Physical measurements: The ability to achieve and maintain specific temperatures (40-45°C) and power outputs.
- Comparison to predicate devices: Establishing substantial equivalence by showing that the subject device operates safely and effectively within similar parameters as legally marketed devices.
8. The sample size for the training set
This is not applicable. This device is a hardware-based therapeutic system, not an AI/machine learning model that undergoes "training." Its "software" (evaluated to IEC 62304) controls the device's functions, but this is a traditional software development and validation process, not machine learning.
9. How the ground truth for the training set was established
This is not applicable for the same reasons as point 8.
Ask a specific question about this device
(17 days)
DigitalDiagnost C50
The DigitalDiagnost C50 system is intended for use in generating radiographic images of human anatomy by qualified/ trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.
The DigitalDiagnost C50 Digital Radiography System (DigitalDiagnost C50) is a flexible digital radiography (DR) system that is designed to provide fast and smooth radiography examinations of sitting, standing or lying patients. The DigitalDiagnost C50 consist of the following components: ceiling suspension with X-ray assembly, wall stand with detector carrier, patient table with detector carrier and floating table top, high voltage generator, and an acquisition and reviewing workstation for post-processing, storage and viewing of images. Images may be transferred via a DICOM network for printing, storage and detailed review.
Here's a breakdown of the acceptance criteria and study information for the DigitalDiagnost C50, based on the provided text.
1. Table of Acceptance Criteria & Reported Device Performance:
Feature | Acceptance Criteria (Predicate) | Reported Device Performance (DigitalDiagnost C50) | Comment on Equivalence |
---|---|---|---|
Basic information | |||
Product Code | KPR | Identical (KPR) | No difference; thus, demonstrating SE. |
Regulation No. | 21 CFR 892.1680 | Identical (21 CFR 892.1680) | No difference; thus, demonstrating SE. |
Device Class | II | Identical (Class II) | No difference; thus, demonstrating SE. |
Electrical Requirement | Input voltage: 3-phase, 200/208/240/380/400/415/440/480/500Vac; Frequency:50/60Hz; Current: Short term: 112A (with generator M-CABINET CXA 50kW), 134A (M-CABINET CXA 65kW); Long term: 10A. | Identical | No difference; thus, demonstrating SE. |
Design characteristic | |||
X-ray Tube | RO 1750 ROT 380 & SRO 33100 ROT 380 | Identical | No difference; thus, demonstrating SE. |
Max Tube Voltage | 150 kV | Identical | No difference; thus, demonstrating SE. |
Focal Spot Size | 0.6mm/1.2mm | Identical | No difference; thus, demonstrating SE. |
Tube Max power | 50KW/100KW (250W equivalent anode input power) | Identical | No difference; thus, demonstrating SE. |
Anode Type | Rotation | Identical | No difference; thus, demonstrating SE. |
Generator | Philips Healthcare (Suzhou), M-CABINET CXA Pro 50kW, M-CABINET CXA Pro 65kW | Identical | No difference; thus, demonstrating SE. |
Max Power | 50KW/65KW | Identical | No difference; thus, demonstrating SE. |
KV range | 40-150 | Identical | No difference; thus, demonstrating SE. |
Milli ampere sec (mAs) product | 0.4 mAs-600 mAs (with AEC control) | Identical | No difference; thus, demonstrating SE. |
Collimator | |||
Operation Mode | Manual collimation | Identical | No difference; thus, demonstrating SE. |
Shape of Beam | Rectangular | Identical | No difference; thus, demonstrating SE. |
Detector | |||
Type | Digital Detector, Fixed RAD Detector, Wireless Static Detector | Identical | No difference; thus, demonstrating SE. |
Fixed RAD Detector | Pixium 4343RG | Pixium 4343RCE (identical to predicate DigitalDiagnost C90 K182973) | The difference between 4343RCE and 4343RG is scintillator material and other minor differences of image area and image matrix. The differences don't affect the safety or effectiveness. Thus, demonstrating SE. |
Wireless Static Detector | Varian PaxScan4336W | SkyPlate E (Trixell 3543DR) (identical to reference MobileDiagnost WDR 2.2 K191813) | The Skyplate E Detector is identical to the Skyplate E Detector of the currently marketed and Reference Device 1, MobileDiagnost WDR 2.2 (K191813- August 2, 2019). Thus, demonstrating SE. |
X-ray Scintillator Material (Fixed) | GdOS (Fixed: Pixium 4343RG) | Cesium Iodide (Fixed: Pixium 4343RCE) | The fixed RAD detector of the proposed DigitalDiagnost C50 and currently marketed and Reference Device 3, DigitalDiagnost C90(K182973- January 11, 2019, Philips Medical Systems DMC) are both fabricated from identical Cesium Iodide material, thus demonstrating SE. |
X-ray Scintillator Material (Wireless) | GdOS (Wireless Varian PaxScan4336W) | Cesium Iodide (Wireless SkyPlate E) | The Wireless Detector of the proposed DigitalDiagnost C50 and the currently marketed and Reference Device 1, MobileDiagnost WDR 2.2 (K191813- August 2, 2019) are both fabricated from identical Cesium Iodide material, thus demonstrating SE. |
Image Area (Fixed) | 42.5cm x 42.5cm (Fixed: Pixium 4343RG) | 42.03cm x 42.54cm (Fixed: Pixium 4343RCE) | The image area of the DigitalDiagnost C50, provided with fixed RAD detector is identical to the image area of the Fixed RAD Detector of the currently marketed and Reference Device 3, DigitalDiagnost C90(K182973- January 11, 2019, Philips Medical Systems DMC), thus demonstrating substantial equivalence. |
Image Area (Wireless) | 42.7 cm x 34.4 cm (Wireless: Varian PaxScan4336W) | 34.5 cm x 42.5cm (Wireless: SkyPlate E) | The image area of the proposed Wireless Detector SkyPlate E is identical to the Wireless Detector SkyPlate E of the currently marketed and reference device 1, MobileDiagnost WDR 2.2 (K191813- August 2, 2019, Sedecal SA). thus demonstrating SE. |
Image Matrix (Fixed) | 2,874 x 2,869 (Fixed: Pixium 4343RG) | 2,874 x 2840 (Fixed: Pixium 4343RCE) | The image matrix of the proposed DigitalDiagnost C50, provided with fixed RAD detector is similar to the image matrix of the currently marketed and Reference Device 3, DigitalDiagnost C90. Thus demonstrating substantial equivalence. Infinitesimal change in the image size (X-ray field) does not impact clinical Image Quality. Therefore, they are equivalent and there is no impact on the safety and effectiveness of the device; thus, demonstrating SE. |
Image Matrix (Wireless) | 3,072 x 2,476 (Wireless: Varian PaxScan4336W) | 2,156 x 2,653 (Wireless-SkyPlate E) | The image matrix of the proposed DigtialDiagnost C50 provided with wireless detector SkyPlate E is similar to the currently marketed and Reference Device 1, MobileDiagnost WDR 2.2. Infinitesimal change in the image size (X-ray field) does not impact clinical Image Quality. Therefore, they are equivalent and there is no impact on the safety and effectiveness of the device; thus, demonstrating SE. |
Pixel Size (Fixed) | 148 μm (Fixed: Pixium 4343RG) | 148 μm (Fixed: Pixium 4343RCE) | The pixel size of the proposed DigtialDiagnost C50 provided with fixed RAD detector is identical to the pixel size of the currently marketed and Reference Device 3, DigitalDiagnost C90. Thus demonstrating substantial equivalence. |
Pixel Size (Wireless) | 139 μm (Wireless: Varian PaxScan4336W) | 160 μm (Wireless-SkyPlate E) | The pixel size of the proposed DigtialDiagnost C50 provided with wireless detector SkyPlate E is identical to the currently marketed and Reference Device 1, MobileDiagnost WDR 2.2. Thus demonstrating SE. |
Analog / Digital (A/D) conversion | 16 bits | Identical | No difference; thus, demonstrating SE. |
Table | |||
Table type | Fixed and Height adjustment | Identical | No difference; thus, demonstrating SE. |
Height adjustment | 51.5 cm to 91.5 cm above floor, motorized adjustment | Identical | No difference; thus, demonstrating SE. |
Tabletop longitudinal travel range | +/- 550mm | Identical | No difference; thus, demonstrating SE. |
Tabletop Lateral travel | +/- 130mm | Identical | No difference; thus, demonstrating SE. |
Loading (patient weight) | 210 Kg | Identical | No difference; thus, demonstrating SE. |
Wall Stand | |||
Vertical movement range | 1500mm | Identical | No difference; thus, demonstrating SE. |
Movement mode | Manual | Identical | No difference; thus, demonstrating SE. |
Tube Stand | |||
Type | Ceiling suspension | Identical | No difference; thus, demonstrating SE. |
Tube rotation | +/- 135 degree | Identical | No difference; thus, demonstrating SE. |
Longitudinal movement range | 1500 mm | Identical | No difference; thus, demonstrating SE. |
Source to Image Distance (SID) | |||
SID | SID depends on different configurations, because the DigitalDiagnost C50 is a ceiling suspension X-ray system. | Identical | No difference; thus, demonstrating SE. |
External Connectivity | |||
DICOM | DICOM 3.0 compatible | Identical | No difference; thus, demonstrating SE. |
Software Platform | |||
Software | Eleva WorkSpot | Identical | No difference; thus, demonstrating SE. |
SkyFlow Software | No | Yes | The proposed Digital Diagnost C50 includes the SkyFlow software used in the currently marketed and reference device Eleva Workspot with SkyFlow, (K153318- December 22, 2015, Philips Medical Systems DMC), thus, demonstrating SE. |
Image Processing Algorithm | UNIQUE | UNIQUE 2 | UNIQUE 2 image processing algorithm provided with DigitalDiagnost C50 was previously cleared with the currently marketed and Reference Device 3, DigitalDiagnost C90 (K182973- January 11, 2019, Philips Medical Systems DMC). Upgrading to UNIQUE2 image processing algorithm does not alter the clinical workflow, hence no impact on the safety or effectiveness of the device; thus, demonstrating SE. |
Operating System | Windows 7 embedded | Windows 10 embedded | Introduction of operating system Windows 10 embedded does not impact clinical image quality. Therefore, there is no impact on the safety and effectiveness of the device; thus, demonstrating SE. |
Study Proving Acceptance Criteria:
The document states that a non-clinical verification/validation study was performed. The basis for substantial equivalence is that the DigitalDiagnost C50 is considered equivalent to the predicate DigitalDiagnost C50 (K163410) based on:
- Identical Indications for Use.
- Fundamental scientific technology (including largely identical components or components identical to other cleared reference devices).
- Non-clinical performance testing (demonstrating compliance with standards).
- Safety and effectiveness (as supported by the non-clinical testing).
The non-clinical verification and validation tests demonstrated that the DigitalDiagnost C50:
- Complies with the aforementioned international and FDA-recognized consensus standards and FDA guidance documents.
- Meets the acceptance criteria and is adequate for its intended use.
2. Sample size used for the test set and the data provenance:
The document explicitly states: "The DigitalDiagnost C50 does not require clinical study..." and "Summary of Clinical Data: Based on the information provided above, the DigitalDiagnost C50 is considered substantially equivalent to the currently marketed and predicate DigitalDiagnost C50 (K163410, January 4, 2017) in terms of fundamental scientific technology."
Therefore, there appears to be no clinical test set, sample size, or specific data provenance for a clinical study as part of this 510(k) submission. The evaluation relies on non-clinical performance data and equivalence to predicate/reference devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Since no clinical study with a test set was performed, this information is not applicable and not provided in the document.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Since no clinical study with a test set was performed, this information is not applicable and not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was mentioned. The device described, DigitalDiagnost C50, is an X-ray system, not an AI-assisted diagnostic device. Therefore, this information is not applicable and not provided in the document.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The DigitalDiagnost C50 is an X-ray imaging system. There is no mention of a standalone algorithm-only performance study, as it's not a standalone diagnostic algorithm. Therefore, this information is not applicable and not provided in the document.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the non-clinical verification and validation tests, the "acceptance criteria" were derived from international and FDA-recognized consensus standards and FDA guidance documents (listed on pages 11-12). These standards themselves implicitly define the "ground truth" or acceptable performance ranges for device characteristics, image quality, electrical safety, radiation protection, software lifecycle, usability, and risk management. No other forms of expert consensus, pathology, or outcomes data are mentioned for establishing ground truth for the device's technical specifications.
8. The sample size for the training set:
Since no clinical study or AI algorithm training is described for this device, this information is not applicable and not provided in the document.
9. How the ground truth for the training set was established:
Since no clinical study or AI algorithm training is described for this device, this information is not applicable and not provided in the document.
Ask a specific question about this device
(281 days)
CAPENERGY - C100, C200, C300, C400, C50
The CAPENERGY C Devices - C100, C200, C300, C400, C50 are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation. The massage device provided is intended to provide a temporary reduction in the appearance of cellulite.
The proposed device is an equipment for diathermy by energy transfer using radio frequency alternating electrical currents applied to the patient by using electrodes or plates. The RF current is coupled to the human body by means of an insulated or non-insulated electrode -the active accessory-, which together with the body and the return electrode -the passive plate- forms an electric circuit, which allows the passage of RF current through the body areas situated between the two electrodes. The number of available channels on the different models of the CAPENERGY product range varies from one to four. When a model has several channels, these operate independently. Each channel has a keyboard that lets you manage the parameters of frequency and channel power. It also has two light columns that report the selected power level and a qualitative indication of the power delivered by the channel. The user interface has a screen and a keyboard that allows you to manage several functions: timer, connected accessories management, state management, energy and temperature measurements and alarms. The user interface is organized in a menu system that makes it possible to choose between the different features available on the equipment.
The provided text is a 510(k) Premarket Notification for the CAPENERGY C Devices (C100, C200, C300, C400, C50). This document is primarily focused on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed clinical study for the device itself. Therefore, much of the requested information regarding acceptance criteria and a study proving the device meets them, particularly in the context of AI/ML performance, is not present in this document.
However, based on the non-clinical data section, we can infer some information about performance testing.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Ability to reach and maintain therapeutic tissue temperature (40-45°C) on the surface of human skin for at least 10 minutes | "demonstrated ability to reach and maintain therapeutic temperature (40-45℃) on the surface of human skin for at least 10 minutes." |
Correct operation according to intended use (general) | "Functional laboratory testing performed in foreseeable operating conditions showed correct operation of the device as per its intended use" |
Electrical Safety (IEC 60601-1:2005 + A1:2012) | Compliance met (implied by non-clinical summary) |
Electromagnetic Compatibility (IEC 60601-1-2:2015) | Compliance met (implied by non-clinical summary) |
Software Validation (IEC 62304:2006, FDA guidance) | "results of this testing conclude the software has met these requirements." |
Biocompatibility of patient-contacting materials (ISO 10993-1:2009) | "confirmed to be biocompatible for their intended use." |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not specified. The document states "bench testing" and "functional laboratory testing," which typically refers to testing of the device itself rather than human or image data. No patient or image test set is mentioned.
- Data Provenance: Not applicable, as no human or image data test set is described. The testing appears to be primarily laboratory-based device performance testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: Not applicable. No ground truth based on expert review of patient data is mentioned.
- Qualifications of Experts: Not applicable.
4. Adjudication method for the test set
- Adjudication Method: Not applicable. No test set requiring adjudication (e.g., image review) is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC comparative effectiveness study was not done. The device description and intended use (diathermy device) do not involve human readers interpreting images or data with or without AI assistance.
- Effect Size of AI Improvement: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: Not applicable. This document describes a physical medical device (diathermy) and its performance, not an AI algorithm. The software validation mentioned is for the device's operational software, not an AI component making diagnostic or treatment recommendations.
7. The type of ground truth used
- Type of Ground Truth: For the performance tests mentioned (temperature maintenance, electrical safety, EMC, software validation, biocompatibility), the "ground truth" would be established by engineering specifications, relevant international standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 10993-1), and internal performance targets.
8. The sample size for the training set
- Sample Size for Training Set: Not applicable. This document does not describe the development or testing of an AI/ML algorithm that would require a training set.
9. How the ground truth for the training set was established
- How Ground Truth was Established: Not applicable, as no training set for an AI/ML algorithm is mentioned.
Ask a specific question about this device
(232 days)
C50 and C80 Multi-parameter Patient Monitor
The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
· The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
· NIBP measurement continual mode is not applicable to neonates;
· Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
The monitors are to be used in general healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. The monitors are not intended for home use.
The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
• The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
• NIBP measurement continual mode is not applicable to neonates;
• Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
Those monitors provide patient monitoring capabilities by using corresponding accessories.
The multi-parameter Patient monitor, model: C50 and C80 have same design principle and technical characteristics:
But the software in the host and modules and components are different.
The differences between C50 and C80 are ICG, BIS and size of monitor.
The provided document describes the acceptance criteria and a study proving the C50 and C80 Multi-parameter Patient Monitor meets these criteria.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established through adherence to recognized international standards and clinical performance metrics. The document details a comparison with a predicate device (K170876, Passport Series Patient Monitors). While explicit "acceptance criteria" for each parameter are not always separately listed with a single performance value from the new device, the "Comparison" column in the table below infers if the new device meets or exceeds the predicate's performance or if sufficient justification is provided for differences.
Feature | Predicate Device (K170876) Performance | Subject Device (C50 & C80) Performance | Comparison to Acceptance Criteria (Predicate) & Justification |
---|---|---|---|
Monitor Size | Passport 17m: 400x370x193mm; Passport 12m: 297x336x187mm | C80: Approx. 344x291x165mm; C50: Approx. 291.7x250x187mm | The monitor size of the subject device is smaller than the predicate device, but it "won't affect the safety and effectiveness of the subject device." Implied acceptance: smaller size is acceptable if safety/effectiveness are not compromised. |
Integrated display & touch screen | Color TFT LCD (17-inch, 1280x1024 pixels for 17m; 12-inch, 800x600 pixels for 12m) | Color TFT LCD (C80: 12.1-inch, 800x600 pixels; C50: 10.4-inch, 800x600 pixels) | Screen size is smaller, but resolution for C80/C50 is the same as Passport 12m. "Same. The screen of Subject device is smaller than predicate device. Complying with IEC 60601-1 and IEC 62366-1 also indicates the clinical use is safe and effective." Implied acceptance: smaller screen size is acceptable due to compliance with relevant IEC standards for safety and usability. |
Power supply | Two rechargeable Lithium-ion batteries or AC for Passport 17m; One rechargeable Lithium-ion battery or AC for Passport 12m | C50 and C80: Powered either by built-in battery or external AC. | "Same, both powered by battery and AC." Implied acceptance: functional equivalence. |
Battery | 11.1V, 4500mAh | 11.1V, 2200mAh/4400mAh | "The battery capacity is different. The batteries of C50/C80 have complied with IEC 62133. C50 and C80 have conformed to IEC 60601-1." Implied acceptance: different capacity is acceptable as long as safety (IEC 62133, IEC 60601-1) is maintained. |
ECG | 3-lead, 5-lead and 12-lead selectable, heart rate (HR) | 3-lead, 5-lead and 12-lead selectable, heart rate (HR) | "Same." Implied acceptance: functional equivalence. |
ECG (Arrhythmia Analysis) | Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, Ron T, Run PVCs, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm, AFib (24 types) | Asystole, ventricular fibrillation, R ON T, VT >2, Couplet, PVC, Bigeminy, Trigeminy, Brady (Bradycardia), PNC (Pacer Not Capture), PNP (Pacer Not Pace), Missed Beats, IHB (Irregular Heart Beat), VTAC (Ventricular Tachycardia), Tachy (Tachycardia), PVC Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm, Heart Pause (20 types) | "C50/C80 has 20 types of arrhythmias in total, 18 of which is same as the predicate device. The other two (underline) is supported by compliance with EC57." Implied acceptance: fewer types are acceptable as long as key arrhythmia detections are present and supported by EC57 compliance. |
Respiration | Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified. | Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified. | "Same." Implied acceptance: functional and performance equivalence. |
Pulse oxygen saturation (SpO2) | Method: red and infrared light method. Masimo SpO2: Range:1 | Method: same. Masimo SpO2: same range & accuracy. Nellcor SpO2: same range & accuracy. Comen SpO2: Range: 0%-100%. Accuracy: 70-100% ±2%(adult/pediatric, non-motion), ±3%(neonate, non-motion). | "Same." (for Masimo and Nellcor modules). For their own Comen SpO2 module, it shows equivalent accuracy. Implied acceptance: performance equivalence with predicate's different OEM modules and new Comen module meets similar accuracy. |
Pulse rate (PR) | From Mindray SpO2 (Range 20-254 bpm, Acc ±3 bpm), Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%). | From Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), Comen SpO2 (Range 20-254 bpm, Acc ±2 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%), NIBP sensor (Range 40-240 bpm, Acc ±3 bpm or ±3%). | "The PR from Comen SpO2 is more accuracy than the PR form Mindray SpO2. The PR form Masimo SpO2, Nellcor SpO2 and IBP sensor of C50 and C80 are the same with the predicate device. In addition, C50 and C80 have one more PR data source (NIBP) than predicate device. C50 and C80 have complied with ISO 80601-2-61." Acceptance criteria met or exceeded; added NIBP source (improvement). Adherence to ISO 80601-2-61. |
Non-invasive blood pressure (NIBP) | Method: Oscillometry. Range: Adult (systolic 25-290mmHg, diastolic 10-250mmHg), pediatric (systolic 25-240mmHg, diastolic 10-200mmHg), neonate (systolic 25-140mmHg, diastolic 10-115mmHg). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg. | Method: same. Range: Adult (systolic 40-270mmHg, diastolic 10-215mmHg), pediatric (systolic 40-200mmHg, diastolic 10-150mmHg), neonate (systolic 40-135mmHg, diastolic 10-100mmHg). Accuracy: 0-300mmHg: ±3mmHg. | "The NIBP measure range for adult/pediatric/neonate in C50 and C80 is smaller than that in predicate device. The C50 and C80 is more accuracy than predicate device. C50 and C80 have conformed to IEC 80601-2-30." Acceptance criteria met or exceeded in accuracy, and justification for smaller range implicitly accepted by meeting IEC 80601-2-30. |
Temperature (Temp) | Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.1°C. | Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.2°C. | "The predicate device is more accuracy than C50 and C80. C50 and C80 have complied with ISO 80601-2-56." Implied acceptance: slightly lower accuracy is acceptable as long as it complies with ISO 80601-2-56. |
Carbon dioxide (CO2) | Method: Infrared absorption. Masimo CO2: Sidestream 0-99mmHg, AwRR 0-120rpm. Microstream 0-99mmHg, AwRR 0-150rpm. Accuracy: Sidestream 0-40mmHg ±2mmHg, 41-76mmHg ±5%, 77-99mmHg ±10%, AwRR ±2rpm. Microstream 0-38mmHg ±2mmHg, 39-99mmHg ±5% +0.08%, AwRR 0-70rpm ±1rpm, 71-120rpm ±2rpm, 121-150rpm ±3rpm. | Method: same. Masimo CO2: Sidestream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Mainstream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Respironics CO2: Sidestream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Mainstream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Accuracy (Masimo): ±(2.25mmHg +reading x 4%), AwRR ±1rpm. Accuracy (Respironics): 0-40mmHg ±2mmHg, etc., AwRR ±1rpm. | "The C50 and C80 are more accuracy and measured wider than predicate device. For C50 and C80, Masimo CO2 module (K123043 and K103604) and Respironics CO2 module (K040183 and K053174) have been cleared. Both modules have complied with ISO 80601-2-55." Acceptance criteria met or exceeded; wider range and higher accuracy; existing modules previously cleared and comply with ISO 80601-2-55. |
Invasive blood pressure (IBP) | Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor). | Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor). | "Same." Implied acceptance: functional and performance equivalence. |
Cardiac output (C.O.) | Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater. | Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater. | "Same." Implied acceptance: functional and performance equivalence. |
Anesthetic gas (AG) | Method: Infrared absorption. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 ±0.3% ABS, N2O ±(8%REL+2%ABS), Other AG 8%REL). | Method: same. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 0-15%: ±(0.2kPa+readingx2%), N2O ±(2 kPa+readingx2%), Hal/Enf/Iso 0-8%: ±(0.15%+readingx5%), Sev 0-10%: ±(0.15%+readingx5%), Des 0-22%: ±(0.15%+readingx5%), O2 ±(1%+readingx2%), Masimo AG AwRR ±1rpm). | "For CO2/Enf/Hal/Iso/Sev/Des, The measurement range of C50 and C80 is smaller than the predicate device. For awRR, the measurement range of C80 is better than the predicate device. The C50 and C80 are more accuracy than predicate device. The measurement range of C50 and C80 is enough for most environments; it won't affect the safety and effectiveness. The C50 and C80 supports two AG modules: MASIMO ISA AX+ Sidestream module and Masimo IRMA AX+ Mainstream module, both of them have been cleared in K103604." Acceptance criteria: different ranges are acceptable with justification of sufficient range for most environments and high accuracy. Utilizes previously cleared and compliant modules. |
BIS (C80 only) | Range: BIS, BIS L, BIS R: 0-100; SQI, SQI L, SQI R: 0-100%; EMG, EMG L, EMG R: 0-100 Db; SR, SR L, SR R: 0-100%; SEF, SEF L, SEF R: 0.5-30.0 Hz; TP, TP L, TP R: 40-100 Db; BC, BC L, BC R: 0-30; sBIS L, sBIS R: 0-10.0; Semg L, Semg R: 0-10.0; ASYM: 0-100%. Accuracy: 1% for BIS, SQI, EMG, ESR (implied). | Range and Accuracy: BIS: same; accuracy: 1%. SQI: same; accuracy: 1%. EMG: same; accuracy: 1%. ESR: 0-100%; accuracy: 1%. | "The underline parameters are what C50 and C80 doesn't have. The BIS module and sensor have been cleared by FDA and its 510k numbers are K040183 and K002734." Acceptance criteria for essential parameters are met by using previously cleared FDA modules. |
ICG (C80 only) | Method: Indirect impedance cardiograph measurement. Range: SV: 5-250 ml; HR: 44-2m; C.O. 1.4-15 L/min. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified. | Method: Indirect impedance cardiograph measurement. Range: HR: 40-250bpm; SV: 0-250mL; C.O.:0-30L/min; TFC: 5-150 /KΩ; SVR:0-3500 dyn・s ・ cm-5. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified. | "The underline parameter is what C50 and C80 doesn't have. The ICG electrode cable and sensors are all cleared by FDA, 510(k) number of which is K110645." Acceptance criteria for essential parameters are met by using previously cleared FDA components. |
Study that Proves the Device Meets Acceptance Criteria
The essential studies to prove the device meets acceptance criteria are divided into Non-Clinical Performance Data and Clinical Studies.
Non-Clinical Performance Data:
- Software Verification and Validation Testing: Conducted in accordance with FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Risk analysis was performed, and the software was deemed a "major" level of concern.
- Electrical safety and Electromagnetic Compatibility (EMC): Testing conducted according to:
- ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 Edition 4: 2014-02
- Bench Testing: Functional and system-level tests were performed, showing the devices meet specifications and perform equivalently to the predicate.
- Biocompatibility Testing: Performed for patient-contacting components (ECG cable, SpO2 probes, temperature probes) that were not previously cleared. Cytotoxicity, sensitization, and irritation testing were conducted based on ISO 10993-1 and FDA's 2016 biocompatibility guidance.
- Conformance with Applicable Standards: The device demonstrated compliance with numerous recognized consensus standards, in addition to those listed above for specific modules (e.g., ISO 80601-2-55 for Respiratory Gas Monitors, ISO 80601-2-56 for Thermometers, ISO 80601-2-61 for Pulse Oximeter Equipment, ANSI AAMI EC57:2012 for Cardiac Rhythm and ST-Segment measurement algorithms).
Clinical Studies:
1. NIBP Clinical Study - for adults and children
- Device Parameter: Non-invasive Blood Pressure (NIBP)
- Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
- Sample Size: 25 patients (19 adults, 6 children).
- Demographics: 10 men, 15 women. 6 patients aged 3-12, 2 aged 12-40, 10 aged 40-60, 7 above 60.
- Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The Second Affiliated Hospital of Guangzhou Medical University and The First Affiliated Hospital of Guangzhou Traditional Chinese Medical University" from 2012 to 2013 (China).
- Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
- Adjudication Method: Not specified, but standard practice for NIBP clinical validation involves multiple observers for reference measurements.
- Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."
2. NIBP Clinical Study - for neonate and infants
- Device Parameter: Non-invasive Blood Pressure (NIBP)
- Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
- Sample Size: 20 subjects.
- Demographics: 10 males, 10 females. 13 subjects
Ask a specific question about this device
(30 days)
DigitalDiagnost C50
The DigitalDiagnost C50 system is intended for use in generating radiographic images of human anatomy by qualified/ trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.
The Philips DigitalDiagnost C50 Digital Radiography System (Philips DigitalDiagnost C50) is a flexible digital radiography (DR) system that is designed to provide fast and smooth radiography examinations of sitting, standing or lying patients.
The Philips DigitalDiagnost C50 consists of the following components: ceiling suspension with X-ray assembly, wall stand with detector carrier, patient table with detector carrier and floating table top, high voltage generator, and an acquisition and reviewing workstation for post-processing, storage and viewing of images. Images may be transferred via a DICOM network for printing, storage and detailed review.
Based on the provided text, the device in question, the Philips DigitalDiagnost C50, is a stationary x-ray system cleared through a 510(k) premarket notification. The document does not describe acceptance criteria or a study that proves the device meets acceptance criteria in the context of an AI/algorithm-driven diagnostic aid.
Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (Philips DuraDiagnost K141381) and a reference device for the wireless detector (Varian Nexus DR™ Digital X-ray Imaging System K161459).
Therefore, I cannot provide the requested information about acceptance criteria and a study proving the device meets acceptance criteria related to AI/algorithm performance, a multi-reader multi-case study, standalone algorithm performance, or the ground truth establishment for a training set. The device is a traditional x-ray system, not an AI-powered diagnostic tool as suggested by some of the detailed questions.
However, I can extract information regarding overall device compliance and performance from the text as it relates to demonstrating substantial equivalence for an X-ray system:
Acceptance Criteria and Device Performance for the Philips DigitalDiagnost C50
The provided text on the Philips DigitalDiagnost C50's 510(k) submission does not include acceptance criteria and a study in the context of an AI-driven diagnostic device. Instead, it focuses on demonstrating substantial equivalence to predicate and reference devices for a stationary X-ray system. The "acceptance criteria" mentioned refer to the device meeting the requirements set forth by international and FDA-recognized consensus standards and device-specific guidance documents, as well as the successful completion of non-clinical verification and validation tests.
1. Table of Acceptance Criteria and Reported Device Performance (as related to Substantial Equivalence and Safety/Effectiveness):
The document details a comparison of technological characteristics between the proposed device (DigitalDiagnost C50) and its predicate (Philips DuraDiagnost) and reference device (Varian Nexus DR™ Digital X-ray Imaging System for the wireless detector). The "acceptance criteria" here are implicitly that the proposed device's characteristics are either identical or that any minor differences do not affect the safety or effectiveness, thus demonstrating substantial equivalence (SE).
Feature | Acceptance Criteria (based on Predicate/Reference Device Equivalence) | Reported Device Performance (Philips DigitalDiagnost C50) |
---|---|---|
X-ray Tube | Similar x-ray tubes, with identical specifications for nominal X-ray tube voltage, nominal focal spot values, maximum tube current, maximum anode heat content, total filtration, maximum and anode heat dissipation. Minor difference in tube housing should not affect performance, safety, and effectiveness. | Uses RO 1750 ROT 380 & SRO 33100 ROT 380, which are similar and have the same specifications as the predicate's tubes (RO 1750 ROT 360 & SRO 33100 ROT 360). The minor tube housing difference has no impact on safety and effectiveness, demonstrating SE. |
Max Tube Voltage | Same as predicate (150kV). | Same (150kV). No difference; demonstrating SE. |
Focal Spot Size | Same as predicate (0.6mm/1.2mm). | Same (0.6mm/1.2mm). No difference; demonstrating SE. |
Tube Max Power | Same as predicate (50KW/100KW, 250W equivalent anode input power). | Same (50KW/100KW, 250W equivalent anode input power). No difference; demonstrating SE. |
Anode Type | Same as predicate (Rotation). | Same (Rotation). No difference; demonstrating SE. |
Generator | Equivalent to predicate (M-CABINET CXA Pro 50kW, M-CABINET CXA Pro 65kW, M-CABINET Pro CXA 80kW, particularly 50kW/65kW). Minor difference in max power (80kW not included) should not affect safety and effectiveness. | Uses M-CABINET CXA Pro 50kW, M-CABINET CXA Pro 65kW. Max Power 50KW/65KW. These are equivalent to the predicate's available generators, with no impact on safety and effectiveness, demonstrating SE. |
kV range | Same as predicate (40-150). | Same (40-150). No difference; demonstrating SE. |
Milliampere second (mAs) product | Same as predicate (0.4 mAs-600 mAs with AEC control). | Same (0.4 mAs-600 mAs with AEC control). No difference; demonstrating SE. |
Collimator Operation Mode | Same as predicate (Manual collimation). | Same (Manual collimation). No difference; demonstrating SE. |
Shape of Beam | Same as predicate (Rectangular). | Same (Rectangular). No difference; demonstrating SE. |
Detector Type | Fixed: Identical to predicate (GdOS). Wireless: Identical to reference device (PaxScan 4336Wv4 in Varian Nexus DR™ Digital X-ray Imaging System). No impact on safety and effectiveness. | Fixed: GdOS (identical to predicate). Wireless: PaxScan 4336Wv4 (GdOS), identical to the reference device. No impact on safety and effectiveness, demonstrating SE. |
X-ray Scintillator Material | Fixed: Identical to predicate (GdOS). Wireless: Identical to reference device (GdOS) which replaces the predicate's Cesium Iodide. No impact on safety and effectiveness. | Fixed: GdOS (identical to predicate). Wireless: GdOS (identical to the reference device). No impact on safety and effectiveness, demonstrating SE. |
Image Area | Fixed: Identical to predicate (42.5cm x 42.5cm). Wireless: Functionally equivalent/identical to reference device (42.7 cm x 34.4 cm vs. reference 42.4cm x 34.8cm). Minor differences should not impact safety and effectiveness. | Fixed: 42.5cm x 42.5cm (identical to predicate). Wireless: 42.7 cm x 34.4 cm (functionally equivalent to reference device's 42.4cm x 34.8cm). No impact on safety and effectiveness, demonstrating SE. |
Image Matrix | Fixed: Identical to predicate (2874 x 2869). Wireless: Functionally equivalent/identical to reference device (3072 x 2476 vs. reference 2866 x 2350). Minor differences should not impact safety and effectiveness. | Fixed: 2874 x 2869 (identical to predicate). Wireless: 3072 x 2476 (functionally equivalent to reference device's 2866 x 2350). No impact on safety and effectiveness, demonstrating SE. |
Analog / Digital (A/D) conversion | Same as predicate (16 bits). | Same (16 bits). No difference; demonstrating SE. |
Source to Image Distance (SID) | Minor differences are acceptable as long as they do not alter application usage and are supported by bench testing results. | SID depends on configurations (ceiling suspension system). Slight difference compared to predicate (Table: 40-115cm; Wallstand: 110-245cm) but bench testing (Section 17) proved no alteration to application usage; thus, no impact on safety and effectiveness, demonstrating SE. |
External Connectivity | Same as predicate (DICOM 3.0 compatible). | Same (DICOM 3.0 compatible). No difference; demonstrating SE. |
Software Platform | Same as predicate (Eleva workspot). | Same (Eleva workspot). No difference; demonstrating SE. |
Compliance with Standards | Complies with specified international and FDA-recognized consensus standards and device-specific guidance documents (e.g., AAMI / ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, IEC 62304, ISO 14971, IEC 62366, CFR 1020.30, CFR 1020.31, FDA guidance documents). | Compliance confirmed through non-clinical verification and validation tests. The device "Meets the acceptance criteria and is adequate for its intended use" based on these tests. |
2. Sample Size and Data Provenance for Test Set (for an AI/Algorithm):
- Not applicable. The submission is for a traditional X-ray system, not an AI/algorithm-driven diagnostic device. The performance data relies on comparing the technical specifications and non-clinical bench testing against established standards and predicate devices.
3. Number of Experts and Qualifications to Establish Ground Truth (for an AI/Algorithm):
- Not applicable. See point 2.
4. Adjudication Method for Test Set (for an AI/Algorithm):
- Not applicable. See point 2.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- Not applicable. The document explicitly states: "The DigitalDiagnost C50 did not require a clinical study since substantial equivalence to the primary currently marketed and predicate device was demonstrated with the following attributes: Design features; Indication for use; Fundamental scientific technology; Non-clinical performance testing; and Safety and effectiveness." There is no mention of an MRMC study or an effect size for human readers with or without AI assistance.
6. Standalone Algorithm Performance:
- Not applicable. The device is an X-ray system, not a standalone algorithm.
7. Type of Ground Truth Used (for an AI/Algorithm):
- Not applicable. For this X-ray system, "ground truth" for performance is established through adherence to recognized standards, technical specifications, and successful non-clinical performance (bench) testing, demonstrating that the images generated are of appropriate quality and the system functions safely and effectively as intended.
8. Sample Size for the Training Set (for an AI/Algorithm):
- Not applicable. See point 2.
9. How the Ground Truth for the Training Set Was Established (for an AI/Algorithm):
- Not applicable. See point 2.
In summary, the Philips DigitalDiagnost C50 is a conventional X-ray system whose acceptance criteria are fundamentally tied to demonstrating substantial equivalence to legally marketed predicate devices through comprehensive non-clinical performance (verification and validation) tests and adherence to established international and FDA-recognized consensus standards. The document does not describe the evaluation of an AI or algorithm.
Ask a specific question about this device
Page 1 of 1